Open Access 01.12.2023 | Research

Evaluation of human T-cell leukemia virus in vitro diagnostics using plasma specimens collected in Japan

verfasst von: Shigeru Kusagawa, Ai Kawana-Tachikawa, Keiji Matsubayashi, Rieko Sobata, Isao Hamaguchi

Erschienen in: BMC Infectious Diseases | Ausgabe 1/2023

Abstract

Background

In vitro diagnostics (IVDs) for primary detection test/screening of human T-cell leukemia virus (HTLV) have recently been updated to new-generation products in Japan. In this study, the performance of these products was evaluated and discussed in terms of the usability of HTLV diagnosis in Japan.

Methods

The performance of 10 HTLV IVDs for primary detection test and confirmatory/discriminatory test was evaluated. Plasma specimens that had been declared ineligible for transfusion were provided by the Japanese Red Cross Blood Center.

Results

The diagnostic specificity of the IVDs was 100% (160/160). Six sandwich assays resulted in all HTLV-1/HTLV-positive specimens being positive (46/46). On the other hand, one sandwich assay, IVD under development 2 (UD2), resulted in one HTLV-1-positive and one HTLV-positive specimen being negative (44/46, 95.7%). One indirect assay, HISCL HTLV-1, could not detect one HTLV-positive specimen (45/46, 97.8%), but the updated product, UD1, correctly detected it (46/46, 100%). Serodia HTLV-I, based on a particle agglutination assay, resulted in 44 of the 46 positive specimens, but could not detect two specimens (44/46, 95.7%). ESPLINE HTLV-I/II, based on an immunochromatography assay (ICA), was able to diagnose all specimens as positive (46/46, 100%).

Conclusions

Six sandwich assays and an ICA demonstrated high diagnostic sensitivity and specificity and are recommended for use in HTLV diagnosis in conjunction with confirmatory/discriminatory test using the INNO-LIA HTLV-I/II Score.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12879-023-08402-w.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
HTLV
Human T-cell leukemia virus
IVD
In vitro diagnostics
S/CO
Signal to cutoff ratio
COI
Cutoff index
ATL
Adult T-cell leukemia
HAM
HTLV-1 associated myelopathy
TSP
Tropical spastic paraparesis
PA
Particle agglutination
ICA
Immunochromatography assay
CLEIA
Chemiluminescent enzyme immunoassay
CLIA
Chemiluminescent immunoassay
ECLIA
Electrochemiluminescence immunoassay
LIA
Line immunoassay
INNO-LIA
INNO-LIA HTLV-I/II Score
UD
IVD under development

Background

Human T-cell leukemia virus (HTLV) belongs to the family Retroviridae, genus Deltaretrovirus. Although most HTLV-1 infections are asymptomatic, some carriers develop adult T-cell leukemia (ATL) [1] and HTLV-1 associated myelopathy/tropical spastic paraparesis (HAM/TSP) [2] after a long asymptomatic period. HTLV-1 infection is endemic in southeast Japan, the Caribbean coast, South America, central Australia, and equatorial Africa [3, 4]. Based on an analysis of Japanese first-time blood donors from 2006 to 2007, approximately 1.08 million individuals were infected with HTLV-1 [5], and the estimated annual number of new HTLV-1 infections was 4,190 at the end of 2011, as reported in a retrospective cohort study of repeat blood donors who were HTLV-1-negative from 2006 to 2007 [6]. HTLV-1 mainly infects vertically through mother-to-infant transmission via breastfeeding, but can also be transmitted sexually or via blood transfusions horizontally. HTLV antibody positivity refers to an HTLV carrier and a source of HTLV infection. The anti-HTLV antibody test is useful for controlling HTLV infections.
A diagnostic test algorithm for HTLV-1 infection was established in Japan [7]. In short, primary detection test/screening of the serum or plasma obtained from peripheral blood specimens is conducted to determine a subject’s HTLV-1 infection status using commercially available HTLV-1 antibody-specific serological assays (particle agglutination (PA), chemiluminescent enzyme immunoassay (CLEIA), chemiluminescent immunoassay (CLIA), and electrochemiluminescence immunoassay (ECLIA)). Positive samples in the primary detection test were tested using a line immunoassay (LIA) as a confirmatory/discriminatory test to exclude pseudo-positive cases, because HTLV-1/2 structural protein-specific antibody reactivity can be detected by LIA. When the result is indeterminate, HTLV-1 PCR is performed to detect proviral DNA [8, 9].
Lately, HTLV in vitro diagnostics (IVDs) for primary detection tests have been updated to new-generation products based on sandwich assay. The HIV sandwich assay shows high performance because using enzymatic or chemically labeled HIV antigen instead of labeled human IgG to detect antigen–antibody complex on a carrier following bound/free separation [10, 11].
In this study, HTLV-positive reference panel, including a low signal-to-cutoff ratio (S/CO) or cutoff index (COI), and HTLV-negative reference panel were recently prepared in Japan. The performance of HTLV diagnostics for primary detection testing was evaluated using the HTLV-positive reference panel and the HTLV-negative reference panel.

