Fetal dynamic phase-contrast MR angiography using ultrasound gating and comparison with Doppler ultrasound measurements

verfasst von: B. P. Schoennagel, J. Yamamura, F. Kording, R. Fischer, P. Bannas, G. Adam, H. Kooijman, C. Ruprecht, K. Fehrs, M. Tavares de Sousa

Erschienen in: European Radiology | Ausgabe 8/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To investigate the feasibility of fetal phase-contrast (PC)-MR angiography of the descending aorta (AoD) using an MR-compatible Doppler ultrasound sensor (DUS) for fetal cardiac gating and to compare velocimetry with Doppler ultrasound measurements.

Methods

In this prospective study, 2D PC-MR angiography was performed in 12 human fetuses (mean gestational age 32.8 weeks) using an MR-compatible DUS for gating of the fetal heart at 1.5 T. Peak flow velocities in the fetal AoD were compared with Doppler ultrasound measurements performed on the same day. Reproducibility of PC-MR measurements was tested by repeated PC-MR in five fetuses.

Results

Dynamic PC-MR angiography in the AoD was successfully performed in all fetuses using the DUS, with an average fetal heart rate of 140 bpm (range 129–163). Time-velocity curves revealed typical arterial blood flow patterns. PC-MR mean flow velocity and mean flux were 21.2 cm/s (range 8.6–36.8) and 8.4 ml/s (range 3.2–14.6), respectively. A positive association between PC-MR mean flux and stroke volume with gestational age was obtained (r = 0.66, p = 0.02 and r = 0.63, p = 0.03). PC-MR and Doppler ultrasound peak velocities revealed a highly significant correlation (r = 0.8, p < 0.002). Peak velocities were lower for PC-MR with 69.1 cm/s (range 39–125) compared with 96.7 cm/s (range 60–142) for Doppler ultrasound (p < 0.001). Reproducibility of PC-MR was high (p > 0.05).

Conclusion

The MR-compatible DUS for fetal cardiac gating allows for PC-MR angiography in the fetal AoD. Comparison with Doppler ultrasound revealed a highly significant correlation of peak velocities with underestimation of PC-MR velocities. This new technique for direct fetal cardiac gating indicates the potential of PC-MR angiography for assessing fetal hemodynamics.

