Open Access 26.08.2022 | Original Paper

Assessing the subjective quality of smartphone anterior segment photography: a non-inferiority study

verfasst von: Raghav Goel, Carmelo Macri, Bobak Bahrami, Robert Casson, Weng Onn Chan

Erschienen in: International Ophthalmology | Ausgabe 2/2023

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Purpose

To assess the subjective quality of anterior segment photos taken from a smartphone camera adapted to the slit lamp compared to a commercial inbuilt slit-lamp camera.

Methods

Non-inferiority study. Five paired images of the anterior segment of normal eyes were taken using an iPhone 11 (Apple, Inc., Calif., USA) camera attached to a universal slit-lamp adaptor and a commercial inbuilt slit-lamp camera (Haag-Streit Diagnostics, Bern, Switzerland). Images were collated into a survey in which ophthalmology students, residents, registrars, and consultants participated to select the image taken from the inbuilt slit-lamp camera. If the image quality was subjectively indistinguishable, we expected a 50:50 split for each photograph that was presented. We selected a 10% non-inferiority margin, with the hypothesis that no less than 40% of images believed to be from the conventional camera were in fact from the smartphone camera.

Results

There were 27 respondents in the survey: ophthalmology consultants (n = 7), registrars (n = 10), residents (n = 7), intern (n = 1) and students (n = 2). The mean correct identification across the respondents was 11.3 out of 25 (45.2%) images. Overall, the smartphone camera was non-inferior to the inbuilt slit-lamp camera (p < 0.001). The non-inferiority of the smartphone camera was significant for consultants (47.4%, p < 0.01), registrars (47.6%, p < 0.001) and residents (37.7%, p < 0.0001).

Conclusions

Anterior segment images obtained with a smartphone camera were non-inferior to the commercial inbuilt slit-lamp camera. Smartphone cameras may be a non-inferior tool for communication of anterior segment images having implications for the ease of access to quality telehealth consultations.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s10792-022-02437-9.
Presented as a poster at the Royal Australian and New Zealand College of Ophthalmologists 52nd Annual Scientific Congress, February 26- March 1 2022 [1].

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

In Australia, approximately 1.4% of all emergency department (ED) presentations are related to eye disorders [2]. Anterior segment pathology accounts for more than 90% of ED presentations related to eye disorders [3]. Specialist consultation is frequent given medical officers report decreasing confidence in slit lamp usage and management of ocular emergencies in the ED [4].
Ophthalmology consults may be performed virtually with the use of high-quality photographs, increasingly relevant with the increased uptake of telemedicine in response to the COVID-19 pandemic [5, 6]. Inbuilt slit-lamp cameras are an effective means for anterior segment photography. Despite a large proportion of EDs having access to slit lamps, medical officers report a lack of training to use them [7]. In addition, rural EDs are less likely to have access to slit lamps [8].
Conversely, smartphones are widely available and offer a wide range of functionalities. The increasing camera quality of smartphones, low cost and universality have sparked interest in their utility in ophthalmology [9]. Slit lamp photography is possible with the use of inexpensive smartphone adaptors that offer comparable image quality to inbuilt slit-lamp cameras. Observations made by previous investigators have demonstrated smartphone cameras offering good image quality, however, reproducibility and agreement on this are lacking [1013].
Given the ease of access to a smartphone camera, there may be a role in using them to image anterior segment pathology in primary and tertiary care centres. This can be used for consultation purposes without the need for specialised slit-lamp cameras. This pilot study was designed to assess whether ophthalmology staff can discriminate between images taken of the anterior segment in healthy eyes from an inbuilt slit-lamp microscope camera compared to that of a smartphone camera attached to a universal slit-lamp adaptor.

