Open Access 01.07.2014 | Original Article

Evaluation of novice reader diagnostic performance in coronary CT angiography using an advanced cardiac software package

verfasst von: Peter Dankerl, Matthias Hammon, Alexey Tsymbal, Alexander Cavallaro, Stephan Achenbach, Michael Uder, Rolf Janka

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 4/2014

Abstract

Purpose

The purpose of this research was to evaluate whether a commercially available advanced cardiac software package for coronary CT angiography (CTA) interpretation may reliably assist inexperienced readers to screen for significant coronary artery stenoses.

Methods

Coronary CTA data sets of 61 consecutive patients with suspected coronary artery disease were evaluated by three novice readers with no experience in cardiac CT interpretation. In the first 15 patients, the novice readers were trained to use the advanced cardiac software package (includes automatic detection of coronary vessels, curved MPR and VRT reconstructions and a measurement too) knowing the results of an expert read. In the next 46 patients, the novice readers had to state whether there is a significant coronary artery stenosis (\(>\)50 %) and if they are confident with their diagnosis. The results of the novice readers were compared to the expert read.

Results

The 46 coronary CTA data sets contained 184 vessels with 15 stenoses in 9 patients. On a per-vessel analysis, novice reader 1/2/3 demonstrated 60 %/100 %/ 93% sensitivity, and 98 %/90 %/86 % specificity. Per patient, the readers diagnosed 36/28/29 cases correctly as free of stenoses, 6/9/8 correctly as having at least one stenosis, missed 3/0/1 cases with a stenosis and overdiagnosed 1/9/8 patients. Cohen’s kappa values for the three readers versus the expert were 0.60, 0.61 and 0.54. The three novice readers felt confident in the diagnosis of 36/33/30 patients. In these patients, they missed one significant stenosis, showed a sensitivity of 100 %/100 %/75 % and a specificity of 100 %/92 %/88 %.

Conclusions

The evaluated advanced cardiac software package successfully assists novice readers in interpreting coronary CTA data sets especially in ruling out significant coronary artery stenosis.
Hinweise
P. Dankerl and M. Hammon contributed equally to this work.

Introduction

Coronary computed tomography angiography (coronary CTA) can be used to identify and rule out coronary artery stenoses in selected patients [13]. In recent years, the number of coronary CTA studies has steadily increased for various reasons. More and more hospitals and radiological practices have access to multi-detector computed tomography systems which permit coronary CT imaging [4]. Furthermore, radiation exposure for routine coronary CTA can be reduced to less than 1 mSv per examination [5, 6]. The majority of patients prefer CT to invasive coronary angiography [7], and an increasing amount of clinical data, for example regarding the use of coronary CTA in acute chest pain patients, demonstrates the method’s efficacy [8].
However, the accurate identification of significant coronary stenoses is a challenging task that requires extensive training and experience, and according to Saur [9] is often constrained by the evaluator’s knowledge and ability. Therefore, the American College of Cardiology and American Heart Association have developed clinical competence criteria to standardize training for interpretation of coronary CTA [10]. Expertise and experience are measured in competence levels (1–3), whereas level 1 can be achieved during a one month training and interpretation of 50 coronary CTA examinations—while level 3 requires 6 month of training, 300 interpreted examinations as well as ongoing teaching and research in the area of coronary CTA. However, some authors [11] claim that the learning curve is substantially longer than suggested by the competence criteria. As indicated by the 15-fold increase in research on cardiac CT between 1996 and 2006 [12], most likely the number of coronary CTA studies dramatically increased, outweighing the number of expert readers.
The purpose of this research was to evaluate whether commercially available advanced cardiac software package for coronary CTA interpretation may reliably assist inexperienced readers to screen for significant coronary artery stenoses.

Materials and methods

The institutional review board approved this study and waived the need for informed consent.

