Skip to main content
Erschienen in: BMC Ophthalmology 1/2020

Open Access 01.12.2020 | Research article

Effect of biomechanical properties on myopia: a study of new corneal biomechanical parameters

verfasst von: Fang Han, Mengdi Li, Pinghui Wei, Jiaonan Ma, Vishal Jhanji, Yan Wang

Erschienen in: BMC Ophthalmology | Ausgabe 1/2020

Abstract

Background

To assess the corneal stress-strain index (SSI), which is a marker for material stiffness and corneal biomechanical parameters, in myopic eyes.

Methods

A total of 1054 myopic patients were included in this study. Corneal visualisation Scheimpflug technology was used to measure the SSI. Corneal biomechanics were assessed using the first and second applanation times (A1-and A2-times); maximum deflection amplitude (DefAmax); deflection area (HCDefArea); the highest concavity peak distance (HC-PD), time (HC-time), and deflection amplitude (HC-DefA); integrated radius (IR); whole eye movement (WEM); stiffness parameter (SP-A1;, biomechanically corrected intraocular pressure (BIOP); and Corvis biomechanical index (CBI). Scheimpflug tomography was used to obtain the mean keratometery (Km) and central corneal thickness (CCT). According to the spherical equivalent (SE) (low myopia: SE ≥ − 3.00D and high myopia: SE ≤ − 6.00D.), the suitable patients were divided into two groups.

Results

The mean SSI value was 0.854 ± 0.004. The SSI had a positive correlation with A1-time ((r = 0.272), HC-time (r = 0.218), WEM (r = 0.288), SP-A1 (r = 0.316), CBI (r = 0.199), CCT (r = 0.125), bIOP (r = 0.230), and SE (r = 0.313) (all p-values<0.01). The SSI had a negative correlation with HCDefA (r = − 0.721), HCDefArea (r = − 0.665), HC-PD(r = − 0.597), IR (r = − 0.555), DefAmax (r = − 0.564), and Km (r = − 0.103) (all p-values<0.01). There were significant differences in SSI (t = 8.960, p<0.01) and IR (t = − 3.509, p<0.01) between the low and high myopia groups.

Conclusions

In different grades of myopia, the SSI values were lower in eyes with higher SEs. It indicates that the mechanical strength of the cornea may be compromised in high myopia. The SSI was positively correlated with the spherical equivalent, and it may provide a new way to study the mechanism of myopia.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
A1-time
The first and applanation time
A2-time
The second applanation time
ARTh
Ambrosio relational thickness horizontal
bIOP
Biomechanically corrected intraocular pressure
CBI
Corvis biomechanical index
CH
Corneal hysteresis
Corvis ST
Corneal visualisation Scheimpflug technology
CCT
Central corneal thickness
CI
Confidence interval
DA
Deformation amplitude
DCR
Dynamic corneal response
DefAmax
Maximum deflection amplitude
HCDefArea
Deflection area
IR
Integrated radius
Km
Mean keratometery
IOP
Intraocular pressure
PD
Peak distance
SE
Spherical equivalent
SP-A1
Stiffness parameter
SP-HC
Stiffness parameter at highest concavity
SSI
Stress-strain index
WEM
Whole eye movement

Background

Quantification of corneal biomechanics has helped us to understand the changes in corneal shape and structure after refractive surgery [1, 2]. The corneal stiffness is a recently described index with clinical significance for the detection in patients who are at risk of ectasia development [3, 4]. Previous studies have shown that the biomechanical properties of the cornea are correlated with many factors such as central corneal thickness (CCT) and intraocular pressure (IOP) [5, 6]. By using finite element models of wide ranges of geometries of human eyeballs affected by different levels of intraocular pressure, Eliasy et al. [7] studied clinical data obtained from two large datasets of healthy participants. According to their algorithm, the study produced a material stiffness parameter-stress-strain index (SSI) as a new parameter does not show significant correlation with CCT and IOP.
However, the distribution characteristics and influencing factors of SSI in myopic patients have not been reported. In this study, we aimed to determine the normative values in myopic patients and the effect of SSI on myopia and to assess its possible correlation with other corneal biomechanical parameters.

Methods

Subjects

This was a retrospective clinical study of 1054 myopic participants (464 women and 590 men) who were scheduled to undergo corneal refractive surgery at Tianjin Eye Hospital, Tianjin Medical University between March 2019 and November 2019. Data from a single representative eye per participant (right eye) were used for the analysis.
According to the spherical equivalent (SE; low myopia: SE ≥ − 3.00 D and high myopia: SE ≤ − 6.00 D), suitable patients were divided into two groups. The study protocol followed the tenets of the Declaration of Helsinki and was approved by the ethics committee of Tianjin Eye Hospital. Written informed consent was obtained from all participants before enrolment. Inclusion criteria included: stable refraction for at least 2 years and the absence of ocular inflammation. Patients were asked to refrain from using soft contact lenses for at least 2 weeks and rigid contact lenses for at least 4 weeks. Exclusion criteria included: history of ophthalmic surgery, ocular trauma, keratoconus, glaucoma, diabetes, systemic connective tissue disease and abnormal immune function.

