Insertion characteristics and placement of the Mid-Scala electrode array in human temporal bones using detailed cone beam computed tomography

verfasst von: Aarno Dietz, Dzemal Gazibegovic, Jyrki Tervaniemi, Veli-Matti Vartiainen, Heikki Löppönen

Erschienen in: European Archives of Oto-Rhino-Laryngology | Ausgabe 12/2016

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to evaluate the insertion results and placement of the new Advanced Bionics HiFocus Mid-Scala (HFms) electrode array, inserted through the round window membrane, in eight fresh human temporal bones using cone beam computed tomography (CBCT). Pre- and post-insertion CBCT scans were registered to create a 3D reconstruction of the cochlea with the array inserted. With an image fusion technique both the bony edges of the cochlea and the electrode array in situ could accurately be determined, thus enabling to identify the exact position of the electrode array within the scala tympani. Vertical and horizontal scalar location was measured at four points along the cochlea base at an angular insertion depth of 90°, 180° and 270° and at electrode 16, the most basal electrode. Smooth insertion through the round window membrane was possible in all temporal bones. The imaging results showed that there were no dislocations from the scala tympani into the scala vestibule. The HFms electrode was positioned in the middle of the scala along the whole electrode array in three out of the eight bones and in 62 % of the individual locations measured along the base of the cochlea. In only one cochlea a close proximity of the electrode with the basilar membrane was observed, indicating possible contact with the basilar membrane. The results and assessments presented in this study appear to be highly accurate. Although a further validation including histopathology is needed, the image fusion technique described in this study represents currently the most accurate method for intracochlear electrode assessment obtainable with CBCT.
Literatur
1.
Zurück zum Zitat Aschendorff A, Kromeier J, Klenzner T, Laszig R (2007) Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear 28(2 Suppl):75S–79SCrossRefPubMed Aschendorff A, Kromeier J, Klenzner T, Laszig R (2007) Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear 28(2 Suppl):75S–79SCrossRefPubMed
2.
Zurück zum Zitat Finley CC, Holden TA, Holden LK et al (2008) Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol 29:920–928CrossRefPubMedPubMedCentral Finley CC, Holden TA, Holden LK et al (2008) Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol 29:920–928CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Carlson ML, Driscoll CL, Gifford RH, Servic GJ, Tombers NM, Hughes-Borst BJ, Neff BA, Beatty CW (2011) Implications of minimizing trauma during conventional cochlear implantation. Otol Neurotol 32(6):962–968CrossRefPubMedPubMedCentral Carlson ML, Driscoll CL, Gifford RH, Servic GJ, Tombers NM, Hughes-Borst BJ, Neff BA, Beatty CW (2011) Implications of minimizing trauma during conventional cochlear implantation. Otol Neurotol 32(6):962–968CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Ramsden JD, Gordon K, Aschendorff A, Borucki L, Bunne M, Burdo S, Garabedian N, Grolman W, Irving R, Lesinski-Schiedat A, Loundon N, Manrique M, Martin J, Raine C, Wouters J, Papsin BC (2012) European bilateral pediatric cochlear implant forum consensus statement. Otol Neurotol 33(4):561–565CrossRefPubMed Ramsden JD, Gordon K, Aschendorff A, Borucki L, Bunne M, Burdo S, Garabedian N, Grolman W, Irving R, Lesinski-Schiedat A, Loundon N, Manrique M, Martin J, Raine C, Wouters J, Papsin BC (2012) European bilateral pediatric cochlear implant forum consensus statement. Otol Neurotol 33(4):561–565CrossRefPubMed
5.
Zurück zum Zitat Briggs RJ, Tykocinski M, Saunders E, Hellier W, Dahm M, Pyman B, Clark GM (2001) Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertion. Cochlear Implants Int 2(2):135–149CrossRefPubMed Briggs RJ, Tykocinski M, Saunders E, Hellier W, Dahm M, Pyman B, Clark GM (2001) Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertion. Cochlear Implants Int 2(2):135–149CrossRefPubMed
6.
Zurück zum Zitat Adunka O, Kiefer J, Unkelbach MH, Lehnert T, Gstoettner W (2004) Development and evaluation of an improved cochlear implant electrode design for electric acoustic stimulation. Laryngoscope 114(7):1237–1241CrossRefPubMed Adunka O, Kiefer J, Unkelbach MH, Lehnert T, Gstoettner W (2004) Development and evaluation of an improved cochlear implant electrode design for electric acoustic stimulation. Laryngoscope 114(7):1237–1241CrossRefPubMed
7.
Zurück zum Zitat Skarzynski H, Lorens A, Matusiak M, Porowski M, Skarzynski PH, James CJ (2012) Partial deafness treatment with the nucleus straight research array cochlear implant. Audiol Neurootol 17(2):82–91CrossRefPubMed Skarzynski H, Lorens A, Matusiak M, Porowski M, Skarzynski PH, James CJ (2012) Partial deafness treatment with the nucleus straight research array cochlear implant. Audiol Neurootol 17(2):82–91CrossRefPubMed
8.
