19.04.2016 | Musculoskeletal

Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation

verfasst von: Pedro Augusto Gondim Teixeira, Anne-Sophie Formery, Gabriela Hossu, Daniel Winninger, Toufik Batch, Alban Gervaise, Alain Blum

Erschienen in: European Radiology | Ausgabe 2/2017

Einloggen, um Zugang zu erhalten

Abstract

Propose

To establish evidence-based recommendations for musculoskeletal kinematic 4D-CT on wide area-detector CT.

Materials and methods

In order to assess factors influencing image quality in kinematic CT studies, a phantom consisting of a polymethylmethacrylate rotating disk with round wells of different sizes was imaged with various acquisition protocols. Cadaveric acquisitions were performed on the ankle joint during motion in two different axes and at different speeds to allow validation of phantom data. Images were acquired with a 320 detector-row CT scanner and were evaluated by two readers.

Results

Motion artefacts were significantly correlated with various parameters (movement axis, distance to centre, rotation speed and volume acquisition speed) (p < 0.0001). The relation between motion artefacts and distance to motion fulcrum was exponential (R2 0.99). Half reconstruction led to a 23 % increase in image noise and a 40 % decrease in motion artefacts. Cadaveric acquisitions confirmed phantom data. Based on these findings, high tube rotation speed and half reconstruction are recommended for kinematic CT. The axis of motion significantly influences image artefacts and should be considered in patient training and evaluation of acquisition protocol suitability.

Conclusion

This study provides evidence-based recommendations for musculoskeletal kinematic 4D-CT.

