Open Access 01.12.2003 | Methodology article

Performance of plate-based cytokine flow cytometry with automated data analysis

verfasst von: Maria A Suni, Holli S Dunn, Patricia L Orr, Rian de Laat, Elizabeth Sinclair, Smita A Ghanekar, Barry M Bredt, John F Dunne, Vernon C Maino, Holden T Maecker

Erschienen in: BMC Immunology | Ausgabe 1/2003

Abstract

Background

Cytokine flow cytometry (CFC) provides a multiparameter alternative to ELISPOT assays for rapid quantitation of antigen-specific T cells. To increase the throughput of CFC assays, we have optimized methods for stimulating, staining, and acquiring whole blood or PBMC samples in 96-well or 24-well plates.

Results

We have developed a protocol for whole blood stimulation and processing in deep-well 24- or 96-well plates, and fresh or cryopreserved peripheral blood mononuclear cell (PBMC) stimulation and processing in conventional 96-well round-bottom plates. Samples from both HIV-1-seronegative and HIV-1-seropositive donors were tested. We show that the percent response, staining intensity, and cell recovery are comparable to stimulation and processing in tubes using traditional methods. We also show the equivalence of automated gating templates to manual gating for CFC data analysis.

Conclusion

When combined with flow cytometry analysis using an automated plate loader and an automated analysis algorithm, these plate-based methods provide a higher throughput platform for CFC, as well as reducing operator-induced variability. These factors will be important for processing the numbers of samples required in large clinical trials, and for epitope mapping of patient responses.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2172-4-9) contains supplementary material, which is available to authorized users.

Authors' contributions

MAS, HSD, PLO, RDL, and ES did the experiments. SAG designed the automated gating tools and edited the manuscript. BMB, JFD, and VCM supervised the work and edited the manuscript. HTM prepared figures and co-wrote the paper with MAS.

Background

Cytokine flow cytometry (CFC) assays use short-term in vitro stimulation and intracellular cytokine staining to quantitate potentially rare populations of antigen-specific T cells (see [1, 2] for reviews). Because of the multi-parametric readout of flow cytometry, CFC is an attractive alternative to ELISPOT assays, allowing the dissection of responses by various phenotypes of T cells [3, 4]. However, ELISPOT assays have traditionally been more amenable to high-throughput analysis due to their plate-based nature and the availability of automated instrumentation to analyze ELISPOT plates.
In order to provide similar advantages to the CFC platform, we optimized the ability to stimulate, process, and acquire CFC samples in 96-well or 24-well plates, using stimulation with the superantigen Staphylococcal enterotoxin B (SEB), as well as cytomegalovirus (CMV) or HIV antigens. The result was two protocols optimized for stimulation and processing of whole blood samples (in deep-well 24- or 96-well plates), and one protocol optimized for stimulation and processing of PBMC samples (in conventional round-bottom 96-well plates). These optimized protocols compared favorably with tube-based methods, and, when combined with automated acquisition and analysis software, resulted in a rapid and highly standardized assay with greatly reduced hands-on time.

Results and Discussion

Comparison of CFC responses in tubes and plates

Both whole blood and PBMC samples are frequently used for CFC assays. While conditions need to be optimized for these two sample types [58], comparable results can be obtained between whole blood and PBMC stimulations [9]. We therefore optimized protocols for each sample type. Processing of whole blood requires lysis of erythrocytes, so deep-well plates were employed to allow for addition of the required volume of lysis buffer. Whole blood samples from HIV-1-seropositive donors can potentially have lowered CD4+ T cell counts, so 24-well deep-well plates were used. These can accommodate up to 1 ml of blood per sample, plus up to 9 ml of lysis buffer.
A representative comparison for each type of assay is shown in Figure 1. Whole blood (200 μl per sample) was stimulated in conical-bottom 96-well deep-well plates. PBMC (106 cells in 200 μl) were stimulated in conventional round-bottom 96-well plates. Finally, whole blood (1 ml per sample) from an HIV-1-seropositive donor was stimulated in round-bottom 24-well deep-well plates. Each sample was stimulated for 6 hours in the presence or absence of a CMV pp65 or HIV-1 Gag p55 peptide mixture. Replicate samples for each assay type were incubated in parallel in 15-ml conical polypropylene tubes, using a standard CFC protocol [5, 6]. As seen in Figure 1, the plate and tube assays give qualitatively and quantitatively similar results.
In order to more accurately quantify the percentage of antigen-specific T cells identified by plate-based versus tube-based assays, a larger number of donors were tested by each method used in Figure 1. The results (Figure 2) show a close correlation between tube- and plate-based assays for the percentage of cytokine-producing cells. Strong correlations were seen for both SEB and CMV or HIV peptide stimulation, and in both CD4 and CD8-gated T cells. The slope of the linear regression lines approximated 1, indicating no particular bias toward higher results in either tubes or plates.

