Is it safe to inject contrast through the side arm of an introducer sheath? An in vitro study
verfasst von:
Garvit D. Khatri, Jeffrey D. Robinson
Erschienen in:
Emergency Radiology
|
Ausgabe 3/2020
Einloggen, um Zugang zu erhalten
Abstract
Purpose
Occasionally, a patient requiring computed tomography (CT) with poor venous access or in an unstable condition may have a sheath as their only form of IV access. Many institutions prohibit contrast power injection through the side ports, for concern for rupture. We hypothesize that under conditions encountered in CT scanning, the side arm of the sheath introducer is safe for power injection of contrast material.
Methods
In this in vitro study, we injected contrast at different flow rates varying from 2 to 8 ml/s, through the side port of the sheath. Time-pressure graphs were obtained for each injection from the injector display. The assembly was observed for any signs of leakage or material failure.
Results
There was not a single event of leak, rupture, or displacement. Maximum pressures were within the range of conventional contrast administrations.
Conclusion
This study suggests that iodinated contrast can safely be injected through the side port of a venous sheath.