Patient acquisition and scan technique

Data sets from 61 consecutive patients undergoing clinically indicated routine coronary CTA (24 male, 37 female; mean age \(54.3\pm 9.4\) years was retrospectively included for evaluation. Multidetector CT had been performed with a 128-section dual source CT system (Somatom Flash\(^\circledR \), Siemens, Forchheim, Germany). Clinical indications for coronary CTA included 3 patients prior and 5 after aortic valve replacement, 3 before coronary bypass operation and 50 patients with chest pain and intermediate pretest likelihood of coronary artery disease. The patients’ mean body weight was \(84.7 \pm 12.8\) kg. Patients with a heart rate greater than 60 beats/min before the examination received oral administration of a \(\beta \)-blocker (atenolol, 100 mg) 1 h prior to the scan. During coronary CTA, the patients’ mean heart rate was \(65.9 \pm 13.6\). Adopted doses of 50–80 ml i.v. contrast agent (350 mg iodine/ml, Imeron\(^\circledR \), Bracco, Friedrichshafen, Germany), followed by 50 ml of saline solution were administered to all patients at a flow rate of 4–7 ml/s (54 patients received 60 ml, 5 patients 50 ml and 2 patients 80 ml contrast media). Images were acquired in prospectively ECG-triggered high-pitch spiral mode, and acquisition was timed to start at 60 % of the patients’ R-peak to R-peak interval [13]. Tube voltage was 100 kV, tube current 320 mAs per rotation. Reconstructed slice thickness was 0.6 mm, slice increment 0.3 mm, and a 3D adaptive noise reduction algorithm termed B26 kernel by the manufacturer [14] was applied.

Advanced cardiac software package

We investigated a commercially available advanced cardiac software package (syngo.via\(^\circledR \), Siemens, Forchheim, Germany). It semi-automatically reformats the original images, segments the coronary arteries and provides a measurement tool for stenosis quantification (Figs. 1, 2).

Image evaluation

Standard of reference

A consensus read of two experienced readers (5 and 6 years experience) and one expert reader (15 years of experience in cardiac CT and Electron-Beam Tomography) supplied the reference standard. Therefore, they evaluated all 61 data sets for the presence of a significant coronary artery stenosis (\(>\)50 % reduction in diameter) in the left main (LM), the left anterior descending (LAD), the left circumflex (CX) and the right coronary artery (RCA), utilizing a standard 3D MPR visualization software (Leonardo, Siemens, Forchheim, Germany).

Readers

Three novice readers (medical students in their last semester of medical school), inexperienced in cardiac imaging and CT interpretation, were trained in the use of the advanced cardiac software package. Under the supervision of a radiologist experienced in using the advanced cardiac software package, the students interpreted the examinations of the first 15 patients applying the software and therefore were aware of the expert’s interpretation.

Workflow of image evaluation applying the advanced cardiac software package

The CT data sets were transferred to a 3D workstation and automatically post-processed by the advanced cardiac software package. For the experimental read, the three novice readers independently interpreted the same 46 examinations without further assistance and they were blinded to the patients’ clinical data. The workflow demanded the readers to apply the advanced cardiac software package detect stenoses and use the stenosis quantification tool. If a stenosis seemed to be overestimated by the advanced cardiac software package, for example due to artifacts, they were encouraged to use the assisted navigation to re-examine the stenosis in the corresponding location in multiplanar views. If in doubt, readers were instructed to base their final diagnosis on their perception of vessel stenosis rather than on the system’s calculated degree of severity, using the software as assistance, but leaving the final diagnosis up to the human readers. Furthermore, the readers were asked to document their confidence about their diagnosis for each vessel (confident/not confident).

Statistical analysis

The statistical analysis on the 46 test patients was calculated on a per-vessel view and on a per-patient view for all three novice readers.

Per vessel

True positives (TP), false positives (FP), true negatives (TN), false negatives (FN), sensitivity, specificity, negative (NPV) and positive predictive values (PPV), as well as diagnostic odds ratio (DOR) and interreader agreement of the novice reader and the expert reader (Cohen’s kappa and McNemar test) were analyzed.

Per patient

TP, FP, TN, FN, sensitivity, specificity, NPV, PPV were calculated.
Additionally, the readers’ per-patient diagnostic performance (sensitivity, specificity, PPV, NPV) depending on their diagnostic confidence were evaluated.