Comparison of staining intensity in tubes and plates

Next, we compared the mean fluorescence intensity of cytokine staining in tube-based versus plate-based assays, using the same samples as above. As seen in Figure 3, the staining intensity for IFNγ compared favorably between tubes and plates, with plates yielding similar or even higher fluorescence intensities than tubes. This was true for both whole blood (deep-well) and PBMC (shallow-well) assays. The differences in the ratio of tube-to-plate staining intensity could be related to staining volume, which is a function of the residual volume after aspiration. This volume tends to be lowest in shallow-well 96-well plates (using our multichannel aspiration manifolds), and highest in 24-well plates.

Effect of overnight resting on cryopreserved PBMC

For cryopreserved PBMC, we tested the effect of an incubation period after thawing but before antigen stimulation in 96-well plates. While we had previously reported that such a "rest" period did not significantly affect the result (percent of cytokine-positive cells) [10], we now show that it does affect the fluorescence intensity of cytokine-positive cells. As shown in Figure 4, we found that IFNγ staining intensity was significantly increased after an overnight incubation of thawed cells in medium, prior to stimulation with CMV pp65 peptide mix. This increased cytokine staining intensity correlated with an increase in functional avidity as measured by responses to lower concentrations of peptide antigen in titration experiments (data not shown).

Comparison of cell recovery in tubes and plates

We also compared cell recoveries in tube- versus plate-based assays. Cell loss in CFC assays is likely to be associated with centrifugation and washing. By using a 12-channel vacuum manifold with custom-length prongs (see Methods), we were able to aspirate supernatants after centrifugation in a rapid and reproducible manner, without risk of aspirating the cell pellets. As such, cell recoveries for both PBMC and whole blood were similar in plates and tubes (Figure 5). For PBMC, recoveries in plates appeared to be slightly better than in tubes, perhaps because of the need to incubate PBMC tubes on a slant [7], and associated difficulties with recovering adherent cells from the side of the tube. Also, in the tube protocol, cells are transferred from a 15-ml stimulation tube to 5-ml staining tubes, which may account for some cell loss. While 5-ml polystyrene tubes can be used for stimulation, they result in reduced cell recovery and slightly reduced activation efficiency with protein antigens (our unpublished data). The plate protocol has the advantage that cells are stimulated, processed, and acquired in a single plate.

