Open Access 01.12.2017 | Research

‘Image-navigated 3-dimensional late gadolinium enhancement cardiovascular magnetic resonance imaging: feasibility and initial clinical results’

verfasst von: Konstantinos Bratis, Markus Henningsson, Chrysanthos Grigoratos, Matteo Dell’Omodarme, Konstantinos Chasapides, Rene Botnar, Eike Nagel

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2017

Abstract

Background

Image-navigated 3-dimensional late gadolinium enhancement (iNAV-3D LGE) is an advanced imaging technique that allows for direct respiratory motion correction of the heart. Its feasibility in a routine clinical setting has not been validated.

Methods

Twenty-three consecutive patients referred for cardiovascular magnetic resonance (CMR) examination including late gadolinium enhancement (LGE) imaging were prospectively enrolled. Image-navigated free-breathing 3-dimensional (3D) T1-weighted gradient-echo LGE and two-dimensional (2D LGE) images were acquired in random order on a 1.5 T CMR system. Images were assessed for global, segmental LGE detection and transmural extent. Objective image quality including signal-to-noise (SNR), contrast-to-noise (CNR) and myocardial/blood sharpness were performed.

Results

Interpretable images were obtained in all 2D–LGE and in 22/23 iNAV-3D LGE exams, resulting in a total of 22 datasets and 352 segments. LGE was detected in 5 patients with ischemic pattern, in 7 with non-ischemic pattern, while it was absent in 10 cases. There was an excellent agreement between 2D and 3D data sets with regard to global, segmental LGE detection and transmurality. Blood-myocardium sharpness measurements were also comparable between the two techniques. SNRblood and CNRblood-myo was significantly higher for 2D LGE (P < 0.001, respectively), while SNRmyo was not statistically significant between 2D LGE and iNAV-3D LGE.

Conclusion

Diagnostic performance of iNAV-3D LGE was comparable to 2D LGE in a prospective clinical setting. SNRblood and CNRblood-myo was significantly lower in the iNAV-3D LGE group.
Abkürzungen
2D
Two-dimensional
3D
Three-dimensional imaging
CMR
Cardiovascular magnetic resonance
CNR
Contrast-to-noise ratio
ECG
Electrocardiogram
iNAV-3D
Image-navigated three-dimensional cardiac magnetic resonance imaging
LGE
Late gadolinium enhancement
PSIR
Phase sensitive inversion recovery
SNR
Signal-to-noise ratio

Background

Cardiovascular magnetic resonance (CMR) late gadolinium enhancement imaging (LGE) is the standard of reference for the detection of myocardial necrosis [1]. The principle of LGE imaging is based on increased signal intensity in the damaged myocardium (compared to healthy) due to the increased extracellular space in necrotic tissue. The development of inversion recovery techniques together with electrocardiographic (ECG)-synchronization and breath holding has dramatically improved contrast between infarcted and normal myocardium [2].
Historically, a two-dimensional (2D) inversion recovery fast spoiled gradient-echo sequence has been applied for LGE imaging, necessitating multiple breath holds with potential slice misregistration and constraints on spatial resolution and signal-to-noise ratio (SNR) [3]. More recently, the development of 3-dimensional (3D) imaging has enabled data acquisition for the entire heart in a single scan [46]. Previously proposed 3D LGE methods included accelerated and extended breath-hold scans with similar constraints on spatial resolution and SNR as 2D LGE [7, 8]. This may be particularly detrimental to the detection of subendocardial scar which requires high resolution LGE. Alternatively, 3D LGE can be acquired during free-breathing which permits high resolution 3D LGE of the whole heart. To reduce respiratory induced motion artifacts, compensation techniques can be employed including the use of respiratory bellows signal [9] or diaphragmatic navigator [1014]. Nevertheless, these techniques have limited motion estimation accuracy as they only indirectly measure the respiratory motion of the heart.
In order to address the limitations of the diaphragmatic navigator, respiratory self-navigation has been developed. With this strategy, the position of the heart itself is monitored over time. Initially developed for coronary artery visualization [15], this technique has been used for assessment of complex congenital cardiac malformations [16, 17]. More recently, image-navigation has been proposed for respiratory motion compensation in CMR, which provides direct respiratory motion tracking of the heart and can be combined with respiratory gating algorithms to improve robustness to motion [18].
The aim of this study was to prospectively examine the feasibility of image-navigated 3D LGE CMR (iNAV-3D LGE) for the detection of LGE in a routine clinical setting and perform a head-to-head comparison against conventional 2D LGE.

Methods

Patients

Consecutive patients with known or suspected heart disease that underwent a CMR examination including LGE imaging at our facility were enrolled from February 2014 to February 2015. Patients were required to be ≥18 years of age and have no contraindications to gadolinium contrast, inclusive of an estimated glomerular filtration rate ≤ 60 ml/min/1.73 m2. Patients with atrial fibrillation were excluded. Written informed consent was obtained from all participants. The institutional ethics board at Guy’s & St Thomas’ Hospital, King’s College London approved the study.

