Fast retrospectively triggered local pulse-wave velocity measurements in mice with CMR-microscopy using a radial trajectory

verfasst von: Patrick Winter, Thomas Kampf, Xavier Helluy, Fabian T Gutjahr, Cord B Meyer, Eberhard Rommel, Wolfgang R Bauer, Peter M Jakob, Volker Herold

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2013

Abstract

Background

The aortic pulse-wave velocity (PWV) is an important indicator of cardiovascular risk. In recent studies MRI methods have been developed to measure this parameter noninvasively in mice. Present techniques require additional hardware for cardiac and respiratory gating. In this work a robust self-gated measurement of the local PWV in mice without the need of triggering probes is proposed.

Methods

The local PWV of 6-months-old wild-type C57BL/6J mice (n=6) was measured in the abdominal aorta with a retrospectively triggered radial Phase Contrast (PC) MR sequence using the flow-area (QA) method. A navigator signal was extracted from the CMR data of highly asymmetric radial projections with short repetition time (TR=3 ms) and post-processed with high-pass and low-pass filters for retrospective cardiac and respiratory gating. The self-gating signal was used for a reconstruction of high-resolution Cine frames of the aortic motion. To assess the local PWV the volume flow Q and the cross-sectional area A of the aorta were determined. The results were compared with the values measured with a triggered Cartesian and an undersampled triggered radial PC-Cine sequence.

Results

In all examined animals a self-gating signal could be extracted and used for retrospective breath-gating and PC-Cine reconstruction. With the non-triggered measurement PWV values of 2.3±0.2 m/s were determined. These values are in agreement with those measured with the triggered Cartesian (2.4±0.2 m/s) and the triggered radial (2.3±0.2 m/s) measurement. Due to the strong robustness of the radial trajectory against undersampling an acceleration of more than two relative to the prospectively triggered Cartesian sampling could be achieved with the retrospective method.

Conclusion

With the radial flow-encoding sequence the extraction of a self-gating signal is feasible. The retrospective method enables a robust and fast measurement of the local PWV without the need of additional trigger hardware.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1532-429X-15-88) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

PW developed the radial flow-encoding sequences and conducted all experiments. ER, WRB, VH and PMJ enrolled and organized this study. TK and VH helped with the data evaluation. XH and VH helped with the pulse-sequence development, XH, CBM and FTG provided the software for trigger signal evaluation. All authors took part in critical review and drafting of the manuscript and have read and approved the final manuscript.
Abkürzungen
PC
Phase-contrast
PWV
Pulse-wave velocity
VENC
Maximum Encoding Velocity
WT
Wild-type
CSA
Cross-sectional area.

Background

Cardiovascular Disease (CVD) is still one of the main causes of death in industrial nations. One important indicator of cardiovascular risk is the aortic pulse-wave velocity (PWV) as a measure of arterial stiffness[1, 2].
In the past years genetically modified mice have turned out to be of great relevance as models for the research of this predictor for the formation of CVD. As a non-invasive method high-field CMR has gained great importance for the measurement of morphological and functional parameters of the cardiovascular system in small animals[3]. A triggered Cartesian PC-Cine sequence was proposed which uses the flow-area (QA) method to determine the local PWV[4]. In this method high spatial and temporal resolutions were achieved by the acquisition of multiple interleaved Cine experiments. With the development of more capable hardware and faster reconstruction algorithms non-Cartesian sequences were developed to accelerate the CMR experiment and improve CMR robustness against flow artifacts. Zhao et al. improved the temporal resolution and acquisition time for a measurement of the regional PWV by measuring flow in multiple slices with a triggered undersampled and asymmetric radial Phase Contrast MR sequence[5].
A drawback of prospectively triggered methods is the necessity of external probes, which monitor cardiac and respiratory motion. The most common method to trigger the CMR acquisition is the use of Electrocardiograms (ECG). However, rapid switching of the imaging gradients can cause interferences with the ECG probes, which can strongly corrupt the trigger signal[6]. In addition, in high field CMR, the small RF coil dimensions can make the attachment of triggering probes difficult. Therefore experience with animal handling is required to ensure a stable trigger signal. The use of the CMR signal itself offers the possibility of a wireless trigger signal, which does not require external probes and is therefore easier to handle. Since the peak value of the echo signal corresponds to the sum of the transverse magnetization across the whole field of view (FOV), moving and changing voxels cause fluctuations of the peak signal intensity corresponding to the motion[7]. The signal variations can be assigned to respiratory and cardiac motion within the excited slice. With the application of fast repetitive radial read-outs a navigator signal can be extracted which can be used for retrospective triggering. This technique was used for the generation of self-gated cardiac Cines in mice without additional hardware[8].
The goal of our work was to design a fast self-gated PC-Cine method for PWV measurements in the aortae of mice, which is independent of additional trigger hardware and is applicable to high field CMR, where susceptibility disturbances are significantly increased. Since a stable trigger signal is required for the determination of this parameter, the self-gated method promises to improve the robustness of the measurement. However, the challenge for flow CMR in the aorta without triggering probes is that the slice of interest is often located remotely from the beating heart and therefore might contain only a small portion of changing voxels, which are able to contribute to a self-gating signal. Additionally, high temporal and spatial resolutions as well as small echo times in order to reduce image artifacts are necessary, which imposes high requirements to the flow compensated spatial encoding gradients. To ensure high-resolution measurements and reduced sensitivity to susceptibility variations, a highly asymmetric radial pulse sequence was developed to enable the acquisition of a self-gating signal without prolonging the read-out.