Automated gating algorithm

In addition to errors that may have been introduced by differential washing, another frequent source of variation in CFC assays is analysis. Relatively minor differences in gating can quantitatively affect the results [2], and the prospect of subjectivity in gating is a concern for many investigators. Analysis also tends to be time-consuming using most conventional software packages. However, using software that contains a cluster-finding algorithm, we were able to build templates for automated CFC gating that identify the required cell populations and report a percentage of CD69-positive, cytokine-positive cells. Samples could then be processed by these templates using batch analysis, and results automatically exported to a spreadsheet. By using the cluster-finding algorithm, much of the subjectivity in gating is avoided.
An example of such a template is shown in Figure 6A. When compared to expert manual gating, the automated template of Figure 6A yielded highly correlated results for a set of 25 whole blood samples stimulated with CMV pp65 peptide mix or SEB (Figure 6B).
The template of Figure 6A also yielded good correlation with manual gating for a set of 39 HIV-1-seropositive whole blood samples stimulated with CMV pp65 peptide mix, HIV p55 peptide mix, or SEB (Figure 7A). However, it was noted that for some CD4+ responses in the HIV-1-seropositive dataset, the automated template undercounted IFNγ+ cells relative to manual gating, resulting in a slope of slightly less than 1 for the linear regression line (Figure 7A, CMV pp65 and SEB responses). In particular, samples with very few CD4+ T cells tended to result in undercounting of IFNγ+ cells by the automated template, due to suboptimal CD4+ cluster identification. Such samples were better analyzed using a template that created a fixed region for CD3+CD4+ gating, tethered to the CD3+CD4- population (Figure 7B). This fixed region was set to avoid the exclusion of CD4dim cells when the total number of CD4+ cells was very low. When this revised template was applied to the determination of CD4+ responses for the HIV-1-seropositive dataset, the resulting linear regressions yielded higher slopes, very close to 1 (Figure 7C). Thus, changes in the gating scheme employed by the automated template can improve the quality of data generated under certain circumstances. In particular, samples containing very few CD4+ cells may be best handled by the modified template of Figure 7B.
Another alternative to the template of Figure 6A that was advantageous in certain circumstances involved changing the initial gating of lymphocytes. Datasets containing poorly defined lymphocyte clusters were best gated by using an initial automated region to identify CD3+ lymphocytes in a CD3 versus side scatter plot, rather than using forward versus side scatter gating (data not shown).
The key features of any of these automated templates include: (a) the ability to track populations that "roam" as a result of donor, reagent, or instrument variations; (b) the ability to create a maximally sized region that will include activated T cells that have partially down-modulated the marker in question (CD3, CD4, and/or CD8); and (c) the ability to tether a manually drawn region to another population cluster to allow identification of potentially rare populations of cells (i.e., cytokine-positive cells).

Gains in assay throughput and standardization

As a result of applying the above methodology, the throughput of CFC assays is increased in a manner proportional to the number of samples per experiment. For example, a complete plate of 96 samples can be processed with significantly less time than an equivalent number of tubes (saving perhaps 1–2 hours of pipetting time). Since acquisition on a plate loader can be unsupervised, this saves an additional 3–5 hours of technician time. Finally, batch analysis with an automated template might save at least one more hour of technician time as compared to manual analysis.
The other major benefit of the methodology described here is standardization. The more rapid sample handling associated with plate-based methods would be expected to reduce sample-to-sample differences in incubation times for each processing step. Thus, variability in results associated with these time differences should also be reduced. Perhaps more significantly, automated analysis using a standard template should standardize differences in gating that occur between users. It is known that gating differences can make relatively large contributions to result variability [2]. In particular, the exclusion of cells with down-modulation of CD4 or CD8 staining due to activation can negatively impact the accurate reporting of cytokine-positive cells. The automated templates described above have been engineered to include these down-modulated cells in a reproducible fashion.

Comparison to ELISPOT assays

By providing plate compatibility and a degree of automation to CFC assays, they become attractive relative to competing platforms such as ELISPOT. These two assays both measure cytokine production, but CFC can provide multiparameter information that clearly distinguishes T cell subsets such as CD4 and CD8 as well as a number of other phenotypic markers on the responding cells. While ELISPOT assays have been reported to have lower limits of detection than CFC assays [11], CFC generally reports higher numbers of positive cells [9, 1214]. Thus, plate-based CFC assays can provide the benefits of multiparameter information, with higher efficiency of detection and comparable throughput to ELISPOT assays.

Conclusions

We conclude that CFC assays can be performed with increased throughput using plate-based methodology. By optimizing the type of plate used for either PBMC or whole blood, we are able to activate, process, and acquire samples in a single plate. This avoids labor and cell loss associated with sample transfers. We also show that data analysis can be streamlined using an automated template that uses cluster-finding algorithms. The data generated from such semi-automated plate-based assays are equivalent to data generated using tube-based methods and manual analysis. In addition, these plate-based methods and automated analysis templates are expected to increase standardization of CFC, such as would be required for its use in large clinical trials. Other applications include epitope mapping using a matrix of overlapping peptide pools [15], where high throughput and an array format are desirable features.

