Abstract

Objective

We tested the hypothesis that Flat Detector computed tomography (FD-CT) with intravenous contrast medium would allow the calculation of whole brain cerebral blood volume (CBV) mapping (FD-CBV) and would correlate with multislice Perfusion CT (PCT).

Methods

Twenty five patients were investigated with FD-CBV and PCT. Correlation of the CBV maps of both techniques was carried out with measurements from six anatomical regions from both sides of the brain. Mean values of each region and the correlation coefficient were calculated. Bland-Altman analysis was performed to compare the two different imaging techniques.

Results

The image and data quality of both PCT and FD-CBV were suitable for evaluation in all patients. The mean CBV values of FD-CBV and PCT showed only minimal differences with overlapping standard deviation. The correlation coefficient was 0.79 (p < 0.01). Bland-Altman analysis showed a mean difference of -0.077 ± 0.48 ml/100 g between FD-CBV and PCT CBV measurements, indicating that FD-CBV values were only slightly lower than those of PCT.

Conclusion

CBV mapping with intravenous contrast medium using Flat Detector CT compared favourably with multislice PCT. The ability to assess cerebral perfusion within the angiographic suite may improve the management of ischaemic stroke and evaluation of the efficacy of dedicated therapies.
Literatur
1.
Zurück zum Zitat Fahrig R, Fox S, Lownie S, Holdsworth DW (1997) Use of a C-arm system to generate true 3-D computed rotational angiograms: Preliminary in vitro and in vivo results. AJNR Am J Neuroradiol 18:1507–1514PubMed Fahrig R, Fox S, Lownie S, Holdsworth DW (1997) Use of a C-arm system to generate true 3-D computed rotational angiograms: Preliminary in vitro and in vivo results. AJNR Am J Neuroradiol 18:1507–1514PubMed
2.
Zurück zum Zitat Akpek S, Brunner T, Benndorf G, Strother C (2005) Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector. Diagn Interv Radiol 11:10–13PubMed Akpek S, Brunner T, Benndorf G, Strother C (2005) Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector. Diagn Interv Radiol 11:10–13PubMed
4.
Zurück zum Zitat Zellerhof M, Scholz B, Rührnschopf E (2005) Low contrast 3D-reconstruction from C-arm data. In Flynn M (ed) Medical Imaging 2005: Physics of Medical Imaging. Proc SPIE 5745:646-655 doi:10.1117/12.593433 Zellerhof M, Scholz B, Rührnschopf E (2005) Low contrast 3D-reconstruction from C-arm data. In Flynn M (ed) Medical Imaging 2005: Physics of Medical Imaging. Proc SPIE 5745:646-655 doi:10.​1117/​12.​593433
5.
Zurück zum Zitat Doelken M, Struffert T, Richter G, Engelhorn T, Nimsky C, Ganslandt O et al (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50:517–523PubMedCrossRef Doelken M, Struffert T, Richter G, Engelhorn T, Nimsky C, Ganslandt O et al (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50:517–523PubMedCrossRef
6.
Zurück zum Zitat Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender W et al (2009) Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19:619–25PubMedCrossRef Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender W et al (2009) Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19:619–25PubMedCrossRef
7.
Zurück zum Zitat Söderman M, Babic D, Holmin S, Andersson T (2008) Brain imaging with a flat detector C-arm: Technique and clinical interest of XperCT. Neuroradiology 50:863–898PubMedCrossRef Söderman M, Babic D, Holmin S, Andersson T (2008) Brain imaging with a flat detector C-arm: Technique and clinical interest of XperCT. Neuroradiology 50:863–898PubMedCrossRef
8.
