Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 10/2009

01.10.2009 | Trauma Surgery

Early results in the treatment of proximal humeral fractures with a polyaxial locking plate

verfasst von: Johannes B. Erhardt, G. Roderer, K. Grob, T. N. Forster, K. Stoffel, M. S. Kuster

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 10/2009

Einloggen, um Zugang zu erhalten

Abstract

Objectives

We report early results using a second generation locking plate, non-contact bridging plate (NCB PH®, Zimmer Inc. Warsaw, IN, USA), for the treatment of proximal humeral fractures. The NCB PH® combines conventional plating technique with polyaxial screw placement and angular stability.

Design

Prospective case series.

Setting

A single level-1 trauma center.

Patients

A total of 50 patients with proximal humeral fractures were treated from May 2004 to December 2005.

Intervention

Surgery was performed in open technique in all cases.

Main outcome measures

Implant-related complications, clinical parameters (duration of surgery, range of motion, Constant–Murley Score, subjective patient satisfaction, complications) and radiographic evaluation [union, implant loosening, implant-related complications and avascular necrosis (AVN) of the humeral head] at 6, 12 and 24 weeks.

Results

All fractures available to follow-up (48 of 50) went to union within the follow-up period of 6 months. One patient was lost to follow-up, one patient died of a cause unrelated to the trauma, four patients developed AVN with cutout, one patient had implant loosening, three patients experienced cutout and one patient had an axillary nerve lesion (onset unknown). The average age- and gender-related Constant Score (n = 35) was 76.

