Early diagnosis of subclinical keratoconus by wavefront parameters using Scheimpflug, Placido and Hartmann–Shack based devices

verfasst von: Zahra Heidari, Mehrdad Mohammadpour, Hassan Hashemi, Ebrahim Jafarzadehpur, Alireza Moghaddasi, Mehdi Yaseri, Akbar Fotouhi

Erschienen in: International Ophthalmology | Ausgabe 7/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the accuracy of three devices (Pentacam, Sirius and OPD-Scan III) to differentiate subclinical keratoconus from normal corneas by wavefront parameters.

Methods

Two hundred and seventeen patients were enrolled in three groups [68 normal, 79 subclinical keratoconus (SKCN) and 70 KCN eyes] in this prospective diagnostic test study. Wavefront indices were evaluated between the groups using Pentacam, Sirius and OPD-Scan III. The accuracy of the parameters was determined by measuring the area under the receiver operating characteristic curve (AUC) for each group.

Results

Front Baiocchi-Calossi-Versaci (BCV) index with Sirius (sensitivity = 87.7%, specificity = 83%, AUC = 0.887), front Vertical Coma (Z3−1) with Pentacam (sensitivity = 75%, specificity = 100%, AUC = 0.857) and Corneal Z3−1 with OPD-Scan III (sensitivity = 100%, specificity = 78.6%, AUC = 0.857) had the highest AUC values for the diagnosis of subclinical KCN. In the KCN group, the highest AUC values were obtained for front higher-order aberration (HOA), front/back Z3−1 and front Secondary Vertical Coma (Z5−1) with Pentacam (sensitivity = 100%, specificity = 100%, AUC = 1.00 for all three), front root mean square values per unit area (RMS/A), HOA, Residual HOA, BCV, RMS Trefoil and RMS Coma with Sirius (sensitivity = 100%, specificity = 100%, AUC = 1.00 for all) and Corneal HOA, RMS total Coma and Z3−1 with OPD-Scan III (sensitivity = 100%, specificity = 93%, AUC = 0.96 for all three).