Methods

HTLV IVDs

The performance of 10 HTLV IVDs for the primary detection test and one for confirmatory/discriminatory test was evaluated in this study. Two IVDs under development (UD1 and UD2) were included. This list is presented in Table 1.
Table 1
List of HTLV IVDs used in this study
IVDs
Method
Indg/Sandh
Manufacturer
Codei
For primary detection test
 Architect rHTLV I/II
CLIAa
Sand
Abbott Japan
A
 Alinity i rHTLV I/II
CLIA
Sand
Abbott Japan
B
 Elecsys HTLV-I/II
ECLIAb
Sand
Roche Diagnostics
C
 Lumipulse HTLV-I/II
CLEIAc
Sand
Fujirebio
D
 Lumipulse Presto HTLV-I/II
CLEIA
Sand
Fujirebio
E
 Serodia HTLV-I
PAd
-
Fujirebio
F
 ESPLINE HTLV-I/II
ICAe
-
Fujirebio
G
 HISCL HTLV-I
CLEIA
Ind
Sysmex
H
 UD1
CLEIA
Sand
Sysmex
I
 UD2
CLEIA
Sand
Non-disclosure
J
For confirmatory/discriminatory test
 INNO-LIA HTLV-I/II Score
LIAf
-
Fujirebio
 
CLIAa Chemiluminescent immunoassay, ECLIAb: electrochemiluminescence immunoassay, CLEIAc chemiluminescent enzyme immunoassay, PAd Particle agglutination, ICAe Immunochromatography assay, LIAf Line immunoassay, Indg Indirect assay, Sandh Sandwich assay, Codei Used in Tables 2, 3 and Additional file 1

HTLV-positive/negative reference panels for evaluation of HTLV IVDs

To prepare HTLV-positive/negative reference panels, plasma specimens that had been declared ineligible for transfusion were provided by the Japanese Red Cross Blood Center. Additional testing after dispensing was performed according to the HTLV testing algorithm in Japan [7] at the National Institute of Infectious Diseases. The specimens were provided following an application for the use of donated blood in Japan per the guidelines on the use of donated blood in research and development. The specimen information was anonymized, and a decoding index was not created, as required by the Helsinki Declaration. Ethical approval was obtained from the Ethical Committee of the National Institute of Infectious Diseases (No. 1082).

Results

Estimation of the diagnostic specificity of 10 HTLV IVDs in primary detection test/screening

A total of 160 HTLV-negative plasma specimens were tested using 10 HTLV IVDs for primary detection test/screening. All IVDs were negative for all tests (Additional File 1). The diagnostic specificity of IVDs was 100%.

