A 3D computer-assisted treatment planning system for breast cancer brachytherapy treatment

verfasst von: Yuchong Rachel Jiang, Edward R. Sykes

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 4/2015

Abstract

Purpose

Brachytherapy is an option for treatment of breast cancer in some cases. This modality requires patient-specific dosimetry based on CT simulator scans. A 3D computer-assisted breast brachytherapy treatment planning system called Vision was developed and tested.

Methods

The brachytherapy treatment planning system used volume estimation and dose analysis with advanced 3D visualization. The patient treatment volume reconstruction was designed to ensure high-volume accuracy requirement of radioactive seed implantation procedure for this treatment. The system enables interactive placement of radioactive seeds embedded in original patient CT images with 3D display.

Results

The system achieved 99.73 % accuracy in volume estimation measured against the true volume and is statistically significantly more accurate than current existing commercial software at the \(p=0.05\) level.

Conclusion

A virtual 3D environment was developed to perform volume measurements, seed placements, and dose distribution planning and analysis based on 2D contours on patient CT images. This system was demonstrated to be feasible and accurate in a clinical setting.

Introduction

Breast cancer is the most common type of cancer in Canadian women [1, 2]. Every year in Ontario, about 7,800 women undergo surgeries to remove cancer from their breasts [35]. Radiation therapy has been established as an effective adjuvant therapy for breast cancer. With radiation therapy, patients no longer need to undergo mastectomy to remove the whole breast [6]. Instead, only a small part of the breast needs to be removed, and the rest is treated by radiation therapy. Lumpectomy and whole breast radiation achieve the same survival rate as mastectomy, but offers much better cosmetic results and improved health-related quality of life [2, 7, 8]. Brachytherapy has emerged as a newer technique that may offer similar control rates as whole breast radiation with the additional benefits such as shorter treatment days and improved convenience [9, 10]. Another benefit of brachytherapy for breast cancer treatment is that it has the potential to overcome the problems associated with external beam radiation therapy, such as burning of the skin [911]. In brachytherapy, radioactive sources are implanted directly into the breast.
At the Toronto-Sunnybrook Regional Cancer Center, several software packages for treatment planning were analyzed, such as Pinnacle and Varian as well as some manual intervention procedures [2, 4, 12]. Furthermore, several software solutions for prostate cancer brachytherapy treatment were also investigated [12]. These systems use ultrasound guidance and have been shown to be very successful. However, these techniques are not suitable for breast brachytherapy because the nature of operation for breast is inherently different than for other organs; furthermore, there are no existing software solutions entirely suitable for this type of treatment [2, 7].
Consequently, we sought out to design and build a system to overcome these shortcomings for breast cancer brachytherapy. A unique system is specifically tailored for breast cancer brachytherapy treatment planning. This 3D virtual simulation environment can be used for seed placement planning, dose analysis, seeds location and spacing examination, modification, visualization, and position verification. This paper presents a report on this new system called Vision that was designed, developed, and refined for breast cancer brachytherapy treatment planning. Vision is based on the technique of adjuvant partial breast irradiation, uses \(^{103}\)Pd permanent seed implants, and was developed at the Toronto-Sunnybrook Regional Cancer Center [6]. The system achieves 99.73 % accuracy in volume estimation measured against the true volume and is statistically significantly more accurate than current existing commercial software at the \(p=0.05\) level.

