Contemporary frameless intracranial biopsy techniques: Might variation in safety and efficacy be expected?

verfasst von: Iris S. C. Verploegh, Victor Volovici, Iain K. Haitsma, Joost W. Schouten, Clemens M. Dirven, Johan M. Kros, Ruben Dammers

Erschienen in: Acta Neurochirurgica | Ausgabe 11/2015

Abstract

Background

Frameless stereotactic neuronavigation has proven to be a feasible technology to acquire brain biopsies with good accuracy and little morbidity and mortality. New systems are constantly introduced into the neurosurgical armamentarium, although few studies have actually evaluated and compared the diagnostic yield, morbidity, and mortality of various manufacturer’s frameless neuronavigation systems. The present study reports our experience with brain biopsy procedures performed using both the Medtronic Stealth TreonTM Vertek® and BrainLAB® Varioguide frameless stereotactic brain biopsy systems.

Patients and methods

All 247 consecutive biopsies from January 2008 until May 2013 were evaluated retrospectively. One hundred two biopsies each were performed using the Medtronic (2008–2009) and BrainLAB® system (2011–2013), respectively. The year 2010 was considered a transition year, in which 43 biopsies were performed with either system. Patient demographics, perioperative characteristics, and histological diagnosis were reviewed, and a comparison was made between the two brain biopsy systems.

Results

The overall diagnostic yield was 94.6 %, i.e., 11 biopsies were nondiagnostic, 5 (4.9 %) with the Medtronic and 6 (5.9 %) with the BrainLAB® system. No differences besides the operating time (108 vs 120 min) were found between the two biopsy methods. On average, 6.6 tissue samples were taken with either technique. Peri- and postoperative complications were seen in 5.3 % and 12.9 %, consisting of three symptomatic hemorrhages (1.2 %). Biopsy-related mortality occurred in 0.8 % of all biopsies.

Conclusions

Regarding diagnostic yield, complication rate, and biopsy-related mortality, there seems to be no difference between the frameless biopsy technique from Medtronic and BrainLAB®. In contemporary time, the neurosurgeon has many tools to choose from, all with a relatively fast learning curve and ever improving feasibility. Thus, the issue of choice involves not the results, but the familiarity, end-user friendliness, and overall comfort when operating the system.
Hinweise
Iris S. C. Verploegh and Victor Volovici contributed equally to this work.

Comments

Intracranial biopsies are an important part of the workload in neurosurgery. In the retrospective study, Verploegh et al. compare their experience with two of the so-called frameless stereotactic systems. The term “frameless” is a misnomer as both systems require the patient to be fixated with pins in a frame; they are just not classical stereotactic frames. The benefit of the frameless systems is that there is no need for an additional scan once the patient has been fixated in the frame. This allows for an improved workflow and convenience for the patient. The drawback is a slight reduction in accuracy compared to the classical frame-based systems. We have recently published that high workflow efficiency and patient comfort can be obtained with the classical frame-based stereotactic system in combination with intraoperative CT [1].
The important finding by Verploegh et al. is the similar diagnostic yield and safety with their current system (Brainlab Varioguide) compared to their previous system (Medtronic Stealth Treon Vertek). The procedure time was approximately 11 % longer with the Brainlab system, with median procedure times around 2 h. All procedures used general anesthesia.
It is our routine in Copenhagen to perform stereotactic biopsies under local anesthesia in adults. With our concept we have median procedure times (patient arrives-patients leaves OR) of 76 min for awake patients and 89 min for intubated pediatric patients [1].
Jannick Brennum
Copenhagen, Denmark
1. Skjøth-Rasmussen J, Jespersen B, Brennum J (2015). The use of Brainlab iCT for frame-based stereotactic procedures. Acta Neurochir 157(8) 1437–1440.
In this well-written article the authors retrospectively analyzed their experience using two different navigation systems for frameless intracranial biopsy, each on 102 patients. The use of neuronavigation to perform stereotactical frameless biopsies has been disseminated worldwide in the last decades, and different strategies have been developed by diverse companies. Overall, the results are excellent with good accuracy and little mortality. This study has the great value of comparing two systems to match up to the diagnostic yield, morbidity, and mortality.
The results show no differences between the two systems besides the different operating time. The issue is of neurosurgical impact and very interesting for daily practice. The results should be confirmed in prospective studies, including other commercially available products (i.e., Stryker, Fiagon, Micromar, Medtech, etc.) and taking into consideration other parameters such as the subjective end-user satisfaction, comfort, and subjective stress during the procedures.
Alex Alfieri, Francesco Ferri,
Neuruppin, Germany

