Analytical performance of a PCR assay for the detection of KRAS mutations (codons 12/13 and 61) in formalin-fixed paraffin-embedded tissue samples of colorectal carcinoma
verfasst von:
Sung Lee, Victoria H. Brophy, Jianli Cao, Margot Velez, Corey Hoeppner, Stephen Soviero, H. Jeffrey Lawrence
KRAS mutation testing is mandatory before prescribing anti-epidermal growth factor monoclonal antibodies in the treatment of advanced colorectal cancer. We describe the performance of a TaqMelt polymerase chain reaction (PCR) assay—the cobas® KRAS Mutation Test—designed to detect 19 mutations in codons 12, 13, and 61. The limit of detection was determined using DNA blends from cell lines, plasmids, and formalin-fixed paraffin-embedded tissue specimens. Assay performance was compared to Sanger sequencing using a panel of 188 specimens. Discordant specimens were subjected to next generation pyrosequencing (454). Assay repeatability was assessed using a panel of six specimens. A >95% correct mutation call rate was obtained in all specimen types with ~5% mutant alleles at DNA inputs of 0.8–6.3 ng per PCR reaction; 100% detection rate was observed at the recommended DNA input of 50 ng. The positive percent agreement with Sanger was 97.5% (79/81) for codons 12/13 and 85.7% (6/7) for codon 61. Negative percent agreement was 94.4% (101/107) for codon 12/13 and 99.4% (180/181) for codon 61. Nine of 10 discordant specimens yielded 454 results consistent with the cobas® results. With repeated testing, the assay showed a correct call rate of 100% (192/192) for all operators, instruments, reagent lots, and days tested. The cobas® test detects KRAS mutations in codons 12, 13, and 61 at a limit of detection of <5%. The PCR assay was more sensitive and specific than Sanger sequencing, and performance was highly reproducible. Test performance was not influenced by various endogenous interfering substances or common gut microbes.
Background
Mutations in the KRAS gene that lead to its constitutive activation have been identified in 24–43% of colorectal cancer (CRC) tumors and are common in other tumor types such as pancreatic, lung, thyroid, and myeloid leukemia [1‐3]. These gain of function mutations lead to epidermal growth factor receptor (EGFR)-independent activation of intracellular signaling pathways, resulting in tumor cell proliferation, protection against apoptosis, increased invasion and metastasis, and activation of tumor-induced angiogenesis [4]. The presence of activating mutations has been shown in randomized clinical trials to play a crucial role in predicting non-responsiveness in patients with advanced CRC who receive the anti-EGFR monoclonal antibodies cetuximab and panitumumab [5‐9]. The majority of activating mutations in CRC tumors occur in codons 12 (~82%) and 13 (~17%) of exon 2 of the KRAS gene; however, mutations in codon 61 (exon 3) have also been described [10]. Codon 61 mutations have been shown to result in constitutive activation of KRAS just as codon 12 and 13 mutations do [11, 12], and limited clinical data indicate that codon 61 mutations may predict non-response to anti-EGFR monoclonal antibody therapy [13‐15].
The high level evidence supporting the clinical utility of KRAS mutation status in predicting response to anti-EGFR monoclonal antibodies has led to specific requirements for KRAS mutation testing by European and US regulatory authorities [16, 17]. Major oncology guidelines, including those of the European Society for Medical Oncology [18] and the American Society of Clinical Oncology [19], have all embraced the recommendation that these antibodies be restricted to patients with KRAS wild-type colorectal cancers.
Anzeige
Bi-directional Sanger sequencing, which theoretically can identify all possible mutations in an exon, has been a common reference method for detecting somatic mutations in tumor specimens. However, Sanger sequencing suffers from limited sensitivity for low-level mutant alleles, particularly in formalin-fixed paraffin-embedded tissue (FFPET) specimens, and has a slow turnaround time [20]. A number of other KRAS mutation detection methods, including allele-specific PCR assays, high-resolution melting analysis, and pyrosequencing, have been developed [21, 22]. These other methods have been reported to have varying levels of sensitivity and specificity for detecting KRAS mutations and there is currently no uniform standard for KRAS mutation testing.