Zurück zum Zitat Gstoettner WK, Adunka O, Franz P, Hamzavi J Jr, Plenk H Jr, Susani M, Baumgartner W, Kiefer J (2001) Perimodiolar electrodes in cochlear implant surgery. Acta Otolaryngol 121:216–219CrossRefPubMed Gstoettner WK, Adunka O, Franz P, Hamzavi J Jr, Plenk H Jr, Susani M, Baumgartner W, Kiefer J (2001) Perimodiolar electrodes in cochlear implant surgery. Acta Otolaryngol 121:216–219CrossRefPubMed
9.
Zurück zum Zitat Gantz BJ, Dunn C, Oleson J, Hansen M, Parkinson A, Turner C (2016) Multicenter clinical trial of the nucleus hybrid S8 cochlear implant: final outcomes. Laryngoscope. (Epub ahead of print) Gantz BJ, Dunn C, Oleson J, Hansen M, Parkinson A, Turner C (2016) Multicenter clinical trial of the nucleus hybrid S8 cochlear implant: final outcomes. Laryngoscope. (Epub ahead of print)
10.
Zurück zum Zitat Helbig S, Settevendemie C, Mack M, Baumann U, Helbig M, Stöver T (2011) Evaluation of an electrode prototype for atraumatic cochlear implantation in hearing preservation candidates: preliminary results from a temporal bone study. Otol Neurotol 32(3):419–423CrossRefPubMed Helbig S, Settevendemie C, Mack M, Baumann U, Helbig M, Stöver T (2011) Evaluation of an electrode prototype for atraumatic cochlear implantation in hearing preservation candidates: preliminary results from a temporal bone study. Otol Neurotol 32(3):419–423CrossRefPubMed
11.
Zurück zum Zitat Hassepass F, Bulla S, Maier W, Laszig R, Arndt S, Beck R, Traser L, Aschendorff A (2014) The new Mid-Scala electrode array: a radiologic and histologic study in human temporal bones. Otol Neurotol 35(8):1415–1420CrossRefPubMed Hassepass F, Bulla S, Maier W, Laszig R, Arndt S, Beck R, Traser L, Aschendorff A (2014) The new Mid-Scala electrode array: a radiologic and histologic study in human temporal bones. Otol Neurotol 35(8):1415–1420CrossRefPubMed
12.
Zurück zum Zitat Frisch CD, Carlson ML, Lane JI, Driscoll CL (2015) Evaluation of a new Mid-Scala cochlear implant electrode using microcomputed tomography. Laryngoscope 125(12):2778–2783CrossRefPubMed Frisch CD, Carlson ML, Lane JI, Driscoll CL (2015) Evaluation of a new Mid-Scala cochlear implant electrode using microcomputed tomography. Laryngoscope 125(12):2778–2783CrossRefPubMed
13.
Zurück zum Zitat Dietz A, Wennström M, Lehtimäki A, Löppönen H, Valtonen H (2015) Electrode migration after cochlear implant surgery: more common than expected? Eur Arch Otorhinolaryngol. (Epub ahead of print) Dietz A, Wennström M, Lehtimäki A, Löppönen H, Valtonen H (2015) Electrode migration after cochlear implant surgery: more common than expected? Eur Arch Otorhinolaryngol. (Epub ahead of print)
14.
Zurück zum Zitat Xu J, Xu SA, Cohen LT, Clark GM (2000) Cochlear view: postoperative radiography for cochlear implantation. Am J Otol 21(1):49–56CrossRefPubMed Xu J, Xu SA, Cohen LT, Clark GM (2000) Cochlear view: postoperative radiography for cochlear implantation. Am J Otol 21(1):49–56CrossRefPubMed
15.
Zurück zum Zitat Cushing SL, Daly MJ, Treaba CG, Chan H, Irish JC, Blaser S, Gordon KA, Papsin BC (2012) High-resolution cone-beam computed tomography: a potential tool to improve atraumatic electrode design and position. Acta Otolaryngol 132(4):361–368CrossRefPubMed Cushing SL, Daly MJ, Treaba CG, Chan H, Irish JC, Blaser S, Gordon KA, Papsin BC (2012) High-resolution cone-beam computed tomography: a potential tool to improve atraumatic electrode design and position. Acta Otolaryngol 132(4):361–368CrossRefPubMed
16.
Zurück zum Zitat Saeed SR, Selvadurai D, Beale T, Biggs N, Murray B, Gibson P, Risi F, Boyd P (2014) The use of cone-beam computed tomography to determine cochlear implant electrode position in human temporal bones. Otol Neurotol 35(8):1338–1344CrossRefPubMed Saeed SR, Selvadurai D, Beale T, Biggs N, Murray B, Gibson P, Risi F, Boyd P (2014) The use of cone-beam computed tomography to determine cochlear implant electrode position in human temporal bones. Otol Neurotol 35(8):1338–1344CrossRefPubMed
17.
Zurück zum Zitat Marx M, Risi F, Escudé B, Durmo I, James C, Lauwers F, Deguine O, Fraysse B (2014) Reliability of cone beam computed tomography in scalar localization of the electrode array: a radio histological study. Eur Arch Otorhinolaryngol 271(4):673–679CrossRefPubMed Marx M, Risi F, Escudé B, Durmo I, James C, Lauwers F, Deguine O, Fraysse B (2014) Reliability of cone beam computed tomography in scalar localization of the electrode array: a radio histological study. Eur Arch Otorhinolaryngol 271(4):673–679CrossRefPubMed
18.
Zurück zum Zitat Güldner C, Wiegand S, Weiss R, Bien S, Sesterhenn A, Teymoortash A, Diogo I (2012) Artifacts of the electrode in cochlea implantation and limits in analysis of deep insertion in cone beam tomography (CBT). Eur Arch Otorhinolaryngol 269(3):767–772CrossRefPubMed Güldner C, Wiegand S, Weiss R, Bien S, Sesterhenn A, Teymoortash A, Diogo I (2012) Artifacts of the electrode in cochlea implantation and limits in analysis of deep insertion in cone beam tomography (CBT). Eur Arch Otorhinolaryngol 269(3):767–772CrossRefPubMed
19.
