To assess the feasibility and safety of a robotic system for percutaneous needle insertion during thermal ablation of liver tumors.
Materials and Methods
This study analyzed the CT-guided percutaneous needle insertion using the EPIONE robotic device (Quantum Surgical, Montpellier, France) for radiofrequency or microwave liver ablation. The main criteria of the study were feasibility (possibility to perform the thermal ablation after needle insertion), the number of needle adjustments (reiteration of robotically assisted needle insertion when initial needle positioning is considered insufficient to perform ablation), and robotic-guided procedure safety (complications related to the needle insertion). Patients were followed up at 6 months post-intervention to assess local tumor control.
Results
Twenty-one patients with 24 tumors, including 6 HCC and 18 metastases measuring 15.6 ± 7.2 mm, were enrolled. One patient (with one tumor) was excluded for protocol deviation. Robotic assisted thermal ablation was feasible for 22/23 lesions (95.7%) and 19/20 patients (95.0%), as validated by a data safety monitoring Board (95% CI [76.39%; 99.11%]) for the per-protocol population. The mean number of needle adjustments per tumor treated was 0.4 (SD: 0.7), with 70.8% of tumors requiring no adjustment. No adverse events were depicted. Rate of local tumor control was 83.3% for patients and 85.7% for tumors, at 6 months.
Conclusion
This bicentric first-in-human pilot study suggests both feasibility and safety of a stereotactic CT-guided EPIONE device for the percutaneous needle insertion during liver tumor thermal ablation.
Hinweise
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Introduction
Percutaneous thermal ablation is one of the treatments of choice for liver tumors less than 3 cm in size [1, 2]. Radiofrequency ablation (RFA) and microwave ablation (MWA) are the most frequently used modalities and are well documented in the current literature [3, 4]. They have shown favorable outcomes for tumor control, with low associated morbidity, and low local tumor progression (LTP) rate [5]. CT-guided freehand insertion of percutaneous needles sometimes requires numerous CT acquisitions for needle progression control, moreover tumor localization itself and/or poor tumor visibility can lead to a challenging procedure with reduced odds of ablation success.
The development of robotic needle guidance in interventional radiology can help to improve needle insertion, reducing the need for needle adjustments. It also allows the physician to set and visualize the planned ablation margin, to obtain more precise ablation zones with less rates of tumor persistence or recurrence. Lastly, robotic solutions generally offer better positioning results compared with freehand needle insertion [6‐10].
Anzeige
We herein report our first in human trial with a CT-guided robotic which is designed to assist interventional radiologists in performing percutaneous procedures in the abdomen. The objective of our work was to assess feasibility and safety for needle insertion.
Materials & Methods
Study Design and Patient Selection
This pilot, prospective and bicentric study was approved by the Ethics Committee (Comité de Protection des Personnes, Ile de France V, France) and was registered on ClinicalTrials.gov (NCT04230642).
The inclusion criteria were percutaneous ablation under CT guidance validated by a multidisciplinary tumor board targeting one or two liver tumors in adult patients who have the ability to understand and sign the informed consent.
Exclusion criteria were pregnancy, breast-feeding women, patients unable to undergo general anesthesia, patients unable to tolerate CT contrast agent, or participating to another clinical study.
Procedures were performed by interventional radiologists with 5 to 25 years of practice in freehand procedures, trained to use the robotic system by carrying out at least one intervention workflow on a phantom model. All interventions were performed under general anesthesia. The choice of the thermal ablation modality was left to the operator’s preference (Fig. 1).
×
Robotic Device
The EPIONE robotic system (Quantum Surgical, Montpellier, France) (Fig. 2) was already detailed in previous animal studies [11, 12]. It is composed of a mobile display cart (workstation), a mobile navigation cart (infra-red camera), a mobile robot cart (robot arm) on which is attached a needle guide, and a patient reference attached to the patient’s skin.
×
The planning and navigation software of the device enables to review images, plan a needle trajectory from skin entrance to target, monitor respiratory motion, and automatically position the needle guide according to the trajectory.
Using a 3D motion analysis of the patient reference, the respiratory monitoring module enables real-time tracking and display of the patient’s respiratory cycle (Fig. 3), allowing the operator to visually check the breathing phase and repeatability of apnea to record a respiratory reference level.
×
The registration between the patient position on the table, the CT images and the robotic device are automatically performed by the system using the patient reference.
Robotic Needle Insertion Planning
A baseline CT was acquired under apnea using an Alphenix or Infinix-i CT scanner (Canon Medical Systems, Otawara, Japan) and then loaded in the software of the robotic system. The type and length of the ablation needle were entered into the planning software. The interventional radiologist defined the target point so that the expected ablation zone covers the tumor and the entry point at the needle puncture site on skin surface. The needle trajectory path was then calculated by the device software and shown on the display screen (Fig. 4, image A). The needle trajectory was carefully confirmed by the operator being sure to avoid critical or bone structures and sent to the robotic arm for execution.
