Ultrasound-based liver tracking utilizing a hybrid template/optical flow approach

verfasst von: Tom Williamson, Wa Cheung, Stuart K. Roberts, Sunita Chauhan

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 10/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

With the ongoing shift toward reduced invasiveness in many surgical procedures, methods for tracking moving targets within the body become vital. Non-invasive treatment methods such as stereotactic radiation therapy and high intensity focused ultrasound, in particular, rely on the accurate localization of targets throughout treatment to ensure optimal treatment provision. This work aims at developing a robust, accurate and fast method for target tracking based on ultrasound images.

Methods

A method for tracking of targets in real-time ultrasound image data was developed, based on the combination of template matching, dense optical flow and image intensity information. A weighting map is generated from each of these approaches which are then normalized, weighted and combined, with the weighted mean position then calculated to predict the current position. The approach was evaluated on the Challenge for Liver Ultrasound Tracking 2015 dataset, consisting of a total of 24 training and 39 test datasets with a total of 53 and 85 annotated targets throughout the liver, respectively.

Results

The proposed method was implemented in MATLAB and achieved an accuracy of \(0.80\pm 0.80\) (95%: 1.91) mm and \(0.74\pm 1.03\) (95%: 1.85) mm on the training and test data, respectively. Tracking frequencies of between 8 and 36 fps (mean of 22 fps) were observed, largely dependent on the size of the region of interest. The achieved results represent an improvement in mean accuracy of approximately 0.3 mm over the reported methods in existing literature.