Methods

For additional details of procedures and notes on critical parameters, please see reference [16].

Stimulation reagents

SEB (Sigma Chemical Co., St. Louis, MO) was dissolved in sterile PBS at 0.5 mg/ml and stored at 4°C. It was used at a final concentration of 1 μg/ml of blood or PBMC suspension. CMV pp65 and HIV p55 gag peptide mixes (15 amino acid residues in length, overlapping by 11 amino acid residues each) have been described previously [10]. They were stored in small aliquots at -80°C, and diluted in sterile PBS on the day of use to achieve a final concentration of approximately 2 μg/ml/peptide. In some experiments, costimulatory antibodies to CD28 and CD49d (FastImmune, BD Biosciences, San Jose, CA) were added at a final concentration of 1 μg/ml each [8]. Brefeldin A (FastImmune or Sigma) was stored in small aliquots at -20°C and diluted in sterile PBS on the day of use to achieve a final concentration of 10 μg/ml.

Plates and accessories

Whole blood was activated in deep-well polypropylene plates, using 96-well plates (BD Discovery Labware, Bedford, MA; or E&K Scientific, Campbell, CA) for 200 μl samples, or 24-well plates (Qiagen, Hilden, Germany) for up to 1 ml samples. PBMC were activated in standard 96-well round-bottom polystyrene tissue culture plates (Falcon, BD Discovery Labware). 12-channel vacuum manifolds with 7 mm prongs (for standard plates) or 35 mm prongs (for deep-well plates) were purchased from V&P Scientific, San Diego, CA. Centrifuge holders compatible with deep-well plates were purchased from Sorvall Instruments, Newtown, CT.

Collection and cryopreservation of PBMC

Informed consent (approved by either the UCSF or the BD Biosciences Institutional Review Board) was provided by all volunteer blood donors. For PBMC preparation, whole blood was collected in Cell Preparation Tubes (CPT™) containing sodium heparin (BD Vacutainer, Franklin Lakes, NJ). PBMC were isolated following the manufacturer's instructions, and washed with RPMI-1640 medium containing 10% heat-inactivated fetal bovine serum and antibiotic/antimycotic solution (cRPMI-10, all components from Sigma). Fresh PBMC were resuspended in a volume of cRPMI-10 equivalent to the original blood volume. In some experiments, PBMC were resuspended for cryopreservation at 2 × 107 cells/ml in RPMI-1640+12.5% human serum albumin (Sigma). An equal volume of cold RPMI-1640 medium+10% human serum albumin+20% dimethyl sulfoxide (Sigma) was slowly added. Cells were transferred to freezing vials (1 ml/vial) and placed at -80°C in a freezing container (Mr. Frosty, Nunc, Naperville, IL). Cryopreserved cells were thawed briefly in a 37°C water bath, then 1 ml of warm (37°C) cRPMI-10 medium was added dropwise to the vial, and the cell suspension transferred to a 50 ml conical polypropylene tube (Falcon) containing 8 ml of warm cRPMI-10 medium. The cells were centrifuged for 7 minutes at 250 × G, then resuspended in cRPMI-10 at 5 × 106 cells/ml.

CFC of tube-based specimens

For comparison to plate assays, heparinized whole blood or PBMC in cRPMI-10 were activated in 15 ml conical polypropylene tubes (Falcon) and processed in 12 × 75 mm round-bottom polystyrene tubes (Falcon). A standard method, previously described [5, 6], was followed.

Cell activation in plates

200 μl of heparinized whole blood or PBMC in cRPMI-10 medium were plated per well in 96-well plates; 0.5 ml or 1 ml of whole blood was plated per well in 24-well plates. Cryopreserved PBMC were then rested at 37°C overnight, while fresh PBMC or whole blood samples were activated immediately. Activation reagents as described above were prepared in a master mix to allow their combined addition to each well in a volume of 20 μl per 200 μl of cell suspension. The cells were incubated at 37°C for 6 hours. In some experiments, the plates were then held overnight at 18°C, using a programmable water bath.

