Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery

verfasst von: Sabong Srivannaboon, Chareenun Chirapapaisan

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 12/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the refractive outcomes following cataract surgery using conventional keratometry (K) and total keratometry (TK) for intraocular lens (IOL) calculation in the SRK/T, HofferQ, Haigis, and Holladay 1 and 2, as well as Barrett and Barrett TK Universal II formulas.

Methods

Sixty eyes of 60 patients from Siriraj Hospital, Thailand, were prospectively enrolled in this comparative study. Eyes were assessed using a swept-source optical biometer (IOLMaster 700; Carl Zeiss Meditec, Jena, Germany). Posterior keratometry, K, TK, central corneal thickness, anterior chamber depth, lens thickness, axial length, and white-to-white corneal diameter were recorded. Emmetropic IOL power was calculated using K and TK in all formulas. Selected IOL power and predicted refractive outcomes were recorded. Postoperative manifest refraction was measured 3 months postoperatively. Mean absolute errors (MAEs), median absolute errors (MedAEs), and percentage of eyes within ± 0.25, ± 0.50, and ± 1.00 D of predicted refraction were calculated for all formulas in both groups.

Results

Mean difference between K and TK was 0.03 D (44.56 ± 1.18 vs. 44.59 ± 1.22 D), showing excellent agreement (ICC = 0.99, all p < 0.001). Emmetropic IOL powers in all formulas for both groups were very similar, with a trend toward lower MAEs and MedAEs for TK when compared with K. The Barrett TK Universal II formula demonstrated the lowest MAEs. Proportion of eyes within ± 0.25, ± 0.50, and ± 1.00 D of predicted refraction were slightly higher in the TK group.