Literature review

Brachytherapy is one of the more well-known methods for cancer treatment that involves placing radioactive sources directly into or next to the area requiring treatment [13]. This enables clinicians to deliver a high dose with minimal impact on surrounding healthy tissues. Brachytherapy has been shown to be a highly successful treatment for cancers of the prostate, cervix, endometrium, breast, skin, bronchus, esophagus, and head and neck, as well as soft tissue sarcomas and several other types of cancer [1215].
Over the last decade, increasing numbers of breast cancer patients are being treated using interstitial radioactive implants [4, 12, 15]. Systems have been created to assist in the planning for brachytherapy treatments that provide a visualization environments based on CT/MRI images [1, 7, 11, 13, 16]. These systems provide doctors with an interactive source layout function, including accurate and quick dosage distribution calculation and dynamical 2D/3D display. These systems also provide features such as data acquisition, registration, segmentation, image processing, 3D reconstruction, and planning report output. Clinical practice has shown they meet the therapeutic demands. However, in these systems, there are deficiencies in the features of the treatment planning interface, validation, and verification modules and in the accuracy of the volume calculation—particularly for the breast treatment [4, 12, 13].
One of the more widespread systems used is GZP6—a high dose rate (HDR) brachytherapy system [2]. In their system, the interface deficiencies identified above were addressed. Their system incorporates an improved treatment planning visual interface component called CTPS [2]. The authors claim that it is an essential component for evaluating the correctness of non-predefined treatments and is easy use for medical staff [2]. Their system is limited; however, in that, their system provides only a partial solution for the planning breast treatments [2]. Other software systems exist but do not allow for virtual planning [1, 2, 11, 12].
In summary, general solutions for brachytherapy are known; however, there are deficiencies in each of these systems (e.g., tools, planning modules, verification modules, etc.) [1, 2, 4, 1113, 15]. These existing solutions do not adequately provide the full range of tools or support for oncologists to effectively and accurately treat breast cancer patients. There is a need for a comprehensive solution for breast brachytherapy. Such a system is presented in this paper. In our system, called Vision, an entire 3D treatment planning and verification solution are provided.

Methods

In the design and development of Vision, the following method was used:
  • reading CT data set, Region of Interests (ROIs), and RT plans from Pinnacle;
  • reslicing the CT data, ROIs, and RT to the oblique direction;
  • reconstructing the volume with high resolution;
  • computing volume interpolation for accurate target volume measurement;
  • simulating the fiducial needle and providing surgical templates;
  • facilitating radiation source editing by designing a customized editor for managing \(^{103}\)Pd source model data;
  • determining a template size for the seed implantation—usually the template must cover the whole area of the ROIs;
  • superimposing the simulated template onto each image that has ROIs along the gantry angle;
  • providing an interface for placing needle/seeds based on the superimposed template;
  • calculating the radial dose distributions on the seeds placed by computer simulation to avoid hot/cold spots to ensure a uniform dose distribution that will cover the entire Planned Target Volume (PTV) by adjusting source activity and the number/position of needles/seeds;
  • projecting the seeds back to the original image data set for 3D viewing.

Data collection and processing

The data collection phase involves a CT scanner on which the patient lies in prone position. The CT images are taken around the chest region traveling in the direction from the head to the feet. The \(Z\) spacing is 5 mm away from the breast and 2 mm on the breast region. Oncologists manually determine the PTV, clinical target volume (CTV), and ISO center and a gantry angle on each slice around the cavity region where the tumor was removed. The gantry angle is parallel with the rib cage to protect needles from going through the rib and lung region during the operation. The oblique reslicing direction is perpendicular to the gantry angle. Seed placement is carried out on the resliced images inside the PTV contours that are reconstructed in the oblique direction. The number of interpolations between two adjacent slices is determined by the distance between them. For example, if the distance is 5 mm, then four interpolations are computed. Our interpolation method guarantees uniform unit spacing between slices.

Reslicing images and ROIs

The reslicing task is done by projecting the data set on the direction perpendicular to the gantry angle after applying interpolation on the CT images and its ROIs. The new reconstructed data can be described as follows:
  • the \(X\) direction is parallel with the original \(Z\) direction
  • the \(Y\) direction is same as the original \(Y\) direction with an angle that is derived from the gantry angle.
From posterior to anterior:
  • the \(Z\) direction parallel with the original \(X\) direction with an angle equal to the gantry angle.
The total number of resliced slices is automatically calculated using the ROIs size along the gantry angle and is described by the following equation:
$$\begin{aligned} \mathrm{Total\,number}_{\mathrm{reslices}} =\left( \frac{\mathrm{ROI\,size\,along\,gantry\,angle}}{\mathrm{\,reslice\,thickness}} \right) +1 \end{aligned}$$
The original data set is recalculated from the view that the gantry angle is specified by oncologist and is suitable for seed placement procedure, see Figs. 1 and 2.

