Comparison of Reconstruction Algorithms for Decreasing the Exposure Dose During Digital Tomosynthesis for Arthroplasty: a Phantom Study

verfasst von: Tsutomu Gomi, Rina Sakai, Masami Goto, Yusuke Watanabe, Tohoru Takeda, Tokuo Umeda

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 4/2016

Einloggen, um Zugang zu erhalten

Abstract

To explore the possibility of decreasing the radiation dose during digital tomosynthesis (DT) for arthroplasty, we compared the image qualities of several reconstruction algorithms, such as filtered back projection (FBP) and two iterative reconstruction (IR), methods maximum likelihood expectation maximization (MLEM) and the simultaneous iterative reconstruction technique (SIRT) under different radiation doses. The three algorithms were implemented using a DT system and experimentally evaluated by contrast-to-noise ratio (CNR), artifact spread function (ASF), and power spectrum measurements on a prosthesis phantom. The CNR and ASF data were statistically analyzed by a one-way analysis of variance. The effectiveness of each technique for enhancing the visibility of the prosthesis phantom was quantified by the CNR (reference dose vs. 20 % reduced dose in FBP, P = 0.62; reference vs. 37 % reduced dose in FBP, P = 0.16; reference vs. 55 % reduced dose in FBP, P < 0.05; reference vs. 20 % reduced dose in IR, P = 0.92; reference vs. 37 % reduced dose in IR, P = 0.40; reference vs. 55 % reduced dose in IR, P < 0.05) and ASF (reference dose vs. 20 % reduced dose in FBP, P = 0.25; reference vs. 37 and 55 % reduced dose in FBP, P < 0.05; reference vs. 20 % reduced dose in IR, P = 0.16; reference vs. 37 and 55 % reduced dose in IR, P < 0.05). The power spectra under the reference and reduced doses are equivalent. In this phantom study, the radiation dose of the reference dose could be decreased by 20 % with FBP and IR for consideration of common factors.
Literatur
1.
Zurück zum Zitat Ziedses des Plante BG: Eine neue methode zur differenzierung in der roentgenographie (planigraphie). Acta Radiol 13:182–192, 1932CrossRef Ziedses des Plante BG: Eine neue methode zur differenzierung in der roentgenographie (planigraphie). Acta Radiol 13:182–192, 1932CrossRef
2.
Zurück zum Zitat Miller ER, McCurry EM, Hruska B: An infinite number of laminagrams from a finite number of radiographs. Radiology 98:249–255, 1971CrossRefPubMed Miller ER, McCurry EM, Hruska B: An infinite number of laminagrams from a finite number of radiographs. Radiology 98:249–255, 1971CrossRefPubMed
3.
Zurück zum Zitat Grant DG: Tomosynthesis. A three-dimensional radiographic imaging technique. IEEE Trans Biomed Eng 19:20–28, 1972CrossRefPubMed Grant DG: Tomosynthesis. A three-dimensional radiographic imaging technique. IEEE Trans Biomed Eng 19:20–28, 1972CrossRefPubMed
4.
5.
Zurück zum Zitat Kruger RA, Nelson JA, Ghosh-Roy D, Miller FJ, Anderson RE, Liu PY: Dynamic tomographic digital subtraction angiography using temporal filteration. Radiology 147:863–867, 1983CrossRefPubMed Kruger RA, Nelson JA, Ghosh-Roy D, Miller FJ, Anderson RE, Liu PY: Dynamic tomographic digital subtraction angiography using temporal filteration. Radiology 147:863–867, 1983CrossRefPubMed
6.
Zurück zum Zitat Sone S, Kasuga T, Sakai S, Aoki J, Izuno I, Tanizaki Y, Shigeta H, Shibata K: Development of a high-resolution digital tomosynthesis system and its clinical application. Radiographics 11:807–822, 1991CrossRefPubMed Sone S, Kasuga T, Sakai S, Aoki J, Izuno I, Tanizaki Y, Shigeta H, Shibata K: Development of a high-resolution digital tomosynthesis system and its clinical application. Radiographics 11:807–822, 1991CrossRefPubMed