Conclusion

Corneal wavefront indices generated from different devices have acceptable validity for differentiating normal cornea from the early form of KCN, and this can be very useful for preoperative screening before refractive surgery. The front BCV with Sirius was the most accurate parameter for diagnosis of SKCN followed by Z3−1 with Pentacam and OPD-Scan III.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Applegate RA, Hilmantel G, Howland HC, Tu EY, Starck T, Zayac EJ (2000) Corneal first surface optical aberrations and visual performance. J Refract Surg 16:507–514CrossRef Applegate RA, Hilmantel G, Howland HC, Tu EY, Starck T, Zayac EJ (2000) Corneal first surface optical aberrations and visual performance. J Refract Surg 16:507–514CrossRef
2.
Zurück zum Zitat Mohammadpour M, Heidari Z, Mohammad-Rabei H, Jafarzadehpur E, Jabbarvand M, Hashemi H, Khabazkhoob M (2016) Correlation of higher order aberrations and components of astigmatism in myopic refractive surgery candidates. J Curr Ophthalmol 28:112–116CrossRef Mohammadpour M, Heidari Z, Mohammad-Rabei H, Jafarzadehpur E, Jabbarvand M, Hashemi H, Khabazkhoob M (2016) Correlation of higher order aberrations and components of astigmatism in myopic refractive surgery candidates. J Curr Ophthalmol 28:112–116CrossRef
3.
4.
Zurück zum Zitat Rabinowitz YS, Nesburn AB, McDonnell PJ (1993) Videokeratography of the fellow eye in unilateral keratoconus. Ophthalmology 100:181–186CrossRef Rabinowitz YS, Nesburn AB, McDonnell PJ (1993) Videokeratography of the fellow eye in unilateral keratoconus. Ophthalmology 100:181–186CrossRef
5.
Zurück zum Zitat Seiler T, Quurke AW (1998) Iatrogenic keratectasia after LASIK in a case of forme fruste keratoconus. J Cataract Refract Surg 24:1007–1009CrossRef Seiler T, Quurke AW (1998) Iatrogenic keratectasia after LASIK in a case of forme fruste keratoconus. J Cataract Refract Surg 24:1007–1009CrossRef
6.
Zurück zum Zitat Chen M, Yoon G (2008) Posterior corneal aberrations and their compensation effects on anterior corneal aberrations in keratoconic eyes. Invest Ophthalmol Vis Sci 49:5645–5652CrossRef Chen M, Yoon G (2008) Posterior corneal aberrations and their compensation effects on anterior corneal aberrations in keratoconic eyes. Invest Ophthalmol Vis Sci 49:5645–5652CrossRef
7.
Zurück zum Zitat Naderan M, Shoar S, Naderan M, Kamaleddin MA, Rajabi MT (2015) Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography. Int J Ophthalmol 8:275–280PubMedPubMedCentral Naderan M, Shoar S, Naderan M, Kamaleddin MA, Rajabi MT (2015) Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography. Int J Ophthalmol 8:275–280PubMedPubMedCentral
8.
Zurück zum Zitat Alio JL, Shabayek MH (2006) Corneal higher order aberrations: a method to grade keratoconus. J Refract Surg 22:539–545CrossRef Alio JL, Shabayek MH (2006) Corneal higher order aberrations: a method to grade keratoconus. J Refract Surg 22:539–545CrossRef
9.
Zurück zum Zitat Buhren J, Kuhne C, Kohnen T (2007) Defining subclinical keratoconus using corneal first-surface higher-order aberrations. Am J Ophthalmol 143:381–389CrossRef Buhren J, Kuhne C, Kohnen T (2007) Defining subclinical keratoconus using corneal first-surface higher-order aberrations. Am J Ophthalmol 143:381–389CrossRef
10.
Zurück zum Zitat Kosaki R, Maeda N, Bessho K, Hori Y, Nishida K, Suzaki A, Hirohara Y, Mihashi T, Fujikado T, Tano Y (2007) Magnitude and orientation of Zernike terms in patients with keratoconus. Invest Ophthalmol Vis Sci 48:3062–3068CrossRef Kosaki R, Maeda N, Bessho K, Hori Y, Nishida K, Suzaki A, Hirohara Y, Mihashi T, Fujikado T, Tano Y (2007) Magnitude and orientation of Zernike terms in patients with keratoconus. Invest Ophthalmol Vis Sci 48:3062–3068CrossRef
11.
Zurück zum Zitat Nakagawa T, Maeda N, Kosaki R, Hori Y, Inoue T, Saika M, Mihashi T, Fujikado T, Tano Y (2009) Higher-order aberrations due to the posterior corneal surface in patients with keratoconus. Invest Ophthalmol Vis Sci 50:2660–2665CrossRef Nakagawa T, Maeda N, Kosaki R, Hori Y, Inoue T, Saika M, Mihashi T, Fujikado T, Tano Y (2009) Higher-order aberrations due to the posterior corneal surface in patients with keratoconus. Invest Ophthalmol Vis Sci 50:2660–2665CrossRef
12.
Zurück zum Zitat Barbero S, Marcos S, Merayo-Lloves J, Moreno-Barriuso E (2002) Validation of the estimation of corneal aberrations from videokeratography in keratoconus. J Refract Surg 18:263–270PubMed Barbero S, Marcos S, Merayo-Lloves J, Moreno-Barriuso E (2002) Validation of the estimation of corneal aberrations from videokeratography in keratoconus. J Refract Surg 18:263–270PubMed
13.
