Skip to main content
Erschienen in: Pediatric Surgery International 9/2020

14.07.2020 | Original Article

CT validation of SVC-RA junction location for pediatric central line placement: is vertebral bodies below the carina accurate?

verfasst von: Jacob R. Hirschl, Samir K. Gadepalli, Brian A. Derstine, Sven A. Holcombe, Ethan A. Smith, Stewart C. Wang, Peter F. Ehrlich, Marcus D. Jarboe

Erschienen in: Pediatric Surgery International | Ausgabe 9/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The key landmark for tip position of a central venous catheter (CVC) is the SVC-RA junction. In adults, localization of the SVC-RA junction may be assessed as a function of vertebral body units (VBU) below the carina during CVC placement. We investigated the relationship between the SVC-RA junction and the carina in children.

Materials and methods

584 CT scans of 0–18 years were analyzed. The carina was marked automatically by software while the SVC-RA junction and vertebrae were marked manually. The SVC-RA junction to carina (JC) distance was the primary study measurement reported in both VBU and mm.

Results

The data show an average JC distance of 1.25 VBU for 0–1 year, 1.27 VBU for 1–4 years, 1.34 VBU for 4–9 years, 1.53 VBU for 9–15 years, and 1.64 VBU for 15–18 years. A positive relationship between weight and JC distance was also demonstrated.