Methods

All the participants underwent comprehensive eye examinations, including uncorrected visual acuity and best corrected visual acuity measurements, subjective refraction, non-contact tonometry, and slit lamp examinations. Corneal thicknesses and mean curvatures were obtained using Scheimpflug imaging (Pentacam, Oculus, Germany). Corneal biomechanical parameters were obtained using the corneal visualisation Scheimpflug technology (Corvis ST) analyser (Oculus, Germany).

Corvis ST and SSI

The CorvisST (software version 1.6r2015) is a non-contact and visual dynamic intraocular pressure analyzer, which integrates ultra-high-speed Seheimpflug technology into a non-contact intraocular pressure measuring instrument to study the whole dynamic process of corneal deformation under external force. The parameters of the deformation process are recorded, and the biomechanical properties of the cornea are analyzed (Fig. 1). The new software used in this study measures new corneal biomechanical index SSI, included in the dynamic corneal response (DCR) parameters, which consist of first applanation (A1) parameters (A1-time, A1-length and A1-velocity), second applanation (A2) parameters (A2-time, A2-length and A2-velocity), highest concavity (HC) parameters (HC-time, HC-radius, HC deformation amplitude (DA), HC peak distance (PD), maximum deflection amplitude (DefAmax), deflection area (HCDefArea), corneal stiffness parameter (SP-A1), biomechanically corrected intraocular pressure (bIOP), and Corvis biomechanical index (CBI) as well.
The value of SSI was estimated by an in-built software using the least squares method, according to numerical modelling using CCT, bIOP, and SP-HC as input and output parameters [7]. It fits the data with the previous numerical analysis results by the regression equation.
$$ \mathrm{SSI}=\mathrm{f}\ \Big(\mathrm{a}1+\mathrm{a}2\mathrm{C}1+\mathrm{a}3\mathrm{C}2+\mathrm{a}4{\mathrm{C}}_1^2+\mathrm{a}5\mathrm{C}1\mathrm{C}2+\mathrm{a}6{\mathrm{C}}_2^2+\mathrm{a}7{\mathrm{C}}_1^3+\mathrm{a}8{\mathrm{C}}_1^2\mathrm{C}2+\mathrm{a}9\mathrm{C}1{\mathrm{C}}_2^2+{\mathrm{C}}_2^3+\ln \left(\mathrm{SP}-\mathrm{HC}\right) $$
where C1 = CCT/545 and C2 = bIOP/20. ln (SP-HC) the natural logarithm of the SP at HC, and a1-a9 constants are determined by fitting the equation to the numerical input and output values. The ssi was considered as 1.0 for the average experimental behaviour obtained for corneal tissue with age = 50 years [8]. Higher values of SSI are indicative of higher tissue stiffness and vice versa.

Statistical analyses

The statistical analyses were performed using SPSS version 26.0 software (IBM, Corp, Armonk, NY, USA). Descriptive statistical results included means, standard deviations, and minimum and maximum values of parameters. The 95% confidence interval (CI) of the overall mean of the parameters was calculated. Normality of all data samples was checked using the Kolmogorov-Smirnov test. Pearson bivariate correlation statistical analysis was used to obtain the linear fit of the correlation among variables. Stepwise multivariate linear regression analysis was applied to assess the correlation between SSI and other corneal properties; p<0.05 was considered to be statistically significant.

Results

Baseline characteristics

Data were collected from 1054 patients. The right eyes were used for the analysis. The characteristics of the participants are summarised in Table 1. The mean age and SE of the participants were 23.9 ± 5.96 years and − 5.03 ± 2.03D, respectively.
Table 1
Characteristics of the participants included in the study
Parameters
Mean ± SD
Range
Age (years)
23.9 ± 5.96
17–45
MRSE (D)
−5.36 ± 2.07
− 0.5- -14.25
Spherical (D)
−5.03 ± 2.03
−0.5- -13.50
Cylinder (D)
−0.66 ± 0.48
0- -1.75
CCT (microns)
553.4 ± 29.9
482–654
Km (D)
43.14 ± 1.35
39.0–46.75
IOPnct (mmHg)
16.39 ± 2.37
9.5–27.5
Values are presented as means (standard deviations) or as ranges
CCT Central corneal thickness, Km Mean keratometry, MRSE Manifest refraction spherical equivalent, IOPnct Intraocular pressure with non-contact tonometry, SD Standard deviation