Conclusions

This work describes an accurate and robust method for the tracking of points of interest within 2D ultrasound data, based on a combination of multi-template matching, dense optical flow and relative image intensity information.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Illing RO, Kennedy JE, Wu F, ter Haar GR, Protheroe AS, Friend PJ, Gleeson FV, Cranston DW, Phillips RR, Middleton MR (2005) The safety and feasibility of extracorporeal high-intensity focused ultrasound (HIFU) for the treatment of liver and kidney tumours in a Western population. Br J Cancer 93(8):890–5CrossRefPubMedPubMedCentral Illing RO, Kennedy JE, Wu F, ter Haar GR, Protheroe AS, Friend PJ, Gleeson FV, Cranston DW, Phillips RR, Middleton MR (2005) The safety and feasibility of extracorporeal high-intensity focused ultrasound (HIFU) for the treatment of liver and kidney tumours in a Western population. Br J Cancer 93(8):890–5CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Chauhan S (2008) Image guided robotic systems for focal ultrasound based surgical applications. In: Medical robotics. I-Tech Education and Publishing, pp 29–40 Chauhan S (2008) Image guided robotic systems for focal ultrasound based surgical applications. In: Medical robotics. I-Tech Education and Publishing, pp 29–40
3.
Zurück zum Zitat Timmerman R, Kavanagh B (2005) Stereotactic body radiation therapy. Curr Probl Cancer 29(3):120–157CrossRefPubMed Timmerman R, Kavanagh B (2005) Stereotactic body radiation therapy. Curr Probl Cancer 29(3):120–157CrossRefPubMed
5.
Zurück zum Zitat Abhilash RH, Chauhan S (2012) Respiration-induced movement correlation for synchronous noninvasive renal cancer surgery. IEEE Trans Ultrason Ferroelectr Freq Control 59(7):1478–86CrossRefPubMed Abhilash RH, Chauhan S (2012) Respiration-induced movement correlation for synchronous noninvasive renal cancer surgery. IEEE Trans Ultrason Ferroelectr Freq Control 59(7):1478–86CrossRefPubMed
6.
Zurück zum Zitat Gierga DP, Brewer J, Sharp GC, Betke M, Willett CG, Chen GTY (2005) The correlation between internal and external markers for abdominal tumors: implications for respiratory gating. Int J Radiat Oncol 61(5):1551–1558CrossRef Gierga DP, Brewer J, Sharp GC, Betke M, Willett CG, Chen GTY (2005) The correlation between internal and external markers for abdominal tumors: implications for respiratory gating. Int J Radiat Oncol 61(5):1551–1558CrossRef
7.
Zurück zum Zitat Pepin EW, Wu H, Zhang Y, Lord B (2011) Correlation and prediction uncertainties in the CyberKnife Synchrony respiratory tracking system. Med Phys 38(7):4036–4044CrossRefPubMedPubMedCentral Pepin EW, Wu H, Zhang Y, Lord B (2011) Correlation and prediction uncertainties in the CyberKnife Synchrony respiratory tracking system. Med Phys 38(7):4036–4044CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Crijns SPM, Raaymakers BW, Lagendijk JJW (2012) Proof of concept of MRI-guided tracked radiation delivery: tracking one-dimensional motion. Phys Med Biol 57(23):7863–7872CrossRefPubMed Crijns SPM, Raaymakers BW, Lagendijk JJW (2012) Proof of concept of MRI-guided tracked radiation delivery: tracking one-dimensional motion. Phys Med Biol 57(23):7863–7872CrossRefPubMed
9.
Zurück zum Zitat Menten MJ, Fast MF, Nill S, Kamerling CP, McDonald F, Oelfke U (2016) Lung stereotactic body radiotherapy with an MR-linac—Quantifying the impact of the magnetic field and real-time tumor tracking. Radiother Oncol 119(3):461–466CrossRefPubMedPubMedCentral Menten MJ, Fast MF, Nill S, Kamerling CP, McDonald F, Oelfke U (2016) Lung stereotactic body radiotherapy with an MR-linac—Quantifying the impact of the magnetic field and real-time tumor tracking. Radiother Oncol 119(3):461–466CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Kim Y (2015) Advances in MR image-guided high-intensity focused ultrasound therapy. Int J Hyperth 31(3):225–232CrossRef Kim Y (2015) Advances in MR image-guided high-intensity focused ultrasound therapy. Int J Hyperth 31(3):225–232CrossRef
11.
Zurück zum Zitat De Luca V, Benz T, Kondo S, König L, Lübke D, Rothlübbers S, Somphone O, Allaire S, Bell ML, Chung DYF, Cifor A (2015) The 2014 liver ultrasound tracking benchmark. Phys Med Biol 60(14):5571CrossRefPubMedPubMedCentral De Luca V, Benz T, Kondo S, König L, Lübke D, Rothlübbers S, Somphone O, Allaire S, Bell ML, Chung DYF, Cifor A (2015) The 2014 liver ultrasound tracking benchmark. Phys Med Biol 60(14):5571CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat O’Shea T, Bamber J, Harris E (2014) Liver feature motion estimation in long high frame rate 2D ultrasound sequences. In: Challenge on liver ultrasound tracking, pp 5–12 O’Shea T, Bamber J, Harris E (2014) Liver feature motion estimation in long high frame rate 2D ultrasound sequences. In: Challenge on liver ultrasound tracking, pp 5–12