Results

Standard of reference

The 15 training coronary CTA data sets consisted of 60 evaluable coronary arteries, of which 7 vessels in 6 patients revealed significant stenoses (0/LM, 4/LAD, 1/CX, 2/RCA).
The 46 experimental coronary CTA data sets contained 184 evaluable coronary arteries, 169 without, and 15 with significant stenoses (0/LM, 6/LAD, 4/CX, 5/RCA). On a per-patient basis, 37 patients had no stenoses, while 9 patients showed at least one significant coronary artery stenosis. No coronary CTA data set was excluded due to artifacts or insufficient image quality.

Performance of the novice readers

The readers where able to process all 46 investigated coronary CTA data sets applying the advanced cardiac software package. No data set had to be excluded from the experiment due to artifacts or insufficient image quality.
On a per-vessel analysis, the three novice readers (reader 1/2/3) demonstrated 4/16/23 false positives, 6/0/1 false negatives and correctly identified 165/153/146 vessels as nonstenotic. This resulted in a sensitivity of 60 %/100 %/93 % and a specificity of 98 %/90 %/86 % and a diagnostic odds ratio (DOR) of 62, 145 and 89. Displaying a Cohen’s kappa of 0.61, reader 2 demonstrated substantial interreader agreement with the reference standard [15], while readers 1 and 3 demonstrated good agreement (0.60/0.54) [16]. McNemar’s p-value did not indicate a significant difference between the diagnostic performances of reader 1 and the reference standard (\(p=0.75\)) (Table 1).
Table 1
Diagnostic performance of all three novice readers in detecting significant coronary artery stenoses compared to the expert read (=reference standard)—per-vessel analysis
 
Reader 1
Reader 2
Reader 3
True positives reference standard n \(=\)15
9
15
14
True negatives reference standard n \(=\)169
165
153
146
False negatives
6
0
1
False positives
4
16
23
Sensitivity (%)
60
100
93
Specificity (%)
98
90
86
Positive predictive value (%)
69
48
36
Negative predictive value (%)
97
100
99
Diagnostic odds ratio (DOR)
61.9
144.9
88.9
Cohen‘s kappa
0.60
0.61
0.54
McNemar \(p\)-value
0.75
\(<\!0.001\)
\(<\!0.001\)
Per patient, the readers diagnosed 36/28/29 cases correctly as free of stenoses, 6/9/8 correctly as having at least one significant stenosis, missed 3/0/1 cases with a significant stenosis and overdiagnosed 1/9/8 patients. On the per-patient basis, sensitivity was 67 %/100 %/89 % and specificity 97 %/76 %/78 % (Table 2).
Table 2
Diagnostic performance of all three novice readers in detecting significant coronary artery stenoses compared to the expert read (=reference standard)—per-patient analysis
 
Reader 1
Reader 2
Reader 3
True positives reference standard \(n=9\)
6
9
8
True negatives reference standard \(n=37\)
36
28
29
False negatives (%)
3
0
1
False positives (%)
1
9
8
Sensitivity (%)
67
100
89
Specificity (%)
97
76
78
Positive predictive value (%)
86
50
50
Negative predictive value (%)
92
100
97
Mean false positives per patient
0.09
0.37
0.5
Overall readers felt uncertain about their diagnosis in 39/138 patients (28 %). In these cases, the sensitivity, specificity, PPV and NPV for detection of patients with significant coronary stenosis ranged between 50 and 63 %. If the readers felt certain about their diagnosis (99/138 patients), sensitivity/specificity/PPV/NPV increased to 93 %/95 %/76 %/99 %. They correctly interpreted 80/99 cases as free of stenosis and missed one significant stenosis in 99 patients (Table 3, Fig. 3).
Table 3
Novice readers’ diagnostic performance in detecting significant coronary artery stenoses depending on their diagnostic confidence—per-patient analysis
 
Confident
Nonconfident
Reader
1
2
3
\(\sum \)
1
2
3
\(\sum \)
True positives
2
8
3
13
4
1
5
10
True negatives
34
23
23
80
2
5
5
12
False positives
0
2
3
5
1
0
6
7
False negatives
0
0
1
1
3
7
0
10
\(\sum \)
36
33
30
99
10
13
16
39
Sensitivity (%)
100
100
75
93
57
13
100
50
Specificity (%)
100
92
88
95
67
100
45
63
Positive predictive value (%)
100
80
50
76
80
100
45
59
Negative predictive value (%)
100
100
96
99
40
42
100
55