Methods

CMR

Pulse sequences

All measurements are carried out at a 17.6 T small animal system (Bruker Avance 750 WB, Bruker BioSpin MRI GmbH, Rheinstetten, Germany) with a 1 T/m gradient system and a home-built transverse electromagnetic (TEM) resonator with an inner diameter of 25 mm. The imaging protocol started with 2D-FLASH experiments to localize the descending aorta. The local PWV was measured in a slice perpendicular to the vessel (see Figure1). Based on the Cartesian Phase Contrast sequence proposed in[9] a motion-compensated radial PC-Cine-FLASH sequence was designed for through-plane velocity encoding (see Figure2). The scan protocol consists of three sequentially applied velocity-encoding steps with varying first gradient moments (M1=0 s/m and M1=±0.3 s/m), corresponding to a maximum encoding velocity (VENC) of 1.7 m/s. The radial sequence enables either a triggered flow-encoding mode based on the interleaved acquisition scheme of[4] or retrospective flow measurements without triggering.
To assess the quality of the retrospectively triggered PC-Cine measurement, prospectively triggered Cartesian and radial PC-Cine experiments were conducted for comparison. The Cartesian measurement was acquired using a 160×160 matrix and a field of view of 24×24 mm2. The radial data was reconstructed on the same 160×160 grid with the same field of view and spatial resolution as with the Cartesian experiment. All images were zero-filled to a matrix size of 256×256 for the following evaluation. The slice thickness was 1 mm in all experiments. Further relevant parameters of all three measurement types are listed in Table1.
Table 1
Scan parameters for the three flow-encoding sequences
Parameters
Cartesian
Radial triggered
Retrospective
TR [ms]
5
3
3
TE [ms]
1.7
1.2
1.2
Sampling bandwidth [kHz]
75.8
75.8
75.8
Flip angle [°]
30-40
30-40
30-40
Echo position
25
10
10
[% of the read-out]
   
Number of read-out points
160
90
90
Number of projections
160
160
16000*
Number of Cine segments
5
3
-
Frames per Cine Segment
8
14
-
Number of velocity-encoding steps
3
3
3
VENC [m/s]
1.7
1.7
1.7
Spatial resolution [ μ m 2 ]
150×150**
150×150
150×150
Field of view [mm 2 ]
24×24
24×24
24×24
Slice thickness [mm]
1
1
1
Measurement time [minutes]
6.7±1.1
4.3±0.6
2.9
*Including 500 dummy scans to establish the steady state. **Zero-filled to a 256 × 256 matrix. Reconstructed on a 160 × 160 grid and zero-filled to a 256 × 256 matrix.

Retrospectively triggered measurement

For the retrospective triggering mode a train of nProj=16000 radial projections per velocity-encoding step was acquired without cardiac and respiratory gating with a short repetition time (TR=3 ms).
To reduce the echo time and therefore the strong influence of B0 inhomogeneities at 17.6T, which are especially significant in vessels close to the lung, the echo was set to 10% of the read-out duration by decreasing the areas of the read dephase gradients (see Figure2). A waiting period of 15 s was applied between each velocity-encoding step to reduce the effect of RF coil heating and SAR due to the fast application of RF pulses[10]. The total scan time including the waiting times was 3 minutes.

Prospectively triggered measurement

All prospectively triggered measurements were done with a pressure sensitive balloon to measure the thoracic motion during breathing and the heartbeats. The trigger signal was post-processed with an amplification unit (Rapid Biomedical, Rimpar, Germany). With the triggered Cartesian measurement an effective temporal resolution of 1 ms was achieved by the acquisition of 5 interleaved Cines with 8 movie frames each, covering 40 ms of the heart cycle around the systole. The total scan time depends on breathing and the heart rate and was 6-8 minutes in this study. For the prospectively triggered radial measurement three interleaved Cines with 14 movie frames each were acquired to obtain the same temporal resolution as for the Cartesian measurement, covering 42 ms of the heart cycle. For each image 160 equally distributed spokes were acquired, therefore the undersampling factor was af ≈3.12, respective to the Nyquist criterion. The total scan time was 3-5 minutes.

Gradient delay correction

As a non-Cartesian imaging method the quality of the radial measurement is compromised by trajectory errors caused by hardware imperfections such as delays of the spatial encoding gradients. To improve image quality, the gradient switching times were determined priorly with a homogenous CuSO4 doped water phantom using two different methods as proposed in[11] and[12], respectively. Measurements using either technique yielded gradient delay values of about t x =8 μ s for the X-, t y =10 μ s for the Y- and t z =8 μ s for the Z-gradient. The measured parameters were afterwards used for a correction of the gradient areas by adjusting the amplitudes of the read dephase gradients, as described in[11].

Trigger signal processing

For the retrospective reconstruction several steps were performed to obtain the trigger signal:

Modified golden angle based radial read-out

In mice anesthetized with isoflurane the heart typically contracts with a repetition rate of 300-600 beats per minute, which is usually too fast for golden-ratio based real time Cine applications[13]. Therefore, to achieve sufficient SNR and temporal resolution, the aortic motion is monitored over several seconds. For the reconstruction of a Cine frame, the projections acquired during many heart beats were afterwards sorted by their phase in the cardiac cycle and used to fill the k-space. Since a standard golden-ratio based angular distribution might result into an undesired clustering of the retrospectively selected projections, a modified golden angle based read-out was precalculated.
Using θ0=0° as initial value, an angle list was calculated by consecutively adding the golden angle Δ θ≈111.246°[13]. Assuming that the variations from the average interval TRR between two heart beats are small, the angles were afterwards sorted into N ≈TRR/TR interleaved sublists using the sorting:
θ n ( k · T RR ) = θ n , 0 + k · Δθ ,
(1)
where n=1,…,N is the number index of the sublists, k = 1 , , K n Proj N the index of the projections per sublist and θn,0=θn-1(K·TRR) the last angle value of the previous sublist. Figure3 and an additional animation in the supplement show the used sorting in more detail (Additional file1). The idea of this sorting of projection angles is that in case of an almost constant heart rate a large amount of retrospectively selected projections for one time frame would belong to the same sublist and would therefore be golden-ratio distributed. An estimate of the average interval TRR was assessed with the pressure sensitive balloon signal, which was observed with the trigger unit to monitor cardiac and respiratory motion of the animal. Depending on the heart rate, the measured projections consist of 33-38 sublists with 420-475 golden angle-distributed projections per sublist.