CMR protocol

All participants were examined in supine position in a 1.5 Tesla CMR system (Gyroscan Intera, Philips Medical Systems, Best, The Netherlands) using a 32-element phased array cardiac synergy coil. The imaging protocol included cine imaging, standard 2D LGE imaging and iNAV-3D LGE imaging. 2D and iNAV-3D LGE image acquisitions were performed in random order to cancel out differences due to acquisition time after contrast agent injection.
Cine images were acquired using a ECG-gated balanced steady-state free precession based pulse sequence in serial short-axis slices from the atrioventricular annulus to the apex at 10 mm intervals, as well as in long-axis orientations (slice thickness 8 mm, gap 2 mm, echo time 1.5 ms, repetition time 3.0 ms, flip angle 50°). For LGE imaging, an intravenous bolus of 0.2 mmol/kg gadolinium (Gadovist®, Bayer Inc., Toronto, Ontario, Canada) was administered. High-resolution (2 mm isotropic) iNav-3D LGE and 2D LGE using T1-weighted RF-spoiled gradient-echo inversion recovery sequences were performed in random order 10 min following contrast administration. The image-navigator approach has been previously implemented for coronary CMR angiography [19], and was acquired using the ramp-up profiles of the 3D LGE sequence, as illustrated in Fig. 1. The ramp-up scheme consisted of linearly increasing amplitude of 10 RF pulses prior to the 3D LGE sequence. Phase encoding gradients were added to the ramp-up profiles to enable iNAV acquisition, using a high-low profile order. A high-low profile order was used to obtain higher signal for the central k-space line of the navigator, which was acquired with the highest flip angle, as well as closest possible temporal distance to the 3D LGE acquisition. As the ramp-up profiles were used to generate the iNAVs, size, location and orientation of the field-of-view for the iNAVs and 3D LGE were identical. To maximize sensitivity to detect respiratory motion for the iNAVs, the field-of-view was oriented in the coronal plane covering the whole heart, with readout in foot-head direction (2 mm resolution) and phase encoding in left-right direction (9 mm resolution). In the slice-encoding direction the iNAVs were projections of the field-of-view resulting in a slice thickness of approximately 100–120 mm. iNAV motion compensation was implemented in real-time on the scanner and has been previously described [20]. In brief, the location of the shim geometry over covering the heart was used to define a region-of-interest in the first iNAV, which was used as reference iNAV. The reference was registered to every subsequent iNAV using 2D normalized cross-correlation, yielding translational motion in foot-head and left-right direction. The estimated motion was used to perform translational correction of the 3D LGE raw data by applying a linear phase shift. Respiratory gating was implemented with constant respiratory efficiency using single end-expiratory threshold CRUISE algorithm [18]. This approach acquires twice as much data as needed to fill 3D LGE k-space (resulting in exactly 50% gating efficiency) and only the half acquired at the most end-expiratory was used to reconstruct the gated image. Data acquisition of the iNAV-3D LGE was ECG-triggered to the mid-diastolic rest period of every cardiac cycle. The 2D LGE was acquired with multiple slices, covering the entire left ventricle in short axis orientation, in addition to single slices in two-, three- and four-chamber views. The 2D LGE consisted of a phase sensitive inversion recovery (PSIR) protocol with inversion pulses every other RR interval. Separate Look-Locker scans were performed prior to iNAV-3D and 2D LGE, to find the optimal inversion times to null healthy myocardium for one and two RR-interval inversion pulse frequencies, respectively. The scan times were recorded for both iNAV-3D and 2D LGE. For the 2D LGE, the scan time was recorded as the time between the acquisition of the first and last slice, including pauses between breath-holds to capture the true scan time of all 2D slices.
Table 1 describes the scan parameters for each method.
Table 1
Scan parameters for iNAV-3D and conventional 2D late gadolinium enhancement (LGE) methods. iNAV-3D: Image-Navigated High-Resolution 3-Dimensional, PSIR: phase sensitive inversion recovery
Parameter
iNAV-3D LGE
2D LGE
Inversion pulse
Every heart beat
Every 2nd heart beat
PSIR
No
Yes
Repetition time (ms)
4.8 ms
5 ms
Echo time (ms)
1.5 ms
1.8 ms
Flip angle
25
25
Parallel imaging factor
2
1.5
Field of view (mm3)
320 × 320 × 120 mm
320 × 320 × 80 mm
Acquired resolution (mm)
2.0 × 2.0 × 2.0 mm3
1.25 × 1.25 × 10 mm3
Reconstructed resolution (mm)
1.0 × 1.0 × 1.0 mm3
0.62 × 0.62 × 10 mm3
Nominal scan duration
1 min 58 s
15 s/ breath holda
a: net duration of acquisition: 1 min 35 s, total scan duration including breath holds: 9 min 57 s

CMR post processing

Post processing was performed with dedicated software (Circle Cardiovascular Imaging 4.2, Circle Cardiovascular Imaging, Calgary, Canada) by two independent cardiologists specializing in CMR imaging (7-years experience in CMR/ SCMR Level 3 accredited and 3-years experience in CMR/ SCMR Level 2 accredited, respectively), who were blinded to the results of prior measurements by the same reader, measurements by other readers and all clinical data, as well as the order in which the 2D and iNAV-3D LGE studies had been acquired. All 2D and 3D data were anonymised and presented in random order.
3D LGE image data were reformatted to short-axis planes to correspond with the 2D LGE slices and optimise comparisons. On both the 2D and iNAV-3D LGE images, endocardial and epicardial contours were automatically placed by the software and reviewed by the reader. After segmentation, diagnostic performance and image quality data analysis of 2D LGE and iNAV-3D images was performed. The acquired images were analyzed with regard to the agreement of the global and segmental detection, pattern and transmurality of LGE for each segment of a 16-segment model of the myocardium proposed by the American Heart Association excluding the apex [21].
Quantitative image sharpness was calculated for the 2D and 3D iNAV-LGE datasets. For each patient, a short-axis slice with the least amount of scar and myocardial trabeculations was selected for sharpness analysis. The corresponding slice in the iNAV-3D LGE was reformatted to the same reconstructed resolution. In each image, 8 profiles were manually selected along the endocardial border, at equidistant locations, as shown in Fig. 2. The sharpness was defined as the distance in pixels between 20% and 80% of the pixel intensity range of the profile, and a lower pixel distance indicate a sharper border. The sharpness from the 8 profiles was averaged for each dataset and patient. To ensure measurements were comparable between patients, all images were reformatted to the same in-plane resolution of 0.67 mm.
SNR was calculated for 2D LGE and iNAV-3D LGE for blood (SNRblood) and myocardium (SNRmyo). The analysis was performed by drawing region of interests in the left ventricular blood pool and septum of the myocardium to obtain signal from the respective tissues. Noise was defined as the standard deviation of a region of air in the lungs (SDair), avoiding pulmonary vessels. The SNR values were calculated as the respective signal divided by SDair, multiplied by a factor of 0.655 to account for the Rayleigh distribution of the noise [22]. Furthermore, blood-myocardium contrast-to-noise (CNRblood-myo) was calculated using the same regions-of-interest as for the SNR calculations, subtracting SNRmyo from SNRblood, dividing by SDair and multiplying by 0.655 to account for the noise distribution.

