Fan beam CT-guided online adaptive external radiotherapy of uterine cervical cancer: a dosimetric evaluation

verfasst von: Haibo Peng, Jie Zhang, Ningyue Xu, Yangang Zhou, Huigang Tan, Tao Ren

Erschienen in: BMC Cancer | Ausgabe 1/2023

Abstract

Purpose

To discuss the dosimetric advantages and reliability of the accurate delivery of online adaptive radiotherapy(online ART) for uterine cervical cancer(UCC).

Methods and materials

Six UCC patients were enrolled in this study. 95% of the planning target volume (PTV) reached 100% of the prescription dose (50.4 Gy/28fractions/6weeks) was required. The patients were scanned with uRT-Linac 506c KV-FBCT then the target volume (TV) and organs at risk (OARs) were delineated by doctors. The dosimeters designed and obtained a routine plan (Plan0). KV-FBCT was used for image guidance before subsequent fractional treatment. The online ART was processed after registration, which acquired a virtual nonadaptive radiotherapy plan (VPlan) and an adaptive plan (APlan). VPlan was the direct calculation of Plan0 on the fractional image, while APlan required adaptive optimization and calculation. In vivo dose monitoring and three-dimensional dose reconstruction were required during the implementation of APlan.

Results

The inter-fractional volumes of the bladder and rectum changed greatly among the treatments. These changes influenced the primary gross tumor volume (GTVp) and the position deviation of GTVp and PTV and positively affected the prescription dose coverage of TV. GTVp decreased gradually along with dose accumulation. The Dmax, D98, D95, D50, and D2 of APlan were superior to those of VPlan in target dose distribution. APlan had good conformal index, homogeneity index and target coverage. The rectum V40 and Dmax, bladder V40, the small bowel V40 and Dmax of APlan were better than that of VPlan. The APlan’s fractional mean γ passing rate was significantly higher than the international standard and the mean γ passing rate of all cases after the three-dimensional reconstruction was higher than 97.0%.

Conclusion

Online ART in external radiotherapy of UCC significantly improved the dose distribution and can become an ideal technology to achieve individualized precise radiotherapy.
Begleitmaterial
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12885-023-11089-6.
Haibo Peng and Jie Zhang contributed equally to this work and share first authorship.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ART
Adaptive radiotherapy
BT
Brachytherapy
CBCT
Cone beam computed tomography
CTV
Clinical target volume
DVH
Dose volume histogram
EBRT
External beam radiotherapy
EPID
Electronic portal imaging device
FBCT
Fan beam computed tomography
GTVn
Gross tumor volume of lymph node
GTVp
Primary gross tumor volume
IMRT
Intensity-modulated radiotherapy
OAR
Organ at risk
PTV
Planning target volume
ROI
Region of interest
SBRT
Stereotactic body radiation therapy
TV
Target volume
UCC
Uterine cervical cancer
VMAT
Volume-modulated arc therapy

Introduction

Uterine cervical cancer (UCC) is the most common gynecological malignant tumor worldwide. Radiotherapy is applicable to all stages of UCC and is a radical treatment for locally advanced UCC and an auxiliary treatment for early- and middle -stage UCC with postoperative risk factors. With the application of intensity-modulated technology which represents precise radiotherapy, the radical rate of UCC has increased from approximately 40–70% in clinical practice and the toxicity and side effects have also been significantly reduced [13]. UCC radiotherapy includes external beam radiotherapy (EBRT), brachytherapy (BT) and their combination. EBRT plays an important role in the treatment of UCC, whose accuracy of dose delivery directly affects the final efficacy [4]. At present, multimodal image fusion improved accuracy of target area delineation, and image-guided technique significantly reduces systematic errors, such as set-up errors [5, 6]. However, due to the special anatomical position of the cervix, adjacent to the bladder and rectum with obvious changes in physiological status (varying due to the degree of filling), coupled with random factors such as tumor retraction, the target area may be deformed and moved between fractional treatments, which may lead to dose distribution deviation [7, 8]. Normally, the above impacts might be reduced by trying to complete positioning, planning and treatment under a relatively stable physiological state within half an hour because the anatomical morphology of the bladder and rectum does not change significantly [912].
Online adaptive radiotherapy (online ART) can modify plans based on real-time images and can be executed immediately to accommodate random transfer of target volume (TV) and organs, as well as the deformation of organs at risk (OARs) in the fractional treatments, to ensure the coverage of the target dose and to control the dose of OARs [1315]. Online ART is particularly suitable for stereotactic radiotherapy of abdominal and pelvic tumors [1621]. Online ART, offline ART and real-time ART are collectively known as adaptive radiotherapy (ART), of which the plan can be dynamically modified according to the anatomical and/or physiological changes of the patient, rather than maintaining a static plan throughout the treatment process, to ensure dose targeting [2225]. Offline ART mainly targets systematic and progressive changes during treatment, such as patient weight loss and tumor morphologic changes. The new plan will be adopted in the follow-up fractional treatment. Offline ART has a significant effect on the clinical efficacy of head and neck and lung tumors [2630]. Real-time ART can automatically modify and implement plans through continuous imaging, replanning and rapid dose calculation during treatment to reduce the impact caused by respiration, internal state changes and organ peristalsis or pulsation and achieve targeted tracking irradiation [3134]. The concept of ART was put forward in 1990s and offline ART has obtained certain positive research results [2630, 35, 36]. However, online ART and real-time ART are limited by technical conditions and other factors, so far their clinical applications are few.
This study conducted online ART for UCC based on daily CT images. We aimed to discuss the necessity of online ART for UCC by analyzing the volume changes of the bladder and rectum, the change rules of primary gross tumor volume (GTVp) and the planning target volume (PTV), and the relative deviation of the target geometric center before each treatment. By comparing the online adaptive plan with the virtual nonadaptive radiotherapy (V-non ART) plan, the dosimetric results and influencing factors of the online ART plan were analyzed, of which the dosimetric advantages were discussed to lay a foundation for adjusting the external boundary from the clinical target volume (CTV) to the PTV and the prescription dose segmentation method. The accuracy of dose delivery was confirmed by the statistical results of in vivo dose monitoring. Finally, our study will provide data support for the clinical promotion of online ART in the treatment of UCC.

Materials and methods

Clinical materials

Six uterine cervical cancer patients with EBRT indications in our department from March 2022 to May 2022 were included in this study. Three patients received postoperative radiotherapy and three patients received external radiotherapy before BT or surgery. Patients did not need para-aortic irradiation, did not have indwelling urinary catheter and can persist in lying flat for more than 30 min. The median age was 54.5 years old and the functional status score (Karnofsky, KPS) was greater than 90.

Equipment

The uRT-linac 506c (United Imaging Healthcare Co., Ltd., Shanghai, China) is a CT-integrated linac guided by high-definition images, which was certified by China Food and Drug Administration(CFDA) as a radiotherapy device for treating systemic lesions in 2018. The uRT-linac 506c and radiotherapy plan management system (uRT-TPOIS, R001) integrate kilovoltage fan beam computed tomography (KV-FBCT) and a 6 MV X-ray linac, which are coaxial in the same bed, simultaneously realizing KV-FBCT image positioning and image guidance. The quality of the obtained image is consistent with the positioning image, which can be directly used for planning. The treatment planning system(TPS) has the functions of image fusion, automatic sketching, automatic planning, deformation and rigid registration and provides planning design methods, including 3-dimensional conformal radiotherapy (3DCRT), intensity-modulated radiotherapy (IMRT), volume-modulated arc therapy (VMAT) and stereotactic body radiation therapy (SBRT), which support online ART.

