Dual X-ray Absorptiometry: Cross-Calibration of a New Fan-Beam System

verfasst von: G. M. Blake, E. J. Harrison, J. E. Adams

Erschienen in: Calcified Tissue International | Ausgabe 1/2004

Einloggen, um Zugang zu erhalten

Abstract

The high precision and stable calibration of dual X-ray absorptiometry (DXA) scanners have led to their widespread use in longitudinal studies for research and the follow-up of individual patients who are receiving treatment for osteoporosis. However, difficulties in maintaining the continuity of the bone mineral density (BMD) calibration scale can arise when an old DXA system is replaced by a newer model. We report the results of an in vivo cross-calibration study performed when a GE-Lunar Prodigy fan-beam system replaced a DPX-L pencil-beam scanner. Lumbar spine and hip DXA scans were performed in 133 patients (104 female, 29 male) attending long-term BMD monitoring. On average, lumbar spine BMD measurements on the two systems agreed closely, with Prodigy values 1% lower than those on the DPX-L. However, after allowing for this difference, the root mean square error (RMSE) of 0.037 g/cm2 was larger than in previous cross-calibration studies reported in the literature, and was 3 times the value expected from the precision of the BMD measurements. Mean femoral neck BMD also agreed closely between the two systems, although for Prodigy, the spread of measurements was 10% smaller than that for the DPX-L. For the trochanter and Ward’s triangle regions, mean BMD was 4% and 6% lower, respectively, on the Prodigy system, and the results were affected by a similar compression of the range of values. RMSE values were 0.037 g/cm2, 0.044 g/cm2, and 0.044 g/cm2, respectively, for the femoral neck, trochanter, and Ward’s triangle sites. When the high value of the RMSE was investigated, it was found that for lumbar spine BMD, patient body weight and the difference between the two systems in the percentage fat reported in the soft tissue reference region explained 40% of the variance. This enabled equations to be developed that significantly improved the agreement between scans performed on the two systems. Smaller improvements were obtained for the femur BMD measurements.
Literatur
1.
Zurück zum Zitat Genant, HK, Engelke, K, Fuerst, T 1996Noninvasive assessment of bone mineral and structure: state of the art. J Bone Miner Res11707730PubMed Genant, HK, Engelke, K, Fuerst, T 1996Noninvasive assessment of bone mineral and structure: state of the art. J Bone Miner Res11707730PubMed
2.
Zurück zum Zitat Baran, DT, Faulkner, KG, Genant, HK 1997Diagnosis and management of osteoporosis: guidelines for the utilization of bone densitometry.Calcif Tissue Int61433440CrossRefPubMed Baran, DT, Faulkner, KG, Genant, HK 1997Diagnosis and management of osteoporosis: guidelines for the utilization of bone densitometry.Calcif Tissue Int61433440CrossRefPubMed
3.
Zurück zum Zitat Kanis, JA Delmas, P Burckhardt, P Cooper, C Torgerson, D on behalf of the European Foundation for Osteoporosis and Bone Disease1997Guidelines for diagnosis and treatment of osteoporosis.Osteoporos Int7390406PubMed Kanis, JA Delmas, P Burckhardt, P Cooper, C Torgerson, D on behalf of the European Foundation for Osteoporosis and Bone Disease1997Guidelines for diagnosis and treatment of osteoporosis.Osteoporos Int7390406PubMed
4.
Zurück zum Zitat Kanis, JA, Glüer, C-C 2000An update on the diagnosis and assessment of osteoporosis with densitometry.Osteoporos Int11192202CrossRefPubMed Kanis, JA, Glüer, C-C 2000An update on the diagnosis and assessment of osteoporosis with densitometry.Osteoporos Int11192202CrossRefPubMed
5.
Zurück zum Zitat Cooper, C, Campion, G, Melton, LJ 1992Hip fractures in the elderly: a world-wide projection.Osteoporos Int2285289PubMed Cooper, C, Campion, G, Melton, LJ 1992Hip fractures in the elderly: a world-wide projection.Osteoporos Int2285289PubMed
6.
