Development of force measurement system for clinical use in minimal access surgery

verfasst von: George B. Hanna, Tim Drew, Graham Arnold, Morkos Fakhry, Alfred Cuschieri

Erschienen in: Surgical Endoscopy | Ausgabe 2/2008

Einloggen, um Zugang zu erhalten

Abstract

Background

Analysis of force in minimal access surgery (MAS) is important for instrument design, surgical simulators, and in the understanding of tissue trauma incurred during surgery. The aim of this study is to develop a force measuring system for use with different instruments in clinical practice.

Methods

Strain gauges were connected to both arms of a standard -5 mm interchangeable forceps handle. A rotational sensor was used to indicate the relative position of the handle arms, and consequently the jaws’ position. A generic force-direction assembly was manufactured to determine the force direction at the port site. Interface electronics included signal conditioning and patient isolation circuits. Dedicated software was used for data acquisition, display, and analysis. To test their performance after sterilization, repeated force measures were obtained with the instruments after 10 cycles of autoclaving. Graduated weights were used to calibrate the strain gauges and a spring balance was employed to calibrate the force applied at the instrument tip. Calibration tests were also carried out to determine the effect of mounting the force direction assembly onto the access port.

Results

Gripping, dissecting, pushing, and pulling forces, along with the vector sum of forces acting at the port site, were synchronously displayed with the operative video record. Repeated autoclaving caused no deterioration in force sensing or signal transmission. The accuracy of the strain gauge readings was ±0.05 V for the jaw force and ±0.1 V for the force at the access port. The additional force created by the force direction assembly force was 7% of the port force alone.

Conclusion

Force measurement system has been developed for clinical use. The system measures the gripping, dissecting, pulling and pushing forces as well as the force vector at port site. It also determines the position of instrument’s jaws.
Literatur
1.
Zurück zum Zitat Berguer R, Forkey DL, Smith WD (1999) Ergonomic problems associated with laparoscopic surgery. Surg Endosc 13:466–468CrossRefPubMed Berguer R, Forkey DL, Smith WD (1999) Ergonomic problems associated with laparoscopic surgery. Surg Endosc 13:466–468CrossRefPubMed
2.
Zurück zum Zitat Bicchi A, Canepa G, De Rossi D, Iacconi P, Scillingo EP (1996) A sensor-based minimally invasive surgery tool for detecting tissue elastic properties. Proceedings of the 1996 IEEE International Conference on Robotics and Automation Minneapolis, Minnesota, pp 884–888 Bicchi A, Canepa G, De Rossi D, Iacconi P, Scillingo EP (1996) A sensor-based minimally invasive surgery tool for detecting tissue elastic properties. Proceedings of the 1996 IEEE International Conference on Robotics and Automation Minneapolis, Minnesota, pp 884–888
3.
Zurück zum Zitat Hanna GB, Shimi S, Cuschieri A (1997) Optimal port locations for endoscopic intracorporeal knotting. Surg Endosc 11:397–401CrossRefPubMed Hanna GB, Shimi S, Cuschieri A (1997) Optimal port locations for endoscopic intracorporeal knotting. Surg Endosc 11:397–401CrossRefPubMed
4.
Zurück zum Zitat Hu T, Tholey G, Desai JP, Castellanos AE (2004) Evaluation of a laparoscopic grasper with force feedback. Surg Endosc 18:863–867PubMed Hu T, Tholey G, Desai JP, Castellanos AE (2004) Evaluation of a laparoscopic grasper with force feedback. Surg Endosc 18:863–867PubMed
5.
Zurück zum Zitat Ilic D, Moix T, Mc Cullough N, Duratti L, Vecerina I, Bleuler H (2005) Real-time haptic interface for VR colonoscopy simulation. Stud Health Technol Inform 111:208–212PubMed Ilic D, Moix T, Mc Cullough N, Duratti L, Vecerina I, Bleuler H (2005) Real-time haptic interface for VR colonoscopy simulation. Stud Health Technol Inform 111:208–212PubMed
6.
Zurück zum Zitat Ma X, Brett PN, Wright MT, Griffiths MV (2004) A flexible digit with tactile feedback for invasive clinical applications. Proc Inst Mech Eng [H] 218:151–157 Ma X, Brett PN, Wright MT, Griffiths MV (2004) A flexible digit with tactile feedback for invasive clinical applications. Proc Inst Mech Eng [H] 218:151–157
7.
Zurück zum Zitat Moix T, Ilic D, Fracheboud B, Zoethout J, Bleuler H (2005) A real-time haptic interface for interventional radiology procedures. Stud Health Technol Inform 111:329–333PubMed Moix T, Ilic D, Fracheboud B, Zoethout J, Bleuler H (2005) A real-time haptic interface for interventional radiology procedures. Stud Health Technol Inform 111:329–333PubMed
8.
Zurück zum Zitat Rosen J, Hannaford B, Richards CG, Sinanan MN (2001) Markov modeling of minimally invasive surgery based on tool/tissue interaction and force/torque signatures for evaluating surgical skills. IEEE Trans Biomed Eng 48:579–591CrossRefPubMed Rosen J, Hannaford B, Richards CG, Sinanan MN (2001) Markov modeling of minimally invasive surgery based on tool/tissue interaction and force/torque signatures for evaluating surgical skills. IEEE Trans Biomed Eng 48:579–591CrossRefPubMed
9.
Zurück zum Zitat Sorensen TS, Mosegaard J (2006) Haptic feedback for the GPU-based surgical simulator. Stud Health Technol Inform 119:523–528PubMed Sorensen TS, Mosegaard J (2006) Haptic feedback for the GPU-based surgical simulator. Stud Health Technol Inform 119:523–528PubMed
10.
Zurück zum Zitat Tang B, Hanna GB, Cuschieri A (2005) Analysis of errors enacted by surgical trainees during skills training courses. Surgery 138:14–20CrossRefPubMed Tang B, Hanna GB, Cuschieri A (2005) Analysis of errors enacted by surgical trainees during skills training courses. Surgery 138:14–20CrossRefPubMed
11.
Zurück zum Zitat Tay BK, Stylopoulos N, De S, Rattner DW, Srinivasan MA (2002) Measurement of in-vivo force response of intra-abdominal soft tissues for surgical simulation. Stud Health Technol Inform 85:514–519PubMed Tay BK, Stylopoulos N, De S, Rattner DW, Srinivasan MA (2002) Measurement of in-vivo force response of intra-abdominal soft tissues for surgical simulation. Stud Health Technol Inform 85:514–519PubMed
12.
Zurück zum Zitat Tholey G, Desai JP, Castellanos AE (2005) Force feedback plays a significant role in minimally invasive surgery: results and analysis. Ann Surg 241:102–109PubMed Tholey G, Desai JP, Castellanos AE (2005) Force feedback plays a significant role in minimally invasive surgery: results and analysis. Ann Surg 241:102–109PubMed
13.
Zurück zum Zitat Zivanovic A, Dibble E, Davies B (2004) Novel force resolver designs for a haptic surgery simulator. Stud Health Technol Inform 98:422–424PubMed Zivanovic A, Dibble E, Davies B (2004) Novel force resolver designs for a haptic surgery simulator. Stud Health Technol Inform 98:422–424PubMed
Metadaten
Titel
Development of force measurement system for clinical use in minimal access surgery
verfasst von
George B. Hanna
Tim Drew
Graham Arnold
Morkos Fakhry
Alfred Cuschieri
Publikationsdatum
01.02.2008
Verlag
Springer-Verlag
Erschienen in
Surgical Endoscopy / Ausgabe 2/2008
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-007-9489-0