Patient model and plan preparation

Patients were instructed to empty their bladders and rectums 2 h before CT localization and treatments and to drink 500 mL water 1 h in advance. KV-FBCT positioning scanning and image-guided scanning of uRT-linac 506c were adopted, with the abdominal helical scanning protocol. The scanning scope included the lower abdomen and pelvis, with a thickness of 5 mm. Seven beams (gantry angle: 180°, 130°, 80°, 30°, 330°, 280° and 230°; collimator angle: 0°; maximum dose rate 600 MU/min) of coplanar dIMRT were used. 95% PTV received 100% of the prescription dose (50.4 Gy/28 fractions/6 weeks) [37, 38]. The region of interest (ROI) template (naming and structural relationship processing of TV and OARs to be sketched), the optimized constraint template and the clinical goal sheet template were prepared (Table S1). The gross tumor volume of lymph node (GTVn) and the GTVp were treated with late-course incremental radiotherapy, which was not included in this study.

Conventional radiotherapy plan formulation

KV-FBCT was used for localization scanning after effectively fixing the patient’s position, and the target area was then delineated according to the guidelines of the Radiation Therapy Oncology Group and the relevant international consensus [3842]. GTVp refers to the primary lesion of cervical tumors. GTVn refers to pelvic metastatic lymph nodes. CTV generally includes vaginal stumps, common iliac, internal iliac, external iliac, obturator and presacral lymphatic drainage areas (non surgical cases also include GTV, cervix, uterine body, parauterine and partial vagina). PTV was formed by the expansion of CTV by 7–15 mm. The corresponding beam template and optimized constraint template were loaded for plan design and optimizing calculation. Finally, the conventional plan, named Plan0, was obtained. Plan0 was transplanted to the phantom for point and surface dose verification and it entered the treatment phase after passing the verification.

Formulation of the online ART and V-non-ART plans

After the first reset of the patient, KV-FBCT image guidance was performed according to the preset scanning template. The positioning error was corrected under registration with the Plan0 image and then the online ART workflow was initiated, entering the patient’s intelligent planning module in the TPS and entering the online ART workflow after automatically loading Plan0. The OARs and couch in Plan0 were automatically delineated by deformation and target volumes (CTV and/or GTVp) in Plan0 were rigidly mapped to the current image and manually modified according to the Plan0 target area delineation principle. After confirming the ROI, an adaptive optimized calculation was selected based on Plan0 to generate the current adaptive radiotherapy plan APlanX (X = 1 ~ 28) according to the modified target area. The adaptive optimization algorithm used the dose distribution of Plan0 and the clinical goal sheet as inputs. The clinical goal sheet contained a wish-list with priorities, which was used to explicit the demands for prescription dose coverage of planning target and dosimetric criteria of OARs. The priorities were set to 1, 2 and 3, in which the number 1 represented the highest priority, and the number 3 represented the least priority (more details in supplementary material). Aplan was evaluated and then transmitted to treatment delivery application (TDA) for implementation once the dosimetric parameters of the ROI reached the pre-set dose objectives. In addition, the direct calculation was selected, and Plan0 was mapped to the current image, which further obtained the current virtual nonadaptive radiotherapy plan VPlanX (X = 1 ~ 28).

Implementation of APlan and in vivo dose monitoring

An electronic portal imaging device (EPID) was used to monitor the in vivo doses of radiation fields. If the global γ passing rate was lower than 88% under the 3%/3mm benchmark, the current treatment was suspended or terminated, and the reasons were recorded and analyzed. The three-dimensional dose reconstruction was finally compared with the planned value.

Statistical comparison of APlan- and VPlan-related parameters

The volume changes in the bladder and rectum were analyzed by comparing APlan and Plan0; the relative volume changes in the bladder, rectum, GTVp and PTV were calculated by the V (APlan-Plan0)/VPlan0 volume normalization method. [APlan offset value (x’, y’, z’) - Plan0 offset value (x, y, z)] and \(\sqrt{{({\text{x}}^{{\prime }} - \text{x})}^{2}+{({\text{y}}^{{\prime }}-\text{y})}^{2}+{({\text{z}}^{{\prime }}-\text{z})}^{2}}\) were used to describe the deviation changes in the relative positions of the geometric center of the PTV and the GTVp. The dosimetric differences between the TV and OARs were observed by comparing APlan and VPlan.

Statistical methods

SPSS 25.0 was used to perform statistical analysis on dosimetric parameters of TV and OARs, expressed by mean ± standard deviation. The Shapiro‒Wilk method was used to test the normality of the data. A paired sample t test was performed for data lines with a normal distribution, while the Wilcoxon signed rank test was performed with a significance level of 0.05.

Results

Volume changes of the bladder and rectum and volume change and position deviation of the PTV and the GTVp

As shown in Fig. 1(a) and Fig. 1(b), the volumes of the bladder and rectum in 6 cases changed significantly among fractional treatments, especially the bladder, but all cases were irregular. For example, the bladder volume in the second fraction of case 6 had the most obvious change with respect to Plan0, reaching 4.06, which is shown in Fig. 1(c), and the overfilled bladder was found in this division; the rectal volume of the fourth fraction of case 2 had the largest difference with respect to Plan0, which was approximately 2.17, as shown in Fig. 1(d), suggesting rectal inflation with the image analysis.
As shown in Fig. 1(e), the change of bladder volume was more obvious than that of the rectum, which was consistent with Fig. 1(a) and Fig. 1(b); the PTV and GTVp did not change significantly, but the GTVp decreased gradually with the accumulation of dose; the relative deviation changes of the geometric center point and “setup” point of the PTV and the GTVp in the lateral (LAT/X), vertical(VRT/Y) and longitudinal(LNG/Z) directions were obvious (the deviation ranges in the X, Y, and Z directions were -0.45~0.29 cm, -0.49~0.57 cm, -0.82~0.76 cm and -0.69~0.62 cm, -0.88~1.26 cm, -0.96~0.88 cm, respectively). As shown in Fig. 1(f), this deviation was affected by the volume changes of the bladder and rectum. The offset value was close to 0 when the volume changes of the bladder and rectum were small, and the points with large changes in the offset value had corresponding large changes in the bladder and rectum volume.