Biomechanical parameters

The mean values of DCR parameters in eyes with corresponding standard deviations and 95% CIs are shown in Table 2.
Table 2
Distribution of normative values of Corvis ST parameters
Parameters
Mean ± SD
95% CI
A1-time (ms)
7.208 ± 0.289
7.194,7.223
A2-time (ms)
21.968 ± 0.373
21.949,21.987
HC-time (ms)
17.032 ± 0.401
17.011,17.052
A1DefA (mm)
0.097 ± 0.006
0.096,0.097
A2DefA (mm)
0.109 ± 0.011
0.108,0.109
HCDefA (mm)
0.941 ± 0.096
0.936,0.946
HCDefArea (mm2)
3.521 ± 0.501
3.495,3.547
DefAmax (mm)
0.953 ± 0.095
0.948,0.957
DAmax (mm)
4.349 ± 0.424
4.328,4.372
PD (mm)
5.181 ± 0.251
5.169,5.194
WEM (mm)
0.201 ± 0.065
0.258,0.265
SP-A1
107.866 ± 14.927
107.091,108.640
ARTh
581.715 ± 112.10
575.918,587.512
IR
7.902 ± 0.977
7.852,7.951
CBI
0.221 ± 0.187
0.212,0.230
SSI
0.854 ± 0.133
0.847,0.860
CCT (μm)
554 ± 32.75
576,587
bIOP (mmHg)
16 ± 2.06
15.9,16.1
Values are presented as means (standard deviations) with 95% confidence intervals
Corvis ST Corneal visualisation Scheimpflug technology, SD Standard deviation, CI Confidence interval, A1-and A2-times Time reaching the first and second applanation, HC-time Highest concavity-time, A1 and A2DefA Displacement of corneal apex at the first or second applanation or at the moment of highest concavity after whole eye motion is removed, HCDefA Amplitude at the highest concavity, HCDefArea Deflection area at the highest concavity, DefAmax Maximum deflection amplitude, DAmax Maximum deformation amplitude, PD Peak distance, WEM Whole eye movement, SP-A1 Stiffness parameter, IR Integrated radius, ARTh Ambrosio relational thickness horizontal, CBI Corvis biomechanical index, SSI Stress-strain index, CCT Central corneal thickness, bIOP Biomechanically corrected intraocular pressure

Correlations of SSI with other biomechanical parameters

Correlation between baseline characteristics and SSI

No statistically significant correlations were observed between SSI and sex (p>0.05). The SSI was negatively correlated with mean keratometry (Km) (r = − 0.103, p<0.01) and positively correlated with SE (r = 0.313, p<0.01) (Fig. 2), CCT (r = 0.125, p<0.01), age (r = 0.198, p<0.01), and bIOP (r = 0.23, p<0.01) (Table 3).
Table 3
Correlations between SSI and characteristics of the participants
Parameters
SSI
r
p
Age (years)
0.198
<0.01
MRSE (D)
0.313
<0.01
CCT (microns)
0.125
<0.01
Km (D)
−0.103
<0.01
bIOP (mmHg)
0.230
<0.01
Statistical significance has been defined as p < 0.05
SSI Stress-strain index, MRSE Manifest refraction spherical equivalent, CCT Central corneal thickness, Km Mean keratometry, BIOP Biomechanically corrected intraocular pressure

Comparison of parameters in low and high myopia

There were no significant differences in age, CCT, BIOP, SP, ambrosio relational thickness horizontal (ARTh), and CBI between the low and high myopia groups. There was a significant difference in SSI (t = 8.960, p<0.01) and integrated radius (IR)(t = − 3.509, p<0.01) (Fig. 3, Table 4).
Table 4
Comparison of the parameters in low and high myopia
Group
Number
Age (years)
CCT (microns)
bIOP
SP-A1
DAmax (mm)
IR (mm)
ARTh
CBI
SSI
Low myopia
161
23.2 ± 6.12
553 ± 32.9
16 ± 2.07
108.520 ± 15.699
4.272 ± 0.448
7.663 ± 1.071
589.402 ± 114.500
0.214 ± 0.184
0.920 ± 0.138
High myopia
506
24.0 ± 5.56
551 ± 29.3
16 ± 1.98
107.787 ± 14.884
4.360 ± 0.413
7.953 ± 0.991
577.718 ± 106.814
0.224 ± 0.183
0.813 ± 0.129
t
 