Estimation of the diagnostic sensitivity of 10 HTLV IVDs in primary detection test/screening

Ten specimens, that were tested positive with relatively low S/CO or COI using multiple IVDs for primary detection test/screening and were HTLV-indeterminate or negative using INNO-LIA HTLV-I/II Score (INNO-LIA), were removed from this IVD evaluation. Forty-four HTLV-1-positive and two HTLV-positive untypable specimens (No. 16 and 26), based on INNO-LIA, were tested for evaluation (Table 2). Six sandwich assays resulted in all specimens being positive. On the other hand, UD2 resulted in one HTLV-1-positive (No. 24) and one HTLV-positive specimen (No. 16) being negative (Table 2, column J). One indirect assay, HISCL HTLV-1, could not detect one HTLV-positive specimen (No. 16), but the updated product, UD1, detected it (Table 2, columns H and I). Serodia HTLV-I detected 44 of the 46 specimens, but could not detect two specimens (No. 16 and 24, Table 2, column F). ESPLINE HTLV-I/II (ICA) was able to diagnose all positive specimens being positive (Table 2, column G). The positive agreement rates are shown in Table 3.
Table 2
Results of HTLV IVDs using 44 HTLV-1 positive and two HTLV-positive untypable specimens
No
IVDs for primary detection test
INNO-LIA HTLV-I/II Score
A
B
C
D
E
F
G
H
I
J
Confirmation
Discrimination
Result
gag
p19 I/II
gag
p24 I/II
env
gp46 I/II
env
gp21 I/II
gag
p19-I
env
gp46-I
env
gp46-II
S/CO
S/CO
COI
COI
COI
D.F.a
Decision
COI
COI
COI
1
103.55
92.10
320.7
 > 50.0
 > 50.0
2048
Positive
 > 100.0
 > 100.0
239.893
3 + 
2 + 
2 + 
2 + 
2 + 
3 + 
 ± 
HTLV-I
2
156.38
141.75
335.0
 > 50.0
 > 50.0
8192
Positive
 > 100.0
 > 100.0
289.531
2 + 
2 + 
-
3 + 
2 + 
3 + 
 ± 
HTLV-I
3
17.63
17.06
58.1
27.9
23.4
64
Positive
3.6
49.9
6.386
1 + 
 ± 
1 + 
1 + 
1 + 
 ± 
-
HTLV-I
4
46.50
43.95
178.2
 > 50.0
 > 50.0
512
Positive
6.8
 > 100.0
6.034
1 + 
-
1 + 
2 + 
-
1 + 
-
HTLV-I
5
119.41
107.43
283.5
 > 50.0
 > 50.0
2048
Positive
23.8
 > 100.0
55.062
2 + 
1 + 
3 + 
2 + 
1 + 
 ± 
-
HTLV-I
6
76.57
73.74
294.0
 > 50.0
 > 50.0
2048
Positive
85.3
 > 100.0
170.378
3 + 
3 + 
1 + 
3 + 
 ± 
3 + 
 ± 
HTLV-I
7
73.14
67.73
292.2
 > 50.0
 > 50.0
1024
Positive
13.8
 > 100.0
33.105
1 + 
1 + 
1 + 
2 + 
 ± 
1 + 
-
HTLV-I
8
51.17