Discussion

As demonstrated by this work, novice readers revealed moderate sensitivity and specificity in detecting significant coronary artery stenosis when using a commercially available advanced cardiac software package. However, when taking readers’ confidence level into consideration, novices displayed excellent sensitivity and NPV for detection of significant coronary artery stenoses. The presented workflow considers the advanced cardiac software as a tool inexperienced readers (e.g., interns) can apply to screen for significant stenoses in coronary CTA. If the expert would have reevaluated only cases which were marked as diseased or uncertain by the novice readers which corresponded to 40 % of all cases, he would have missed only one significant stenosis. Our results stand in accordance with the papers of Meyer et al. and Arnoldi et al. [17, 18], who demonstrated computer-aided systems to be helpful for detection of coronary stenosis by less-experienced readers. In contrast to our work, the software was applied as second reader, whereas the setup in our experiment simulates utilization of the software in the initial read. Additionally, in our experiment, the software never suggested a diagnosis. Instead, in the presented study, inexperienced readers apply the advanced cardiac software package as semi-automatic coronary artery visualization and segmentation tool, with the final decision up to the human interpreter. Our work stands in consistency with Goldenberg et al. [19] who propose a CADx as an initial analyzer (rather than a second reader), but leaves the final decision to the physician. As displayed by the excellent sensitivity and NPV, as well as the acceptable specificity and PPV, the presented method fulfills demands of a screening test [20].
According to Anders et al. and Ferencik et al. [21, 22], the diagnostic accuracy of prerendered images, for example curved MPR and VRT, is inferior to conventional interactive reading of coronary CTA data. However, the investigated software offers interactive 3D multiplanar view and prerendered images, as well as real-time navigation between corresponding locations from the prerendered images and the interactive 3D multiplanar viewer. This provides the speed and convenience of post-processed images and the precision of interactive 3D multiplanar reading.
Pugliese et al. [11] studied the learning curve of members of a cCTA fellowship. They also investigated novices in reading coronary CTA. However, since all readers were radiology/cardiology fellows, in contrast to our study, they all had at least some level of experience in reading CT images or diagnosing coronary artery disease. After one year of fellowship, investigated readers reached competence level 3 for the number of interpreted examinations [10]. Although the results from [11] (naturally using different data sets), and our study (due to different experimental protocols, varying study hypotheses, distinct analyzing approaches, etc.) have to be compared very carefully and do not display a perfect concordance, we might compare the diagnostic odds ratio (DOR), a single and powerful indicator of test performance [23]. Even without considering level of confidence for our readers, they demonstrated a higher DOR (62, 145, 89) compared to [11] (between 15 and 27). Reported sensitivity (68–75 %) and specificity (88–92 %), both after one year of training were similar or worse to the overall performance of our readers.
Further, novices’ diagnostic performance is validated by Cohen’s kappa, attesting moderate to good interreader agreement to the reference standard. McNemar’s value suggests high correlation between reader 1 and the reference standard. However, the need for dedicated and structured training in interpreting coronary CTA cannot be overemphasized.
One limitation of our study is that the novice readers did not initially analyze the coronary CTA data sets without the assistance of the advanced cardiac software. So we do not know if and how much the advanced software package has increased the diagnostic performance of the novice readers. Another limitation is that the expert reading was not verified by invasive coronary angiography and significant CTA stenoses could not be verified to be hemodynamically relevant or insignificant. It is a well-known fact that coronary CTA has a limited positive predictive value, i.e., too many patients are incorrectly suspected of having a significant coronary artery stenosis [24]. However, the aim of our research was mainly set on comparing the interpretation results of the novices with an expert read. The clinical effect of coronary CTA stenosis possibly displaying hemodynamic significance could be the next step in future work. Moreover, our study focused on the evaluation of consecutive, unselected routine coronary CTA data sets. Most investigated patients presented with chest pain and intermediate pretest likelihood of coronary artery disease and after negative CTA did not receive further invasive coronary angiography. Finally, the patient number was relatively small and the number of diseased arteries relatively low. This low incidence, however, represents a clinical cohort of patients with an intermediate pretest likelihood of coronary artery disease better than a preselected group of diseased patients. Although coronary artery disease with 20 % accounts for the leading cause of death in the western world, our incidence indeed is way higher than the average 30 % lifetime prevalence for men and 15 % for women in Western countries [25].