Cell processing in plates

20 μl of 20 mM EDTA were added per 200 μl of cell suspension, and incubated for 15 minutes at room temperature, followed by vigorous pipetting to dislodge adherent cells. For PBMC, the plates were then centrifuged at 250 × G for 5 minutes, and supernatants aspirated using a vacuum manifold (V&P Scientific, see above). Cells were resuspended in 100 μl/well of FACS Lysing Solution (BD Biosciences). For whole blood, 8–10 volumes of FACS Lysing Solution were added directly to the EDTA-treated blood. For both PBMC and whole blood, cells were incubated in FACS Lysing Solution for 10 minutes at room temperature, then centrifuged for 5 minutes at 500 × G. Supernatants were aspirated using an appropriate-length vacuum manifold (described above). Cells were resuspended in FACS Permeabilizing Solution 2 (BD Biosciences) (200 μl for PBMC; 1 ml for deep-well 96-well plates; 2 ml for 24-well plates). After an additional 10 minute incubation at room temperature, sample wells of deep-well plates were filled with wash buffer (PBS+0.5% bovine serum albumin+0.1% NaN3) and all plates were centrifuged for 5 minutes at 500 × g. Supernatants were aspirated as above, and the cells washed a second time. PBMC plates were subjected to a third wash, then all samples were resuspended in 20 μl of antibody cocktail containing anti-IFNγ FITC/CD69 PE/CD4 or CD8 PerCP-Cy5.5/CD3 APC (FastImmune, BD Biosciences). In experiments of Figure 7, CD4 PE-Cy5 (Beckman Coulter, Miami, FL) was used in place of CD4 PerCP-Cy5.5. Samples were incubated for 60 minutes at room temperature in the dark, then washed twice as above before resuspending in a final volume of 200 μl of 1% paraformaldehyde in PBS (250 μl for 24-well plates).

Flow cytometric analysis

Samples were acquired within 24 h of staining using a FACSCalibur flow cytometer and CellQuest Pro software (BD Biosciences). Plate-based samples were acquired on the FACSCalibur using a Multiwell Autosampler and Multiwell Plate Manager software (BD Biosciences). In general, 40,000 CD4+ or CD8+ events were collected. For plate-based samples, sample volume was set to 200 μl and mixing volume to 100 μl in Multiwell Plate Manager. Acquisition criteria were set to 40,000 events of interest (CD3+CD4+ or CD3+CD8+ lymphocytes), or 180 seconds. Manual analysis was done in CellQuest Pro or FloJo software (Tree Star, Inc., San Carlos, CA) by gating on both small lymphocytes and CD3+CD4+ (or CD3+CD8+) cells. A gated dot plot displaying CD69 versus IFNγ from an SEB-stimulated sample was then used to set a "response region" around double-positive cells. This response region was then applied to all samples to determine the percentage of cytokine-positive cells. Automated analysis was done as described in Figure 6A or 7B, using "Snap-to" gating and tethered region tools in CellQuest Pro 5.0.1.

Determination of cell recovery

The absolute counts of CD3+ T cells in whole blood and PBMC were determined using TruCOUNT Control beads and TriTest CD4 FITC/CD8 PE/CD3 PerCP antibody cocktail (BD Biosciences). For the samples representing 100% cell recovery, 50 μl of uncultured whole blood or PBMC were aliquoted by reverse pipetting into 12 × 75 mm polystyrene tubes containing 20 μl of TriTest CD4 FITC/CD8 PE/CD3 PerCP. Each cell sample was stained in triplicate. The samples were mixed well and incubated for 30 minutes at room temperature in the dark, and then lysed/fixed using 450 μl of FACS Lysing Solution. After a 15-minute incubation at room temperature, 50 μl of TruCOUNT "Low" or "Medium" beads were added by reverse pipetting. The samples were mixed well and acquired immediately using a FACSCalibur flow cytometer.
CFC samples were also stained in triplicate. EDTA-treated activated whole blood in deep-well plates and tubes (200 μl) or 24-well plates (1000 μl) and PBMC in shallow-well plates and tubes (200 μl) were lysed/fixed and permeabilized as described above. The samples were stained with 20 μl of TriTest CD4 FITC/CD8 PE/CD3 PerCP for 30 minutes at room temperature in the dark, washed twice and resuspended for acquisition in 200 μl of wash buffer. TruCOUNT "Low" or "Medium" beads were then added by reverse pipetting at 50 μl per sample. The samples were mixed well and acquired immediately.
Samples were acquired on a FACSCalibur flow cytometer using an FL3 (for CD3 PerCP) threshold. The CD3+ lymphocytes were identified in FL3 vs SSC and region R1 was drawn to include all CD3+ cells. The acquisition was carried out until 10,000 CD3+ events were collected, however, all events passing the FL3 threshold were stored in order to include the beads, which were then identified and gated (R2) in FL1 vs FL2. The absolute numbers of CD3+ lymphocytes were calculated as follows: (# of events in R1)/(# of events in R2) × (# of beads per test)/(test volume).

