Skip to main content
Erschienen in: European Spine Journal 4/2012

01.06.2012 | Case Report

Clinical, radiological, histological and retrieval findings of Activ-L and Mobidisc total disc replacements: a study of two patients

verfasst von: Shennah Austen, Ilona M. Punt, Jack P. M. Cleutjens, Paul C. Willems, Steven M. Kurtz, Daniel W. MacDonald, Lodewijk W. van Rhijn, André van Ooij

Erschienen in: European Spine Journal | Sonderheft 4/2012

Einloggen, um Zugang zu erhalten

Abstract

Introduction

This study evaluates the short-term clinical outcome, radiological, histological and device retrieval findings of two patients with second generation lumbar total disc replacement (TDR).

Materials and methods

The first patient had a single level L4-L5 Activ-L TDR, the second patient a L4-L5 Mobidisc and L5-S1 Activ-L TDR. The TDRs were implanted elsewhere and had implantation times between 1.3 and 2.8 years.

Results

Plain radiographs and CT-scanning showed slight subsidence of the Activ-L TDR in both patients and facet joint degeneration. The patients underwent revision surgery because of recurrent back and leg pain. After removal of the TDR and posterolateral fusion, the pain improved. Histological examination revealed large ultrahigh molecular weight polyethylene (UHMWPE) particles and giant cells in the retrieved tissue surrounding the Mobidisc. The particles in the tissue samples of the Activ-L TDR were smaller and contained in macrophages. Retrieval analysis of the UHMWPE cores revealed evidence of minor adhesive and abrasive wear with signs of impingement in both TDR designs.

