Purpose

The purpose of this study is to evaluate a new device providing real-time monitoring on radiation exposure during fluoroscopy procedures intending to reduce radiation in an interventional radiology setting.

Materials and Methods

In one interventional suite, a new system providing a real-time radiation dose display and five individual wireless dosimeters were installed. The five dosimeters were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure taking place in that suite. During the first 6-week interval the dose display was off (closed phase) and activated thereafter, for a 6-week learning phase (learning phase) and a 10-week open phase (open phase). During these phases, the staff dose and the individual dose for each procedure were recorded from the wireless dosimeter and correlated with the fluoroscopy time. Further subanalysis for dose exposure included diagnostic versus interventional as well as short (<10 min) versus long (>10 min) procedures.

Results

A total of 252 procedures were performed (n = 88 closed phase, n = 50 learning phase, n = 114 open phase). The overall mean staff dose per fluoroscopic minute was 42.79 versus 19.81 µSv/min (p < 0.05) comparing the closed and open phase. Thereby, anesthesiologists were the only individuals attaining a significant dose reduction during open phase 16.9 versus 8.86 µSv/min (p < 0.05). Furthermore, a significant reduction of total staff dose was observed for short 51 % and interventional procedures 45 % (p < 0.05, for both).

Conclusion

A real-time qualitative display of radiation exposure may reduce team radiation dose. The process may take a few weeks during the learning phase but appears sustained, thereafter.
Literatur
5.
Zurück zum Zitat Jensen K, Zangani L, Martinsen AC, Sandbaek G (2011) Changes in dose-area product, entrance surface dose, and lens dose to the radiologist in a vascular interventional laboratory when an old X-ray system is exchanged with a new system. Cardiovasc Intervent Radiol 34(4):717–722. doi:10.1007/s00270-010-0017-6 CrossRefPubMed Jensen K, Zangani L, Martinsen AC, Sandbaek G (2011) Changes in dose-area product, entrance surface dose, and lens dose to the radiologist in a vascular interventional laboratory when an old X-ray system is exchanged with a new system. Cardiovasc Intervent Radiol 34(4):717–722. doi:10.​1007/​s00270-010-0017-6 CrossRefPubMed
7.
Zurück zum Zitat Miller DL, Vano E, Bartal G, Balter S, Dixon R, Padovani R et al (2010) Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. J Vasc Interv Radiol 21(5):607–615. doi:10.1016/j.jvir.2010.01.007 CrossRefPubMed Miller DL, Vano E, Bartal G, Balter S, Dixon R, Padovani R et al (2010) Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. J Vasc Interv Radiol 21(5):607–615. doi:10.​1016/​j.​jvir.​2010.​01.​007 CrossRefPubMed
10.
Zurück zum Zitat External Dosimetry: Operational Quantities and their Measurement, 11th International Congress of the International Radiation Protection Association (IRPA) Sess. (May, 2004) External Dosimetry: Operational Quantities and their Measurement, 11th International Congress of the International Radiation Protection Association (IRPA) Sess. (May, 2004)
11.
Zurück zum Zitat Williams J (2004) Is there a benefit in promoting the concept of radiation risk? Br J Radiol 77(919):545–546CrossRefPubMed Williams J (2004) Is there a benefit in promoting the concept of radiation risk? Br J Radiol 77(919):545–546CrossRefPubMed
12.
Zurück zum Zitat Racadio J, Nachabe R, Carelsen B, Racadio J, Hilvert N, Johnson N et al (2014) Effect of real-time radiation dose feedback on pediatric interventional radiology staff radiation exposure. J Vascul Interv Radiol 25(1):119–126. doi:10.1016/j.jvir.2013.08.015 CrossRef Racadio J, Nachabe R, Carelsen B, Racadio J, Hilvert N, Johnson N et al (2014) Effect of real-time radiation dose feedback on pediatric interventional radiology staff radiation exposure. J Vascul Interv Radiol 25(1):119–126. doi:10.​1016/​j.​jvir.​2013.​08.​015 CrossRef
Metadaten
Titel
The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting
verfasst von
Frederic Baumann
Barry T. Katzen
Bart Carelsen
Nicolas Diehm
James F. Benenati
Constantino S. Peña
Publikationsdatum
01.10.2015
Verlag
Springer US
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 5/2015
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-015-1075-6

Weitere Artikel der Ausgabe 5/2015

CardioVascular and Interventional Radiology 5/2015 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.