Acknowledgements

The authors thank C. Lorrie Epling and Joseph M. McCune for helpful discussions. This work was supported in part by grants from the National Institutes of Health, MO1 RR00083, P30 AI27763, R01 AI47062; U54 CA90818; and from the California AIDS Research Center, CC99-SF-001.

Authors' contributions

MAS, HSD, PLO, RDL, and ES did the experiments. SAG designed the automated gating tools and edited the manuscript. BMB, JFD, and VCM supervised the work and edited the manuscript. HTM prepared figures and co-wrote the paper with MAS.
Literatur
1.
Zurück zum Zitat Ghanekar SA, Maecker H: Cytokine flow cytometry: multiparameter approach to immune function analysis. Cytotherapy. 2003, 5: 1-6. 10.1080/14653240310000029.CrossRefPubMed Ghanekar SA, Maecker H: Cytokine flow cytometry: multiparameter approach to immune function analysis. Cytotherapy. 2003, 5: 1-6. 10.1080/14653240310000029.CrossRefPubMed
2.
Zurück zum Zitat Maecker HT, Maino VC: Flow cytometric analysis of cytokines. Manual of Clinical Laboratory Immunology, 6th ed. Edited by: Hooks J. 2002, Washington, DC, ASM Press Maecker HT, Maino VC: Flow cytometric analysis of cytokines. Manual of Clinical Laboratory Immunology, 6th ed. Edited by: Hooks J. 2002, Washington, DC, ASM Press
3.
Zurück zum Zitat Maecker HT, Maino VC, Picker LJ: Immunofluorescence analysis of T-cell responses in health and disease. J Clin Immunol. 2000, 20: 391-399. 10.1023/A:1026403724413.CrossRefPubMed Maecker HT, Maino VC, Picker LJ: Immunofluorescence analysis of T-cell responses in health and disease. J Clin Immunol. 2000, 20: 391-399. 10.1023/A:1026403724413.CrossRefPubMed
4.
Zurück zum Zitat Shacklett BL: Beyond 51Cr release: new methods for assessing HIV-1-specific CD8+ T cell responses in peripheral blood and mucosal tissues. Clin Exp Immunol. 2002, 130: 172-182. 10.1046/j.1365-2249.2002.01981.x.PubMedCentralCrossRefPubMed Shacklett BL: Beyond 51Cr release: new methods for assessing HIV-1-specific CD8+ T cell responses in peripheral blood and mucosal tissues. Clin Exp Immunol. 2002, 130: 172-182. 10.1046/j.1365-2249.2002.01981.x.PubMedCentralCrossRefPubMed
5.
Zurück zum Zitat Suni MA, Picker LJ, Maino VC: Detection of antigen-specific T cell cytokine expression in whole blood by flow cytometry. J Immunol Methods. 1998, 212: 89-98. 10.1016/S0022-1759(98)00004-0.CrossRefPubMed Suni MA, Picker LJ, Maino VC: Detection of antigen-specific T cell cytokine expression in whole blood by flow cytometry. J Immunol Methods. 1998, 212: 89-98. 10.1016/S0022-1759(98)00004-0.CrossRefPubMed
6.
Zurück zum Zitat Nomura LE, Walker JM, Maecker HT: Optimization of whole blood antigen-specific cytokine assays for CD4(+) T cells. Cytometry. 2000, 40: 60-68. 10.1002/(SICI)1097-0320(20000501)40:1<60::AID-CYTO8>3.0.CO;2-J.CrossRefPubMed Nomura LE, Walker JM, Maecker HT: Optimization of whole blood antigen-specific cytokine assays for CD4(+) T cells. Cytometry. 2000, 40: 60-68. 10.1002/(SICI)1097-0320(20000501)40:1<60::AID-CYTO8>3.0.CO;2-J.CrossRefPubMed