Volume interpolation

The interpolation component is designed to reconstruct image data and treatment volumes from the original slice thickness and spacing (can be any size) into 1 mm in spacing. The reconstructed target volume is built from the new set of contours—original plus interpolated contours. We assign the spacing equal to the number of interpolations. Therefore, the new resolution is always equal to 1 mm. With 5 mm spacing between two adjacent slices, four interpolated images and contours are created to achieve 1 mm of unit spacing.
When creating interpolation contours, two types of interpolation process are encountered, overlapped and intersected types. The intersect type is that one contour is inside the other; the overlapped type is that one contour partially inside the other. The interpolation contour is the dividing line that is equally dividing the region between two contours when only one interpolation is required. Figure 3 describes the mathematical operation for deriving the dividing line.

Probability function for volume interpolation

When 2D contours are used to define treatment volumes, a partially subjective process is underlined [18]. The accuracy of volume definition depends on numerous image parameters such as the contrast to noise ratio and the signal-to-noise ratio. The subjective process and imaging parameters might cause inaccurate target volume creation. The volume definition using 2D contours is especially difficult at the edges of the target because of the partial volume effect that is related mostly to the image slice thickness [18]. A low inter- and intra-observer reproducibility was reported in different clinical studies, mainly because of the subjective and objective factors [18]. Consequently, a probability function was designed in Vision to study the partial volume effect of the volume definition caused by the 2D contouring method. The function is presented below using 26 for the total number of voxels in the neighborhood:
$$\begin{aligned} P=\frac{\mathrm{number}\, \mathrm{of}\, \mathrm{voxels}\,\mathrm{in}\, \mathrm{the}\, \mathrm{target}\, \mathrm{volume}}{26} \end{aligned}$$
The function is able to measure how many voxels belong to PTV in the neighborhood of current voxel. When the entire neighborhood of the current voxel belongs to the target volume, the probability value (output of the probability function) is 1; otherwise, it is 0. If part of the neighborhood voxels is inside the target volume and part is outside of the target volume, the probability value is between 0 and 1. A threshold value can be set for the probability function. The treatment volume is made up of all the voxels whose neighborhood probability value is greater than the threshold. When the threshold value is larger, the estimated volume is smaller (more neighborhood voxels are excluded from the target volume), see Fig. 4. The volumes generated using this probability function is more accurate than other methods when the threshold is set to 0.5.

Radiation source configuration in vision

Task Group 43 (TG-43) from the American Association of Physicists in Medicine (AAPM) is currently used at the Toronto-Sunnybrook Regional Cancer Center [6]. Other kinds of radiation seeds (also called radial sources) can be used as radiation sources [6]. The radial sources are simulated by following AAPM TG-43 guidelines for brachytherapy to calculate dose distribution on different radiation sources. A source editor interface allows the user to input the radiation source and dose distribution charts. Currently, [\(^{103}\)Pd (2335) Best] is used for breast cancer brachytherapy [18].

Virtual template implementation

Two templates guiding the insertion of needles loaded with seeds are simulated in Vision. A fiducial needle used to hook the surgical cavity, which is then attached to a template that guides the insertion of needles, is also simulated. Figure 5 shows the simulated template. Following image shows a real-time implantation operation using a needle one of the templates.

Seed placement planning and dose distribution analysis interface

Vision provides an accurate, fast, and easy-to-use interface for seed placement on resliced CT data. Source placement spacing calculation in both 2D and 3D and dose volume histogram (DVH) is available. After seeds are placed on the resliced CT data set, a 3D view with the seeds can be generated for intuitive seed spacing analysis. This is a unique feature of Vision that does not exist among commercial brachytherapy treatment planning software systems. This virtual 3D environment displays the seeds in the original data set that provides an intuitive viewing environment. Figure 6 shows an example of the 3D volume rendering.