Methods

All participants were recruited from the Royal Adelaide Hospital (RAH) Ophthalmology Department. Ethics approval was acquired from the local Human Research Ethics Committee at the Central Adelaide Local Health Network Research office (reference 13716). Consent was obtained from five staff volunteers in the department to acquire images of their anterior segment. An inexpensive adjustable slit-lamp phone adaptor was utilised as depicted in Fig. 1 [14]. The iPhone 11 (Apple, Inc., Cupertino, Calif., USA) was mounted onto the adaptor and placed over the right eyepiece of the slit lamp (Fig. 2). At the time, iOS™ (14.3) software was installed on the iPhone 11.
A survey with each pair of images for all subjects was created using Google Forms (Alphabet Inc., California, USA). An invitation link to the survey was provided to medical students, interns, ophthalmology residents (i.e., unaccredited trainees), registrars (i.e., accredited trainees), consultant ophthalmologists. Participants were given the option of answering anonymously and recording their level of expertise. The survey consisted of 25 comparison questions and was blinded to individuals who were not involved in the imaging process. The questions asked the participants to select image captured by the commercial slit-lamp camera (Figs. 3 and 4). The correct responses were recorded.
Survey design sample size calculations allowing for correlated data indicated that approximately 650 data points would be required to detect a 10% margin in non-inferiority where the null hypothesis assumed a 50:50 split with 80% power and alpha at 0.05. The p-values were calculated with a z-test and the correct variance estimate on the proportion was calculated by modelling the study as a clustered design using the survey package in R (version 4.0) with the individual as the unit of a cluster [15, 16]. A Bonferroni correction was applied for sub-group analyses.
The slit-lamp images were taken using the integrated Haag-Streit IM900 Imaging module with Eyesuite™ software (Haag-Streit Diagnostics, Bern, Switzerland) [17]. The same iPhone and slit lamp were used throughout the study. Two medical officers (MO) took sets of three photographs using both cameras, for five separate imaging techniques. These techniques were diffuse illumination, narrow corneal slit, sclerotic scatter, lens optical section, and iris retro-illumination (see Appendix, Table 1s@@). The anterior segment photos were captured simultaneously using the two cameras where one MO captured photos using the smartphone on the slit lamp, whilst the other using the Eyesuite™ software on the computer.
The settings enabled on the iPhone camera included ‘high dynamic range (HDR)’ and ‘autofocus’. The slit lamp settings were adjusted based on the imaging technique but were kept constant for all the subjects (see Appendix, Table 1s@@). The best quality images were chosen from a set of three, to simulate out-of-study conditions in which the best photos of multiple would be selected. The pairs of photos were cropped using Adobe Lightroom™ 4.1.1 (Adobe, San Jose, California, USA) to display similar structures in each photo. The cropped images were then light adjusted using Lightroom’s Auto function, in conjunction with adjusting the Exposure slider, to display visibly equal exposure. This was done to ensure survey participants would only use photo quality to discriminate the image pairs and prevent different light exposure or field angles to influence the result. The time taken for this was approximately a minute for each set of photos.