×
Robotic-Guided Needle Insertion
The robotic arm was manually pre-positioned close to the patient’s abdomen with a footswitch command control. Under apnea, the operator inserted the needle through the needle guide until the end stop was reached (Fig. 4, image B). The needle was then released from the needle guide and the robot arm was manually withdrawn. A post-insertion CT-scan image was performed to assess the feasibility (possibility to perform the procedure related to the needle positioned by the robot) (Fig. 4, image C).
Post-operative Assessment and Ablation Success
A CT-scan acquisition was performed at the end of the intervention to ensure the absence of complication, and to assess whether the tumor was completely ablated or not. The patients were then reviewed periodically every 3 months using either CT-scan or MRI to assess LTP, as per standard practice guidelines [13]. The follow-up performed after a minimal 6-month period was to collect data to assess the local tumor control rate, as routine follow-up.
Outcome Measures
Ablation Feasibility
Feasibility was defined as a needle placement judged by the operator as adequate to perform ablation. The operator visually estimated on the post-insertion CT-scan if the needle was positioned accurately enough compared to the planned trajectory. To confirm feasibility assessment, the post-insertion CT-scan was ultimately reviewed by the Data and Safety Monitoring Board (DSMB), constituted of 2 experts in interventional radiology and one expert in digestive surgery who worked in different institutions.
Needle Placement Adjustment
If the ablation was not considered feasible following the first needle insertion, the operator may choose to adjust the position by planning a new trajectory. The number of needle placement adjustments (attempts required to reach a position that met the feasibility criteria) was recorded. After a second failed adjustment attempt, the procedure was considered unsuccessful and converted into a freehand insertion.
Safety
Safety was evaluated as the number of major adverse events attributable to the needle insertion are reported using the Society of Interventional Radiology (SIR) scale [14, 15] and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE) classification [16]. The adverse events were monitored and reviewed by the DSMB to confirm the safety evaluation by the operator.
Statistical Analysis
All statistical analyses were performed using SAS® 9.4 software (Cary, NC, USA). Description of relevant study variables was: n (total number of data), number of missing data, mean, standard deviation (SD), median, min and max for quantitative variables and using number of missing data, frequency and percentage of non-missing for qualitative variables. The statistical analysis took into account data clustering (several ablations for a given patient) in case of inferential statistics. Data were analyzed according to both the intention-to-treat (ITT) and per protocol (PP) principles.
Results
Patient Flow and Baseline Data
From May 2020 to January 2021, 22 patients were enrolled with 26 liver tumors. Patients were either treated at the institution A (n = 16) or B (n = 6) (Fig. 1). During the study period, one patient (with two tumors) for which the ablation procedure was not maintained was considered as a screening failure and was therefore excluded from the study. Of the 21 patients with 24 tumors remaining in the ITT population, one patient (with one tumor) was excluded from the PP analysis due to protocol deviation (the CT-scan was performed later than planned in the protocol).
Baseline characteristics of the 21 patients (9 women, 12 men) of ITT analysis are given in Table 1. Patients had a mean body mass index of 23.8 ± 5. kg/m2 and a mean age of 64.8 ± 11.9 years. The liver targeted tumors were BCLC stages A hepatocellular carcinomas in 5 (23.8%) patients and metastases in 16 (76.2%) patients. The mean largest diameter of the tumors was 15.6 ± 7.2 mm (range 5–32 mm). Eleven (45.8%) targeted tumors were judged as challenging by the operators in regard with their location including 9 (81.8%) lesions located in the hepatic dome and two (18.2%) subcapsular lesions. Eighteen (75%) trajectories were not in the axial plane and fifteen (62.5%) trajectories had a double angulation (i.e., craniocaudal and lateral).