Conclusion

Although wear was unrelated to the reason for revision, this study demonstrates the presence of UHMWPE particles and inflammatory cells in second generation TDR. Long-term follow-up after TDR is indicated for monitoring wear and implant status.
Literatur
1.
Zurück zum Zitat Yajun W, Yue Z, Xiuxin H, Cui C (2010) A meta-analysis of artificial total disc replacement versus fusion for lumbar degenerative disc disease. Eur Spine J 19:1250–1261PubMedCrossRef Yajun W, Yue Z, Xiuxin H, Cui C (2010) A meta-analysis of artificial total disc replacement versus fusion for lumbar degenerative disc disease. Eur Spine J 19:1250–1261PubMedCrossRef
2.
3.
Zurück zum Zitat Kurtz SM (2009) Ultra high molecular weight polyethylene in total joint replacement and medical devices. In: UHMWPE Biomaterials Handbook, 2nd edn. Academic Press, New York Kurtz SM (2009) Ultra high molecular weight polyethylene in total joint replacement and medical devices. In: UHMWPE Biomaterials Handbook, 2nd edn. Academic Press, New York
4.
Zurück zum Zitat Harrop JS, Youssef JA, Maltenfort M, Vorwald P, Jabbour P, Bono CM, Goldfarb N, Vaccaro AR, Hilibrand AS (2008) Lumbar adjacent segment degeneration and disease after arthrodesis and total disc arthroplasty. Spine (Phila Pa 1976) 33:1701–1707CrossRef Harrop JS, Youssef JA, Maltenfort M, Vorwald P, Jabbour P, Bono CM, Goldfarb N, Vaccaro AR, Hilibrand AS (2008) Lumbar adjacent segment degeneration and disease after arthrodesis and total disc arthroplasty. Spine (Phila Pa 1976) 33:1701–1707CrossRef
5.
Zurück zum Zitat Chung SK, Kim YE, Wang KC (2009) Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis. Spine (Phila Pa 1976) 34:1281–1286CrossRef Chung SK, Kim YE, Wang KC (2009) Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis. Spine (Phila Pa 1976) 34:1281–1286CrossRef
6.
Zurück zum Zitat Huang RC, Girardi FP, Cammisa FP Jr, Wright TM (2003) The implications of constraint in lumbar total disc replacement. J Spinal Disord Tech 16:412–417PubMedCrossRef Huang RC, Girardi FP, Cammisa FP Jr, Wright TM (2003) The implications of constraint in lumbar total disc replacement. J Spinal Disord Tech 16:412–417PubMedCrossRef
7.
Zurück zum Zitat Steib J, Aubourg L, Beaurain J (2008) Mobidisc prosthesis. In: Yue JJ (ed) Motion preservation surgery of the spine. Saunders, Philadelphia, pp 326–329CrossRef Steib J, Aubourg L, Beaurain J (2008) Mobidisc prosthesis. In: Yue JJ (ed) Motion preservation surgery of the spine. Saunders, Philadelphia, pp 326–329CrossRef
8.
Zurück zum Zitat van den Eerenbeemt KD, Ostelo RW, van Royen BJ, Peul WC, van Tulder MW (2010) Total disc replacement surgery for symptomatic degenerative lumbar disc disease: a systematic review of the literature. Eur Spine J 19:1262–1280PubMedCrossRef van den Eerenbeemt KD, Ostelo RW, van Royen BJ, Peul WC, van Tulder MW (2010) Total disc replacement surgery for symptomatic degenerative lumbar disc disease: a systematic review of the literature. Eur Spine J 19:1262–1280PubMedCrossRef
9.
Zurück zum Zitat Ha SK, Kim SH, Kim DH, Park JY, Lim DJ, Lee SK (2009) Biomechanical study of lumbar spinal arthroplasty with a semi-constrained artificial disc (activ L) in the human cadaveric spine. J Korean Neurosurg Soc 45:169–175PubMedCrossRef Ha SK, Kim SH, Kim DH, Park JY, Lim DJ, Lee SK (2009) Biomechanical study of lumbar spinal arthroplasty with a semi-constrained artificial disc (activ L) in the human cadaveric spine. J Korean Neurosurg Soc 45:169–175PubMedCrossRef
10.
Zurück zum Zitat Zander T, Rohlmann A, Bergmann G (2009) Influence of different artificial disc kinematics on spine biomechanics. Clin Biomech (Bristol, Avon) 24:135–142CrossRef Zander T, Rohlmann A, Bergmann G (2009) Influence of different artificial disc kinematics on spine biomechanics. Clin Biomech (Bristol, Avon) 24:135–142CrossRef
11.