13.
Zurück zum Zitat Somphone O, Allaire S, Mory B, Dufour C (2014) Live feature tracking in ultrasound liver sequences with sparse demons. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 53–60 Somphone O, Allaire S, Mory B, Dufour C (2014) Live feature tracking in ultrasound liver sequences with sparse demons. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 53–60
14.
Zurück zum Zitat Luebke D, Grozea C (2014) MICCAI CLUST 2014—Bayesian real-time liver feature ultrasound tracking. In: MICCAI workshop: challenge on Liver Ultrasound Tracking, pp 37–44 Luebke D, Grozea C (2014) MICCAI CLUST 2014—Bayesian real-time liver feature ultrasound tracking. In: MICCAI workshop: challenge on Liver Ultrasound Tracking, pp 37–44
15.
Zurück zum Zitat Benz T, Kowarschik M, Navab N (2014) Kernel-based tracking in ultrasound sequences of liver. In: Proceedings of challenge on liver ultrasound tracking, MICCAI workshop, pp 21–28 Benz T, Kowarschik M, Navab N (2014) Kernel-based tracking in ultrasound sequences of liver. In: Proceedings of challenge on liver ultrasound tracking, MICCAI workshop, pp 21–28
16.
Zurück zum Zitat Kondo S (2014) Liver ultrasound tracking using long-term and short-term template matching. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 13–20 Kondo S (2014) Liver ultrasound tracking using long-term and short-term template matching. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 13–20
17.
Zurück zum Zitat König L, Kipshagen T, Rühaak J (2014) A non-linear image registration scheme for real-time liver ultrasound tracking using normalized gradient fields. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 29–36 König L, Kipshagen T, Rühaak J (2014) A non-linear image registration scheme for real-time liver ultrasound tracking using normalized gradient fields. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 29–36
18.
Zurück zum Zitat Rothlübbers S, Schwaab J, Jenne J, Günther M (2014) MICCAI CLUST 2014—Bayesian real-time liver feature ultrasound tracking. In: MICCAI workshop: challenge on Liver Ultrasound Tracking, pp 45–52 Rothlübbers S, Schwaab J, Jenne J, Günther M (2014) MICCAI CLUST 2014—Bayesian real-time liver feature ultrasound tracking. In: MICCAI workshop: challenge on Liver Ultrasound Tracking, pp 45–52
19.
Zurück zum Zitat Nouri D, Rothberg A (2015) Liver ultrasound tracking using a learned distance metric. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 5–12 Nouri D, Rothberg A (2015) Liver ultrasound tracking using a learned distance metric. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 5–12
20.
Zurück zum Zitat Kondo S (2015) Liver ultrasound tracking using kernelized correlation filter with adaptive window size selection. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 13–19 Kondo S (2015) Liver ultrasound tracking using kernelized correlation filter with adaptive window size selection. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 13–19
21.
Zurück zum Zitat Makhinya M, Goksel O (2015) Motion tracking in 2D ultrasound using vessel models and robust optic-flow. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 20–27 Makhinya M, Goksel O (2015) Motion tracking in 2D ultrasound using vessel models and robust optic-flow. In: MICCAI workshop: challenge on liver ultrasound tracking, pp 20–27
22.
Zurück zum Zitat Hallack A, Papiez W B, Cifor A, Gooding MJ, Schnabel J A (2015) Robust liver ultrasound tracking using dense distinctive image features. In: Proceedings MICCAI workshop: challenge on liver ultrasound tracking, pp 28–35 Hallack A, Papiez W B, Cifor A, Gooding MJ, Schnabel J A (2015) Robust liver ultrasound tracking using dense distinctive image features. In: Proceedings MICCAI workshop: challenge on liver ultrasound tracking, pp 28–35
23.
Zurück zum Zitat Ozkan E, Tanner C, Kastelic M, Mattausch O, Makhinya M, Goksel O (2017) Robust motion tracking in liver from 2D ultrasound images using supporters. Int J Comput Assist Radiol Surg 12(6):941–950CrossRefPubMed Ozkan E, Tanner C, Kastelic M, Mattausch O, Makhinya M, Goksel O (2017) Robust motion tracking in liver from 2D ultrasound images using supporters. Int J Comput Assist Radiol Surg 12(6):941–950CrossRefPubMed