Diagnostic quality scores

To obtain a head-to-head direct comparison between the two sequences only complete and diagnostic sequence sets were used for further analysis.
At first, visual perusal of the images was performed to determine whether infarction was present or absent. Positive finding in terms of a detectable scar in the LGE images was defined as a visible late enhancement. Global and segmental LGE detection were subsequently characterized in terms of LGE pattern (subendocardial/ ischemic, subepicardial, patchy, diffuse, involving right ventricular insertion points). In datasets with discrepancy between the two techniques, the results were compared against a panel standard. LGE transmurality was visually assessed by the reader as the segmental spatial extend of LGE within each segment and graded by using a five-point scale (0: 0%; 1: 1%–24%; 2: 25%–49%; 3, 50%–74%; 4, 75%–100%; 5, striae; 6- diffuse).

Statistical analysis

Global agreement between the two acquisition techniques was assessed by means of McNemar test to account for patient pairing. Agreement between the two techniques in segmental scar detection was determined with a marginal homogeneity test. The difference in proportions for image quality, recoded as excellent/ non excellent, were evaluated by the McNemar test. The scan times were compared using paired t-test. The statistical analysis was performed using R 3.1.2. (R Project, Vienna, Austria). The statistical significance was assumed for P < 0.05.

Results

Baseline characteristics

A total of 23 patients were examined. Interpretable images were obtained in all 2D and in 22/23 iNAV-3D exams, resulting in a total of 22 complete sequence datasets and 352 segments. One iNAV-3D LGE dataset had to be discarded due to fast arrhythmia. The average acquisition time for iNAV-3D LGE was 3 min and 52 s with a standard deviation of 1 min and 39 s. Corresponding values for the stack of 2D LGE slices was 9 min and 57 s ± 2 min and 34 s. The difference in scan time was statistically significant (P < 0.001).

Diagnostic quality

LGE was detected in 5 patients with ischemic pattern (23%), in 7 with non-ischemic pattern (32%), while it was absent in 10 cases (45%). There was excellent agreement between 2D and 3D data sets with regard to global LGE detection. There were no significant differences between the two techniques in segmental LGE detection and transmural extent (P = N.S.). The 95% confidence intervals and global tests showed no significant differences in the LGE pattern among all the considered combinations (P = N.S.). Similar observations were noted when the effect of the order of each sequence acquisition was considered.
Table 2 summarizes the results of diagnostic quality assessment. Typical case examples in patients with LGE are provided in Fig. 3.
Table 2
Comparison of the main diagnostic quality scores for 2D and iNAV-3D LGE iNAV-3D LGE: Image-Navigated Late Gadolinium Enhancement, n: number, NS: not significant
Diagnostic performance
Global LGE detection
2D (n = 22)
55% (12)
iNAV-3D (n = 22)
55% (12)
Segmental LGE detectiona (P = 0.28)
 
0
1
2
3
4
5
2D (n = 352)
75.9% (267)
10.2% (36)
3.7% (13)
7.7% (27)
2% (7)
0.6% (2)
iNAV-3D (n = 352)
78.1% (275)
8.5% (30)
2% (7)
7.4% (26)
3.4% (12)
0.6% (2)
LGE transmural extentb (P = NS)
 
0
1
2
3
4
5
6
2D (n = 352)
268
6
7
10
13
14
34
iNAV-3D (n = 352)
275
3
3
5
19
30
17
aSegmental LGE detection: 0 = no LGE, 1 = ischaemic, 2 = patchy, 3 = subepicardial, 4 = mid wall, 5 = RV insertion points bTransmural extension: 0, 1 = 1–25%, 2 = 26–50%, 3 = 51–75%, 4 = 76–100%, 5 = striae, 6 = diffuse

Blood-myocardium sharpness

The mean blood-myocardium sharpness ± standard deviation was 8.5 ± 3.6 for 2D LGE and 9.4 ± 3.0 for iNAV-3D LGE. No statistically significant differences were found between 2D LGE and iNAV 3D LGE for the blood-myocardium sharpness measurements (P = N.S.). Examples from two patients with improved image sharpness obtained using iNAV-3D LGE compared to 2D LGE are shown in Fig. 4.

