Effective luminance deterioration of medical liquid crystal displays in clinical use

verfasst von: Keita Takahashi, Shinichi Awamoto, Shinya Takarabe, Kazuhisa Ogawa, Yasuhiko Nakamura

Erschienen in: Radiological Physics and Technology | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

The objective of this work was to evaluate the maximum luminance (L max) level of medical liquid crystal displays (LCDs) as a function of backlight hours (BLH) annually. The L max values for 249 2-megapixel color LCDs (RadiForce RX210, EIZO Corporation) were measured in February 2014, 2015, and 2016. Four near-range luminance meters and the built-in type luminance meters, each with an LCD, were used for the measurements. The average and standard deviation (SD) of BLH measured in 2014 was 15,371 ± 8219 h. Four, twenty, and thirty-nine LCDs failed in the constancy tests performed in February 2014, 2015, and 2016, respectively, i.e., they were unable to output 170 cd/m2. The SD of L max increased each year and as BLH became longer. In conclusion, evaluation of L max as a function of BLH during constancy testing will help predict the decrease in L max of a clinically used medical color LCD.
Literatur
1.
Zurück zum Zitat Silosky M, Marsh RM. Constancy of built-in luminance meter measurements in diagnostic displays. Med Phys. 2013;40(12):121902.CrossRefPubMed Silosky M, Marsh RM. Constancy of built-in luminance meter measurements in diagnostic displays. Med Phys. 2013;40(12):121902.CrossRefPubMed
2.
Zurück zum Zitat Takarabe S, Morishita J, Yabuuchi H, et al. A preliminary study for exploring the luminance ratio of liquid-crystal displays required for display of radiographs. Radiolphystechnol. 2014;7(1):73–8. Takarabe S, Morishita J, Yabuuchi H, et al. A preliminary study for exploring the luminance ratio of liquid-crystal displays required for display of radiographs. Radiolphystechnol. 2014;7(1):73–8.
3.
Zurück zum Zitat Lehmann W. Phosphor deterioration in fluorescent lamps. J ElectrochemSoc. 1983;130(2):426.CrossRef Lehmann W. Phosphor deterioration in fluorescent lamps. J ElectrochemSoc. 1983;130(2):426.CrossRef
4.
Zurück zum Zitat Hellen-Halme K, Hellen-Halme B, Wenzel A. The effect of aging on luminance of standard liquid crystal display (LCD) monitors. Oral Surg Oral Med Oral Pathol Oral Radiol Endodontol. 2011;112(2):237–42.CrossRef Hellen-Halme K, Hellen-Halme B, Wenzel A. The effect of aging on luminance of standard liquid crystal display (LCD) monitors. Oral Surg Oral Med Oral Pathol Oral Radiol Endodontol. 2011;112(2):237–42.CrossRef
5.
Zurück zum Zitat National Electrical Manufacturers Association (NEMA). Digital imaging and Communications in Medicine (DICOM) Part 14: grayscale standard display function PS 3.14 2011. Rosslyn, VA: NEMA. National Electrical Manufacturers Association (NEMA). Digital imaging and Communications in Medicine (DICOM) Part 14: grayscale standard display function PS 3.14 2011. Rosslyn, VA: NEMA.
8.
Zurück zum Zitat Krupinski EA, Roehrig H, Fan J. Does the age of liquid crystal displays influence observer performance? Acad Radiol. 2007;14(4):463–67.CrossRefPubMed Krupinski EA, Roehrig H, Fan J. Does the age of liquid crystal displays influence observer performance? Acad Radiol. 2007;14(4):463–67.CrossRefPubMed
Metadaten
Titel
Effective luminance deterioration of medical liquid crystal displays in clinical use
verfasst von
Keita Takahashi
Shinichi Awamoto
Shinya Takarabe
Kazuhisa Ogawa
Yasuhiko Nakamura
Publikationsdatum
02.05.2017
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 3/2017
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-017-0401-x

Weitere Artikel der Ausgabe 3/2017

Radiological Physics and Technology 3/2017 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.