Weitere Artikel der Ausgabe 2/2008

Surgical Endoscopy 2/2008 Zur Ausgabe

Echinokokkose medikamentös behandeln oder operieren?

06.05.2024 DCK 2024 Kongressbericht

Die Therapie von Echinokokkosen sollte immer in spezialisierten Zentren erfolgen. Eine symptomlose Echinokokkose kann – egal ob von Hunde- oder Fuchsbandwurm ausgelöst – konservativ erfolgen. Wenn eine Op. nötig ist, kann es sinnvoll sein, vorher Zysten zu leeren und zu desinfizieren. 

Recycling im OP – möglich, aber teuer

05.05.2024 DCK 2024 Kongressbericht

Auch wenn sich Krankenhäuser nachhaltig und grün geben – sie tragen aktuell erheblich zu den CO2-Emissionen bei und produzieren jede Menge Müll. Ein Pilotprojekt aus Bonn zeigt, dass viele Op.-Abfälle wiederverwertet werden können.

Im OP der Zukunft läuft nichts mehr ohne Kollege Roboter

04.05.2024 DCK 2024 Kongressbericht

Der OP in der Zukunft wird mit weniger Personal auskommen – nicht, weil die Technik das medizinische Fachpersonal verdrängt, sondern weil der Personalmangel es nötig macht.

Wo hapert es noch bei der Umsetzung der POMGAT-Leitlinie?

03.05.2024 DCK 2024 Kongressbericht

Seit November 2023 gibt es evidenzbasierte Empfehlungen zum perioperativen Management bei gastrointestinalen Tumoren (POMGAT) auf S3-Niveau. Vieles wird schon entsprechend der Empfehlungen durchgeführt. Wo es im Alltag noch hapert, zeigt eine Umfrage in einem Klinikverbund.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.