SNR and CNR

The SNRblood, SNRmyo and CNRblood-myo for 2D LGE and iNAV-3D LGE are summarised in Table 3. Statistically significant differences were found for SNRblood (P < 0.001) and CNRblood-myo (P < 0.001), showing higher SNRblood and CNRblood-myo for 2D LGE, while SNRmyo was not statistically significant between 2D LGE and iNAV-3D LGE.
Table 3
Signal-to-noise (SNR) and contrast-to-noise (CNR) for blood and myocardium
 
2D LGE
iNAV-3D LGE
P
SNRblood
26.1 ± 12.2
12.0 ± 3.8
0.001a
SNRmyo
2.4 ± 1.1
2.1 ± 1.0
NS
CNRblood-myo
23.7 ± 11.1
9.9 ± 3.3
0.001a
aDenotes statistically significant differences. NS: not significant

Discussion

In our study, we demonstrated the feasibility of using image-navigated 3D LGE imaging in a clinical setting for the detection of LGE. The diagnostic and image quality of the iNAV-3D LGE method were in very good agreement with conventional 2D LGE imaging.
Image-based navigation has previously been restricted to 3D coronary CMR angiography, but here we have adapted and extended the method to allow free-breathing high resolution 3D LGE as well. This study describes for the first time an advanced, image-based motion compensation technique deployed for 3D LGE in a clinical setting. The sequence was shown to shorten LGE scan time, while providing similar image quality based on a subjective assessment. With the image-navigated technique the respiratory position of the heart itself is monitored over time. The position of the heart at the beginning of each data segment is thus compared with a reference position, i.e. the position of the ventricular blood-pool at the very first data segment, and automatically corrected for respiratory motion prior to image reconstruction. The acquired volumetric data with isotropic resolution can be retrospectively reformatted and specific 2D images can be extracted in any plane orientation during post-processing.
Significant differences in SNRblood and CNRblood-myo were found between 2D LGE and iNAV-3D LGE, in favor of 2D LGE. This may be attributed to the smaller voxel size of iNAV-3D LGE (8 mm3) compared to 2D LGE (15.6 mm3). Furthermore, 2D LGE was acquired using inversion pulses every other RR-interval, leading to larger contrast between blood and myocardium compared to inversion repetitions every RR-interval that was used for iNAV-3D LGE. A limitation of the SNR and CNR analysis is that the noise estimation may be inaccurate due to the use of parallel imaging. Ideally, a noise scan should be acquired to accurately estimate the noise if parallel imaging is employed [23]. However, due to the additional scan time required to acquire the noise scan, this could not be included in this clinical study.
The image-navigated 3D cardiac imaging sequence is characterized by improved ease of use and a low operator interaction. This sequence yields high-resolution images and besides the detection of LGE, its quality supports its use to provide accurate estimates of total scar burden and to generate volumetric models of scar distribution. This is essential as the burden of myocardial scar in ischemic heart disease is an important predictor of functional recovery after revascularization [24] and particularly useful in the modern clinical trials where CMR infarct size is an indispensable surrogate endpoint.
The temporal resolution of iNAV-3D LGE can be individually adapted to the heart rate, resulting in a 3 to 5 min acquisition duration, which can be preset, compared to approximately 10 min for the stack of 2D LGE images. The significantly increased scan time using a stack of 2D LGE images is largely due to the need for patient recovery time between breath-holds, which substantially increases the overall scan time. Although the isotropic 3D scan requires reformatting to identify the conventional scan planes, this is done retrospectively and does not extend the scan time.
Furthermore, due to the non-isotropic resolution of 2D LGE, additional 2D slices are often required to visualize LGE, which is not adequately captured in standard short axis, two-, three- or four-chamber views, which further increase scan time. In contrast, iNAV-3D LGE is acquired with high isotropic resolution and with whole-heart coverage that enables retrospective reconstruction of the data in any view. The proposed free-breathing technique may potentially be of particular value as an alternative to conventional breath-held 2D LGE in patients with poor compliance.
A limitation of the proposed iNAV 3D LGE technique is that motion compensation is restricted to foot-head and left-right motion, ignoring any anterior-posterior motion of the heart. In 18/22 iNAV-3D LGE cases residual respiratory motion caused visible blurring (Fig. 3, f-h). As 2D LGE images were acquired during breath-hold, and therefore are motion free, in practice it may be technically extremely challenging to similarly completely eliminate respiratory motion in a generalizable motion compensation approach. The residual motion artifacts may explain the increased transmural extent of LGE measurements using iNAV-3D LGE, compared to 2D LGE, due to motion related blurring of the signal. Technical improvements including the development of 3D motion compensation algorithms [2527] and 3D non-rigid motion correction [28, 29] may further improve image quality.
A further challenge of 3D LGE, which has not been addressed here, is the change in inversion time due to contrast washout during the scan [30]. Artifacts arising from sub-optimal inversion time from contrast washout may be exacerbated by the use of respiratory gating which prolongs the scan duration. Accelerated imaging may mitigate artifacts due to changes in inversion time and heart rate variability [31].
One dataset was discarded due to arrhythmia for iNAV 3D LGE and additional technical development is necessary to improve robustness to heart-rate variability and arrhythmia. This could include extension to PSIR LGE which is inherently more robust as the inversion pulse is performed every two heart-beats, leading to more stable signal even in the event of varying RR-intervals. Image-based navigation is particularly sensitive to arrhythmia, compared to conventional diaphragmatic navigation, as cardiac motion may be interpreted as respiratory motion. Additionally, signal and contrast fluctuations are more dramatic for inversion-recovery sequences in the event of arrhythmia, and this may also cause iNAV registration errors. Further work will focus on the implementation of arrhythmia rejection in conjunction with iNAV 3D LGE.
An advantage of the iNAV approach compared to conventional diaphragmatic navigator is that no navigator re-inversion pulse is required to restore the signal. This may be particularly useful for LGE of the pulmonary veins and left atrium where the navigator restore pulses cause signal enhancement mimicking LGE [32]. Nevertheless, due to the absence of navigator restore pulse the iNAV is susceptible to the same magnetization evolution as the 3D LGE, leading to a lower iNAV SNR. This may in turn impact the iNAV motion estimation precision. In this context, within the 10–25 min time window for LGE, performing iNAV-3D LGE earlier after contrast rather than later may be beneficial to improve signal from the blood pool, which could aid iNAV motion estimation. A limitation of this study is that no direct comparison between iNAV-3D LGE and diaphragmatic navigator was possible due to the time constraints of the CMR protocol. The diaphragmatic navigator approach, with a fixed, narrow gating window may result in low scan efficiency in the case of irregular breathing, leading to excessive scan times. In this study in patients referred for 2D LGE adding both iNAV-3D LGE and 3D LGE using conventional diaphragmatic navigator was not feasible. Further studies are required to directly compare the performance of iNAV and diaphragmatic navigator for 3D LGE.
There was no pathological reference standard for true infarct assessment. As our aim was to validate the feasibility of iNAV-3D method in the detection of late gadolinium enhancement in clinical practice comparison against a validated, conventional technique deemed sufficient. Patients with atrial fibrillation were excluded from the study, as conventional 2D LGE is known to be prone to arrhythmia-mediated artifacts and therefore the results would be significantly biased. The small sample size can hide some differences because of lack of statistical power. While the scope of the study was mainly to demonstrate feasibility of iNAV-3D LGE, further ongoing research projects recruiting larger number patients aim to more accurately determine diagnostic performance of this new technique.

