Introduction

The aim of this study was to test the mechanical stability of LOQTEQ® HTO plate in comparison to the TomoFix™ internal plate fixator.

Materials and methods

An unstable fracture gap model with two standardized polyoxymethylene (POM) cylinders was used for this study. With this model 5 LOQTEQ® HTO plates and 5 TomoFix™ were cyclically loaded. The start load was 100 N and increased by 50 N after each successful completed cycle. Each load step included 250,000 cycles. All tests were performed in accordance with the American Society for Testing and Materials (ASTM) Standard Specification and Test Method for Metallic Bone Plates F382–99(2003)e1.

Results

The mean fatigue strength of the LOQTEQ® HTO Plate was 155 % of the value gained by the TomoFix™ implant. This difference was statistically significant (t test <0.001). There was also a significant difference in the cycles but both implants survived. The LOQTEQ® HTO Plate system survived 1,520,336.8 (±121,687.7) cycles, whereas the TomoFix™ group passed on average 847,802,600 (±134,660.7) cycles. In both groups, the only failure mode was plate breakage in a shaft hole.

Conclusions

The results of the present study showed that fatigue strength of the LOQTEQ® HTO plate was significantly higher in comparison to TomoFix™. These results justify the clinical use of the LOQTEQ® HTO plate.
Literatur
1.
Zurück zum Zitat Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14(3):291–300PubMedCrossRef Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14(3):291–300PubMedCrossRef
2.
Zurück zum Zitat Ahmad M, Nanda R, Bajwa AS, Candal-Couto J, Green S, Hui AC (2007) Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? Injury 38(3):358–364PubMedCrossRef Ahmad M, Nanda R, Bajwa AS, Candal-Couto J, Green S, Hui AC (2007) Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? Injury 38(3):358–364PubMedCrossRef
3.
Zurück zum Zitat Annual book of ASTM Standards, ASTM International (2004) American Society for Testing & Materials Annual book of ASTM Standards, ASTM International (2004) American Society for Testing & Materials
4.
Zurück zum Zitat Bode G, von Heyden J, Pestka J, Schmal H, Salzmann G, Südkamp N, Niemeyer P (2013) Prospective 5-year survival rate data following open-wedge valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 19 (Epub ahead of print) Bode G, von Heyden J, Pestka J, Schmal H, Salzmann G, Südkamp N, Niemeyer P (2013) Prospective 5-year survival rate data following open-wedge valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 19 (Epub ahead of print)
5.
Zurück zum Zitat Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90(12):1548–1557PubMedCrossRef Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90(12):1548–1557PubMedCrossRef
6.
Zurück zum Zitat Estes C, Rhee P, Shrader MW, Csavina K, Jacofsky MC, Jacofsky DJ (2008) Biomechanical strength of the Peri-Loc proximal tibial plate: a comparison of all-locked versus hybrid locked/nonlocked screw configurations. J Orthop Trauma 22(5):312–316PubMedCrossRef Estes C, Rhee P, Shrader MW, Csavina K, Jacofsky MC, Jacofsky DJ (2008) Biomechanical strength of the Peri-Loc proximal tibial plate: a comparison of all-locked versus hybrid locked/nonlocked screw configurations. J Orthop Trauma 22(5):312–316PubMedCrossRef
7.
Zurück zum Zitat Floerkemeier S, Staubli AE, Schroeter S, Goldhahn S (2013) Lobenhoffer P (2013) Outcome after high tibial open-wedge osteotomy: a retrospective evaluation of 533 patients. Knee Surg Sports Traumatol Arthrosc 21(1):170–180PubMedCrossRef Floerkemeier S, Staubli AE, Schroeter S, Goldhahn S (2013) Lobenhoffer P (2013) Outcome after high tibial open-wedge osteotomy: a retrospective evaluation of 533 patients. Knee Surg Sports Traumatol Arthrosc 21(1):170–180PubMedCrossRef
8.