Source editor: an interface for seed placement and dose calculation

Vision’s source editor provides an interface that allows sources to be input and used. Currently, \(^{103}\)Pd (2335) Best is used for breast cancer brachytherapy. Figure 7 shows an example of source placement editing in 2D and 3D and its dose volume histogram (DVH). The dose at a particular point is calculated by summing the dose of all the seeds that affect the point. \(^{103}\)Pd (2335) Best seed is a cylindrically symmetric source, and dose distribution is a two-dimensional and can be described in terms of a polar coordinate system with its origin at the source center where \(r\) is the distance to the point of interest and \(q\) is the angle with respect to the long axis of the source.
The dose rate, \(P(r,\theta )\), at point (r, \(\theta \)) can be written as follows:
$$\begin{aligned} P(r,\theta )=S_{k}^{\left[ \frac{G(r,\theta )}{G(ro,\theta o)}\right] g(r,\theta )},\hbox {where} \end{aligned}$$
\(r\) is the distance to the point of interest, \(\theta \) is the angle with respect to the long axis of the source, \(S_{k}\) is air kerma strength of the source; \(\left[ {\frac{G\left( {r,\theta } \right) }{G\left( {r0,\theta 0} \right) }} \right] g\left( r \right) F(r,\theta ) \) is dose rate constant, \(G(r,\theta )\) is geometry factor, \(g(r)\) is radial dose function, \(F(r, \theta )\) is anisotropy function.

Reslice back with virtual seeds

One method of verifying the plan is by projecting the virtual implants back to their original data set. This innovative feature of our Vision system allows oncologists to view the implant position and spacing in 3D intuitively. In addition, Vision enables seed placement results to be viewed and validated.

Results

We have tested the system using more than 30 data sets of breast cancer patients selected from the Toronto-Sunnybrook Regional Cancer Center after breast cancer treatment. Three phantom data sets were also tested.
The dose distribution calculation is performed in real time while seeds are added. After seed implantation, 3D viewing of target volume containing all the placed seeds is displayed. The reslice calculation is also verified by projecting the image back to its original scanned direction. Our 3D reconstruction algorithm performs volume interpolation because of the non-uniform resolution of the scanned data sets. Normally, the in-plane (\(X\)\(Y\)) resolution of a data set is greater than the out-of-plane (\(Z\)) resolution. This difference can range from a factor of 2–10. Our volume interpolation algorithm in Vision is able to adjust the number of interpolations between two slices automatically and then interpolate accordingly (see Fig. 8). The result spacing and thickness are always a unit of 1 mm between slices. Mathematical morphology operation for medial axis finding [17] and level set method [19, 20] are employed in the interpolation process.
Our volume estimation method produces 99.73 % accuracy on sphere phantom data sets. When the probability threshold of volume estimation method is set to 0.5, the volume estimation gives the highest accurate volume estimation result. The result was compared with the Pinnacle system (Philips Medical imaging system). A statistically significant accuracy improvement on volume estimation was achieved (99.38 % from 92.38 % at the \(p=0.05\) level). The true volume for this sphere phantom is as follows: 22.45 cm\(^{3} \quad ({\mathrm{diameter}=3.5\,\mathrm{cm},\mathrm{volume}=\frac{4\pi 3^{3}}{3}}={22.45\,\mathrm{cm}^{3}})\). When using the interpolation method, more consistent volume estimation output is achieved. Our interpolation method provides the most accurate volume estimate as depicted in Fig. 9. Table 1 shows the testing results for seven data sets. The volume output from the existing commercial software overestimated the true volume (see Table 2; Figs. 9, 10).
Table 1
Volume estimations using patient CT data sets using different methods (in cm\(^{3})\)
Case no.
Pinnacle
No interpolation
Interpolation
Reslice without interpolation
Reslice with interpolation
1
10.01
9.66
8.78
9.86
8.85
2
32.59
32.52
28.83
26.90
28.30
3
13.15
12.64
10.89
13.02
10.86
4
16.69
16.01
13.90
15.74
13.65
5
11.28
10.60
9.53
10.67
9.48
6
4.09
3.86
3.49
4.08
3.70
7
4.12
3.89
3.31
4.01
3.30
Table 2
Volumes (in cm\(^{3}\)) and reconstructed volumes after reslice by our method with interpolation and without interpolation on phantom object with known volume (22.45 cm\(^{3})\)
Pinnacle
No interpolation
Interpolation
Reslice without interpolation
Reslice with interpolation
24.16
24.23
23.13
24.34
22.61