−1.616
0.770
−1.354
0.538
−2.328
−3.509
1.190
−0.609
8.960
p
 
0.107
0.441
0.176
0.590
0.020
<0.01
0.234
0.543
<0.01
Statistical significance has been defined as p < 0.05
CCT Central corneal thickness, bIOP Biomechanically corrected intraocular pressure, SP-A1 Stiffness parameter, DAmax Maximum deformation amplitude, IR Integrated radius, ARTh Ambrosio relational thickness horizontal, CBI Corvis biomechanical index, SSI Stress-strain index

Regression analysis of SSI and baseline characteristics

Multiple linear stepwise regression analysis was performed with SSI as the dependent variable and SE, Km, CCT, bIOP, and age as independent variables (Table 5). The following regression equation was obtained:
$$ \mathrm{SSI}=0.768+0.021\mathrm{SE}+0.02\mathrm{bIOP}+0.006\mathrm{AGE}-0.013\mathrm{KM}+0.001\mathrm{CCT} $$
Table 5
Multiple linear stepwise regression analysis with SSI as the dependent variable
 
β
t
P
95% CI
Constant
0.768
6.429
<0.01
0.534,1.002
MRSE (D)
0.021
14.114
<0.01
0.018,0.024
bIOP
0.020
12.752
<0.01
0.017,0.023
Age (year)
0.006
11.352
<0.01
0.005,0.007
Km (D)
-0.013
−5.680
<0.01
−0.018, − 0.009
CCT (μm)
0.001
5.558
<0.01
0.000,0.001
Values are presented 95% confidence intervals, and statistical significance has been defined as p < 0.05
CI Confidence interval, MRSE Manifest refraction spherical equivalent, bIOP Biomechanically corrected intraocular pressure, Km Mean keratometry, CCT Central corneal thickness, SSI Stress-strain index

Correlation between corneal biomechanical parameters and SSI

A1-time, HC-time, A2DefA, WEM, SP-A1, ARTh, CBI, IR, and bIOP were weak positively correlated with SSI. DAmax, A2-time, and A2DefA were weakly negatively correlated with SSI. HCDefA, HCDefArea, PD, IR, and DefAmax were strongly negatively correlated with SSI. No significant correlation was found between A1DefA and SSI (Table 6, Fig. 4).
Table 6
Correlations between SSI and Corvis ST parameters
Parameters
SSI
R
p
A1-time (ms)
0.272
<0.01
A2-time (ms)
−0.323
<0.01
HC-time (ms)
0.218
<0.01
A1DefA (mm)
−0.007
0.798
A2DefA (mm)
0.081
<0.01
HCDefA (mm)
−0.721
<0.01
HCDefArea (mm2)
−0.665
<0.01
DefAmax (mm)
−0.564
<0.01
DAmax (mm)
−0.388
<0.01
PD (mm)
−0.597
<0.01
WEM (mm)
0.288
<0.01
SP-A1
0.316
<0.01
ARTh
0.113
<0.01
IR
−0.555
<0.01
CBI
0.199
<0.01
bIOP
0.230
<0.01
Statistical significance has been defined as p < 0.05
SSI Stress-strain index, Corvis ST Corneal visualiation Scheimpflug technology, A1-and A2-times Time reaching the first and second applanation, HC-time Highest concavity-time, A1 and A2DefA Displacement of corneal apex at the first or second applanation or at the moment of highest concavity after whole eye motion is removed, HCDefA Amplitude at the highest concavity, HCDefArea Deflection area at the highest concavity, DefAmax Maximum deflection amplitude, DAmax Maximum deformation amplitude, PD Peak distance, WEM Whole eye movement, SP-A1 Stiffness parameter, IR Integrated radius, ARTh Ambrosio relational thickness horizontal, CBI Corvis biomechanical index, bIOP Biomechanically corrected intraocular pressure