24.
Zurück zum Zitat Golemati S, Sassano A, Lever MJ, Bharath AA, Dhanjil S, Nicolaides AN (2003) Carotid artery wall motion estimated from b-mode ultrasound using region tracking and block matching. Ultrasound Med Biol 29(3):387–399CrossRefPubMed Golemati S, Sassano A, Lever MJ, Bharath AA, Dhanjil S, Nicolaides AN (2003) Carotid artery wall motion estimated from b-mode ultrasound using region tracking and block matching. Ultrasound Med Biol 29(3):387–399CrossRefPubMed
25.
Zurück zum Zitat Li M, Kambhamettu C, Stone M (2005) Automatic contour tracking in ultrasound images. Clin Linguist Phon 19(6–7):545–554CrossRefPubMed Li M, Kambhamettu C, Stone M (2005) Automatic contour tracking in ultrasound images. Clin Linguist Phon 19(6–7):545–554CrossRefPubMed
26.
Zurück zum Zitat Zhang X, Günther M, Bongers A (2010) Real-time organ tracking in ultrasound imaging using active contours and conditional density propagation, pp 286–294 Zhang X, Günther M, Bongers A (2010) Real-time organ tracking in ultrasound imaging using active contours and conditional density propagation, pp 286–294
27.
Zurück zum Zitat De Luca V, Tschannen M, Székely G, Tanner C (2013) A learning-based approach for fast and robust vessel tracking in long ultrasound sequences, pp 518–525 De Luca V, Tschannen M, Székely G, Tanner C (2013) A learning-based approach for fast and robust vessel tracking in long ultrasound sequences, pp 518–525
28.
Zurück zum Zitat Yeung Fai, Levinson SF, Fu Dongshan, Parker KJ (1998) Feature-adaptive motion tracking of ultrasound image sequences using a deformable mesh. IEEE Trans Med Imaging 17(6):945–956CrossRefPubMed Yeung Fai, Levinson SF, Fu Dongshan, Parker KJ (1998) Feature-adaptive motion tracking of ultrasound image sequences using a deformable mesh. IEEE Trans Med Imaging 17(6):945–956CrossRefPubMed
29.
Zurück zum Zitat Abolmaesumi P, Salcudean SE, Zhu Wen-Hong, Sirouspour MR, DiMaio SP (2002) Image-guided control of a robot for medical ultrasound. IEEE Trans Robot Autom 18(1):11–23CrossRef Abolmaesumi P, Salcudean SE, Zhu Wen-Hong, Sirouspour MR, DiMaio SP (2002) Image-guided control of a robot for medical ultrasound. IEEE Trans Robot Autom 18(1):11–23CrossRef
30.
Zurück zum Zitat Ortmaier T, Vitrani M-A, Morel G, Pinault S Robust real-time instrument tracking in ultrasound images for visual servoing. In: Proceedings of the 2005 IEEE international conference on robotics and automation, pp 2167–2172 Ortmaier T, Vitrani M-A, Morel G, Pinault S Robust real-time instrument tracking in ultrasound images for visual servoing. In: Proceedings of the 2005 IEEE international conference on robotics and automation, pp 2167–2172
31.
Zurück zum Zitat Farnebäck Gunnar (2003) Two-frame motion estimation based on polynomial expansion. In: Scandinavian conference on image analysis, pp 363–370CrossRef Farnebäck Gunnar (2003) Two-frame motion estimation based on polynomial expansion. In: Scandinavian conference on image analysis, pp 363–370CrossRef
32.
Zurück zum Zitat Idzenga T, Gaburov E, Vermin W, Menssen J, De Korte C (2014) Fast 2-D ultrasound strain imaging: the benefits of using a GPU. IEEE Trans Ultrason Ferroelectr Freq Control 61(1):207–213CrossRefPubMed Idzenga T, Gaburov E, Vermin W, Menssen J, De Korte C (2014) Fast 2-D ultrasound strain imaging: the benefits of using a GPU. IEEE Trans Ultrason Ferroelectr Freq Control 61(1):207–213CrossRefPubMed
33.
Zurück zum Zitat Castaño-Díez D, Moser D, Schoenegger A, Pruggnaller S, Frangakis AS (2008) Performance evaluation of image processing algorithms on the GPU. J Struct Biol 164(1):153–160CrossRefPubMed Castaño-Díez D, Moser D, Schoenegger A, Pruggnaller S, Frangakis AS (2008) Performance evaluation of image processing algorithms on the GPU. J Struct Biol 164(1):153–160CrossRefPubMed
Metadaten
Titel
Ultrasound-based liver tracking utilizing a hybrid template/optical flow approach
verfasst von
Tom Williamson
Wa Cheung
Stuart K. Roberts
Sunita Chauhan
Publikationsdatum
05.06.2018
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 10/2018
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1780-0

Weitere Artikel der Ausgabe 10/2018

International Journal of Computer Assisted Radiology and Surgery 10/2018 Zur Ausgabe

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.