Two-dimensional XD-GRASP provides better image quality than conventional 2D cardiac cine MRI for patients who cannot suspend respiration

verfasst von: Eve Piekarski, Teodora Chitiboi, Rebecca Ramb, Larry A. Latson Jr, Puneet Bhatla, Li Feng, Leon Axel

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Residual respiratory motion degrades image quality in conventional cardiac cine MRI (CCMRI). We evaluated whether a free-breathing (FB) radial imaging CCMRI sequence with compressed sensing reconstruction [extradimensional (e.g. cardiac and respiratory phases) golden-angle radial sparse parallel, or XD-GRASP] could provide better image quality than a conventional Cartesian breath-held (BH) sequence in an unselected population of patients undergoing clinical CCMRI.

Materials and methods

One hundred one patients who underwent BH and FB imaging in a midventricular short-axis plane at a matching location were included. Visual and quantitative image analysis was performed by two blinded experienced readers, using a five-point qualitative scale to score overall image quality and visual signal-to-noise ratio (SNR) grade, with measures of noise and sharpness. End-diastolic and end-systolic left ventricular areas were also measured and compared for both BH and FB images.

Results

Image quality was generally better with the BH cines (overall quality grade for BH vs FB images 4 vs 2.9, p < 0.001; noise 0.06 vs 0.08 p < 0.001; SNR grade 4.1 vs 3, p < 0.001), except for sharpness (p = 0.48). There were no significant differences between BH and FB images regarding end-diastolic or end-systolic areas (p = 0.35 and p = 0.12). Eighteen of the 101 patients had poor BH image quality (grade 1 or 2). In this subgroup, the quality of the FB images was better (p = 0.0032), as was the SNR grade (p = 0.003), but there were no significant differences regarding noise and sharpness (p = 0.45 and p = 0.47).