Signal processing

For retrospective triggering the k-space center of the radial projections was taken. The real part, the imaginary part and the magnitude values were investigated to find the trigger signal with the strongest modulation due to blood flow. The first 1.5 s (i.e., 500 projections) of the data were excluded from data processing since during this time the signal shows a strong transient to the steady state.
Since the original signal is often compromised by high harmonic disturbances (Figure4a), a Butterworth low-pass filter[14] was applied to suppress high-frequency interferences (F4b). Afterwards a Gaussian-shaped window function was used to reduce low-frequency modulations caused by respiratory motion (F4c). F4d and e show the magnitude responses of the used frequency filters. For the Butterworth filter a pass band frequency of 10-13 Hz and a stop band frequency of 15-20 Hz were found to provide the best filtering results. The high-pass filter has the shape w(f)=1-exp(-(f/σ)2), with a window width of σ=4-8 Hz.

K-space selection

For the assignment of the acquired projections to a corresponding heart cycle phase, the trigger signal extracted from the radial k-space data, which was acquired with a sampling rate of TR=3 ms, was interpolated to 1/10 of TR using a linear interpolation. To select a particular phase within the heart cycle, trigger points were determined with thresholds (threshold value: 90% of the local maximum of the interpolated self-gating signal). The first value above the threshold was chosen as the trigger point (Figure5a). After determination of the trigger positions each radial projection was assigned to a relative position in the heart cycle between 0 and 1 (see Figure5b). Data points during respiratory motion were gated out (relative position = -1). For the reconstruction of relative time frames, the relative heart cycle positions were divided into 100 equidistant selection intervals (Figure5b). Projections located within a temporal distance of ±0.5 ms around the corresponding heart cycle phase were assigned to each selection interval. Using this sliding window selection, an average number of about 120 projections per image were assigned to 100 time frames of the aortic motion covering the whole heart cycle.

Reconstruction

All radial reconstructions were done in MATLAB (The MathWorks, Inc., Natick, MA) with a Non Uniform Fast Fourier Transform (NUFFT,[15]) using the software toolbox provided by Fessler et al.[16]. The complex k-space data was re-gridded with a Kaiser-Bessel window[17]. The density compensation function was calculated using the iterative algorithm described by Pipe et al.[18].
Due to imperfect shimming a retrospective correction of B0 offsets and linear inhomogeneities was applied for reconstruction. Therefore an off-resonance map was determined through the acquisition of two triggered gradient-echo images ϱ1 and ϱ2 with a relative echo time difference of Δ TE=0.5 ms, as described in[19]. With the phase of the conjugate complex product, ΔΦ ( r ) = angle ( ϱ 1 · ϱ 2 ) , the off-resonance frequencies can be calculated using Δ ω 0 ( r ) = ΔΦ ( r ) Δ TE . A phase unwrapping was applied, when necessary[20]. Using this a priori information the off-resonance corrected image ϱ ̂ ( r ) can be estimated with the conjugated phase reconstruction proposed in ([21], Eq. 7):
ϱ ̂ ( r ) = j s ( t j ) e ω 0 ( r ) t j C ( j , r ) .
(2)
Hereby j denotes a summation over the measured k-space points, s(t j ) the measured CMR signal and t j the discrete sampling times of the measurement. C ( j , r ) = e i k ( t j ) r w j is a complex weighting matrix which contains the information about the k-space trajectory k ( t j ) and the real-valued sampling densities w j calculated with the Pipe algorithm[18]. Since an exact execution of the conjugated phase reconstruction has high computational requirements due to the large dimensions of the trajectory and the off-resonance term[21], Equation (2) is solved numerically using the NUFFT and an approximation of the exponential term e ω 0 ( r ) t j , which is described in detail in[22, 23].

PWV calculation

For the calculation of the local pulse-wave velocity the flow-area (QA) method proposed by[24] was used. Assuming that a flow pulse through the aorta during the early systolic wave is reflection-free and unidirectional[24], the PWV can be determined using the approximation:
PWV = dQ dA ,
(3)
where dQ is the change of the volume flow during the early systole and dA is the change of the cross-sectional area of the vessel. For a segmentation of the aorta all images were zero-filled to a matrix size of 256×256. The cross-sectional area A(t) and the volume flow Q(t) were determined using AMIRA (Mercury Computer Systems, Chelmsford, USA) for a segmentation of the vessel for each time frame. All segmentations were done 4 times by the same observer to reduce the standard error. The velocity information was calculated with a linear fit to the phase data of each pixel within the aorta.
For a PWV determination the early upstroke of the measured Q(A)-curve was fitted with a linear function. Since the Q and A data estimated from the reconstructed Cine images is superimposed with high frequency random noise, both curves were smoothed using a moving average filter[25] in the temporal domain, as mentioned in[4].

Animal handling

A group of 6 C57BL/6J mice (Charles River Laboratories) aged 6 months was imaged in vivo. Anesthetization was done with isoflurane (1.5-2% volume 02 (2L/min)). The temperature was kept constant at 35°C using the gradient cooling unit. All mice were measured vertically (head up). All experimental procedures were in accordance with institutional and internationally recognized guidelines and were approved by the Regierung von Unterfranken (Government of Lower Franconia, Würzburg, Germany) to comply with German animal protection law. The reference number of the permit of the animal experiments is 55.2-2531.01-14/10.