7.
Zurück zum Zitat Waldrop SL, Pitcher CJ, Peterson DM, Maino VC, Picker LJ: Determination of antigen-specific memory/effector CD4+ T cell frequencies by flow cytometry: evidence for a novel, antigen-specific homeostatic mechanism in HIV-associated immunodeficiency. J Clin Invest. 1997, 99: 1739-1750.PubMedCentralCrossRefPubMed Waldrop SL, Pitcher CJ, Peterson DM, Maino VC, Picker LJ: Determination of antigen-specific memory/effector CD4+ T cell frequencies by flow cytometry: evidence for a novel, antigen-specific homeostatic mechanism in HIV-associated immunodeficiency. J Clin Invest. 1997, 99: 1739-1750.PubMedCentralCrossRefPubMed
8.
Zurück zum Zitat Waldrop SL, Davis KA, Maino VC, Picker LJ: Normal human CD4+ memory T cells display broad heterogeneity in their activation threshold for cytokine synthesis. J Immunol. 1998, 161: 5284-5295.PubMed Waldrop SL, Davis KA, Maino VC, Picker LJ: Normal human CD4+ memory T cells display broad heterogeneity in their activation threshold for cytokine synthesis. J Immunol. 1998, 161: 5284-5295.PubMed
9.
Zurück zum Zitat Maino VC, Maecker HT: Cytokine Flow Cytometry: A Multiparametric Approach For Assessing Cellular Immune Responses To Viral Antigens. Clin Immunol. 2003, Maino VC, Maecker HT: Cytokine Flow Cytometry: A Multiparametric Approach For Assessing Cellular Immune Responses To Viral Antigens. Clin Immunol. 2003,
10.
Zurück zum Zitat Maecker HT, Dunn HS, Suni MA, Khatamzas E, Pitcher CJ, Bunde T, Persaud N, Trigona W, Fu TM, Sinclair E, Bredt BM, McCune JM, Maino VC, Kern F, Picker LJ: Use of overlapping peptide mixtures as antigens for cytokine flow cytometry. J Immunol Methods. 2001, 255: 27-40. 10.1016/S0022-1759(01)00416-1.CrossRefPubMed Maecker HT, Dunn HS, Suni MA, Khatamzas E, Pitcher CJ, Bunde T, Persaud N, Trigona W, Fu TM, Sinclair E, Bredt BM, McCune JM, Maino VC, Kern F, Picker LJ: Use of overlapping peptide mixtures as antigens for cytokine flow cytometry. J Immunol Methods. 2001, 255: 27-40. 10.1016/S0022-1759(01)00416-1.CrossRefPubMed
11.
Zurück zum Zitat Helms T, Boehm BO, Asaad RJ, Trezza RP, Lehmann PV, Tary-Lehmann M: Direct visualization of cytokine-producing recall antigen-specific CD4 memory T cells in healthy individuals and HIV patients. J Immunol. 2000, 164: 3723-3732.CrossRefPubMed Helms T, Boehm BO, Asaad RJ, Trezza RP, Lehmann PV, Tary-Lehmann M: Direct visualization of cytokine-producing recall antigen-specific CD4 memory T cells in healthy individuals and HIV patients. J Immunol. 2000, 164: 3723-3732.CrossRefPubMed
12.
Zurück zum Zitat Kuzushima K, Hoshino Y, Fujii K, Yokoyama N, Fujita M, Kiyono T, Kimura H, Morishima T, Morishima Y, Tsurumi T: Rapid determination of Epstein-Barr virus-specific CD8(+) T-cell frequencies by flow cytometry. Blood. 1999, 94: 3094-3100.PubMed Kuzushima K, Hoshino Y, Fujii K, Yokoyama N, Fujita M, Kiyono T, Kimura H, Morishima T, Morishima Y, Tsurumi T: Rapid determination of Epstein-Barr virus-specific CD8(+) T-cell frequencies by flow cytometry. Blood. 1999, 94: 3094-3100.PubMed