Zurück zum Zitat Schurch, M-A, Rizzoli, R, Mermillod, B, Vasey, H, Michel, JP, Bonjour, J-P 1996A prospective study on socio-economic aspects of fracture of the proximal femur.Bone Miner Res1119351942 Schurch, M-A, Rizzoli, R, Mermillod, B, Vasey, H, Michel, JP, Bonjour, J-P 1996A prospective study on socio-economic aspects of fracture of the proximal femur.Bone Miner Res1119351942
7.
Zurück zum Zitat Ray, NF, Chan, JK, Thamer, M, Melton, LJ 1997Medical expenditures for the treatment of osteoporotic fractures in the United States in 1995: report from the National Osteoporosis Foundation.J Bone Miner Res122435PubMed Ray, NF, Chan, JK, Thamer, M, Melton, LJ 1997Medical expenditures for the treatment of osteoporotic fractures in the United States in 1995: report from the National Osteoporosis Foundation.J Bone Miner Res122435PubMed
8.
Zurück zum Zitat Black, DM, Cummings, SR, Karpf, DB 1996Randomised trial of the effect of alendronate on risk of fracture in women with existing vertebral fractures.Lancet34815351541CrossRefPubMed Black, DM, Cummings, SR, Karpf, DB 1996Randomised trial of the effect of alendronate on risk of fracture in women with existing vertebral fractures.Lancet34815351541CrossRefPubMed
9.
Zurück zum Zitat Ettinger, B, Black, DM, Mitlak, BH 1999Reduction of vertebral fracture risk in postmenopausal women with osteoporosis treated with raloxifene: results from a 3-year randomized clinical trial.JAMA282637645CrossRefPubMed Ettinger, B, Black, DM, Mitlak, BH 1999Reduction of vertebral fracture risk in postmenopausal women with osteoporosis treated with raloxifene: results from a 3-year randomized clinical trial.JAMA282637645CrossRefPubMed
10.
Zurück zum Zitat Harris, ST, Watts, NB, Genant, HK 1999Effects of risedronate treatment on vertebral and nonvertebral fractures in women with postmenopausal osteoporosis.JAMA28213441352CrossRefPubMed Harris, ST, Watts, NB, Genant, HK 1999Effects of risedronate treatment on vertebral and nonvertebral fractures in women with postmenopausal osteoporosis.JAMA28213441352CrossRefPubMed
11.
Zurück zum Zitat Neer, RM, Arnaud, CD, Zanchetta, JR 2001Effect of recombinant human parathyroid hormone (1-34) fragment on spine and non-spine fractures and bone mineral density in postmenopausal osteoporosis.N Engl J Med34414341441CrossRefPubMed Neer, RM, Arnaud, CD, Zanchetta, JR 2001Effect of recombinant human parathyroid hormone (1-34) fragment on spine and non-spine fractures and bone mineral density in postmenopausal osteoporosis.N Engl J Med34414341441CrossRefPubMed
12.
Zurück zum Zitat Royal College of Physicians1999Osteoporosis: clinical guidelines for prevention and treatment.RCPLondon, England Royal College of Physicians1999Osteoporosis: clinical guidelines for prevention and treatment.RCPLondon, England
13.
Zurück zum Zitat WHO Technical Report Series 8431994Assessment of fracture risk and its application to screening for postmenopausal osteoporosis.World Health OrganizationGeneva WHO Technical Report Series 8431994Assessment of fracture risk and its application to screening for postmenopausal osteoporosis.World Health OrganizationGeneva
14.
Zurück zum Zitat Royal College of Physicians2000Osteoporosis: clinical guidelines for prevention and treatment. Update on pharmacological interventions and an algorithm for management.RCPLondon, England Royal College of Physicians2000Osteoporosis: clinical guidelines for prevention and treatment. Update on pharmacological interventions and an algorithm for management.RCPLondon, England
16.
Zurück zum Zitat Liberman, UA, Weiss, SR, Bröll, J 1995Effect of oral alendronate on bone mineral density and the incidence of fractures in postmenopausal osteoporosis.N Engl J Med33314371443CrossRefPubMed Liberman, UA, Weiss, SR, Bröll, J 1995Effect of oral alendronate on bone mineral density and the incidence of fractures in postmenopausal osteoporosis.N Engl J Med33314371443CrossRefPubMed
17.
Zurück zum Zitat Tonino, RP, Meunier, PJ, Emkey, R 2000Skeletal benefits of alendronate: 7-year treatment of postmenopausal osteoporotic women.J Clin Endocrinol Metab8531093115CrossRefPubMed Tonino, RP, Meunier, PJ, Emkey, R 2000Skeletal benefits of alendronate: 7-year treatment of postmenopausal osteoporotic women.J Clin Endocrinol Metab8531093115CrossRefPubMed