Results

Figure6a displays a representative section of the self-gating signal of a motion compensated measurement (M1=0 s/m). The signal modulations due to breathing and the blood flow through the aorta are clearly visible. Figure6b shows a histogram of the time intervals between each found trigger point, which was calculated from the navigator signals of all three velocity-encoding steps. The average interval between two trigger peaks was 107.2±4.1 ms (mean value and variation estimated with a Gaussian fit). The distribution of the trigger point distances is in range of natural variations of the heart rate. Through the reconstruction of 100 movie frames of the aortic motion an average frame rate of 933±36 frames/s was achieved. In Figure7 representative flow compensated magnitude images of the aorta during the systolic heart phase are shown for the Cartesian, the triggered radial and the retrospective radial measurement (Figure7a to c upper row). The scan time was 7 minutes for the Cartesian measurement and 4 minutes for the triggered radial measurement. The retrospective measurement was done in only 3 minutes. For the non-triggered reconstruction an average number of 120 spokes could be used, which corresponds to an undersampling factor of more than 4 respective to the Nyquist criterion. The bottom images of Figure7a-c show phase maps corresponding to the same measurement. In the phase map of the retrospectively triggered measurement streaking artifacts are visible, however, the aortic flow is only slightly affected by these. The Figures8a and b display an overlay of the flow (Q) and the area (A) waveform for the triggered and retrospective measurement. The area curves of the three measurements are slightly shifted among each other due to uncertainties of the aorta segmentation. Also smaller values of the peak volume flow are noticeable with the prospectively triggered measurements, which might be the result of instabilities of the balloon signal. However, for a determination of the PWV only relative changes of the area- and flow curves are important[4]. All three flow-area plots show consistency with the hypothesized linear behavior during the early systole (Figure8c-e). For a determination of the pulse-wave velocity only the first data points after the beginning of the upstroke are relevant. With each the triggered Cartesian and the triggered radial measurement 8 data points (8 ms) of the early upstroke of the systolic-flow pulse could be used for a linear fit. The frame rate of the retrospective measurement was high enough to take 7 data points (≈7.5 ms) for the fit. Table2 shows the averaged results for the calculated parameters of all 6 monitored animals. All three methods have sufficient temporal resolution to sample the upstroke of the systolic volume flow with 5-8 data points, respectively. The average PWV determined with the Cartesian measurement was 2.4±0.2 m/s and for the prospectively triggered radial measurement 2.3±0.2 m/s. The PWV measured with the retrospectively triggered method was 2.3±0.2 m/s, which is in good agreement with the results of the prospectively triggered measurements.

Discussion

PWV values

This work shows that a self-gated measurement of the local pulse-wave velocity in the aorta is feasible. The average PWV values of the retrospectively triggered measurement are in good agreement with the triggered Cartesian and radial measurements. In[4] PWV values of 2.6±0.2 m/s are published for 8-months-old WT-mice using the prospectively triggered Cartesian PC-Cine measurement, which is in the same range as the data presented here. Zhao et al.[5] published values of 3.54±0.97 m/s for the PWV of 9-months-old WT-mice. In ultrasound measurements PWV values of 2.86±0.14 m/s are found for 8-month old B6D2F1 mice[26].
Table 2
Average results of the local PWV measurements in the abdominal aorta, determined with 6 animals
Parameters
Results
Weight (g)
24.9±3.1
Peak CSA [ mm 2 ]
 
Cartesian Measurement
0.75±0.05
Radial Triggered Measurement
0.75±0.04
Retrospective Measurement
0.73±0.04
Peak volume flow [ cm 3 /s]
 
Cartesian Measurement
0.27±0.04
Radial Triggered Measurement
0.28±0.03
Retrospective Measurement
0.28±0.04
PWV [m/s]
 
Cartesian Measurement
2.4±0.2
Radial Triggered Measurement
2.3±0.2
Retrospective Measurement
2.3±0.2
Results are presented as mean ± SE. PWV: Pulse-wave velocity. CSA: Cross-sectional-area.

Trigger quality

In contrast to measurements close to the beating heart only view moving voxels can be found in a slice in the abdominal region. Therefore the waveform of the navigator signal is less distinctive then in a cardiac Cine measurement and depends on the scale of the pulsating vessel. Since the navigator signal depends on changes of blood signal within the slice, the trigger quality is sensitive to the applied flip angle. Robust trigger signals were only found with flip angles of 30° and higher. With flip angles less then 20°, the retrospective triggering often failed.

Temporal resolution

The monitoring of the pulse-wave propagation requires high temporal resolution. With the triggered interleaved PC-Cine measurements described in[4] a frame rate of 1000 frames/s can be achieved. Since always a fixed number of 100 time frames were reconstructed, the frame rate of the retrospective measurement depends on the heart rate. In the examined group the average cardiac interval was 101-115 ms, therefore the absolute frame rate was 870-990 frames/s, which was sufficient to take 5-8 data points during the upstroke of the systolic flow for the linear fit. For measurements of animals with lower heart rates (175 - 200 ms) one should increase the number of reconstructed Cine frames to ensure sufficient temporal resolution, which might result into longer measurement times.
The native temporal resolution of the retrospective measurement is restricted to the repetition time of the read-out, which was 3 ms in this experiment. However, since the heart rate shows small variations due to naturalfluctuations of the heart function, the sampled states in the cardiac cycle slightly vary over the time, allowing a denser sampling of the aortic motion. As a result the radial projections could be assigned to relative positions in the heart cycle using a sliding window width of only 1 ms. Furthermore one could implement small TR variations in the timing of the pulse sequence to randomize even more the sampling of the heart periods. In Figure9 the average positions in the heart cycle and the standard deviations for the heart cycle positions of the selected radial projections of one retrospective Cine measurement are plotted as a function of the frame index (between 1 und 100). The mean variation per frame was ±0.75% of the heart cycle, which corresponds to a deviation of ≈±0.8 ms. Since the temporal frames therefore had an overlap of ≈50%, the reconstructed Cine is slightly blurred and the measured area-and flow-curves appear to be smoothed. However, a low-pass filter is usually applied to the flow data of the triggered measurements as well to suppress high frequency random noise[4], which results into a similar smoothing effect.