Zurück zum Zitat Pinero DP, Saenz Gonzalez C, Alio JL (2009) Intraobserver and interobserver repeatability of curvature and aberrometric measurements of the posterior corneal surface in normal eyes using Scheimpflug photography. J Cataract Refract Surg 35:113–120CrossRef Pinero DP, Saenz Gonzalez C, Alio JL (2009) Intraobserver and interobserver repeatability of curvature and aberrometric measurements of the posterior corneal surface in normal eyes using Scheimpflug photography. J Cataract Refract Surg 35:113–120CrossRef
14.
Zurück zum Zitat Shankar H, Taranath D, Santhirathelagan CT, Pesudovs K (2008) Repeatability of corneal first-surface wavefront aberrations measured with Pentacam corneal topography. J Cataract Refract Surg 34:727–734CrossRef Shankar H, Taranath D, Santhirathelagan CT, Pesudovs K (2008) Repeatability of corneal first-surface wavefront aberrations measured with Pentacam corneal topography. J Cataract Refract Surg 34:727–734CrossRef
15.
Zurück zum Zitat Pinero DP, Alio JL, Aleson A, Escaf M, Miranda M (2009) Pentacam posterior and anterior corneal aberrations in normal and keratoconic eyes. Clin Exp Optom 92:297–303CrossRef Pinero DP, Alio JL, Aleson A, Escaf M, Miranda M (2009) Pentacam posterior and anterior corneal aberrations in normal and keratoconic eyes. Clin Exp Optom 92:297–303CrossRef
16.
Zurück zum Zitat Buhren J, Kook D, Yoon G, Kohnen T (2010) Detection of subclinical keratoconus by using corneal anterior and posterior surface aberrations and thickness spatial profiles. Invest Ophthalmol Vis Sci 51:3424–3432CrossRef Buhren J, Kook D, Yoon G, Kohnen T (2010) Detection of subclinical keratoconus by using corneal anterior and posterior surface aberrations and thickness spatial profiles. Invest Ophthalmol Vis Sci 51:3424–3432CrossRef
17.
Zurück zum Zitat Xu Z, Li W, Jiang J, Zhuang X, Chen W, Peng M, Wang J, Lu F, Shen M, Wang Y (2017) Characteristic of entire corneal topography and tomography for the detection of sub-clinical keratoconus with Zernike polynomials using Pentacam. Sci Rep 7:16486CrossRef Xu Z, Li W, Jiang J, Zhuang X, Chen W, Peng M, Wang J, Lu F, Shen M, Wang Y (2017) Characteristic of entire corneal topography and tomography for the detection of sub-clinical keratoconus with Zernike polynomials using Pentacam. Sci Rep 7:16486CrossRef
18.
Zurück zum Zitat Hashemi H, Beiranvand A, Yekta A, Maleki A, Yazdani N, Khabazkhoob M (2016) Pentacam top indices for diagnosing subclinical and definite keratoconus. J Curr Ophthalmol 28:21–22CrossRef Hashemi H, Beiranvand A, Yekta A, Maleki A, Yazdani N, Khabazkhoob M (2016) Pentacam top indices for diagnosing subclinical and definite keratoconus. J Curr Ophthalmol 28:21–22CrossRef
19.
Zurück zum Zitat Colak HN, Kantarci FA, Yildirim A, Tatar MG, Goker H, Uslu H, Gurler B (2016) Comparison of corneal topographic measurements and high order aberrations in keratoconus and normal eyes. Cont Lens Anterior Eye 39:380–384CrossRef Colak HN, Kantarci FA, Yildirim A, Tatar MG, Goker H, Uslu H, Gurler B (2016) Comparison of corneal topographic measurements and high order aberrations in keratoconus and normal eyes. Cont Lens Anterior Eye 39:380–384CrossRef
20.
Zurück zum Zitat Prakash G, Suhail M, Srivastava D (2016) Predictive analysis between topographic, pachymetric and wavefront parameters in keratoconus, suspects and normal eyes: creating unified equations to evaluate keratoconus. Curr Eye Res 41:334–342PubMed Prakash G, Suhail M, Srivastava D (2016) Predictive analysis between topographic, pachymetric and wavefront parameters in keratoconus, suspects and normal eyes: creating unified equations to evaluate keratoconus. Curr Eye Res 41:334–342PubMed
21.
Zurück zum Zitat Safarzadeh M, Nasiri N (2016) Anterior segment characteristics in normal and keratoconus eyes evaluated with a combined Scheimpflug/Placido corneal imaging device. J Curr Ophthalmol 28:106–111CrossRef Safarzadeh M, Nasiri N (2016) Anterior segment characteristics in normal and keratoconus eyes evaluated with a combined Scheimpflug/Placido corneal imaging device. J Curr Ophthalmol 28:106–111CrossRef
22.
Zurück zum Zitat Naderan M, Jahanrad A, Farjadnia M (2018) Ocular, corneal, and internal aberrations in eyes with keratoconus, forme fruste keratoconus, and healthy eyes. Int Ophthalmol 38:1565–1573CrossRef Naderan M, Jahanrad A, Farjadnia M (2018) Ocular, corneal, and internal aberrations in eyes with keratoconus, forme fruste keratoconus, and healthy eyes. Int Ophthalmol 38:1565–1573CrossRef
23.