Zurück zum Zitat Ruivo J, Mermuys K, Bacher K, Kuhweide R, Offeciers E, Casselman JW (2009) Cone beam computed tomography, a low-dose imaging technique in the postoperative assessment of cochlear implantation. Otol Neurotol 30(3):299–303CrossRefPubMed Ruivo J, Mermuys K, Bacher K, Kuhweide R, Offeciers E, Casselman JW (2009) Cone beam computed tomography, a low-dose imaging technique in the postoperative assessment of cochlear implantation. Otol Neurotol 30(3):299–303CrossRefPubMed
20.
Zurück zum Zitat Theunisse HJ, Joemai RM, Maal TJ, Geleijns J, Mylanus EA, Verbist BM (2015) Cone-beam CT versus multi-slice CT systems for postoperative imaging of cochlear implantation—a phantom study on image quality and radiation exposure using human temporal bones. Otol Neurotol 36(4):592–599CrossRefPubMed Theunisse HJ, Joemai RM, Maal TJ, Geleijns J, Mylanus EA, Verbist BM (2015) Cone-beam CT versus multi-slice CT systems for postoperative imaging of cochlear implantation—a phantom study on image quality and radiation exposure using human temporal bones. Otol Neurotol 36(4):592–599CrossRefPubMed
21.
Zurück zum Zitat Majdani O et al (2009) Temporal bone imaging: comparison of flat panel volume CT and multisection CT. Am J Neuroradiol 30:1419–1424CrossRefPubMed Majdani O et al (2009) Temporal bone imaging: comparison of flat panel volume CT and multisection CT. Am J Neuroradiol 30:1419–1424CrossRefPubMed
22.
Zurück zum Zitat Husstedt HW, Aschendorff A, Richter B, Laszig R, Schumacher M (2002) Nondestructive three-dimensional analysis of electrode to modiolus proximity. Otol Neurotol 23(1):49–52CrossRefPubMed Husstedt HW, Aschendorff A, Richter B, Laszig R, Schumacher M (2002) Nondestructive three-dimensional analysis of electrode to modiolus proximity. Otol Neurotol 23(1):49–52CrossRefPubMed
23.
Zurück zum Zitat Stutzki M, Jahns E, Mandapathil MM, Diogo I, Werner JA, Güldner C (2015) Indications of cone beam CT in head and neck imaging. Acta Otolaryngol 135(12):1337–1343CrossRefPubMed Stutzki M, Jahns E, Mandapathil MM, Diogo I, Werner JA, Güldner C (2015) Indications of cone beam CT in head and neck imaging. Acta Otolaryngol 135(12):1337–1343CrossRefPubMed
25.
Zurück zum Zitat Pieper S, Lorensen B, Schroeder W, Kikinis R (2006) The NA-MIC kit: ITK, VTK, pipelines, grids and 3D slicer as an open platform for the medical image computing community. In: 3rd IEEE international symposium on biomedical imaging: macro to nano, pp 698–701 Pieper S, Lorensen B, Schroeder W, Kikinis R (2006) The NA-MIC kit: ITK, VTK, pipelines, grids and 3D slicer as an open platform for the medical image computing community. In: 3rd IEEE international symposium on biomedical imaging: macro to nano, pp 698–701
26.
Zurück zum Zitat Pieper S, Halle M, Kikinis R (2004) 3D slicer. In: 2nd IEEE international symposium on biomedical imaging: macro to nano (IEEE Cat No. 04EX821), vol 2, pp 632–635 Pieper S, Halle M, Kikinis R (2004) 3D slicer. In: 2nd IEEE international symposium on biomedical imaging: macro to nano (IEEE Cat No. 04EX821), vol 2, pp 632–635
27.
Zurück zum Zitat Johnson H, Harris G (2007) BRAINSFit: mutual information registrations of whole-brain 3D images, using the insight toolkit. Insight J Johnson H, Harris G (2007) BRAINSFit: mutual information registrations of whole-brain 3D images, using the insight toolkit. Insight J
28.
Zurück zum Zitat Kurzweg T, Dalchow CV, Bremke M, Majdani O, Kureck I, Knecht R, Werner JA, Teymoortash A (2010) The value of digital volume tomography in assessing the position of cochlear implant arrays in temporal bone specimens. Ear Hear 31(3):413–419CrossRefPubMed Kurzweg T, Dalchow CV, Bremke M, Majdani O, Kureck I, Knecht R, Werner JA, Teymoortash A (2010) The value of digital volume tomography in assessing the position of cochlear implant arrays in temporal bone specimens. Ear Hear 31(3):413–419CrossRefPubMed
29.
Zurück zum Zitat Postnov A, Zarowski A, De Clerck N, Vanpoucke F, Offeciers FE, Van Dyck D, Peeters S (2006) High resolution micro-CT scanning as an innovative tool for evaluation of the surgical positioning of cochlear implant electrodes. Acta Otolaryngol 126(5):467–474CrossRefPubMed Postnov A, Zarowski A, De Clerck N, Vanpoucke F, Offeciers FE, Van Dyck D, Peeters S (2006) High resolution micro-CT scanning as an innovative tool for evaluation of the surgical positioning of cochlear implant electrodes. Acta Otolaryngol 126(5):467–474CrossRefPubMed
30.
Zurück zum Zitat Faccioli N, Barillari M, Guariglia S, Zivelonghi E, Rizzotti A, Cerini R, Mucelli RP (2009) Radiation dose saving through the use of cone-beam CT in hearing-impaired patients. Radiol Med 114(8):1308–1318CrossRefPubMed Faccioli N, Barillari M, Guariglia S, Zivelonghi E, Rizzotti A, Cerini R, Mucelli RP (2009) Radiation dose saving through the use of cone-beam CT in hearing-impaired patients. Radiol Med 114(8):1308–1318CrossRefPubMed
Metadaten
Titel
Insertion characteristics and placement of the Mid-Scala electrode array in human temporal bones using detailed cone beam computed tomography
verfasst von
Aarno Dietz
Dzemal Gazibegovic
Jyrki Tervaniemi
Veli-Matti Vartiainen
Heikki Löppönen
Publikationsdatum
19.05.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Archives of Oto-Rhino-Laryngology / Ausgabe 12/2016
Print ISSN: 0937-4477
Elektronische ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-016-4099-x

