30.12.2022 | Research Article

Characteristics of the deep learning-based virtual monochromatic image with fast kilovolt-switching CT: a phantom study

verfasst von: Yuna Katsuyama, Tsukasa Kojima, Takashi Shirasaka, Masatoshi Kondo, Toyoyuki Kato

Erschienen in: Radiological Physics and Technology | Ausgabe 1/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

We assessed the physical properties of virtual monochromatic images (VMIs) obtained with different energy levels in various contrast settings and radiation doses using deep learning-based spectral computed tomography (DL-Spectral CT) and compared the results with those from single-energy CT (SECT) imaging.

Materials and methods

A Catphan® 600 phantom was scanned by DL-Spectral CT at various radiation doses. We reconstructed the VMIs obtained at 50, 70, and 100 keV. SECT (120 kVp) images were acquired at the same radiation doses. The standard deviations of the CT number and noise power spectrum (NPS) were calculated for noise characterization. We evaluated the spatial resolution by determining the 10% task-based transfer function (TTF) level, and we assessed the task-based detectability index (d’).

Results

Regardless of the radiation dose, the noise was the lowest at 70 keV VMI. The NPS showed that the noise amplitude at all spatial frequencies was the lowest among other VMI and 120 kVp images. The spatial resolution was higher for 70 keV VMI compared to the other VMIs, except for high-contrast objects. The d’ of 70 keV VMI was the highest among the VMI and 120 kVp images at all radiation doses and contrast settings. The d’ of the 70 keV VMIs at the minimum dose was higher than that at the maximum dose in any other image.

