Skip to main content
Erschienen in: Pediatric Surgery International 3/2016

01.03.2016 | Technical Innovation

Development of an operative suspension system for the performance of MRI-OR-guided laparoscopic anoplasty

verfasst von: Tina Thomas, Daniel H. Teitelbaum, Ranjith Vellody, Dragan Spremo, Samual Elkins, Robert Ladouceur, David Nagy, Marcus D. Jarboe

Erschienen in: Pediatric Surgery International | Ausgabe 3/2016

Einloggen, um Zugang zu erhalten

Abstract

Background

MRI-guided laparoscopic assisted anorectoplasty (MRI-LAARP), a new approach for surgical correction of high imperforate anus, does not bisect the sphincter complex as in the PSARP and is able to pull the neorectum through the entire sphincter complex unlike the LAARP. There is no available MRI-compatible device to position and transport patients during this procedure. We report on the design of such a device here.

Methods and device

The device was constructed from 1.0″ polyvinylchloride tubing and poly-methyl methacrylate (Plexiglass®) platform. The device has a stable, rigid base on which platform is secure. An adjustable and removable superstructure is secured to this base to suspend legs for lithotomy position.

Results

MRI-LAARP has been performed on 6 patients. The device has performed well and meets requirements set forth in development including construction with MRI-compatible materials, size fitting in the MRI bore, ability to hold patient in lithotomy position, ability to position and support MRI flex coils, and providing stability while transporting to a separate OR with needle in position.

Conclusions

This device provides a stable structure to position and transport a patient with a needle in a tenuous position without dislodgement allowing this procedure, and potentially other procedures, to be done in hospitals without MROR capability.
Literatur
1.
Zurück zum Zitat Iwai N, Fumino S (2013) Surgical treatment of anorectal malformations. Surg Today 43:955–962CrossRefPubMed Iwai N, Fumino S (2013) Surgical treatment of anorectal malformations. Surg Today 43:955–962CrossRefPubMed
2.
Zurück zum Zitat Pena A, Devries PA (1982) Posterior sagittal anorectoplasty: important technical considerations and new applications. J Pediatr Surg 17:796–811CrossRefPubMed Pena A, Devries PA (1982) Posterior sagittal anorectoplasty: important technical considerations and new applications. J Pediatr Surg 17:796–811CrossRefPubMed
3.
Zurück zum Zitat Georgeson KE, Inge TH, Albanese CT (2000) Laparoscopically assisted anorectal pull-through for high imperforate anus—a new technique. J Pediatr Surg 35:927–930CrossRefPubMed Georgeson KE, Inge TH, Albanese CT (2000) Laparoscopically assisted anorectal pull-through for high imperforate anus—a new technique. J Pediatr Surg 35:927–930CrossRefPubMed
4.
Zurück zum Zitat Raschbaum GR, Bleacher JC, Damien Grattan-Smith J, Jones RA (2010) Magnetic resonance imaging-guided laparoscopic-assisted anorectoplasty for imperforate anus. J Pediatr Surg 45:220–223CrossRefPubMed Raschbaum GR, Bleacher JC, Damien Grattan-Smith J, Jones RA (2010) Magnetic resonance imaging-guided laparoscopic-assisted anorectoplasty for imperforate anus. J Pediatr Surg 45:220–223CrossRefPubMed
5.
Zurück zum Zitat Alsam A, Grier DJ, Duncan AW, Spicer RD (1998) The role of magnetic resonance imaging in the preoperative assessment of anorectal anomalies. Pediatr Surg Int 14:71–736CrossRef Alsam A, Grier DJ, Duncan AW, Spicer RD (1998) The role of magnetic resonance imaging in the preoperative assessment of anorectal anomalies. Pediatr Surg Int 14:71–736CrossRef
6.
Zurück zum Zitat Guo M, Gao C, Li D, Guo W, Shafik AA, Zbar AP, Pescatori M (2010) MRI anatomy of the anal region. Dis Colon Rectum 53(11):1542–1548CrossRefPubMed Guo M, Gao C, Li D, Guo W, Shafik AA, Zbar AP, Pescatori M (2010) MRI anatomy of the anal region. Dis Colon Rectum 53(11):1542–1548CrossRefPubMed
7.
Zurück zum Zitat Black PM, Moriarty T, Alexander E 3rd, Stieg P, Woodard EJ, Gleason PL, Martin CH, Kikinis R, Schwartz RB, Jolesz FA (1997) Development and implementation of intraoperative magnetic resonance imaging and its neurosurgical applications. Neurosurgery 41(4):831–842 (discussion 842–845) CrossRefPubMed Black PM, Moriarty T, Alexander E 3rd, Stieg P, Woodard EJ, Gleason PL, Martin CH, Kikinis R, Schwartz RB, Jolesz FA (1997) Development and implementation of intraoperative magnetic resonance imaging and its neurosurgical applications. Neurosurgery 41(4):831–842 (discussion 842–845) CrossRefPubMed
8.
Zurück zum Zitat Hamrick M, Helmrath M, Bischoff A, Eradi B, Levitt M, Pena A (2011) Correspondence: letter to the Editor regarding comparison of anorectal angle and continence after Georgeson and Pena Procedures for high/imperforate anus. J Pediatr Surg 46:1019–1020CrossRefPubMed Hamrick M, Helmrath M, Bischoff A, Eradi B, Levitt M, Pena A (2011) Correspondence: letter to the Editor regarding comparison of anorectal angle and continence after Georgeson and Pena Procedures for high/imperforate anus. J Pediatr Surg 46:1019–1020CrossRefPubMed
Metadaten
Titel
Development of an operative suspension system for the performance of MRI-OR-guided laparoscopic anoplasty
verfasst von
Tina Thomas
Daniel H. Teitelbaum
Ranjith Vellody
Dragan Spremo
Samual Elkins
Robert Ladouceur
David Nagy
Marcus D. Jarboe
Publikationsdatum
01.03.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Surgery International / Ausgabe 3/2016
Print ISSN: 0179-0358
Elektronische ISSN: 1437-9813
DOI
https://doi.org/10.1007/s00383-015-3838-1

Weitere Artikel der Ausgabe 3/2016

Pediatric Surgery International 3/2016 Zur Ausgabe

Neuer Typ-1-Diabetes bei Kindern am Wochenende eher übersehen

23.04.2024 Typ-1-Diabetes Nachrichten

Wenn Kinder an Werktagen zum Arzt gehen, werden neu auftretender Typ-1-Diabetes und diabetische Ketoazidosen häufiger erkannt als bei Arztbesuchen an Wochenenden oder Feiertagen.

Neue Studienergebnisse zur Myopiekontrolle mit Atropin

22.04.2024 Fehlsichtigkeit Nachrichten

Augentropfen mit niedrig dosiertem Atropin können helfen, das Fortschreiten einer Kurzsichtigkeit bei Kindern zumindest zu verlangsamen, wie die Ergebnisse einer aktuellen Studie mit verschiedenen Dosierungen zeigen.

Spinale Muskelatrophie: Neugeborenen-Screening lohnt sich

18.04.2024 Spinale Muskelatrophien Nachrichten

Seit 2021 ist die Untersuchung auf spinale Muskelatrophie Teil des Neugeborenen-Screenings in Deutschland. Eine Studie liefert weitere Evidenz für den Nutzen der Maßnahme.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.