There is a clear need for a well-validated, robust, and reproducible method. The European Union in vitro Diagnostic (IVD) Medical Device Directive 98/79/EC requires that IVDs “achieve the performance…stated by the manufacturer” for a variety of test attributes such as limits of detection, analytical sensitivity and specificity, reproducibility/repeatability, and potential interfering and cross-reacting substances. Aligning with the European Union IVD directives and optimizing the analytical performance of a mutation assay requires thorough validation testing on different specimen types (including the intended use of FFPET specimens) as well as a comprehensive understanding of the impact of specimen attributes, such as the tumor content, the levels of amplifiable DNA, necrosis, and the influence of endogenous and exogenous interfering substances.
We set out to design a robust, accurate molecular diagnostic test to identify patients who have CRC tumors with somatic mutations in codons 12 and 13 of exon 2 or 61 of exon 3 of the KRAS gene and who are therefore unlikely to benefit from therapy with anti-EGFR monoclonal antibodies. The cobas® KRAS Mutation Test (Roche Molecular Systems, Branchburg, NJ, USA), for use with the cobas® 4800 System v2.0 is a TaqMelt™ real-time PCR test intended to detect 19 mutations in codons 12 and 13 of exon 2 and 61 of exon 3 of the KRAS gene using DNA derived from FFPET from human CRC tumors. In this report, we present data on the analytic performance of the cobas® test.
Material and methods
Materials
FFPET specimens from CRC tumors were purchased from US commercial vendors: Discovery Life Sciences, Inc. (Los Osos, CA, USA); BioServe (Beltsville, MD, USA); ProteoGenex (Culver City, CA, USA); CureLine, Inc. (San Francisco, CA, USA); and Indivumed, Inc. (Kensington, MD, USA). The FFPET specimens from vendors were from CRC stages I to IV with a majority from stages II and III. Estimated tumor content ranged from 25% to 80%.
Anzeige
Cobas® KRAS mutation test
The cobas® KRAS Mutation Test kit (cobas® test) is a TaqMelt™-based PCR assay (Fig. 1). The testing process consists of two main steps: manual specimen preparation to obtain genomic DNA from FFPET and PCR amplification and detection of target DNA using complementary primer pairs and two oligonucleotide probes labeled with different fluorescent dyes. The PCR primers are designed to amplify defined sections of exons 2 and 3 of the KRAS gene, and fluorescently labeled oligonucleotide probes detect mutations in codons 12, 13, and 61, which fall within these exons.
×
DNA extraction: DNA is extracted from a single, deparaffinized 5 μm section of FFPET following the standard procedure described in the cobas® DNA Sample Preparation Kit package insert. A deparaffinized specimen is lysed by incubation at an elevated temperature with a protease and chaotropic lysis/binding buffer that releases nucleic acid and protects the released genomic DNA from DNases. Subsequently, isopropanol is added to the lysis mixture that is then centrifuged through a column with a glass fiber filter. Unbound substances, such as salts, proteins, and other cellular impurities, are removed by centrifugation. The adsorbed nucleic acids are washed and then eluted with an aqueous solution. The amount of genomic DNA is spectrometrically determined and adjusted to a fixed concentration of 2 ng/μL.
PCR amplification and detection: The target DNA is amplified and detected on the cobas® 4800 system, which measures the fluorescence generated by specific PCR products, using the amplification and detection reagents provided in the cobas® KRAS Mutation Test kit (Fig. 1). Mutation detection is achieved by melting curve analysis, using an automated result interpretation software package (KRAS Assay Specific Analysis Package v1.0). Based on melting temperatures and peak heights for mutants and wild types, mutation status of each specimen was determined. A mutant control, a negative control, and a calibrator are included in each run to confirm the validity of the run. Amplification, detection, quality control analysis, and result interpretation are automated on the system; raw data and melting curves are not reported. The DNA isolation, amplification/detection, and result reporting can be performed in less than 8 h. The test kit is configured such that users running a minimum of three specimens can achieve the full number of 24 reportable results per kit; as many as 45 specimens can be processed in a single test run.