Table 1
Baseline characteristics
Parameter
N (%)
Mean (SD)
Min; Max
Demographics
Age of consent (years old)
64.8 (11.9)
35.0; 79.5
Gender
Male
12 (57.1)
Female
9 (42.9)
23.8 (5.8)
15.6; 40.5
BMI (kg/m2)
Targeted lesion
Lesion entity
Hepatocellular carcinoma
5 (23.8)
Liver metastases
16 (76.2)
Number of liver lesion(s) per patient
1
18 (85.7)
2
3 (14.3)
Total number of lesions
24
Lesion Size (diameter in mm)
–
15.6 (7.2)
5.0; 32.0
Lesion location
2 (8.3)
Segment II
1 (4.2)
Segments II VIII
2 (8.3)
Segment III
3 (12.5)
Segments IV V
5 (20.8)
Segments V VI
1 (4.2)
Segment VI
2 (8.3)
Segment VII
4 (16.7)
Segment VIII
4 (16.7)
Challenging lesions
_
No
13 (54.2)
Yes
11 (45.8)
Lesion located in hepatic dome
9 (85.9)
Subcapsular lesion
2 (18.2)
Needle placement
Trajectory
Axial (in plane)
6 (25.0)
Oblique (out of plane)
18 (75.0)
Angulations
Lateral (°)
_
– 33.5; 78.8
Cranio-caudal (°)
_
– 16.9; 61.4
Needle type
Dophi (Surgnova)
18 (75.0)
Octopus (Starmed)
1 (4.2)
Solero (Angiodynamics)
5 (20.8)
Needle length
150mm
9 (37.5)
190mm
5 (20.8)
200mm
10 (41.7)
Needle diameter
15G
23 (95.8)
17G
1 (4.2)
Ablation parameters per lesion
Type of thermal ablation
Radiofrequency
1 (4.2)
Microwave
23 (95.8)
Cumulative ablation time (min)
_
7.5 (5.7)
3.0; 31.0
Cumulative power for microwave ablation (watts)
81.3 (32.2)
50.0; 160.0
Operator
≤10 years of experience
Patients treated
6 (28.6)
Lesions treated
7 (29.2)
>10 years of experience
Patients treated
15 (71.4)
Lesions treated
17 (70.8)
Min; Max, Minimum; Maximum
SD, Standard Deviation
MWA was used in 23 (95.8%) tumors, using a 15G Dophi (Surgnova, HDTech, Lorient, France) needle in 18 (75%) tumors and a 15G Solero (Angiodynamics, New York, USA) needle in 5 (20.8%) tumors. RFA was used in one (4.2%) tumor with a 17G Octopus (Starmed, Gyeonggi-do, South Korea) needle. The needle length was 150mm in 9 (37.5%) lesions, 190mm in 5 (20.8%) and 200mm in 10 (41.7%).
All 24 tumors were treated by 4 operators. Seventeen (70.8%) lesions in 15 (71.4%) patients were treated by 2 senior operators having more than 10 years of experience, the remaining 7 (29.2%) lesions in 6 (28.6%) patients were treated by 2 less experienced operators.
The mean (SD) overall procedure duration from first pre-interventional CT-scan to last post-ablation CT-scan was 73.8 (29.2) min.
Feasibility
Among the total number of 23 ablations included in the PP analysis, the robotic needle placement was judged adequate to perform ablation for 22 lesions, giving a 95.7% feasibility rate (Table 2), confirmed by the DSMB blind review. The unique failure was recorded when the initially planned needle trajectory was not accessible with the robot arm. After planning a new trajectory, insertion of the ablation needle resulted in a sliding of the needle on the liver’s capsule surface, and the procedure was ultimately resumed freehand under CT guidance.
Table 2
Number of patients with no major AE related to the CT-guided ablation procedure AND with the needle positioned accurately enough to allow the next step of the procedure to be performed by patient and by lesion (PP population)
Parameter
N (%)
95 CI
Combined “feasibility” and “no major Adverse Event”
By lesion
23 (100)
65.9%-99.6%
Yes
22 (95.7)
No
1 (4.3)
Concordance of response between operator and DSMB related to feasibility
Both “yes”
22 (100)
Both “No”
1 (100)
Investigator “No” and DSMB “Yes”
0
Investigator “Yes” and DSMB “No”
0
DSMB, Data Safety Monitoring Board
95 CI, Confidence Interval at 95%
Anzeige
Needle Placement Adjustment
No needle placement adjustment was needed in 17 (70.8%) tumors, while 6 (25%) lesions required 1 adjustment and 1 (4.2%) lesion required 3 adjustments. The mean number of adjustments per lesion was 0.4 ± 0.7 (Table 3).
Table 3
Number of needle placement re-adjustments
Parameter
N (%)
Mean (SD)
Median (Q1; Q3)
Number of needle adjustments
By lesion
24 (100)
0.4 (0.7)
0.0 (0.0; 1.0)
0
17 (70.8)
1
6 (25.0)
2
–
3
1 (4.2)
Q1; Q3: Inferior Quartile; Superior Quartile
SD, Standard Deviation
Safety
No procedure-related complications were observed on post-procedural CT-scan, and no adverse events were detected. Twenty patients were discharged from hospital the day after the procedure, one patient after 2 days.
At 6 months of follow-up, two patients died before review, determining an overall survival of 90.0% in the PP population, results are reported in Table 4. The causes of death were not related to the ablation procedure or the robot use, but to cancer progression.