Zurück zum Zitat Moumene M, Geisler FH (2007) Comparison of biomechanical function at ideal and varied surgical placement for two lumbar artificial disc implant designs: mobile-core versus fixed-core. Spine (Phila Pa 1976) 32:1840–1851CrossRef Moumene M, Geisler FH (2007) Comparison of biomechanical function at ideal and varied surgical placement for two lumbar artificial disc implant designs: mobile-core versus fixed-core. Spine (Phila Pa 1976) 32:1840–1851CrossRef
12.
Zurück zum Zitat Delécrin J, Allain J, Steib J, Beucrain J, Dufour T, Chataigner H, Gau M, Huppert J, Onimus M, Zeegers W (2008) Mobidisc lumbar arthroplasty: Multi-center study with 2 years follow-up. In: 9th Congress of the European Federation of National Associations of Orthopaedics and Traumatology Nice, France Delécrin J, Allain J, Steib J, Beucrain J, Dufour T, Chataigner H, Gau M, Huppert J, Onimus M, Zeegers W (2008) Mobidisc lumbar arthroplasty: Multi-center study with 2 years follow-up. In: 9th Congress of the European Federation of National Associations of Orthopaedics and Traumatology Nice, France
13.
Zurück zum Zitat Yue JJ (2010) A prospective clinical comparison of 3 biomechanical types of lumbar total disc replacements: A semi-constrained device, a controlled translation device, and an unconstrained device, minimum 2 year follow-up. In: Annual Conference of the International Society for the Advancement of Spine Surgery. Perpignan, France Yue JJ (2010) A prospective clinical comparison of 3 biomechanical types of lumbar total disc replacements: A semi-constrained device, a controlled translation device, and an unconstrained device, minimum 2 year follow-up. In: Annual Conference of the International Society for the Advancement of Spine Surgery. Perpignan, France
14.
Zurück zum Zitat Punt IM, Cleutjens JP, de Bruin T, Willems PC, Kurtz SM, van Rhijn LW, Schurink GW, van Ooij A (2009) Periprosthetic tissue reactions observed at revision of total intervertebral disc arthroplasty. Biomaterials 30:2079–2084PubMedCrossRef Punt IM, Cleutjens JP, de Bruin T, Willems PC, Kurtz SM, van Rhijn LW, Schurink GW, van Ooij A (2009) Periprosthetic tissue reactions observed at revision of total intervertebral disc arthroplasty. Biomaterials 30:2079–2084PubMedCrossRef
15.
Zurück zum Zitat Kurtz SM, van Ooij A, Ross R, de Waal Malefijt J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7:12–21PubMedCrossRef Kurtz SM, van Ooij A, Ross R, de Waal Malefijt J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7:12–21PubMedCrossRef
16.
Zurück zum Zitat Punt I, Baxter R, van Ooij A, Willems P, van Rhijn L, Kurtz S, Steinbeck M (2011) Submicron sized ultra-high molecular weight polyethylene wear particle analysis from revised SB Charite III total disc replacements. Acta Biomater 7:3404–3411PubMedCrossRef Punt I, Baxter R, van Ooij A, Willems P, van Rhijn L, Kurtz S, Steinbeck M (2011) Submicron sized ultra-high molecular weight polyethylene wear particle analysis from revised SB Charite III total disc replacements. Acta Biomater 7:3404–3411PubMedCrossRef
17.
Zurück zum Zitat Punt IM, Austen S, Cleutjens JP, Kurtz SM, Ten Broeke RH, van Rhijn LW, Willems PC, Ooij AV (2011) Are periprosthetic tissue reactions observed after revision of total disc replacement comparable to the reactions observed after total hip or knee revision surgery? Spine (Phila Pa 1976) (in press) Punt IM, Austen S, Cleutjens JP, Kurtz SM, Ten Broeke RH, van Rhijn LW, Willems PC, Ooij AV (2011) Are periprosthetic tissue reactions observed after revision of total disc replacement comparable to the reactions observed after total hip or knee revision surgery? Spine (Phila Pa 1976) (in press)
18.
Zurück zum Zitat Gallo J, Raska M, Mrazek F, Petrek M (2008) Bone remodeling, particle disease and individual susceptibility to periprosthetic osteolysis. Physiol Res 57:339–349PubMed Gallo J, Raska M, Mrazek F, Petrek M (2008) Bone remodeling, particle disease and individual susceptibility to periprosthetic osteolysis. Physiol Res 57:339–349PubMed
19.
