Comparing different planimetric methods on volumetric estimations by using cone beam computed tomography

verfasst von: Alaettin Koç, Ömer Said Sezgin, Saadettin Kayıpmaz

Erschienen in: La radiologia medica | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this study was to compare the accuracy of the planimetric methods on volume estimations by using cone beam computed tomography (CBCT).

Materials and methods

Thirty-one prepared intraosseous bone defects from thirteen bovine femur condyles were scanned with CBCT. The defect volumes were estimated by point counting (PC), manual segmentation (MS) and semiautomatic segmentation (SAS) methods at 0.3-mm section thickness without any intersection gap. The estimated volumes were compared with the results of the Archimedes’ method. The planimetric methods were analyzed using a Friedman’s two-way analysis of variance test.

Results

The estimated volumes of MS and SAS methods were compatible with the volumes of Archimedes’ method (p = 0.768, p = 0.140, respectively), but the volumes from the PC method were not compatible with Archimedes’ method (p < 0.001).

Conclusion

SAS was approximately 2.5 times faster than MS. Both MS and SAS are valid methods for volume estimation; however, SAS may be preferred due to its practicability.
Literatur
1.
Zurück zum Zitat Scarfe WC, Farman AG (2008) What is cone-beam CT and how does it work? Dent Clin N Am 52:707–730CrossRef Scarfe WC, Farman AG (2008) What is cone-beam CT and how does it work? Dent Clin N Am 52:707–730CrossRef
2.
Zurück zum Zitat Oh SH, Kang JH, Seo YK, Lee SR, Choi HY, Choi YS et al (2018) Linear accuracy of cone-beam computed tomography and a 3-dimensional facial scanning system: an anthropomorphic phantom study. Imaging Sci Dent 48:111–119CrossRef Oh SH, Kang JH, Seo YK, Lee SR, Choi HY, Choi YS et al (2018) Linear accuracy of cone-beam computed tomography and a 3-dimensional facial scanning system: an anthropomorphic phantom study. Imaging Sci Dent 48:111–119CrossRef
3.
Zurück zum Zitat Koç A (2018) Comparison of the horizontal condyle angle of the dentulous and edentulous patients using cone beam computed tomography. East J Med 23:254–257CrossRef Koç A (2018) Comparison of the horizontal condyle angle of the dentulous and edentulous patients using cone beam computed tomography. East J Med 23:254–257CrossRef
4.
Zurück zum Zitat Oberoi S, Chigurupati R, Gill P, Hoffman WY, Vargervik K (2009) Volumetric assessment of secondary alveolar bone grafting using cone beam computed tomography. Cleft Palate Craniofac J 46:503–511CrossRef Oberoi S, Chigurupati R, Gill P, Hoffman WY, Vargervik K (2009) Volumetric assessment of secondary alveolar bone grafting using cone beam computed tomography. Cleft Palate Craniofac J 46:503–511CrossRef
5.
Zurück zum Zitat Sezgin OS, Kayıpmaz S, Sahin B (2013) The effect of slice thickness on the assessment of bone defect volumes by the Cavalieri principle using cone beam computed tomography. J Digit Imaging 26:115–118CrossRef Sezgin OS, Kayıpmaz S, Sahin B (2013) The effect of slice thickness on the assessment of bone defect volumes by the Cavalieri principle using cone beam computed tomography. J Digit Imaging 26:115–118CrossRef
6.
Zurück zum Zitat Kayipmaz S, Sezgin OS, Saricaoğlu ST, Bas O, Sahin B, Küçük M (2011) The estimation of the volume of sheep mandibular defects using cone-beam computed tomography images and a stereological method. Dentomaxillofac Radiol 40:165–169CrossRef Kayipmaz S, Sezgin OS, Saricaoğlu ST, Bas O, Sahin B, Küçük M (2011) The estimation of the volume of sheep mandibular defects using cone-beam computed tomography images and a stereological method. Dentomaxillofac Radiol 40:165–169CrossRef