Zurück zum Zitat Saad A, Gatinel D (2012) Evaluation of total and corneal wavefront high order aberrations for the detection of forme fruste keratoconus. Invest Ophthalmol Vis Sci 53:2978–2992CrossRef Saad A, Gatinel D (2012) Evaluation of total and corneal wavefront high order aberrations for the detection of forme fruste keratoconus. Invest Ophthalmol Vis Sci 53:2978–2992CrossRef
24.
Zurück zum Zitat Maeda N, Klyce SD, Smolek MK, Thompson HW (1994) Automated keratoconus screening with corneal topography analysis. Invest Ophthalmol Vis Sci 35:2749–2757PubMed Maeda N, Klyce SD, Smolek MK, Thompson HW (1994) Automated keratoconus screening with corneal topography analysis. Invest Ophthalmol Vis Sci 35:2749–2757PubMed
25.
Zurück zum Zitat Rabinowitz YS, McDonnell PJ (1989) Computer-assisted corneal topography in keratoconus. Refract corneal surg 5:400–408PubMed Rabinowitz YS, McDonnell PJ (1989) Computer-assisted corneal topography in keratoconus. Refract corneal surg 5:400–408PubMed
26.
Zurück zum Zitat Wang Q, Savini G, Hoffer KJ, Xu Z, Feng Y, Wen D, Hua Y, Yang F, Pan C, Huang J (2012) A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices. PLoS ONE 7:e45607CrossRef Wang Q, Savini G, Hoffer KJ, Xu Z, Feng Y, Wen D, Hua Y, Yang F, Pan C, Huang J (2012) A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices. PLoS ONE 7:e45607CrossRef
27.
Zurück zum Zitat Bayhan HA, Aslan Bayhan S, Muhafiz E, Can I (2014) Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer. J Cataract Refract Surg 40:269–275CrossRef Bayhan HA, Aslan Bayhan S, Muhafiz E, Can I (2014) Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer. J Cataract Refract Surg 40:269–275CrossRef
28.
Zurück zum Zitat Asgari S, Hashemi H, Jafarzadehpur E, Mohamadi A, Rezvan F, Fotouhi A (2016) OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus. Int Ophthalmol 36:697–705CrossRef Asgari S, Hashemi H, Jafarzadehpur E, Mohamadi A, Rezvan F, Fotouhi A (2016) OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus. Int Ophthalmol 36:697–705CrossRef
29.
Zurück zum Zitat Gordon-Shaag A, Millodot M, Ifrah R, Shneor E (2012) Aberrations and topography in normal, keratoconus-suspect, and keratoconic eyes. Optom Vis Sci 89:411–418CrossRef Gordon-Shaag A, Millodot M, Ifrah R, Shneor E (2012) Aberrations and topography in normal, keratoconus-suspect, and keratoconic eyes. Optom Vis Sci 89:411–418CrossRef
30.
Zurück zum Zitat Maeda N, Fujikado T, Kuroda T, Mihashi T, Hirohara Y, Nishida K, Watanabe H, Tano Y (2002) Wavefront aberrations measured with Hartmann–Shack sensor in patients with keratoconus. Ophthalmology 109:1996–2003CrossRef Maeda N, Fujikado T, Kuroda T, Mihashi T, Hirohara Y, Nishida K, Watanabe H, Tano Y (2002) Wavefront aberrations measured with Hartmann–Shack sensor in patients with keratoconus. Ophthalmology 109:1996–2003CrossRef
31.
Zurück zum Zitat Bantis LE, Nakas CT, Reiser B (2014) Construction of confidence regions in the ROC space after the estimation of the optimal Youden index-based cut-off point. Biometrics 70:212–223CrossRef Bantis LE, Nakas CT, Reiser B (2014) Construction of confidence regions in the ROC space after the estimation of the optimal Youden index-based cut-off point. Biometrics 70:212–223CrossRef
32.
Zurück zum Zitat Shetty R, Rao H, Khamar P, Sainani K, Vunnava K, Jayadev C, Kaweri L (2017) Keratoconus screening indices and their diagnostic ability to distinguish normal from ectatic corneas. Am J Ophthalmol 181:140–148CrossRef Shetty R, Rao H, Khamar P, Sainani K, Vunnava K, Jayadev C, Kaweri L (2017) Keratoconus screening indices and their diagnostic ability to distinguish normal from ectatic corneas. Am J Ophthalmol 181:140–148CrossRef
33.
Zurück zum Zitat Jafarinasab MR, Shirzadeh E, Feizi S, Karimian F, Akaberi A, Hasanpour H (2015) Sensitivity and specificity of posterior and anterior corneal elevation measured by orbscan in diagnosis of clinical and subclinical keratoconus. J Ophthalmic Vis Res 10:10–15CrossRef Jafarinasab MR, Shirzadeh E, Feizi S, Karimian F, Akaberi A, Hasanpour H (2015) Sensitivity and specificity of posterior and anterior corneal elevation measured by orbscan in diagnosis of clinical and subclinical keratoconus. J Ophthalmic Vis Res 10:10–15CrossRef
Metadaten
Titel
Early diagnosis of subclinical keratoconus by wavefront parameters using Scheimpflug, Placido and Hartmann–Shack based devices
verfasst von
Zahra Heidari
Mehrdad Mohammadpour
Hassan Hashemi
Ebrahim Jafarzadehpur
Alireza Moghaddasi
Mehdi Yaseri
Akbar Fotouhi
Publikationsdatum
27.03.2020
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 7/2020
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-020-01334-3

Weitere Artikel der Ausgabe 7/2020

International Ophthalmology 7/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.