Conclusion

The physical properties of the DL-Spectral CT VMIs varied with the energy level. The 70 keV VMI had the highest detectability by far among the VMI and 120-kVp images. DL-Spectral CT may be useful to reduce radiation doses.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Thieme SF, Graute V, Nikolaou K, Maxien D, Reiser MF, Hacker M, et al. Dual energy CT lung perfusion imaging—Correlation with SPECT/CT. Eur J Radiol. 2012;81:360–5.CrossRefPubMed Thieme SF, Graute V, Nikolaou K, Maxien D, Reiser MF, Hacker M, et al. Dual energy CT lung perfusion imaging—Correlation with SPECT/CT. Eur J Radiol. 2012;81:360–5.CrossRefPubMed
2.
Zurück zum Zitat Hu R, Daftari Besheli L, Young J, Wu M, Pomerantz S, Lev MH, et al. Dual-energy head CT enables accurate distinction of intraparenchymal hemorrhage from calcification in emergency department patients. Radiology. 2016;280:177–83.CrossRefPubMed Hu R, Daftari Besheli L, Young J, Wu M, Pomerantz S, Lev MH, et al. Dual-energy head CT enables accurate distinction of intraparenchymal hemorrhage from calcification in emergency department patients. Radiology. 2016;280:177–83.CrossRefPubMed
3.
Zurück zum Zitat Lee SH, Lee JM, Kim KW, et al. Dual-energy computed tomography to assess tumor response to hepatic radiofrequency ablation: potential diagnostic value of virtual noncontrast images and iodine maps. Invest Radiol. 2011;46:77–84.CrossRefPubMed Lee SH, Lee JM, Kim KW, et al. Dual-energy computed tomography to assess tumor response to hepatic radiofrequency ablation: potential diagnostic value of virtual noncontrast images and iodine maps. Invest Radiol. 2011;46:77–84.CrossRefPubMed
4.
Zurück zum Zitat Shuman WP, Chan KT, Busey JM, Mitsumori LM, Koprowicz KM. Dual-energy CT aortography with 50% reduced iodine dose versus single-energy CT aortography with standard iodine dose. Acad Radiol. 2016;23:611–8.CrossRefPubMed Shuman WP, Chan KT, Busey JM, Mitsumori LM, Koprowicz KM. Dual-energy CT aortography with 50% reduced iodine dose versus single-energy CT aortography with standard iodine dose. Acad Radiol. 2016;23:611–8.CrossRefPubMed
5.
Zurück zum Zitat Bhosale P, Le O, Balachandran A, Fox P, Paulson E, Tamm E. Quantitative and qualitative comparison of single-source dual-energy computed tomography and 120-kVp computed tomography for the assessment of pancreatic ductal adenocarcinoma. J Comput Assist Tomogr. 2015;39:907–13.CrossRefPubMedPubMedCentral Bhosale P, Le O, Balachandran A, Fox P, Paulson E, Tamm E. Quantitative and qualitative comparison of single-source dual-energy computed tomography and 120-kVp computed tomography for the assessment of pancreatic ductal adenocarcinoma. J Comput Assist Tomogr. 2015;39:907–13.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Mileto A, Nelson RC, Samei E, Jaffe TA, Paulson EK, Barina A, et al. Impact of dual-energy multi-detector row CT with virtual monochromatic imaging on renal cyst pseudoenhancement: in vitro and in vivo study. Radiology. 2014;272:767–76.CrossRefPubMed Mileto A, Nelson RC, Samei E, Jaffe TA, Paulson EK, Barina A, et al. Impact of dual-energy multi-detector row CT with virtual monochromatic imaging on renal cyst pseudoenhancement: in vitro and in vivo study. Radiology. 2014;272:767–76.CrossRefPubMed
8.
Zurück zum Zitat Ohira S, Karino T, Ueda Y, Nitta Y, Kanayama N, Miyazaki M, et al. How well does dual-energy CT with fast kilovoltage switching quantify CT number and iodine and calcium concentrations? Acad Radiol. 2018;25:519–28.CrossRefPubMed Ohira S, Karino T, Ueda Y, Nitta Y, Kanayama N, Miyazaki M, et al. How well does dual-energy CT with fast kilovoltage switching quantify CT number and iodine and calcium concentrations? Acad Radiol. 2018;25:519–28.CrossRefPubMed
9.
Zurück zum Zitat Washio H, Ohira S, Karino T, Nitta Y, Hayashi M, Miyazaki M, et al. Accuracy of quantification of iodine and hounsfield unit values on virtual monochromatic imaging using dual-energy computed tomography: comparison of dual-layer computed tomography with fast kilovolt-switching computed tomography. J Comput Assist Tomogr. 2018;42:965–71.CrossRefPubMed Washio H, Ohira S, Karino T, Nitta Y, Hayashi M, Miyazaki M, et al. Accuracy of quantification of iodine and hounsfield unit values on virtual monochromatic imaging using dual-energy computed tomography: comparison of dual-layer computed tomography with fast kilovolt-switching computed tomography. J Comput Assist Tomogr. 2018;42:965–71.CrossRefPubMed
11.
Zurück zum Zitat Kojima T, Shirasaka T, Kondo M, Kato T, Nishie A, Ishigami K, et al. A novel fast kilovoltage switching dual-energy CT with deep learning: accuracy of CT number on virtual monochromatic imaging and iodine quantification. Phys Med. 2021;81:253–61.CrossRefPubMed Kojima T, Shirasaka T, Kondo M, Kato T, Nishie A, Ishigami K, et al. A novel fast kilovoltage switching dual-energy CT with deep learning: accuracy of CT number on virtual monochromatic imaging and iodine quantification. Phys Med. 2021;81:253–61.CrossRefPubMed
13.