Results

Twenty-seven respondents completed the survey comprising twenty-five image questions and two quality control questions. There was a total of 675 data points extracted from the survey. Overall, the respondents achieved a mean correct score of 11.3 out of 25 (45.2%, SD 14.0) in the survey for correctly identifying the inbuilt slit lamp photos. Thus, 54.8% of respondents believed the smartphone camera was the commercial slit-lamp camera image, signifying that the smartphone camera is non-inferior to the inbuilt slit-lamp camera (z-test, p-value < 0.001). The mean score for each subset of images is shown in Fig. 5.
The survey also noted the category of the respondent as a consultant, registrar, resident, intern or student. The survey was mostly completed by the consultants 25.9% (n = 7), registrars 37% (n = 10) and residents 25.9% (n = 7). The other respondents were interns 3.7% (n = 1) and students 7.4% (n = 2). The consultant group were accurate 47.4% of the time incorrectly identifying slit-lamp camera image (p-value < 0.01). Similarly, registrars had a correct guess rate of 47.6% for slit-lamp images (p-value < 0.001) of the time and the residents strongly favoured the smartphone camera with only 37.7% correct guess rate for slit lamp images (p-value < 0.0001). The correct guess rate for slit-lamp photos is depicted in Table 1.
Table 1
Mean scores of the respondents for correctly identifying the inbuilt slit-lamp image and non-inferiority significance level
Seniority
Respondents
Mean (%)
Mean score (SD)
95% confidence interval
P-Value
Consultant
7
47.4%
11.9 (3.5)
11.8–12.0
P = 0.0085
Registrar
10
47.6%
11.9 (3.1)
11.8–12.0
P < 0.001
Resident
7
37.7%
9.4 (3.7)
9.3–9.5
P < 0.0001
Overall
27
45.2%
11.3 (3.4)
11.25–11.35
P < 0.0001
The data for intern (n = 2) and students (n = 1) was omitted due to insufficient sample size
Respondents were asked if the images were of comparable quality and 20 out of 27 respondents (74.1%) believed they were. The respondents who did not agree with this finding achieved a mean score of 9.3 out of 25 (n = 7, 37.2%), compared to those that agreed had a mean score of 12 (n = 20, 48%) for correctly identifying slit lamp photos. Respondents also reported the difficulty experienced in comparing images as per Fig. 6.