Zurück zum Zitat Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplast 17:649–661CrossRef Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplast 17:649–661CrossRef
20.
Zurück zum Zitat Mirra JM, Amstutz HC, Matos M, Gold R (1976) The pathology of the joint tissues and its clinical relevance in prosthesis failure. Clin Orthop Relat Res 117:221–240PubMed Mirra JM, Amstutz HC, Matos M, Gold R (1976) The pathology of the joint tissues and its clinical relevance in prosthesis failure. Clin Orthop Relat Res 117:221–240PubMed
21.
Zurück zum Zitat de Maat GH, Punt IM, van Rhijn LW, Schurink GW, van Ooij A (2009) Removal of the Charite lumbar artificial disc prosthesis: surgical technique. J Spinal Disord Tech 22:334–339PubMedCrossRef de Maat GH, Punt IM, van Rhijn LW, Schurink GW, van Ooij A (2009) Removal of the Charite lumbar artificial disc prosthesis: surgical technique. J Spinal Disord Tech 22:334–339PubMedCrossRef
22.
Zurück zum Zitat Kurtz SM, MacDonald D, Ianuzzi A, van Ooij A, Isaza J, Ross ER, Regan J (2009) The natural history of polyethylene oxidation in total disc replacement. Spine (Phila Pa 1976) 34:2369–2377CrossRef Kurtz SM, MacDonald D, Ianuzzi A, van Ooij A, Isaza J, Ross ER, Regan J (2009) The natural history of polyethylene oxidation in total disc replacement. Spine (Phila Pa 1976) 34:2369–2377CrossRef
23.
Zurück zum Zitat Costa L, Bracco P, Brach del Prever EM, Kurtz SM, Gallinaro P (2006) Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components. J Biomed Mater Res B Appl Biomater 78:20–26PubMed Costa L, Bracco P, Brach del Prever EM, Kurtz SM, Gallinaro P (2006) Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components. J Biomed Mater Res B Appl Biomater 78:20–26PubMed
24.
Zurück zum Zitat Kurtz SM (2008) Wear in highly crosslinked polyethylenes. Curr Orthop 22:392–399CrossRef Kurtz SM (2008) Wear in highly crosslinked polyethylenes. Curr Orthop 22:392–399CrossRef
25.
Zurück zum Zitat Minoda Y, Kobayashi A, Iwaki H, Miyaguchi M, Kadoya Y, Ohashi H, Takaoka K (2004) Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties. J Biomed Mater Res B Appl Biomater 71:1–6PubMedCrossRef Minoda Y, Kobayashi A, Iwaki H, Miyaguchi M, Kadoya Y, Ohashi H, Takaoka K (2004) Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties. J Biomed Mater Res B Appl Biomater 71:1–6PubMedCrossRef
26.
Zurück zum Zitat Hirakawa K, Bauer TW, Stulberg BN, Wilde AH, Borden LS (1996) Characterization of debris adjacent to failed knee implants of 3 different designs. Clin Orthop Relat Res 331:151–158PubMedCrossRef Hirakawa K, Bauer TW, Stulberg BN, Wilde AH, Borden LS (1996) Characterization of debris adjacent to failed knee implants of 3 different designs. Clin Orthop Relat Res 331:151–158PubMedCrossRef
27.
Zurück zum Zitat Kurtz SM, Austin MS, Azzam K, Sharkey PF, MacDonald DW, Medel FJ, Hozack WJ (2010) Mechanical properties, oxidation, and clinical performance of retrieved highly cross-linked Crossfire liners after intermediate-term implantation. J Arthroplast 25(614–623):e611–e612 Kurtz SM, Austin MS, Azzam K, Sharkey PF, MacDonald DW, Medel FJ, Hozack WJ (2010) Mechanical properties, oxidation, and clinical performance of retrieved highly cross-linked Crossfire liners after intermediate-term implantation. J Arthroplast 25(614–623):e611–e612
28.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW (2007) Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am 89:2023–2029PubMedCrossRef Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW (2007) Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am 89:2023–2029PubMedCrossRef
29.
Zurück zum Zitat Nam D, Kepler CK, Nho SJ, Craig EV, Warren RF, Wright TM (2010) Observations on retrieved humeral polyethylene components from reverse total shoulder arthroplasty. J Shoulder Elb Surg 19:1003–1012CrossRef Nam D, Kepler CK, Nho SJ, Craig EV, Warren RF, Wright TM (2010) Observations on retrieved humeral polyethylene components from reverse total shoulder arthroplasty. J Shoulder Elb Surg 19:1003–1012CrossRef
30.