Zurück zum Zitat Engelhorn T, Struffert T, Richter G, Doelken M, Ganslandt O, Kalender W et al (2008) Flat-panel detector angiographic CT in the management of aneurysmal rupture during coil embolization. AJNR Am J Neuroradiol 29:1581–1584PubMedCrossRef Engelhorn T, Struffert T, Richter G, Doelken M, Ganslandt O, Kalender W et al (2008) Flat-panel detector angiographic CT in the management of aneurysmal rupture during coil embolization. AJNR Am J Neuroradiol 29:1581–1584PubMedCrossRef
9.
Zurück zum Zitat Zellerhoff M, Deuerling-Zheng Y, Strother CM, Ahmed AS, Pulfer K, Redel T, et al. (2009) Measurement of cerebral blood volume using angiographic C-arm systems. In Hu XP, Clough AV (eds) Medical Imaging 2009: Biomedical Applications in Molecular, Structural, and Functional Imaging. Proc SPIE 7262:72620H-72620H-8 Zellerhoff M, Deuerling-Zheng Y, Strother CM, Ahmed AS, Pulfer K, Redel T, et al. (2009) Measurement of cerebral blood volume using angiographic C-arm systems. In Hu XP, Clough AV (eds) Medical Imaging 2009: Biomedical Applications in Molecular, Structural, and Functional Imaging. Proc SPIE 7262:72620H-72620H-8
10.
Zurück zum Zitat Hamberg LM, Hunter GJ, Kierstead D, Lo EH, Gonzalez RG, Wolf GL (1996) Measurement of cerebral blood volume with subtraction three-dimensional functional CT. AJNR Am J Neuroradiol 17:1861–1869PubMed Hamberg LM, Hunter GJ, Kierstead D, Lo EH, Gonzalez RG, Wolf GL (1996) Measurement of cerebral blood volume with subtraction three-dimensional functional CT. AJNR Am J Neuroradiol 17:1861–1869PubMed
11.
Zurück zum Zitat Hunter GJ, Silvennoinen HM, Hamberg LM, Koroshetz WJ, Buonanno FS, Schwamm LH et al (2003) Whole-brain CT perfusion measurement of perfused cerebral blood volume in acute ischemic stroke: probability curve for regional infarction. Radiology 227:725–730PubMedCrossRef Hunter GJ, Silvennoinen HM, Hamberg LM, Koroshetz WJ, Buonanno FS, Schwamm LH et al (2003) Whole-brain CT perfusion measurement of perfused cerebral blood volume in acute ischemic stroke: probability curve for regional infarction. Radiology 227:725–730PubMedCrossRef
12.
Zurück zum Zitat Kloska S, Fischer T, Nabavi D, Dittrich R, Ditt H, Klotz E et al (2007) Color-coded perfused blood volume imaging using multidetector CT: initial results of whole-brain perfusion analysis in acute cerebral ischemia. Eur Radiol 17:2352–2358PubMedCrossRef Kloska S, Fischer T, Nabavi D, Dittrich R, Ditt H, Klotz E et al (2007) Color-coded perfused blood volume imaging using multidetector CT: initial results of whole-brain perfusion analysis in acute cerebral ischemia. Eur Radiol 17:2352–2358PubMedCrossRef
13.
Zurück zum Zitat Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y et al (2009) C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol 30:917–922PubMedCrossRef Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y et al (2009) C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol 30:917–922PubMedCrossRef
14.
Zurück zum Zitat Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y et al (2010) C-Arm CT measurement of cerebral blood volume in ischemic stroke: an experimental study in canines. AJNR Am J Neuroradiol 31:536–540PubMedCrossRef Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y et al (2010) C-Arm CT measurement of cerebral blood volume in ischemic stroke: an experimental study in canines. AJNR Am J Neuroradiol 31:536–540PubMedCrossRef
15.
Zurück zum Zitat Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender W et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. doi:10.3174/ajnr.A2083 Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender W et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. doi:10.​3174/​ajnr.​A2083
16.
Zurück zum Zitat Klotz E, Koenig M (1999) Perfusion measurements of the brain: using dynamic CT for the quantitative assessment of cerebral ischemia in acute stroke. Eur J Radiol 30:170–184PubMedCrossRef Klotz E, Koenig M (1999) Perfusion measurements of the brain: using dynamic CT for the quantitative assessment of cerebral ischemia in acute stroke. Eur J Radiol 30:170–184PubMedCrossRef
