Between-Scanner and Between-Visit Variation in Normal White Matter Apparent Diffusion Coefficient Values in the Setting of a Multi-Center Clinical Trial
verfasst von:
J. Huo, J. Alger, PhD, H. Kim, M. Brown, K. Okada, W. Pope, J. Goldin
Erschienen in:
Clinical Neuroradiology
|
Ausgabe 4/2016
Einloggen, um Zugang zu erhalten
Abstract
Purpose
To study the between-scanner variation and the between-visit reproducibility of brain apparent diffusion coefficient (ADC) measurements in the setting of a multi-center chemotherapy clinical trial for glioblastoma multiforme.
Methods and Materials
ADC maps of 52 patients at six sites were calculated in-house from diffusion-weighted images obtained by seven individual scanner models of two vendors. The median and coefficient of variation (CV) of normal brain white matter ADC values from a defined region of interest were used to evaluate the differences among scanner models, vendors, magnetic fields, as well as successive visits. All patients participating in this study signed institutional review board approved informed consent. Data acquisition was performed in compliance with all applicable Health Insurance Portability and Accountability Act regulations. The study spanned from August 1, 2006, to January 29, 2008.
Results
For baseline median ADC, no difference was observed between the different scanner models, different vendors, and different magnetic field strength. For baseline ADC CV, a significant difference was found between different scanner models (p = 0.0002). No between-scanner difference was observed in ADC changes between two visits. For between-visit reproducibility, significant difference was seen between the ADC values measured at two successive visits for the whole patient group.
Conclusion
The CVs varied significantly between scanners, presumably due to image noise. Consistent scanner parameter setup can improve reproducibility of the ADC measurements between visits.