Zurück zum Zitat Florin M, Arzdorf M, Linke B, Auer JA (2005) Assessment of stiffness and strength of 4 different implants available for equine fracture treatment: a study on a 20 degrees oblique long-bone fracture model using a bone substitute. Vet Surg 34(3):231–238PubMedCrossRef Florin M, Arzdorf M, Linke B, Auer JA (2005) Assessment of stiffness and strength of 4 different implants available for equine fracture treatment: a study on a 20 degrees oblique long-bone fracture model using a bone substitute. Vet Surg 34(3):231–238PubMedCrossRef
9.
Zurück zum Zitat Gaebler C, Speitling A, Milne EL, Stanzl-Tschegg S, Vecsei V, Latta LL (2001) A new modular testing system for biomechanical evaluation of tibial intramedullary fixation devices. Injury 32(9):708–712PubMedCrossRef Gaebler C, Speitling A, Milne EL, Stanzl-Tschegg S, Vecsei V, Latta LL (2001) A new modular testing system for biomechanical evaluation of tibial intramedullary fixation devices. Injury 32(9):708–712PubMedCrossRef
12.
Zurück zum Zitat Irubetagoyena I, Verset M, Palierne S, Swider P, Autefage A (2013) Ex vivo cyclic mechanical behaviour of 2.4 mm locking plates compared with 2.4 mm limited contact plates in a cadaveric diaphyseal gap model. Vet Comp Orthop Traumatol 26(6):479–488PubMedCrossRef Irubetagoyena I, Verset M, Palierne S, Swider P, Autefage A (2013) Ex vivo cyclic mechanical behaviour of 2.4 mm locking plates compared with 2.4 mm limited contact plates in a cadaveric diaphyseal gap model. Vet Comp Orthop Traumatol 26(6):479–488PubMedCrossRef
13.
Zurück zum Zitat Jung WH, Chun CW, Lee JH, Ha JH, Kim JH, Jeong JH (2013) Comparative study of medial opening-wedge high tibial osteotomy using 2 different implants. Arthroscopy 29(6):1063–1071PubMedCrossRef Jung WH, Chun CW, Lee JH, Ha JH, Kim JH, Jeong JH (2013) Comparative study of medial opening-wedge high tibial osteotomy using 2 different implants. Arthroscopy 29(6):1063–1071PubMedCrossRef
14.
Zurück zum Zitat Kobayashi S, Ohgoe Y, Ozeki K, Hirakuri K, Aoki H (2007) Dissolution effect and cytotoxicity of diamond-like carbon coatings on orthodontic archwires. J Mater Sci Mater Med 18(12):2263–2268PubMedCrossRef Kobayashi S, Ohgoe Y, Ozeki K, Hirakuri K, Aoki H (2007) Dissolution effect and cytotoxicity of diamond-like carbon coatings on orthodontic archwires. J Mater Sci Mater Med 18(12):2263–2268PubMedCrossRef
15.
Zurück zum Zitat Lobenhoffer P, Agneskirchner J, Zoch W (2004) Open valgus alignment osteotomy of the proximal tibia with fixation by medial plate fixator. Orthopade 33(2):153–160PubMedCrossRef Lobenhoffer P, Agneskirchner J, Zoch W (2004) Open valgus alignment osteotomy of the proximal tibia with fixation by medial plate fixator. Orthopade 33(2):153–160PubMedCrossRef
16.
Zurück zum Zitat Luo CA, Hua SY, Lin SC, Chen CM, Tseng CS (2013) Stress and stability comparison between different systems for high tibial osteotomies. BMC Musculoskelet Disord 25(14):110. doi:10.1186/1471-2474-14-110 CrossRef Luo CA, Hua SY, Lin SC, Chen CM, Tseng CS (2013) Stress and stability comparison between different systems for high tibial osteotomies. BMC Musculoskelet Disord 25(14):110. doi:10.​1186/​1471-2474-14-110 CrossRef
17.
Zurück zum Zitat Luites JW, Brinkman JM, Wymenga AB, van Heerwaarden RJ (2009) Fixation stability of opening- versus closing-wedge high tibial osteotomy: a randomised clinical trial using radiostereometry. J Bone Joint Surg Br 91(11):1459–1465PubMedCrossRef Luites JW, Brinkman JM, Wymenga AB, van Heerwaarden RJ (2009) Fixation stability of opening- versus closing-wedge high tibial osteotomy: a randomised clinical trial using radiostereometry. J Bone Joint Surg Br 91(11):1459–1465PubMedCrossRef
18.
Zurück zum Zitat Matsushita K, Inoue N, Totsuka Y (2011) In vitro biomechanical evaluation of the effect of an additional L-shaped plate on straight or box plate fixation in sagittal split ramus osteotomy using a bioabsorbable plate system. Oral Maxillofac Surg 15(3):139–146PubMedCrossRef Matsushita K, Inoue N, Totsuka Y (2011) In vitro biomechanical evaluation of the effect of an additional L-shaped plate on straight or box plate fixation in sagittal split ramus osteotomy using a bioabsorbable plate system. Oral Maxillofac Surg 15(3):139–146PubMedCrossRef