Measurement time

A drawback of the triggered Cartesian PWV measurement is the long scan time. To achieve a temporal resolution of 1 ms, 5 segmented Cines were necessary. The average scan time of the Cartesian experiment was 6.7±1.1 minutes (mean ± SE). Since the early systolic time frame is very short, only few data points can be used for linear fitting. Therefore the heart rate has to be kept constant during the whole measurement. With the use of a highly asymmetric radial trajectory, the scan time of the triggered PWV measurement could be reduced to 4.3±0.6 minutes without losing temporal and spatial resolution. Using the retrospective measurement the whole PWV experiment can be done in only 2.9 minutes, which is more than two times faster than the triggered Cartesian protocol.

Contrast

Compared to the prospectively triggered methods, where the MRI signal does not reach the steady state due to cardiac and respiratory gating, the MRI signal of the retrospective flow measurement is in steady state all the time. Therefore each retrospectively reconstructed image contains a comparable contrast, while contrast changes are noticeable with the triggered and respiratory-gated measurements. In addition, since the surrounding static tissue appears to be darker than with the triggered sequences, the retrospectively triggered method enables an easier segmentation of the aorta.

Deviations of the radial trajectory

The retrospective technique does not require external ECG probes and is therefore less sensitive to animal handling. However, off-resonance and local field inhomogeneities can compromise the quality of the reconstructed images and lead to deviations of the radial trajectory. Using highly asymmetric projections, the echo time can be shortened from 1.7 ms to 1.2 ms, reducing the effect of inhomogeneities and flow on the measurement. With the aid of B0 maps remaining susceptibility artifacts can be corrected retrospectively. To demonstrate the effect of the B0 correction an exemplary map of the off-resonance frequencies Δf ( r ) = Δ ω 0 ( r ) 2 π is shown in Figure10 (top: complete map, bottom: section around the aorta). Around the aorta the frequency offset lies around 600 Hz, causing severe blurring artifacts in the uncorrected radial image which make an segmentation of the vessel impossible (see images in the middle). Using the measured B0 map and the algorithm mentioned above the blurring can be significantly reduced, allowing a better distinction of the vessel from the surrounding tissue (see right images). However, the acquisition of the required triggered gradient-echo images for this optional B0 correction results into an additional measurement time of 1-2 minutes. Even with correction, B0 inhomogeneities remain the main cause for uncertainties of aorta segmentation and difficulties with PWV measurements in vessels close to lung tissue. However, issues with susceptibility artifacts in vessels nearby lung tissue were also reported for the triggered Cartesian measurement described in[4]. Another cause of trajectory deviations are timing errors caused by delays of the spatial encoding gradients[11]. By measuring the gradient switching times trajectory distortions can be reduced, however, residual trajectory errors and errors of the phase contrast maps caused by eddy currents[12, 27] can remain.

Advantages of the radial acquisition

In[28] and[29] a Cartesian self-gating method was proposed which measures a navigator echo during refocusing of the slice-selection gradient without the need of non-uniform trajectories. In[8] the rectilinear self-gating signal could be demonstrated to be more robust than the radial signal since it does not suffer from trajectory off-center effects. However, in this method the acquisition of the navigator signal was achieved by a temporal separation of the slice-refocusing gradient and the read dephase and phase encoding gradients. With a flow-encoding sequence based on a tripolar slice-selection gradient and a bipolar flow-encoding gradient this separation would result in an undesirable prolonging of the read-out duration and in a longer echo time. With the radial self-gating sequence proposed in this work a self-gating signal can be acquired without extending the timings of the flow-encoding sequence. Through the application of high echo asymmetry and short read dephase gradients the echo time could be reduced, which helps to reduce misregistration effects due to blood flow[30]. In addition, since every radial projection samples the k-space center and therefore has equal information content, an easier reconstruction of time frames is possible compared to the retrospective Cartesian sampling. As a further advantage, the proposed self-gated flow-encoding sequence profits from the advantages of the radial trajectory, due to the higher incoherency of the flow and motion artifacts[31, 32] and a higher robustness against undersampling[33].

Conclusions

We present a robust self-gated PC-Cine method that enables a measurement of the local pulse-wave velocity in the murine aorta without the need of additional triggering hardware. This new technique allows the reconstruction of high-resolution Cines of the aortic motion with a frame rate comparable to triggered methods achieving similar results for the PWV values.

Acknowledgements

This study was supported by the German Research Foundation (DFG) in the scope of Sonderforschungsbereich 688 (TP B4, B5, Z2), the Bundesministerium für Forschung und Bildung in the scope of BMBF01 E01004, the University of Würzburg in the funding program Open Access Publishing and the Comprehensive Heart Failure Center (Deutsches Zentrum für Herzinsuffizienz). We thank Carmen Bundschuh for the preparation of the animals, Norbert Elsaesser and Stephan Richter for developing a Matlab tool for phase unwrapping and David Alexander Kann and Katrin Winter for the help in preparing the manuscript.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License(http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