Conclusions

The NCB PH® combines conventional plating technique with polyaxial screw placement and angular stability. Although the complication rate was 19%, with a reoperation rate of 12%, the early results show that the NCB PH® is a safe implant for the treatment of proximal humeral fractures.
Literatur
1.
Zurück zum Zitat Bathis H, Tingart M, Bouillon B, Tiling T (2001) Surgical treatment of proximal humeral fractures. Is the T-plate still adequate osteosynthesis procedure? Zentralbl Chir 126(3):211–216. doi:10.1055/s-2001-12498 PubMedCrossRef Bathis H, Tingart M, Bouillon B, Tiling T (2001) Surgical treatment of proximal humeral fractures. Is the T-plate still adequate osteosynthesis procedure? Zentralbl Chir 126(3):211–216. doi:10.​1055/​s-2001-12498 PubMedCrossRef
2.
Zurück zum Zitat Charalambous CP, Siddique I, Valluripalli K, Kovacevic M, Panose P, Srinivasan M, Marynissen H (2007) Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures. Arch Orthop Trauma Surg 127(3):205–210. doi:10.1007/s00402-006-0256-9 PubMedCrossRef Charalambous CP, Siddique I, Valluripalli K, Kovacevic M, Panose P, Srinivasan M, Marynissen H (2007) Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures. Arch Orthop Trauma Surg 127(3):205–210. doi:10.​1007/​s00402-006-0256-9 PubMedCrossRef
3.
Zurück zum Zitat Constant CR (1997) An evaluation of the Constant–Murley shoulder assessment. J Bone Joint Surg Br 79(4):695–696PubMed Constant CR (1997) An evaluation of the Constant–Murley shoulder assessment. J Bone Joint Surg Br 79(4):695–696PubMed
4.
Zurück zum Zitat Constant CR, Murley AH (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed Constant CR, Murley AH (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed
5.
Zurück zum Zitat Edwards SL, Wilson NA, Zhang LQ, Flores S, Merk BR (2006) Two-part surgical neck fractures of the proximal part of the humerus. A biomechanical evaluation of two fixation techniques. J Bone Joint Surg Am 88(10):2258–2264. doi:10.2106/JBJS.E.00757 PubMedCrossRef Edwards SL, Wilson NA, Zhang LQ, Flores S, Merk BR (2006) Two-part surgical neck fractures of the proximal part of the humerus. A biomechanical evaluation of two fixation techniques. J Bone Joint Surg Am 88(10):2258–2264. doi:10.​2106/​JBJS.​E.​00757 PubMedCrossRef
7.
Zurück zum Zitat Gardner MJ, Griffith MH, Demetrakopoulos D, Brophy RH, Grose A, Helfet DL, Lorich DG (2006) Hybrid locked plating of osteoporotic fractures of the humerus. J Bone Joint Surg Am 88(9):1962–1967. doi:10.2106/JBJS.E.00893 PubMedCrossRef Gardner MJ, Griffith MH, Demetrakopoulos D, Brophy RH, Grose A, Helfet DL, Lorich DG (2006) Hybrid locked plating of osteoporotic fractures of the humerus. J Bone Joint Surg Am 88(9):1962–1967. doi:10.​2106/​JBJS.​E.​00893 PubMedCrossRef
11.
Zurück zum Zitat Hente R, Kampshoff J, Kinner B, Fuchtmeier B, Nerlich M (2004) Treatment of dislocated 3- and 4-part fractures of the proximal humerus with an angle-stabilizing fixation plate. Unfallchirurg 107(9):769–782. doi:10.1007/s00113-004-0818-7 PubMedCrossRef Hente R, Kampshoff J, Kinner B, Fuchtmeier B, Nerlich M (2004) Treatment of dislocated 3- and 4-part fractures of the proximal humerus with an angle-stabilizing fixation plate. Unfallchirurg 107(9):769–782. doi:10.​1007/​s00113-004-0818-7 PubMedCrossRef
13.
Zurück zum Zitat Hessmann M, Baumgaertel F, Gehling H, Klingelhoeffer I, Gotzen L (1999) Plate fixation of proximal humeral fractures with indirect reduction: surgical technique and results utilizing three shoulder scores. Injury 30(7):453–462. doi:10.1016/S0020-1383(99)00111-4 PubMedCrossRef Hessmann M, Baumgaertel F, Gehling H, Klingelhoeffer I, Gotzen L (1999) Plate fixation of proximal humeral fractures with indirect reduction: surgical technique and results utilizing three shoulder scores. Injury 30(7):453–462. doi:10.​1016/​S0020-1383(99)00111-4 PubMedCrossRef
15.
Zurück zum Zitat Kettler M, Biberthaler P, Braunstein V, Zeiler C, Kroetz M, Mutschler W (2006) Treatment of proximal humeral fractures with the PHILOS angular stable plate: presentation of 225 cases of dislocated fractures. Unfallchirurg 109(12):1032–1040. doi:10.1007/s00113-006-1165-7 PubMedCrossRef Kettler M, Biberthaler P, Braunstein V, Zeiler C, Kroetz M, Mutschler W (2006) Treatment of proximal humeral fractures with the PHILOS angular stable plate: presentation of 225 cases of dislocated fractures. Unfallchirurg 109(12):1032–1040. doi:10.​1007/​s00113-006-1165-7 PubMedCrossRef
17.
Zurück zum Zitat Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN (2003) Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg 123(2–3):74–81PubMed Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN (2003) Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg 123(2–3):74–81PubMed
18.
22.
Zurück zum Zitat Robinson CM, Page RS (2004) Severely impacted valgus proximal humeral fractures. J Bone Joint Surg Am 86-A(Suppl 1(Pt 2)):143–155PubMed Robinson CM, Page RS (2004) Severely impacted valgus proximal humeral fractures. J Bone Joint Surg Am 86-A(Suppl 1(Pt 2)):143–155PubMed
23.
Zurück zum Zitat Roderer G, Gebhard F, Erhardt J, Abouelsoud M, Kinzl L (2007) The non-contact bridging plate: a new fixed-angle device for the minimally-invasive treatment of fractures of the proximal humerus—technique and preliminary results. Unfallchirurg 110(6):505–512PubMedCrossRef Roderer G, Gebhard F, Erhardt J, Abouelsoud M, Kinzl L (2007) The non-contact bridging plate: a new fixed-angle device for the minimally-invasive treatment of fractures of the proximal humerus—technique and preliminary results. Unfallchirurg 110(6):505–512PubMedCrossRef
24.
Metadaten
Titel
Early results in the treatment of proximal humeral fractures with a polyaxial locking plate
verfasst von
Johannes B. Erhardt
G. Roderer
K. Grob
T. N. Forster
K. Stoffel
M. S. Kuster
Publikationsdatum
01.10.2009
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 10/2009
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-009-0924-7

Weitere Artikel der Ausgabe 10/2009

Archives of Orthopaedic and Trauma Surgery 10/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.