13.
Zurück zum Zitat Moretto WJ, Drohan LA, Nixon DF: Rapid quantification of SIV-specific CD8 T cell responses with recombinant vaccinia virus ELISPOT or cytokine flow cytometry. Aids. 2000, 14: 2625-2627. 10.1097/00002030-200011100-00034.CrossRefPubMed Moretto WJ, Drohan LA, Nixon DF: Rapid quantification of SIV-specific CD8 T cell responses with recombinant vaccinia virus ELISPOT or cytokine flow cytometry. Aids. 2000, 14: 2625-2627. 10.1097/00002030-200011100-00034.CrossRefPubMed
14.
Zurück zum Zitat Karlsson AC, Martin JN, Younger SR, Bredt BM, Epling L, Ronquillo R, Varma A, Deeks SG, McCune JM, Nixon DF, Sinclair E: Comparison of the ELISPOT and cytokine flow cytometry assays for the enumeration of antigen-specific T cells. J Immunol Methods. 2003, Karlsson AC, Martin JN, Younger SR, Bredt BM, Epling L, Ronquillo R, Varma A, Deeks SG, McCune JM, Nixon DF, Sinclair E: Comparison of the ELISPOT and cytokine flow cytometry assays for the enumeration of antigen-specific T cells. J Immunol Methods. 2003,
15.
Zurück zum Zitat Kern F, Surel IP, Faulhaber N, Frommel C, Schneider-Mergener J, Schonemann C, Reinke P, Volk HD: Target structures of the CD8(+)-T-cell response to human cytomegalovirus: the 72-kilodalton major immediate-early protein revisited. J Virol. 1999, 73: 8179-8184.PubMedCentralPubMed Kern F, Surel IP, Faulhaber N, Frommel C, Schneider-Mergener J, Schonemann C, Reinke P, Volk HD: Target structures of the CD8(+)-T-cell response to human cytomegalovirus: the 72-kilodalton major immediate-early protein revisited. J Virol. 1999, 73: 8179-8184.PubMedCentralPubMed
16.
Zurück zum Zitat Maecker HT: Cytokine flow cytometry. Flow Cytometry Protocols. Edited by: Hawley T S and Hawley R G. 2003, Humana Press, 2nd Maecker HT: Cytokine flow cytometry. Flow Cytometry Protocols. Edited by: Hawley T S and Hawley R G. 2003, Humana Press, 2nd
Metadaten
Titel
Performance of plate-based cytokine flow cytometry with automated data analysis
verfasst von
Maria A Suni
Holli S Dunn
Patricia L Orr
Rian de Laat
Elizabeth Sinclair
Smita A Ghanekar
Barry M Bredt
John F Dunne
Vernon C Maino
Holden T Maecker
Publikationsdatum
01.12.2003
Verlag
BioMed Central
Erschienen in
BMC Immunology / Ausgabe 1/2003
Elektronische ISSN: 1471-2172
DOI
https://doi.org/10.1186/1471-2172-4-9

Weitere Artikel der Ausgabe 1/2003

BMC Immunology 1/2003 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Reizdarmsyndrom: Diäten wirksamer als Medikamente

29.04.2024 Reizdarmsyndrom Nachrichten

Bei Reizdarmsyndrom scheinen Diäten, wie etwa die FODMAP-arme oder die kohlenhydratreduzierte Ernährung, effektiver als eine medikamentöse Therapie zu sein. Das hat eine Studie aus Schweden ergeben, die die drei Therapieoptionen im direkten Vergleich analysierte.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.