Statistical results of dosimetric parameters

In terms of target dose distribution, as shown in Table 1, Dmax, D98, D95, D50 and D2 of APlan were superior to those of VPlan, with significant differences (p < 0.05), among which the D95 (5041.18 ± 42.02) of APlan met the prescription requirements, while D95 (4972.67 ± 159.17) of VPlan failed to meet the prescription requirements. Figure 2(a) also shows that the D95 of APlan reached 100% of the prescription dose, while most implementations of VPlan did not; APlan had good conformal index (CI = 0.92 ± 0.02), homogeneity index (HI = 0.05 ± 0.00) and target coverage (V100%=0.96 ± 0.01), while VPlan had significant differences with its corresponding parameters (0.82 ± 0.06, 0.10 ± 0.06, 0.90 ± 0.04, respectively) (p < 0.05).
Table 1
Comparison of dosimetric parameters of various planning target volumes
Parameter
Plan0
VPlan
APlan
VPlan V.S. APlan P values
Dmax(cGy)
5370.36 ± 12.65
5433.23 ± 35.48
5338.57 ± 18.66
<0.05
D98(cGy)
5010.42 ± 4.21
4835.04 ± 327.39
5005.33 ± 8.69
<0.05
D95(cGy)
5041.62 ± 1.75
4972.67 ± 159.17
5041.18 ± 42.02
<0.05
D50(cGy)
5182.73 ± 3.70
5178.26 ± 23.45
5155.27 ± 10.66
<0.05
D2(cGy)
5307.27 ± 2.98
5333.12 ± 28.48
5273.42 ± 13.33
<0.05
CI
0.90 ± 0.01
0.82 ± 0.06
0.92 ± 0.02
<0.05
HI
0.06 ± 0.00
0.10 ± 0.06
0.05 ± 0.00
<0.05
V100%
0.95 ± 0.00
0.90 ± 0.04
0.96 ± 0.01
<0.05
Abbreviations:Dmax: maximum dose; D98, D95, D50, D2: the dose received by 98%, 95%, 50%, 2% of PTV; CI: conformity index=(Vt,ref/Vt)·(Vt,ref/Vref), Vt,ref is the target volume covered by the reference isodose line, Vt is the target volume, Vref is the total volume covered by the reference isodose line; HI: homogeneity index=(D2-D98)/D50; V100%: coverage of 100% prescription dose on PTV.
The protection of OARs is shown in Table 2. The rectum V40 (13.42 ± 7.88) and bladder V40 (24.75 ± 7.63) of APlan were superior to the rectum V40 (15.91 ± 12.87) and bladder V40 (24.01 ± 14.66) of VPlan. The small bowel Dmax, V40 and rectum Dmax of APlan were better than those of VPlan (p < 0.05). Other parameters were equivalent. Figure 2(b) and Fig. 2(c) show the doses of bladder V40 and rectum V40 in 168 fractions of 6 patients, which suggested that the bladder V40 and rectum V40 of APlan were better than those of VPlan, but without statistical significance.
Table 2
Comparison of dosimetric parameters of various organs at risk
OAR
parameter
Plan0
VPlan
APlan
VPlan V.S. APlan p values
Bladder-PTV
V40(%)
22.41 ± 8.69
24.01 ± 14.66
24.75 ± 7.63
0.409
Rectum-PTV
Dmax(cGy)
4270.27 ± 181.21
4906.02 ± 316.91
4403.62 ± 85.33
<0.05
V40(%)
2.74 ± 1.35
15.91 ± 12.87
13.42 ± 7.88
0.148
Small_
Bowel-PTV
Dmax(cGy)
4445.53 ± 173.81
4924.67 ± 235.47
4455.98 ± 40.73
<0.05
V40(%)
0.65 ± 0.32
3.34 ± 3.27
2.50 ± 2.31
<0.05
Femur Head_L
V30(%)
10.89 ± 3.04
11.66 ± 3.75
11.82 ± 3.03
0.088
V50(%)
0 ± 0
0.04 ± 0.16
0.02 ± 0.07
0.793
Femur Head_R
V30(%)
12.11 ± 1.55
12.16 ± 2.22
11.99 ± 2.16
0.174
V50(%)
0 ± 0
0.01 ± 0.04
0.02 ± 0.12
0.372
Abbreviations:V30, V40, and V50: volume ratios of 30 Gy, 40 and 50 Gy covering the corresponding OAR.
The dose volume histogram (DVH) in Fig. 2(d) intuitively shows that the dosimetric characteristics of APlan are equivalent to those of Plan0 and superior to those of VPlan.

Relative relationship between target dose and volume changes in the bladder and rectum

As shown in Fig. 3(a), the relative change in the VPlan target dose was positively related to the volume changes in the bladder and rectum, which directly affected the target dose distribution. Figure 3(b ~ d) shows the dose distribution in the target area under three different conditions (rectal volume reduction, GTVp reduction and rectal enlargement due to inflation). It was found that the target moved backward due to rectal volume reduction, and a low-dose area appeared in VPlan (Fig. 3(b)), while it was suggested in Fig. 3(c) that CTV (PTV) decreased due to GTVp shrinking and uterine body sinking, and the irradiated volumes of normal tissues and organs in VPlan were too large. Furthermore, Fig. 3(d) suggested that the rectal volume increased due to inflation and expansion, entering into the original treatment area in VPlan, with increasing irradiated dose and volume, which could be avoided by APlan.

In vivo dose monitoring and three-dimensional dose reconstruction analysis

As shown in Fig. 4(a ~ c), the 168 fractions’ mean γ passing rate of APlan of 6 cases was significantly higher than the international standard [43, 44]. The mean passing rate of case 6 was slightly lower (97.6%) than others, of which the fractional passing rate was generally greater than 96.5%. The mean passing rates of the other cases were more than 98.5%, and their fractional passing rates were generally greater than 98.0%. The lowest passing rate was from case 4, though it was still close to 95.0%. The statistical results of all angles’ γ passing rate showed that the mean γ passing rate at 280° was approximately 97.5%, which was obviously lower than the other angles.
Three-dimensional reconstruction was carried out for the collected two-dimensional dose information, and the mean γ passing rates of each case were slightly lower than the corresponding two-dimensional mean γ passing rates, which were all higher than 97.0%. The dosimetric parameters of the TV and OARs in the three-dimensional reconstruction dose field were not significantly different from those of APlan (Fig. 4(d ~ f)).

