The best of two worlds? Pulmonary vein isolation using a novel radiofrequency ablation catheter incorporating contact force sensing technology and 56-hole porous tip irrigation

verfasst von: Tilman Maurer, Laura Rottner, Hisaki Makimoto, Bruno Reissmann, Christian-H. Heeger, Christine Lemes, Thomas Fink, Johannes Riedl, Francesco Santoro, Peter Wohlmuth, Marius Volkmer, Shibu Mathew, Andreas Metzner, Feifan Ouyang, Karl-Heinz Kuck, Christian Sohns

Erschienen in: Clinical Research in Cardiology | Ausgabe 11/2018

Einloggen, um Zugang zu erhalten

Abstract

Aims

This study aimed to evaluate feasibility and safety as well as 1-year clinical outcome of pulmonary vein isolation (PVI) using a unique radiofrequency ablation catheter (“Thermocool SmartTouch SurroundFlow”; STSF) incorporating both, contact force (CF) sensing technology and enhanced tip irrigation with 56 holes, in one device.

Methods

A total of 110 patients suffering from drug-refractory atrial fibrillation underwent wide area circumferential PVI using either the STSF ablation catheter (75 consecutive patients, study group) or a CF catheter with conventional tip irrigation (“Thermocool SmartTouch”, 35 consecutive patients, control group). For each ablation lesion, a target CF of ≥ 10–39 g and a force time integral (FTI) of > 400 g s was targeted.

Results

Acute PVI was achieved in all patients with target CF obtained in > 85% of ablation points when using either device. Mean procedure time (131.3 ± 33.7 min in the study group vs. 133.0 ± 42.0 min in the control group; p = 0.99), mean fluoroscopy time (14.0 ± 6 vs. 13.5 ± 6.6 min; p = 0.56) and total ablation time were not significantly different (1751.0 ± 394.0 vs. 1604.6 ± 287.8 s; p = 0.2). However, there was a marked reduction in total irrigation fluid delivery by 51.7% (265.52 ± 64.4 vs. 539.6 ± 118.2 ml; p < 0.01). The Kaplan–Meier estimate 12-month arrhythmia–free survival after the index procedure following a 3-month blanking period was 79.9% (95% CI 70.4%, 90.4%) for the study group and 66.7% for the control group (95% CI 50.2%, 88.5%). This finding did not reach statistical significance (p = 0.18). Major complications occurred in 2/75 patients (2.7%; one pericardial tamponade and one transient ischemic attack) in the study group and no patient in the control group (p = 18).