19.
Zurück zum Zitat Niemeyer P, Schmal H, Hauschild O, von Heyden J, Südkamp NP, Köstler W (2010) Open-wedge osteotomy using an internal plate fixator in patients with medial-compartment gonarthritis and varus malalignment: 3-year results with regard to preoperative arthroscopic and radiographic findings. Arthroscopy 26(12):1607–1616PubMedCrossRef Niemeyer P, Schmal H, Hauschild O, von Heyden J, Südkamp NP, Köstler W (2010) Open-wedge osteotomy using an internal plate fixator in patients with medial-compartment gonarthritis and varus malalignment: 3-year results with regard to preoperative arthroscopic and radiographic findings. Arthroscopy 26(12):1607–1616PubMedCrossRef
20.
Zurück zum Zitat Schröter S, Gonser CE, Konstantinidis L, Helwig P, Albrecht D (2011) High complication rate after biplanar open wedge high tibial osteotomy stabilized with a new spacer plate (Position HTO plate) without bone substitute. Arthroscopy 27(5):644–652PubMedCrossRef Schröter S, Gonser CE, Konstantinidis L, Helwig P, Albrecht D (2011) High complication rate after biplanar open wedge high tibial osteotomy stabilized with a new spacer plate (Position HTO plate) without bone substitute. Arthroscopy 27(5):644–652PubMedCrossRef
21.
Zurück zum Zitat Spahn G (2004) Complications in high tibial (medial opening wedge) osteotomy. Arch Orthop Trauma Surg 124(10):649–653PubMedCrossRef Spahn G (2004) Complications in high tibial (medial opening wedge) osteotomy. Arch Orthop Trauma Surg 124(10):649–653PubMedCrossRef
22.
Zurück zum Zitat Staubli AE (2011) Open-wedge high tibial osteotomy with a locking plate (tomofix) for treatment of medial monocompartment osteoarthritis. Elsevier 21(2):156–162 Staubli AE (2011) Open-wedge high tibial osteotomy with a locking plate (tomofix) for treatment of medial monocompartment osteoarthritis. Elsevier 21(2):156–162
23.
Zurück zum Zitat Staubli A, Hilaire A, Jacob C (2010) Evolution of open-wedge high-tibial osteotomy: experience with a special angular stable device for internal fixation without interposition material. Int Orthop 34(2):167–172PubMedCrossRefPubMedCentral Staubli A, Hilaire A, Jacob C (2010) Evolution of open-wedge high-tibial osteotomy: experience with a special angular stable device for internal fixation without interposition material. Int Orthop 34(2):167–172PubMedCrossRefPubMedCentral
24.
Zurück zum Zitat Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):B55–B62PubMedCrossRef Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):B55–B62PubMedCrossRef
25.
Zurück zum Zitat Uhl JM, Seguin B, Kapatkin AS, Schulz KS, Garcia TC, Stover SM (2008) Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models. Vet Surg 37(7):663–673PubMedCrossRef Uhl JM, Seguin B, Kapatkin AS, Schulz KS, Garcia TC, Stover SM (2008) Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models. Vet Surg 37(7):663–673PubMedCrossRef
26.
Zurück zum Zitat van den Bekerom MP, Patt TW, Kleinhout MY, van der Vis HM, Albers GH (2008) Early complications after high tibial osteotomy: a comparison of two techniques. J Knee Surg 21:68–74 van den Bekerom MP, Patt TW, Kleinhout MY, van der Vis HM, Albers GH (2008) Early complications after high tibial osteotomy: a comparison of two techniques. J Knee Surg 21:68–74
Metadaten
Titel
Stability of two angular stable locking plates for open wedge high tibial osteotomy (HTO): TomoFix™ versus LOQTEQ® HTO plate
verfasst von
Wolf Petersen
Alexander Wall
Thomas Paulin
Hi Un Park
Lydia Heymann
Publikationsdatum
01.10.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 10/2014
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-014-2042-4

Weitere Artikel der Ausgabe 10/2014

Archives of Orthopaedic and Trauma Surgery 10/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.