Discussion

Uterine cervical cancer (UCC) is a typical disease that is suitable and necessary for ART. It has been noted in UCC radiotherapy practice that tumor shrinkage, the movement and volume changes of peri-cervical OARs between fractions negatively affect the delivery of planned doses, which has the potential to reduce local tumor control and/or increase normal tissue toxicities [13, 45]. Based on IMRT and IGRT, performing the online ART may be the best means to achieve UCC clinical dosimetric goals. In this study, we found significant volume changes in the bladder and rectum relative to localized images through 168 fractional images of 6 cases and more significantly in the bladder, and all were irregular, consistent with other reports in the literature [7, 8, 46]. This may be related to factors such as altered human metabolism, surgery and radiation inflammation [47]. The volume changes of PTV generated by CTV expansion are not obvious, as the anatomy of the pelvis is solid and most of the lymphatic drainage areas in UCC are distributed along the pelvis wall [48]; approximately 80% of UCC cases are squamous cell carcinoma, which is relatively sensitive to radiation, which explains why the GTVp shrinks to different degrees with the accumulation of radiation dose – for example, by 60–80% in the first 3–4 weeks of radiotherapy [4951]. This study also presented similar results. In addition, the volume changes of the bladder and rectum were found to affect the PTV and the GTVp, resulting in changes in the relative offset between the geometric center point and the “setup” point in the three-dimensional direction, and the GTVp was particularly significant in the vertical direction, reaching 2.14 cm. The points with large changes in offset values corresponded to larger changes in bladder and rectal volumes and vice versa.
Online ART has obvious dosimetric advantages in UCC radiotherapy and can deliver dose accurately. With the introduction of the ART concept, offline ART was gradually used in clinical research and practice, especially after the advent of kilovoltage cone beam computed tomography (KV-CBCT). The value of offline ART in adjusting the TV dose distribution and reducing the CTV-PTV border has been proved, and partly reduced the impact of tumor shrinkage to some extent. However, relative to the anatomical information of the thoracoabdominal and pelvic cavities of localization, offline ART cannot cope with the impact of random changes at each fractional treatment [52, 53]. To balance both of these effects, online ART was developed. Due to the technical characteristics and factors such as less fractions of BT for UCC, online ART had been successfully applied in the BT stage of UCC. In the last two decades, the technique has gradually matured and become routine [5456]. Currently, EBRT for UCC is usually performed using coplanar IMRT or VMAT with conventional splitting of dose and 23–28 fractions, which is characterized by multiple fractions and long duration. In recent years, some scholars have randomized a few online ART studies based on CT guidance during the stage of EBRT for UCC, while there has not yet been reported on whole course online ART [57, 58]. EBRT was prescribed at a dose of 50.4 Gy for 28 fractions, five fractions per week, in all 6 cases of this study. Each time, the real-time images were reacquired, the TVs and OARs were redrawn based on the anatomical structure of the image feedback, and finally, the dose was optimized and the treatment was administered, completing the “localization–planning–treatment” regimen while the patient was almost stationary. In each case, 28 groups of APlan and VPlan and 1 group of Plan0 were obtained, reflecting the dosimetric differences between the two radiotherapy modalities. The dose distribution of Plan0 was directly mapped to the adaptive target area to form VPlan, and then the actual received dose of real-time TV and OARs was analyzed, which mainly showed that the target area coverage of the prescribed dose was insufficient, and the small bowel Dmax and rectum Dmax and V40 exceeded the standard or were even higher than the prescribed dose; the dose-optimized calculation was performed for real-time TV and OARs to obtain APlan. Its dosimetric parameters almost all met the standard, which was equivalent to Plan0, achieving a static effect with a dynamic adjustment.
Image quality and deformable alignment accuracy are key elements of Online ART, and device functionality is its foundation. The uRT-Linac 506c used in this study combines a diagnostic-grade KV-FBCT and a 6MV-X linac into a single device with two functions, image positioning and radiotherapy, and is equipped with integrated ART function software. KV-FBCT is guided by the same scan parameters as the positioning CT, and the acquired image sequence shares a CT value-relative electron density conversion curve with the positioning image so it can be used not only for pose calibration but also directly for planning design, which is the key to realizing online ART and is lacking in other forms of modal image guidance (e.g. CBCT, MR). In addition, the mechanical design of the CT and the linac are coaxial, with the same couch and the same coordinate system, and the couch board sinking monitoring and automatic correction function are added to keep the spatial position of the patient constant throughout the whole process. The Demons algorithm is used for variable image alignment, thus ensuring high image quality and high accuracy of alignment [59, 60].
EPID-based in vivo dose monitoring is a reliable guarantee of accurate APlan dose delivery. As the patient was placed on the couch during the whole process of online ART, pretreatment planned dose verification was not possible. Therefore, during the execution of APlan, in vivo dose monitoring was performed using EPID to analyze the γ passing rate in each field in real time and to set thresholds for warnings and interruptions according to the corresponding criteria [43, 44, 6164]. The mean γ passing rate of 168 events in this study was significantly higher than the international standard, except for a slightly lower mean γ passing rate in the 280° incidence direction (possibly due to the variation of the equivalent thickness of the transmitted body), and the mean γ passing rates of the remaining angles were above 98.5%. The two-dimensional dose comparison was not sufficient to confirm its accuracy, so a three-dimensional reconstruction was also needed. The reconstructed dose distribution was analyzed by three-dimensional comparison with the planned presented dose distribution. The three-dimensional mean γ passing rates of all cases in this study were higher than 97.0%, indicating that there was no significant change in the anatomical environment of the target area during the process from APlan fractionated image acquisition to execution, and the APlan dose could be delivered accurately with the error controlled within 5% of the expected dose.
Online ART may change the clinical strategy in UCC. The shape and location of GTVp during online ART are largely determined, so simultaneous integrated boost of GTVp, during the EBRT stage, increasing the dose of the primary site, adjusting the dose splitting pattern to shorten the treatment time, all of which have the potential to improve the efficacy. The main driver of target area location shift comes from volume changes in the bladder and rectum. Even if the preparation before each radiotherapy treatment is consistent with that before positioning, the reproducibility of the shape of bladder and rectum cannot be improved or even vary significantly, with the largest relative changes in this study reaching 405.74% and 217.20%, respectively. Thus, online ART helps to omit the cumbersome preparation process, such as filling the bladder and emptying the rectum before positioning and treatment, thus improving patient comfort.
In this research, APlan presented dosimetric results with significant advantages and was accurately implemented. However, due to the small number of cases and the short follow-up period, it is not yet possible to confirm whether the final clinical benefit was achieved. There are still further detailed explorations are needed such as what are the trigger conditions of online ART, what are the principles of target volume adjustment, how to ensure the accuracy of dose accumulation in fading tumors and what are the possible risks of increasing the radiation dose of each KV-FBCT [65, 66]?, etc. All of these questions need to be further considered and studied. In addition, during the whole process of online ART, since patients are always immobilized on the couch, the time of the online ART workflow needs to be shortened as much as possible. Through this study, TV modification and APlan optimization calculations were time-consuming and the total time of online ART was 21 min on average. Therefore, more in-depth research is also needed in developing intelligent registration algorithms and automatic dose optimization algorithms to strive for further optimization and iteration [67].

Conclusions

The application of online adaptive radiotherapy in UCC external radiotherapy can significantly improve the dosimetric distribution of radiotherapy for UCC; dose monitoring during treatment and 3D reconstruction of the dose after treatment can effectively guarantee the correct implementation of online adaptive radiotherapy. Further studies on image quality, deformation alignment and dose optimization algorithms can promote the application of online ART in UCC radiotherapy and may change some clinical strategies of UCC radiotherapy. Online ART has significant dosimetric advantages in UCC radiotherapy and is an ideal technique for achieving individualized and precise radiotherapy, with clinical promotional value.

Acknowledgements

Thank you to Chunyan Dai, Dong Gao, Tengda Wang and Binghuan Li, engineers from the United Imaging corporation, for providing technological support.