Sanger sequencing
DNA prepared from FFPET specimens using the same extraction method as described for the cobas® test were sequenced by 2X bi-directional Sanger sequencing by a Clinical Laboratory Improvement Amendments (CLIA)-certified laboratory (SeqWright, Houston, TX) using a validated protocol.
Specimens which gave discordant cobas® test and Sanger test results were retested using a quantitative massively parallel pyrosequencing method (454 GS Titanium, 454 Life Sciences, Branford, CT) by a CLIA-certified laboratory (SeqWright, Houston, TX) using a validated protocol. Additionally, a randomly selected subset of specimens with concordant cobas® test and Sanger results were tested by 454 sequencing to confirm the accuracy of mutation status. The validated limit of detection for the 454 KRAS protocol was 1%.
Limit of detection
The limit of detection (LOD) of the cobas® test was evaluated for one predominant mutation in codon 12 (GAT), codon 13 (GAC), and codon 61 (CAC) of the KRAS gene with ~5% mutation. Testing on 24–72 replicates per panel member at different levels of DNA input was conducted using the following specimen types:
FFPET specimens,
DNA extracted from CRC FFPET mutant and wild-type specimens,
DNA extracted and blended from two CRC mutant and wild-type cell lines for codons 12 and 13, and
DNA from one mutant plasmid for codon 61 blended with wild-type KRAS cell line DNA.
Testing was performed using three lots of the cobas® test kit and one lot of the cobas® DNA Sample Preparation Kit. DNA extracted from each specimen type was then serially diluted to DNA input levels ranging from 50.0 to 0.4 ng per reaction well, in order to simulate varying degrees of DNA degradation that may occur with formalin fixation. 454 sequencing was used to assess percent mutant alleles for DNA extracted from each specimen type. Sensitivity was determined based on a correct mutation call rate of ≥95% and by Probit analysis.
In a separate analysis, the LOD was evaluated for 16 additional KRAS mutation genotypes. Due to the lack of naturally occurring specimens with ~5% mutant alleles, FFPET DNA and plasmid blends at ~5% mutation (as assessed by 454 sequencing) were prepared including:
Five FFPET blends representing five KRAS codon 12 mutations,
Five blends (four plasmid and one FFPET) representing five KRAS codon 13 mutations, and
Six blends (two plasmid and four FFPET) representing six KRAS codon 61 mutations.
Anzeige
Before cobas® testing, the mutation status and mutation percentage of the specimens were verified by 2X bi-directional Sanger sequencing and 454 sequencing, respectively.
Analytic sensitivity and specificity—methods correlation with Sanger sequencing
The analytic performance of the cobas® test was evaluated by comparing two lots of the cobas® KRAS test to 2X bi-directional Sanger Sequencing using 188 vendor-purchased CRC FFPET specimens. Each cobas® test reagent lot was paired with a separate lot of the cobas® DNA Sample Preparation Kit. Specimens were tested in a blinded fashion for the presence of codon 12, 13, and 61 mutations. Positive and negative percent agreements between the cobas® test and Sanger test results were determined. Discrepant analysis was performed by testing specimens with discordant cobas® test and Sanger results with 454 sequencing for resolution. In addition, 50 randomly selected specimens with concordant Sanger and cobas® test results were tested by 454 sequencing in order to provide further verification of the correct call rate.
Repeatability
The repeatability of the cobas® test was evaluated across different reagent lots, days, operators, and instruments with six FFPET CRC specimens, two KRAS wild-type, and four KRAS mutation positive. The materials used in the repeatability testing are:
One cobas® DNA Sample Preparation Kit lot,
Two cobas® KRAS Mutation Test Kit lots,
Two operators, and
Four cobas® 4800 System v.2.0.
Each operator performed one run per reagent lot per day for 4 days for a total of 16 runs with two replicates of each of the six specimens.