Conclusions

This paper described Vision, a 3D computer-assisted treatment planning system we developed for breast cancer brachytherapy treatment. The system was developed using mathematical theories for accurate volume estimation and dose analysis, as well as advanced 3D visualization technologies. The patient treatment volume reconstructed by a method developed in this project significantly improved the accuracy from existing methods and guarantees the high-volume accuracy requirement of radioactive seed implantation procedure for this treatment. The Vision system is a virtual 3D environment that allows radiation oncologists to perform volume measurement, seed placement, and dose distribution planning and analysis based on a set of 2D contours on patient CT images. Vision is able to embed placed seeds on original patient CT images and displayed in an institutive 3D environment that can be easily manipulated by the oncologist.
Using our 3D neighborhood probability function to estimate the treatment volume, we were able to rectify the redundant voxels located outside of the target region and minimize the discrepancies in volume definition by 2D manually contouring between adjacent slices. In this research project, advanced mathematical theories were researched and applied, such as mathematical morphology theory, partial differential equations, probability statistics, and mathematical modeling. Our system achieves 99.73 % accuracy in volume estimation measured against the true volume and is statistically significantly more accurate than current existing commercial software at the \(p=0.05\) level.
Future work will focus on automatic source seed placement that could be adjusted by an oncologist if necessary. The purpose of this work was to optimize the seed placements such that radial dose distributions will be uniformly distributed over the entire PTV to avoid hot/cold spots. This, in turn, will assist an oncologist by minimizing the number of seeds required and potentially increasing the effectiveness of the treatment.

Acknowledgments

This work was made possible by SHARCNET (Shared Hierarchical Academic Research Computing Network). Please see: www.​sharcnet.​ca.

