Low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system—First clinical results

verfasst von: Boris Schell, Ralf W. Bauer, Thomas Lehnert, J. Matthias Kerl, Markus Hambek, Angelika May, Thomas J. Vogl, Martin G. Mack

Erschienen in: European Radiology | Ausgabe 1/2011

Einloggen, um Zugang zu erhalten

Abstract

Objective

Computed tomography (CT) of the paranasal sinus is the standard diagnostic tool for a wide range of indications in mostly younger patients. This study aims to assess the image quality of CT of the sinus by using a high-pitch dual-source technique with special regard to the radiation dose.

Methods

Examinations were performed on a second-generation dual-source CT with a pitch factor of 3.0 (dual-source mode). Images were compared with those with a pitch factor of 0.9 on the same system (single-source mode) and with those of 16-slice CT. Image quality was evaluated by four blinded readers using a 5-point scale (1 = poor, 5 = excellent). Comparison of the dose length product (DLP) was used to estimate radiation exposure.

Results

Seventy-three consecutive patients underwent imaging with the proposed CT protocols. The viewers rated the image quality of the dual-source image sets as nearly as good (3.62) as the single-source images on the same device (4.18) and those on 16-slice CT (3.7). DLP was cut to half of the dose [51 mGycm vs. 97.8 mGycm vs. 116.9 mGycm (p < 0.01)].

