Dual modality intravascular optical coherence tomography (OCT) and near-infrared fluorescence (NIRF) imaging: a fully automated algorithm for the distance-calibration of NIRF signal intensity for quantitative molecular imaging

verfasst von: Giovanni J. Ughi, Johan Verjans, Ali M. Fard, Hao Wang, Eric Osborn, Tetsuya Hara, Adam Mauskapf, Farouc A. Jaffer, Guillermo J. Tearney

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 2/2015

Einloggen, um Zugang zu erhalten

Abstract

Intravascular optical coherence tomography (IVOCT) is a well-established method for the high-resolution investigation of atherosclerosis in vivo. Intravascular near-infrared fluorescence (NIRF) imaging is a novel technique for the assessment of molecular processes associated with coronary artery disease. Integration of NIRF and IVOCT technology in a single catheter provides the capability to simultaneously obtain co-localized anatomical and molecular information from the artery wall. Since NIRF signal intensity attenuates as a function of imaging catheter distance to the vessel wall, the generation of quantitative NIRF data requires an accurate measurement of the vessel wall in IVOCT images. Given that dual modality, intravascular OCT–NIRF systems acquire data at a very high frame-rate (>100 frames/s), a high number of images per pullback need to be analyzed, making manual processing of OCT–NIRF data extremely time consuming. To overcome this limitation, we developed an algorithm for the automatic distance-correction of dual-modality OCT–NIRF images. We validated this method by comparing automatic to manual segmentation results in 180 in vivo images from six New Zealand White rabbit atherosclerotic after indocyanine-green injection. A high Dice similarity coefficient was found (0.97 ± 0.03) together with an average individual A-line error of 22 µm (i.e., approximately twice the axial resolution of IVOCT) and a processing time of 44 ms per image. In a similar manner, the algorithm was validated using 120 IVOCT clinical images from eight different in vivo pullbacks in human coronary arteries. The results suggest that the proposed algorithm enables fully automatic visualization of dual modality OCT–NIRF pullbacks, and provides an accurate and efficient calibration of NIRF data for quantification of the molecular agent in the atherosclerotic vessel wall.
Literatur
1.
Zurück zum Zitat Go AS, Mozaffarian D, Roger VL et al (2013) Heart disease and stroke statistics—2013 update: a report from the American Heart Association. Circulation 127(1):e6–e245CrossRefPubMed Go AS, Mozaffarian D, Roger VL et al (2013) Heart disease and stroke statistics—2013 update: a report from the American Heart Association. Circulation 127(1):e6–e245CrossRefPubMed
2.
Zurück zum Zitat Tearney GJ, Regar E, Akasaka T et al (2012) Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for intravascular optical coherence tomography standardization and validation. J Am Coll Cardiolol 59(12):1058–1072CrossRef Tearney GJ, Regar E, Akasaka T et al (2012) Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for intravascular optical coherence tomography standardization and validation. J Am Coll Cardiolol 59(12):1058–1072CrossRef
3.
Zurück zum Zitat Yun SH, Tearney GJ, Vakoc BJ, Shishkov M, Oh WY, Desjardins AE, Suter MJ, Chan RC, Evans JA, Jang I, Nishioka NS, de Boer JF, Bouma BE (2006) Comprehensive volumetric optical microscopy in vivo. Nat Med 12:1429–1433CrossRefPubMedCentralPubMed Yun SH, Tearney GJ, Vakoc BJ, Shishkov M, Oh WY, Desjardins AE, Suter MJ, Chan RC, Evans JA, Jang I, Nishioka NS, de Boer JF, Bouma BE (2006) Comprehensive volumetric optical microscopy in vivo. Nat Med 12:1429–1433CrossRefPubMedCentralPubMed
4.