7.
Zurück zum Zitat Bayram M, Kayipmaz S, Sezgin OS, Küçük M (2012) Volumetric analysis of the mandibular condyle using cone beam computed tomography. Eur J Radiol 81:1812–1816CrossRef Bayram M, Kayipmaz S, Sezgin OS, Küçük M (2012) Volumetric analysis of the mandibular condyle using cone beam computed tomography. Eur J Radiol 81:1812–1816CrossRef
8.
Zurück zum Zitat Agbaje JO, Jacobs R, Maes F, Michiels K, Van Steenberghe D (2007) Volumetric analysis of extraction sockets using cone beam computed tomography: a pilot study on ex vivo jaw bone. J Clin Periodontol 34:985–990CrossRef Agbaje JO, Jacobs R, Maes F, Michiels K, Van Steenberghe D (2007) Volumetric analysis of extraction sockets using cone beam computed tomography: a pilot study on ex vivo jaw bone. J Clin Periodontol 34:985–990CrossRef
9.
Zurück zum Zitat Clatterbuck RE, Sipos EP (1997) The efficient calculation of neurosurgically relevant volumes from computed tomographic scans using Cavalieri’s direct estimator. Neurosurgery 40:339–343CrossRef Clatterbuck RE, Sipos EP (1997) The efficient calculation of neurosurgically relevant volumes from computed tomographic scans using Cavalieri’s direct estimator. Neurosurgery 40:339–343CrossRef
10.
Zurück zum Zitat Mazonakis M, Damilakis J, Varveris H (1998) Bladder and rectum volume estimations using CT and stereology. Comput Med Imaging Graph 22:195–201CrossRef Mazonakis M, Damilakis J, Varveris H (1998) Bladder and rectum volume estimations using CT and stereology. Comput Med Imaging Graph 22:195–201CrossRef
11.
Zurück zum Zitat Sahin B, Ergur H (2006) Assessment of the optimum section thickness for the estimation of liver volume using magnetic resonance images: a stereological gold standard study. Eur J Radiol 57:96–101CrossRef Sahin B, Ergur H (2006) Assessment of the optimum section thickness for the estimation of liver volume using magnetic resonance images: a stereological gold standard study. Eur J Radiol 57:96–101CrossRef
12.
Zurück zum Zitat Gong QY, Tan LT, Romaniuk CS, Jones B, Brunt JN, Roberts N (1999) Determination of tumour regression rates during radiotherapy for cervical carcinoma by serial MRI: comparison of two measurement techniques and examination of intraobserver and interobserver variability. Br J Radiol 72:62–72CrossRef Gong QY, Tan LT, Romaniuk CS, Jones B, Brunt JN, Roberts N (1999) Determination of tumour regression rates during radiotherapy for cervical carcinoma by serial MRI: comparison of two measurement techniques and examination of intraobserver and interobserver variability. Br J Radiol 72:62–72CrossRef
13.
Zurück zum Zitat Kong Z, Li T, Luo J, Xu S (2019) Automatic tissue image segmentation based on image processing and deep learning. J Healthc Eng 2019:2912458CrossRef Kong Z, Li T, Luo J, Xu S (2019) Automatic tissue image segmentation based on image processing and deep learning. J Healthc Eng 2019:2912458CrossRef
14.
Zurück zum Zitat Chu C-C, Aggarwal JK (1993) The integration of image segmentation maps using region and edge information. IEEE Trans Pattern Anal Mach Intell 15:1241–1252CrossRef Chu C-C, Aggarwal JK (1993) The integration of image segmentation maps using region and edge information. IEEE Trans Pattern Anal Mach Intell 15:1241–1252CrossRef
15.