Conclusions

In this study, we have examined the feasibility of an image-navigated 3D LGE method in a clinical routine setting. The novel technique provides high isotropic spatial resolution LGE in a shorter scan time compared to conventional 2D LGE. The diagnostic quality of this imaging technique supports its use in clinical practice for accurate estimates of scar burden.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Acknowledgements

Not applicable.

Funding

The authors acknowledge financial support from the Department of Health through the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award to Guy’s & St Thomas’ NHS Foundation Trust in partnership with King’s College London and King’s College Hospital NHS Foundation Trust. The Division of Imaging Sciences receives also support as the Centre of Excellence in Medical Engineering (funded by the Welcome Trust and EPSRC; grant number WT 088641/Z/09/Z) as well as the British Heart Foundation (RG/12/1/29262) and the BHF Centre of Excellence (RE/08/03). Professor Eike Nagel is supported by the German Centre for Cardiovascular Research (DZHK).

Availability of data and materials

The datasets used and/or analyzed during the current study available from the corresponding author on reasonable request.
Written informed consent was obtained from all participants. The institutional ethics board of Guy’s & St Thomas’ Hospital, King’s College London approved the study.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Kim HW, Farzaneh-Far A, Kim RJ. Cardiovascular magnetic resonance in patients with myocardial infarction: current and emerging applications. J Am Coll Cardiol. 2009;55(1):1–16.CrossRefPubMed Kim HW, Farzaneh-Far A, Kim RJ. Cardiovascular magnetic resonance in patients with myocardial infarction: current and emerging applications. J Am Coll Cardiol. 2009;55(1):1–16.CrossRefPubMed
2.
Zurück zum Zitat Wagner A, Mahrholdt H, Thomson L, Hager S, Meinhardt G, Rehwald W, Parker M, Shah D, Sechtem U, Kim RJ, Judd RM. Effects of time, dose, and inversion time for acute myocardial infarct size measurements based on magnetic resonance imaging-delayed contrast enhancement. J Am Coll Cardiol. 2006;47(10):2027–33.CrossRefPubMed Wagner A, Mahrholdt H, Thomson L, Hager S, Meinhardt G, Rehwald W, Parker M, Shah D, Sechtem U, Kim RJ, Judd RM. Effects of time, dose, and inversion time for acute myocardial infarct size measurements based on magnetic resonance imaging-delayed contrast enhancement. J Am Coll Cardiol. 2006;47(10):2027–33.CrossRefPubMed
3.
Zurück zum Zitat Kim RJ, Shah DJ, Judd RM. How we perform delayed enhancement imaging. J Cardiovasc Magn Reson. 2003;5(3):505–14.CrossRefPubMed Kim RJ, Shah DJ, Judd RM. How we perform delayed enhancement imaging. J Cardiovasc Magn Reson. 2003;5(3):505–14.CrossRefPubMed
4.
Zurück zum Zitat Foo TK, Stanley DW, Castillo E, Rochitte CE, Wang Y, Lima JA, Bluemke DA, Wu KC. Myocardial viability: breath-hold 3D MR imaging of delayed hyperenhancement with variable sampling in time. Radiology. 2004;230(3):845–51.CrossRefPubMed Foo TK, Stanley DW, Castillo E, Rochitte CE, Wang Y, Lima JA, Bluemke DA, Wu KC. Myocardial viability: breath-hold 3D MR imaging of delayed hyperenhancement with variable sampling in time. Radiology. 2004;230(3):845–51.CrossRefPubMed
5.
Zurück zum Zitat Dewey M, Laule M, Taupitz M, Kaufels N, Hamm B, Kivelitz D. Myocardial viability: assessment with three-dimensional MR imaging in pigs and patients. Radiology. 2006;239(3):703–9.CrossRefPubMed Dewey M, Laule M, Taupitz M, Kaufels N, Hamm B, Kivelitz D. Myocardial viability: assessment with three-dimensional MR imaging in pigs and patients. Radiology. 2006;239(3):703–9.CrossRefPubMed
6.
Zurück zum Zitat Bratis K, Henningsson M, Grigoratos C, Omodarme MD, Chasapides K, Botnar R, Nagel E. Clinical evaluation of three-dimensional late enhancement MRI. J Magn Reson Imaging. 2017;45(6):1675–83.CrossRefPubMed Bratis K, Henningsson M, Grigoratos C, Omodarme MD, Chasapides K, Botnar R, Nagel E. Clinical evaluation of three-dimensional late enhancement MRI. J Magn Reson Imaging. 2017;45(6):1675–83.CrossRefPubMed
7.
Zurück zum Zitat Goetti R, Kozerke S, Donati OF, Surder D, Stolzmann P, Kaufmann PA, Luscher TF, Corti R, Manka R. Acute, subacute, and chronic myocardial infarction: quantitative comparison of 2D and 3D late gadolinium enhancement MR imaging. Radiology. 2011;259(3):704–11.CrossRefPubMed Goetti R, Kozerke S, Donati OF, Surder D, Stolzmann P, Kaufmann PA, Luscher TF, Corti R, Manka R. Acute, subacute, and chronic myocardial infarction: quantitative comparison of 2D and 3D late gadolinium enhancement MR imaging. Radiology. 2011;259(3):704–11.CrossRefPubMed