Key Points

• The developed MR-compatible Doppler ultrasound sensor allows direct fetal cardiac gating and can be used for prenatal dynamic cardiovascular MRI.
• The MR-compatible Doppler ultrasound sensor was successfully applied to perform intrauterine phase-contrast MR angiography of the fetal aorta, which revealed a highly significant correlation with Doppler ultrasound measurements.
• As fetal flow hemodynamics is an important parameter in the diagnosis and management of fetal pathologies, fetal phase-contrast MR angiography may offer an alternative imaging method in addition to Doppler ultrasound and develop as a second line tool in the evaluation of fetal flow hemodynamics.
Literatur
3.
Zurück zum Zitat Gill RW (1985) Measurement of blood flow by ultrasound: accuracy and sources of error. Ultrasound Med Biol 11:625–641CrossRefPubMed Gill RW (1985) Measurement of blood flow by ultrasound: accuracy and sources of error. Ultrasound Med Biol 11:625–641CrossRefPubMed
4.
Zurück zum Zitat Ranke C, Hendrickx P, Roth U, Brassel F, Creutzig A, Alexander K (1992) Color and conventional image-directed Doppler ultrasonography: accuracy and sources of error in quantitative blood flow measurements. J Clin Ultrasound 20:187–193 Ranke C, Hendrickx P, Roth U, Brassel F, Creutzig A, Alexander K (1992) Color and conventional image-directed Doppler ultrasonography: accuracy and sources of error in quantitative blood flow measurements. J Clin Ultrasound 20:187–193
15.
Zurück zum Zitat Rebergen SA, Chin JG, Ottenkamp J, van der Wall EE, de Roos A (1993) Pulmonary regurgitation in the late postoperative follow-up of tetralogy of Fallot. Volumetric quantitation by nuclear magnetic resonance velocity mapping. Circulation 88:2257–2266 Rebergen SA, Chin JG, Ottenkamp J, van der Wall EE, de Roos A (1993) Pulmonary regurgitation in the late postoperative follow-up of tetralogy of Fallot. Volumetric quantitation by nuclear magnetic resonance velocity mapping. Circulation 88:2257–2266
20.
21.
Zurück zum Zitat Feinstein JA, Epstein FH, Arai AE et al (1997) Using cardiac phase to order reconstruction (CAPTOR): a method to improve diastolic images. J Magn Reson Imaging 7:794–798CrossRefPubMed Feinstein JA, Epstein FH, Arai AE et al (1997) Using cardiac phase to order reconstruction (CAPTOR): a method to improve diastolic images. J Magn Reson Imaging 7:794–798CrossRefPubMed
25.
Zurück zum Zitat Seitz J, Strotzer M, Wild T et al (2001) Quantification of blood flow in the carotid arteries: comparison of Doppler ultrasound and three different phase-contrast magnetic resonance imaging sequences. Invest Radiol 36:642–647CrossRefPubMed Seitz J, Strotzer M, Wild T et al (2001) Quantification of blood flow in the carotid arteries: comparison of Doppler ultrasound and three different phase-contrast magnetic resonance imaging sequences. Invest Radiol 36:642–647CrossRefPubMed
27.
Zurück zum Zitat Shibata M, Sakuma H, Isaka N, Takeda K, Higgins CB, Nakano T (1999) Assessment of coronary flow reserve with fast cine phase contrast magnetic resonance imaging: comparison with measurement by Doppler guide wire. J Magn Reson Imaging 10:563–568 Shibata M, Sakuma H, Isaka N, Takeda K, Higgins CB, Nakano T (1999) Assessment of coronary flow reserve with fast cine phase contrast magnetic resonance imaging: comparison with measurement by Doppler guide wire. J Magn Reson Imaging 10:563–568
28.
Zurück zum Zitat Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E (2009) Accelerated phase-contrast MR imaging: comparison of k-t BLAST with SENSE and Doppler ultrasound for velocity and flow measurements in the aorta. J Magn Reson Imaging 29:817–824. https://doi.org/10.1002/jmri.21706 Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E (2009) Accelerated phase-contrast MR imaging: comparison of k-t BLAST with SENSE and Doppler ultrasound for velocity and flow measurements in the aorta. J Magn Reson Imaging 29:817–824. https://​doi.​org/​10.​1002/​jmri.​21706
29.
Zurück zum Zitat Engvall J, Sjöqvist L, Nylander E, Thuomas KA, Wranne B (1995) Biplane transoesophageal echocardiography, transthoracic Doppler, and magnetic resonance imaging in the assessment of coarctation of the aorta. Eur Heart J 16:1399–1409 Engvall J, Sjöqvist L, Nylander E, Thuomas KA, Wranne B (1995) Biplane transoesophageal echocardiography, transthoracic Doppler, and magnetic resonance imaging in the assessment of coarctation of the aorta. Eur Heart J 16:1399–1409