Introduction

The treatment of various intracranial lesions, despite progress in radiological techniques, largely still depends on the histological diagnosis. When open surgical biopsy or resection is not considered appropriate, brain needle biopsy has proven to be a safe and effective means of acquiring a diagnosis. Diagnostic yield has been shown to be between 83.6 and 100 %, with morbidity and mortality ranging from 0.7 to 16.1 % and 0 to 3.9 %, respectively [4]. Nevertheless, a question remains about the diagnostic accuracy, ranging 73 to 97 %, which might be due to sampling error and a lack of standardized criteria of diagnostic yield [4, 9]. Therefore, many techniques are currently employed to increase the yield as well as accuracy to improve the clinical significance and prognostic value of brain biopsy methods, such as radiological imaging techniques, intraoperative imaging, frozen section analysis, and even robotics [1, 2, 4, 8, 1014, 16, 17].
There is no question that frameless stereotactic neuronavigation has been successfully proven to be a feasible technology to acquire brain biopsies with good accuracy and little mortality [4, 8, 11, 16]. Furthermore, its use fans out into others areas, such as placement of deep brain electrodes for epilepsy surgery or the treatment of Parkinson’s disease [5, 6]. Manufacturers of neuronavigation devices are rapidly introducing new frameless stereotaxy systems into the neurosurgical armamentarium. There are, however, no studies of which we are aware that have actually evaluated and compared the diagnostic yield, morbidity, and mortality of various manufacturer’s frameless neuronavigation systems in stereotactic frameless brain needle biopsies.
The present study reports our experience with brain biopsy procedures performed with both the Medtronic Stealth TreonTM Vertek® and BrainLAB® Varioguide frameless stereotactic brain biopsy systems.

Patients and methods

No IRB approval is needed for retrospective cohorts in academic centers in The Netherlands, so no informed consent was required from any of the patients. Files of all consecutive patients undergoing frameless image-guided stereotactic brain biopsy procedures at the Erasmus Medical Centre from January 2008 until and including May 2013 were reviewed. A total of 247 biopsies were performed in 242 patients. All biopsies performed before January 2010 were performed with the Medtronic Stealth TreonTM Vertek® frameless stereotactic brain biopsy system. In this year the BrainLAB® Varioguide frameless stereotactic brain biopsy system was introduced in our clinic, which would replace the former. In the “transition” period between January and December 2010, all 43 biopsies were performed using one of the two frame-based techniques (19 with Medtronic Stealth TreonTM Vertek® and 23 with Brainlab® Varioguide; for one patient the system was not reported in the operative report). Since January 2011 brain biopsies have been exclusively performed using the BrainLAB® Varioguide neuronavigation system. After the transition period, we had the Kolibri® system at our disposal, but we eventually switched to the Curve® BrainLAB® system, running these two in parallel in late 2012 and early 2013. The biopsy system itself, however, remained unchanged, viz. the Brainlab® Varioguide was used with both.

Patient and preoperative neuroimaging characteristics

Age, gender, and biopsy-related complications (morbidity) and death (mortality) as well as operating time, biopsy method used, number of biopsies, and complications were extracted from case notes and operative reports. The anatomical site of the targeted lesion was obtained from preoperative computed tomography or magnetic resonance imaging scans that were acquired no more than 24 h preoperatively. The eventual diagnosis was obtained from the pathological report, prepared or supervised by JMK, from which the diagnostic yield was determined.
The issue of diagnostic yield is a delicate one, as detailed in several papers [24, 8, 9, 11]. Tumor biopsy is a process involving two potential sampling error points, the operation itself, depending on which area of the tumor the samples were taken from, and the pathological investigation, depending on what microscopic area of the acquired samples will be investigated. Therefore, we considered three possible diagnostic options, a full diagnosis, a partial diagnosis (pathologic tissue but no WHO classification possible), or no diagnosis at all. Because in the cases where the diagnosis was only partial a treatment could be initiated based on other characteristics of the specific patient (radiological aspect, speed of progression, oncological history), all biopsies where a treatment could be initiated and tumor tissue was obtained were considered diagnostic.
As such, a stereotactic biopsy serves its purpose if a specific patient can get the best specific treatment given all available data, and not only if it delivers a perfectly classifiable result, this sometimes being not possible because of shortcomings in the WHO classification.
The diagnostic yield of the biopsies was independently reviewed by two authors (RD and VV), and the results were discussed and validated to a common consensus.