Conclusions

The applied advanced cardiac software successfully assists novice readers in interpreting coronary CTA data sets. Evaluated software and displayed manner of use demonstrates moderate sensitivity and specificity for novice readers. When further taking readers’ confidence level into account, sensitivity of detecting coronary stenosis was excellent. Therefore, the advanced cardiac software may be useful for novice readers as an initial analysis tool for screening and ruling out significant stenoses in coronary CT angiography.

Acknowledgments

This study was supported by the German Federal Ministry of Education and Research (project number 01EX1012B, “Spitzencluster Medical Valley”).

Conflict of interest

The authors from the University Hospital Erlangen (PD, MH, AC, SA, MU, RJ) do not have a conflict of interest. Mr. Alexey Tsymbal is a PhD statistician working for Siemens AG. He was commissioned to do the statistics in this research.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef
2.
Zurück zum Zitat Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms temporal resolution. J Am Coll Cardiol 45:123–127PubMedCrossRef Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms temporal resolution. J Am Coll Cardiol 45:123–127PubMedCrossRef
3.
Zurück zum Zitat Achenbach S, Ropers D, Pohle F-K, Raaz D, von Erffa J, Yilmaz A, Muschiol G, Daniel WG (2005) Detection of coronary artery stenoses using multi-detector CT with 16 x 0.75 collimation and 375 ms rotation. Eur Heart J 26:1978–1986PubMedCrossRef Achenbach S, Ropers D, Pohle F-K, Raaz D, von Erffa J, Yilmaz A, Muschiol G, Daniel WG (2005) Detection of coronary artery stenoses using multi-detector CT with 16 x 0.75 collimation and 375 ms rotation. Eur Heart J 26:1978–1986PubMedCrossRef
4.
Zurück zum Zitat Yeow T, Raju V, Venkatanarasimha N, Fox B, Roobottom C (2011) Pictorial review: computed tomography features of cardiovascular emergencies and associated imminent decompensation. Emerg Radiol 18:127–138PubMedCrossRef Yeow T, Raju V, Venkatanarasimha N, Fox B, Roobottom C (2011) Pictorial review: computed tomography features of cardiovascular emergencies and associated imminent decompensation. Emerg Radiol 18:127–138PubMedCrossRef
5.
Zurück zum Zitat Lell M, Marwan M, Schepis T, Pflederer T, Anders K, Flohr T, Allmendinger T, Kalender W, Ertel D, Thierfelder C, Kuettner A, Ropers D, Daniel WG, Achenbach S (2009) Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience. Eur Radiol 19:2576–2583PubMedCrossRef Lell M, Marwan M, Schepis T, Pflederer T, Anders K, Flohr T, Allmendinger T, Kalender W, Ertel D, Thierfelder C, Kuettner A, Ropers D, Daniel WG, Achenbach S (2009) Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience. Eur Radiol 19:2576–2583PubMedCrossRef
6.
Zurück zum Zitat Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM (2010) Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 31:340–346PubMedCrossRef Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM (2010) Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 31:340–346PubMedCrossRef
7.
Zurück zum Zitat Dewey M, Zimmermann E, Deissenrieder F, Laule M, Dübel H-P, Schlattmann P, Knebel F, Rutsch W, Hamm B (2009) Noninvasive coronary angiography by 320-row computed tomography With lower radiation exposure and maintained diagnostic accuracy comparison of results With cardiac catheterization in a head-to-head pilot investigation. Circulation 120:867–875PubMedCrossRef Dewey M, Zimmermann E, Deissenrieder F, Laule M, Dübel H-P, Schlattmann P, Knebel F, Rutsch W, Hamm B (2009) Noninvasive coronary angiography by 320-row computed tomography With lower radiation exposure and maintained diagnostic accuracy comparison of results With cardiac catheterization in a head-to-head pilot investigation. Circulation 120:867–875PubMedCrossRef