Anzeige
Interfering endogenous substances
The effects of endogenous substances, hemoglobin and triglycerides, on the performance of the cobas® test were investigated in 10 CRC FFPET specimens. The specimens were selected for mutation status based on Sanger and/or 454 sequencing. Five specimens were KRAS mutation positive and five were wild type. Four 5-μm sections were extracted for each specimen, one each in the following:
Tissue Lysis Buffer + Proteinase K (TLB/PK) (control),
TLB/PK + saline (solvent control),
TLB/PK + 2 g/L hemoglobin (CLSI-recommended test concentrations), and
TLB/PK + 37 mM triglycerides.
Hemoglobin and triglycerides were added during the lysis step of DNA extraction and were not subjected to deparaffinization. All DNA extracts were tested for interference of mutation detection using the cobas® test.
Interference by common colon-related microorganisms
The effects of colon-related microorganisms on the performance of the cobas® test were investigated in 10 FFPET CRC specimens. The specimens were selected for mutation status based on Sanger and/or 454 sequencing. Five specimens were KRAS mutation positive and five were wild type. Approximately one million colony-forming units of Bacteroides caccae, Prevotella intermedia, and Escherichia coli was added during the lysis step of DNA extraction. Phosphate-buffered saline and blank controls were also tested. One 5-μm section of each specimen was extracted and all DNA extracts were tested for interference of mutation detection using the cobas® test.
Cross-reactivity
The cross-reactivity of the cobas® test with a panel of KRAS silent mutations and KRAS homolog plasmids was assessed. Plasmids containing one each of three silent codon 12 mutations and three silent codon 13 mutations were prepared in a background of wild-type cell line DNA and tested in the presence and absence of 5% KRAS mutant DNA. Additionally, cross-reactivity and interference of the cobas® test were assessed with highly homologous RAS gene sequences present in the human genome (HRAS, NRAS, and pseudogene KRAS1P). Six plasmids (KRAS codon 12/13 pseudogene, codon 61 pseudogene, NRAS exon 2, NRAS exon 3, HRAS exon 2, and HRAS exon 3) were blended with KRAS wild-type DNA (K562 cell line DNA) and tested in the presence and absence of 5% KRAS mutant DNA. Each blend was tested in triplicate.
Anzeige
Necrosis
The impact of tissue necrosis on the performance of the cobas® test was evaluated in duplicate in 20 FFPET CRC specimens with varying percentages of necrotic tissue (5–70%) as assessed by an external pathologist. Mutation status was verified by Sanger sequencing. The FFPET specimens used were:
Ten wild-type specimens with 30–70% necrotic tissue,
Five KRAS codon 12 mutants with 30–45% necrotic tissue,
Four KRAS codon 13 mutants with 15–50% necrotic tissue, and
One KRAS codon 61 mutant with 5% necrotic tissue.
Results
Limit of detection
For the LOD analysis for the three predominant mutations in codons 12, 13, and 61, ≥95% correct mutation call rates were obtained across all specimen types with approximately 5% mutant alleles (as assessed by 454) at a DNA input of 0.8–6.3 ng per reaction well (Table 1). A 100% correct mutation call rate for the remaining 16 mutations in codons 12, 13, and 61 for all specimens was obtained at the recommended DNA input of 50 ng per reaction well (Table 2).
