Beneficial aspects of real time flow measurements for the management of acute right ventricular heart failure following continuous flow ventricular assist device implantation

verfasst von: Sotirios Spiliopoulos, Dilek Guersoy, Reiner Koerfer, Gero Tenderich

Erschienen in: Journal of Cardiothoracic Surgery | Ausgabe 1/2012

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Optimal management of acute right heart failure following the implantation of a left ventricular assist device requires a reliable estimation of left ventricular preload and contractility. This is possible by real-time pump blood flow measurements.

Clinical case

We performed implantation of a continuous flow left ventricular assist device in a 66 years old female patient with an end-stage heart failure on the grounds of a dilated cardiomyopathy. Real-time pump blood flow was directly measured by an ultrasonic flow probe placed around the outflow graft.

Diagnosis

The progressive decline of real time flow and the loss of pulsatility were associated with an increase of central venous pressure, inotropic therapy and progressive renal failure suggesting the presence of an acute right heart failure. Diagnosis was validated by echocardiography and thermodilution measurements.

Treatment

Temporary mechanical circulatory support of the right ventricle was successfully performed. Real time flow measurement proved to be a useful tool for the diagnosis and ultimately for the management of right heart failure including the weaning from extracorporeal membrane oxygenation.
Hinweise

Competing interest

The authors exclude any conflict of interest.

Authors’ contributions

SS drafted the manuscript. DG critically reviewed the manuscript. RK and GT approved the manuscript for submission. All authors read and approved the final manuscript.

Background

Right heart failure is a common complication following left ventricular circulatory support[1]. Diagnosis and management are usually based upon echocardiographic findings and measurements of mixed venous saturation which provide information about left ventricular preload and left ventricular contractility. Alternatively this information could be easily obtained by the pump blood flow estimated by the left ventricular assist device. However this would not be accurate since currently available assist devices do not measure real-time blood flow but only calculate it using the rotational speed of the pump and the power consumption of the motor. Furthermore this calculation would be susceptible to external factors such as blood viscosity[2]. In contrast, real-time blood flow measurements are accurate. In this paper we describe a case of an acute right heart failure following left ventricular circulatory support which was diagnosed and successfully managed based upon information obtained through real-time blood flow measurements.

Case presentation

A continuous flow left ventricular assist device (HeartAssist 5; Micromed Cardiovascular Inc. Houston, USA) was implanted in a 66 years old female patient with an end-stage heart failure NYHA IV with a history of dilated cardiomyopathy. Preoperative echocardiography showed a severe impairment of the left ventricular function (EF 10%), a secondary pulmonary hypertension with an estimated systolic pulmonary artery pressure of 48 mmHg and a right ventricular dysfunction with a tricuspid annular plane systolic excursion (TAPSE) of 15 mm. The cardiac output, determined by right heart catheterization, was 3,5 L/min (Cardiac Index 1,92 L/min/m2) with a systemic peripheral vascular resistance of 1350 dyn.sec.cm5 and pulmonary vascular resistance of 383 dyn·s/cm5). After four days intravenous therapy with milrinone (14 μg/kg/hr)., cardiac output improved to 4,6 L/min (Cardiac Index 2,5 L/min/m2) with a systemic vascular resistance of 1000 dyn·s/cm5 and pulmonary vascular resistance decreased to 190 dyn·s/ cm5. After implant of the HeartAssist5, the thorax could not be completely closed, due to severe fluid retention and oedema. Complete closure resulted in severe right ventricular failure. During the first and the second postoperative day the mean real-time flow was 4,8 L/min (range: 4,4 L/min-5,2 L/min) at a fixed pump speed of 8400 rpm (range: 8300–8500 rpm) and the flow waveforms showed very pulsatile flows (2.0 L/min in diastole and 6.5 L/min in systole). In the third postoperative day the mean real-time flow decreased to 4,3 L/min (range: 3,9-5,1 L/min) and the flow curve became flat. Real time flow was correlated to cardiac output measured by thermodilution (4,07 L/min on a systemic peripheral vascular resistance of 766 dyn·s/cm5). Initialy a low-dose inotropic therapy with milrinone (3 μg/kg/hr), noradrenalin (1 μg/kg/hr) and dopamine (200 μg/kg/hr) was enough to restore cardiac output (4,7 L/min with a systemic vascular resistance of 938 dyn·s/cm5). Unfortunately, in the eighth postoperative day, the central venous pressure increased from 9 mmHg to a mean 16 mmHg (range 14–20 mmHg) and the real-time flow decreased (mean 3,9 L/min, range 3,5-4,7 L/min). 24 hours later the patient developed renal failure, necessitating haemodialysis and a high dose inotropic therapy became mandatory (norepinephrine 8 μg/kg/hr, Milrinone 11 μg/kg/hr). Despite maximal inotropic therapy, flows continued to decline and on the tenth postoperative day the patient was placed on venoarterial extracorporeal membrane oxygenation (Levitronix Centrimag, Zurich, Switzerland). The two-stage venous cannula was inserted to the right atrium and the arterial cannula in the main pulmonary artery. Real time flow & pulsatility recovered promptly. Mean flow values over the first 24 hours ranged from 4,8 L/min to 6,2 L/min at a fixed pump speed of 8700 rpm. After four days of ECMO, the inotropes were weaned. At the sixth day of ECMO support, weaning from ECMO was started. Mean real time flow remained stable over the weaning procedure (5 L/min; range 4,3-5,5 L/min). The ECMO circuit was disconnected at the 14th post-operative day. Three days after ECMO removal, the Thorax was closed, and the further course was uneventful.

