Phantom validation of 17 and 11 heartbeat MOLLI T1 mapping sequence at 3T

verfasst von: Songtao Liu, Justin Huang, Martin Ugander, Christopher Sibley, Abiola Oki, Jing Han, Andreas Greiser, Daniel R Messroghli, Peter Kellman, Andrew E Arai, David A Bluemke

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2011

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background and objective

The Modified Look-Locker Inversion-Recovery (MOLLI) sequence (Messroghli, et al, JMRI, 2007) was optimized for myocardial T1 mapping on 1.5T. However, little data exists on T1 mapping at 3T. The standard MOLLI sequence uses three inversion-recovery blocks to acquire 11 images over 17 heartbeats (HB). The long breath hold could limit its clinical application in patients with cardiorespiratory compromise. A new 11 HB MOLLI protocol with two inversion-recovery blocks was introduced recently, which reduces scan time by 35%. This aim of this study is to verify both 17 HB and 11 HB MOLLI sequences at 3T.

Methods

Agarose gel-cupric sulfate phantoms were scanned using both 17 HB and 11 HB MOLLI sequence at 3T (Verio, Siemens) with simulated heart rates from 40-110 beats per minute (bpm). An inversion-recovery spin-echo (IR-SE) sequence with TR of 10sec, TI varying from 10ms to 2500ms was used in the same configuration. The phantoms were also scanned using a spin-echo sequence with varying TE for T2 quantification. T1 maps were generated by MRMap (Messroghli, BMC Medical Imaging, 2010) and a region of interest inside each phantom was measured by ImageJ. Results were compared to each other and to IR-SE, which served as the reference standard. The phantoms studied had T1 times that ranged from 174-2660ms and T2 times ranging from 51-188ms

Results

T1 estimates by both 17HB and 11HB MOLLI matched measurements by IR-SE up to ~500ms, but underestimated longer T1 times. This underestimation increased for higher heart rates (see figure). After correction, the difference between 17HB MOLLI and IR-SE was reduced to -0.8+/-2.3%, and difference between 11HB MOLLI and IR-SE was -0.4%+/-2.2%. The mean difference between 17HB MOLLI and 11HB MOLLI was 0.4+/-1.2% at 60 bpm, and this was similar at other heart rates.

Conclusions

T1 mapping by MOLLI is feasible and accurate at 3T. Calibrated heart rate correction is necessary for T1 times longer than 500ms. Within our sample, 11HB MOLLI was an equally accurate estimator of IR-SE compared to 17HB MOLLI after correction was applied. Based on this, 17HB MOLLI and 11HB MOLLI can be used interchangeably.
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Metadaten
Titel
Phantom validation of 17 and 11 heartbeat MOLLI T1 mapping sequence at 3T
verfasst von
Songtao Liu
Justin Huang
Martin Ugander
Christopher Sibley
Abiola Oki
Jing Han
Andreas Greiser
Daniel R Messroghli
Peter Kellman
Andrew E Arai
David A Bluemke
Publikationsdatum
01.12.2011
Verlag
BioMed Central
DOI
https://doi.org/10.1186/1532-429X-13-S1-P10

Weitere Artikel der Sonderheft 1/2011

Journal of Cardiovascular Magnetic Resonance 1/2011 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.