Discussion

In this study conducted by medical personnel with varying ophthalmic expertise, the smartphone camera was non-inferior to the commercial inbuilt slit-lamp camera. This suggests that adequate anterior segment imaging can be obtained without specialist equipment, offering more accessible telehealth opportunities for consultation and specialist opinion. This may reduce the time for specialist involvement in patient care and provide cost savings by reducing the need for expensive photographic equipment.
The pandemic era paves the way for smartphone devices to be used as a valuable tool for telehealth consultations in ophthalmology. Uses range from a tool to detect ptosis, cataracts, measurement of toric alignment, evaluation of globe anatomy and trachoma surveillance [1823]. Previous pilot studies have shown that anterior segment photos taken through smartphone cameras such as the iPhone 4 s and 5 s were all of poorer quality when compared to the Zeiss photo slit lamp camera (Carl Zeiss Meditec, Dublin, Calif., USA) [24]. Others have studied anterior segment images post-cataract operation or with the use of different smartphone attachments such as the 90-dioptre lens [2529]. None of these studies have compared if the subjective quality of anterior segment images captured by a smartphone camera and an inbuilt slit-lamp camera.
In contrast to the above studies, this study shows that anterior segment images taken from iPhone 11 are non-inferior to the Haag Streit BQ-900 imaging module as perceived by the ophthalmology staff. Interestingly, the resident group tended to strongly favour the smartphone images, as they correctly identified the slit lamp photos only 37.7% of the time. This preference of the resident group may be due to their education level, age, and smartphone usage which can cause deviation in the results.
The reasons for non-inferiority are multifactorial. The 12-megapixel camera on the iPhone 11 is comparable to the high sensitivity and a wide dynamic range provided by the Haag Streit Imaging Module (IM) 900. The resolution offered by iPhone 11 camera (4032-by-3024 pixels) is superior to the Haag-Streit IM 900 module (1920-by-1200 pixels) [17, 30, 31]. The other key differences are the software and functionalities of the two cameras. Similar to other smartphones, the iPhone 11 offers features such as auto-focus and optical image stabilisation capabilities that the IM 900 module does not. However, it can be argued that the IM 900 has superior stabilisation abilities as it is integrated into the slit lamp as compared to the smartphone attached to an adaptor which requires adjustments. Furthermore, the software of the iPhone 11 obtains up to 10 updates every year as compared to just an annual update for the IM 900 module which may enhance its camera’s performance [32, 33].
This study shows that the combination of a slit-lamp adaptor and smartphone has its advantages. Firstly, the adaptors are inexpensive, such as the one used in this pilot study (AUD $20) as compared to inbuilt slit-lamp cameras which costs upwards of AUD $20,000 (excluding costs of ongoing maintenance) [14, 34]. Secondly, the slit lamp illumination system provides clear image quality at higher magnification due to the reduced focusing distance to the eye [35]. The smartphone uses the slit lamp’s illumination at magnification to capture sharper images as compared to using the zoom function on the smartphone alone. No additional light source from the smartphone also results in decreased reflection on the ocular surface [36]. Lastly, the adaptors can be attached to most smartphones, are transportable and suitable for most slit lamps.
Applications of images captured from this setup are numerous. Sink, Blatt, Yoo et al. compared diagnosis achieved through remote smartphone photographs to those of in-office exams for common ophthalmology presentations such as external eye diseases or red-eye pathology. The results showed that the remote and in-office diagnosis made by different specialists were in agreement 93% of the time [37]. This pilot study supports the use of anterior segment images acquired with a smartphone camera to diagnose anterior segment pathology as they are similar to the inbuilt slit lamp camera. The inbuilt slit lamp camera is usually the benchmark for taking ocular photos and as the ophthalmology staff are unable to distinguish between the two cameras, they will likely find the smartphone images to be sufficient to make a diagnosis. Furthermore, new smartphone cameras providing stereoscopic images allow for tele-ophthalmology viewers to see in 3D similar to a slit lamp, furthering the potential applicability of these mounts [38, 39]. This can also be integrated with machine learning in the future to assess for anterior chamber depth and grading of cataracts which has potential generalizability to the primary care setting [39, 40].
There are some issues with the two imaging systems that can hinder the quality of photographs. The inbuilt slit-lamp cameras are often subject to availability and are only present in well-established ophthalmology clinics. They require integrated storage space on local computers and images are not as portable. Secondly, in our experience we found a small delay between pressing the capture button to when the image is acquired. This may interfere with the quality of the image due to patient movement. Conversely, the adaptor has inherent issues such as difficulty in attachment to the rubber ring on the eyepiece. If this occurs, an alternative is to place double sided tape on the adapter to allow for better fixation. Moreover, the depth of eye piece fixation ring is often shallow and may not always bear the weight of smartphone. Thus, removal of any protective cases or accessories on the phone may allow the adapter to hold the weight of the smartphone. Also, we found that having one hand on the smartphone and other on the slit lamp for manoeuvring provided stability in capturing photos. The quality of photos may differ between each doctor using the adapter, however, with basic slit lamp examination skills it should still produce good quality photos with a short learning curve.
The limitations of the study are that it does not apply to fundus photographs captured by a smartphone camera as others have previously compared [41]. The sample size of the study was dependent on the limited number of ophthalmology staff in the department. Only ophthalmology staff were used in the survey due to their more experience and knowledge of the slit lamp and ocular anatomical structures. However, this study could be replicated at a larger scale including other disciplines in the future. Lastly, the software used for light-adjusting photos was not completely automated and subject to user bias.

Conclusion

This pilot study demonstrates that photos captured by a smartphone camera are non-inferior to those taken with an inbuilt slit lamp camera as discriminated by medical staff in our Ophthalmology department. Anterior segment images can be captured by clinicians with the use of an inexpensive slit lamp adaptor in the absence of inbuilt photography modules. This increases the accessibility of tele-ophthalmic consultation for expert opinion in settings without proximate ophthalmologists or specialised ophthalmic imaging equipment.

Declarations

Conflicts of interest

No conflicts of interest to declare.
Volunteers were asked to sign a consent process prior to agreeing for eyes to be photographed.