8.
9.
Zurück zum Zitat Saur S, Alkadhi H, Stolzmann P, Baumüller S, Leschka S, Scheffel H, Desbiolles L, Fuchs TJ, Szekely G, Cattin PC (2012) Effect of reader experience on variability, evaluation time and accuracy of coronary plaque detection with computed tomography coronary angiography. Eur Radiol 20:1599–1606CrossRef Saur S, Alkadhi H, Stolzmann P, Baumüller S, Leschka S, Scheffel H, Desbiolles L, Fuchs TJ, Szekely G, Cattin PC (2012) Effect of reader experience on variability, evaluation time and accuracy of coronary plaque detection with computed tomography coronary angiography. Eur Radiol 20:1599–1606CrossRef
10.
Zurück zum Zitat Budoff MJ, Achenbach S, Berman DS, Fayad ZA, Poon M, Taylor AJ, Uretsky BF, Williams KA (2008) Task force 13 training in advanced cardiovascular imaging (computed tomography). J Cardiovasc Comput Tomogr 2:130–135PubMedCrossRef Budoff MJ, Achenbach S, Berman DS, Fayad ZA, Poon M, Taylor AJ, Uretsky BF, Williams KA (2008) Task force 13 training in advanced cardiovascular imaging (computed tomography). J Cardiovasc Comput Tomogr 2:130–135PubMedCrossRef
11.
Zurück zum Zitat Pugliese F, Hunink MGM, Gruszczynska K, Alberghina F, Malagó R, Van Pelt N (2009) Learning curve for coronary CT angiography: what constitutes sufficient training? Radiology 251:359–368 Pugliese F, Hunink MGM, Gruszczynska K, Alberghina F, Malagó R, Van Pelt N (2009) Learning curve for coronary CT angiography: what constitutes sufficient training? Radiology 251:359–368
12.
Zurück zum Zitat Itagaki MW, Suh RD, Goldin JG (2009) Cardiac CT research: exponential growth. Radiology 252:468–476 Itagaki MW, Suh RD, Goldin JG (2009) Cardiac CT research: exponential growth. Radiology 252:468–476
13.
Zurück zum Zitat Achenbach S, Manolopoulos M, Schuhbäck A, Ropers D, Rixe J, Schneider C, Krombach GA, Uder M, Hamm C, Daniel WG, Lell M (2012) Influence of heart rate and phase of the cardiac cycle on the occurrence of motion artifact in dual-source CT angiography of the coronary arteries. J Cardiovasc Comput Tomogr 6:91–98 Achenbach S, Manolopoulos M, Schuhbäck A, Ropers D, Rixe J, Schneider C, Krombach GA, Uder M, Hamm C, Daniel WG, Lell M (2012) Influence of heart rate and phase of the cardiac cycle on the occurrence of motion artifact in dual-source CT angiography of the coronary arteries. J Cardiovasc Comput Tomogr 6:91–98
14.
Zurück zum Zitat McCollough CH, Primak AN, Saba O, Bruder H, Stierstorfer K, Raupach R, Suess C, Schmidt B, Ohnesorge BM, Flohr TG (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243:775–784PubMedCrossRef McCollough CH, Primak AN, Saba O, Bruder H, Stierstorfer K, Raupach R, Suess C, Schmidt B, Ohnesorge BM, Flohr TG (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243:775–784PubMedCrossRef
15.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
16.
Zurück zum Zitat Fleiss JL (1973) Statistical methods for rates and proportions. Wiley, New York Fleiss JL (1973) Statistical methods for rates and proportions. Wiley, New York
17.
Zurück zum Zitat Meyer M, Schoepf UJ, Fink C, Goldenberg R, Apfaltrer P, Gruettner J, Vajcs D, Schoenberg SO, Henzler T (2013) Diagnostic performance evaluation of a computer-aided simple triage system for coronary CT angiography in patients with intermediate risk for acute coronary syndrome. Acad Radiol 20:980–986PubMedCrossRef Meyer M, Schoepf UJ, Fink C, Goldenberg R, Apfaltrer P, Gruettner J, Vajcs D, Schoenberg SO, Henzler T (2013) Diagnostic performance evaluation of a computer-aided simple triage system for coronary CT angiography in patients with intermediate risk for acute coronary syndrome. Acad Radiol 20:980–986PubMedCrossRef
18.