Operative technique

The techniques of frameless image-guided stereotactic biopsies have been described previously for the Stealth TreonTM Vertek® system (Medtronic Inc., Minneapolis, MN, USA) [3] as well as the BrainLAB® Varioguide system (BrainLAB, Feldkirchen, Germany) [7, 19]. All biopsies were performed using the frameless stereotaxy protocol under general anesthesia and head fixation in a three-point Mayfield clamp. Intraoperative image guidance and the surgical plan were obtained using the Stealth Treon™ Vertek® system in 102 patients from January 2008 until December 2009 and with the BrainLAB® Varioguide system in 102 patients from January 2011 until May 2013, respectively. In the transition year 2010, 19 and 23 biopsies were performed with the Stealth TreonTM Vertek® and BraiLAB® Varioguide system, respectively. For one patient, the biopsy system had not been reported in the operative report. As to the surgeon’s preference either fiducial markers and/or surface-merging techniques were used to enable operating room neuronavigational registration. The surgery was performed either by neurosurgeons or neurosurgical trainees under supervision of neurosurgeons. The trainees performing the surgery were split into two categories, junior residents, PGY (postgraduate years) 1–3, and senior residents, PGY 4–6. Tissue samples approximately 8 mm long and 1 mm thick were thus obtained.
In general, four biopsies were obtained at the preoperatively suggested target, as well as two to four more biopsies at a site proximal to the target on the same biopsy trajectory. The macroscopic appearance of the acquired tissue sample was assessed as described previously and, when appropriate, prompted to intraoperative freeze sectioning (10/247 = 4.0 %) [4]. If diagnostic tissue was not obtained, all parameters and measurements were rechecked to ensure accurate localization, after which a repeat biopsy was performed via the same or a new trajectory if needed.

Statistical analysis

Database collection and statistical analysis were performed with SPSS 15.0 for Windows (SPSS Inc., Chicago, IL, USA). For comparison between the two frameless stereotaxy biopsy systems, continuous data were compared using one-way analysis of variance (ANOVA) testing for parametric data. For nonparametric data the Kruskal-Wallis one-way ANOVA test was used. These are presented as mean values ± standard deviations. For multiple comparisons the Bonferroni post-hoc correction was used. Proportions were compared with chi-square and Fisher’s exact test where appropriate and presented as percentages. P-values < 0.05 were considered statistically significant.

Results

Table 1 summarizes the patient and perioperative characteristics with respect to the frameless stereotactic biopsy system used, i.e., Medtronic TreonTM Vertek® vs. BrainLAB® Varioguide vs. the 2010 transition period. A higher proportion of males and a higher mean age was observed in the BrainLAB® group compared to the Medtronic TreonTM group. The number of biopsies taken and perioperative complications were equal with both biopsy systems. Remarkably, the total operating time, including anesthetic preparation, was significantly longer using the BrainLAB® Varioguide.
Table 1
Patient and perioperative characteristics related to biopsy method
 
Stealth TreonTM
(n = 102)
Transition period
(n = 43)
BrainLAB®
(n = 102)
Male (%)
49.0*
62.8
65.7
Age (years ± SD)
54.0 ± 19.7*
56.3 ± 20.0
60.1 ± 15.3
Operating time (min ± SD)
108 ± 36*
121 ± 26
120 ± 28
Number of biopsies ± SD
7.0 ± 2.9
6.6 ± 2.9
6.4 ± 2.8
Complication (%)
4.9
4.7
5.9
 Hemorrhage
2.9
2.3
4.9
 Technical failure
2.0
2.3
1.0
Shown are percentages and mean values ± standard deviations (SD). *p < 0.05
The distribution of the different tumor sites related to the biopsy system used is shown in Table 2. There were no statistically relevant differences between the system used for frameless brain biopsy and the various tumor locations.
Table 2
Preoperative anatomical localization of the intended lesion chosen for biopsy
 