8.
Zurück zum Zitat Roujol S, Basha TA, Akcakaya M, Foppa M, Chan RH, Kissinger KV, Goddu B, Berg S, Manning WJ, Nezafat R. 3D late gadolinium enhancement in a single prolonged breath-hold using supplemental oxygenation and hyperventilation. Magn Reson Med. 2014;72(3):850–7.CrossRefPubMed Roujol S, Basha TA, Akcakaya M, Foppa M, Chan RH, Kissinger KV, Goddu B, Berg S, Manning WJ, Nezafat R. 3D late gadolinium enhancement in a single prolonged breath-hold using supplemental oxygenation and hyperventilation. Magn Reson Med. 2014;72(3):850–7.CrossRefPubMed
9.
Zurück zum Zitat Peters DC, Shaw JL, Knowles BR, Moghari MH, Manning WJ. Respiratory bellows-gated late gadolinium enhancement of the left atrium. J Magn Reson Imaging. 2013;38(5):1210–4.CrossRefPubMed Peters DC, Shaw JL, Knowles BR, Moghari MH, Manning WJ. Respiratory bellows-gated late gadolinium enhancement of the left atrium. J Magn Reson Imaging. 2013;38(5):1210–4.CrossRefPubMed
10.
Zurück zum Zitat Keegan J, Drivas P, Firmin DN. Navigator artifact reduction in three-dimensional late gadolinium enhancement imaging of the atria. Magn Reson Med. 2014;72(3):779–85.CrossRefPubMed Keegan J, Drivas P, Firmin DN. Navigator artifact reduction in three-dimensional late gadolinium enhancement imaging of the atria. Magn Reson Med. 2014;72(3):779–85.CrossRefPubMed
11.
Zurück zum Zitat Kino A, Zuehlsdorff S, Sheehan JJ, Weale PJ, Carroll TJ, Jerecic R, Carr JC. Three-dimensional phase-sensitive inversion-recovery turbo FLASH sequence for the evaluation of left ventricular myocardial scar. AJR Am J Roentgenol. 2009;193(5):W381–8.CrossRefPubMed Kino A, Zuehlsdorff S, Sheehan JJ, Weale PJ, Carroll TJ, Jerecic R, Carr JC. Three-dimensional phase-sensitive inversion-recovery turbo FLASH sequence for the evaluation of left ventricular myocardial scar. AJR Am J Roentgenol. 2009;193(5):W381–8.CrossRefPubMed
12.
Zurück zum Zitat Nguyen TD, Spincemaille P, Weinsaft JW, Ho BY, Cham MD, Prince MR, Wang Y. A fast navigator-gated 3D sequence for delayed enhancement MRI of the myocardium: comparison with breathhold 2D imaging. J Magn Reson Imaging. 2008;27(4):802–8.CrossRefPubMed Nguyen TD, Spincemaille P, Weinsaft JW, Ho BY, Cham MD, Prince MR, Wang Y. A fast navigator-gated 3D sequence for delayed enhancement MRI of the myocardium: comparison with breathhold 2D imaging. J Magn Reson Imaging. 2008;27(4):802–8.CrossRefPubMed
13.
Zurück zum Zitat Peters DC, Appelbaum EA, Nezafat R, Dokhan B, Han Y, Kissinger KV, Goddu B, Manning WJ. Left ventricular infarct size, peri-infarct zone, and papillary scar measurements: a comparison of high-resolution 3D and conventional 2D late gadolinium enhancement cardiac MR. J Magn Reson Imaging. 2009;30(4):794–800.CrossRefPubMedPubMedCentral Peters DC, Appelbaum EA, Nezafat R, Dokhan B, Han Y, Kissinger KV, Goddu B, Manning WJ. Left ventricular infarct size, peri-infarct zone, and papillary scar measurements: a comparison of high-resolution 3D and conventional 2D late gadolinium enhancement cardiac MR. J Magn Reson Imaging. 2009;30(4):794–800.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat van den Bosch HC, Westenberg JJ, Post JC, Yo G, Verwoerd J, Kroft LJ, de Roos A. Free-breathing MRI for the assessment of myocardial infarction: clinical validation. AJR Am J Roentgenol. 2009;192(6):W277–81.CrossRefPubMed van den Bosch HC, Westenberg JJ, Post JC, Yo G, Verwoerd J, Kroft LJ, de Roos A. Free-breathing MRI for the assessment of myocardial infarction: clinical validation. AJR Am J Roentgenol. 2009;192(6):W277–81.CrossRefPubMed
15.
Zurück zum Zitat Stehning C, Bornert P, Nehrke K, Eggers H, Stuber M. Free-breathing whole-heart coronary MRA with 3D radial SSFP and self-navigated image reconstruction. Magn Reson Med. 2005;54(2):476–80.CrossRefPubMed Stehning C, Bornert P, Nehrke K, Eggers H, Stuber M. Free-breathing whole-heart coronary MRA with 3D radial SSFP and self-navigated image reconstruction. Magn Reson Med. 2005;54(2):476–80.CrossRefPubMed
16.