Zurück zum Zitat Arnoldi E, Schoepf U, Rosenblum M, Zwerner PL, Goldenberg R, Reiser MF, Thilo C (2011) Computer-aided detection of coronary artery stenosis at computed tomography angiography: effect on performance of readers with different experience levels ECR 2011 Book of Abstracts-B- Scientific Sessions. Insights Imaging, 135–324 Arnoldi E, Schoepf U, Rosenblum M, Zwerner PL, Goldenberg R, Reiser MF, Thilo C (2011) Computer-aided detection of coronary artery stenosis at computed tomography angiography: effect on performance of readers with different experience levels ECR 2011 Book of Abstracts-B- Scientific Sessions. Insights Imaging, 135–324
19.
Zurück zum Zitat Goldenberg R, Peled N (2011) Computer-aided simple triage. Int J Comput Assist Radiol Surg 6:705–711PubMedCrossRef Goldenberg R, Peled N (2011) Computer-aided simple triage. Int J Comput Assist Radiol Surg 6:705–711PubMedCrossRef
20.
21.
Zurück zum Zitat Anders K, Ropers U, Kuettner A, Wechsel M, Daniel WG, Uder M, Achenbach S (2011) Individually adapted, interactive multiplanar reformations vs. semi-automated coronary segmentation and curved planar reformations for stenosis detection in coronary computed tomography angiography. Eur J Radiol 80:89–95PubMedCrossRef Anders K, Ropers U, Kuettner A, Wechsel M, Daniel WG, Uder M, Achenbach S (2011) Individually adapted, interactive multiplanar reformations vs. semi-automated coronary segmentation and curved planar reformations for stenosis detection in coronary computed tomography angiography. Eur J Radiol 80:89–95PubMedCrossRef
22.
Zurück zum Zitat Ferencik M, Ropers D, Abbara S, Cury RC, Hoffmann U, Nieman K, Brady TJ, Moselwski F, Daniel WG, Achenbach S (2007) Diagnostic accuracy of image postprocessing methods for the detection of coronary artery stenoses by using multidetector CT. Radiology 243:696–702PubMedCrossRef Ferencik M, Ropers D, Abbara S, Cury RC, Hoffmann U, Nieman K, Brady TJ, Moselwski F, Daniel WG, Achenbach S (2007) Diagnostic accuracy of image postprocessing methods for the detection of coronary artery stenoses by using multidetector CT. Radiology 243:696–702PubMedCrossRef
23.
Zurück zum Zitat Glas AS, Lijmer JG, Prins MH, Bonsel GJ, Bossuyt PMM (2003) The diagnostic odds ratio: a single indicator of test performance. J Clin Epidemiol 56:1129–1135PubMedCrossRef Glas AS, Lijmer JG, Prins MH, Bonsel GJ, Bossuyt PMM (2003) The diagnostic odds ratio: a single indicator of test performance. J Clin Epidemiol 56:1129–1135PubMedCrossRef
24.
Zurück zum Zitat Vogler N, Meyer M, Fink C, Schoepf UJ, Schönberg SO, Henzler T (2013) Predictive value of zero calcium score and low-end percentiles for the presence of significant coronary artery stenosis in stable patients with suspected coronary artery disease. Rofo 185:726–732PubMedCrossRef Vogler N, Meyer M, Fink C, Schoepf UJ, Schönberg SO, Henzler T (2013) Predictive value of zero calcium score and low-end percentiles for the presence of significant coronary artery stenosis in stable patients with suspected coronary artery disease. Rofo 185:726–732PubMedCrossRef
25.
Zurück zum Zitat Herold G (2013) Innere Medizin. Gerd Herold, Köln Herold G (2013) Innere Medizin. Gerd Herold, Köln
Metadaten
Titel
Evaluation of novice reader diagnostic performance in coronary CT angiography using an advanced cardiac software package
verfasst von
Peter Dankerl
Matthias Hammon
Alexey Tsymbal
Alexander Cavallaro
Stephan Achenbach
Michael Uder
Rolf Janka
Publikationsdatum
01.07.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 4/2014
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-013-0953-0

Weitere Artikel der Ausgabe 4/2014

International Journal of Computer Assisted Radiology and Surgery 4/2014 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.