Application of a deformable registration technique to investigate breath-hold reproducibility

verfasst von: Nobuyoshi Fukumitsu, Yasutaka Hayashi

Erschienen in: Japanese Journal of Radiology | Ausgabe 12/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To develop an objective and less invasive method for evaluating breath-hold status, and to investigate breath-hold reproducibility during voluntary breath-holding irradiation.

Materials and methods

We enrolled 20 patients who were treated using a voluntary breath-holding technique. Four or five sessions of cone beam computed tomography (CT) were performed during the radiotherapy session. An image of the lung with respiratory tract was extracted from the CT findings. The rigid registration of subsequent CT findings was completed using the first fraction CT findings. Next, subsequent CT images already subjected to rigid registration were deformed using the first CT images. We compared the differences in the subsequent CT images before and after the deformable registration.

Results

We were able to complete the method to evaluate breath-hold status without having to consider set-up uncertainty, manual processes, visual estimates, or excessive patient cooperation. Tumor dislocation was almost within 3 mm in all directions and deformation was much smaller than 30 % period of free breathing.

Conclusion

This method we have developed may have great potential for the objective verification of breath-hold reproducibility in a less invasive manner. This method assumed that the voluntary breath-hold status was reproducible enough at a clinically satisfactory level.
Literatur
1.
Zurück zum Zitat Onishi H, Kawakami H, Marino K, Komiyama T, Kuriyama K, Araya M, et al. A simple respiratory indicator for irradiation during voluntary breath holding: a one-touch device without electronic materials. Radiology. 2010;255:917–23.PubMedCrossRef Onishi H, Kawakami H, Marino K, Komiyama T, Kuriyama K, Araya M, et al. A simple respiratory indicator for irradiation during voluntary breath holding: a one-touch device without electronic materials. Radiology. 2010;255:917–23.PubMedCrossRef
2.
Zurück zum Zitat Tarohda TI, Ishiguro M, Hasegawa K, Kohda Y, Onishi H, Aoki T, et al. The management of tumor motions in the stereotactic irradiation to lung cancer under the use of Abches to control active breathing. Med Phys. 2011;38:4141–6.PubMedCrossRef Tarohda TI, Ishiguro M, Hasegawa K, Kohda Y, Onishi H, Aoki T, et al. The management of tumor motions in the stereotactic irradiation to lung cancer under the use of Abches to control active breathing. Med Phys. 2011;38:4141–6.PubMedCrossRef
3.
Zurück zum Zitat Weiss E, Robertson SP, Mukhopadhyay N, Hugo GD. Tumor, lymph node, and lymph node-to-tumor displacements over a radiotherapy series: analysis of interfraction and intrafraction variations using active breathing control (ABC) in lung cancer. Int J Radiat Oncol Biol Phys. 2012;82:e639–45.PubMedCentralPubMedCrossRef Weiss E, Robertson SP, Mukhopadhyay N, Hugo GD. Tumor, lymph node, and lymph node-to-tumor displacements over a radiotherapy series: analysis of interfraction and intrafraction variations using active breathing control (ABC) in lung cancer. Int J Radiat Oncol Biol Phys. 2012;82:e639–45.PubMedCentralPubMedCrossRef
4.