Discussion

Myopia is a public health problem [9] with a high prevalence, especially in the Far East [10]. The onset and progression of myopia has been associated with genetic and environmental factors [11, 12]. Previous studies have noted that myopia is correlated with an increase in corneal curvature and a decrease in corneal thickness [13]. Animal studies have shown a change in the length of the eye and shape of the anterior cornea during the process of myopia modelling [14, 15]. It is also known that high myopes have lower corneal hysteresis than emmetropes [16]. However, it is difficult to detect the ocular biomechanical properties in vivo [17]. The Corvis ST provides information on corneal deformation parameters by visualising the dynamic reaction of the cornea to a single puff of air [18].
In our study, one of the new Corvis ST parameters—SSI, was evaluated in myopic eyes. We demonstrated that the SSI was positively correlated with SE (r = 0.313, p<0.01) (Fig. 1). When comparing eyes with low and high myopia, there were no significant differences in CCT, bIOP, SP, ARTh, and CBI, although there was a significant difference in SSI(t = 8.960, p<0.01) and IR(t = − 3.509, p<0.01 (Fig. 2, Table 4) values. The results showed that the SSI of high myopia was lower than that of low myopia, suggesting that the biomechanical properties of the cornea changed and corneal hardness decreased with an increase in the SE.
Inmaculada Bueno-Gimeno et al. used ocular response analyser and suggested that corneal biomechanical properties appear to be compromised in myopia from an early age, especially in high myopia [19]. Another study showed a weak although significant correlation between corneal hysteresis (CH) and refractive error, with CH being lower in both moderate and high myopia than in emmetropia and low myopia [20]. Wu et al. [21] reported a difference in corneal biomechanical properties between 835 low myopic eyes and 1027 high myopic eyes. Low CH and corneal resistance factor and high cornea-compensated and Goldmann-correlated IOPs were suggested to be associated with high myopia. However, the correlation of the biomechanics of myopia is controversial. Some studies reported no significant correlation between myopia and CH [22, 23]. The results of our study showed a strong negative correlation of SSI with HCDefA(r = − 0.721, p<0.01), HCDefArea(r = − 0.665, p<0.01), PD(r = − 0.597, p<0.01), IR(r = − 0.555, p<0.01), and DefAmax(r = − 0.564, p<0.01) (Table 6, Fig. 3). Wang et al. [24] found that eyes with high myopia had a larger corneal DA than eyes with mild-to-moderate myopia, and A2-time and HC-radius were positively correlated with SE. Eyes with high myopia also showed longer DA and smaller HC-radius. Similar results were reported by Miaohe et al. [5]. These findings are consistent with our results.
Previous studies have shown that the biomechanical properties of the cornea are correlated with CCT. Eyes with thick CCT exhibited strong corneal resistance to external force and are less prone to deformation [25]. Higher intraocular pressure may mask abnormal corneal biomechanical properties, resulting in apparently normal HCDA measurements [26]. The introduction of SSI resolved this issue because it estimates the material stiffness [27]. Eliasy et al. [7] used a numerical model in the study of SSI parameter, which not only covered a wide range of changes in IOP, CCT, geometry and material parameters, but also covered even slightly extended beyond the scope reported in clinical studies. Through the consideration of a Corvis parameter—SP-HC, it is more strongly correlated with corneal stiffness than IOP. As a new index independent of IOP and corneal geometry, SSI could specifically detect high-risk or susceptible patients with ectasia after refractive surgery, remind physicians of the possible risks caused by the decrease of corneal biomechanical properties and could aid in surgery planning.
In this study, we observed a weak correlation among BIOP (r = 0.23, p<0.01), CCT (r = 0.125, p<0.01), and SSI (Table 3). It indicated that despite correction, the effect of corneal biomechanics cannot be completely independent of IOP and thickness, which is consistent with our clinical experience and previous studies. Effects of IOP and corneal biomechanics on eye behaviour are difficult to separate; IOP also affects the immediate corneal stiffness. It is generally believed that sex has no significant effect on corneal biomechanics [28]. Our study also suggested that the corneal biomechanical properties of myopia may have nothing to do with sex. Correlation between stress-strain behaviour and age was reported [8, 29], although this study found that there was no strong significant correlation between age (r = 0.198, p<0.01) and SSI, which may be correlated with the concentration of the individual age included.
It is noteworthy that the mean CCT value measured by Corvis ST (553 ± 29.96 um) was slightly lower than that measured by Pentacam corneal topography (554 ± 31.04 um) (t = 4.970 p<0.01). However, it has been shown that Corvis-ST CCT measurements have good repeatability [30].
The main limitation of the current study is lack of eye axis parameters and a control group with emmetropia, despite a large sample size with myopic participants, which will be improved and supplemented in future studies.

Conclusions

In conclusion, our study showed that there was a positive correlation between SSI and SE. It may provide a new way to study the mechanism of myopia. In different grades of myopia, the SSI values were lower in eyes with higher SE. This indicates that the mechanical strength of the cornea may by compromised in high myopia. Future studies can corroborate the findings of our study. A longitudinal study in progressive and stable myopic participants is warranted.