49.42
15.0
42.4
48.8
64
Positive
4.5
39.4
11.979
1 + 
-
2 + 
1 + 
 ± 
 ± 
-
HTLV-I
9
54.26
51.43
130.7
 > 50.0
 > 50.0
256
Positive
15.5
 > 100.0
39.815
2 + 
1 + 
2 + 
2 + 
2 + 
 ± 
 ± 
HTLV-I
10
67.95
64.61
100.2
 > 50.0
 > 50.0
256
Positive
5.1
 > 100.0
40.375
1 + 
1 + 
2 + 
2 + 
 ± 
2 + 
-
HTLV-I
11
61.10
59.17
41.9
 > 50.0
 > 50.0
128
Positive
12.8
 > 100.0
51.587
2 + 
1 + 
3 + 
1 + 
 ± 
2 + 
-
HTLV-I
12
108.69
102.12
286.7
 > 50.0
 > 50.0
1024
Positive
43.3
 > 100.0
114.827
2 + 
1 + 
1 + 
3 + 
2 + 
3 + 
-
HTLV-I
13
111.35
103.79
153.4
 > 50.0
 > 50.0
512
Positive
32.6
 > 100.0
39.667
2 + 
1 + 
3 + 
2 + 
2 + 
3 + 
-
HTLV-I
14
71.64
72.32
146.3
 > 50.0
 > 50.0
1024
Positive
10.2
 > 100.0
20.430
2 + 
 ± 
1 + 
2 + 
1 + 
2 + 
-
HTLV-I
15
117.15
110.85
311.5
 > 50.0
 > 50.0
4096
Positive
50.3
 > 100.0
147.137
3 + 
2 + 
2 + 
3 + 
3 + 
3 + 
-
HTLV-I
16
2.99
2.78
1.1
3.7
1.5
0
Positive
0.0
2.2
0.205
 ± 
-
-
 ± 
-
-
-
HTLV
17
25.70
23.20
107.3
23.9
29.5
256
Positive
8.1
79.1
14.375
2 + 
2 + 
2 + 
2 + 
-
2 + 
-
HTLV-I
18
40.27
38.15
56.7
45.2
42.6
128
Positive
9.5
74.2
9.735
1 + 
-
2 + 
2 + 
1 + 
1 + 
-
HTLV-I
19
36.10
36.02
45.0
 > 50.0
37.0
256
Positive
35.2
62.8
45.865
2 + 
1 + 
1 + 
2 + 
2 + 
-
-
HTLV-I
20
59.80
56.40
103.3
 > 50.0
 > 50.0
256
Positive
12.8
 > 100.0
17.582
1 + 
 ± 
2 + 
2 + 
1 + 
1 + 
-
HTLV-I
21
81.13
75.12
112.1
 > 50.0
 > 50.0
512
Positive
8.2
 > 100.0
21.974
2 + 
 ± 
1 + 
2 + 
2 + 
1 + 
-
HTLV-I
22
12.07
11.87
56.2
21.6
22.5
128
Positive
6.5
52.2
4.381
2 + 
 ± 
2 + 
2 + 
-
2 + 
-
HTLV-I
23
50.77
48.20
139.9
 > 50.0
 > 50.0
512
Positive
14.7
 > 100.0
12.135
2 + 
-
1 + 
2 + 
2 + 
 ± 
-
HTLV-I
24
3.46
3.41
7.7
2.9
1.7
0
Positive
4.7
3.5
0.542
 ± 
-
-
1 + 
1 + 
-
-
HTLV-I
25
2.72
2.68
53.3
6.5
6.1
32
Positive
19.4
13.7
2.509
2 + 
-
1 + 
2 + 
 ± 
 ± 
-
HTLV-I
26
58.14
57.77
65.6
 > 50.0
 > 50.0
256
Positive
7.7
 > 100.0
16.409
2 + 
-
-
2 + 
-
-
-
HTLV
27
98.19
88.51
237.5
 > 50.0
 > 50.0
1024
Positive
26.6
 > 100.0
28.569
2 + 
 ± 
2 + 
2 + 
2 + 
2 + 
-
HTLV-I
28
106.43
98.42
182.1
 > 50.0
 > 50.0
1024
Positive
68.4
 > 100.0
107.693
2 + 
2 + 
2 + 
2 + 
2 + 
3 + 
 ± 
HTLV-I
29
104.05
94.04
325.5
 > 50.0
 > 50.0
2048
Positive
35.0
 > 100.0
61.918
2 + 
1 + 
2 + 
3 + 
-
2 + 
-
HTLV-I
30
143.36