Zurück zum Zitat Sorensen TS, Korperich H, Greil GF, Eichhorn J, Barth P, Meyer H, Pedersen EM, Beerbaum P. Operator-independent isotropic three-dimensional magnetic resonance imaging for morphology in congenital heart disease: a validation study. Circulation. 2004;110(2):163–9.CrossRefPubMed Sorensen TS, Korperich H, Greil GF, Eichhorn J, Barth P, Meyer H, Pedersen EM, Beerbaum P. Operator-independent isotropic three-dimensional magnetic resonance imaging for morphology in congenital heart disease: a validation study. Circulation. 2004;110(2):163–9.CrossRefPubMed
17.
Zurück zum Zitat Monney P, Piccini D, Rutz T, Vincenti G, Coppo S, Koestner SC, Sekarski N, Di Bernardo S, Bouchardy J, Stuber M, Schwitter J. Single centre experience of the application of self navigated 3D whole heart cardiovascular magnetic resonance for the assessment of cardiac anatomy in congenital heart disease. J Cardiovasc Magn Reson. 2015;17:55.CrossRefPubMedPubMedCentral Monney P, Piccini D, Rutz T, Vincenti G, Coppo S, Koestner SC, Sekarski N, Di Bernardo S, Bouchardy J, Stuber M, Schwitter J. Single centre experience of the application of self navigated 3D whole heart cardiovascular magnetic resonance for the assessment of cardiac anatomy in congenital heart disease. J Cardiovasc Magn Reson. 2015;17:55.CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Henningsson M, Smink J, van Ensbergen G, Botnar R. Coronary MR angiography using image-based respiratory motion compensation with inline correction and fixed gating efficiency. Magn Reson Med. 2017. doi:10.1002/mrm.26678 Henningsson M, Smink J, van Ensbergen G, Botnar R. Coronary MR angiography using image-based respiratory motion compensation with inline correction and fixed gating efficiency. Magn Reson Med. 2017. doi:10.​1002/​mrm.​26678
19.
Zurück zum Zitat Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole-heart coronary MR angiography with 2D self-navigated image reconstruction. Magn Reson Med. 2012;67(2):437–45.CrossRefPubMed Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole-heart coronary MR angiography with 2D self-navigated image reconstruction. Magn Reson Med. 2012;67(2):437–45.CrossRefPubMed
20.
Zurück zum Zitat Henningsson M, Hussain T, Vieira MS, Greil GF, Smink J, Ensbergen GV, Beck G, Botnar RM. Whole-heart coronary MR angiography using image-based navigation for the detection of coronary anomalies in adult patients with congenital heart disease. J Magn Reson Imaging. 2016;43(4):947–55.CrossRefPubMed Henningsson M, Hussain T, Vieira MS, Greil GF, Smink J, Ensbergen GV, Beck G, Botnar RM. Whole-heart coronary MR angiography using image-based navigation for the detection of coronary anomalies in adult patients with congenital heart disease. J Magn Reson Imaging. 2016;43(4):947–55.CrossRefPubMed
21.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the cardiac imaging Committee of the Council on clinical cardiology of the American Heart Association. Circulation. 2002;105(4):539–42.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the cardiac imaging Committee of the Council on clinical cardiology of the American Heart Association. Circulation. 2002;105(4):539–42.CrossRefPubMed
22.
Zurück zum Zitat Kaufman L, Kramer DM, Crooks LE, Ortendahl DA. Measuring signal-to-noise ratios in MR imaging. Radiology. 1989;173(1):265–7.CrossRefPubMed Kaufman L, Kramer DM, Crooks LE, Ortendahl DA. Measuring signal-to-noise ratios in MR imaging. Radiology. 1989;173(1):265–7.CrossRefPubMed
23.
Zurück zum Zitat Yu J, Agarwal H, Stuber M, Schar M. Practical signal-to-noise ratio quantification for sensitivity encoding: application to coronary MR angiography. J Magn Reson Imaging. 2011;33(6):1330–40.CrossRefPubMedPubMedCentral Yu J, Agarwal H, Stuber M, Schar M. Practical signal-to-noise ratio quantification for sensitivity encoding: application to coronary MR angiography. J Magn Reson Imaging. 2011;33(6):1330–40.CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Selvanayagam JB, Kardos A, Francis JM, Wiesmann F, Petersen SE, Taggart DP, Neubauer S. Value of delayed-enhancement cardiovascular magnetic resonance imaging in predicting myocardial viability after surgical revascularization. Circulation. 2004;110(12):1535–41.CrossRefPubMed Selvanayagam JB, Kardos A, Francis JM, Wiesmann F, Petersen SE, Taggart DP, Neubauer S. Value of delayed-enhancement cardiovascular magnetic resonance imaging in predicting myocardial viability after surgical revascularization. Circulation. 2004;110(12):1535–41.CrossRefPubMed
25.