Stealth TreonTM(n = 102)
Transition period (n = 43)
BrainLAB® (n = 102)
Frontal
23
10
21
Temporal
11
5
16
Parietal
21
3
11
Occipital
5
1
4
Frontotemporal*
2
1
3
Frontoparietal
5
3
7
Parietooccipital
7
0
9
Parietotemporal
7
3
4
Temporooccipital
0
0
2
Thalamus/basal ganglia*
16
12
18
Pons/medulla oblongata cerebellopontine angle
1
3
2
Cerebellum
0
0
2
Supra-/intrasellar**
3
2
2
Third ventricle
0
0
1

Length of operating time

Over the years that the BrainLAB® system was used (2010–2013) there was a gradual but nonsignificant decrease in operating time from a median of 130 min to 115 min in 2010 and 2013, respectively. Nevertheless, total operating time remained significantly longer for the BrainLAB® system. Correction for age, tumor site, or surgeon experience level (i.e., PGY 1–3, PGY 4–6, or neurosurgeon) did not change this finding. Therefore, it seems that a biopsy procedure with the BrainLAB® Varioguide on average takes some 10 min longer than a biopsy with the Medtronic TreonTM Vertek®.

Complications and mortality

Postoperative complications were encountered in 32 procedures (12.9 %), summarized in Table 3. Of these, three symptomatic hemorrhages (1.2 %) were diagnosed via postoperative computed tomography scanning. Furthermore, symptomatic postoperative edema was present in 6 (2.4 %) patients and early focal neurologic deterioration after biopsy was seen in 15 (6.1 %), of which 9 were observed in the BrainLAB® Varioguide group (8.8 %, p > 0.05, compared to the Medtronic TreonTM group). The incidence of postoperative symptomatic hemorrhage between the two techniques was equal.
Table 3
Postoperative complications
(%)
Stealth TreonTM(n = 102)
Transition period (n = 43)
BrainLAB® (n = 102)
None
91.2
83.7
84.3
Hemorrhage
1.0
0
2.0
Edema formation
3.9
2.3
1.0
Focal neurology
2.9
7.0
8.8
Others*
1.0
4.7
3.9
Biopsy-related death
0
0
2.0
Data are presented as percentages. There are no differences between the biopsy procedures. *Others include psychosis, hyperglycemia, epilepsy, hypertension, and diabetes insipidus
Biopsy-related death was defined as death occurring within 30 days as a result of symptomatic postoperative hemorrhage or edema formation and was observed in two patients, i.e., 0.8 %. There were no differences between the two biopsy systems.

Diagnostic yield

The overall diagnostic yield was 94.6 %, i.e., a histological diagnosis was made in 193 of 204 patients or in 11 cases (5.4 %) the biopsy was nondiagnostic. This was not statistically different between the Medtronic TreonTM (4.9 %) and BrainLAB® (5.9 %) stereotactic biopsy techniques. When including the transition period patients, the overall diagnostic yield was 93.5 % (231/247). A nondiagnostic biopsy was observed in 11.6 % in the transition period (p > 0.05). The lesion locations of nondiagnostic biopsies were frontal (n = 2), frontotemporal (n = 1), parietooccipital (n = 2), and in the third ventricle (n = 1) for BrainLAB®; parietal (n = 2), occipital (n = 1), parietooccipital (n = 1), and temporal (n = 1) for Medtronic TreonTM; frontoparietal (n = 1), temporal (n = 3), and thalamus (n = 1) for the transition period.
When a nondiagnostic biopsy was encountered, follow-up surgery (either stereotactic or open biopsy or tumor decompression) was recommended to nine patients, of whom two underwent three more stereotactic biopsies before a diagnosis could be made (lymphoma, n = 3; glioblastoma, n = 3; metastasis, n = 2; ependymoma, n = 1). These nondiagnostic biopsies were included in the analysis. Of the remaining four patients, two were successfully treated for radionecrosis and one for intracerebral abscess; the fourth was concluded to have an infarction after meticulous analysis for a demyelinating disease.
On average, 6.6 (range: 1 – 20) tissue samples were taken during each procedure with either frameless stereotaxy technique (Table 1). Table 4 shows the histological diagnoses that were made on the tissue samples taken by either system. In ten cases, seven in Medtronic TreonTM (6.9 %) and three in BrainLAB® (2.9 %; p < 0.05), an intraoperative frozen-section diagnosis was requested. All of these included diagnostic biopsies.
Table 4
Histological diagnoses made on tissue samples acquired by stereotactic biopsy
 