Conclusions

Currently available continuous flow ventricular assist devices estimate the pump blood flow with a calculation using the power consumption of the motor and the rotational speed of the pump[3]. However, it is known that the estimated blood flow depends to a large extent from external factors such as blood viscosity[2], rotational speed and loading of the ventricle. The estimated blood flow is therefore not accurate. Right heart failure is a common complication following left ventricular circulatory support[1]. The main reasons are a shift of the interventricular septum to the left and the inability of a dysfunctional right ventricle to cope with the increased venous return, delivered by the LVAD. It is therefore essential for the management of right ventricular failure to precisely determine the left ventricular preload and contractility. This is quite well possible by the means of transesophageal echocardiography and mixed venous saturation but is only available in the early post-op phase. The HeartAssist 5 left ventricular assist device instead has an ultrasonic flow probe placed around the outflow graft which provides real-time flow measurement[4]. This measurement is reliable and gives information about the left ventricular preload and the left ventricular contractility[5]. In our case real time flow was repeatedly validated by haemodynamic measurements obtained by thermodilution. The progressive decline of real time flow and loss of pulsatility was associated with an increase of central venous pressure, inotropic therapy and progressive renal failure requiring ultimately a mechanical circulatory support of the right ventricle. Furthermore the real time flow measurement proved to be a useful tool for the weaning from the extracorporeal membrane oxygenation since it enabled the monitoring of the left ventricular preload & contractility (0.6 – 8.3 L/min) throughout the post-operative course. The loss of pulsatility in the flow curve is an early warning sign for insufficient preload of the LV, which will allow earlier intervention and therefore may help to better manage critically ill patients. Furthermore real-time flow measurements proved to be a useful tool and a reliable alternative to conventional techniques for the measurement of cardiac output in the clinical setting.
Written informed consent was obtained from the patient for publication of this report.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interest

The authors exclude any conflict of interest.

Authors’ contributions

SS drafted the manuscript. DG critically reviewed the manuscript. RK and GT approved the manuscript for submission. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Dang NC, Topkara VK, Mercando M: Right heart failure after left ventricular assist device implantation in patients with chronic congestive heart failure. J Heart Lung Transplant. 2006, 25 (1): 1-6. 10.1016/j.healun.2005.07.008.CrossRefPubMed Dang NC, Topkara VK, Mercando M: Right heart failure after left ventricular assist device implantation in patients with chronic congestive heart failure. J Heart Lung Transplant. 2006, 25 (1): 1-6. 10.1016/j.healun.2005.07.008.CrossRefPubMed
2.
Zurück zum Zitat Tsukiya T, Taenaka Y, Nishinaka T: Application of indirect flow measurements using motor driving signals to a centrifugal blood pump with an integrated motor. Artif Organs. 2001, 25: 692-696. 10.1046/j.1525-1594.2001.06858.x.CrossRefPubMed Tsukiya T, Taenaka Y, Nishinaka T: Application of indirect flow measurements using motor driving signals to a centrifugal blood pump with an integrated motor. Artif Organs. 2001, 25: 692-696. 10.1046/j.1525-1594.2001.06858.x.CrossRefPubMed
3.
Zurück zum Zitat Thunberg CA, Gaitan BD, Arabia FA: Ventricular Assist devices today and tomorrow. J Cardiothorac Vasc Anaesth. 2010, 24 (3): 656-680.CrossRef Thunberg CA, Gaitan BD, Arabia FA: Ventricular Assist devices today and tomorrow. J Cardiothorac Vasc Anaesth. 2010, 24 (3): 656-680.CrossRef
4.
Zurück zum Zitat Noon GP, Loebe M: Current status of the MicroMed DeBakey Noon Ventricular Assist device. Tex Heart Inst J. 2010, 37 (6): 652-663.PubMedPubMedCentral Noon GP, Loebe M: Current status of the MicroMed DeBakey Noon Ventricular Assist device. Tex Heart Inst J. 2010, 37 (6): 652-663.PubMedPubMedCentral
5.
Zurück zum Zitat Benkowski R, Lynch B, Morello G: Real time flow measurement for a continuous flow VAD. ASAIO. 2001, 47 (2): 133-CrossRef Benkowski R, Lynch B, Morello G: Real time flow measurement for a continuous flow VAD. ASAIO. 2001, 47 (2): 133-CrossRef
Metadaten
Titel
Beneficial aspects of real time flow measurements for the management of acute right ventricular heart failure following continuous flow ventricular assist device implantation
verfasst von
Sotirios Spiliopoulos
Dilek Guersoy
Reiner Koerfer
Gero Tenderich
Publikationsdatum
01.12.2012
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2012
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-7-119

Weitere Artikel der Ausgabe 1/2012

Journal of Cardiothoracic Surgery 1/2012 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Was nützt die Kraniektomie bei schwerer tiefer Hirnblutung?

17.05.2024 Hirnblutung Nachrichten

Eine Studie zum Nutzen der druckentlastenden Kraniektomie nach schwerer tiefer supratentorieller Hirnblutung deutet einen Nutzen der Operation an. Für überlebende Patienten ist das dennoch nur eine bedingt gute Nachricht.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.