Acknowledgements

Not applicable.
The study adhered to the tenets of the Declaration of Helsinki and was approved by the ethics committee at Tianjin Eye Hospital. Additionally, written informed consent was obtained from each participant.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Oner V, Tas M, Ozkaya E, Oruc Y. Effect of pathological myopia on biomechanical properties: a study by ocular response analyzer. Int J Ophthalmol. 2015;8(2):365–8.PubMedPubMedCentral Oner V, Tas M, Ozkaya E, Oruc Y. Effect of pathological myopia on biomechanical properties: a study by ocular response analyzer. Int J Ophthalmol. 2015;8(2):365–8.PubMedPubMedCentral
2.
Zurück zum Zitat Blum M, Täubig K, Gruhn C, Sekundo W, Kunert KS. Five-year results of small incision lenticule extraction (ReLEx SMILE). Br J Ophthalmol. 2016;100(9):1192–5.CrossRef Blum M, Täubig K, Gruhn C, Sekundo W, Kunert KS. Five-year results of small incision lenticule extraction (ReLEx SMILE). Br J Ophthalmol. 2016;100(9):1192–5.CrossRef
3.
Zurück zum Zitat Tian L, Ko MW, Wang LK, Zhang JY, Li TJ, Huang YF, Zheng YP. Assessment of ocular biomechanics using dynamic ultra high-speed Scheimpflug imaging in keratoconic and normal eyes. J Refract Surg. 2014;30(11):785–91.CrossRef Tian L, Ko MW, Wang LK, Zhang JY, Li TJ, Huang YF, Zheng YP. Assessment of ocular biomechanics using dynamic ultra high-speed Scheimpflug imaging in keratoconic and normal eyes. J Refract Surg. 2014;30(11):785–91.CrossRef
4.
Zurück zum Zitat Qiu K, Zhang R, Wang G, Zhang M. Relationship of corneal hysteresis and optic nerve parameters in healthy myopic subjects. Sci Rep. 2017;7(1):17538–7.CrossRef Qiu K, Zhang R, Wang G, Zhang M. Relationship of corneal hysteresis and optic nerve parameters in healthy myopic subjects. Sci Rep. 2017;7(1):17538–7.CrossRef
5.
Zurück zum Zitat He M, Wang W, Ding H, Zhong X. Corneal biomechanical properties in high myopia measured by dynamic Scheimpflug imaging technology. Optom Vis Sci. 2017;94(12):1074–80.CrossRef He M, Wang W, Ding H, Zhong X. Corneal biomechanical properties in high myopia measured by dynamic Scheimpflug imaging technology. Optom Vis Sci. 2017;94(12):1074–80.CrossRef
6.
Zurück zum Zitat Jia X, Yu J, Liao SH, Duan XC. Biomechanics of the sclera and effects on intraocular pressure. Int J Ophthalmol. 2016;9(12):1824–31.PubMedPubMedCentral Jia X, Yu J, Liao SH, Duan XC. Biomechanics of the sclera and effects on intraocular pressure. Int J Ophthalmol. 2016;9(12):1824–31.PubMedPubMedCentral
7.
Zurück zum Zitat Xu H, Zong Y, Zhai R, Kong X, Jiang C, Sun X. Intereye and intraeye asymmetry analysis of retinal microvascular and neural structure parameters for diagnosis of primary open-angle glaucoma. Eye (Lond). 2019;33(10):1596–605.CrossRef Xu H, Zong Y, Zhai R, Kong X, Jiang C, Sun X. Intereye and intraeye asymmetry analysis of retinal microvascular and neural structure parameters for diagnosis of primary open-angle glaucoma. Eye (Lond). 2019;33(10):1596–605.CrossRef
8.
Zurück zum Zitat Elsheikh A, Geraghty B, Rama P, Campanelli M, Meek KM. Characterization of age-related variation in corneal biomechanical properties. J R Soc Interface. 2010;7(51):1475–85.CrossRef Elsheikh A, Geraghty B, Rama P, Campanelli M, Meek KM. Characterization of age-related variation in corneal biomechanical properties. J R Soc Interface. 2010;7(51):1475–85.CrossRef
9.
Zurück zum Zitat Holden BA, Fricke TR, Wilson DA, Jong M, Naidoo KS, Sankaridurg P, Wong TY, Naduvilath TJ, Resnikoff S. Global prevalence of myopia and high myopia and temporal trends from 2000 through 2050. Ophthalmology. 2016;123(5):1036–42.CrossRef Holden BA, Fricke TR, Wilson DA, Jong M, Naidoo KS, Sankaridurg P, Wong TY, Naduvilath TJ, Resnikoff S. Global prevalence of myopia and high myopia and temporal trends from 2000 through 2050. Ophthalmology. 2016;123(5):1036–42.CrossRef
10.
