The commercial use of telemetric intracranial pressure (ICP) monitoring has been in use for over a decade now. The first device which was licensed is the Neurovent-P-tel probe® (Raumedic, Helmbrecht, Germany) in 2009 followed by reports of its first clinical use [2, 3, 8, 24]. Preclinical studies showed a sufficient long-term stability of the ICP measurement with only a minimal zero drift with good accuracy [9, 11, 12].
The second device followed with the Sensor Reservoir® which was approved for clinical use in 2015 (Christoph Miethke GmbH & Co.KG, Potsdam, Germany). The Sensor Reservoir® is to be integrated into the shunt system, and there is no limitation about the approved implantation time in opposition to the P-tel probe whose approved implantation time is limited to 90 days. The first experience with its clinical use was published in 2017 [6], and the first larger study appeared in an article in 2018 describing the experience of one of its biggest technical potentials—the intracranial pressure–guided shunt valve adjustment [1].
Nowadays, telemetric ICP monitoring is feasible and useable for a wide spectrum of pathologies.
First of all, compared to conventional ICP monitoring methods, it has the advantage of being a closed system decreasing the risk of infection [3, 13]. This is particularly an argument for using the P-tel device for ICP monitoring in the neuro-intensive care setting which is usually the application field for cabled ICP probes [15, 21].
In addition, the closed system and the long capable implantation duration facilitate different areas of application.
Long-term ICP monitoring with the Neurovent-P-tel® after ETV for 8 to 12 weeks is especially useful for treatment response prediction since the average time of ETV failure occurs within this time period [4, 10, 23].
Furthermore, it is a helpful tool for the diagnosis and prediction of shunt responsiveness in patients with complex hydrocephalus and ICP-related diseases, like idiopathic intracranial hypertension—in both adults and children [2, 14, 19, 20].
Additionally, it allows the monitoring of the shunt effect and the verification of a proper shunt function over a longer time period in complex shunt patients [1, 2, 19, 20].
Beyond that, the possibility of doing measurements in supine and vertical positions and over several measurement periods provides a clear distinction between over- and underdrainage. The potential of ICP-guided optimization of valve setting can be exploited to full extent due to use of both adjustable differential and anti-gravitational valves [1, 8, 1820], which even a reduction in surgical revisions and radiation exposure due to a reduced necessity of imaging can be achieved [19]. Also, it provides a lower rate of hospitalization and there is even an alternative for home-telemonitoring [17, 19, 22].
“Does it change management” ask the authors of the study presented in the following, where a series of twelve shunt treated patients before and after insertion of a Sensor Reservoir® were investigated. Without giving away too much of the answer provided by this study, the authors found an improvement of symptoms in 75% of patients, a reduction of radiation exposure and hospitalization, and increased cost-effectiveness, similar to results which previous studies have yielded [1, 5, 19].
But telemetric ICP measurement also has its limitations and disadvantages.
First of all, the sampling frequency is significantly lower compared to cable bound devices (Shunt Reservoir® 40 Hz, Neurovent-P-tel® 5Hz). A single pressure curve analysis is therefore not possible, although measurement of pulse pressure amplitude with both devices is feasible. A further obstacle is the risk of zero drifting which obviously threatens the accuracy of the results. This problem occurs not only in telemetric ICP monitoring but also in conventional measurement. Nevertheless, because of the long durability of the implant, the time-dependent risk is higher in telemetric than in conventional measurements. The median shift from the baseline was in a study with P-tel® devices 2.5 mmHg on average (implantation time was often longer than the CE-approval time of 90 days) [16]. Therefore, analyzing dynamic ICP values like vasogenic slow waves and the pulse pressure amplitude is much more essential than static ICP values like the mean ICP [7]. Gathering this information is feasible with both devices but its analysis is not automatically done and therefore depends on the experience of the neurosurgeon and is time consuming. And in the case of the Shunt Sensor®, there is so far a lack of an appropriate analysis program. Here is room for improvement.
Nevertheless, telemetric ICP monitoring is already a valuable tool. The choice of the most suitable device depends on diagnostic goal in the individual case: for long-duration measurements, for example, for diagnostic purposes or for monitoring response to ETV, the Neurovent-P-tel® is to be preferred. For continuous, repeated ambulatory follow-up measurements intending to verify subtle drainage-related shunt failure and to control the valve setting adjustment in sequential outpatient presentations over years the Shunt Sensor® is more suitable.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
download
DOWNLOAD
print
DRUCKEN

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
14.
Metadaten
Titel
Benefits and obstacles of telemetric ICP monitoring
verfasst von
Joachim M. K. Oertel
Matthias J. M. Huelser
Publikationsdatum
07.02.2021
Verlag
Springer Vienna
Erschienen in
Acta Neurochirurgica / Ausgabe 4/2021
Print ISSN: 0001-6268
Elektronische ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-021-04730-5

Weitere Artikel der Ausgabe 4/2021

Acta Neurochirurgica 4/2021 Zur Ausgabe

Letter to the editor - Neurosurgical technique evaluation

Endoscopic microvascular decompression for glossopharyngeal neuralgia

Editorial (by Invitation) - Obituary

Evandro de Oliveira, the “Brain Whisperer”

Original Article - CSF Circulation

The Sensor Reservoir—does it change management?

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Frühe Alzheimertherapie lohnt sich

25.04.2024 AAN-Jahrestagung 2024 Nachrichten

Ist die Tau-Last noch gering, scheint der Vorteil von Lecanemab besonders groß zu sein. Und beginnen Erkrankte verzögert mit der Behandlung, erreichen sie nicht mehr die kognitive Leistung wie bei einem früheren Start. Darauf deuten neue Analysen der Phase-3-Studie Clarity AD.

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.