Weitere Artikel der Ausgabe 12/2016

European Archives of Oto-Rhino-Laryngology 12/2016 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

HNO-Op. auch mit über 90?

16.04.2024 HNO-Chirurgie Nachrichten

Mit Blick auf das Risiko für Komplikationen nach elektiven Eingriffen im HNO-Bereich scheint das Alter der Patienten kein ausschlaggebender Faktor zu sein. Entscheidend ist offenbar, wie fit die Betroffenen tatsächlich sind.

Intrakapsuläre Tonsillektomie gewinnt an Boden

16.04.2024 Tonsillektomie Nachrichten

Gegenüber der vollständigen Entfernung der Gaumenmandeln hat die intrakapsuläre Tonsillektomie einige Vorteile, wie HNO-Fachleute aus den USA hervorheben. Sie haben die aktuelle Literatur zu dem Verfahren gesichtet.

Bilateraler Hörsturz hat eine schlechte Prognose

15.04.2024 Hörsturz Nachrichten

Die Mehrzahl der Menschen mit Hörsturz ist einseitig betroffen, doch auch ein beidseitiger Hörsturz ist möglich. Wie häufig solche Fälle sind und wie sich ihr Verlauf darstellt, hat eine HNO-Expertenrunde aus den USA untersucht.

Update HNO

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.