Table 1
Summary of limit of detection (LOD) results for the three most predominant KRAS mutations
KRAS mutation
Specimen type and ID
Mutation (%)
Lowest DNA input (ng/PCR) for a mutation rate ≥95%
Codon 12 (exon 2) GGT>GAT
FFPET specimen
4.3
3.1
FFPET blend #1
4.2
3.1
FFPET blend #2
4.7
3.1
Cell line blend
5.3
0.8
Codon 13 (exon 2) GGC>GAC
FFPET blend #1
5.0
3.1
FFPET blend #2
4.6
1.6
FFPET blend #3
7.2
1.6
Cell line blend
4.9
1.6
Codon 61 (exon 3) CAA>CAC
FFPET blend #1
4.4
3.1
FFPET blend #2
5.5
3.1
FFPET blend #3
3.8
6.3
Plasmid blend
5.8
6.3
FFPET formalin-fixed paraffin-embedded tissue
Table 2
Limit of detection (LOD) results for the remaining 16 KRAS mutations at ~5% at recommended DNA input of 50 ng per reaction well
Correct call rate
Sample
Codon
Mutation
Valid replicates
KRAS mutation test results (codon)
Correct mutation calls
Accuracy (%)
1
12
GGT>AGT
24
12/13
24
100
2
12
GGT>CGT
24
12/13
24
100
3
12
GGT>TGT
24
12/13
24
100
4
12
GGT>GCT
24
12/13
24
100
5
12
GGT>GTT
24
12/13
24
100
6
13
GGC>AGC
24
12/13
24
100
7
13
GGC>CGC
24
12/13
24
100
8
13
GGC>TGC
24
12/13
24
100
9
13
GGC>GCC
24
12/13
24
100
10
13
GGC>GTC
24
12/13
24
100
11
61
CAA>AAA
24
61
24
100
12
61
CAA>GAA
24
61
24
100
13
61
CAA>CTA
24
61
24
100
14
61
CAA>CGA
26
61
26
100
15
61
CAA>CCA
24
61
24
100
16
61
CAA>CAT
24
61
24
100
Methods correlation
The analytic performance of the cobas® test and 2X bi-directional Sanger sequencing was compared with a total of 188 FFPET vendor-purchased specimens using two different cobas® test reagent lots. Concordance between the cobas® test and Sanger sequencing for each lot was 97.5% positive percent agreement (PPA) and 94.4% negative percent agreement (NPA) for codon 12/13 and 85.7% PPA and 99.4% NPA for codon 61 (Table 3, columns A and B). Eight discordant results for codon 12/13 and two for codon 61 were subjected to 454 sequencing. Six specimens were reported mutation-detected codon 12/13 by the cobas® test and mutation not detected by Sanger. 454 results agreed with the cobas® test in five of the six cases. The other two specimens were reported mutation not detected codon 12/13 by the cobas® test and mutation detected by Sanger. 454 agreed with the cobas® test in both cases. Therefore, 454 sequencing results agreed with the cobas® test for seven of the eight discrepant specimens. One unique specimen per lot remained discordant for codon 12/13 mutation status after 454 sequencing. Both discordant specimens displayed low peak heights in the cobas assay, suggesting the presence of very low levels of mutant alleles. After discordant resolution with 454 sequencing, the PPA for codon 12/13 was 100% and the NPA was 99% (Table 3, column C). 454 sequencing results of the two discordant specimens for codon 61 agreed with the cobas® test. After discordant resolution with 454 sequencing, the PPA and the NPA for codon 61 were 100% (Table 3, column D). In addition to the 10 discrepant specimens, 50 specimens with concordant cobas® test and Sanger sequencing results (26 codon 12/13 mutations, 6 codon 61 mutations, and 18 mutation not detected specimens) were subjected to 454 sequencing. The 454 sequencing results for all 50 concordant specimens were concordant with the Sanger sequencing and cobas® KRAS test results.
Table 3
Methods correlation between mutation findings with the cobas® mutation test and Sanger sequencing (columns A and B) and combined Sanger and 454/GS20 sequencing resolution (columns C and D)
A
B
C
D
Codon 12/13
2X bi-directional Sanger sequencing
Codon 61
2X bi-directional Sanger sequencing
Codon 12/13
Sanger sequencing, resolved with 454 sequencing
Codon 61
Sanger sequencing, resolved with 454 sequencing
MD
MND
Totals
MD
MND
Totals
MD
MND
Totals
MD
MND
Totals
cobas® KRAS
MD
79
6
85
cobas® KRAS
MD
6
1
7
cobas® KRAS
MD
84
1
85
cobas® KRAS
MD
7
0
7
MND
2
101
103
MND
1
180
181
MND
0
103
103
MND
0
181
181
Totals
81
107
188
Totals
7
181
188
Totals
84
104
188
Totals
7
181
188
Positive agreement, 97.5% (95% CI, 91.4–99.3)
Positive agreement, 85.7% (95% CI, 48.7–97.4)
Positive agreement, 100% (95% CI, 95.6–100)
Positive agreement, 100% (95% CI, 64.6–100)
Negative agreement, 94.4% (95% CI, 88.3–97.4)
Negative agreement, 99.4% (95% CI, 96.9–99.9)
Negative agreement, 99.0% (95% CI, 94.8–99.8)
Negative agreement, 100% (95% CI, 97.9–100)
Overall agreement, 95.7% (95% CI, 91.8–97.8)
Overall agreement, 98.9% (95% CI, 96.2–99.7)
Overall agreement, 99.5% (95% CI, 97.0–99.9)
Overall agreement, 100% (95% CI, 98.0–100)
MD mutation detected, MND wild type for codons 12/13 or 61
Repeatability
All 16 repeatability runs were valid. The cobas® test had a mutation result accuracy of 100% (192/192 replicates) across all specimens, reagent lots, operators, and instruments combined (Table 4).