Stealth TreonTM(n = 102)
Transition period (n = 43)
BrainLAB® (n = 102)
Nondiagnostic
5
5
6
Malignant glioma
59
21
42
Low-grade glioma
9
5
15
Lymphoma
11
9
20
Metastasis
2
1
7
Infection/vasculitis/MS/abscess
12
1
9
Craniopharyngioma cysts*
2
1
1
Medulloblastoma
0
0
2
Others
2
0
0
MS = multiple sclerosis; others include malignant peripheral nerve sheath tumor and hematoma

Discussion

The main purpose of stereotactical biopsy systems, frameless or otherwise, is to produce a safe, fast, reliable and easy diagnostic biopsy. For this purpose, many different systems are available to the neurosurgeon, each with its distinctive set of qualities. We present a retrospective comparison between the Medtronic TreonTM Vertek® and the BrainLAB® Varioguide systems. Our diagnosis yield overall was 94.6 %. Morbidity and mortality profiles were also comparable with reports in the literature [1, 2, 710, 12, 16].
There was no difference in diagnostic yield between the two systems, which we defined as the likelihood of a biopsy producing pathological tissue to be analyzed, regardless of whether it could be included in the World Health Organization WHO/Central Nervous System CNS classifications [9], as there are still pathological entities that cannot be classified according to this system.
A descending trend can be noticed in the number of times the team asked for a frozen section histology examination to be performed during the operation, possibly reflecting the increase in experience and confidence using the frameless stereotactic systems. However, the operating time was significantly longer using the BrainLAB® Varioguide, in part probably pointing toward the learning curve, not pertaining to the act of biopsying, but to the system itself, as shown by a gradual decrease of 12.5 % in median operating time over the years. Nevertheless, in our cohort the total operating time remained approximately 11 %, or 10 min, longer when using the BrainLAB® system, even after correction for some well-established confounders.
Patients undergoing biopsy with the BrainLAB® Varioguide show significantly more frequent focal neurological signs postoperatively, but this was not correlated with increased hemorrhage or perilesional edema. Only speculations are possible as to why this has occurred, statistical evidence not pointing to any reason in particular. Some reports [18] suggest that the size of the needle correlates with the risk of hemorrhage, but this could not be established in our series (outer needle diameter 2.2 and 1.8 mm and cutting window length 7 and 10 mm for the Stealth TreonTM Vertek® and Brainlab® Varioguide, respectively).
Two deaths occurred, one concerning an 81-year-old male with a parieto-occipital glioblastoma who suffered a basilar artery thrombosis 8 days after the operation. The second concerned a 53-year-old male with a multiloculated mixed anaplastic oligo-astrocytoma expanding the better part of the right cerebral hemisphere who suffered a large, deep intracerebral hematoma and subsequently expired 9 days after the operation. These cases are thus not attributable to the stereotactic systems themselves.
Using both systems we were able to reach a diagnostic yield that was higher even than those usually reported in the literature [2, 7, 9, 10, 15]. Counterintuitively, the intraoperative MRI-enhanced biopsies, which actively provide feedback as to the changes the lesion to be biopsied is undergoing, yielded the worst diagnostic yield [9].
We can thus conclude that both systems are effective, safe, and reliable tools, which when used properly have an excellent diagnostic yield, above 95 %. Despite minor differences, these tools are both part of a well-crafted diagnostic armamentarium available to contemporary neurosurgeons, which should be put to work in order to yield quick and precise diagnostic results. Future prospective studies are needed to compare the effectiveness of these and other available products, including parameters such as end-user satisfaction, subjective intraoperative stress, and overall comfort, among others.