Zurück zum Zitat Samei E, Bakalyar D, Boedeker KL, Brady S, Fan J, Leng S, et al. Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233. Med Phys. 2019;46:e735–56.CrossRefPubMed Samei E, Bakalyar D, Boedeker KL, Brady S, Fan J, Leng S, et al. Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233. Med Phys. 2019;46:e735–56.CrossRefPubMed
14.
Zurück zum Zitat Samei E, Richard S. Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys. 2015;42:314–23.CrossRefPubMed Samei E, Richard S. Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys. 2015;42:314–23.CrossRefPubMed
15.
Zurück zum Zitat Shirasaka T, Kojima T, Funama Y, Sakai Y, Kondo M, Mikayama R, et al. Image quality improvement with deep learning-based reconstruction on abdominal ultrahigh-resolution CT: a phantom study. J Appl Clin Med Phys. 2021;22:286–96.CrossRefPubMedPubMedCentral Shirasaka T, Kojima T, Funama Y, Sakai Y, Kondo M, Mikayama R, et al. Image quality improvement with deep learning-based reconstruction on abdominal ultrahigh-resolution CT: a phantom study. J Appl Clin Med Phys. 2021;22:286–96.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Boedeker KL, Cooper VN, McNitt-Gray MF. Application of the noise power spectrum in modern diagnostic MDCT: part I. measurement of noise power spectra and noise equivalent quanta. Phys Med Biol. 2007;52:4027–46.CrossRefPubMed Boedeker KL, Cooper VN, McNitt-Gray MF. Application of the noise power spectrum in modern diagnostic MDCT: part I. measurement of noise power spectra and noise equivalent quanta. Phys Med Biol. 2007;52:4027–46.CrossRefPubMed
17.
Zurück zum Zitat Urikura A, Ichikawa K, Hara T, Nishimaru E, Nakaya Y. Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography. Radiol Phys Technol. 2014;7:358–66.CrossRefPubMed Urikura A, Ichikawa K, Hara T, Nishimaru E, Nakaya Y. Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography. Radiol Phys Technol. 2014;7:358–66.CrossRefPubMed
18.
Zurück zum Zitat Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39:4115–22.CrossRefPubMed Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39:4115–22.CrossRefPubMed
19.
Zurück zum Zitat Greffier J, Si-Mohamed S, Dabli D, de Forges H, Hamard A, Douek P, et al. Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data. Eur Radiol. 2021;31:5324–34.CrossRefPubMed Greffier J, Si-Mohamed S, Dabli D, de Forges H, Hamard A, Douek P, et al. Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data. Eur Radiol. 2021;31:5324–34.CrossRefPubMed
20.
Zurück zum Zitat Eckstein M, Bartroff J, Abbey C, Whiting J, Bochud F. Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks. Opt Express. 2003;11:460–75.CrossRefPubMed Eckstein M, Bartroff J, Abbey C, Whiting J, Bochud F. Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks. Opt Express. 2003;11:460–75.CrossRefPubMed
21.
Zurück zum Zitat Greffier J, Frandon J, Larbi A, Beregi JP, Pereira F. CT iterative reconstruction algorithms: a task-based image quality assessment. Eur Radiol. 2020;30:487–500.CrossRefPubMed Greffier J, Frandon J, Larbi A, Beregi JP, Pereira F. CT iterative reconstruction algorithms: a task-based image quality assessment. Eur Radiol. 2020;30:487–500.CrossRefPubMed
22.
Zurück zum Zitat Greffier J, Larbi A, Frandon J, Moliner G, Beregi JP, Pereira F. Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from three manufacturers. Diagn Interv Imaging. 2019;100:401–10.CrossRefPubMed Greffier J, Larbi A, Frandon J, Moliner G, Beregi JP, Pereira F. Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from three manufacturers. Diagn Interv Imaging. 2019;100:401–10.CrossRefPubMed
23.
Zurück zum Zitat Ehman EC, Yu L, Manduca A, Hara AK, Shiung MM, Jondal D, et al. Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT. Radiographics. 2014;34:849–62.CrossRefPubMed Ehman EC, Yu L, Manduca A, Hara AK, Shiung MM, Jondal D, et al. Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT. Radiographics. 2014;34:849–62.CrossRefPubMed
24.
Zurück zum Zitat Geyer LL, Schoepf UJ, Meinel FG, Nance JW, J, Bastarrika G, Leipsic JA, et al. State of the art: iterative CT reconstruction techniques. Radiology. 2015;276:339–57.CrossRefPubMed Geyer LL, Schoepf UJ, Meinel FG, Nance JW, J, Bastarrika G, Leipsic JA, et al. State of the art: iterative CT reconstruction techniques. Radiology. 2015;276:339–57.CrossRefPubMed
Metadaten
Titel
Characteristics of the deep learning-based virtual monochromatic image with fast kilovolt-switching CT: a phantom study
verfasst von
Yuna Katsuyama
Tsukasa Kojima
Takashi Shirasaka
Masatoshi Kondo
Toyoyuki Kato
Publikationsdatum
30.12.2022
Verlag
Springer Nature Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 1/2023
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-022-00695-x

Weitere Artikel der Ausgabe 1/2023

Radiological Physics and Technology 1/2023 Zur Ausgabe

PET kann infarktgefährdete Koronararterien entdecken

04.06.2024 Koronare Herzerkrankung Nachrichten

Der Nachweis aktiver Plaques mittels 18F-Natriumfluorid-PET hilft nicht nur, infarktgefährdete Patienten, sondern auch infarktgefährdete Koronararterien zu erkennen. Von einer gezielten Behandlung vulnerabler Plaques ist man trotzdem weit entfernt.

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.