116.25
336.7
 > 50.0
 > 50.0
2048
Positive
 > 100.0
 > 100.0
242.534
2 + 
2 + 
2 + 
2 + 
2 + 
3 + 
1 + 
HTLV-I
31
109.65
100.68
272.5
 > 50.0
 > 50.0
2048
Positive
6.1
 > 100.0
112.804
2 + 
2 + 
2 + 
2 + 
 ± 
1 + 
-
HTLV-I
32
110.22
105.59
312.9
 > 50.0
 > 50.0
2048
Positive
11.8
 > 100.0
47.743
2 + 
2 + 
 ± 
2 + 
-
1 + 
-
HTLV-I
33
141.18
128.81
323.1
 > 50.0
 > 50.0
4096
Positive
61.2
 > 100.0
160.308
2 + 
2 + 
2 + 
3 + 
1 + 
3 + 
 ± 
HTLV-I
34
104.07
90.33
164.4
 > 50.0
 > 50.0
512
Positive
28.2
 > 100.0
22.931
2 + 
-
2 + 
2 + 
1 + 
2 + 
-
HTLV-I
35
142.34
129.40
304.5
 > 50.0
 > 50.0
8192
Positive
98.0
 > 100.0
101.134
2 + 
2 + 
2 + 
3 + 
2 + 
3 + 
1 + 
HTLV-I
36
123.75
111.83
336.8
 > 50.0
 > 50.0
4096
Positive
58.0
 > 100.0
227.879
2 + 
2 + 
2 + 
2 + 
2 + 
2 + 
 ± 
HTLV-I
37
134.56
125.49
194.8
 > 50.0
 > 50.0
512
Positive
11.5
 > 100.0
72.161
2 + 
1 + 
-
2 + 
1 + 
-
-
HTLV-I
38
93.86
87.15
326.8
 > 50.0
 > 50.0
4096
Positive
77.6
 > 100.0
158.547
2 + 
2 + 
2 + 
3 + 
2 + 
3 + 
-
HTLV-I
39
91.70
78.26
107.8
 > 50.0
 > 50.0
256
Positive
22.8
 > 100.0
51.753
1 + 
2 + 
2 + 
2 + 
1 + 
2 + 
-
HTLV-I
40
128.72
121.58
355.3
 > 50.0
 > 50.0
2048
Positive
5.2
 > 100.0
26.333
1 + 
-
2 + 
2 + 
1 + 
-
-
HTLV-I
41
103.66
92.73
80.3
 > 50.0
 > 50.0
256
Positive
1.8
 > 100.0
38.196
2 + 
1 + 
1 + 
2 + 
1 + 
-
-
HTLV-I
42
130.78
115.42
333.7
 > 50.0
 > 50.0
4096
Positive
69.7
 > 100.0
127.553
3 + 
2 + 
2 + 
3 + 
2 + 
3 + 
 ± 
HTLV-I
43
99.55
83.48
291.8
 > 50.0
 > 50.0
2048
Positive
72.4
 > 100.0
112.354
3 + 
2 + 
2 + 
2 + 
2 + 
2 + 
-
HTLV-I
44
40.07
38.19
230.3
 > 50.0
 > 50.0
512
Positive
9.0
 > 100.0
11.480
2 + 
 ± 
2 + 
2 + 
-
2 + 
-
HTLV-I
45
63.71
61.02
106.0
 > 50.0
 > 50.0
512
Positive
11.1
 > 100.0
15.413
2 + 
 ± 
2 + 
2 + 
2 + 
2 + 
-
HTLV-I
46
77.28
71.96
272.4
 > 50.0
 > 50.0
2048
Positive
83.0
 > 100.0
193.647
3 + 
3 + 
2 + 
2 + 
2 + 
2 + 
-
HTLV-I
A: Architect rHTLV I/II; B: Alinity I rHTLV I/II; C: Elecsys HTLV-I/II; D: Lumipulse HTLV-I/II; E: Lumipulse Presto HTLV-I/II; F: Serodia HTLV-I; G: ESPLINE HTLV-I/II; H: HISCL HTLV-I; I: UD1; J: UD2
D.F.a: dilution factor, D.F. ≥ 16: Positive, S/CO, COI ≧ 1.0: Positive
Table 3
Positive agreement rate of HTLV IVDs for primary detection test/screening
 