Zurück zum Zitat Farooq V, Serruys PW, Heo JH, Gogas BD, Onuma Y, Perkins LE, Diletti R, Radu MD, Raber L, Bourantas CV, Zhang Y, van Remortel E, Pawar R, Rapoza RJ, Powers JC, van Beusekom HM, Garcia-Garcia HM, Virmani R (2013) Intracoronary optical coherence tomography and histology of overlapping everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: the potential implications for clinical practice. J Am Coll Cardiol Intv 6(5):523–532CrossRef Farooq V, Serruys PW, Heo JH, Gogas BD, Onuma Y, Perkins LE, Diletti R, Radu MD, Raber L, Bourantas CV, Zhang Y, van Remortel E, Pawar R, Rapoza RJ, Powers JC, van Beusekom HM, Garcia-Garcia HM, Virmani R (2013) Intracoronary optical coherence tomography and histology of overlapping everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: the potential implications for clinical practice. J Am Coll Cardiol Intv 6(5):523–532CrossRef
5.
Zurück zum Zitat Farooq V, Gogas BD, Okamura T, Heo JH, Magro M, Gomez-Lara J, Onuma Y, Radu MD, Brugaletta S, van Bochove G, van Geuns RJ, Garcia-Garcia HM, Serruys PW (2011) Three-dimensional optical frequency domain imaging in conventional percutaneous coronary intervention: the potential for clinical application. Eur Heart J 34(12):875–885CrossRefPubMed Farooq V, Gogas BD, Okamura T, Heo JH, Magro M, Gomez-Lara J, Onuma Y, Radu MD, Brugaletta S, van Bochove G, van Geuns RJ, Garcia-Garcia HM, Serruys PW (2011) Three-dimensional optical frequency domain imaging in conventional percutaneous coronary intervention: the potential for clinical application. Eur Heart J 34(12):875–885CrossRefPubMed
6.
Zurück zum Zitat Jaffer FA, Calfon MA, Rosenthal A, Mallas G, Razansky N, Mauskampf A, Weissleder R, Libby P, Ntziachristos V (2011) Two-dimensional intravascular near-infrared fluorescence molecular imaging of inflammation in atherosclerosis and stent-induced vascular injury. J Am Coll Cardiol 57(25):2516–2525CrossRefPubMedCentralPubMed Jaffer FA, Calfon MA, Rosenthal A, Mallas G, Razansky N, Mauskampf A, Weissleder R, Libby P, Ntziachristos V (2011) Two-dimensional intravascular near-infrared fluorescence molecular imaging of inflammation in atherosclerosis and stent-induced vascular injury. J Am Coll Cardiol 57(25):2516–2525CrossRefPubMedCentralPubMed
7.
Zurück zum Zitat Vinegoni C, Botnaru I, Aikawa E, Calfon MA, Iwamoto Y, Folco EY, Ntziachristos V, Weissleder R, Libby P, Jaffer FA (2011) Indocyanine green enables near-infrared fluorescence imaging of lipid-rich, inflamed atherosclerotic plaques. Sci Transl Med 3(84):84ra45CrossRefPubMedCentralPubMed Vinegoni C, Botnaru I, Aikawa E, Calfon MA, Iwamoto Y, Folco EY, Ntziachristos V, Weissleder R, Libby P, Jaffer FA (2011) Indocyanine green enables near-infrared fluorescence imaging of lipid-rich, inflamed atherosclerotic plaques. Sci Transl Med 3(84):84ra45CrossRefPubMedCentralPubMed
8.
Zurück zum Zitat Osborn EA, Jaffer FA (2013) The advancing clinical impact of molecular imaging in CVD. JACC Cardiovasc Imaging 6(12):1327–1341CrossRefPubMed Osborn EA, Jaffer FA (2013) The advancing clinical impact of molecular imaging in CVD. JACC Cardiovasc Imaging 6(12):1327–1341CrossRefPubMed
10.
Zurück zum Zitat Yoo H, Kim JW, Shishkov M, Namati E, Morse T, Shubockin R, McCarthy JR, Ntziachristos V, Bouma BE, Jaffer FA, Tearney GJ (2011) Intra-arterial catheter for simultaneous microstructural and molecular imaging in vivo. Nat Med 17:1680–1684CrossRefPubMedCentralPubMed Yoo H, Kim JW, Shishkov M, Namati E, Morse T, Shubockin R, McCarthy JR, Ntziachristos V, Bouma BE, Jaffer FA, Tearney GJ (2011) Intra-arterial catheter for simultaneous microstructural and molecular imaging in vivo. Nat Med 17:1680–1684CrossRefPubMedCentralPubMed
11.
Zurück zum Zitat Stefanini GG, Holmes DR (2013) Drug-eluting coronary-artery stents. N Engl J Med 368:254–265CrossRefPubMed Stefanini GG, Holmes DR (2013) Drug-eluting coronary-artery stents. N Engl J Med 368:254–265CrossRefPubMed
12.
Zurück zum Zitat Camezind E, Steg P, Wijns W (2007) Stent thrombosis late after implantation of first-generation drug-eluting stents: a cause of concern. Circulation 115:1440–1455CrossRef Camezind E, Steg P, Wijns W (2007) Stent thrombosis late after implantation of first-generation drug-eluting stents: a cause of concern. Circulation 115:1440–1455CrossRef
13.
14.
Zurück zum Zitat Leitgeb R, Hitzenberger C, Fercher A (2003) Performance of fourier domain vs. time domain optical coherence tomography. Opt Express 11(8):889–894CrossRefPubMed Leitgeb R, Hitzenberger C, Fercher A (2003) Performance of fourier domain vs. time domain optical coherence tomography. Opt Express 11(8):889–894CrossRefPubMed