Table 4
Repeatability—cobas® KRAS mutation test accuracy by test day, reagent lot, operator, and instrument
Test day
Reagent lot
Run
Operator
Instrument ID
Incorrect results
Valid results
Accuracy (%)
1
1
1
A
1
0
12
100
2
B
2
0
12
100
2
3
A
3
0
12
100
4
B
4
0
12
100
2
1
5
A
3
0
12
100
6
B
2
0
12
100
2
7
A
1
0
12
100
8
B
4
0
12
100
3
1
9
A
1
0
12
100
10
B
2
0
12
100
2
11
A
3
0
12
100
12
B
4
0
12
100
4
1
13
A
3
0
12
100
14
B
2
0
12
100
2
15
A
1
0
12
100
16
B
4
0
12
100
Total
0
192
100
Interfering substances
No interference was observed for any of the 10 specimens tested for potential endogenous interfering substances (hemoglobin and triglycerides) at Clinical and Laboratory Standards Institute recommended test concentrations of 2 g/L and 37 mM, respectively. None of the exogenous colon-related microorganisms tested interfered with the cobas® test. All five of the KRAS wild-type specimens spiked with an exogenous substance gave valid results of “mutation not detected”, while all five KRAS-mutant specimens spiked with an exogenous substance gave valid results of “mutation detected” for the correct codon. Both study controls gave valid and accurate mutation results (data not shown).
Cross-reactivity
The cobas® test did not cross-react with either KRAS silent mutations in codons 12/13 or highly homologous RAS gene sequences. When tested in a wild-type background, all plasmid blends tested with either KRAS silent mutations or KRAS homologs resulted in “mutation not detected”. When tested in a background of 5% KRAS-mutant DNA, all plasmid blends resulted in “mutation detected” for the correct codon (data not shown).
Necrosis
No interference from necrotic tissue was observed when evaluating the performance of the cobas® test. Results for all specimens were concordant with Sanger and 454 sequencing results. Necrosis up to 70% did not affect cobas® test performance (data not shown).
Discussion
KRAS testing is now mandatory for patients with CRC being considered for treatment with anti-EGFR monoclonal antibodies such as cetuximab (Erbitux®, Bristol Myers Squibb and Merck-Serono) and panitumumab (Vectibix®, Amgen). Several methods for KRAS testing currently exist, from commercially available CE-marked test kits to laboratory-developed tests, but there are limited data to support the performance of individual methods over others [21‐23]. With the lack of standardization for KRAS testing, there is a widely recognized need for external quality assessment programs and standards [24, 25]. Indeed, a number of recent studies have highlighted the variability of mutation test results from different clinical laboratories [26, 27]. As results directly affect patient therapy, KRAS tests should be thoroughly validated on clinically relevant specimen types (including FFPET) with optimal analytical performance that encompasses a comprehensive understanding of the impact of specimen attributes, such as the levels of amplifiable DNA; the tumor content of FFPET specimens used for analysis; tissue necrosis; and the influence of endogenous and exogenous interfering substances.
The cobas® test was thoroughly evaluated with 23 unique performance verification studies and in 188 CRC clinical specimens. Here, we have reported results on six of these studies, including methods correlation with Sanger sequencing, limit of detection, repeatability, interfering substances, cross-reactivity, and necrosis. These verification tests were performed in FFPET specimens as well as DNA derived from cell lines and plasmids. Specimen attributes that were evaluated were: tumor cell content, potential effects of endogenous and exogenous interfering substances, cross-reactivity with KRAS silent mutations and KRAS homologs, and proportion of necrotic tissue in the FFPET specimen. These analytic performance attributes have not been published for other KRAS testing methods.