Conclusion

JC distance is a useful predictor of SVC-RA junction location in children. Significant relationships were shown between JC distance and both age and weight. Due to small differences between age groups, however, average JC distance for all comers (1.48 VBU, 95% CI 0.7 – 2.3) can be used for SVC-RA junction identification in CVC placement.
Literatur
1.
Zurück zum Zitat Somme S, Bronsert M, Morrato E, Ziegler M (2013) Frequency and variety of inpatient pediatric surgical procedures in the United States. Pediatrics 132(6):e1466–e1472CrossRef Somme S, Bronsert M, Morrato E, Ziegler M (2013) Frequency and variety of inpatient pediatric surgical procedures in the United States. Pediatrics 132(6):e1466–e1472CrossRef
2.
Zurück zum Zitat Hsu J-H, Wang C-K, Chu K-S et al (2006) Comparison of radiographic landmarks and the echocardiographic svc/ra junction in the positioning of long-term central venous catheters. Acta Anaesthesiol Scand 50:731–735CrossRef Hsu J-H, Wang C-K, Chu K-S et al (2006) Comparison of radiographic landmarks and the echocardiographic svc/ra junction in the positioning of long-term central venous catheters. Acta Anaesthesiol Scand 50:731–735CrossRef
3.
Zurück zum Zitat Gish J, Wright T, Gadepalli S, Jarboe M (2016) Avoiding postoperative malposition of upper body tunneled central venous catheters in children: evaluating technique and depth of placement. J Pediatr Surg 51(8):1336–1340CrossRef Gish J, Wright T, Gadepalli S, Jarboe M (2016) Avoiding postoperative malposition of upper body tunneled central venous catheters in children: evaluating technique and depth of placement. J Pediatr Surg 51(8):1336–1340CrossRef
4.
Zurück zum Zitat Nayeemuddin M, Pherwani A, Asquith J (2013) Imaging and management of complications of central venous catheters. Clin Radiol 68(5):529–544CrossRef Nayeemuddin M, Pherwani A, Asquith J (2013) Imaging and management of complications of central venous catheters. Clin Radiol 68(5):529–544CrossRef
5.
Zurück zum Zitat Collier PE, Blocker SH, Graff DM, Doyle P (1998) Cardiac tamponade from central venous catheters. Am J Surg 176:212–214CrossRef Collier PE, Blocker SH, Graff DM, Doyle P (1998) Cardiac tamponade from central venous catheters. Am J Surg 176:212–214CrossRef
6.
Zurück zum Zitat Dane T, King E (1975) Fatal cardiac tamponade and other mechanical complications of central venous catheters. Br J Surg 62:6–10CrossRef Dane T, King E (1975) Fatal cardiac tamponade and other mechanical complications of central venous catheters. Br J Surg 62:6–10CrossRef
7.
Zurück zum Zitat Chan TY, England A, Meredith SM, McWilliams RG (2016) Radiologist variability in assessing the position of the cavoatrial junction on chest radiographs. Br J Radiol 89(1065):20150965CrossRef Chan TY, England A, Meredith SM, McWilliams RG (2016) Radiologist variability in assessing the position of the cavoatrial junction on chest radiographs. Br J Radiol 89(1065):20150965CrossRef
8.
Zurück zum Zitat Baskin KM, Jimenez RM, Cahill AM et al (2008) Cavoatrial junction and central venous anatomy: implications for central venous access tip position. J Vasc Interv Radiol 19:359–365CrossRef Baskin KM, Jimenez RM, Cahill AM et al (2008) Cavoatrial junction and central venous anatomy: implications for central venous access tip position. J Vasc Interv Radiol 19:359–365CrossRef
9.
Zurück zum Zitat Song Y, Byun J, Hwang S et al (2015) Use of vertebral body units to locate the cavoatrial junction for optimum central venous catheter tip positioning. Br J Anaesth 115:252–257CrossRef Song Y, Byun J, Hwang S et al (2015) Use of vertebral body units to locate the cavoatrial junction for optimum central venous catheter tip positioning. Br J Anaesth 115:252–257CrossRef
10.
Zurück zum Zitat Englesbe MJ, Terjimanian MN, Lee JS et al (2013) Morphometric age and surgical risk. J Am Coll Surg 216:976–985CrossRef Englesbe MJ, Terjimanian MN, Lee JS et al (2013) Morphometric age and surgical risk. J Am Coll Surg 216:976–985CrossRef
11.
Zurück zum Zitat Harbaugh CM, Terjimanian MN, Lee JS et al (2013) Abdominal aortic calcification and surgical outcomes in patients with no known cardiovascular risk factors. Ann Surg 257:774–781CrossRef Harbaugh CM, Terjimanian MN, Lee JS et al (2013) Abdominal aortic calcification and surgical outcomes in patients with no known cardiovascular risk factors. Ann Surg 257:774–781CrossRef
12.
Zurück zum Zitat Lee JS, He K, Harbaugh CM et al (2011) Frailty, core muscle size, and mortality in patients undergoing open abdominal aortic aneurysm repair. J Vasc Surg 53:912–917CrossRef Lee JS, He K, Harbaugh CM et al (2011) Frailty, core muscle size, and mortality in patients undergoing open abdominal aortic aneurysm repair. J Vasc Surg 53:912–917CrossRef
13.
Zurück zum Zitat Lee JS, Terjimanian MN, Tishberg LM et al (2011) Surgical site infection and analytic morphometric assessment of body composition in patients undergoing midline laparotomy. J Am Coll Surg 213:236–244CrossRef Lee JS, Terjimanian MN, Tishberg LM et al (2011) Surgical site infection and analytic morphometric assessment of body composition in patients undergoing midline laparotomy. J Am Coll Surg 213:236–244CrossRef
14.
Zurück zum Zitat Harbaugh CM, Zhang P, Henderson B et al (2017) Personalized medicine: enhancing our understanding of pediatric growth with analytic morphomics. J Pediatr Surg 52(5):837–842CrossRef Harbaugh CM, Zhang P, Henderson B et al (2017) Personalized medicine: enhancing our understanding of pediatric growth with analytic morphomics. J Pediatr Surg 52(5):837–842CrossRef
15.
Zurück zum Zitat Parenteau CS, Ehrlich P, Ma L et al (2013) The quantification of liver anatomical changes and assessment of occupant liver injury patterns. Stapp Car Crash J 57:267–283PubMed Parenteau CS, Ehrlich P, Ma L et al (2013) The quantification of liver anatomical changes and assessment of occupant liver injury patterns. Stapp Car Crash J 57:267–283PubMed
16.
Zurück zum Zitat Zhang P, Parenteau C, Wang L et al (2013) Prediction of thoracic injury severity in frontal impacts by selected anatomical morphomic variables through model-averaged logistic regression approach. Accid Anal Prev 60:172–180CrossRef Zhang P, Parenteau C, Wang L et al (2013) Prediction of thoracic injury severity in frontal impacts by selected anatomical morphomic variables through model-averaged logistic regression approach. Accid Anal Prev 60:172–180CrossRef
17.
Zurück zum Zitat Core Team R (2016) R: a language and environment for statistical computing. R Foundation for Statistical Computing, Vienna Core Team R (2016) R: a language and environment for statistical computing. R Foundation for Statistical Computing, Vienna
Metadaten
Titel
CT validation of SVC-RA junction location for pediatric central line placement: is vertebral bodies below the carina accurate?
verfasst von
Jacob R. Hirschl
Samir K. Gadepalli
Brian A. Derstine
Sven A. Holcombe
Ethan A. Smith
Stewart C. Wang
Peter F. Ehrlich
Marcus D. Jarboe
Publikationsdatum
14.07.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Surgery International / Ausgabe 9/2020
Print ISSN: 0179-0358
Elektronische ISSN: 1437-9813
DOI
https://doi.org/10.1007/s00383-020-04712-1

Weitere Artikel der Ausgabe 9/2020

Pediatric Surgery International 9/2020 Zur Ausgabe

Neuer Typ-1-Diabetes bei Kindern am Wochenende eher übersehen

23.04.2024 Typ-1-Diabetes Nachrichten

Wenn Kinder an Werktagen zum Arzt gehen, werden neu auftretender Typ-1-Diabetes und diabetische Ketoazidosen häufiger erkannt als bei Arztbesuchen an Wochenenden oder Feiertagen.

Neue Studienergebnisse zur Myopiekontrolle mit Atropin

22.04.2024 Fehlsichtigkeit Nachrichten

Augentropfen mit niedrig dosiertem Atropin können helfen, das Fortschreiten einer Kurzsichtigkeit bei Kindern zumindest zu verlangsamen, wie die Ergebnisse einer aktuellen Studie mit verschiedenen Dosierungen zeigen.

Spinale Muskelatrophie: Neugeborenen-Screening lohnt sich

18.04.2024 Spinale Muskelatrophien Nachrichten

Seit 2021 ist die Untersuchung auf spinale Muskelatrophie Teil des Neugeborenen-Screenings in Deutschland. Eine Studie liefert weitere Evidenz für den Nutzen der Maßnahme.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.