Declarations

The Ethics Committee of the First Affiliated Hospital of Chengdu Medical College had approved the study and the corresponding opinion letter NO. was 2022CYFYIRB-BA-Sep04. The informed consent had been obtained from each patient before enrollment and the whole process conformed to the guideline of cervical cancer of American Society for Radiation Oncology (ASTRO). All methods were performed in accordance with the relevant guidelines and regulations.
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Siegel RL, Miller KD, Fuchs HE, et al. Cancer Statistics, 2021. CA Cancer J Clin. 2021;71(1):7–33.PubMedCrossRef Siegel RL, Miller KD, Fuchs HE, et al. Cancer Statistics, 2021. CA Cancer J Clin. 2021;71(1):7–33.PubMedCrossRef
2.
Zurück zum Zitat Suneja G, Viswanathan A. Gynecologic malignancies. Hematol Oncol Clin North Am. 2020;34(1):71–89.PubMedCrossRef Suneja G, Viswanathan A. Gynecologic malignancies. Hematol Oncol Clin North Am. 2020;34(1):71–89.PubMedCrossRef
5.
Zurück zum Zitat Boda-Heggemann J, Lohr F, Wenz F, et al. kV cone-beam CT-based IGRT: a clinical review. Strahlenther Onkol. 2011;187(5):284–91.PubMedCrossRef Boda-Heggemann J, Lohr F, Wenz F, et al. kV cone-beam CT-based IGRT: a clinical review. Strahlenther Onkol. 2011;187(5):284–91.PubMedCrossRef
6.
Zurück zum Zitat Garibaldi C, Fodor C, Riva G, et al. Cone-beam CT-based inter-fraction localization errors for tumors in the pelvic region. Phys Med. 2018;46:59–66.PubMedCrossRef Garibaldi C, Fodor C, Riva G, et al. Cone-beam CT-based inter-fraction localization errors for tumors in the pelvic region. Phys Med. 2018;46:59–66.PubMedCrossRef
7.
Zurück zum Zitat Eminowicz G, Rompokos V, Stacey C, et al. Understanding the impact of pelvic organ motion on dose delivered to target volumes during IMRT for cervical cancer. Radiother Oncol. 2017;122(1):116–21.PubMedCrossRef Eminowicz G, Rompokos V, Stacey C, et al. Understanding the impact of pelvic organ motion on dose delivered to target volumes during IMRT for cervical cancer. Radiother Oncol. 2017;122(1):116–21.PubMedCrossRef
8.
Zurück zum Zitat Chen J, Liu P, Chen W, et al. Early changes of volume and spatial location in target and normal tissues caused by IMRT for cervical cancer. Tumori. 2016;102(6):610–3.PubMedCrossRef Chen J, Liu P, Chen W, et al. Early changes of volume and spatial location in target and normal tissues caused by IMRT for cervical cancer. Tumori. 2016;102(6):610–3.PubMedCrossRef
9.
Zurück zum Zitat Heijkoop ST, Langerak TR, Quint S, et al. Quantification of intra-fraction changes during radiotherapy of cervical cancer assessed with pre- and post-fraction cone Beam CT scans. Radiother Oncol. 2015;117(3):536–41.PubMedCrossRef Heijkoop ST, Langerak TR, Quint S, et al. Quantification of intra-fraction changes during radiotherapy of cervical cancer assessed with pre- and post-fraction cone Beam CT scans. Radiother Oncol. 2015;117(3):536–41.PubMedCrossRef
10.
Zurück zum Zitat Shortall J, Vasquez Osorio E, Cree A, et al. Inter- and intra-fractional stability of rectal gas in pelvic cancer patients during MRIgRT. Med Phys. 2021;48(1):414–26.PubMedCrossRef Shortall J, Vasquez Osorio E, Cree A, et al. Inter- and intra-fractional stability of rectal gas in pelvic cancer patients during MRIgRT. Med Phys. 2021;48(1):414–26.PubMedCrossRef
11.
Zurück zum Zitat Li X, Wang L, Cui Z, et al. Online MR evaluation of inter- and intra-fraction uterus motions and bladder volume changes during cervical cancer external beam radiotherapy. Radiat Oncol. 2021;16(1):179.PubMedPubMedCentralCrossRef Li X, Wang L, Cui Z, et al. Online MR evaluation of inter- and intra-fraction uterus motions and bladder volume changes during cervical cancer external beam radiotherapy. Radiat Oncol. 2021;16(1):179.PubMedPubMedCentralCrossRef
12.
Zurück zum Zitat Luo H, Jin F, Yang D, et al. Interfractional variation in bladder volume and its impact on cervical cancer radiotherapy: clinical significance of portable bladder scanner. Med Phys. 2016;43(7):4412.PubMedCrossRef Luo H, Jin F, Yang D, et al. Interfractional variation in bladder volume and its impact on cervical cancer radiotherapy: clinical significance of portable bladder scanner. Med Phys. 2016;43(7):4412.PubMedCrossRef
13.
Zurück zum Zitat Lim-Reinders S, Keller BM, Al-Ward S, et al. Online Adaptive Radiation Therapy. Int J Radiat Oncol Biol Phys. 2017;99(4):994–1003.PubMedCrossRef Lim-Reinders S, Keller BM, Al-Ward S, et al. Online Adaptive Radiation Therapy. Int J Radiat Oncol Biol Phys. 2017;99(4):994–1003.PubMedCrossRef
14.
Zurück zum Zitat Kim H, Lee P, Tree AC, et al. Adaptive Radiation Therapy Physician Guidelines: recommendations from an Expert users’ panel. Pract Radiat Oncol. 2022;12(5):e355–62.PubMedCrossRef Kim H, Lee P, Tree AC, et al. Adaptive Radiation Therapy Physician Guidelines: recommendations from an Expert users’ panel. Pract Radiat Oncol. 2022;12(5):e355–62.PubMedCrossRef
16.
Zurück zum Zitat Han EY, Wang H, Briere TM, et al. Brain stereotactic radiosurgery using MR-guided online adaptive planning for daily setup variation: an end-to-end test. J Appl Clin Med Phys. 2022;23(3):e13518.PubMedPubMedCentralCrossRef Han EY, Wang H, Briere TM, et al. Brain stereotactic radiosurgery using MR-guided online adaptive planning for daily setup variation: an end-to-end test. J Appl Clin Med Phys. 2022;23(3):e13518.PubMedPubMedCentralCrossRef
17.
Zurück zum Zitat Poon DMC, Yang B, Geng H, et al. Analysis of online plan adaptation for 1.5T magnetic resonance-guided stereotactic body radiotherapy (MRgSBRT) of prostate cancer. J Cancer Res Clin Oncol. 2022;24:1–10. Poon DMC, Yang B, Geng H, et al. Analysis of online plan adaptation for 1.5T magnetic resonance-guided stereotactic body radiotherapy (MRgSBRT) of prostate cancer. J Cancer Res Clin Oncol. 2022;24:1–10.
18.
Zurück zum Zitat Ikushima H, Balter P, Komaki R, et al. Daily alignment results of in-room computed tomography-guided stereotactic body radiation therapy for lung cancer. Int J Radiat Oncol Biol Phys. 2011;79(2):473–80.PubMedCrossRef Ikushima H, Balter P, Komaki R, et al. Daily alignment results of in-room computed tomography-guided stereotactic body radiation therapy for lung cancer. Int J Radiat Oncol Biol Phys. 2011;79(2):473–80.PubMedCrossRef
19.
Zurück zum Zitat Gong W, Yao Y, Ni J, et al. Deep learning-based low-dose CT for adaptive radiotherapy of abdominal and pelvic tumors. Front Oncol. 2022;12:968537.PubMedPubMedCentralCrossRef Gong W, Yao Y, Ni J, et al. Deep learning-based low-dose CT for adaptive radiotherapy of abdominal and pelvic tumors. Front Oncol. 2022;12:968537.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Kensen CM, Janssen TM, Betgen A, et al. Effect of intrafraction adaptation on PTV margins for MRI guided online adaptive radiotherapy for rectal cancer. Radiat Oncol. 2022;17(1):110.PubMedPubMedCentralCrossRef Kensen CM, Janssen TM, Betgen A, et al. Effect of intrafraction adaptation on PTV margins for MRI guided online adaptive radiotherapy for rectal cancer. Radiat Oncol. 2022;17(1):110.PubMedPubMedCentralCrossRef
21.
Zurück zum Zitat Nierer L, Eze C, da Silva Mendes V, et al. Dosimetric benefit of MR-guided online adaptive radiotherapy in different tumor entities: liver, lung, abdominal lymph nodes, pancreas and prostate. Radiat Oncol. 2022;17(1):53.PubMedPubMedCentralCrossRef Nierer L, Eze C, da Silva Mendes V, et al. Dosimetric benefit of MR-guided online adaptive radiotherapy in different tumor entities: liver, lung, abdominal lymph nodes, pancreas and prostate. Radiat Oncol. 2022;17(1):53.PubMedPubMedCentralCrossRef
22.
Zurück zum Zitat Yan D, Vicini F, Wong J, et al. Adaptive radiation therapy. Phys Med Biol. 1997;42(1):123–32.PubMedCrossRef Yan D, Vicini F, Wong J, et al. Adaptive radiation therapy. Phys Med Biol. 1997;42(1):123–32.PubMedCrossRef
23.
Zurück zum Zitat Glide-Hurst CK, Lee P, Yock AD, et al. Adaptive Radiation Therapy (ART) strategies and technical considerations: a state of the ART Review from NRG Oncology. Int J Radiat Oncol Biol Phys. 2021;109(4):1054–75.PubMedCrossRef Glide-Hurst CK, Lee P, Yock AD, et al. Adaptive Radiation Therapy (ART) strategies and technical considerations: a state of the ART Review from NRG Oncology. Int J Radiat Oncol Biol Phys. 2021;109(4):1054–75.PubMedCrossRef
24.