Conclusion

Although FB XD-GRASP CCMRI was visually inferior to conventional BH CCMRI in general, it provided improved image quality in the subgroup of patients with respiratory-motion-induced artifacts on BH images.
Literatur
1.
Zurück zum Zitat Pennell DJ, Sechtem UP, Higgins CB et al (2004) Clinical indications for cardiovascular magnetic resonance (CMR): consensus panel report. Eur Heart J 25(21):1940–1965CrossRefPubMed Pennell DJ, Sechtem UP, Higgins CB et al (2004) Clinical indications for cardiovascular magnetic resonance (CMR): consensus panel report. Eur Heart J 25(21):1940–1965CrossRefPubMed
2.
Zurück zum Zitat Hundley WG, Bluemke DA, Finn JP et al (2010) ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents. J Am Coll Cardiol 55(23):2614–2662CrossRefPubMed Hundley WG, Bluemke DA, Finn JP et al (2010) ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents. J Am Coll Cardiol 55(23):2614–2662CrossRefPubMed
4.
Zurück zum Zitat Saloner D, Liu J, Haraldsson H (2015) MR physics in practice: how to optimize acquisition quality and time for cardiac MR imaging. Magn Reson Imaging Clin N Am 23(1):1–6CrossRefPubMedPubMedCentral Saloner D, Liu J, Haraldsson H (2015) MR physics in practice: how to optimize acquisition quality and time for cardiac MR imaging. Magn Reson Imaging Clin N Am 23(1):1–6CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat van Heeswijk RB, Bonanno G, Coppo S, Coristine A, Kober T, Stuber M (2012) Motion compensation strategies in magnetic resonance imaging. Crit Rev Biomed Eng 40(2):99–119CrossRefPubMed van Heeswijk RB, Bonanno G, Coppo S, Coristine A, Kober T, Stuber M (2012) Motion compensation strategies in magnetic resonance imaging. Crit Rev Biomed Eng 40(2):99–119CrossRefPubMed
6.
Zurück zum Zitat Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R (2016) XD GRASP: golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med 75(2):775–788CrossRefPubMed Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R (2016) XD GRASP: golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med 75(2):775–788CrossRefPubMed
7.
Zurück zum Zitat Chen C, Li Y, Axel L, Huang J (2016) Real time dynamic MRI by exploiting spatial and temporal sparsity. Magn Reson Imaging 34(4):473–482CrossRefPubMed Chen C, Li Y, Axel L, Huang J (2016) Real time dynamic MRI by exploiting spatial and temporal sparsity. Magn Reson Imaging 34(4):473–482CrossRefPubMed
8.
Zurück zum Zitat Wech T, Pickl W, Tran-Gia J, Ritter C, Beer M, Hahn D, Köstler H (2014) Whole-heart cine MRI in a single breath-hold—a compressed sensing accelerated 3D acquisition technique for assessment of cardiac function. Rofo 186(1):37–41PubMed Wech T, Pickl W, Tran-Gia J, Ritter C, Beer M, Hahn D, Köstler H (2014) Whole-heart cine MRI in a single breath-hold—a compressed sensing accelerated 3D acquisition technique for assessment of cardiac function. Rofo 186(1):37–41PubMed
9.
Zurück zum Zitat Weber MA (2013) Magnetic resonance imaging of the skeletal musculature. Springer, Berlin Weber MA (2013) Magnetic resonance imaging of the skeletal musculature. Springer, Berlin
10.
Zurück zum Zitat Glover GH, Pauly JM (1992) Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med 28(2):275–289CrossRefPubMed Glover GH, Pauly JM (1992) Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med 28(2):275–289CrossRefPubMed
11.
Zurück zum Zitat Rudin LO, Osher S, Fatemi E (1992) Nonlinear total variation based noise removal algorithms. Physica D 60(1–4):259–268CrossRef Rudin LO, Osher S, Fatemi E (1992) Nonlinear total variation based noise removal algorithms. Physica D 60(1–4):259–268CrossRef
12.
Zurück zum Zitat Adluru G, Awate SP, Tasdizen T, Whitaker RT, Dibella EV (2007) Temporally constrained reconstruction of dynamic cardiac perfusion MRI. Magn Reson Med 57(6):1027–1036CrossRefPubMed Adluru G, Awate SP, Tasdizen T, Whitaker RT, Dibella EV (2007) Temporally constrained reconstruction of dynamic cardiac perfusion MRI. Magn Reson Med 57(6):1027–1036CrossRefPubMed
13.
Zurück zum Zitat Block KT, Uecker M, Frahm J (2007) Undersampled radial MRI with multiple coils. iterative image reconstruction using a total variation constraint. Magn Reson Med 57(6):1086–1098CrossRefPubMed Block KT, Uecker M, Frahm J (2007) Undersampled radial MRI with multiple coils. iterative image reconstruction using a total variation constraint. Magn Reson Med 57(6):1086–1098CrossRefPubMed
15.
Zurück zum Zitat Thompson RB, McVeigh ER (2006) Cardiorespiratory-resolved magnetic resonance imaging: measuring respiratory modulation of cardiac function. Magn Reson Med 56(6):1301–1310CrossRefPubMedPubMedCentral Thompson RB, McVeigh ER (2006) Cardiorespiratory-resolved magnetic resonance imaging: measuring respiratory modulation of cardiac function. Magn Reson Med 56(6):1301–1310CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Lustig M, Donoho D, Santos J, Pauly JM (2008) Compressed sensing MRI. IEEE Signal Process Mag 25(2):72–82CrossRef Lustig M, Donoho D, Santos J, Pauly JM (2008) Compressed sensing MRI. IEEE Signal Process Mag 25(2):72–82CrossRef
17.
Zurück zum Zitat Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58(6):1182–1195CrossRefPubMed Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58(6):1182–1195CrossRefPubMed
19.
