Correlation of ductus arteriosus length and morphology between computed tomographic angiography and catheter angiography and their relation to ductal stent length

verfasst von: Siddharth P. Jadhav, Varun Aggarwal, Prakash M. Masand, Eric Diaz, Wei Zhang, Athar M. Qureshi

Erschienen in: Pediatric Radiology | Ausgabe 6/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

Patent ductus arteriosus (PDA) stent placement in infants with ductal-dependent pulmonary blood flow is being increasingly used in clinical practice.

Objective

To correlate computed tomographic (CT) angiography morphology and length of the PDA with catheter angiography and its relation to eventual PDA stent length.

Materials and methods

We retrospectively identified all pediatric patients who underwent PDA stenting at our institute from 2004 to 2018. We included children who had CT angiography prior to stenting. PDA length was measured by a radiologist blinded to the catheter angiography data, using Syngo-via post-processing software (Siemens, Erlangen, Germany). Vessel centerline technique was used. We measured the actual length of the duct as well as straight length between aortic and pulmonary ends. PDA morphology tortuosity index was classified as straight (Type I), mildly tortuous with 1 turn (Type II) and tortuous with >1 turn (Type III), and the PDA origin was noted. The PDA was also measured and morphology classified on catheter angiography by an interventional cardiologist blinded to the CT angiography findings. We compared the CT angiography and catheter angiography lengths, straight lengths and stent length using scatter plots and intraclass correlation coefficient (ICC).

Results

A total of 83 children who had PDA stenting were identified, of whom 17 had prior CT angiography. Fifteen of these were neonates. There was agreement between CT angiography and catheter angiography regarding the PDA morphology tortuosity index in 94% of cases and PDA origin in 100% of cases. There was moderate agreement between CT angiography and catheter angiography actual and straight PDA lengths, with ICC coefficients of 0.65 and 0.68, respectively. There was moderate agreement between CT angiography actual length, CT angiography straight length, catheter angiography actual length and eventual stented PDA length, with ICCs of 0.57, 0.67 and 0.73, respectively. There was poor agreement between catheter angiography straight length and eventual stented PDA length, with an ICC of 0.39.