Ethical approval

Gained from the Central Adelaide Local Health Network Human Research and Ethics Committee, Reference number 13716.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
3.
Zurück zum Zitat Channa R, Zafar SN, Canner JK, Haring RS, Schneider EB, Friedman DS (2016) Epidemiology of eye-related emergency department visits. JAMA Ophthalmol 134(3):312–319CrossRefPubMed Channa R, Zafar SN, Canner JK, Haring RS, Schneider EB, Friedman DS (2016) Epidemiology of eye-related emergency department visits. JAMA Ophthalmol 134(3):312–319CrossRefPubMed
4.
Zurück zum Zitat Sim D, Hussain A, Tebbal A, Daly S, Pringle E, Ionides A (2008) National survey of the management of eye emergencies in the accident and emergency departments by senior house officers: 10 years on—has anything changed? Emerg Med J 25(2):76–77CrossRefPubMed Sim D, Hussain A, Tebbal A, Daly S, Pringle E, Ionides A (2008) National survey of the management of eye emergencies in the accident and emergency departments by senior house officers: 10 years on—has anything changed? Emerg Med J 25(2):76–77CrossRefPubMed
5.
Zurück zum Zitat Sommer AC, Blumenthal EZ (2020) Telemedicine in ophthalmology in view of the emerging COVID-19 outbreak. Graefes Arch Clin Exp Ophthalmol 258(11):2341–2352CrossRefPubMedPubMedCentral Sommer AC, Blumenthal EZ (2020) Telemedicine in ophthalmology in view of the emerging COVID-19 outbreak. Graefes Arch Clin Exp Ophthalmol 258(11):2341–2352CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Sim PY, La CJ, Than J, Ho J (2020) National survey of the management of eye emergencies in the accident and emergency department by foundation doctors: has anything changed over the past 15 years? Eye 34(6):1094–1099CrossRefPubMed Sim PY, La CJ, Than J, Ho J (2020) National survey of the management of eye emergencies in the accident and emergency department by foundation doctors: has anything changed over the past 15 years? Eye 34(6):1094–1099CrossRefPubMed
8.
Zurück zum Zitat Tauber J, Ayoub S, Shah P, Wu M, Tsui E, Schuman JS et al (2020) Assessing the demand for teleophthalmology in florida emergency departments. Telemed e-Health 26(12):1500–1506CrossRef Tauber J, Ayoub S, Shah P, Wu M, Tsui E, Schuman JS et al (2020) Assessing the demand for teleophthalmology in florida emergency departments. Telemed e-Health 26(12):1500–1506CrossRef
9.
Zurück zum Zitat Chan JB, Ho HC, Ngah NF, Hussein E (2014) DIY - smartphone slit-lamp adaptor. J Mobile Technol Med 3(1):16–22CrossRef Chan JB, Ho HC, Ngah NF, Hussein E (2014) DIY - smartphone slit-lamp adaptor. J Mobile Technol Med 3(1):16–22CrossRef
10.
Zurück zum Zitat Bhatter P, Cao L, Crochetiere A, Raefsky SM, Cuevas LR, Enendu K et al (2020) Using a macro lens for anterior segment imaging in rural panama. Telemed J E Health 26(11):1414–1418CrossRefPubMedPubMedCentral Bhatter P, Cao L, Crochetiere A, Raefsky SM, Cuevas LR, Enendu K et al (2020) Using a macro lens for anterior segment imaging in rural panama. Telemed J E Health 26(11):1414–1418CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Pujari A, Mukhija R, Singh AB, Chawla R, Sharma N, Kumar A (2018) Smartphone-based high definition anterior segment photography. Indian J Ophthalmol 66(9):1375–1376CrossRefPubMedPubMedCentral Pujari A, Mukhija R, Singh AB, Chawla R, Sharma N, Kumar A (2018) Smartphone-based high definition