Zurück zum Zitat Henningsson M, Prieto C, Chiribiri A, Vaillant G, Razavi R, Botnar RM. Whole-heart coronary MRA with 3D affine motion correction using 3D image-based navigation. Magn Reson Med. 2014;71(1):173–81.CrossRefPubMed Henningsson M, Prieto C, Chiribiri A, Vaillant G, Razavi R, Botnar RM. Whole-heart coronary MRA with 3D affine motion correction using 3D image-based navigation. Magn Reson Med. 2014;71(1):173–81.CrossRefPubMed
26.
Zurück zum Zitat Pang J, Sharif B, Arsanjani R, Bi X, Fan Z, Yang Q, Li K, Berman DS, Li D. Accelerated whole-heart coronary MRA using motion-corrected sensitivity encoding with three-dimensional projection reconstruction. Magn Reson Med. 2015;73(1):284–91.CrossRefPubMed Pang J, Sharif B, Arsanjani R, Bi X, Fan Z, Yang Q, Li K, Berman DS, Li D. Accelerated whole-heart coronary MRA using motion-corrected sensitivity encoding with three-dimensional projection reconstruction. Magn Reson Med. 2015;73(1):284–91.CrossRefPubMed
27.
Zurück zum Zitat Aitken AP, Henningsson M, Botnar RM, Schaeffter T, Prieto C. 100% efficient three-dimensional coronary MR angiography with two-dimensional beat-to-beat translational and bin-to-bin affine motion correction. Magn Reson Med. 2015;74(3):756–64.CrossRefPubMed Aitken AP, Henningsson M, Botnar RM, Schaeffter T, Prieto C. 100% efficient three-dimensional coronary MR angiography with two-dimensional beat-to-beat translational and bin-to-bin affine motion correction. Magn Reson Med. 2015;74(3):756–64.CrossRefPubMed
28.
Zurück zum Zitat Cruz G, Atkinson D, Henningsson M, Botnar RM, Prieto C. Highly efficient nonrigid motion-corrected 3D whole-heart coronary vessel wall imaging. Magn Reson Med. 2017;77(5):1894–908.CrossRefPubMed Cruz G, Atkinson D, Henningsson M, Botnar RM, Prieto C. Highly efficient nonrigid motion-corrected 3D whole-heart coronary vessel wall imaging. Magn Reson Med. 2017;77(5):1894–908.CrossRefPubMed
29.
Zurück zum Zitat Luo J, Addy NO, Ingle RR, Baron CA, Cheng JY, BS H, Nishimura DG. Nonrigid motion correction with 3D image-based navigators for coronary MR angiography. Magn Reson Med. 2017;77(5):1884–93.CrossRefPubMed Luo J, Addy NO, Ingle RR, Baron CA, Cheng JY, BS H, Nishimura DG. Nonrigid motion correction with 3D image-based navigators for coronary MR angiography. Magn Reson Med. 2017;77(5):1884–93.CrossRefPubMed
30.
Zurück zum Zitat Keegan J, Gatehouse PD, Haldar S, Wage R, Babu-Narayan SV, Firmin DN. Dynamic inversion time for improved 3D late gadolinium enhancement imaging in patients with atrial fibrillation. Magn Reson Med. 2015;73(2):646–54.CrossRefPubMed Keegan J, Gatehouse PD, Haldar S, Wage R, Babu-Narayan SV, Firmin DN. Dynamic inversion time for improved 3D late gadolinium enhancement imaging in patients with atrial fibrillation. Magn Reson Med. 2015;73(2):646–54.CrossRefPubMed
31.
Zurück zum Zitat Akcakaya M, Rayatzadeh H, Basha TA, Hong SN, Chan RH, Kissinger KV, Hauser TH, Josephson ME, Manning WJ, Nezafat R. Accelerated late gadolinium enhancement cardiac MR imaging with isotropic spatial resolution using compressed sensing: initial experience. Radiology. 2012;264(3):691–9.CrossRefPubMedPubMedCentral Akcakaya M, Rayatzadeh H, Basha TA, Hong SN, Chan RH, Kissinger KV, Hauser TH, Josephson ME, Manning WJ, Nezafat R. Accelerated late gadolinium enhancement cardiac MR imaging with isotropic spatial resolution using compressed sensing: initial experience. Radiology. 2012;264(3):691–9.CrossRefPubMedPubMedCentral
32.
Zurück zum Zitat Peters DC, Wylie JV, Hauser TH, Kissinger KV, Botnar RM, Essebag V, Josephson ME, Manning WJ. Detection of pulmonary vein and left atrial scar after catheter ablation with three-dimensional navigator-gated delayed enhancement MR imaging: initial experience. Radiology. 2007;243(3):690–5.CrossRefPubMed Peters DC, Wylie JV, Hauser TH, Kissinger KV, Botnar RM, Essebag V, Josephson ME, Manning WJ. Detection of pulmonary vein and left atrial scar after catheter ablation with three-dimensional navigator-gated delayed enhancement MR imaging: initial experience. Radiology. 2007;243(3):690–5.CrossRefPubMed
Metadaten
Titel
‘Image-navigated 3-dimensional late gadolinium enhancement cardiovascular magnetic resonance imaging: feasibility and initial clinical results’
verfasst von
Konstantinos Bratis
Markus Henningsson
Chrysanthos Grigoratos
Matteo Dell’Omodarme
Konstantinos Chasapides
Rene Botnar
Eike Nagel
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2017
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-017-0418-7

Weitere Artikel der Ausgabe 1/2017

Journal of Cardiovascular Magnetic Resonance 1/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.