Conclusion

PDA length and morphology description on CT angiography correlates well with catheter angiography and can be a reliable guide for the interventional cardiologist in decision-making regarding appropriate choice of PDA stent length.
Literatur
1.
Zurück zum Zitat Gibbs JL, Rothman MT, Rees MR et al (1992) Stenting of the arterial duct: a new approach to palliation for pulmonary atresia. Br Heart J 67:240–245CrossRef Gibbs JL, Rothman MT, Rees MR et al (1992) Stenting of the arterial duct: a new approach to palliation for pulmonary atresia. Br Heart J 67:240–245CrossRef
2.
Zurück zum Zitat Glatz AC, Petit CJ, Goldstein BH et al (2018) Comparison between patent ductus arteriosus stent and modified Blalock-Taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow: insights from the congenital catheterization research collaborative. Circulation 137:589–601CrossRef Glatz AC, Petit CJ, Goldstein BH et al (2018) Comparison between patent ductus arteriosus stent and modified Blalock-Taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow: insights from the congenital catheterization research collaborative. Circulation 137:589–601CrossRef
3.
Zurück zum Zitat Bentham JR, Zava NK, Harrison WJ et al (2018) Duct stenting versus modified Blalock-Taussig shunt in neonates with duct-dependent pulmonary blood flow: associations with clinical outcomes in a multicenter national study. Circulation 137:581–588CrossRef Bentham JR, Zava NK, Harrison WJ et al (2018) Duct stenting versus modified Blalock-Taussig shunt in neonates with duct-dependent pulmonary blood flow: associations with clinical outcomes in a multicenter national study. Circulation 137:581–588CrossRef
4.
Zurück zum Zitat Celebi A, Yucel IK, Bulut MO et al (2017) Stenting of the ductus arteriosus in infants with functionally univentricular heart disease and ductal-dependent pulmonary blood flow: a single-center experience. Catheter Cardiovasc Interv 89:699–708CrossRef Celebi A, Yucel IK, Bulut MO et al (2017) Stenting of the ductus arteriosus in infants with functionally univentricular heart disease and ductal-dependent pulmonary blood flow: a single-center experience. Catheter Cardiovasc Interv 89:699–708CrossRef
5.
Zurück zum Zitat Udink Ten Cate FE, Sreeram N, Hamza H et al (2013) Stenting the arterial duct in neonates and infants with congenital heart disease and duct-dependent pulmonary blood flow: a multicenter experience of an evolving therapy over 18 years. Catheter Cardiovasc Interv 82:E233–E243CrossRef Udink Ten Cate FE, Sreeram N, Hamza H et al (2013) Stenting the arterial duct in neonates and infants with congenital heart disease and duct-dependent pulmonary blood flow: a multicenter experience of an evolving therapy over 18 years. Catheter Cardiovasc Interv 82:E233–E243CrossRef
6.
Zurück zum Zitat Alwi M (2012) Stenting the patent ductus arteriosus in duct-dependent pulmonary circulation: techniques, complications and follow-up issues. Futur Cardiol 8:237–250CrossRef Alwi M (2012) Stenting the patent ductus arteriosus in duct-dependent pulmonary circulation: techniques, complications and follow-up issues. Futur Cardiol 8:237–250CrossRef
7.
Zurück zum Zitat Gibbs JL (2008) Ductal stenting for restricted pulmonary blood flow in neonates: 15 years on but still a very limited place in clinical practice. Heart 94:834–835CrossRef Gibbs JL (2008) Ductal stenting for restricted pulmonary blood flow in neonates: 15 years on but still a very limited place in clinical practice. Heart 94:834–835CrossRef
8.
Zurück zum Zitat Santoro G, Gaio G, Giugno L et al (2015) Ten-years, single-center experience with arterial duct stenting in duct-dependent pulmonary circulation: early results, learning-curve changes, and mid-term outcome. Catheter Cardiovasc Interv 86:249–257CrossRef Santoro G, Gaio G, Giugno L et al (2015) Ten-years, single-center experience with arterial duct stenting in duct-dependent pulmonary circulation: early results, learning-curve changes, and mid-term outcome. Catheter Cardiovasc Interv 86:249–257CrossRef
9.
Zurück zum Zitat Alwi M, Choo KK, Latiff HA et al (2004) Initial results and medium-term follow-up of stent implantation of patent ductus arteriosus in duct-dependent pulmonary circulation. J Am Coll Cardiol 44:438–445CrossRef Alwi M, Choo KK, Latiff HA et al (2004) Initial results and medium-term follow-up of stent implantation of patent ductus arteriosus in duct-dependent pulmonary circulation. J Am Coll Cardiol 44:438–445CrossRef
10.
Zurück zum Zitat Vida VL, Speggiorin S, Maschietto N et al (2010) Cardiac operations after patent ductus arteriosus stenting in duct-dependent pulmonary circulation. Ann Thorac Surg 90:605–609CrossRef Vida VL, Speggiorin S, Maschietto N et al (2010) Cardiac operations after patent ductus arteriosus stenting in duct-dependent pulmonary circulation. Ann Thorac Surg 90:605–609CrossRef
11.
Zurück zum Zitat Jadhav SP, Golriz F, Atweh LA et al (2015) CT angiography of neonates and infants: comparison of radiation dose and image quality of target mode prospectively ECG-gated 320-MDCT and ungated helical 64-MDCT. AJR Am J Roentgenol 204:W184–W191CrossRef Jadhav SP, Golriz F, Atweh LA et al (2015) CT angiography of neonates and infants: comparison of radiation dose and image quality of target mode prospectively ECG-gated 320-MDCT and ungated helical 64-MDCT. AJR Am J Roentgenol 204:W184–W191CrossRef
12.
Zurück zum Zitat Qureshi AM, Goldstein BH, Glatz AC et al (2019) Classification scheme for ductal morphology in cyanotic patients with ductal dependent pulmonary blood flow and association with outcomes of patent ductus arteriosus stenting. Catheter Cardiovasc Interv 93:933–943CrossRef Qureshi AM, Goldstein BH, Glatz AC et al (2019) Classification scheme for ductal morphology in cyanotic patients with ductal dependent pulmonary blood flow and association with outcomes of patent ductus arteriosus stenting. Catheter Cardiovasc Interv 93:933–943CrossRef
13.
Zurück zum Zitat Bauser-Heaton H, Qureshi AM, Goldstein BH et al (2019) Use of novel "flip technique" aids in percutaneous carotid artery approach in neonates. JACC Cardiovasc Interv 12:1630–1631CrossRef Bauser-Heaton H, Qureshi AM, Goldstein BH et al (2019) Use of novel "flip technique" aids in percutaneous carotid artery approach in neonates. JACC Cardiovasc Interv 12:1630–1631CrossRef
Metadaten
Titel
Correlation of ductus arteriosus length and morphology between computed tomographic angiography and catheter angiography and their relation to ductal stent length
verfasst von
Siddharth P. Jadhav
Varun Aggarwal
Prakash M. Masand
Eric Diaz
Wei Zhang
Athar M. Qureshi
Publikationsdatum
13.03.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 6/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-020-04624-1

Weitere Artikel der Ausgabe 6/2020

Pediatric Radiology 6/2020 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.