anterior segment photography. Indian J Ophthalmol 66(9):1375–1376CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Pujari A, Behera AK, Agarwal D, Sihota R, Dada T (2020) A New technique of iPhone 11 pro max smartphone-aided angle video and standstill image documentation. J Glaucoma 29(5):e28–e30CrossRefPubMed Pujari A, Behera AK, Agarwal D, Sihota R, Dada T (2020) A New technique of iPhone 11 pro max smartphone-aided angle video and standstill image documentation. J Glaucoma 29(5):e28–e30CrossRefPubMed
16.
Zurück zum Zitat Lumley T (2020) "survey: analysis of complex survey samples". R package version 4.0. Lumley T (2020) "survey: analysis of complex survey samples". R package version 4.0.
17.
Zurück zum Zitat Haag-Streit (2016) Slit lamp imaging guide, 9th edn. Haag-Streit, Switzerland Haag-Streit (2016) Slit lamp imaging guide, 9th edn. Haag-Streit, Switzerland
18.
Zurück zum Zitat Hogarty DT, Hogarty JP, Hewitt AW (2020) Smartphone use in ophthalmology: what is their place in clinical practice? Surv Ophthalmol 65(2):250–262CrossRefPubMed Hogarty DT, Hogarty JP, Hewitt AW (2020) Smartphone use in ophthalmology: what is their place in clinical practice? Surv Ophthalmol 65(2):250–262CrossRefPubMed
19.
Zurück zum Zitat Kumar N, Bandello F, Sharma A (2019) Smartphone-based Gonio-imaging: a novel addition to glaucoma screening tools. J Glaucoma 28(9):e149–e150CrossRefPubMed Kumar N, Bandello F, Sharma A (2019) Smartphone-based Gonio-imaging: a novel addition to glaucoma screening tools. J Glaucoma 28(9):e149–e150CrossRefPubMed
20.
Zurück zum Zitat Snyder BM, Sié A, Tapsoba C, Dah C, Ouermi L, Zakane SA et al (2019) Smartphone photography as a possible method of post-validation trachoma surveillance in resource-limited settings. Int Health 11(6):613–615CrossRefPubMedPubMedCentral Snyder BM, Sié A, Tapsoba C, Dah C, Ouermi L, Zakane SA et al (2019) Smartphone photography as a possible method of post-validation trachoma surveillance in resource-limited settings. Int Health 11(6):613–615CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Brennan PF, McNeil AJ, Jing M, Awuah A, Finlay DD, Blighe K et al (2019) Quantitative assessment of the conjunctival microcirculation using a smartphone and slit-lamp biomicroscope. Microvasc Res 126:103907CrossRefPubMed Brennan PF, McNeil AJ, Jing M, Awuah A, Finlay DD, Blighe K et al (2019) Quantitative assessment of the conjunctival microcirculation using a smartphone and slit-lamp biomicroscope. Microvasc Res 126:103907CrossRefPubMed
22.
Zurück zum Zitat Pujari A, Selvan H, Asif MI, Gupta B, Dada T (2019) Smartphone-aided quantification of iridocorneal angle. J Glaucoma 28(9):e153–e155CrossRefPubMed Pujari A, Selvan H, Asif MI, Gupta B, Dada T (2019) Smartphone-aided quantification of iridocorneal angle. J Glaucoma 28(9):e153–e155CrossRefPubMed
23.
Zurück zum Zitat Nesemann JM, Seider MI, Snyder BM, Maamari RN, Fletcher DA, Haile BA et al (2020) Comparison of smartphone photography, single-lens reflex photography, and field-grading for trachoma. Am J Trop Med Hyg 103(6):2488–2491CrossRefPubMedPubMedCentral Nesemann JM, Seider MI, Snyder BM, Maamari RN, Fletcher DA, Haile BA et al (2020) Comparison of smartphone photography, single-lens reflex photography, and field-grading for trachoma. Am J Trop Med Hyg 103(6):2488–2491CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Jalil M, Ferenczy SR, Shields CL (2017) iPhone 4s and iPhone 5s imaging of the eye. Ocul Oncol Pathol 3(1):49–55CrossRefPubMed Jalil M, Ferenczy SR, Shields CL (2017) iPhone 4s and iPhone 5s imaging of the eye. Ocul Oncol Pathol 3(1):49–55CrossRefPubMed