IVDs
A, B, C,
D, E, G, I
H
F, J
HTLV(-1) positive
46
46
46
Test positive
46
45
44
Positive agreement rate (%)
100
97.8
95.7
A: Architect rHTLV I/II; B: Alinity I rHTLV I/II; C: Elecsys HTLV-I/II; D: Lumipulse HTLV-I/II; E: Lumipulse Presto HTLV-I/II; F: Serodia HTLV-I; G: ESPLINE HTLV-I/II; H: HISCL HTLV-I; I: UD1; J: UD2

Distribution of S/CO or COI in HTLV-1 and HTLV positive specimens

The distribution of the S/CO or COI from the CLEIA, CLIA, and ECLIA tests with HTLV-1 and HTLV-positive specimens is shown in Fig. 1. The COI of UD1, based on the sandwich assay, showed a widespread distribution (Fig. 1g) compared with the approved indirect assay product, HISCL HTLV-1 (Fig. 1f). The distribution of other IVDs, based on sandwich assay, showed a similar to that of UD1 (Fig. 1a–e), except for UD2 (Fig. 1h).

Discussion

In the present study, HTLV-positive and HTLV-negative specimen reference panels were prepared from plasma specimens that had been declared ineligible for transfusion and used for the evaluation of HTLV IVDs. The S/CO and COI of HTLV-positive specimens were widely distributed, with values around and far from the cutoff values. This specimen reference panel is useful for the evaluation of IVDs.
The diagnostic specificity of IVDs for the primary detection test/screening was examined using 160 HTLV-negative specimens. The negative agreement rate for all IVDs used in this study was 100%. These results are acceptable.
The diagnostic sensitivity of the IVDs for the primary detection test/screening was examined using 44 HTLV-1-positive and two HTLV-positive specimens, which resulted from INNO-LIA. HTLV CLEIA/CLIA/ECLIA IVDs, which were approved recently in Japan, have been updated to new-generation products based on sandwich assay. Six of the seven sandwich assays showed 100% sensitivity (46/46). HISCL HTLV-I, indirect CLEIA, could not detect one HTLV-positive specimen, and the positive agreement rate was 97.8% (45/46). However, their updated product, UD1, detected all of the specimens, and the COI was widely distributed to that of HISCL HTLV-I. The distribution of the other five sandwich assays was similar to that of UD1. The performances of the HTLV CLEIA/CLIA/ECLIA IVDs were also examined and have been highly acclaimed in previous reports [1218]. These results indicated that the update to sandwich assay was effective for the improvement of HTLV IVD performance, similar to HIV IVDs [10, 11], and the updated products were recommended for primary detection test/screening of HTLV infection.
Specimens no.16 and 24 showed negative results using some IVDs for primary detection test/screening. No.16 showed p19 I/II ( ±), gp21 I/II ( ±), and no.24 was p19 I/II ( ±), gp21 I/II (1 +), p19 I (1 +) using INNO-LIA. HISCL HTLV-I, that decided specimen no.16 to be negative, use HTLV-1 p19 and gp46 for detection of anti-HTLV-1 antibodies. Serodia HTLV-I use purified HTLV-1 viral antigen and UD2 use HTLV-1/2 gp21 and p24. Those IVDs tested negative both no.16 and 24. These results suggested that the kind of the antigen used in these kits is not necessarily related to the contradiction with the test result of INNO-LIA.
Eight of the ten HTLV IVDs for primary detection test/screening used in this study are based on chemiluminescence detection and require a specific analyzer. These IVDs have the following advantages: (1) The process after setting the specimens is performed automatically, preventing human error and enabling tests to be performed with consistent quality. (2) More specimens can be tested with fewer personnel. (3) Use of the IVDs does not require special manual skills, reducing the burden of education and training. (4) They are compatible with the automation of examination reception and result output. The sensitivity and specificity of IVDs are high, particularly in sandwich assays; thus, the performance is considered to be adequate. Because a system for transporting clinical specimens is in place in Japan, consolidating laboratory operations in large-scale medical facilities and testing laboratories facilitates more efficient processing of multiple specimens.
By contrast, using these IVDs in small- and medium-sized laboratories and areas with low HTLV prevalence is difficult because they are expensive to purchase and maintain. ESPLINE HTLV-I/II is based on ICA and may be suitable for HTLV primary detection test/screening in small- and medium-sized diagnostic laboratories. Its performance was superior to that of Serodia HTLV-I, the only manual-use IVD approved in Japan. Its performance was equivalent to the new-generation CLEIA/CLIA/ECLIA tests in this study. Incorporating ESPLINE HTLV-I/II into the HTLV testing algorithm will be a subject in the future.
INNO-LIA can detect HTLV-1/2 structural protein-specific antibody responses and is suitable for confirmatory/discriminatory test. In previous studies, INNO-LIA has been reported to display superior sensitivity and specificity compared to the previously used WB-based diagnostics [7, 1921] and has been incorporated into the recommended testing algorithm in Japan [7]. In the present study, 10 specimens tested positive for relatively low S/CO or COI in multiple IVDs in primary detection test/screening and were HTLV-indeterminate or negative upon INNO-LIA testing. Primary detection test/screening positive and INNO-LIA-indeterminate or negative cases have also been reported [7, 1721]. Because these were plasma specimens, proviral testing could not be performed, and these specimens could not be identified as HTLV-positive. Therefore, we believe it is reasonable to exclude these specimens from IVD evaluation for primary detection test/screening.
We have included an entry of IVDs under development in this project and provided its evaluation results. We believe this type of research project contributes to the performance confirmation of approved IVDs and to improving the performance of new IVDs. We are still collecting new specimens. We will plan to continue evaluating approved HTLV IVDs periodically.