Zurück zum Zitat Yan D. Adaptive radiotherapy: merging principle into clinical practice. Semin Radiat Oncol. 2010;20(2):79–83.PubMedCrossRef Yan D. Adaptive radiotherapy: merging principle into clinical practice. Semin Radiat Oncol. 2010;20(2):79–83.PubMedCrossRef
25.
Zurück zum Zitat Chetty IJ, Fontenot J. Adaptive Radiation Therapy: Off-Line, On-Line, and In-Line? Int J Radiat Oncol Biol Phys. 2017;99(3):689–91.PubMedCrossRef Chetty IJ, Fontenot J. Adaptive Radiation Therapy: Off-Line, On-Line, and In-Line? Int J Radiat Oncol Biol Phys. 2017;99(3):689–91.PubMedCrossRef
26.
Zurück zum Zitat Yan D, Liang J. Expected treatment dose construction and adaptive inverse planning optimization: implementation for offline head and neck cancer adaptive radiotherapy. Med Phys. 2013;40(2):021719.PubMedCrossRef Yan D, Liang J. Expected treatment dose construction and adaptive inverse planning optimization: implementation for offline head and neck cancer adaptive radiotherapy. Med Phys. 2013;40(2):021719.PubMedCrossRef
27.
Zurück zum Zitat Lee VS, SchettIno G, Nisbet A. UK adaptive radiotherapy practices for head and neck cancer patients. BJR Open. 2020;2(1):20200051.PubMedPubMedCentral Lee VS, SchettIno G, Nisbet A. UK adaptive radiotherapy practices for head and neck cancer patients. BJR Open. 2020;2(1):20200051.PubMedPubMedCentral
28.
Zurück zum Zitat Zhang B, Lee SW, Chen S, et al. Action levels on dose and anatomic variation for adaptive Radiation Therapy using Daily Offline Plan evaluation: preliminary results. Pract Radiat Oncol. 2019;9(1):49–54.PubMedCrossRef Zhang B, Lee SW, Chen S, et al. Action levels on dose and anatomic variation for adaptive Radiation Therapy using Daily Offline Plan evaluation: preliminary results. Pract Radiat Oncol. 2019;9(1):49–54.PubMedCrossRef
29.
Zurück zum Zitat Yap ML, Sun A, Higgins J, et al. Adaptive dose escalation using serial four-dimensional Positron Emission Tomography/Computed tomography scans during Radiotherapy for locally Advanced Non-small Cell Lung Cancer. Clin Oncol (R Coll Radiol). 2016;28(12):e199–e205.PubMedCrossRef Yap ML, Sun A, Higgins J, et al. Adaptive dose escalation using serial four-dimensional Positron Emission Tomography/Computed tomography scans during Radiotherapy for locally Advanced Non-small Cell Lung Cancer. Clin Oncol (R Coll Radiol). 2016;28(12):e199–e205.PubMedCrossRef
30.
Zurück zum Zitat Teo TP, Ahmed SB, Kawalec P, et al. Feasibility of predicting tumor motion using online data acquired during treatment and a generalized neural network optimized with offline patient tumor trajectories. Med Phys. 2018;45(2):830–45.PubMedCrossRef Teo TP, Ahmed SB, Kawalec P, et al. Feasibility of predicting tumor motion using online data acquired during treatment and a generalized neural network optimized with offline patient tumor trajectories. Med Phys. 2018;45(2):830–45.PubMedCrossRef
32.
Zurück zum Zitat Muurholm CG, Ravkilde T, Skouboe S, et al. Real-time dose-guidance in radiotherapy: Proof of principle. Radiother Oncol. 2021;164:175–82.PubMedCrossRef Muurholm CG, Ravkilde T, Skouboe S, et al. Real-time dose-guidance in radiotherapy: Proof of principle. Radiother Oncol. 2021;164:175–82.PubMedCrossRef
33.
Zurück zum Zitat Keall P, Poulsen P, Booth JT. See, think, and Act: real-time adaptive radiotherapy. Semin Radiat Oncol. 2019;29(3):228–35.PubMedCrossRef Keall P, Poulsen P, Booth JT. See, think, and Act: real-time adaptive radiotherapy. Semin Radiat Oncol. 2019;29(3):228–35.PubMedCrossRef
34.
Zurück zum Zitat Mann P, Witte M, Mercea P, et al. Feasibility of markerless fluoroscopic real-time tumor detection for adaptive radiotherapy: development and end-to-end testing. Phys Med Biol. 2020;65(11):115002.PubMedCrossRef Mann P, Witte M, Mercea P, et al. Feasibility of markerless fluoroscopic real-time tumor detection for adaptive radiotherapy: development and end-to-end testing. Phys Med Biol. 2020;65(11):115002.PubMedCrossRef
35.
Zurück zum Zitat Ayyalusamy A, Vellaiyan S, Shanmugam S, et al. Feasibility of offline head & neck adaptive radiotherapy using deformed planning CT electron density mapping on weekly cone beam computed tomography. Br J Radiol. 2017;90(1069):20160420.PubMedCrossRef Ayyalusamy A, Vellaiyan S, Shanmugam S, et al. Feasibility of offline head & neck adaptive radiotherapy using deformed planning CT electron density mapping on weekly cone beam computed tomography. Br J Radiol. 2017;90(1069):20160420.PubMedCrossRef
36.
Zurück zum Zitat Veiga C, McClelland J, Moinuddin S, et al. Toward adaptive radiotherapy for head and neck patients: feasibility study on using CT-to-CBCT deformable registration for “dose of the day” calculations. Med Phys. 2014;41(3):031703.PubMedCrossRef Veiga C, McClelland J, Moinuddin S, et al. Toward adaptive radiotherapy for head and neck patients: feasibility study on using CT-to-CBCT deformable registration for “dose of the day” calculations. Med Phys. 2014;41(3):031703.PubMedCrossRef
37.
Zurück zum Zitat Klopp AH, Moughan J, Portelance L, et al. Hematologic toxicity in RTOG 0418: a phase 2 study of postoperative IMRT for gynecologic cancer. Int J Radiat Oncol Biol Phys. 2013;86(1):83–90.PubMedPubMedCentralCrossRef Klopp AH, Moughan J, Portelance L, et al. Hematologic toxicity in RTOG 0418: a phase 2 study of postoperative IMRT for gynecologic cancer. Int J Radiat Oncol Biol Phys. 2013;86(1):83–90.PubMedPubMedCentralCrossRef
38.
Zurück zum Zitat Wang W, Liu X, Meng Q, et al. Prophylactic extended-field irradiation for patients with cervical Cancer treated with concurrent chemoradiotherapy: a propensity-score matching analysis. Int J Gynecol Cancer. 2018;28(8):1584–91.PubMedPubMedCentralCrossRef Wang W, Liu X, Meng Q, et al. Prophylactic extended-field irradiation for patients with cervical Cancer treated with concurrent chemoradiotherapy: a propensity-score matching analysis. Int J Gynecol Cancer. 2018;28(8):1584–91.PubMedPubMedCentralCrossRef
39.
Zurück zum Zitat Taylor A, Rockall AG, Reznek RH, et al. Mapping pelvic lymph nodes: guidelines for delineation in intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2005;63(5):1604–12.PubMedCrossRef Taylor A, Rockall AG, Reznek RH, et al. Mapping pelvic lymph nodes: guidelines for delineation in intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2005;63(5):1604–12.PubMedCrossRef
40.
Zurück zum Zitat Torabi M, Aquino SL, Harisinghani MG. Current concepts in lymph node imaging. J Nucl Med. 2004;45(9):1509–18.PubMed Torabi M, Aquino SL, Harisinghani MG. Current concepts in lymph node imaging. J Nucl Med. 2004;45(9):1509–18.PubMed
41.
Zurück zum Zitat Small W Jr, Mell LK, Anderson P, et al. Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy in postoperative treatment of endometrial and cervical cancer. Int J Radiat Oncol Biol Phys. 2008;71(2):428–34.PubMedCrossRef Small W Jr, Mell LK, Anderson P, et al. Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy in postoperative treatment of endometrial and cervical cancer. Int J Radiat Oncol Biol Phys. 2008;71(2):428–34.PubMedCrossRef
42.
Zurück zum Zitat Lim K, Small W Jr, Portelance L, et al. Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy for the definitive treatment of cervix cancer. Int J Radiat Oncol Biol Phys. 2011;79(2):348–55.PubMedCrossRef Lim K, Small W Jr, Portelance L, et al. Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy for the definitive treatment of cervix cancer. Int J Radiat Oncol Biol Phys. 2011;79(2):348–55.PubMedCrossRef
43.
Zurück zum Zitat Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359–73.PubMedCrossRef Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359–73.PubMedCrossRef
44.
Zurück zum Zitat Wu Y, Cao R, Hu L, et al. Development and validation of dynamic intensity modulated Accurate Radiotherapy System KylinRay-IMRT. Zhongguo Yi Liao Qi Xie Za Zhi. 2018;42(1):7–10. Chinese.PubMed Wu Y, Cao R, Hu L, et al. Development and validation of dynamic intensity modulated Accurate Radiotherapy System KylinRay-IMRT. Zhongguo Yi Liao Qi Xie Za Zhi. 2018;42(1):7–10. Chinese.PubMed
45.
Zurück zum Zitat Tanderup K, Georg D, Pötter R, et al. Adaptive management of cervical cancer radiotherapy. Semin Radiat Oncol. 2010r;20(2):121–9.PubMedCrossRef Tanderup K, Georg D, Pötter R, et al. Adaptive management of cervical cancer radiotherapy. Semin Radiat Oncol. 2010r;20(2):121–9.PubMedCrossRef
46.