7.
Zurück zum Zitat Sone S, Kasuga T, Sakai F, Kawai T, Oguchi K, Hirano H, Li F, Kubo K, Honda T, Hniuda M, Takemura K, Hosoba M: Image processing in the digital tomosynthesis for pulmonary imaging. Eur Radiol 5:96–101, 1995CrossRef Sone S, Kasuga T, Sakai F, Kawai T, Oguchi K, Hirano H, Li F, Kubo K, Honda T, Hniuda M, Takemura K, Hosoba M: Image processing in the digital tomosynthesis for pulmonary imaging. Eur Radiol 5:96–101, 1995CrossRef
8.
Zurück zum Zitat Machida H, Yuhara T, Mori T, Ueno E, Moribe Y, Sabol JM: Optimizing parameters for flat-panel detector digital tomosynthesis. Radiographics 30:549–562, 2010CrossRefPubMed Machida H, Yuhara T, Mori T, Ueno E, Moribe Y, Sabol JM: Optimizing parameters for flat-panel detector digital tomosynthesis. Radiographics 30:549–562, 2010CrossRefPubMed
9.
Zurück zum Zitat Vikgren J, Zachrisson S, Svalkvist A, Johnsson AA, Boijsen M, Flinck A, Kheddache S, Båth M: Comparison of chest tomosynthesis and chest radiography for detection of pulmonary nodules: human observer study of clinical cases. Radiology 249:1034–1041, 2008CrossRefPubMed Vikgren J, Zachrisson S, Svalkvist A, Johnsson AA, Boijsen M, Flinck A, Kheddache S, Båth M: Comparison of chest tomosynthesis and chest radiography for detection of pulmonary nodules: human observer study of clinical cases. Radiology 249:1034–1041, 2008CrossRefPubMed
10.
Zurück zum Zitat Skaane P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U, Izadi M, Jebsen IN, Jahr G, Krager M, Niklason LT, Hofvind S, Gur D: Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56, 2013CrossRefPubMed Skaane P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U, Izadi M, Jebsen IN, Jahr G, Krager M, Niklason LT, Hofvind S, Gur D: Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56, 2013CrossRefPubMed
11.
Zurück zum Zitat Stiel G, Stiel LG, Klotz E, Nienaber CA: Digital flashing tomosynthesis: a promising technique for angiographic screening. IEEE Trans Med Imaging 12:314–321, 1993CrossRefPubMed Stiel G, Stiel LG, Klotz E, Nienaber CA: Digital flashing tomosynthesis: a promising technique for angiographic screening. IEEE Trans Med Imaging 12:314–321, 1993CrossRefPubMed
12.
Zurück zum Zitat Duryea J, Dobbins JT, Lynch JA: Digital tomosynthesis of hand joints for arthritis assessment. Med Phys 30:325–33, 2003CrossRefPubMed Duryea J, Dobbins JT, Lynch JA: Digital tomosynthesis of hand joints for arthritis assessment. Med Phys 30:325–33, 2003CrossRefPubMed
13.
Zurück zum Zitat Niklason LT, Christian BT, Niklason LE, Kopans DB, Castleberry DE, Opsahl-Ong BH, Landberg CE, Slanetz PJ, Giardino AA, Moore R, Albagli D, DeJule MC, Fitzgerald PF, Fobare DF, Giambattista BW, Kwasnick RF, Liu J, Lubowski SJ, Possin GE, Richotte JF, Wei CY, Wirth RF: Digital tomosynthesis in breast imaging. Radiology 205:399–406, 1994CrossRef Niklason LT, Christian BT, Niklason LE, Kopans DB, Castleberry DE, Opsahl-Ong BH, Landberg CE, Slanetz PJ, Giardino AA, Moore R, Albagli D, DeJule MC, Fitzgerald PF, Fobare DF, Giambattista BW, Kwasnick RF, Liu J, Lubowski SJ, Possin GE, Richotte JF, Wei CY, Wirth RF: Digital tomosynthesis in breast imaging. Radiology 205:399–406, 1994CrossRef
14.
Zurück zum Zitat Dobbins III, JT, Godfrey DJ: Digital X-ray tomosynthesis: current state of the art and clinical potential. Phys Med Biol 48:R65–106, 2003CrossRefPubMed Dobbins III, JT, Godfrey DJ: Digital X-ray tomosynthesis: current state of the art and clinical potential. Phys Med Biol 48:R65–106, 2003CrossRefPubMed
15.