Disclosure of competing interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements) or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Comments

Intracranial biopsies are an important part of the workload in neurosurgery. In the retrospective study, Verploegh et al. compare their experience with two of the so-called frameless stereotactic systems. The term “frameless” is a misnomer as both systems require the patient to be fixated with pins in a frame; they are just not classical stereotactic frames. The benefit of the frameless systems is that there is no need for an additional scan once the patient has been fixated in the frame. This allows for an improved workflow and convenience for the patient. The drawback is a slight reduction in accuracy compared to the classical frame-based systems. We have recently published that high workflow efficiency and patient comfort can be obtained with the classical frame-based stereotactic system in combination with intraoperative CT [1].
The important finding by Verploegh et al. is the similar diagnostic yield and safety with their current system (Brainlab Varioguide) compared to their previous system (Medtronic Stealth Treon Vertek). The procedure time was approximately 11 % longer with the Brainlab system, with median procedure times around 2 h. All procedures used general anesthesia.
It is our routine in Copenhagen to perform stereotactic biopsies under local anesthesia in adults. With our concept we have median procedure times (patient arrives-patients leaves OR) of 76 min for awake patients and 89 min for intubated pediatric patients [1].
Jannick Brennum
Copenhagen, Denmark
1. Skjøth-Rasmussen J, Jespersen B, Brennum J (2015). The use of Brainlab iCT for frame-based stereotactic procedures. Acta Neurochir 157(8) 1437–1440.
In this well-written article the authors retrospectively analyzed their experience using two different navigation systems for frameless intracranial biopsy, each on 102 patients. The use of neuronavigation to perform stereotactical frameless biopsies has been disseminated worldwide in the last decades, and different strategies have been developed by diverse companies. Overall, the results are excellent with good accuracy and little mortality. This study has the great value of comparing two systems to match up to the diagnostic yield, morbidity, and mortality.
The results show no differences between the two systems besides the different operating time. The issue is of neurosurgical impact and very interesting for daily practice. The results should be confirmed in prospective studies, including other commercially available products (i.e., Stryker, Fiagon, Micromar, Medtech, etc.) and taking into consideration other parameters such as the subjective end-user satisfaction, comfort, and subjective stress during the procedures.
Alex Alfieri, Francesco Ferri,
Neuruppin, Germany