PW developed the radial flow-encoding sequences and conducted all experiments. ER, WRB, VH and PMJ enrolled and organized this study. TK and VH helped with the data evaluation. XH and VH helped with the pulse-sequence development, XH, CBM and FTG provided the software for trigger signal evaluation. All authors took part in critical review and drafting of the manuscript and have read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Blacher J, Guerin AP, Pannnier B, Marchais SJ, Safar ME, London GM: Impact of aortic stiffness on survival in end-stage renal disease. Circulation. 1999, 99: 2434-9. 10.1161/01.CIR.99.18.2434.CrossRefPubMed Blacher J, Guerin AP, Pannnier B, Marchais SJ, Safar ME, London GM: Impact of aortic stiffness on survival in end-stage renal disease. Circulation. 1999, 99: 2434-9. 10.1161/01.CIR.99.18.2434.CrossRefPubMed
2.
Zurück zum Zitat Laurent S, Katsahian S, Fassot C, Tropeano AI, Gautier I, Laloux B, Boutouyrie P: Aortic Stiffness is an independent Predictor of fatal Stroke in essential hypertension. Stroke. 2003, 34: 1203-6. 10.1161/01.STR.0000065428.03209.64.CrossRefPubMed Laurent S, Katsahian S, Fassot C, Tropeano AI, Gautier I, Laloux B, Boutouyrie P: Aortic Stiffness is an independent Predictor of fatal Stroke in essential hypertension. Stroke. 2003, 34: 1203-6. 10.1161/01.STR.0000065428.03209.64.CrossRefPubMed
3.
Zurück zum Zitat Greve JM, Les AS, Tang BT, Draney Blomme MT, Wilson NM, Dalman RL, Pelc NJ, Taylor CA: Allometric scaling of wall shear stress from mice to humans: quantification using cine phase-contrast MRI and computational fluid dynamics. Am J Physiol: Heart and Circ Physiol. 2006, 291: H1700-7. 10.1152/ajpheart.00274.2006. Greve JM, Les AS, Tang BT, Draney Blomme MT, Wilson NM, Dalman RL, Pelc NJ, Taylor CA: Allometric scaling of wall shear stress from mice to humans: quantification using cine phase-contrast MRI and computational fluid dynamics. Am J Physiol: Heart and Circ Physiol. 2006, 291: H1700-7. 10.1152/ajpheart.00274.2006.
4.
Zurück zum Zitat Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM: In vivo measurement of local aortic pulse-wave velocity in mice with MR microscopy at 17.6 tesla. Magn Reson Med. 2009, 61: 1293-9. 10.1002/mrm.21957.CrossRefPubMed Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM: In vivo measurement of local aortic pulse-wave velocity in mice with MR microscopy at 17.6 tesla. Magn Reson Med. 2009, 61: 1293-9. 10.1002/mrm.21957.CrossRefPubMed
5.
Zurück zum Zitat Zhao X, Pratt R, Wansapura J: Quantification of aortic compliance in mice using radial phase contrast MRI. J Magn Reson Imaging. 2009, 30: 286-91. 10.1002/jmri.21846.CrossRefPubMed Zhao X, Pratt R, Wansapura J: Quantification of aortic compliance in mice using radial phase contrast MRI. J Magn Reson Imaging. 2009, 30: 286-91. 10.1002/jmri.21846.CrossRefPubMed
6.
Zurück zum Zitat Polson MJR, Barker AT, Gardiner S: The effect of rapid rise-time magnetic fields on the ECG of the rat. Clin Phys Physiol Meas. 1982, 3 (3): 231-4. 10.1088/0143-0815/3/3/008.CrossRefPubMed Polson MJR, Barker AT, Gardiner S: The effect of rapid rise-time magnetic fields on the ECG of the rat. Clin Phys Physiol Meas. 1982, 3 (3): 231-4. 10.1088/0143-0815/3/3/008.CrossRefPubMed
7.
8.
Zurück zum Zitat Hiba B, Richard N, Janier M, Croisille P: Cardiac and respiratory self-gated cine MRI in the mouse at 7T. Magn Reson Med. 2006, 55: 506-13. 10.1002/mrm.20815.CrossRefPubMed Hiba B, Richard N, Janier M, Croisille P: Cardiac and respiratory self-gated cine MRI in the mouse at 7T. Magn Reson Med. 2006, 55: 506-13. 10.1002/mrm.20815.CrossRefPubMed
9.
Zurück zum Zitat Herold V, Mörchel P, Faber C, Rommel E, Haase A, Jakob PM: In vivo quantitative three-dimensional motion mapping of the murine myocardium with PC-MRI at 17.6T. Magn Reson Med. 2006, 55: 1058-64. 10.1002/mrm.20866.CrossRefPubMed Herold V, Mörchel P, Faber C, Rommel E, Haase A, Jakob PM: In vivo quantitative three-dimensional motion mapping of the murine myocardium with PC-MRI at 17.6T. Magn Reson Med. 2006, 55: 1058-64. 10.1002/mrm.20866.CrossRefPubMed
10.
Zurück zum Zitat Swoap SJ, Overton M, Garber G: Effect of ambient temperature on cardiovascular parameters in rats and mice: a comparative approach. Am J Physiol: Regul, Integr Comp Physiol. 2004, 287: R391-6. 10.1152/ajpregu.00731.2003. Swoap SJ, Overton M, Garber G: Effect of ambient temperature on cardiovascular parameters in rats and mice: a comparative approach. Am J Physiol: Regul, Integr Comp Physiol. 2004, 287: R391-6. 10.1152/ajpregu.00731.2003.
11.