Zurück zum Zitat Day JS, Macdonald DW, Olsen M, Getz C, Williams GR, Kurtz SM (2011) Polyethylene wear in retrieved reverse total shoulder components. J Shoulder Elbow Surg (in press) Day JS, Macdonald DW, Olsen M, Getz C, Williams GR, Kurtz SM (2011) Polyethylene wear in retrieved reverse total shoulder components. J Shoulder Elbow Surg (in press)
31.
Zurück zum Zitat Scholz J, Woolf CJ (2007) The neuropathic pain triad: neurons, immune cells and glia. Nat Neurosci 10:1361–1368PubMedCrossRef Scholz J, Woolf CJ (2007) The neuropathic pain triad: neurons, immune cells and glia. Nat Neurosci 10:1361–1368PubMedCrossRef
32.
Zurück zum Zitat Thacker MA, Clark AK, Marchand F, McMahon SB (2007) Pathophysiology of peripheral neuropathic pain: immune cells and molecules. Anesth Analg 105:838–847PubMedCrossRef Thacker MA, Clark AK, Marchand F, McMahon SB (2007) Pathophysiology of peripheral neuropathic pain: immune cells and molecules. Anesth Analg 105:838–847PubMedCrossRef
33.
Zurück zum Zitat Siepe CJ, Zelenkov P, Sauri-Barraza JC, Szeimies U, Grubinger T, Tepass A, Stabler A, Mayer MH (2010) The fate of facet joint and adjacent level disc degeneration following total lumbar disc replacement: a prospective clinical, X-ray, and magnetic resonance imaging investigation. Spine (Phila Pa 1976) 35:1991–2003CrossRef Siepe CJ, Zelenkov P, Sauri-Barraza JC, Szeimies U, Grubinger T, Tepass A, Stabler A, Mayer MH (2010) The fate of facet joint and adjacent level disc degeneration following total lumbar disc replacement: a prospective clinical, X-ray, and magnetic resonance imaging investigation. Spine (Phila Pa 1976) 35:1991–2003CrossRef
34.
Zurück zum Zitat Schmidt H, Midderhoff S, Adkins K, Wilke HJ (2009) The effect of different design concepts in lumbar total disc arthroplasty on the range of motion, facet joint forces and instantaneous center of rotation of a L4–5 segment. Eur Spine J 18:1695–1705PubMedCrossRef Schmidt H, Midderhoff S, Adkins K, Wilke HJ (2009) The effect of different design concepts in lumbar total disc arthroplasty on the range of motion, facet joint forces and instantaneous center of rotation of a L4–5 segment. Eur Spine J 18:1695–1705PubMedCrossRef
35.
Zurück zum Zitat Siepe CJ, Korge A, Grochulla F, Mehren C, Mayer HM (2008) Analysis of post-operative pain patterns following total lumbar disc replacement: results from fluoroscopically guided spine infiltrations. Eur Spine J 17:44–56PubMedCrossRef Siepe CJ, Korge A, Grochulla F, Mehren C, Mayer HM (2008) Analysis of post-operative pain patterns following total lumbar disc replacement: results from fluoroscopically guided spine infiltrations. Eur Spine J 17:44–56PubMedCrossRef
36.
Zurück zum Zitat Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC (2003) Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model. Spine (Phila Pa 1976) 28:S110–S117CrossRef Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC (2003) Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model. Spine (Phila Pa 1976) 28:S110–S117CrossRef
37.
Zurück zum Zitat Ahn DK, Park HS, Choi DJ, Kim KS, Yang SJ (2010) Survival and prognostic analysis of adjacent segments after spinal fusion. Clin Orthop Surg 2:140–147PubMedCrossRef Ahn DK, Park HS, Choi DJ, Kim KS, Yang SJ (2010) Survival and prognostic analysis of adjacent segments after spinal fusion. Clin Orthop Surg 2:140–147PubMedCrossRef
38.
Zurück zum Zitat Lee CS, Hwang CJ, Lee SW, Ahn YJ, Kim YT, Lee DH, Lee MY (2009) Risk factors for adjacent segment disease after lumbar fusion. Eur Spine J 18:1637–1643PubMedCrossRef Lee CS, Hwang CJ, Lee SW, Ahn YJ, Kim YT, Lee DH, Lee MY (2009) Risk factors for adjacent segment disease after lumbar fusion. Eur Spine J 18:1637–1643PubMedCrossRef
39.
Zurück zum Zitat Ching AC, Birkenmaier C, Hart RA (2010) Short segment coronal plane deformity after two-level lumbar total disc replacement. Spine (Phila Pa 1976) 35:44–50CrossRef Ching AC, Birkenmaier C, Hart RA (2010) Short segment coronal plane deformity after two-level lumbar total disc replacement. Spine (Phila Pa 1976) 35:44–50CrossRef