18.
Zurück zum Zitat Cummings, SR, Black, DM, Thompson, DE 1998Effect of alendronate on risk of fracture in women with low bone density but without vertebral fracture: results from the Fracture Intervention Trial.JAMA28020772082CrossRefPubMed Cummings, SR, Black, DM, Thompson, DE 1998Effect of alendronate on risk of fracture in women with low bone density but without vertebral fracture: results from the Fracture Intervention Trial.JAMA28020772082CrossRefPubMed
19.
Zurück zum Zitat Reginster, J-Y, Minne, HW, Sorensen, OH 2000Randomised trial of the effects of risedronate on vertebral fractures in women with established postmenopausal osteoporosis.Osteoporos Int118391CrossRefPubMed Reginster, J-Y, Minne, HW, Sorensen, OH 2000Randomised trial of the effects of risedronate on vertebral fractures in women with established postmenopausal osteoporosis.Osteoporos Int118391CrossRefPubMed
20.
Zurück zum Zitat Chesnut, CH, Silverman, S, Andriano, K 2000A randomized trial of nasal spray salmon calcitonin in postmenopausal women with established osteoporosis: the prevent recurrence of osteoporotic fractures study.Am J Med109267276CrossRefPubMed Chesnut, CH, Silverman, S, Andriano, K 2000A randomized trial of nasal spray salmon calcitonin in postmenopausal women with established osteoporosis: the prevent recurrence of osteoporotic fractures study.Am J Med109267276CrossRefPubMed
21.
Zurück zum Zitat McClung, MR, Geusens, P, Miller, PD 2001Effect of risedronate treatment on hip fracture risk in elderly women.N Engl J Med344333340CrossRefPubMed McClung, MR, Geusens, P, Miller, PD 2001Effect of risedronate treatment on hip fracture risk in elderly women.N Engl J Med344333340CrossRefPubMed
22.
Zurück zum Zitat Reid, IR, Brown, JP, Burckhardt, P 2002Intravenous zolendronic acid in postmenopausal women with low bone mineral density.N Engl J Med346653661CrossRefPubMed Reid, IR, Brown, JP, Burckhardt, P 2002Intravenous zolendronic acid in postmenopausal women with low bone mineral density.N Engl J Med346653661CrossRefPubMed
23.
Zurück zum Zitat Gluer, C-C, Faulkner, KG, Estilo, MJ, Engelke, K, Rosin, J, Genant, HK 1993Quality assurance for bone densitometry research studies: concept and impact.Osteoporos Int3227235PubMed Gluer, C-C, Faulkner, KG, Estilo, MJ, Engelke, K, Rosin, J, Genant, HK 1993Quality assurance for bone densitometry research studies: concept and impact.Osteoporos Int3227235PubMed
24.
Zurück zum Zitat Faulkner, KG, McClung, MR 1995Quality control of DXA instruments in multicenter trials.Osteoporos Int5218227PubMed Faulkner, KG, McClung, MR 1995Quality control of DXA instruments in multicenter trials.Osteoporos Int5218227PubMed
25.
Zurück zum Zitat Genant, HK, Grampp, S, Glüer, C-C 1994Universal standardization for dual X-ray absorptiometry: patient and phantom cross-calibration results.J Bone Miner Res915031514 Genant, HK, Grampp, S, Glüer, C-C 1994Universal standardization for dual X-ray absorptiometry: patient and phantom cross-calibration results.J Bone Miner Res915031514
26.
Zurück zum Zitat Faulkner, KG, Gluer, C-C, Estilo, MJ, Genant, HK 1993Cross-calibration of DXA equipment: upgrading from a Hologic QDR-1000/W to a QDR-2000.Calcif Tissue Int527984PubMed Faulkner, KG, Gluer, C-C, Estilo, MJ, Genant, HK 1993Cross-calibration of DXA equipment: upgrading from a Hologic QDR-1000/W to a QDR-2000.Calcif Tissue Int527984PubMed
27.
Zurück zum Zitat Blake, GM, Parker, JC, Buxton, FMA, Fogelman, I 1993Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans.Br J Radiol66902906PubMed Blake, GM, Parker, JC, Buxton, FMA, Fogelman, I 1993Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans.Br J Radiol66902906PubMed
28.
Zurück zum Zitat Mazess, RB, Collick, B, Trempe, J, Barden, HS, Hanson, JA 1989Performance evaluation of a dual-energy X-ray bone densitometer.Calcif Tissue Int44228232PubMed Mazess, RB, Collick, B, Trempe, J, Barden, HS, Hanson, JA 1989Performance evaluation of a dual-energy X-ray bone densitometer.Calcif Tissue Int44228232PubMed