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Bekelis K, Radwan TA, Desai A, Roberts DW (2012) Frameless robotically targeted stereotactic brain biopsy: feasibility, diagnostic yield, and safety. J Neurosurg 116(5):1002–1006CrossRefPubMed Bekelis K, Radwan TA, Desai A, Roberts DW (2012) Frameless robotically targeted stereotactic brain biopsy: feasibility, diagnostic yield, and safety. J Neurosurg 116(5):1002–1006CrossRefPubMed
2.
Zurück zum Zitat Burkhardt JK, Neidert MC, Woernle CM, Bozinov O, Bernays RL (2013) Intraoperative low-field MR-guided frameless stereotactic biopsy for intracerebral lesions. Acta Neurochir (Wien) 155(4):721–726CrossRef Burkhardt JK, Neidert MC, Woernle CM, Bozinov O, Bernays RL (2013) Intraoperative low-field MR-guided frameless stereotactic biopsy for intracerebral lesions. Acta Neurochir (Wien) 155(4):721–726CrossRef
3.
Zurück zum Zitat Dammers R, Haitsma IK, Schouten JW, Kros JM, Avezaat CJ, Vincent AJ (2008) Safety and efficacy of frameless and frame-based intracranial biopsy techniques. Acta Neurochir (Wien) 150(1):23–29CrossRef Dammers R, Haitsma IK, Schouten JW, Kros JM, Avezaat CJ, Vincent AJ (2008) Safety and efficacy of frameless and frame-based intracranial biopsy techniques. Acta Neurochir (Wien) 150(1):23–29CrossRef
4.
Zurück zum Zitat Dammers R, Schouten JW, Haitsma IK, Vincent AJ, Kros JM, Dirven CM (2010) Towards improving the safety and diagnostic yield of stereotactic biopsy in a single centre. Acta Neurochir (Wien) 152(11):1915–1921CrossRef Dammers R, Schouten JW, Haitsma IK, Vincent AJ, Kros JM, Dirven CM (2010) Towards improving the safety and diagnostic yield of stereotactic biopsy in a single centre. Acta Neurochir (Wien) 152(11):1915–1921CrossRef
6.
Zurück zum Zitat Fridley J, Thomas JG, Navarro JC, Yoshor D (2012) Brain stimulation for the treatment of epilepsy. Neurosurg Focus 32(3):E13CrossRefPubMed Fridley J, Thomas JG, Navarro JC, Yoshor D (2012) Brain stimulation for the treatment of epilepsy. Neurosurg Focus 32(3):E13CrossRefPubMed
7.
Zurück zum Zitat Gempt J, Buchmann N, Ryang YM, Krieg S, Kreutzer J, Meyer B, Ringel F (2012) Frameless image-guided stereotaxy with real-time visual feedback for brain biopsy. Acta Neurochir (Wien) 154(9):1663–1667CrossRef Gempt J, Buchmann N, Ryang YM, Krieg S, Kreutzer J, Meyer B, Ringel F (2012) Frameless image-guided stereotaxy with real-time visual feedback for brain biopsy. Acta Neurochir (Wien) 154(9):1663–1667CrossRef
8.
Zurück zum Zitat Harrisson SE, Shooman D, Grundy PL (2012) A prospective study of the safety and efficacy of frameless, pinless electromagnetic image-guided biopsy of cerebral lesions. Neurosurgery 70(1 Suppl Operative):29–33, discussion 33PubMed Harrisson SE, Shooman D, Grundy PL (2012) A prospective study of the safety and efficacy of frameless, pinless electromagnetic image-guided biopsy of cerebral lesions. Neurosurgery 70(1 Suppl Operative):29–33, discussion 33PubMed
9.
Zurück zum Zitat Khatab S, Spliet W, Woerdeman PA (2014) Frameless image-guided stereotactic brain biopsies: emphasis on diagnostic yield. Acta Neurochir (Wien) 156(8):1441–1450CrossRef Khatab S, Spliet W, Woerdeman PA (2014) Frameless image-guided stereotactic brain biopsies: emphasis on diagnostic yield. Acta Neurochir (Wien) 156(8):1441–1450CrossRef
10.
Zurück zum Zitat Lefranc M, Capel C, Pruvot-Occean AS, Fichten A, Desenclos C, Toussaint P, Le Gars D, Peltier J (2015) Frameless robotic stereotactic biopsies: a consecutive series of 100 cases. J Neurosurg 122(2):342–352CrossRefPubMed Lefranc M, Capel C, Pruvot-Occean AS, Fichten A, Desenclos C, Toussaint P, Le Gars D, Peltier J (2015) Frameless robotic stereotactic biopsies: a consecutive series of 100 cases. J Neurosurg 122(2):342–352CrossRefPubMed
11.
Zurück zum Zitat Livermore LJ, Ma R, Bojanic S, Pereira EA (2014) Yield and complications of frame-based and frameless stereotactic brain biopsy--the value of intra-operative histological analysis. Br J Neurosurg 28(5):637–644CrossRefPubMed Livermore LJ, Ma R, Bojanic S, Pereira EA (2014) Yield and complications of frame-based and frameless stereotactic brain biopsy--the value of intra-operative histological analysis. Br J Neurosurg 28(5):637–644CrossRefPubMed
12.
Zurück zum Zitat Lu Y, Yeung C, Radmanesh A, Wiemann R, Black PM, Golby AJ (2014) Comparative effectiveness of frame-based, frameless, and intraoperative magnetic resonance imaging-guided brain biopsy techniques. World Neurosurg 83(3):261-268 Lu Y, Yeung C, Radmanesh A, Wiemann R, Black PM, Golby AJ (2014) Comparative effectiveness of frame-based, frameless, and intraoperative magnetic resonance imaging-guided brain biopsy techniques. World Neurosurg 83(3):261-268