Conclusion

PVI using the STSF catheter is safe and effective and results in beneficial 1-year clinical outcome. The improved tip irrigation leads to a significant reduction in procedural fluid burden.
Literatur
1.
Zurück zum Zitat Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B et al (2016) 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Europace 18(11):1609–1678CrossRef Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B et al (2016) 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Europace 18(11):1609–1678CrossRef
3.
Zurück zum Zitat Makimoto H, Heeger CH, Lin T, Rillig A, Metzner A, Wissner E et al (2015) Comparison of contact force-guided procedure with non-contact force-guided procedure during left atrial mapping and pulmonary vein isolation: impact of contact force on recurrence of atrial fibrillation. Clin Res Cardiol 104(10):861–870CrossRef Makimoto H, Heeger CH, Lin T, Rillig A, Metzner A, Wissner E et al (2015) Comparison of contact force-guided procedure with non-contact force-guided procedure during left atrial mapping and pulmonary vein isolation: impact of contact force on recurrence of atrial fibrillation. Clin Res Cardiol 104(10):861–870CrossRef
4.
Zurück zum Zitat Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T et al (2010) Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation 122(23):2368–2377CrossRef Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T et al (2010) Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation 122(23):2368–2377CrossRef
7.
Zurück zum Zitat Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N et al (2008) Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 1(5):354–362CrossRef Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N et al (2008) Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 1(5):354–362CrossRef
8.
Zurück zum Zitat Makimoto H, Lin T, Rillig A, Metzner A, Wohlmuth P, Arya A et al (2014) In vivo contact force analysis and correlation with tissue impedance during left atrial mapping and catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol 7(1):46–54CrossRef Makimoto H, Lin T, Rillig A, Metzner A, Wohlmuth P, Arya A et al (2014) In vivo contact force analysis and correlation with tissue impedance during left atrial mapping and catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol 7(1):46–54CrossRef
9.
Zurück zum Zitat Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C et al (2012) The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 9(11):1789–1795CrossRef Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C et al (2012) The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 9(11):1789–1795CrossRef
10.
Zurück zum Zitat Stabile G, Solimene F, Calo L, Anselmino M, Castro A, Pratola C et al (2015) Catheter-tissue contact force values do not impact mid-term clinical outcome following pulmonary vein isolation in patients with paroxysmal atrial fibrillation. J Interv Card Electrophysiol 42(1):21–26CrossRef Stabile G, Solimene F, Calo L, Anselmino M, Castro A, Pratola C et al (2015) Catheter-tissue contact force values do not impact mid-term clinical outcome following pulmonary vein isolation in patients with paroxysmal atrial fibrillation. J Interv Card Electrophysiol 42(1):21–26CrossRef
11.
Zurück zum Zitat Reddy V, Pollak S, Lindsay B, McElderry H, Natale A, Kantipudi C et al (2016) Relationship between catheter stability and 12-month success after pulmonary vein isolation—a subanalysis of the SMART-AF Trial. JACC 2(6):700–702 Reddy V, Pollak S, Lindsay B, McElderry H, Natale A, Kantipudi C et al (2016) Relationship between catheter stability and 12-month success after pulmonary vein isolation—a subanalysis of the SMART-AF Trial. JACC 2(6):700–702
12.
Zurück zum Zitat Park CI, Lehrmann H, Keyl C, Weber R, Schurr P, Schiebeling-Romer J et al (2013) Enhanced efficiency of a novel porous tip irrigated RF ablation catheter for pulmonary vein isolation. J Cardiovasc Electrophysiol 24(12):1328–1335CrossRef Park CI, Lehrmann H, Keyl C, Weber R, Schurr P, Schiebeling-Romer J et al (2013) Enhanced efficiency of a novel porous tip irrigated RF ablation catheter for pulmonary vein isolation. J Cardiovasc Electrophysiol 24(12):1328–1335CrossRef
13.
Zurück zum Zitat Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, De Simone A et al (2013) Comparison of ThermoCool(R) Surround Flow catheter versus ThermoCool(R) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol 24(3):269–273CrossRef Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, De Simone A et al (2013) Comparison of ThermoCool(R) Surround Flow catheter versus ThermoCool(R) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol 24(3):269–273CrossRef
14.
Zurück zum Zitat Marrouche NF, Brachmann J, Andresen D, Siebels J, Boersma L, Jordaens L et al (2018) Catheter ablation for atrial fibrillation with heart failure. N Engl J Med 378(5):417–427CrossRef Marrouche NF, Brachmann J, Andresen D, Siebels J, Boersma L, Jordaens L et al (2018) Catheter ablation for atrial fibrillation with heart failure. N Engl J Med 378(5):417–427CrossRef