Zurück zum Zitat Jadon R, Pembroke CA, Hanna CL, et al. A systematic review of organ motion and image-guided strategies in external beam radiotherapy for cervical cancer. Clin Oncol (R Coll Radiol). 2014;26(4):185–96.PubMedCrossRef Jadon R, Pembroke CA, Hanna CL, et al. A systematic review of organ motion and image-guided strategies in external beam radiotherapy for cervical cancer. Clin Oncol (R Coll Radiol). 2014;26(4):185–96.PubMedCrossRef
47.
Zurück zum Zitat Hanfmann B, Engels M, Dörr W. Radiation-induced impairment of urinary bladder function. Assessment of micturition volumes. Strahlenther Onkol. 1998;174(Suppl 3):96–8.PubMed Hanfmann B, Engels M, Dörr W. Radiation-induced impairment of urinary bladder function. Assessment of micturition volumes. Strahlenther Onkol. 1998;174(Suppl 3):96–8.PubMed
48.
Zurück zum Zitat Agur AMR, Dalley AF. Grant’s Atlas of Anatomy. 12th ed. [M]. Philadelphia: Lippincott,Williams Wilkins Press, 2009. Agur AMR, Dalley AF. Grant’s Atlas of Anatomy. 12th ed. [M]. Philadelphia: Lippincott,Williams Wilkins Press, 2009.
49.
Zurück zum Zitat National Health Commission Of The People’s Republic Of China. National guidelines for diagnosis and treatment of cervical cancer 2022 in China (English version). Chin J Cancer Res. 2022;34(3):256–69.PubMedPubMedCentralCrossRef National Health Commission Of The People’s Republic Of China. National guidelines for diagnosis and treatment of cervical cancer 2022 in China (English version). Chin J Cancer Res. 2022;34(3):256–69.PubMedPubMedCentralCrossRef
50.
Zurück zum Zitat Mayr NA, Taoka T, Yuh WT, et al. Method and timing of tumor volume measurement for outcome prediction in cervical cancer using magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2002;52(1):14–22.PubMedCrossRef Mayr NA, Taoka T, Yuh WT, et al. Method and timing of tumor volume measurement for outcome prediction in cervical cancer using magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2002;52(1):14–22.PubMedCrossRef
51.
Zurück zum Zitat Arnesen MR, Hellebust TP, Malinen E. Impact of dose escalation and adaptive radiotherapy for cervical cancers on tumour shrinkage-a modelling study. Phys Med Biol. 2017;62(6):N107–19.PubMedCrossRef Arnesen MR, Hellebust TP, Malinen E. Impact of dose escalation and adaptive radiotherapy for cervical cancers on tumour shrinkage-a modelling study. Phys Med Biol. 2017;62(6):N107–19.PubMedCrossRef
52.
Zurück zum Zitat Buijs M, Bloemers MCWM, Remeijer P. Impact of a fiducial marker based ART strategy on margins in postoperative IMRT of gynecological tumors. Radiother Oncol. 2021;158:1–6.PubMedCrossRef Buijs M, Bloemers MCWM, Remeijer P. Impact of a fiducial marker based ART strategy on margins in postoperative IMRT of gynecological tumors. Radiother Oncol. 2021;158:1–6.PubMedCrossRef
53.
Zurück zum Zitat Sun R, Mazeron R, Chargari C, et al. CTV to PTV in cervical cancer: from static margins to adaptive radiotherapy. Cancer Radiother. 2016;20(6–7):622–8.PubMedCrossRef Sun R, Mazeron R, Chargari C, et al. CTV to PTV in cervical cancer: from static margins to adaptive radiotherapy. Cancer Radiother. 2016;20(6–7):622–8.PubMedCrossRef
54.
Zurück zum Zitat Portelance L, Corradini S, Erickson B, et al. Online magnetic resonance-guided Radiotherapy (oMRgRT) for gynecological cancers. Front Oncol. 2021;11:628131.PubMedPubMedCentralCrossRef Portelance L, Corradini S, Erickson B, et al. Online magnetic resonance-guided Radiotherapy (oMRgRT) for gynecological cancers. Front Oncol. 2021;11:628131.PubMedPubMedCentralCrossRef
55.
Zurück zum Zitat Ecker S, Zimmermann L, Heilemann G, et al. Neural network-assisted automated image registration for MRI-guided adaptive brachytherapy in cervical cancer. Z Med Phys. 2022;12:0939. -3889(22)00057 – 5. Ecker S, Zimmermann L, Heilemann G, et al. Neural network-assisted automated image registration for MRI-guided adaptive brachytherapy in cervical cancer. Z Med Phys. 2022;12:0939. -3889(22)00057 – 5.
56.
Zurück zum Zitat Wang B, Hu W, Shan G, et al. Estimating the accumulative dose uncertainty for intracavitary and interstitial brachytherapy. Biomed Eng Online. 2021;20(1):106.PubMedPubMedCentralCrossRef Wang B, Hu W, Shan G, et al. Estimating the accumulative dose uncertainty for intracavitary and interstitial brachytherapy. Biomed Eng Online. 2021;20(1):106.PubMedPubMedCentralCrossRef
57.
Zurück zum Zitat Bondar ML, Hoogeman MS, Mens JW, et al. Individualized nonadaptive and online-adaptive intensity-modulated radiotherapy treatment strategies for cervical cancer patients based on pretreatment acquired variable bladder filling computed tomography scans. Int J Radiat Oncol Biol Phys. 2012;83(5):1617–23.PubMedCrossRef Bondar ML, Hoogeman MS, Mens JW, et al. Individualized nonadaptive and online-adaptive intensity-modulated radiotherapy treatment strategies for cervical cancer patients based on pretreatment acquired variable bladder filling computed tomography scans. Int J Radiat Oncol Biol Phys. 2012;83(5):1617–23.PubMedCrossRef
58.
Zurück zum Zitat Heijkoop ST, Langerak TR, Quint S, et al. Clinical implementation of an online adaptive plan-of-the-day protocol for nonrigid motion management in locally advanced cervical cancer IMRT. Int J Radiat Oncol Biol Phys. 2014;90(3):673–9.PubMedCrossRef Heijkoop ST, Langerak TR, Quint S, et al. Clinical implementation of an online adaptive plan-of-the-day protocol for nonrigid motion management in locally advanced cervical cancer IMRT. Int J Radiat Oncol Biol Phys. 2014;90(3):673–9.PubMedCrossRef
59.
Zurück zum Zitat Yu L, Zhao J, Zhang Z, et al. Commissioning of and preliminary experience with a new fully integrated computed tomography linac. J Appl Clin Med Phys. 2021;22(7):208–23.PubMedPubMedCentralCrossRef Yu L, Zhao J, Zhang Z, et al. Commissioning of and preliminary experience with a new fully integrated computed tomography linac. J Appl Clin Med Phys. 2021;22(7):208–23.PubMedPubMedCentralCrossRef
60.
Zurück zum Zitat Thirion JP. Image matching as a diffusion process: an analogy with Maxwell’s demons. Med Image Anal. 1998;2(3):243–60.PubMedCrossRef Thirion JP. Image matching as a diffusion process: an analogy with Maxwell’s demons. Med Image Anal. 1998;2(3):243–60.PubMedCrossRef
61.
Zurück zum Zitat Zwan BJ, Caillet V, Booth JT, et al. Toward real-time verification for MLC tracking treatments using time-resolved EPID imaging. Med Phys. 2021;48(3):953–64.PubMedCrossRef Zwan BJ, Caillet V, Booth JT, et al. Toward real-time verification for MLC tracking treatments using time-resolved EPID imaging. Med Phys. 2021;48(3):953–64.PubMedCrossRef
62.
Zurück zum Zitat McCowan PM, Van Uytven E, Van Beek T, et al. An in vivo dose verification method for SBRT-VMAT delivery using the EPID. Med Phys. 2015;42(12):6955–63.PubMedCrossRef McCowan PM, Van Uytven E, Van Beek T, et al. An in vivo dose verification method for SBRT-VMAT delivery using the EPID. Med Phys. 2015;42(12):6955–63.PubMedCrossRef
63.
Zurück zum Zitat Van Elmpt W, Nijsten S, Petit S, et al. 3D in vivo dosimetry using megavoltage cone-beam CT and EPID dosimetry. Int J Radiat Oncol Biol Phys. 2009;73(5):1580–7.PubMedCrossRef Van Elmpt W, Nijsten S, Petit S, et al. 3D in vivo dosimetry using megavoltage cone-beam CT and EPID dosimetry. Int J Radiat Oncol Biol Phys. 2009;73(5):1580–7.PubMedCrossRef
65.
Zurück zum Zitat Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys. 2007;34(10):4041–63.PubMedCrossRef Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys. 2007;34(10):4041–63.PubMedCrossRef
66.
Zurück zum Zitat Steiner E, Stock M, Kostresevic B, et al. Imaging dose assessment for IGRT in particle beam therapy. Radiother Oncol. 2013;109(3):409–13.PubMedCrossRef Steiner E, Stock M, Kostresevic B, et al. Imaging dose assessment for IGRT in particle beam therapy. Radiother Oncol. 2013;109(3):409–13.PubMedCrossRef
67.
Zurück zum Zitat Fan J, Cao X, Xue Z, et al. Adversarial Similarity Network for evaluating Image Alignment in Deep Learning based Registration. Med Image Comput Comput Assist Interv. 2018;11070:739–46.PubMedPubMedCentral Fan J, Cao X, Xue Z, et al. Adversarial Similarity Network for evaluating Image Alignment in Deep Learning based Registration. Med Image Comput Comput Assist Interv. 2018;11070:739–46.PubMedPubMedCentral
Metadaten
Titel
Fan beam CT-guided online adaptive external radiotherapy of uterine cervical cancer: a dosimetric evaluation
verfasst von
Haibo Peng
Jie Zhang
Ningyue Xu
Yangang Zhou
Huigang Tan
Tao Ren
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Cancer / Ausgabe 1/2023
Elektronische ISSN: 1471-2407
DOI
https://doi.org/10.1186/s12885-023-11089-6