29.
Zurück zum Zitat Mazess, RB, Hanson, JA, Payne, R, Nord, R, Wilson, M 2000Axial and total-body bone densitometry using a narrow-angle fan-beam.Osteoporos Int11158166PubMed Mazess, RB, Hanson, JA, Payne, R, Nord, R, Wilson, M 2000Axial and total-body bone densitometry using a narrow-angle fan-beam.Osteoporos Int11158166PubMed
30.
Zurück zum Zitat Kalender, WA, Felsenberg, D, Genant, HK, Fischer, M, Dequeker, J, Reeve, J 1995The European Spine Phantom—a tool for standardization and quality control in spinal bone mineral measurements by DXA and OCT.Eur J Radiol208392CrossRefPubMed Kalender, WA, Felsenberg, D, Genant, HK, Fischer, M, Dequeker, J, Reeve, J 1995The European Spine Phantom—a tool for standardization and quality control in spinal bone mineral measurements by DXA and OCT.Eur J Radiol208392CrossRefPubMed
31.
Zurück zum Zitat Mazess, RB, Tempe, JA, Bisek, JP, Hanson, JA 1991Calibration of dual-energy X-ray absorptiometry (DEXA) for bone density.J Bone Miner Res6799806PubMed Mazess, RB, Tempe, JA, Bisek, JP, Hanson, JA 1991Calibration of dual-energy X-ray absorptiometry (DEXA) for bone density.J Bone Miner Res6799806PubMed
32.
Zurück zum Zitat Hanson, J 1997Standardization of femur BMD [letter].J Bone Miner Res1213161317PubMed Hanson, J 1997Standardization of femur BMD [letter].J Bone Miner Res1213161317PubMed
33.
Zurück zum Zitat Altman, DG 1991Practical statistics for medical research.Chapman HallLondon Altman, DG 1991Practical statistics for medical research.Chapman HallLondon
34.
Zurück zum Zitat Blake, GM 1996Replacing DXA scanners: cross-calibration with phantoms may be misleading.Calcif Tissue Int5915CrossRef Blake, GM 1996Replacing DXA scanners: cross-calibration with phantoms may be misleading.Calcif Tissue Int5915CrossRef
35.
Zurück zum Zitat Blake, GM, Fogelman, I 1995Replacing dual X-ray absorptiometry scanners: cross-calibration of a new multidetector array.J Bone Miner Res10S267 Blake, GM, Fogelman, I 1995Replacing dual X-ray absorptiometry scanners: cross-calibration of a new multidetector array.J Bone Miner Res10S267
36.
Zurück zum Zitat Steel, SA, Thorpe, J 2002The GE-Lunar Prodigy: in vivo precision and cross-calibration.Calcif Tissue Int71277278 Steel, SA, Thorpe, J 2002The GE-Lunar Prodigy: in vivo precision and cross-calibration.Calcif Tissue Int71277278
37.
Zurück zum Zitat Bolotin, HH, Sievanen, H, Grashuis, JL, Kuiper, JW, Jarvinen, TLN 2001Inaccuracies inherent in patient-specific dual-energy X-ray absorptiometry bone mineral density measurements: comprehensive phantom-based evaluation.J Bone Miner Res16417426PubMed Bolotin, HH, Sievanen, H, Grashuis, JL, Kuiper, JW, Jarvinen, TLN 2001Inaccuracies inherent in patient-specific dual-energy X-ray absorptiometry bone mineral density measurements: comprehensive phantom-based evaluation.J Bone Miner Res16417426PubMed
38.
Zurück zum Zitat Wilson, CR, Carrera, GF, Bergersen, J 2002The effect of patient weight on BMD measurement precision.Calcif Tissuer Int71281282 Wilson, CR, Carrera, GF, Bergersen, J 2002The effect of patient weight on BMD measurement precision.Calcif Tissuer Int71281282
Metadaten
Titel
Dual X-ray Absorptiometry: Cross-Calibration of a New Fan-Beam System
verfasst von
G. M. Blake
E. J. Harrison
J. E. Adams
Publikationsdatum
01.07.2004
Erschienen in
Calcified Tissue International / Ausgabe 1/2004
Print ISSN: 0171-967X
Elektronische ISSN: 1432-0827
DOI
https://doi.org/10.1007/s00223-004-0169-y

Weitere Artikel der Ausgabe 1/2004

Calcified Tissue International 1/2004 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Reizdarmsyndrom: Diäten wirksamer als Medikamente

29.04.2024 Reizdarmsyndrom Nachrichten

Bei Reizdarmsyndrom scheinen Diäten, wie etwa die FODMAP-arme oder die kohlenhydratreduzierte Ernährung, effektiver als eine medikamentöse Therapie zu sein. Das hat eine Studie aus Schweden ergeben, die die drei Therapieoptionen im direkten Vergleich analysierte.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.