32.
Zurück zum Zitat Konje JC, Abrams K, Bell SC, de Chazal RC, Taylor DJ (2000) The application of color power angiography to the longitudinal quantification of blood flow volume in the fetal middle cerebral arteries, ascending aorta, descending aorta, and renal arteries during gestation. Am J Obstet Gynecol 182:393–400 Konje JC, Abrams K, Bell SC, de Chazal RC, Taylor DJ (2000) The application of color power angiography to the longitudinal quantification of blood flow volume in the fetal middle cerebral arteries, ascending aorta, descending aorta, and renal arteries during gestation. Am J Obstet Gynecol 182:393–400
33.
Zurück zum Zitat Walther FJ, Siassi B, King J, Wu PY (1986) Blood flow in the ascending and descending aorta in term newborn infants. Early Hum Dev 13:21–25CrossRefPubMed Walther FJ, Siassi B, King J, Wu PY (1986) Blood flow in the ascending and descending aorta in term newborn infants. Early Hum Dev 13:21–25CrossRefPubMed
34.
Zurück zum Zitat Marsál K, Lindblad A, Lingman G, Eik-Nes SH (1984) Blood flow in the fetal descending aorta; intrinsic factors affecting fetal blood flow, i.e. fetal breathing movements and cardiac arrhythmia. Ultrasound Med Biol 10:339–348CrossRefPubMed Marsál K, Lindblad A, Lingman G, Eik-Nes SH (1984) Blood flow in the fetal descending aorta; intrinsic factors affecting fetal blood flow, i.e. fetal breathing movements and cardiac arrhythmia. Ultrasound Med Biol 10:339–348CrossRefPubMed
35.
Zurück zum Zitat Davidson S, Sokolover N, Erlich A, Litwin A, Linder N, Sirota L (2008) New and improved Israeli reference of birth weight, birth length, and head circumference by gestational age: a hospital-based study. Isr Med Assoc J 10:130–134 Davidson S, Sokolover N, Erlich A, Litwin A, Linder N, Sirota L (2008) New and improved Israeli reference of birth weight, birth length, and head circumference by gestational age: a hospital-based study. Isr Med Assoc J 10:130–134
36.
Zurück zum Zitat van Eyck J, Wladimiroff JW, Noordam MJ, Tonge HM, Prechtl HF (1985) The blood flow velocity waveform in the fetal descending aorta: its relationship to fetal behavioural states in normal pregnancy at 37-38 weeks. Early Hum Dev 12:137–143 van Eyck J, Wladimiroff JW, Noordam MJ, Tonge HM, Prechtl HF (1985) The blood flow velocity waveform in the fetal descending aorta: its relationship to fetal behavioural states in normal pregnancy at 37-38 weeks. Early Hum Dev 12:137–143
37.
Zurück zum Zitat Hecher K, Campbell S, Doyle P, Harrington K, Nicolaides K (1995) Assessment of fetal compromise by Doppler ultrasound investigation of the fetal circulation. Arterial, intracardiac, and venous blood flow velocity studies. Circulation 91:129–138 Hecher K, Campbell S, Doyle P, Harrington K, Nicolaides K (1995) Assessment of fetal compromise by Doppler ultrasound investigation of the fetal circulation. Arterial, intracardiac, and venous blood flow velocity studies. Circulation 91:129–138
38.
Zurück zum Zitat Morris RK, Malin G, Robson SC, Kleijnen J, Zamora J, Khan KS (2011) Fetal umbilical artery Doppler to predict compromise of fetal/neonatal wellbeing in a high-risk population: systematic review and bivariate meta-analysis. Ultrasound Obstet Gynecol 37:135–142. https://doi.org/10.1002/uog.7767 Morris RK, Malin G, Robson SC, Kleijnen J, Zamora J, Khan KS (2011) Fetal umbilical artery Doppler to predict compromise of fetal/neonatal wellbeing in a high-risk population: systematic review and bivariate meta-analysis. Ultrasound Obstet Gynecol 37:135–142. https://​doi.​org/​10.​1002/​uog.​7767
42.
Zurück zum Zitat Sarwar A, Boland G, Monks A, Kruskal JB (2015) Metrics for radiologists in the era of value-based health care delivery. Radiographics 35:866–876CrossRefPubMed Sarwar A, Boland G, Monks A, Kruskal JB (2015) Metrics for radiologists in the era of value-based health care delivery. Radiographics 35:866–876CrossRefPubMed
Metadaten
Titel
Fetal dynamic phase-contrast MR angiography using ultrasound gating and comparison with Doppler ultrasound measurements
verfasst von
B. P. Schoennagel
J. Yamamura
F. Kording
R. Fischer
P. Bannas
G. Adam
H. Kooijman
C. Ruprecht
K. Fehrs
M. Tavares de Sousa
Publikationsdatum
07.01.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5940-y

Weitere Artikel der Ausgabe 8/2019

European Radiology 8/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.