Concordance between the cobas® test and sequencing (Sanger plus 454) was 99.5% for codon 12/13 and 100% for codon 61. The cobas® test had a ≥95% correct mutation call rate across all specimen types with ~5% mutant alleles at a DNA input of 0.8–6.3ng/PCR, and a 100% correct mutation call rate across all specimen types for the recommended DNA input of 50 ng per reaction well. The ability of this assay to detect KRAS mutations at lower DNA inputs reduces the likelihood that mutations will be missed in poor-quality FFPET samples with limited amounts of amplifiable DNA. Because we did not perform a systematic assessment of samples with <5% mutant alleles, the detectability of lower levels of mutation with this assay is uncertain.
The cobas® test had a mutation result accuracy of 100% for all specimens and operators combined in the repeatability study. Neither endogenous interfering factors (hemoglobin or triglycerides) nor the presence of exogenous colon-related microorganisms interfered with the test results. The cobas® test does not cross-react with KRAS silent mutations or KRAS homologs from the RAS gene family. The cobas® test gave 100% correct mutation call in FFPET specimens with up to 70% necrosis and ≥95% correct mutation call rate in specimens with low tumor content at ≤50 ng per reaction well.
In contrast to other KRAS testing kits, such as the therascreen® KRAS test (Qiagen, Manchester, UK), which identifies seven somatic mutations in codons 12 (six mutations) and 13 (one mutation), the cobas® test identifies 19 mutations in codons 12, 13, and 61. Based on data from the Catalogue of Somatic Mutations in Cancer (COSMIC) database [10], the additional 12 mutations detected by the cobas test represent approximately 2% of all KRAS mutations or approximately 1% of all CRC cases. However, recent studies suggest that codon 61 mutations are more prevalent than reflected in the COSMIC data [28, 29]. Thus, more comprehensive KRAS mutation coverage may help to improve the selection of patients for this treatment [15]. The analytical studies presented in this paper demonstrate that the cobas® KRAS Mutation Test is a thoroughly validated, robust, and reproducible companion diagnostic test for the detection of the KRAS mutation in CRC tumors from patients under consideration for treatment with anti-EGFR monoclonal antibodies.
Acknowledgments
We acknowledge Felice Shieh for her contributions to manuscript preparation.
Conflict of interest
All authors are full-time employees of Roche Molecular Systems, Inc., Pleasanton, CA.
Open Access
This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://creativecommons.org/licenses/by-nc/2.0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Analytical performance of a PCR assay for the detection of KRAS mutations (codons 12/13 and 61) in formalin-fixed paraffin-embedded tissue samples of colorectal carcinoma
verfasst von
Sung Lee Victoria H. Brophy Jianli Cao Margot Velez Corey Hoeppner Stephen Soviero H. Jeffrey Lawrence
Um auch an kleinen Gewebeproben zuverlässige und reproduzierbare Ergebnisse zu gewährleisten ist eine strenge Qualitätskontrolle in jedem Schritt des Arbeitsablaufs erforderlich. Eine nicht ordnungsgemäße Prüfung oder Behandlung des …
Die vergleichende experimentelle Pathologie („comparative experimental pathology“) ist ein Fachbereich an der Schnittstelle von Human- und Veterinärmedizin. Sie widmet sich der vergleichenden Erforschung von Gemeinsamkeiten und Unterschieden von …
Gastrointestinale Stromatumoren (GIST) stellen seit über 20 Jahren ein Paradigma für die zielgerichtete Therapie mit Tyrosinkinaseinhibitoren dar. Eine elementare Voraussetzung für eine mögliche neoadjuvante oder adjuvante Behandlung bei …
Aufgrund des erheblichen technologischen Fortschritts in der molekularen und genetischen Diagnostik sowie zunehmender Erkenntnisse über die molekulare Pathogenese von Krankheiten hat in den letzten zwei Jahrzehnten ein grundlegender …