Conflict of interest

The authors certify that we have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements) in the subject matter or materials discussed in this manuscript.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Xiang H, Zhuang T (2002) CT/MRI-based system for 3-D brachytherapy treatment planning. In: 24th annual conference and the annual fall meeting of the biomedical engineering society EMBS/BMES conference, pp 1196–1197 Xiang H, Zhuang T (2002) CT/MRI-based system for 3-D brachytherapy treatment planning. In: 24th annual conference and the annual fall meeting of the biomedical engineering society EMBS/BMES conference, pp 1196–1197
2.
Zurück zum Zitat Hariri S (2010) Asl AK introducing a complementary treatment planning software for GZP6 brachytherapy system. In: 3rd international conference on biomedical engineering and informatics, pp 16–18 Hariri S (2010) Asl AK introducing a complementary treatment planning software for GZP6 brachytherapy system. In: 3rd international conference on biomedical engineering and informatics, pp 16–18
3.
Zurück zum Zitat Fisher B, Anderson S, Bryant J, Margolese RG, Deutsch M, Fisher ER, Jeong JH, Wolmark N (2002) Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer. N Engl J Med 347(16):1233–1241CrossRefPubMed Fisher B, Anderson S, Bryant J, Margolese RG, Deutsch M, Fisher ER, Jeong JH, Wolmark N (2002) Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer. N Engl J Med 347(16):1233–1241CrossRefPubMed
4.
Zurück zum Zitat Kuske RR, Young SS (2013) Breast brachytherapy versus whole-breast irradiation: reported differences may be statistically significant but clinically trivial. Int J Radiat Oncol Biol Phys 88(2):266–268CrossRef Kuske RR, Young SS (2013) Breast brachytherapy versus whole-breast irradiation: reported differences may be statistically significant but clinically trivial. Int J Radiat Oncol Biol Phys 88(2):266–268CrossRef
6.
Zurück zum Zitat Pignol J-P, Keller B, Rakovitch E, Sankreacha R, Easton H, Que W (2006) First report of a permanent breast \(^{103}\)Pd seed implant as adjuvant radiation treatment for early-stage breast cancer. Int J Radiat Oncol Biol Phys 64(1):176–181CrossRefPubMed Pignol J-P, Keller B, Rakovitch E, Sankreacha R, Easton H, Que W (2006) First report of a permanent breast \(^{103}\)Pd seed implant as adjuvant radiation treatment for early-stage breast cancer. Int J Radiat Oncol Biol Phys 64(1):176–181CrossRefPubMed
7.
Zurück zum Zitat Zhang T, Patel R (2011) Optimization-based dosimetry planning for brachytherapy. In: 33rd annual international conference of the IEEE EMBS, Boston, Massachusetts, USA, 2011. IEEE, pp 5569–5572 Zhang T, Patel R (2011) Optimization-based dosimetry planning for brachytherapy. In: 33rd annual international conference of the IEEE EMBS, Boston, Massachusetts, USA, 2011. IEEE, pp 5569–5572
8.
Zurück zum Zitat Jiang RY (2007) Software tool for breast cancer brachytherapy planning using VTK. Paper presented at the 6th IEEE international conference on cognitive informatics, Lake Tahoe, USA Jiang RY (2007) Software tool for breast cancer brachytherapy planning using VTK. Paper presented at the 6th IEEE international conference on cognitive informatics, Lake Tahoe, USA
9.
Zurück zum Zitat Curran D, van Dongen JP, Aaronson NK, Kiebert G, Fentiman IS, Mignolet F, Bartelink H (1998) Quality of life of early-stage breast cancer patients treated with radical mastectomy or breast-conserving procedures: results of EORTC Trial 10801. The European Organization for Research and Treatment of Cancer (EORTC), Breast Cancer Co-operative Group (BCCG). Eur J Cancer 34(4):307–314CrossRefPubMed Curran D, van Dongen JP, Aaronson NK, Kiebert G, Fentiman IS, Mignolet F, Bartelink H (1998) Quality of life of early-stage breast cancer patients treated with radical mastectomy or breast-conserving procedures: results of EORTC Trial 10801. The European Organization for Research and Treatment of Cancer (EORTC), Breast Cancer Co-operative Group (BCCG). Eur J Cancer 34(4):307–314CrossRefPubMed
10.
Zurück zum Zitat van Dongen JA, Voogd AC, Fentiman IS, Legrand C, Sylvester RJ, Tong D, van der Schueren E, Helle PA, van Zijl K, Bartelink H (2000) Long-term results of a randomized trial comparing breast-conserving therapy with mastectomy: European Organization for Research and Treatment of Cancer 10801 trial. J Natl Cancer Inst 92(14):1143–1150CrossRefPubMed van Dongen JA, Voogd AC, Fentiman IS, Legrand C, Sylvester RJ, Tong D, van der Schueren E, Helle PA, van Zijl K, Bartelink H (2000) Long-term results of a randomized trial comparing breast-conserving therapy with mastectomy: European Organization for Research and Treatment of Cancer 10801 trial. J Natl Cancer Inst 92(14):1143–1150CrossRefPubMed
11.