15.
Zurück zum Zitat Chinitz LA, Melby DP, Marchlinski FE, Delaughter C, Fishel RS, Monir G et al (2017) Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial. Europace https://doi.org/10.1093/europace/eux264 Chinitz LA, Melby DP, Marchlinski FE, Delaughter C, Fishel RS, Monir G et al (2017) Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial. Europace https://​doi.​org/​10.​1093/​europace/​eux264
16.
Zurück zum Zitat Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D et al (2017) Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol 28(7):762–767CrossRef Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D et al (2017) Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol 28(7):762–767CrossRef
17.
Zurück zum Zitat Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT et al (2014) Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol 64(7):647–656CrossRef Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT et al (2014) Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol 64(7):647–656CrossRef
18.
Zurück zum Zitat Das M, Loveday JJ, Wynn GJ, Gomes S, Saeed Y, Bonnett LJ et al (2016) Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values. Europace 19(5):775–783. https://doi.org/10.1093/europace/euw105 CrossRef Das M, Loveday JJ, Wynn GJ, Gomes S, Saeed Y, Bonnett LJ et al (2016) Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values. Europace 19(5):775–783. https://​doi.​org/​10.​1093/​europace/​euw105 CrossRef
19.
Zurück zum Zitat Williams S, Harrison JL, Chubb H, Bloch L, Andersen N, Dam H et al (2015) The effect of contact force in atrial radiofrequency ablation electroanatomical, cardiovascular magnetic resonance, and histological assessment in a chronic porcine model. JACC 1(5):421–431PubMed Williams S, Harrison JL, Chubb H, Bloch L, Andersen N, Dam H et al (2015) The effect of contact force in atrial radiofrequency ablation electroanatomical, cardiovascular magnetic resonance, and histological assessment in a chronic porcine model. JACC 1(5):421–431PubMed
21.
Zurück zum Zitat Oza SR, Hunter TD, Biviano AB, Dandamudi G, Herweg B, Patel AM et al (2014) Acute safety of an open-irrigated ablation catheter with 56-hole porous tip for radiofrequency ablation of paroxysmal atrial fibrillation: analysis from 2 observational registry studies. J Cardiovasc Electrophysiol 25(8):852–858CrossRef Oza SR, Hunter TD, Biviano AB, Dandamudi G, Herweg B, Patel AM et al (2014) Acute safety of an open-irrigated ablation catheter with 56-hole porous tip for radiofrequency ablation of paroxysmal atrial fibrillation: analysis from 2 observational registry studies. J Cardiovasc Electrophysiol 25(8):852–858CrossRef
22.
Zurück zum Zitat Theis C, Rostock T, Mollnau H, Sonnenschein S, Himmrich E, Kampfner D et al (2015) The incidence of audible steam pops is increased and unpredictable with the ThermoCool(R) surround flow catheter during left atrial catheter ablation: a prospective observational study. J Cardiovasc Electrophysiol. https://doi.org/10.1111/jce.12721 CrossRef Theis C, Rostock T, Mollnau H, Sonnenschein S, Himmrich E, Kampfner D et al (2015) The incidence of audible steam pops is increased and unpredictable with the ThermoCool(R) surround flow catheter during left atrial catheter ablation: a prospective observational study. J Cardiovasc Electrophysiol. https://​doi.​org/​10.​1111/​jce.​12721 CrossRef
23.
Zurück zum Zitat Quallich SG, Van Heel M, Iaizzo PA (2015) Optimal contact forces to minimize cardiac perforations before, during, and/or after radiofrequency or cryothermal ablations. Heart Rhythm 12(2):291–296CrossRef Quallich SG, Van Heel M, Iaizzo PA (2015) Optimal contact forces to minimize cardiac perforations before, during, and/or after radiofrequency or cryothermal ablations. Heart Rhythm 12(2):291–296CrossRef
25.
Zurück zum Zitat Cappato R, Calkins H, Chen SA, Davies W, Iesaka Y, Kalman J et al (2010) Updated worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circ Arrhythm Electrophysiol 3(1):32–38CrossRef Cappato R, Calkins H, Chen SA, Davies W, Iesaka Y, Kalman J et al (2010) Updated worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circ Arrhythm Electrophysiol 3(1):32–38CrossRef
Metadaten
Titel
The best of two worlds? Pulmonary vein isolation using a novel radiofrequency ablation catheter incorporating contact force sensing technology and 56-hole porous tip irrigation
verfasst von
Tilman Maurer
Laura Rottner
Hisaki Makimoto
Bruno Reissmann
Christian-H. Heeger
Christine Lemes
Thomas Fink
Johannes Riedl
Francesco Santoro
Peter Wohlmuth
Marius Volkmer
Shibu Mathew
Andreas Metzner
Feifan Ouyang
Karl-Heinz Kuck
Christian Sohns
Publikationsdatum
08.05.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Research in Cardiology / Ausgabe 11/2018
Print ISSN: 1861-0684
Elektronische ISSN: 1861-0692
DOI
https://doi.org/10.1007/s00392-018-1270-y

Weitere Artikel der Ausgabe 11/2018

Clinical Research in Cardiology 11/2018 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.