Conclusions

Conventional K and TK for IOL calculation showed strong agreement with a trend toward better refractive outcomes using TK. The same IOL constant can be used for both K and TK.
Literatur
1.
Zurück zum Zitat Sano M, Hiraoka T, Ueno Y (2016) Influence of posterior corneal astigmatism on postoperative refractive astigmatism in pseudophakic eyes after cataract surgery. BMC Ophthalmol 16:212CrossRef Sano M, Hiraoka T, Ueno Y (2016) Influence of posterior corneal astigmatism on postoperative refractive astigmatism in pseudophakic eyes after cataract surgery. BMC Ophthalmol 16:212CrossRef
2.
Zurück zum Zitat Reitblat O, Levy A, Kleinmann G, Abulafia A, Assia EI (2016) Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: comparison of methodologies. J Cataract Refract Surg 42:217–225CrossRef Reitblat O, Levy A, Kleinmann G, Abulafia A, Assia EI (2016) Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: comparison of methodologies. J Cataract Refract Surg 42:217–225CrossRef
3.
Zurück zum Zitat Savini G, Næser K (2015) An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with toric intraocular lenses. Invest Ophthalmol Vis Sci 13:827–835CrossRef Savini G, Næser K (2015) An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with toric intraocular lenses. Invest Ophthalmol Vis Sci 13:827–835CrossRef
4.
Zurück zum Zitat Savini G, Hoffer KJ, Lomoriello DS, Ducoli P (2017) Simulated keratometry versus total corneal power by ray tracing: a comparison in prediction accuracy of intraocular lens power. Cornea 36:1368–1372CrossRef Savini G, Hoffer KJ, Lomoriello DS, Ducoli P (2017) Simulated keratometry versus total corneal power by ray tracing: a comparison in prediction accuracy of intraocular lens power. Cornea 36:1368–1372CrossRef
5.
Zurück zum Zitat Kirgiz A, Atalay K, Kaldirim H, Cabuk KS, Akdemir MO, Taskapili M (2017) Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation. Int Ophthalmol 37:781–786CrossRef Kirgiz A, Atalay K, Kaldirim H, Cabuk KS, Akdemir MO, Taskapili M (2017) Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation. Int Ophthalmol 37:781–786CrossRef
6.
Zurück zum Zitat Saad E, Shammas MC, Shammas HJ (2013) Scheimpflug corneal power measurements for intraocular lens power calculation in cataract surgery. Am J Ophthalmol 156:460–467CrossRef Saad E, Shammas MC, Shammas HJ (2013) Scheimpflug corneal power measurements for intraocular lens power calculation in cataract surgery. Am J Ophthalmol 156:460–467CrossRef
7.
Zurück zum Zitat Srivannaboon S, Chotikavanich S, Chirapapaisan C, Kasemson S, Po-ngam W (2012) Precision analysis of posterior corneal topography measured by Visante Omni: repeatability, reproducibility, and agreement with Orbscan II. J Refract Surg 28:133–138CrossRef Srivannaboon S, Chotikavanich S, Chirapapaisan C, Kasemson S, Po-ngam W (2012) Precision analysis of posterior corneal topography measured by Visante Omni: repeatability, reproducibility, and agreement with Orbscan II. J Refract Surg 28:133–138CrossRef
8.
Zurück zum Zitat Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, Romay A, Vigara I (2012) Dual versus single Scheimpflug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg 38:1934–1949CrossRef Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, Romay A, Vigara I (2012) Dual versus single Scheimpflug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg 38:1934–1949CrossRef
10.
Zurück zum Zitat Rosner B (2000) Fundamental of biostatistic. Duxbury, California Rosner B (2000) Fundamental of biostatistic. Duxbury, California
11.
Zurück zum Zitat Shajari M, Kolb CM, Petermann K, Böhm M, Herzog M, de’Lorenzo N, Schönbrunn S, Kohnen T (2018) Comparison of 9 modern intraocular lens power calculation formulas for a quadrifocal intraocular lens. J Cataract Refract Surg 44:942–948CrossRef Shajari M, Kolb CM, Petermann K, Böhm M, Herzog M, de’Lorenzo N, Schönbrunn S, Kohnen T (2018) Comparison of 9 modern intraocular lens power calculation formulas for a quadrifocal intraocular lens. J Cataract Refract Surg 44:942–948CrossRef
12.
Zurück zum Zitat Fabian E, Wehner W (2019) Prediction accuracy of total keratometry compared to standard keratometry using different intraocular lens power formulas. J Refract Surg 35:362–368CrossRef Fabian E, Wehner W (2019) Prediction accuracy of total keratometry compared to standard keratometry using different intraocular lens power formulas. J Refract Surg 35:362–368CrossRef
13.
Zurück zum Zitat Wang L, Koch DD, Hill W, Abulafia A (2017) Pursuing perfection in intraocular lens calculations: III. Criteria for analyzing outcomes. J Cataract Refract Surg 43:999–1002CrossRef Wang L, Koch DD, Hill W, Abulafia A (2017) Pursuing perfection in intraocular lens calculations: III. Criteria for analyzing outcomes. J Cataract Refract Surg 43:999–1002CrossRef
14.
Zurück zum Zitat Aristodemou P, Cartwright NK, Sparrow JM, Johnston R (2011) Intraocular lens calculations. Ophthalmology 118:1221CrossRef Aristodemou P, Cartwright NK, Sparrow JM, Johnston R (2011) Intraocular lens calculations. Ophthalmology 118:1221CrossRef
15.
Zurück zum Zitat Lachin JM (1992) Power and sample size evaluation for the McNemar test with application to matched case-control studies. Stat Med 11:1239–1251CrossRef Lachin JM (1992) Power and sample size evaluation for the McNemar test with application to matched case-control studies. Stat Med 11:1239–1251CrossRef
16.
Zurück zum Zitat Tang M, Wang L, Koch DD, Li Y, Huang D (2012) Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography. J Cataract Refract Surg 238:589–594CrossRef Tang M, Wang L, Koch DD, Li Y, Huang D (2012) Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography. J Cataract Refract Surg 238:589–594CrossRef
17.
Zurück zum Zitat Qazi MA, Cua IY, Roberts CJ, Pepose JS (2007) Determining corneal power using Orbscan II videokeratography for intraocular lens calculation after excimer laser surgery for myopia. J Cataract Refract Surg 33:21–30CrossRef Qazi MA, Cua IY, Roberts CJ, Pepose JS (2007) Determining corneal power using Orbscan II videokeratography for intraocular lens calculation after excimer laser surgery for myopia. J Cataract Refract Surg 33:21–30CrossRef
18.
Zurück zum Zitat Kwitko S, Marinho DR, Rymer S, Severo N, Arce CG (2012) Orbscan II and double-K method for IOL calculation after refractive surgery. Graefes Arch Clin Exp Ophthalmol 2250:1029–1034CrossRef Kwitko S, Marinho DR, Rymer S, Severo N, Arce CG (2012) Orbscan II and double-K method for IOL calculation after refractive surgery. Graefes Arch Clin Exp Ophthalmol 2250:1029–1034CrossRef
19.
Zurück zum Zitat Shammas HJ, Hoffer KJ, Shammas MC (2009) Scheimpflug photography keratometry readings for routine intraocular lens power calculation. J Cataract Refract Surg 35:330–334CrossRef Shammas HJ, Hoffer KJ, Shammas MC (2009) Scheimpflug photography keratometry readings for routine intraocular lens power calculation. J Cataract Refract Surg 35:330–334CrossRef
20.
Zurück zum Zitat Ayala M, Strandås R (2015) Accuracy of optical coherence tomography (OCT) in pachymetry for glaucoma patients. BMC Ophthalmol 15:124 Ayala M, Strandås R (2015) Accuracy of optical coherence tomography (OCT) in pachymetry for glaucoma patients. BMC Ophthalmol 15:124
21.
Zurück zum Zitat Gullstrand A (1911) Einfihrung in die Methoden der Dioptrik des Auges des Menschen. Hirzel, Leipzig Gullstrand A (1911) Einfihrung in die Methoden der Dioptrik des Auges des Menschen. Hirzel, Leipzig
22.
Zurück zum Zitat Stenstrom S (1964) Optics and the eye. Goteborg, Akademiforlaget Stenstrom S (1964) Optics and the eye. Goteborg, Akademiforlaget
23.
Zurück zum Zitat Olsen T (1986) On the calculation of power from curvature of the cornea. Br J Ophthalmol 70:152–154CrossRef Olsen T (1986) On the calculation of power from curvature of the cornea. Br J Ophthalmol 70:152–154CrossRef
24.
Zurück zum Zitat Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, Bentow S (2015) Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol 160:403–405CrossRef Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, Bentow S (2015) Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol 160:403–405CrossRef
25.
Zurück zum Zitat Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Intraocular lens formula constant optimization and partial coherence interferometry biometry: refractive outcomes in 8108 eyes after cataract surgery. J Cataract Refract Surg 37:50–62CrossRef Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Intraocular lens formula constant optimization and partial coherence interferometry biometry: refractive outcomes in 8108 eyes after cataract surgery. J Cataract Refract Surg 37:50–62CrossRef
Metadaten
Titel
Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery
verfasst von
Sabong Srivannaboon
Chareenun Chirapapaisan
Publikationsdatum
05.09.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 12/2019
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-019-04443-7

Weitere Artikel der Ausgabe 12/2019

Graefe's Archive for Clinical and Experimental Ophthalmology 12/2019 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.