Zurück zum Zitat Panakis N, McNair HA, Christian JA, Mendes R, Symonds-Tayler JR, Knowles C, et al. Defining the margins in the radical radiotherapy of non-small cell lung cancer (NSCLC) with active breathing control (ABC) and the effect on physical lung parameters. Radiother Oncol. 2008;87:65–73.PubMedCrossRef Panakis N, McNair HA, Christian JA, Mendes R, Symonds-Tayler JR, Knowles C, et al. Defining the margins in the radical radiotherapy of non-small cell lung cancer (NSCLC) with active breathing control (ABC) and the effect on physical lung parameters. Radiother Oncol. 2008;87:65–73.PubMedCrossRef
5.
Zurück zum Zitat Cheung PC, Sixel KE, Tirona R, Ung YC. Reproducibility of lung tumor position and reduction of lung mass within the planning target volume using active breathing control (ABC). Int J Radiat Oncol Biol Phys. 2003;57:1437–42.PubMedCrossRef Cheung PC, Sixel KE, Tirona R, Ung YC. Reproducibility of lung tumor position and reduction of lung mass within the planning target volume using active breathing control (ABC). Int J Radiat Oncol Biol Phys. 2003;57:1437–42.PubMedCrossRef
6.
Zurück zum Zitat Shen Y, Zhang H, Wang J, Zhong R, Jiang X, Xu Q, et al. Hypofractionated radiotherapy for lung tumors with online cone beam CT guidance and active breathing control. Radiat Oncol. 2010;5:19.PubMedCentralPubMedCrossRef Shen Y, Zhang H, Wang J, Zhong R, Jiang X, Xu Q, et al. Hypofractionated radiotherapy for lung tumors with online cone beam CT guidance and active breathing control. Radiat Oncol. 2010;5:19.PubMedCentralPubMedCrossRef
7.
Zurück zum Zitat Mah D, Hanley J, Rosenzweig KE, Yorke E, Braban L, Ling CC, et al. Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. Int J Radiat Oncol Biol Phys. 2000;48:1175–85.PubMedCrossRef Mah D, Hanley J, Rosenzweig KE, Yorke E, Braban L, Ling CC, et al. Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. Int J Radiat Oncol Biol Phys. 2000;48:1175–85.PubMedCrossRef
8.
Zurück zum Zitat Sarrut D, Boldea V, Ayadi M, Badel JN, Ginestet C, Clippe S, et al. Nonrigid registration method to assess reproducibility of breath-holding with ABC in lung cancer. Int J Radiat Oncol Biol Phys. 2005;61:594–607.PubMedCrossRef Sarrut D, Boldea V, Ayadi M, Badel JN, Ginestet C, Clippe S, et al. Nonrigid registration method to assess reproducibility of breath-holding with ABC in lung cancer. Int J Radiat Oncol Biol Phys. 2005;61:594–607.PubMedCrossRef
9.
Zurück zum Zitat Fukumitsu N, Hashimoto T, Okumura T, Mizumoto M, Tohno E, Fukuda K, et al. Investigation of the geometric accuracy of proton beam irradiation in the liver. Int J Radiat Oncol Biol Phys. 2012;82:826–33.PubMedCrossRef Fukumitsu N, Hashimoto T, Okumura T, Mizumoto M, Tohno E, Fukuda K, et al. Investigation of the geometric accuracy of proton beam irradiation in the liver. Int J Radiat Oncol Biol Phys. 2012;82:826–33.PubMedCrossRef
10.
Zurück zum Zitat Gierga DP, Turcotte JC, Tong LW, Chen YL, DeLaney TF. Analysis of setup uncertainties for extremity sarcoma patients using surface imaging. Pract Radiat Oncol. 2014;4:261–6.PubMedCrossRef Gierga DP, Turcotte JC, Tong LW, Chen YL, DeLaney TF. Analysis of setup uncertainties for extremity sarcoma patients using surface imaging. Pract Radiat Oncol. 2014;4:261–6.PubMedCrossRef
11.
Zurück zum Zitat Uh J, Merchant TE, Li Y, Li X, Hua C. MRI-based treatment planning with pseudo CT generated through atlas registration. Med Phys. 2014;41:051711.PubMedCrossRef Uh J, Merchant TE, Li Y, Li X, Hua C. MRI-based treatment planning with pseudo CT generated through atlas registration. Med Phys. 2014;41:051711.PubMedCrossRef
12.