Table 4
Survival rates at 6 months follow-up in the PP population
Parameters at 6 months
Patients
Lesions
Overall survival
18/20 (90.0%)
21/23 (91.3%)
Disease free survival
15/18 (83.3%)
18/21 (85.7%)
Ablation Success
Among the 18 patients with 21 lesions that were followed up to 6 months, a local tumor control was obtained in 15 (83.3%) patients and 21 (85.7%) tumors, three patients have developed LTP and were retreated with MWA.
Anzeige
Discussion
No adverse events were reported in this study in relation with the use of the device, confirming the safety data collected in animal pre-clinical studies [11, 12]. Only few other devices are available for robotic needle placements, and no specific complications were reported in prospective clinical investigations involving these devices [3, 7, 17‐19]. The short hospital stay with quick recovery of our patients is similar to results reported in the literature on navigational devices for liver ablations [10, 20‐23].
In our cohort, feasibility was achieved for 95.7% of the lesions. This result is consistent with the rate of 89% in 28 lesions reported by Engstrand et al. with another navigation device [20]. In our PP population, we observed one case of non-feasibility, whereas the first trajectory could not be reached by the robot arm which was stopped before a possible contact with the patient, the second attempt resulted in the sliding of the needle along the liver capsule due to a steep approach angle. Therefore, in future it would be relevant to plan trajectories as perpendicular as possible to the liver capsule. In our study, the mean number of needle adjustment per lesion was 0.4 (± 0.7) and 29.2% of lesions requiring needle adjustment. It is lower than rates from other reports of robotic needle placement with manual repositioning required in 48.7% to 60% of the interventions, and a mean number of adjustment from 0.8 (± 0.8) to 1.1 (± 0.7) per lesion [7, 17]. Needle placement has a major impact on the treatment outcomes since any adjustment may increase radiation dose, procedure time, tissue trauma, and possibly the risk of bleeding and tumor seeding; but moreover approximative needle placement will decrease the probability of local tumor control [20, 22, 24, 25].
The targeting potentialities are highly dependent on tumor visibility and location of the lesion within the liver, with a third of lesions judged “unablatable” in the clinical setting [21, 26]. We reported that 46% of our patients had tumors judged as challenging to target (Fig. 5), but tumor location did not influence feasibility. Lachenmayer et al. [21] and Tinguely et al. [22] using a CT-guided percutaneous system, and Odisio et al. [27] using freehand insertion under CT guidance for liver ablations recently reported similar results. Tinguely et al. showed complex targeting trajectories such as intercostal trajectories and steep trajectory angles have no impact on targeting accuracy [22]. In our study, eighteen (75%) trajectories were out of plane with a minimum and maximum craniocaudal angulations from − 16.9 to 61.4°, demonstrating the ability of the device to target lesions requiring significant angulation without impacting feasibility.
×
There are several limitations and potential shortcomings to the present study. First and foremost, this is not a comparative and randomized trial. We chose to design a single-arm study to collect the first data in Human with a new robotic device. A simple design was therefore preferred to demonstrate the main outcomes which were feasibility and safety. Additionally, the number of patients and lesions may be too small to allow robust statistical analyses and identify differences in our results compared to other data from the literature. Furthermore, the radiation dose received by the patient was not analyzed in the present study although the low rate of needle adjustment demonstrated in this study highlights the fact that the majority of patients had only 3 CT-scan acquisitions (treatment planning CT, needle placement assessment CT, and post-ablation CT), and no radiation was delivered to the operator. These could be interesting parameter to evaluate in the future.
Anzeige
Conclusion
Robotic arm assistance for needle positioning during thermal ablation was safe and allowed a single insertion in 70.8% of targeted tumors, with an overall mean number of adjustments of 0.4 ± 0.7 per lesion.
Our present first-in-human limited series suggests both feasibility and safety of a stereotactic CT-guided device for the percutaneous needle insertion for liver tumors treatment. Other CT-guided interventions such as biopsies and other organs in abdomen might benefit from such needle guidance.
Declarations
Conflict of interest
The authors of this manuscript declare relationships with the following companies: Thierry de Baère is consultant for and receive research grant from Quantum Surgical, consultant and speaker for Boston Scientific, GE Healthcare, AstraZeneca and Boris Guiu is consultant for Quantum Surgical, Boston Scientific, Canon Medical, Roche, Ipsen, Terumo and received research grant from his work.
Ethical Approval
The present study was submitted and approved by the Ethics Committee (Comité de Protection des Personnes, Ile de France V, France) and was registered on ClinicalTrials.gov (NCT04230642).
Informed Consent
Informed consent was obtained from all individual participants included in the study.
Consent for Publication
Patients signed informed consent regarding publishing their data and photographs.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.
Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.
Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.
Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.
Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.
In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.
Update Radiologie
Bestellen Sie unseren Fach-Newsletterund bleiben Sie gut informiert.