17.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–3010PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–3010PubMed
18.
Zurück zum Zitat Schellinger PD, Fiebach JB, Hacke W (2003) Imaging-based decision making in thrombolytic therapy for ischemic stroke: present status. Stroke 34:575–583PubMedCrossRef Schellinger PD, Fiebach JB, Hacke W (2003) Imaging-based decision making in thrombolytic therapy for ischemic stroke: present status. Stroke 34:575–583PubMedCrossRef
19.
Zurück zum Zitat Muir KW, Buchan A, von Kummer R, Rother J, Baron JC (2005) Imaging of acute stroke. Lancet Neurol 5:755–768CrossRef Muir KW, Buchan A, von Kummer R, Rother J, Baron JC (2005) Imaging of acute stroke. Lancet Neurol 5:755–768CrossRef
20.
Zurück zum Zitat Wintermark M, Flanders AE, Velthuis B, Meuli R, van Leeuwen M, Goldsher D et al (2006) Perfusion-CT assessment of infarct core and penumbra: receiver operating characteristic curve analysis in 130 patients suspected of acute hemispheric stroke. Stroke 37:979–985PubMedCrossRef Wintermark M, Flanders AE, Velthuis B, Meuli R, van Leeuwen M, Goldsher D et al (2006) Perfusion-CT assessment of infarct core and penumbra: receiver operating characteristic curve analysis in 130 patients suspected of acute hemispheric stroke. Stroke 37:979–985PubMedCrossRef
21.
Zurück zum Zitat Parsons MW, Pepper EM, Bateman GA, Wang Y, Levi CR (2007) Identification of the penumbra and infarct core on hyperacute noncontrast and perfusion CT. Neurology 68:730–736PubMedCrossRef Parsons MW, Pepper EM, Bateman GA, Wang Y, Levi CR (2007) Identification of the penumbra and infarct core on hyperacute noncontrast and perfusion CT. Neurology 68:730–736PubMedCrossRef
22.
Zurück zum Zitat Murphy BD, Fox AJ, Lee DH, Sahlas DJ, Black SE, Hogan MJ et al (2008) White matter thresholds for ischemic penumbra and infarct core in patients with acute stroke: CT perfusion study. Radiology 247:818–825PubMedCrossRef Murphy BD, Fox AJ, Lee DH, Sahlas DJ, Black SE, Hogan MJ et al (2008) White matter thresholds for ischemic penumbra and infarct core in patients with acute stroke: CT perfusion study. Radiology 247:818–825PubMedCrossRef
23.
Zurück zum Zitat Broderick JP (2009) Endovascular therapy for acute ischemic stroke. Stroke 40(3 Suppl):S103–106PubMedCrossRef Broderick JP (2009) Endovascular therapy for acute ischemic stroke. Stroke 40(3 Suppl):S103–106PubMedCrossRef
24.
Zurück zum Zitat Khatri P, Abruzzo T, Yeatts SD, Nichols C, Broderick JP, Tomsick TA, IMS I, Investigators II (2009) Good clinical outcome after ischemic stroke with successful revascularization is time-dependent. Neurology 73:1066–1072PubMedCrossRef Khatri P, Abruzzo T, Yeatts SD, Nichols C, Broderick JP, Tomsick TA, IMS I, Investigators II (2009) Good clinical outcome after ischemic stroke with successful revascularization is time-dependent. Neurology 73:1066–1072PubMedCrossRef
25.
Zurück zum Zitat Orth R, Wallace M, Kuo M (2008) C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 19:814–820PubMedCrossRef Orth R, Wallace M, Kuo M (2008) C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 19:814–820PubMedCrossRef
Metadaten
Titel
Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT
verfasst von
Tobias Struffert
Yu Deuerling-Zheng
Stephan Kloska
Tobias Engelhorn
Jan Boese
Michael Zellerhoff
Stefan Schwab
Arnd Doerfler
Publikationsdatum
01.04.2011
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 4/2011
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-1957-6

Weitere Artikel der Ausgabe 4/2011

European Radiology 4/2011 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.