Zurück zum Zitat Wei S, Sun Y, Li S, Hu J, Yang X, Lin C, Cao K, Du J, Guo J, Li H, et al. Refractive errors in university students in Central China: the Anyang University Students Eye Study. Invest Ophthalmol Vis Sci. 2018;59(11):4691–700.CrossRef Wei S, Sun Y, Li S, Hu J, Yang X, Lin C, Cao K, Du J, Guo J, Li H, et al. Refractive errors in university students in Central China: the Anyang University Students Eye Study. Invest Ophthalmol Vis Sci. 2018;59(11):4691–700.CrossRef
11.
Zurück zum Zitat Hung GK, Mahadas K, Mohammad F. Eye growth and myopia development: unifying theory and Matlab model. Comput Biol Med. 2016;70:106–18.CrossRef Hung GK, Mahadas K, Mohammad F. Eye growth and myopia development: unifying theory and Matlab model. Comput Biol Med. 2016;70:106–18.CrossRef
12.
Zurück zum Zitat Lee MW, Kim JM, Shin YI, Jo YJ, Kim JY. Longitudinal changes in peripapillary retinal nerve fiber layer thickness in high myopia: a prospective, observational study. Ophthalmology. 2019;126(4):522–8.CrossRef Lee MW, Kim JM, Shin YI, Jo YJ, Kim JY. Longitudinal changes in peripapillary retinal nerve fiber layer thickness in high myopia: a prospective, observational study. Ophthalmology. 2019;126(4):522–8.CrossRef
13.
Zurück zum Zitat Leung TW, Lam AK, Kee CS. Corneal shapes of Chinese emmetropes and myopic astigmats aged 10 to 45 years. Optom Vis Sci. 2013;90(11):1259–66.CrossRef Leung TW, Lam AK, Kee CS. Corneal shapes of Chinese emmetropes and myopic astigmats aged 10 to 45 years. Optom Vis Sci. 2013;90(11):1259–66.CrossRef
14.
Zurück zum Zitat Cohen Y, Belkin M, Yehezkel O, Avni I, Polat U. Light intensity modulates corneal power and refraction in the chick eye exposed to continuous light. Vis Res. 2008;48(21):2329–35.CrossRef Cohen Y, Belkin M, Yehezkel O, Avni I, Polat U. Light intensity modulates corneal power and refraction in the chick eye exposed to continuous light. Vis Res. 2008;48(21):2329–35.CrossRef
15.
Zurück zum Zitat Rucker F, Britton S, Spatcher M, Hanowsky S. Blue light protects against temporal frequency sensitive refractive changes. Invest Ophthalmol Vis Sci. 2015;56(10):6121–31.CrossRef Rucker F, Britton S, Spatcher M, Hanowsky S. Blue light protects against temporal frequency sensitive refractive changes. Invest Ophthalmol Vis Sci. 2015;56(10):6121–31.CrossRef
16.
Zurück zum Zitat Wong YZ, Lam AK. The roles of cornea and axial length in corneal hysteresis among emmetropes and high myopes: a pilot study. Curr Eye Res. 2015;40(3):282–9.CrossRef Wong YZ, Lam AK. The roles of cornea and axial length in corneal hysteresis among emmetropes and high myopes: a pilot study. Curr Eye Res. 2015;40(3):282–9.CrossRef
17.
Zurück zum Zitat Eliasy A, Chen KJ, Vinciguerra R, Maklad O, Vinciguerra P, Ambrósio R Jr, Roberts CJ, Elsheikh A. Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST. Exp Eye Res. 2018;175:98–102.CrossRef Eliasy A, Chen KJ, Vinciguerra R, Maklad O, Vinciguerra P, Ambrósio R Jr, Roberts CJ, Elsheikh A. Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST. Exp Eye Res. 2018;175:98–102.CrossRef
18.
Zurück zum Zitat Roberts CJ. Concepts and misconceptions in corneal biomechanics. J Cataract Refract Surg. 2014;40(6):862–9.CrossRef Roberts CJ. Concepts and misconceptions in corneal biomechanics. J Cataract Refract Surg. 2014;40(6):862–9.CrossRef
19.
Zurück zum Zitat Bueno-Gimeno I, España-Gregori E, Gene-Sampedro A, Lanzagorta-Aresti A, Piñero-Llorens DP. Relationship among corneal biomechanics, refractive error, and axial length. Optom Vis Sci. 2014;91(5):507–13.CrossRef Bueno-Gimeno I, España-Gregori E, Gene-Sampedro A, Lanzagorta-Aresti A, Piñero-Llorens DP. Relationship among corneal biomechanics, refractive error, and axial length. Optom Vis Sci. 2014;91(5):507–13.CrossRef
20.
Zurück zum Zitat Shen M, Fan F, Xue A, Wang J, Zhou X, Lu F. Biomechanical properties of the cornea in high myopia. Vis Res. 2008;48(21):2167–71.CrossRef Shen M, Fan F, Xue A, Wang J, Zhou X, Lu F. Biomechanical properties of the cornea in high myopia. Vis Res. 2008;48(21):2167–71.CrossRef
21.
Zurück zum Zitat Ha A, Kim YK, Baek SU, Park KH, Jeoung JW. Optic disc microhemorrhage in primary open-angle glaucoma: clinical implications for visual field progression. Invest Ophthalmol Vis Sci. 2019;60(6):1824–32.CrossRef Ha A, Kim YK, Baek SU, Park KH, Jeoung JW. Optic disc microhemorrhage in primary open-angle glaucoma: clinical implications for visual field progression. Invest Ophthalmol Vis Sci. 2019;60(6):1824–32.CrossRef