Zurück zum Zitat Smith DS, Li X, Abramson RG, Quarles CC, Yankeelov TE, Welch EB (2013) Potential of compressed sensing in quantitative MR imaging of cancer. Cancer Imaging 13(4):633–644CrossRefPubMedPubMedCentral Smith DS, Li X, Abramson RG, Quarles CC, Yankeelov TE, Welch EB (2013) Potential of compressed sensing in quantitative MR imaging of cancer. Cancer Imaging 13(4):633–644CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Candes EJ, Wakin MB (2008) An introduction to compressive sampling. IEEE Signal Process Mag 25(2):21–30CrossRef Candes EJ, Wakin MB (2008) An introduction to compressive sampling. IEEE Signal Process Mag 25(2):21–30CrossRef
22.
Zurück zum Zitat Ahmad R, Ding Y, Simonetti OP (2015) Edge sharpness assessment by parametric modeling: application to magnetic resonance imaging. Concepts Magn Reson Part A 44(3):138–149CrossRef Ahmad R, Ding Y, Simonetti OP (2015) Edge sharpness assessment by parametric modeling: application to magnetic resonance imaging. Concepts Magn Reson Part A 44(3):138–149CrossRef
23.
Zurück zum Zitat Xue Y, Yu J, Kang HS, Englander S, Rosen MA, Song HK (2012) Automatic coil selection for streak artifact reduction in radial MRI. Magn Reson Med 67(2):470–476CrossRefPubMed Xue Y, Yu J, Kang HS, Englander S, Rosen MA, Song HK (2012) Automatic coil selection for streak artifact reduction in radial MRI. Magn Reson Med 67(2):470–476CrossRefPubMed
24.
Zurück zum Zitat Chan RW, Ramsay EA, Cheung EY, Plewes DB (2012) The influence of radial undersampling schemes on compressed sensing reconstruction in breast MRI. Magn Reson Med 67(2):363–377CrossRefPubMed Chan RW, Ramsay EA, Cheung EY, Plewes DB (2012) The influence of radial undersampling schemes on compressed sensing reconstruction in breast MRI. Magn Reson Med 67(2):363–377CrossRefPubMed
25.
Zurück zum Zitat Riffel P, Zoellner FG, Budjan J et al (2016) “One-stop shop”: free-breathing dynamic contrast-enhanced magnetic resonance imaging of the kidney using iterative reconstruction and continuous golden-angle radial sampling. Investig Radiol 51(11):714–719CrossRef Riffel P, Zoellner FG, Budjan J et al (2016) “One-stop shop”: free-breathing dynamic contrast-enhanced magnetic resonance imaging of the kidney using iterative reconstruction and continuous golden-angle radial sampling. Investig Radiol 51(11):714–719CrossRef
26.
Zurück zum Zitat Lin W, Huang F, Simonotto E, Duensing GR, Reykowski A (2012) Off-resonance artifacts correction with convolution in k-space (ORACLE). Magn Reson Med 67:1547–1555CrossRefPubMed Lin W, Huang F, Simonotto E, Duensing GR, Reykowski A (2012) Off-resonance artifacts correction with convolution in k-space (ORACLE). Magn Reson Med 67:1547–1555CrossRefPubMed
27.
Zurück zum Zitat Wei Li W, Storey P, Chen Q, Li BSY, Prasad PV, Edelman RR (2004) Dark flow artifacts with steady-state free precession cine MR technique: causes and implications for cardiac MR imaging. Radiology 230:569–575CrossRefPubMed Wei Li W, Storey P, Chen Q, Li BSY, Prasad PV, Edelman RR (2004) Dark flow artifacts with steady-state free precession cine MR technique: causes and implications for cardiac MR imaging. Radiology 230:569–575CrossRefPubMed
28.
Zurück zum Zitat Pang J, Sharif B, Fan Z et al (2014) ECG and navigator-free four-dimensional whole-heart coronary MRA for simultaneous visualization of cardiac anatomy and function. Magn Reson Med. doi:10.1002/mrm.25450 Pang J, Sharif B, Fan Z et al (2014) ECG and navigator-free four-dimensional whole-heart coronary MRA for simultaneous visualization of cardiac anatomy and function. Magn Reson Med. doi:10.​1002/​mrm.​25450
29.
30.
Zurück zum Zitat Piekarski E, Chitiboi T, Ramb R, Feng L, Axel L (2016) Use of self-gated radial cardiovascular magnetic resonance to detect and classify arrhythmias (atrial fibrillation and premature ventricular contraction). J Cardiovasc Magn Reson 18(1):83CrossRefPubMedPubMedCentral Piekarski E, Chitiboi T, Ramb R, Feng L, Axel L (2016) Use of self-gated radial cardiovascular magnetic resonance to detect and classify arrhythmias (atrial fibrillation and premature ventricular contraction). J Cardiovasc Magn Reson 18(1):83CrossRefPubMedPubMedCentral
32.
Zurück zum Zitat Ruan X, Liu C, Li P (2011) Detection of atrial fibrillation using R-R interval signal. In: 4th international conference on biomedical engineering and informatics, BMEI 2011, Shanghai, China, October 15–17, 2011, at Shanghai, China. doi: 10.1109/BMEI.2011.6098492 Ruan X, Liu C, Li P (2011) Detection of atrial fibrillation using R-R interval signal. In: 4th international conference on biomedical engineering and informatics, BMEI 2011, Shanghai, China, October 15–17, 2011, at Shanghai, China. doi: 10.​1109/​BMEI.​2011.​6098492
33.
Zurück zum Zitat Francone M, Dymarkowski S, Kalantzi M, Bogaert J (2005) Real-time cine MRI of ventricular septal motion: a novel approach to assess ventricular coupling. J Magn Reson Imaging 21(3):305–309CrossRefPubMed Francone M, Dymarkowski S, Kalantzi M, Bogaert J (2005) Real-time cine MRI of ventricular septal motion: a novel approach to assess ventricular coupling. J Magn Reson Imaging 21(3):305–309CrossRefPubMed
Metadaten
Titel
Two-dimensional XD-GRASP provides better image quality than conventional 2D cardiac cine MRI for patients who cannot suspend respiration
verfasst von
Eve Piekarski
Teodora Chitiboi
Rebecca Ramb
Larry A. Latson Jr
Puneet Bhatla
Li Feng
Leon Axel
Publikationsdatum
24.10.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 1/2018
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-017-0655-7

Weitere Artikel der Ausgabe 1/2018

Magnetic Resonance Materials in Physics, Biology and Medicine 1/2018 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.