Key points

Motion artefacts can hamper the quality and interpretation of dynamic joint studies
The recommendations presented here help increase image quality
Patient training and preparation can be improved
The artefact-free distance concept helps protocol adaptation and comparison
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Kwong Y, Mel AO, Wheeler G, Troupis JM (2015) Four-dimensional computed tomography (4DCT): a review of the current status and applications. J Med Imaging Radiat Oncol. doi:10.1111/1754-9485.12326 PubMed Kwong Y, Mel AO, Wheeler G, Troupis JM (2015) Four-dimensional computed tomography (4DCT): a review of the current status and applications. J Med Imaging Radiat Oncol. doi:10.​1111/​1754-9485.​12326 PubMed
2.
Zurück zum Zitat Wassilew GI, Janz V, Heller MO et al (2013) Real time visualization of femoroacetabular impingement and subluxation using 320-slice computed tomography. J Orthop Res 31:275–281CrossRefPubMed Wassilew GI, Janz V, Heller MO et al (2013) Real time visualization of femoroacetabular impingement and subluxation using 320-slice computed tomography. J Orthop Res 31:275–281CrossRefPubMed
3.
Zurück zum Zitat Totterman S, Tamez-Pena J, Kwok E et al (1998) 3D visual presentation of shoulder joint motion. Stud Health Technol Inform 50:27–33PubMed Totterman S, Tamez-Pena J, Kwok E et al (1998) 3D visual presentation of shoulder joint motion. Stud Health Technol Inform 50:27–33PubMed
4.
Zurück zum Zitat Edirisinghe Y, Troupis JM, Patel M et al (2014) Dynamic motion analysis of dart throwers motion visualized through computerized tomography and calculation of the axis of rotation. J Hand Surg Eur Vol 39:364–372CrossRefPubMed Edirisinghe Y, Troupis JM, Patel M et al (2014) Dynamic motion analysis of dart throwers motion visualized through computerized tomography and calculation of the axis of rotation. J Hand Surg Eur Vol 39:364–372CrossRefPubMed
5.
Zurück zum Zitat Shapeero LG, Dye SF, Lipton MJ et al (1988) Functional dynamics of the knee joint by ultrafast, cine-CT. Invest Radiol 23:118–123CrossRefPubMed Shapeero LG, Dye SF, Lipton MJ et al (1988) Functional dynamics of the knee joint by ultrafast, cine-CT. Invest Radiol 23:118–123CrossRefPubMed
7.
Zurück zum Zitat Choi YS, Lee YH, Kim S et al (2013) Four-dimensional real-time cine images of wrist joint kinematics using dual source ct with minimal time increment scanning. Yonsei Med J 54:1026–1032CrossRefPubMedPubMedCentral Choi YS, Lee YH, Kim S et al (2013) Four-dimensional real-time cine images of wrist joint kinematics using dual source ct with minimal time increment scanning. Yonsei Med J 54:1026–1032CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Gondim Teixeira PA, Gervaise A, Louis M et al (2015) Musculoskeletal wide detector. Principles, techniques and applications in clinical practice and research. Eur J Radiol, CT. doi:10.1016/j.ejrad.2014.12.033 Gondim Teixeira PA, Gervaise A, Louis M et al (2015) Musculoskeletal wide detector. Principles, techniques and applications in clinical practice and research. Eur J Radiol, CT. doi:10.​1016/​j.​ejrad.​2014.​12.​033
10.
Zurück zum Zitat Chen B, Lambrou T, Offiah AC et al (2013) An in vivo subject-specific 3D functional knee joint model using combined MR imaging. Int J Comput Assist Radiol Surg 8:741–750CrossRefPubMed Chen B, Lambrou T, Offiah AC et al (2013) An in vivo subject-specific 3D functional knee joint model using combined MR imaging. Int J Comput Assist Radiol Surg 8:741–750CrossRefPubMed
11.
Zurück zum Zitat Raghuraman S, Schrauth JHX, Weber DL et al (2013) Dynamic MR imaging of a minipig’s knee using a high-density multi-channel receive array and a movement device. MAGMA 26:215–228CrossRefPubMed Raghuraman S, Schrauth JHX, Weber DL et al (2013) Dynamic MR imaging of a minipig’s knee using a high-density multi-channel receive array and a movement device. MAGMA 26:215–228CrossRefPubMed
12.
Zurück zum Zitat Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301CrossRefPubMed Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301CrossRefPubMed
14.
Zurück zum Zitat Grosjean R, Sauer B, Guerra RM et al (2007) Degradation of the z- resolution due to a longitudinal motion with a 64-channel CT scanner. Conf Proc IEEE Eng Med Biol Soc 2007:4429–4432PubMed Grosjean R, Sauer B, Guerra RM et al (2007) Degradation of the z- resolution due to a longitudinal motion with a 64-channel CT scanner. Conf Proc IEEE Eng Med Biol Soc 2007:4429–4432PubMed
15.
Zurück zum Zitat Tay SC, Primak AN, Fletcher JG et al (2008) Understanding the relationship between image quality and motion velocity in gated computed tomography: preliminary work for 4-dimensional musculoskeletal imaging. J Comput Assist Tomogr 32:634–639CrossRefPubMedPubMedCentral Tay SC, Primak AN, Fletcher JG et al (2008) Understanding the relationship between image quality and motion velocity in gated computed tomography: preliminary work for 4-dimensional musculoskeletal imaging. J Comput Assist Tomogr 32:634–639CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Gondim Teixeira PA, Gervaise A, Louis M et al (2015) Musculoskeletal Wide detector CT: Principles, Techniques and Applications in Clinical Practice and. Research, European Journal of Radiology Gondim Teixeira PA, Gervaise A, Louis M et al (2015) Musculoskeletal Wide detector CT: Principles, Techniques and Applications in Clinical Practice and. Research, European Journal of Radiology
17.
Zurück zum Zitat Hastie T, Tibshirani R, Friedman J (2009) The Elements of Statistical Learning. Springer New York, New York, NYCrossRef Hastie T, Tibshirani R, Friedman J (2009) The Elements of Statistical Learning. Springer New York, New York, NYCrossRef
18.
Zurück zum Zitat Farshad-Amacker NA, Alkadhi H, Leschka S, Frauenfelder T (2013) Effect of high-pitch dual-source CT to compensate motion artifacts: a phantom study. Acad Radiol 20:1234–1239CrossRefPubMed Farshad-Amacker NA, Alkadhi H, Leschka S, Frauenfelder T (2013) Effect of high-pitch dual-source CT to compensate motion artifacts: a phantom study. Acad Radiol 20:1234–1239CrossRefPubMed
19.
Zurück zum Zitat Dawson P, Lees WR (2001) Multi-slice technology in computed tomography. Clin Radiol 56:302–309CrossRefPubMed Dawson P, Lees WR (2001) Multi-slice technology in computed tomography. Clin Radiol 56:302–309CrossRefPubMed
20.
Zurück zum Zitat Bardo DME, Brown P (2008) Cardiac multidetector computed tomography: basic physics of image acquisition and clinical applications. Curr Cardiol Rev 4:231–243CrossRefPubMedPubMedCentral Bardo DME, Brown P (2008) Cardiac multidetector computed tomography: basic physics of image acquisition and clinical applications. Curr Cardiol Rev 4:231–243CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Yu L, Pan X (2003) Half-scan fan-beam computed tomography with improved noise and resolution properties. Med Phys 30:2629–2637CrossRefPubMed Yu L, Pan X (2003) Half-scan fan-beam computed tomography with improved noise and resolution properties. Med Phys 30:2629–2637CrossRefPubMed
Metadaten
Titel
Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation
verfasst von
Pedro Augusto Gondim Teixeira
Anne-Sophie Formery
Gabriela Hossu
Daniel Winninger
Toufik Batch
Alban Gervaise
Alain Blum
Publikationsdatum
19.04.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2017
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4362-y

Weitere Artikel der Ausgabe 2/2017

European Radiology 2/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.