25.
Zurück zum Zitat Fliotsos MJ, Deljookorani S, Dzhaber D, Chandan S, Ighani M, Eghrari AO (2020) Qualitative and quantitative analysis of the corneal endothelium with smartphone specular microscopy. Cornea 39(7):924–929CrossRefPubMedPubMedCentral Fliotsos MJ, Deljookorani S, Dzhaber D, Chandan S, Ighani M, Eghrari AO (2020) Qualitative and quantitative analysis of the corneal endothelium with smartphone specular microscopy. Cornea 39(7):924–929CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Sanguansak T, Morley K, Morley M, Kusakul S, Lee R, Shieh E et al (2017) Comparing smartphone camera adapters in imaging post-operative cataract patients. J Telemed Telecare 23(1):36–43CrossRefPubMed Sanguansak T, Morley K, Morley M, Kusakul S, Lee R, Shieh E et al (2017) Comparing smartphone camera adapters in imaging post-operative cataract patients. J Telemed Telecare 23(1):36–43CrossRefPubMed
28.
Zurück zum Zitat Mohammadpour M, Mohammadpour L, Hassanzad M (2016) Smartphone assisted slit lamp free anterior segment imaging: a novel technique in teleophthalmology. Cont Lens Anterior Eye 39(1):80–81CrossRefPubMed Mohammadpour M, Mohammadpour L, Hassanzad M (2016) Smartphone assisted slit lamp free anterior segment imaging: a novel technique in teleophthalmology. Cont Lens Anterior Eye 39(1):80–81CrossRefPubMed
29.
Zurück zum Zitat Barsam A, Bhogal M, Morris S, Little B (2010) Anterior segment slitlamp photography using the iPhone. J Cataract Refract Surg 36(7):1240–1241CrossRefPubMed Barsam A, Bhogal M, Morris S, Little B (2010) Anterior segment slitlamp photography using the iPhone. J Cataract Refract Surg 36(7):1240–1241CrossRefPubMed
31.
Zurück zum Zitat Haag-Streit (2019) Slit lamp BQ 900. Haag-Streit, Zurich Haag-Streit (2019) Slit lamp BQ 900. Haag-Streit, Zurich
33.
Zurück zum Zitat Haag-Streit (2019) Imaging modules IM 900 & IM 600, 8th edn. Haag-Streit, Switzerland Haag-Streit (2019) Imaging modules IM 900 & IM 600, 8th edn. Haag-Streit, Switzerland
34.
Zurück zum Zitat Chan JB, Ho HC, Ngah NF, Hussein E (2014) DIY-smartphone slit-lamp adaptor. J Mobile Technol Med 3(1):16–22CrossRef Chan JB, Ho HC, Ngah NF, Hussein E (2014) DIY-smartphone slit-lamp adaptor. J Mobile Technol Med 3(1):16–22CrossRef
35.
Zurück zum Zitat Kalra G, Ichhpujani P, Thakur S, Singh RB, Sharma U, Kumar S (2021) A pilot study for smartphone photography to assess bleb morphology and vasculature post- trabeculectomy. Int Ophthalmol 41:483–490CrossRefPubMed Kalra G, Ichhpujani P, Thakur S, Singh RB, Sharma U, Kumar S (2021) A pilot study for smartphone photography to assess bleb morphology and vasculature post- trabeculectomy. Int Ophthalmol 41:483–490CrossRefPubMed
36.
Zurück zum Zitat Myung D, Jais A, He L, Chang RT (2014) Simple, low-cost smartphone adapter for rapid, high quality ocular anterior segment imaging: a photo diary. J Mobile Technol Med 3(1):2–8CrossRef Myung D, Jais A, He L, Chang RT (2014) Simple, low-cost smartphone adapter for rapid, high quality ocular anterior segment imaging: a photo diary. J Mobile Technol Med 3(1):2–8CrossRef
37.