Zurück zum Zitat Peters DC, Derbyshire JA, McVeigh ER: Centering the projection reconstruction trajectory: reducing gradient delay errors. Magn Reson Med. 2003, 50: 1-6. 10.1002/mrm.10501.PubMedCentralCrossRefPubMed Peters DC, Derbyshire JA, McVeigh ER: Centering the projection reconstruction trajectory: reducing gradient delay errors. Magn Reson Med. 2003, 50: 1-6. 10.1002/mrm.10501.PubMedCentralCrossRefPubMed
12.
Zurück zum Zitat Duyn JH, Yang Y, Frank JA, van der Veen JW: Simple correction method for k-space trajectory deviations in MRI. J Magn Reson. 1998, 132: 150-3. 10.1006/jmre.1998.1396.CrossRefPubMed Duyn JH, Yang Y, Frank JA, van der Veen JW: Simple correction method for k-space trajectory deviations in MRI. J Magn Reson. 1998, 132: 150-3. 10.1006/jmre.1998.1396.CrossRefPubMed
13.
Zurück zum Zitat Winkelmann S, Schaeffter T, Koehler T, Eggers H, Doessel O: An optimal radial profile order based on the golden ratio for time-resolved MRI. IEEE Trans Med Imaging. 2007, 26: 68-76.CrossRefPubMed Winkelmann S, Schaeffter T, Koehler T, Eggers H, Doessel O: An optimal radial profile order based on the golden ratio for time-resolved MRI. IEEE Trans Med Imaging. 2007, 26: 68-76.CrossRefPubMed
14.
Zurück zum Zitat Butterworth S: On the theory of filter amplifiers. Exp Wireless and the Wireless Eng. 1930, 7: 536-41. Butterworth S: On the theory of filter amplifiers. Exp Wireless and the Wireless Eng. 1930, 7: 536-41.
15.
Zurück zum Zitat Fessler JA, Sutton BP: Nonuniform fast fourier transform using min-max interpolation. IEEE Trans Signal Process. 2003, 51: 560-74. 10.1109/TSP.2002.807005.CrossRef Fessler JA, Sutton BP: Nonuniform fast fourier transform using min-max interpolation. IEEE Trans Signal Process. 2003, 51: 560-74. 10.1109/TSP.2002.807005.CrossRef
17.
Zurück zum Zitat Jackson JI, Meyer CH, Nishimura DG, Macovski A: Selection of a convolution function for fourier inversion using gridding. IEEE Trans Med Imaging. 1991, 10: 473-8. 10.1109/42.97598.CrossRefPubMed Jackson JI, Meyer CH, Nishimura DG, Macovski A: Selection of a convolution function for fourier inversion using gridding. IEEE Trans Med Imaging. 1991, 10: 473-8. 10.1109/42.97598.CrossRefPubMed
18.
Zurück zum Zitat Pipe JG, Menon P: Sampling density compensation in MRI: Rationale and an iterative numerical solution. Magn Reson Med. 1999, 41: 179-86. 10.1002/(SICI)1522-2594(199901)41:1<179::AID-MRM25>3.0.CO;2-V.CrossRefPubMed Pipe JG, Menon P: Sampling density compensation in MRI: Rationale and an iterative numerical solution. Magn Reson Med. 1999, 41: 179-86. 10.1002/(SICI)1522-2594(199901)41:1<179::AID-MRM25>3.0.CO;2-V.CrossRefPubMed
19.
Zurück zum Zitat Schneider E, Glover G: Rapid in vivo proton shimming. Magn Reson Med. 1991, 18: 335-47. 10.1002/mrm.1910180208.CrossRefPubMed Schneider E, Glover G: Rapid in vivo proton shimming. Magn Reson Med. 1991, 18: 335-47. 10.1002/mrm.1910180208.CrossRefPubMed
20.
Zurück zum Zitat Jenkinson M: Fast, automated, N-dimensional phase-unwrapping algorithm. Magn Reson Med. 2003, 49: 193-7. 10.1002/mrm.10354.CrossRefPubMed Jenkinson M: Fast, automated, N-dimensional phase-unwrapping algorithm. Magn Reson Med. 2003, 49: 193-7. 10.1002/mrm.10354.CrossRefPubMed
21.
Zurück zum Zitat Noll DC, Meyer CH, Pauly JM, Nishimura DG, Macovski A: A homogeneity correction method for magnetic resonance imaging with time-varying gradients. IEEE Trans Med Imaging. 1991, 10: 629-37. 10.1109/42.108599.CrossRefPubMed Noll DC, Meyer CH, Pauly JM, Nishimura DG, Macovski A: A homogeneity correction method for magnetic resonance imaging with time-varying gradients. IEEE Trans Med Imaging. 1991, 10: 629-37. 10.1109/42.108599.CrossRefPubMed
22.
Zurück zum Zitat Fessler JA, Sutton BP, Noll DC: Fast, iterative image reconstruction for MRI in the presence of field inhomogeneities. IEEE Trans Med Imaging. 2003, 22: 178-88. 10.1109/TMI.2002.808360.CrossRefPubMed Fessler JA, Sutton BP, Noll DC: Fast, iterative image reconstruction for MRI in the presence of field inhomogeneities. IEEE Trans Med Imaging. 2003, 22: 178-88. 10.1109/TMI.2002.808360.CrossRefPubMed
23.
Zurück zum Zitat Fessler JA, Lee S, Olafsson V, Shi HR, Noll DC: Toeplitz-based iterative image reconstruction for MRI with correction for magnetic field inhomogeneity. IEEE Trans Signal Process. 2005, 53 (9): 3393-402.CrossRef Fessler JA, Lee S, Olafsson V, Shi HR, Noll DC: Toeplitz-based iterative image reconstruction for MRI with correction for magnetic field inhomogeneity. IEEE Trans Signal Process. 2005, 53 (9): 3393-402.CrossRef
24.