22.
Zurück zum Zitat Lim L, Gazzard G, Chan YH, Fong A, Kotecha A, Sim EL, Tan D, Tong L, Saw SM. Cornea biomechanical characteristics and their correlates with refractive error in Singaporean children. Investig Ophthalmol Vis Sci. 2008;49(9):3852–7.CrossRef Lim L, Gazzard G, Chan YH, Fong A, Kotecha A, Sim EL, Tan D, Tong L, Saw SM. Cornea biomechanical characteristics and their correlates with refractive error in Singaporean children. Investig Ophthalmol Vis Sci. 2008;49(9):3852–7.CrossRef
23.
Zurück zum Zitat Kamiya K, Hagishima M, Fujimura F, Shimizu K. Factors affecting corneal hysteresis in normal eyes. Graefes Arch Clin Exp Ophthalmol. 2008;246(10):1491–4.CrossRef Kamiya K, Hagishima M, Fujimura F, Shimizu K. Factors affecting corneal hysteresis in normal eyes. Graefes Arch Clin Exp Ophthalmol. 2008;246(10):1491–4.CrossRef
24.
Zurück zum Zitat Wang J, Li Y, Jin Y, Yang X, Zhao C, Long Q. Corneal biomechanical properties in myopic eyes measured by a dynamic Scheimpflug analyzer. J Ophthalmol. 2015;2015:161869.PubMedPubMedCentral Wang J, Li Y, Jin Y, Yang X, Zhao C, Long Q. Corneal biomechanical properties in myopic eyes measured by a dynamic Scheimpflug analyzer. J Ophthalmol. 2015;2015:161869.PubMedPubMedCentral
25.
Zurück zum Zitat Çevik SG, Kıvanç SA, Akova-Budak B, Tok-Çevik M. Relationship among corneal biomechanics, anterior segment parameters, and geometric corneal parameters. J Ophthalmol. 2016;2016:8418613.CrossRef Çevik SG, Kıvanç SA, Akova-Budak B, Tok-Çevik M. Relationship among corneal biomechanics, anterior segment parameters, and geometric corneal parameters. J Ophthalmol. 2016;2016:8418613.CrossRef
26.
Zurück zum Zitat Ma J, Wang Y, Hao W, Jhanji V. Comparative analysis of biomechanically corrected intraocular pressure with corneal visualization Scheimpflug technology versus conventional noncontact intraocular pressure. Int Ophthalmol. 2020;40(1):117–24.CrossRef Ma J, Wang Y, Hao W, Jhanji V. Comparative analysis of biomechanically corrected intraocular pressure with corneal visualization Scheimpflug technology versus conventional noncontact intraocular pressure. Int Ophthalmol. 2020;40(1):117–24.CrossRef
27.
Zurück zum Zitat Vinciguerra R, Ambrosio RJ, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, Azzolini C, Vinciguerra P. Detection of Keratoconus with a new biomechanical index. J Refract Surg. 2016;32(12):803–10.CrossRef Vinciguerra R, Ambrosio RJ, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, Azzolini C, Vinciguerra P. Detection of Keratoconus with a new biomechanical index. J Refract Surg. 2016;32(12):803–10.CrossRef
28.
Zurück zum Zitat Shah S, Laiquzzaman M, Yeung I, Pan X, Roberts C. The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery. Cont Lens Anterior Eye. 2009;32(3):123–8.CrossRef Shah S, Laiquzzaman M, Yeung I, Pan X, Roberts C. The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery. Cont Lens Anterior Eye. 2009;32(3):123–8.CrossRef
29.
Zurück zum Zitat Geraghty B, Jones SW, Rama P, Akhtar R, Elsheikh A. Age-related variations in the biomechanical properties of human sclera. J Mech Behav Biomed Mater. 2012;16:181–91.CrossRef Geraghty B, Jones SW, Rama P, Akhtar R, Elsheikh A. Age-related variations in the biomechanical properties of human sclera. J Mech Behav Biomed Mater. 2012;16:181–91.CrossRef
30.
Zurück zum Zitat Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55(6):3651–9.CrossRef Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55(6):3651–9.CrossRef
Metadaten
Titel
Effect of biomechanical properties on myopia: a study of new corneal biomechanical parameters
verfasst von
Fang Han
Mengdi Li
Pinghui Wei
Jiaonan Ma
Vishal Jhanji
Yan Wang
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2020
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-020-01729-x

Weitere Artikel der Ausgabe 1/2020

BMC Ophthalmology 1/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.