Conclusions

CLEIA/CLIA/ECLIA IVDs allow for automatic and rapid multi-specimen results without complex manipulation. New-generation HTLV IVDs, based on sandwich assay, demonstrated HTLV testing capabilities with high sensitivity and specificity. They are recommended for use in HTLV primary detection test/screening. However, these IVDs are expensive to purchase and maintain for small- and medium-sized laboratories. The use of ICA, which showed as same high performance as these IVDs in this study may overcome this issue.

Acknowledgements

The plasma specimens used in this study were provided following an application for the use of donated blood in Japan based on the guidelines on the use of donated blood in research and development. We thank Abbott Japan LLC., Roche Diagnostics Inc., Fujirebio Inc., Sysmex Inc., and Ortho Clinical Diagnostics Inc. for their cooperation and advice during the Japan Association of Clinical Reagents Industries conference.

Declarations

All the methods followed in the study were in accordance with Declaration of Helsinki. The information of the specimens was anonymized, and a decoding index was not created. Ethical approval was obtained from the Ethical Committee of the National Institute of Infectious Diseases (No. 1082).
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
7.
Zurück zum Zitat Okuma K, Kuramitsu M, Niwa T, Taniguchi T, Masaki Y, Ueda G, et al. Establishment of a novel diagnostic test algorithm for human T-cell leukemia virus type 1 infection with line immunoassay replacement of western blotting: a collaborative study for performance evaluation of diagnostic assays in Japan. Retrovirology. 2020;17(1):26. https://doi.org/10.1186/s12977-020-00534-0. (PMID: 32831150).CrossRefPubMedPubMedCentral Okuma K, Kuramitsu M, Niwa T, Taniguchi T, Masaki Y, Ueda G, et al. Establishment of a novel diagnostic test algorithm for human T-cell leukemia virus type 1 infection with line immunoassay replacement of western blotting: a collaborative study for performance evaluation of diagnostic assays in Japan. Retrovirology. 2020;17(1):26. https://​doi.​org/​10.​1186/​s12977-020-00534-0. (PMID: 32831150).CrossRefPubMedPubMedCentral
Metadaten
Titel
Evaluation of human T-cell leukemia virus in vitro diagnostics using plasma specimens collected in Japan
verfasst von
Shigeru Kusagawa
Ai Kawana-Tachikawa
Keiji Matsubayashi
Rieko Sobata
Isao Hamaguchi
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Infectious Diseases / Ausgabe 1/2023
Elektronische ISSN: 1471-2334
DOI
https://doi.org/10.1186/s12879-023-08402-w

Weitere Artikel der Ausgabe 1/2023

BMC Infectious Diseases 1/2023 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.