Zurück zum Zitat Vulliémoz S, Stergiopulos N, Meuli R: Estimation of local aortic elastic properties with MRI. Magn Reson Med. 2002, 47: 649-54. 10.1002/mrm.10100.CrossRefPubMed Vulliémoz S, Stergiopulos N, Meuli R: Estimation of local aortic elastic properties with MRI. Magn Reson Med. 2002, 47: 649-54. 10.1002/mrm.10100.CrossRefPubMed
25.
Zurück zum Zitat Smith SW: Moving average filters. The Scientist and Engineer’s Guide to Digital Signal Processing. 1997, San Diego: California Technical Publishing, Chapter 15:277-284. [http://www.dspguide.com], Smith SW: Moving average filters. The Scientist and Engineer’s Guide to Digital Signal Processing. 1997, San Diego: California Technical Publishing, Chapter 15:277-284. [http://​www.​dspguide.​com],
26.
Zurück zum Zitat Reddy AK, Li YH, Pham TT, Ochoa LN, Trevino MT, Hartley CJ, Michael LH, Entman ML, Taffet GE: Measurement of aortic input impedance in mice: effects of age on aortic stiffness. Am J Physiol- Heart Circ Physiol. 2003, 285: H1464-70.CrossRefPubMed Reddy AK, Li YH, Pham TT, Ochoa LN, Trevino MT, Hartley CJ, Michael LH, Entman ML, Taffet GE: Measurement of aortic input impedance in mice: effects of age on aortic stiffness. Am J Physiol- Heart Circ Physiol. 2003, 285: H1464-70.CrossRefPubMed
27.
Zurück zum Zitat Bernstein MA, Zhou XJ, Polzin JA, King KF, Ganin A, Pelc NJ, Glover GH: Concomitant gradient terms in phase contrast MR: analysis and correction. Magn Reson Med. 1998, 39: 300-8. 10.1002/mrm.1910390218.CrossRefPubMed Bernstein MA, Zhou XJ, Polzin JA, King KF, Ganin A, Pelc NJ, Glover GH: Concomitant gradient terms in phase contrast MR: analysis and correction. Magn Reson Med. 1998, 39: 300-8. 10.1002/mrm.1910390218.CrossRefPubMed
28.
Zurück zum Zitat Esparaza-Coss E, Ramirez MS, Bankson JA: Wireless self-gated multiple-mouse cardiac cine MRI. Magn Reson Med. 2008, 59: 1203-6. 10.1002/mrm.21562.CrossRef Esparaza-Coss E, Ramirez MS, Bankson JA: Wireless self-gated multiple-mouse cardiac cine MRI. Magn Reson Med. 2008, 59: 1203-6. 10.1002/mrm.21562.CrossRef
29.
Zurück zum Zitat Bovens SM, te Boekhorst BCM, den Ouden K, van de Kolk KWA, Nauert A, Nederhoff MGJ, Pasterkamp G, ten Hove M, van Echteld CJA: Evaluation of infarcted murine heart function: Comparison of prospectively triggered with self-gated MRI. NMR in Biomed. 2010, 24: 307-315.CrossRef Bovens SM, te Boekhorst BCM, den Ouden K, van de Kolk KWA, Nauert A, Nederhoff MGJ, Pasterkamp G, ten Hove M, van Echteld CJA: Evaluation of infarcted murine heart function: Comparison of prospectively triggered with self-gated MRI. NMR in Biomed. 2010, 24: 307-315.CrossRef
30.
Zurück zum Zitat Firmin DN, Nayler GL, Kilner PJ, Longmore DB: The application of phase shifts in NMR of flow measurement. Magn Reson Med. 1990, 14: 230-41. 10.1002/mrm.1910140209.CrossRefPubMed Firmin DN, Nayler GL, Kilner PJ, Longmore DB: The application of phase shifts in NMR of flow measurement. Magn Reson Med. 1990, 14: 230-41. 10.1002/mrm.1910140209.CrossRefPubMed
31.
Zurück zum Zitat Nishimura DG, Jackson JI, Pauly JM: On the nature and reduction of the displacement artifact in flow images. Magn Reson Med. 1991, 22: 481-492. 10.1002/mrm.1910220255.CrossRefPubMed Nishimura DG, Jackson JI, Pauly JM: On the nature and reduction of the displacement artifact in flow images. Magn Reson Med. 1991, 22: 481-492. 10.1002/mrm.1910220255.CrossRefPubMed
32.
Zurück zum Zitat Glover GH, Pauly JM: Projection Reconstruction Techniques for Reduction of Motion Effects in MRI. Magn Reson Med. 1992, 28: 275-89. 10.1002/mrm.1910280209.CrossRefPubMed Glover GH, Pauly JM: Projection Reconstruction Techniques for Reduction of Motion Effects in MRI. Magn Reson Med. 1992, 28: 275-89. 10.1002/mrm.1910280209.CrossRefPubMed
33.
Zurück zum Zitat Peters DC, Korosec FR, Grist TM, Block WF, Holden JE, Vigen KK, Mistretta CA: Undersampled projection reconstruction applied to MR angiography. Magn Reson Med. 2000, 43: 91-101. 10.1002/(SICI)1522-2594(200001)43:1<91::AID-MRM11>3.0.CO;2-4.CrossRefPubMed Peters DC, Korosec FR, Grist TM, Block WF, Holden JE, Vigen KK, Mistretta CA: Undersampled projection reconstruction applied to MR angiography. Magn Reson Med. 2000, 43: 91-101. 10.1002/(SICI)1522-2594(200001)43:1<91::AID-MRM11>3.0.CO;2-4.CrossRefPubMed
Metadaten
Titel
Fast retrospectively triggered local pulse-wave velocity measurements in mice with CMR-microscopy using a radial trajectory
verfasst von
Patrick Winter
Thomas Kampf
Xavier Helluy
Fabian T Gutjahr
Cord B Meyer
Eberhard Rommel
Wolfgang R Bauer
Peter M Jakob
Volker Herold
Publikationsdatum
01.12.2013
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2013
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-88

Weitere Artikel der Ausgabe 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.