Zurück zum Zitat Sahin B, Emirzeoglu M, Uzun A, Incesu L, Bek Y, Bilgic S et al (2003) Unbiased estimation of the liver volume by the Cavalieri principle using magnetic resonance images. Eur J Radiol 47:164–170CrossRef Sahin B, Emirzeoglu M, Uzun A, Incesu L, Bek Y, Bilgic S et al (2003) Unbiased estimation of the liver volume by the Cavalieri principle using magnetic resonance images. Eur J Radiol 47:164–170CrossRef
16.
Zurück zum Zitat De Faria Vasconcelos K, Evangelista KM, Rodrigues C, Estrela C, De Sousa TO, Silva MA (2012) Detection of periodontal bone loss using cone beam CT and intraoral radiography. Dentomaxillofac Radiol 41:64–69CrossRef De Faria Vasconcelos K, Evangelista KM, Rodrigues C, Estrela C, De Sousa TO, Silva MA (2012) Detection of periodontal bone loss using cone beam CT and intraoral radiography. Dentomaxillofac Radiol 41:64–69CrossRef
17.
Zurück zum Zitat Koç A, Kavut İ, Uğur M (2019) Assessment of buccal bone thickness in the anterior maxilla: a cone beam computed tomography study. Cumhur Dent J 22:102–107CrossRef Koç A, Kavut İ, Uğur M (2019) Assessment of buccal bone thickness in the anterior maxilla: a cone beam computed tomography study. Cumhur Dent J 22:102–107CrossRef
18.
Zurück zum Zitat Acer N, Sahin B, Usanmaz M, Tatoğlu H, Irmak Z (2008) Comparison of point counting and planimetry methods for the assessment of cerebellar volume in human using magnetic resonance imaging: a stereological study. Surg Radiol Anat 30:335–339CrossRef Acer N, Sahin B, Usanmaz M, Tatoğlu H, Irmak Z (2008) Comparison of point counting and planimetry methods for the assessment of cerebellar volume in human using magnetic resonance imaging: a stereological study. Surg Radiol Anat 30:335–339CrossRef
19.
Zurück zum Zitat Rana M, Modrow D, Keuchel J, Chui C, Rana M, Wagner M et al (2015) Development and evaluation of an automatic tumor segmentation tool: a comparison between automatic, semi-automatic and manual segmentation of mandibular odontogenic cysts and tumors. J Craniomaxillofac Surg 43:355–359CrossRef Rana M, Modrow D, Keuchel J, Chui C, Rana M, Wagner M et al (2015) Development and evaluation of an automatic tumor segmentation tool: a comparison between automatic, semi-automatic and manual segmentation of mandibular odontogenic cysts and tumors. J Craniomaxillofac Surg 43:355–359CrossRef
20.
Zurück zum Zitat Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A et al (2013) Variability of dental cone beam CT grey values for density estimations. Br J Radiol 86:20120135CrossRef Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A et al (2013) Variability of dental cone beam CT grey values for density estimations. Br J Radiol 86:20120135CrossRef
21.
Zurück zum Zitat Pauwels R, Jacobs R, Singer SR, Mupparapu M (2015) CBCT-based bone quality assessment: are Hounsfield units applicable? Dentomaxillofac Radiol 44:20140238CrossRef Pauwels R, Jacobs R, Singer SR, Mupparapu M (2015) CBCT-based bone quality assessment: are Hounsfield units applicable? Dentomaxillofac Radiol 44:20140238CrossRef
22.
Zurück zum Zitat Bastami F, Shahab S, Parsa A, Abbas FM, Noori Kooshki MH, Namdari M et al (2018) Can gray values derived from CT and cone beam CT estimate new bone formation? An in vivo study. Oral Maxillofac Surg 22:13–20CrossRef Bastami F, Shahab S, Parsa A, Abbas FM, Noori Kooshki MH, Namdari M et al (2018) Can gray values derived from CT and cone beam CT estimate new bone formation? An in vivo study. Oral Maxillofac Surg 22:13–20CrossRef
23.
Zurück zum Zitat Albuquerque MA, Gaia BF, Cavalcanti MG (2011) Comparison between multislice and cone-beam computerized tomography in the volumetric assessment of cleft palate. Oral Surg Oral Med Oral Pathol Oral Radiol Endodontol 112:249–257CrossRef Albuquerque MA, Gaia BF, Cavalcanti MG (2011) Comparison between multislice and cone-beam computerized tomography in the volumetric assessment of cleft palate. Oral Surg Oral Med Oral Pathol Oral Radiol Endodontol 112:249–257CrossRef