Zurück zum Zitat Gomi T, Hirano H: Clinical potential of digital linear tomosynthesis imaging of total joint arthroplasty. J Digit Imaging 21:312–322, 2008CrossRefPubMed Gomi T, Hirano H: Clinical potential of digital linear tomosynthesis imaging of total joint arthroplasty. J Digit Imaging 21:312–322, 2008CrossRefPubMed
16.
Zurück zum Zitat Gomi T, Hirano H, Umeda T: Evaluation of the X-ray digital linear tomosynthesis reconstruction processing method for metal artifact reduction. Comput Med Imaging Graph 33:257–274, 2009CrossRef Gomi T, Hirano H, Umeda T: Evaluation of the X-ray digital linear tomosynthesis reconstruction processing method for metal artifact reduction. Comput Med Imaging Graph 33:257–274, 2009CrossRef
17.
Zurück zum Zitat White LM, Buckwalder KA: Technical considerations: CT and MR imaging in the postoperative orthopaedic patient. Semin Musculoskelet Radiol 6:5–17, 2002CrossRefPubMed White LM, Buckwalder KA: Technical considerations: CT and MR imaging in the postoperative orthopaedic patient. Semin Musculoskelet Radiol 6:5–17, 2002CrossRefPubMed
18.
Zurück zum Zitat Gomi T: Comparison of metal artifact in digital tomosynthesis and computed tomography for evaluation of phantoms. J Biomed Sci Eng 6:722–731, 2013CrossRef Gomi T: Comparison of metal artifact in digital tomosynthesis and computed tomography for evaluation of phantoms. J Biomed Sci Eng 6:722–731, 2013CrossRef
19.
Zurück zum Zitat Gordon R, Bender R, Hermen GT: Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and X-ray photography. J Theor Biol 29:471–481, 1970CrossRefPubMed Gordon R, Bender R, Hermen GT: Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and X-ray photography. J Theor Biol 29:471–481, 1970CrossRefPubMed
20.
Zurück zum Zitat Wu T, Stewart A, Stanton M, McCauley T, Phillips W, Kopans DB, et al: Tomographic mammography using a limited number of low-dose cone-beam projection images. Med Phys 30:365–380, 2003CrossRefPubMed Wu T, Stewart A, Stanton M, McCauley T, Phillips W, Kopans DB, et al: Tomographic mammography using a limited number of low-dose cone-beam projection images. Med Phys 30:365–380, 2003CrossRefPubMed
21.
Zurück zum Zitat Tapiovaara M, Siiskonen T. A Monte Carlo program for calculating patient doses in medical X-ray examinations, 2nd edition. Report No. STUK-A231 STUK 2008; Helsinki, Finland Tapiovaara M, Siiskonen T. A Monte Carlo program for calculating patient doses in medical X-ray examinations, 2nd edition. Report No. STUK-A231 STUK 2008; Helsinki, Finland
22.
Zurück zum Zitat Bleuet P, Guillemaud R, Magnin I, Desbat L: An adapted fan volume sampling scheme for 3D algebraic reconstruction in linear tomosynthesis. IEEE Trans Nucl Sci 3:1720–1724, 2001 Bleuet P, Guillemaud R, Magnin I, Desbat L: An adapted fan volume sampling scheme for 3D algebraic reconstruction in linear tomosynthesis. IEEE Trans Nucl Sci 3:1720–1724, 2001
24.
Zurück zum Zitat Rangayyan RM: Biomedical Image Analysis. CRC Press, Florida, 2000, pp 612–621 Rangayyan RM: Biomedical Image Analysis. CRC Press, Florida, 2000, pp 612–621
Metadaten
Titel
Comparison of Reconstruction Algorithms for Decreasing the Exposure Dose During Digital Tomosynthesis for Arthroplasty: a Phantom Study
verfasst von
Tsutomu Gomi
Rina Sakai
Masami Goto
Yusuke Watanabe
Tohoru Takeda
Tokuo Umeda
Publikationsdatum
04.03.2016
Verlag
Springer International Publishing
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 4/2016
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-016-9876-y

Weitere Artikel der Ausgabe 4/2016

Journal of Digital Imaging 4/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.