Conclusions

Using the proposed dual-source mode when examining the paranasal sinus, diagnostic image quality can be achieved while drastically lowering the patient's radiation exposure.
Literatur
1.
Zurück zum Zitat Vogl TJ, Mack MG, Balzer J (2000) Chronic infections of the paranasal sinuses. Radiologe 40:500–506CrossRefPubMed Vogl TJ, Mack MG, Balzer J (2000) Chronic infections of the paranasal sinuses. Radiologe 40:500–506CrossRefPubMed
2.
Zurück zum Zitat Dammann F (2007) Imaging of paranasal sinuses today. Radiologe 47(576):578–583 Dammann F (2007) Imaging of paranasal sinuses today. Radiologe 47(576):578–583
3.
Zurück zum Zitat Hart D, Wall BF (2004) UK population dose from medical X-ray examinations. Eur J Radiol 50:285–291CrossRefPubMed Hart D, Wall BF (2004) UK population dose from medical X-ray examinations. Eur J Radiol 50:285–291CrossRefPubMed
4.
Zurück zum Zitat Brenner DJ, Hall EJ (2007) Computed tomography-an increasing source of radiation exposure. N Engl J Med 357:2277–2284CrossRefPubMed Brenner DJ, Hall EJ (2007) Computed tomography-an increasing source of radiation exposure. N Engl J Med 357:2277–2284CrossRefPubMed
6.
Zurück zum Zitat Petersilka M, Bruder H, Krauss B, Stierstorfer K, Flohr TG (2008) Technical principles of dual source CT. Eur J Radiol 68:362–368CrossRefPubMed Petersilka M, Bruder H, Krauss B, Stierstorfer K, Flohr TG (2008) Technical principles of dual source CT. Eur J Radiol 68:362–368CrossRefPubMed
7.
Zurück zum Zitat Leschka S, Stolzmann P, Desbiolles L, Baumueller S, Goetti R, Schertler T, Scheffel H, Plass A, Falk V, Feuchtner G, Marincek B, Alkadhi H (2009) Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience. Eur Radiol 19:2896–2903CrossRefPubMed Leschka S, Stolzmann P, Desbiolles L, Baumueller S, Goetti R, Schertler T, Scheffel H, Plass A, Falk V, Feuchtner G, Marincek B, Alkadhi H (2009) Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience. Eur Radiol 19:2896–2903CrossRefPubMed
8.
Zurück zum Zitat Sommer WH, Schenzle JC, Becker CR, Nikolaou K, Graser A, Michalski G, Neumaier K, Reiser MF, Johnson TR (2009) Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique. Invest Radiol 45:64–71CrossRef Sommer WH, Schenzle JC, Becker CR, Nikolaou K, Graser A, Michalski G, Neumaier K, Reiser MF, Johnson TR (2009) Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique. Invest Radiol 45:64–71CrossRef
9.
Zurück zum Zitat Achenbach S, Marwan M, Schepis T, Pflederer T, Bruder H, Allmendinger T, Petersilka M, Anders K, Lell M, Kuettner A, Ropers D, Daniel WG, Flohr T (2009) High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr 3:117–121CrossRefPubMed Achenbach S, Marwan M, Schepis T, Pflederer T, Bruder H, Allmendinger T, Petersilka M, Anders K, Lell M, Kuettner A, Ropers D, Daniel WG, Flohr T (2009) High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr 3:117–121CrossRefPubMed
10.
Zurück zum Zitat Simmen D, Schuknecht B (1997) Computerized tomography of paranasal sinuses–a preoperative check list. Laryngorhinootologie 76:8–13CrossRefPubMed Simmen D, Schuknecht B (1997) Computerized tomography of paranasal sinuses–a preoperative check list. Laryngorhinootologie 76:8–13CrossRefPubMed
11.
Zurück zum Zitat Bongartz G, Golding SJ, Jurik AG, Leonardi M, Van Persijn van Meerten E, Geleijns J, Jessen KA, Panzer W, Shrimpton PC, Tosi G (1999) European Guidelines on Quality Criteria for Computed Tomography. Report EUR 16262 Bongartz G, Golding SJ, Jurik AG, Leonardi M, Van Persijn van Meerten E, Geleijns J, Jessen KA, Panzer W, Shrimpton PC, Tosi G (1999) European Guidelines on Quality Criteria for Computed Tomography. Report EUR 16262
12.
Zurück zum Zitat Dammann F, Momino-Traserra E, Remy C, Pereira PL, Baumann I, Koitschev A, Claussen CD (2000) Radiation exposure during spiral-CT of the paranasal sinuses. Rofo 172:232–237PubMed Dammann F, Momino-Traserra E, Remy C, Pereira PL, Baumann I, Koitschev A, Claussen CD (2000) Radiation exposure during spiral-CT of the paranasal sinuses. Rofo 172:232–237PubMed
13.
Zurück zum Zitat Little MP, Wakeford R, Tawn EJ, Bouffler SD, Berrington de Gonzalez A (2009) Risks associated with low doses and low dose rates of ionizing radiation: why linearity may be (almost) the best we can do. Radiology 251:6–12CrossRefPubMed Little MP, Wakeford R, Tawn EJ, Bouffler SD, Berrington de Gonzalez A (2009) Risks associated with low doses and low dose rates of ionizing radiation: why linearity may be (almost) the best we can do. Radiology 251:6–12CrossRefPubMed
14.
Zurück zum Zitat Tubiana M, Feinendegen LE, Yang C, Kaminski JM (2009) The linear no-threshold relationship is inconsistent with radiation biologic and experimental data. Radiology 251:13–22CrossRefPubMed Tubiana M, Feinendegen LE, Yang C, Kaminski JM (2009) The linear no-threshold relationship is inconsistent with radiation biologic and experimental data. Radiology 251:13–22CrossRefPubMed
15.
Zurück zum Zitat Cohen MD (2009) Pediatric CT radiation dose: how low can you go? AJR Am J Roentgenol 192:1292–1303CrossRefPubMed Cohen MD (2009) Pediatric CT radiation dose: how low can you go? AJR Am J Roentgenol 192:1292–1303CrossRefPubMed
16.
Zurück zum Zitat Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de Gonzalez A, Miglioretti DL (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086CrossRefPubMed Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de Gonzalez A, Miglioretti DL (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086CrossRefPubMed
17.
Zurück zum Zitat Martin CJ, Sutton DG, Sharp PF (1999) Balancing patient dose and image quality. Appl Radiat Isot 50:1–19CrossRefPubMed Martin CJ, Sutton DG, Sharp PF (1999) Balancing patient dose and image quality. Appl Radiat Isot 50:1–19CrossRefPubMed
18.
Zurück zum Zitat Wall BF, Hart D (1997) Revised radiation doses for typical X-ray examinations. Report on a recent review of doses to patients from medical X-ray examinations in the UK by NRPB. National Radiological Protection Board. Br J Radiol 70:437–439PubMed Wall BF, Hart D (1997) Revised radiation doses for typical X-ray examinations. Report on a recent review of doses to patients from medical X-ray examinations in the UK by NRPB. National Radiological Protection Board. Br J Radiol 70:437–439PubMed
19.
Zurück zum Zitat Aalokken TM, Hagtvedt T, Dalen I, Kolbenstvedt A (2003) Conventional sinus radiography compared with CT in the diagnosis of acute sinusitis. Dentomaxillofac Radiol 32:60–62CrossRefPubMed Aalokken TM, Hagtvedt T, Dalen I, Kolbenstvedt A (2003) Conventional sinus radiography compared with CT in the diagnosis of acute sinusitis. Dentomaxillofac Radiol 32:60–62CrossRefPubMed
20.
Zurück zum Zitat Klose KC, Elies W, Sondermann U (1991) The accuracy of plain-film radiology in demonstrating the shadows of the pneumatic spaces of the skull–a comparison with computed tomography. Rofo 155:199–206PubMed Klose KC, Elies W, Sondermann U (1991) The accuracy of plain-film radiology in demonstrating the shadows of the pneumatic spaces of the skull–a comparison with computed tomography. Rofo 155:199–206PubMed
21.
Zurück zum Zitat Batra PS, Kanowitz SJ, Citardi MJ (2008) Clinical utility of intraoperative volume computed tomography scanner for endoscopic sinonasal and skull base procedures. Am J Rhinol 22:511–515CrossRefPubMed Batra PS, Kanowitz SJ, Citardi MJ (2008) Clinical utility of intraoperative volume computed tomography scanner for endoscopic sinonasal and skull base procedures. Am J Rhinol 22:511–515CrossRefPubMed
22.
Zurück zum Zitat Yu L, Vrieze TJ, Bruesewitz MR, Kofler JM, DeLone DR, Pallanch JF, Lindell EP, McCollough CH (2010) Dose and image quality evaluation of a dedicated cone-beam CT system for high-contrast neurologic applications. AJR Am J Roentgenol 194:193–201CrossRef Yu L, Vrieze TJ, Bruesewitz MR, Kofler JM, DeLone DR, Pallanch JF, Lindell EP, McCollough CH (2010) Dose and image quality evaluation of a dedicated cone-beam CT system for high-contrast neurologic applications. AJR Am J Roentgenol 194:193–201CrossRef
23.
Zurück zum Zitat Fahrig R, Dixon R, Payne T, Morin RL, Ganguly A, Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med Phys 33:4541–4550CrossRefPubMed Fahrig R, Dixon R, Payne T, Morin RL, Ganguly A, Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med Phys 33:4541–4550CrossRefPubMed
Metadaten
Titel
Low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system—First clinical results
verfasst von
Boris Schell
Ralf W. Bauer
Thomas Lehnert
J. Matthias Kerl
Markus Hambek
Angelika May
Thomas J. Vogl
Martin G. Mack
Publikationsdatum
01.01.2011
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 1/2011
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-1892-6

Weitere Artikel der Ausgabe 1/2011

European Radiology 1/2011 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.