Weitere Artikel der Ausgabe 1/2023

BMC Cancer 1/2023 Zur Ausgabe

Alphablocker schützt vor Miktionsproblemen nach der Biopsie

16.05.2024 alpha-1-Rezeptorantagonisten Nachrichten

Nach einer Prostatabiopsie treten häufig Probleme beim Wasserlassen auf. Ob sich das durch den periinterventionellen Einsatz von Alphablockern verhindern lässt, haben australische Mediziner im Zuge einer Metaanalyse untersucht.

Mammakarzinom: Senken Statine das krebsbedingte Sterberisiko?

15.05.2024 Mammakarzinom Nachrichten

Frauen mit lokalem oder metastasiertem Brustkrebs, die Statine einnehmen, haben eine niedrigere krebsspezifische Mortalität als Patientinnen, die dies nicht tun, legen neue Daten aus den USA nahe.

Labor, CT-Anthropometrie zeigen Risiko für Pankreaskrebs

13.05.2024 Pankreaskarzinom Nachrichten

Gerade bei aggressiven Malignomen wie dem duktalen Adenokarzinom des Pankreas könnte Früherkennung die Therapiechancen verbessern. Noch jedoch klafft hier eine Lücke. Ein Studienteam hat einen Weg gesucht, sie zu schließen.

Viel pflanzliche Nahrung, seltener Prostata-Ca.-Progression

12.05.2024 Prostatakarzinom Nachrichten

Ein hoher Anteil pflanzlicher Nahrung trägt möglicherweise dazu bei, das Progressionsrisiko von Männern mit Prostatakarzinomen zu senken. In einer US-Studie war das Risiko bei ausgeprägter pflanzlicher Ernährung in etwa halbiert.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.