Zurück zum Zitat Poulin E, Fekete C-AC, Létourneau M, Fenster A, Pouliot J, Beaulieu L (2004) Virtual planning of multicatheter brachytherapy implants for accelerated partial breast irradiation. In: 26th annual international conference of the engineering in medicine and biology society conference. IEEE, pp 1–5 Poulin E, Fekete C-AC, Létourneau M, Fenster A, Pouliot J, Beaulieu L (2004) Virtual planning of multicatheter brachytherapy implants for accelerated partial breast irradiation. In: 26th annual international conference of the engineering in medicine and biology society conference. IEEE, pp 1–5
12.
Zurück zum Zitat Aneja S, Yu JB (2014) Comparative effectiveness research in radiation oncology: stereotactic radiosurgery, hypofractionation, and brachytherapy. Semin Radiat Oncol 24(1):35–42CrossRefPubMed Aneja S, Yu JB (2014) Comparative effectiveness research in radiation oncology: stereotactic radiosurgery, hypofractionation, and brachytherapy. Semin Radiat Oncol 24(1):35–42CrossRefPubMed
13.
Zurück zum Zitat Smith GL, Jiang J, Buchholz TA, Xu Y, Hoffman KE, Giordano SH, Hunt KK, Smith BD (2014) Benefit of adjuvant brachytherapy versus external beam radiation for early breast cancer: impact of patient stratification on breast preservation. Int J Radiat Oncol Biol Phys 88(2):274–284CrossRefPubMedCentralPubMed Smith GL, Jiang J, Buchholz TA, Xu Y, Hoffman KE, Giordano SH, Hunt KK, Smith BD (2014) Benefit of adjuvant brachytherapy versus external beam radiation for early breast cancer: impact of patient stratification on breast preservation. Int J Radiat Oncol Biol Phys 88(2):274–284CrossRefPubMedCentralPubMed
14.
Zurück zum Zitat Stewart AJ, Khan AJ, Devlin PM (2010) Partial breast irradiation: a review of techniques and indications. Br J Radiol 83(989):369–378 Stewart AJ, Khan AJ, Devlin PM (2010) Partial breast irradiation: a review of techniques and indications. Br J Radiol 83(989):369–378
15.
Zurück zum Zitat Vargo JA, Verma V, Kim H, Kalash R, Heron DE, Johnson R, Beriwal S (2014) Extended (5-year) outcomes of accelerated partial breast irradiation using mammosite balloon brachytherapy: patterns of failure, patient selection, and dosimetric correlates for late toxicity. Int J Radiat Oncol Biol Phys 88(2):285–291CrossRefPubMed Vargo JA, Verma V, Kim H, Kalash R, Heron DE, Johnson R, Beriwal S (2014) Extended (5-year) outcomes of accelerated partial breast irradiation using mammosite balloon brachytherapy: patterns of failure, patient selection, and dosimetric correlates for late toxicity. Int J Radiat Oncol Biol Phys 88(2):285–291CrossRefPubMed
16.
Zurück zum Zitat Foppiano F, Guatelli S, Moscicki J (2003) Pia MG From DICOM to GRID: a dosimetric system for brachytherapy born from HEP. In: Nuclear science symposium conference record. pp 1746– 1750 Foppiano F, Guatelli S, Moscicki J (2003) Pia MG From DICOM to GRID: a dosimetric system for brachytherapy born from HEP. In: Nuclear science symposium conference record. pp 1746– 1750
17.
Zurück zum Zitat Barrett W, Mortenson E, Taylor D (1994) An image space algorithm for morphological contour interpolation. Paper presented at the graphics interface conference Barrett W, Mortenson E, Taylor D (1994) An image space algorithm for morphological contour interpolation. Paper presented at the graphics interface conference
18.
Zurück zum Zitat Caudrelier JM, Vial S, Gibon D (2003) MRI definition of target volumes using fuzzy logic method for 3D conformal radiation therapy. Int Radiat Oncol Biol Phys 55(1):225–233 Caudrelier JM, Vial S, Gibon D (2003) MRI definition of target volumes using fuzzy logic method for 3D conformal radiation therapy. Int Radiat Oncol Biol Phys 55(1):225–233
19.
Zurück zum Zitat Sapiro G (2001) Geometric partial differential equations and image analysis. Cambridge University Press, New YorkCrossRef Sapiro G (2001) Geometric partial differential equations and image analysis. Cambridge University Press, New YorkCrossRef
20.
Zurück zum Zitat Sethian JA (1999) Level set methods and fast marching methods: evolving interfaces in computational geometry, fluid mechanics, computer vision, and materials science. Cambridge Press, Cambridge. Sethian JA (1999) Level set methods and fast marching methods: evolving interfaces in computational geometry, fluid mechanics, computer vision, and materials science. Cambridge Press, Cambridge.
Metadaten
Titel
A 3D computer-assisted treatment planning system for breast cancer brachytherapy treatment
verfasst von
Yuchong Rachel Jiang
Edward R. Sykes
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 4/2015
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-014-1092-y

Weitere Artikel der Ausgabe 4/2015

International Journal of Computer Assisted Radiology and Surgery 4/2015 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.