40.
Zurück zum Zitat Siepe CJ, Mayer HM, Heinz-Leisenheimer M, Korge A (2007) Total lumbar disc replacement: different results for different levels. Spine (Phila Pa 1976) 32:782–790CrossRef Siepe CJ, Mayer HM, Heinz-Leisenheimer M, Korge A (2007) Total lumbar disc replacement: different results for different levels. Spine (Phila Pa 1976) 32:782–790CrossRef
41.
Zurück zum Zitat Huppert J, Beaurain J, Steib JP, Bernard P, Dufour T, Hovorka I, Stecken J, Dam-Hieu P, Fuentes JM, Vital JM, Vila T, Aubourg L (2011) Comparison between single- and multi-level patients: clinical and radiological outcomes 2 years after cervical disc replacement. Eur Spine J 20:1417–1426PubMedCrossRef Huppert J, Beaurain J, Steib JP, Bernard P, Dufour T, Hovorka I, Stecken J, Dam-Hieu P, Fuentes JM, Vital JM, Vila T, Aubourg L (2011) Comparison between single- and multi-level patients: clinical and radiological outcomes 2 years after cervical disc replacement. Eur Spine J 20:1417–1426PubMedCrossRef
42.
Zurück zum Zitat Gstoettner M, Heider D, Liebensteiner M, Bach CM (2008) Footprint mismatch in lumbar total disc arthroplasty. Eur Spine J 17:1470–1475PubMedCrossRef Gstoettner M, Heider D, Liebensteiner M, Bach CM (2008) Footprint mismatch in lumbar total disc arthroplasty. Eur Spine J 17:1470–1475PubMedCrossRef
43.
Zurück zum Zitat Spivak JM, Petrizzo AM (2010) Revision of a lumbar disc arthroplasty following late infection. Eur Spine J 19:677–681PubMedCrossRef Spivak JM, Petrizzo AM (2010) Revision of a lumbar disc arthroplasty following late infection. Eur Spine J 19:677–681PubMedCrossRef
44.
Zurück zum Zitat Tropiano P (2010) Expert’s comment concerning Grand Rounds case entitled “Revision of a lumbar disc arthroplasty following late infection” (by Jeffrey M. Spivak and Anthony M. Petrizzo). Eur Spine J 19:682–684PubMedCrossRef Tropiano P (2010) Expert’s comment concerning Grand Rounds case entitled “Revision of a lumbar disc arthroplasty following late infection” (by Jeffrey M. Spivak and Anthony M. Petrizzo). Eur Spine J 19:682–684PubMedCrossRef
45.
Zurück zum Zitat Patel AA, Brodke DS, Pimenta L, Bono CM, Hilibrand AS, Harrop JS, Riew KD, Youssef JA, Vaccaro AR (2008) Revision strategies in lumbar total disc arthroplasty. Spine (Phila Pa 1976) 33:1276–1283CrossRef Patel AA, Brodke DS, Pimenta L, Bono CM, Hilibrand AS, Harrop JS, Riew KD, Youssef JA, Vaccaro AR (2008) Revision strategies in lumbar total disc arthroplasty. Spine (Phila Pa 1976) 33:1276–1283CrossRef
46.
Zurück zum Zitat Punt IM, Visser VM, van Rhijn LW, Kurtz SM, Antonis J, Schurink GW, van Ooij A (2008) Complications and reoperations of the SB Charite lumbar disc prosthesis: experience in 75 patients. Eur Spine J 17:36–43PubMedCrossRef Punt IM, Visser VM, van Rhijn LW, Kurtz SM, Antonis J, Schurink GW, van Ooij A (2008) Complications and reoperations of the SB Charite lumbar disc prosthesis: experience in 75 patients. Eur Spine J 17:36–43PubMedCrossRef
Metadaten
Titel
Clinical, radiological, histological and retrieval findings of Activ-L and Mobidisc total disc replacements: a study of two patients
verfasst von
Shennah Austen
Ilona M. Punt
Jack P. M. Cleutjens
Paul C. Willems
Steven M. Kurtz
Daniel W. MacDonald
Lodewijk W. van Rhijn
André van Ooij
Publikationsdatum
01.06.2012
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe Sonderheft 4/2012
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-011-2141-7

Weitere Artikel der Sonderheft 4/2012

European Spine Journal 4/2012 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Lever-Sign-Test hilft beim Verdacht auf Kreuzbandriss

15.05.2024 Vordere Kreuzbandruptur Nachrichten

Mit dem Hebelzeichen-Test lässt sich offenbar recht zuverlässig feststellen, ob ein vorderes Kreuzband gerissen ist. In einer Metaanalyse war die Vorhersagekraft vor allem bei positivem Testergebnis hoch.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.