Zurück zum Zitat Abdelnour AF, Nehmeh SA, Pan T, Humm JL, Vernon P, Schoder H, et al. Phase and amplitude binning for 4D-CT imaging. Phys Med Biol. 2007;52:3515–29.PubMedCrossRef Abdelnour AF, Nehmeh SA, Pan T, Humm JL, Vernon P, Schoder H, et al. Phase and amplitude binning for 4D-CT imaging. Phys Med Biol. 2007;52:3515–29.PubMedCrossRef
13.
Zurück zum Zitat Kashani R, Hub M, Balter JM, Kessler ML, Dong L, Zhang L, et al. Objective assessment of deformable image registration in radiotherapy: a multi-institution study. Med Phys. 2008;35:5944–53.PubMedCentralPubMedCrossRef Kashani R, Hub M, Balter JM, Kessler ML, Dong L, Zhang L, et al. Objective assessment of deformable image registration in radiotherapy: a multi-institution study. Med Phys. 2008;35:5944–53.PubMedCentralPubMedCrossRef
14.
Zurück zum Zitat Nie K, Chuang C, Kirby N, Braunstein S, Pouliot J. Site-specific deformable imaging registration algorithm selection using patient-based simulated deformations. Med Phys. 2013;40:041911.PubMedCrossRef Nie K, Chuang C, Kirby N, Braunstein S, Pouliot J. Site-specific deformable imaging registration algorithm selection using patient-based simulated deformations. Med Phys. 2013;40:041911.PubMedCrossRef
15.
Zurück zum Zitat Kirby N, Chuang C, Ueda U, Pouliot J. The need for application-based adaptation of deformable image registration. Med Phys. 2013;40:011702.PubMedCrossRef Kirby N, Chuang C, Ueda U, Pouliot J. The need for application-based adaptation of deformable image registration. Med Phys. 2013;40:011702.PubMedCrossRef
16.
Zurück zum Zitat Pirozzi S, Piper J, Nelson A, Duchateau M, Verellen D, De Ridder M. A novel framework for user-intervened correction of deformable registration. Int J Radiat Oncol Biol Phys. 2013;87:S144.CrossRef Pirozzi S, Piper J, Nelson A, Duchateau M, Verellen D, De Ridder M. A novel framework for user-intervened correction of deformable registration. Int J Radiat Oncol Biol Phys. 2013;87:S144.CrossRef
17.
18.
Zurück zum Zitat Oborski MJ, Laymon CM, Qian Y, Lieberman FS, Nelson AD, Mountz JM. Challenges and approaches to quantitative therapy response assessment in glioblastoma multiforme using the novel apoptosis positron emission tomography tracer F-18 ML-10. Transl Oncol. 2014;7:111–9.PubMedCentralPubMedCrossRef Oborski MJ, Laymon CM, Qian Y, Lieberman FS, Nelson AD, Mountz JM. Challenges and approaches to quantitative therapy response assessment in glioblastoma multiforme using the novel apoptosis positron emission tomography tracer F-18 ML-10. Transl Oncol. 2014;7:111–9.PubMedCentralPubMedCrossRef
19.
Zurück zum Zitat Yakoumakis N, Winey B, Killoran J, Mayo C, Niedermayr T, Panayiotakis G, et al. Using four-dimensional computed tomography images to optimize the internal target volume when using volume-modulated arc therapy to treat moving targets. J Appl Clin Med Phys. 2012;13:3850.PubMed Yakoumakis N, Winey B, Killoran J, Mayo C, Niedermayr T, Panayiotakis G, et al. Using four-dimensional computed tomography images to optimize the internal target volume when using volume-modulated arc therapy to treat moving targets. J Appl Clin Med Phys. 2012;13:3850.PubMed
Metadaten
Titel
Application of a deformable registration technique to investigate breath-hold reproducibility
verfasst von
Nobuyoshi Fukumitsu
Yasutaka Hayashi
Publikationsdatum
01.12.2014
Verlag
Springer Japan
Erschienen in
Japanese Journal of Radiology / Ausgabe 12/2014
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-014-0369-y

Weitere Artikel der Ausgabe 12/2014

Japanese Journal of Radiology 12/2014 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.