15.
Zurück zum Zitat Otsu N (1979) A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern 9(1):62–66CrossRef Otsu N (1979) A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern 9(1):62–66CrossRef
16.
Zurück zum Zitat Serra JP, Cressie NAC (1982) Image analysis and mathematical morphology. Academic Press, London Serra JP, Cressie NAC (1982) Image analysis and mathematical morphology. Academic Press, London
19.
Zurück zum Zitat Zou KH, Warfield SK, Bharatha A, Tempany CMC, Kaus MR, Haker SJ, Wells WM, Jolesz FA, Kikinis R (2004) Statistical validation of image segmentation quality based on a Spatial Overlap Index. Acad Radiol 11(2):178–189CrossRefPubMedCentralPubMed Zou KH, Warfield SK, Bharatha A, Tempany CMC, Kaus MR, Haker SJ, Wells WM, Jolesz FA, Kikinis R (2004) Statistical validation of image segmentation quality based on a Spatial Overlap Index. Acad Radiol 11(2):178–189CrossRefPubMedCentralPubMed
21.
Zurück zum Zitat van Soest G, Bosch JG, van der Steen AFW (2008) Azimuthal registration of image sequences affected by nonuniform rotation distortion. IEEE Trans Inf Technol Biomed 12(3):348–355CrossRefPubMed van Soest G, Bosch JG, van der Steen AFW (2008) Azimuthal registration of image sequences affected by nonuniform rotation distortion. IEEE Trans Inf Technol Biomed 12(3):348–355CrossRefPubMed
22.
Zurück zum Zitat Ha J, Yoo H, Tearney GJ, Bouma BE (2012) Compensation of motion artifacts in intracoronary optical frequency domain imaging and optical coherence tomography. Int J Cardiovasc Imaging 28(6):1299–1304CrossRefPubMedCentralPubMed Ha J, Yoo H, Tearney GJ, Bouma BE (2012) Compensation of motion artifacts in intracoronary optical frequency domain imaging and optical coherence tomography. Int J Cardiovasc Imaging 28(6):1299–1304CrossRefPubMedCentralPubMed
23.
Zurück zum Zitat Wang T, Wieser W, Springeling G, Beurskens R, Lancee CT, Pfeiffer T, van der Steen AFW, Huber RA, van Soest G (2013) Ultrahigh-speed intravascular optical coherence tomography imaging at 3200 frames per second. SPIE Proceed 8802. doi:10.1117/12.2032723 Wang T, Wieser W, Springeling G, Beurskens R, Lancee CT, Pfeiffer T, van der Steen AFW, Huber RA, van Soest G (2013) Ultrahigh-speed intravascular optical coherence tomography imaging at 3200 frames per second. SPIE Proceed 8802. doi:10.​1117/​12.​2032723
24.
Zurück zum Zitat Schmitt JM, Bezerra H, Petroff C, Gophinat A, Trachtenberg JC (2013) Method and apparatus for automated determination of a lumen contour of a blood vessel. Software, WO/2014/092755 Schmitt JM, Bezerra H, Petroff C, Gophinat A, Trachtenberg JC (2013) Method and apparatus for automated determination of a lumen contour of a blood vessel. Software, WO/2014/092755
25.
Zurück zum Zitat Ughi GJ, Adriaenssens T, Onsea K, Kayaert P, Dubois C, Sinnaeve P, Coosemans M, Desmet W, D’hooge J (2012) Automatic segmentation of in vivo intra-coronary optical coherence tomography images to assess stent strut apposition and coverage. Int J Cardiovasc Imaging 28(2):229–241CrossRefPubMed Ughi GJ, Adriaenssens T, Onsea K, Kayaert P, Dubois C, Sinnaeve P, Coosemans M, Desmet W, D’hooge J (2012) Automatic segmentation of in vivo intra-coronary optical coherence tomography images to assess stent strut apposition and coverage. Int J Cardiovasc Imaging 28(2):229–241CrossRefPubMed
Metadaten
Titel
Dual modality intravascular optical coherence tomography (OCT) and near-infrared fluorescence (NIRF) imaging: a fully automated algorithm for the distance-calibration of NIRF signal intensity for quantitative molecular imaging
verfasst von
Giovanni J. Ughi
Johan Verjans
Ali M. Fard
Hao Wang
Eric Osborn
Tetsuya Hara
Adam Mauskapf
Farouc A. Jaffer
Guillermo J. Tearney
Publikationsdatum
01.02.2015
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 2/2015
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-014-0556-z

Weitere Artikel der Ausgabe 2/2015

The International Journal of Cardiovascular Imaging 2/2015 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.