13.
Zurück zum Zitat Misch M, Guggemos A, Driever PH, Koch A, Grosse F, Steffen IG, Plotkin M, Thomale UW (2015) (18)F-FET-PET guided surgical biopsy and resection in children and adolescence with brain tumors. Childs Nerv Syst 31(2):261–267CrossRefPubMed Misch M, Guggemos A, Driever PH, Koch A, Grosse F, Steffen IG, Plotkin M, Thomale UW (2015) (18)F-FET-PET guided surgical biopsy and resection in children and adolescence with brain tumors. Childs Nerv Syst 31(2):261–267CrossRefPubMed
14.
Zurück zum Zitat Pirotte BJ, Lubansu A, Massager N, Wikler D, Goldman S, Levivier M (2007) Results of positron emission tomography guidance and reassessment of the utility of and indications for stereotactic biopsy in children with infiltrative brainstem tumors. J Neurosurg 107(5 Suppl):392–399PubMed Pirotte BJ, Lubansu A, Massager N, Wikler D, Goldman S, Levivier M (2007) Results of positron emission tomography guidance and reassessment of the utility of and indications for stereotactic biopsy in children with infiltrative brainstem tumors. J Neurosurg 107(5 Suppl):392–399PubMed
15.
Zurück zum Zitat Ringel F, Ingerl D, Ott S, Meyer B (2009) VarioGuide: a new frameless image-guided stereotactic system—accuracy study and clinical assessment. Neurosurgery 64(5 Suppl 2):365–371, discussion 371–3PubMed Ringel F, Ingerl D, Ott S, Meyer B (2009) VarioGuide: a new frameless image-guided stereotactic system—accuracy study and clinical assessment. Neurosurgery 64(5 Suppl 2):365–371, discussion 371–3PubMed
16.
Zurück zum Zitat Shooman D, Belli A, Grundy PL (2010) Image-guided frameless stereotactic biopsy without intraoperative neuropathological examination. J Neurosurg 113(2):170–178CrossRefPubMed Shooman D, Belli A, Grundy PL (2010) Image-guided frameless stereotactic biopsy without intraoperative neuropathological examination. J Neurosurg 113(2):170–178CrossRefPubMed
17.
Zurück zum Zitat Widhalm G, Minchev G, Woehrer A, Preusser M, Kiesel B, Furtner J, Mert A, Di Ieva A, Tomanek B, Prayer D, Marosi C, Hainfellner JA, Knosp E, Wolfsberger S (2012) Strong 5-aminolevulinic acid-induced fluorescence is a novel intraoperative marker for representative tissue samples in stereotactic brain tumor biopsies. Neurosurg Rev 35(3):381–391, discussion 391CrossRefPubMed Widhalm G, Minchev G, Woehrer A, Preusser M, Kiesel B, Furtner J, Mert A, Di Ieva A, Tomanek B, Prayer D, Marosi C, Hainfellner JA, Knosp E, Wolfsberger S (2012) Strong 5-aminolevulinic acid-induced fluorescence is a novel intraoperative marker for representative tissue samples in stereotactic brain tumor biopsies. Neurosurg Rev 35(3):381–391, discussion 391CrossRefPubMed
18.
Zurück zum Zitat Yuen J, Zhu CX, Chan DT, Ng RY, Nia W, Poon WS, Ng HK, Mok VC, Wong LK, Cheung TC, Siu DY (2014) A sequential comparison on the risk of haemorrhage with different sizes of biopsy needles for stereotactic brain biopsy. Stereotact Funct Neurosurg 92(3):160–169PubMed Yuen J, Zhu CX, Chan DT, Ng RY, Nia W, Poon WS, Ng HK, Mok VC, Wong LK, Cheung TC, Siu DY (2014) A sequential comparison on the risk of haemorrhage with different sizes of biopsy needles for stereotactic brain biopsy. Stereotact Funct Neurosurg 92(3):160–169PubMed
19.
Zurück zum Zitat Zhang JS, Chen XL, Li FY, Li JJ, Zheng G, Zhang T, Hu S, Xu BN (2012) Implementation of VarioGuide in stereotactic brain biopsy: a preliminary experience. Zhonghua Yi Xue Za Zhi 92(21):1468–1471PubMed Zhang JS, Chen XL, Li FY, Li JJ, Zheng G, Zhang T, Hu S, Xu BN (2012) Implementation of VarioGuide in stereotactic brain biopsy: a preliminary experience. Zhonghua Yi Xue Za Zhi 92(21):1468–1471PubMed
Metadaten
Titel
Contemporary frameless intracranial biopsy techniques: Might variation in safety and efficacy be expected?
verfasst von
Iris S. C. Verploegh
Victor Volovici
Iain K. Haitsma
Joost W. Schouten
Clemens M. Dirven
Johan M. Kros
Ruben Dammers
Publikationsdatum
01.11.2015
Verlag
Springer Vienna
Erschienen in
Acta Neurochirurgica / Ausgabe 11/2015
Print ISSN: 0001-6268
Elektronische ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-015-2543-0

Weitere Artikel der Ausgabe 11/2015

Acta Neurochirurgica 11/2015 Zur Ausgabe

Letter to the editor - Neurosurgical Techniques

Poor man-rich man fluorescence. Is this really the problem?

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.