24.
Zurück zum Zitat Ahlowalia MS, Patel S, Anwar HM, Cama G, Austin RS, Wilson R et al (2013) Accuracy of CBCT for volumetric measurement of simulated periapical lesions. Int Endod J 46:538–546CrossRef Ahlowalia MS, Patel S, Anwar HM, Cama G, Austin RS, Wilson R et al (2013) Accuracy of CBCT for volumetric measurement of simulated periapical lesions. Int Endod J 46:538–546CrossRef
25.
Zurück zum Zitat Forst D, Nijjar S, Flores-Mir C, Carey J, Secanell M, Lagravere M (2014) Comparison of in vivo 3D cone-beam computed tomography tooth volume measurement protocols. Prog Orthod 15:69CrossRef Forst D, Nijjar S, Flores-Mir C, Carey J, Secanell M, Lagravere M (2014) Comparison of in vivo 3D cone-beam computed tomography tooth volume measurement protocols. Prog Orthod 15:69CrossRef
26.
Zurück zum Zitat Xi T, Schreurs R, Heerink WJ, Bergé SJ, Maal TJ (2014) A novel region-growing based semi-automatic segmentation protocol for three-dimensional condylar reconstruction using cone beam computed tomography (CBCT). PLoS ONE 9:e111126CrossRef Xi T, Schreurs R, Heerink WJ, Bergé SJ, Maal TJ (2014) A novel region-growing based semi-automatic segmentation protocol for three-dimensional condylar reconstruction using cone beam computed tomography (CBCT). PLoS ONE 9:e111126CrossRef
27.
Zurück zum Zitat Acer N, Ilıca AT, Turgut AT, Özçelik Ö, Yıldırım B, Turgut M (2012) Comparison of three methods for the estimation of pineal gland volume using magnetic resonance imaging. Sci World J 2012:123412CrossRef Acer N, Ilıca AT, Turgut AT, Özçelik Ö, Yıldırım B, Turgut M (2012) Comparison of three methods for the estimation of pineal gland volume using magnetic resonance imaging. Sci World J 2012:123412CrossRef
28.
Zurück zum Zitat Bolat D, Bahar S, Tipirdamaz S, Selcuk M (2013) Comparison of the morphometric features of the left and right horse kidneys: a stereological approach. Anat Histol Embryol 42:448–452CrossRef Bolat D, Bahar S, Tipirdamaz S, Selcuk M (2013) Comparison of the morphometric features of the left and right horse kidneys: a stereological approach. Anat Histol Embryol 42:448–452CrossRef
29.
Zurück zum Zitat Sahin B, Acer N, Sonmez O, Emirzeoglu M, Basaloglu H, Uzun A et al (2007) Comparison of four methods for the estimation of intracranial volume: a gold standard study. Clin Anat 20:766–773CrossRef Sahin B, Acer N, Sonmez O, Emirzeoglu M, Basaloglu H, Uzun A et al (2007) Comparison of four methods for the estimation of intracranial volume: a gold standard study. Clin Anat 20:766–773CrossRef
30.
Zurück zum Zitat Schulze R, Heil U, Groβ D, Bruellmann D, Dranischnikow E, Schwanecke U et al (2011) Artefacts in CBCT: a review. Dentomaxillofac Radiol 40:265–273CrossRef Schulze R, Heil U, Groβ D, Bruellmann D, Dranischnikow E, Schwanecke U et al (2011) Artefacts in CBCT: a review. Dentomaxillofac Radiol 40:265–273CrossRef
Metadaten
Titel
Comparing different planimetric methods on volumetric estimations by using cone beam computed tomography
verfasst von
Alaettin Koç
Ömer Said Sezgin
Saadettin Kayıpmaz
Publikationsdatum
01.04.2020
Verlag
Springer Milan
Erschienen in
La radiologia medica / Ausgabe 4/2020
Print ISSN: 0033-8362
Elektronische ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-019-01131-8

Weitere Artikel der Ausgabe 4/2020

La radiologia medica 4/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.