Zurück zum Zitat Sink J, Blatt S, Yoo D, Henry M, Yang SD, Vasaiwala R et al (2020) A novel telemedicine technique for evaluation of ocular exam findings via smartphone images. J Telemed Telecare 28:197–202CrossRefPubMed Sink J, Blatt S, Yoo D, Henry M, Yang SD, Vasaiwala R et al (2020) A novel telemedicine technique for evaluation of ocular exam findings via smartphone images. J Telemed Telecare 28:197–202CrossRefPubMed
38.
Zurück zum Zitat Solyman O, Ahmad M, Arora K, Henderson AD, Carey A (2021) Stereoscopic three-dimensional (3D) slit-lamp photography using a compact 3D digital camera. Indian J Ophthalmol 69(5):1303–1305CrossRefPubMedPubMedCentral Solyman O, Ahmad M, Arora K, Henderson AD, Carey A (2021) Stereoscopic three-dimensional (3D) slit-lamp photography using a compact 3D digital camera. Indian J Ophthalmol 69(5):1303–1305CrossRefPubMedPubMedCentral
39.
Zurück zum Zitat Hu S, Wu H, Luan X, Wang Z, Adu M, Wang X et al (2020) Portable handheld slit-lamp based on a smartphone camera for cataract screening. J Ophthalmol 2020:1037689CrossRefPubMedPubMedCentral Hu S, Wu H, Luan X, Wang Z, Adu M, Wang X et al (2020) Portable handheld slit-lamp based on a smartphone camera for cataract screening. J Ophthalmol 2020:1037689CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Chen D, Ho Y, Sasa Y, Lee J, Yen CC, Tan C (2021) Machine learning-guided prediction of central anterior chamber depth using slit lamp images from a portable smartphone device. Biosensors 11(6):182CrossRefPubMedPubMedCentral Chen D, Ho Y, Sasa Y, Lee J, Yen CC, Tan C (2021) Machine learning-guided prediction of central anterior chamber depth using slit lamp images from a portable smartphone device. Biosensors 11(6):182CrossRefPubMedPubMedCentral
41.
Zurück zum Zitat Bifolck E, Fink A, Pedersen D, Gregory T (2018) Smartphone imaging for the ophthalmic examination in primary care. Jaapa 31(8):34–38CrossRefPubMed Bifolck E, Fink A, Pedersen D, Gregory T (2018) Smartphone imaging for the ophthalmic examination in primary care. Jaapa 31(8):34–38CrossRefPubMed
42.
Zurück zum Zitat Kalra G, Ichhpujani P, Thakur S, Singh RB, Sharma U, Kumar S (2021) A pilot study for smartphone photography to assess bleb morphology and vasculature post- trabeculectomy. Int Ophthalmol. 41:483–490. Kalra G, Ichhpujani P, Thakur S, Singh RB, Sharma U, Kumar S (2021) A pilot study for smartphone photography to assess bleb morphology and vasculature post- trabeculectomy. Int Ophthalmol. 41:483–490.
43.
Zurück zum Zitat Sink J, Blatt S, Yoo D, Henry M, Yang SD, Vasaiwala R et al (2020) A novel telemedicine technique for evaluation of ocular exam findings via smartphone images. J Telemed Telecare 28:197–202 Sink J, Blatt S, Yoo D, Henry M, Yang SD, Vasaiwala R et al (2020) A novel telemedicine technique for evaluation of ocular exam findings via smartphone images. J Telemed Telecare 28:197–202
Metadaten
Titel
Assessing the subjective quality of smartphone anterior segment photography: a non-inferiority study
verfasst von
Raghav Goel
Carmelo Macri
Bobak Bahrami
Robert Casson
Weng Onn Chan
Publikationsdatum
26.08.2022
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 2/2023
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-022-02437-9

Weitere Artikel der Ausgabe 2/2023

International Ophthalmology 2/2023 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.