Conventionally Tissue Phase Mapping (TPM) is carried out in one 2D slice. 4D PCMR has not previously been used for TPM due to long acquisition times. The purpose of this study was to develop a highly accelerated 4D PCMR sequence to acquire TPM data across the entire left ventricle (LV), during a breath-hold.
Methods
An undersampled, prospectively triggered, stack-of-spirals, 4D PCMR sequence was developed, and reconstructed using SENSE-UNFOLD. A uniform-density spiral trajectory was used in kx-ky with 8 interleaves required to fill k-space, with acceleration of R = 4 (resulting in 2 spiral interleaves acquired per kz position). In kz, 16 slices were used, with acceleration of R = 2, resulting in 9 kz positions being acquired for each cardiac phase (calculated as 16÷2 plus one to ensure that the middle kz position is always acquired). The other sequence parameters were; TE/TR:4.0/14.4 ms, matrix size:160 × 160, FOV:500 × 500 mm, slices:16, slice thickness:8 mm, venc(x/y/z):30/30/30 cm/s. This resulted in a temporal resolution of 57.5 ms, a spatial resolution of 3.1 × 3.1 × 8 mm, and a scan time of 16 heartbeats (breath-hold), plus 7.3 s for acquisition of fully sampled coil sensitivity maps with 4NSAs (free breathing, at the end of the TPM acquisition). This 4D PCMR TPM data was acquired in 5 healthy volunteers (4M:1F, mean age:38.6 ± 6.6 yrs), and 1 patient with mild LV dyssynchrony (F, 43.3 yrs). Data Analysis: Bulk motion correction was performed, before transformation of the velocities to an internal polar coordinate system allowing motion to be described in terms of contraction (using radial velocities; Vr) and shortening (using longitudinal velocities; Vz). Peak velocities in the S (systolic) and E (early diastolic) waves were measured for the longitudinal and radial velocities within the entire myocardium, at a basal, mid and apical slice.
Anzeige
Results
Data was successfully acquired and analyzed in all subjects. Figure 1 shows the calculated velocity data in one subject. In the volunteers quantitatively, the longitudinal velocities in the base/mid/apical slices were 4.9 ± 1.2/3.5 ± 1.3/2.4 ± 0.6 cm/s at the S wave, and -6.2 ± 1.3/-3.9 ± 1.4/-2.43 ± 1.5 cm/s at the E wave. The radial velocities in the base/mid/apical slices were 2.2 ± 0.4/2.7 ± 0.2/2.1 ± 0.2 cm/s at the S wave, and -2.5 ± 0.4/-3.1 ± 0.5/-2.9 ± 0.1 cm/s at the E wave. In the patient the longitudinal velocities in the base/mid/apical slices were significantly lower at the S wave (1.1/1.0/1.4 cm/s). Furthermore, there was evidence of dyssynchrony in early LV diastole as shown in Figure 2.
×
×
Conclusions
We have shown that using a novel spiral UNFOLD SENSE PCMR sequence it is possible to obtain 4D TPM data in a breath hold. This sequence may have significant value in the assessment of LV dyssynchrony, particularly in relation to resynchronization therapy.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.
Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.
Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.
Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.
Update Radiologie
Bestellen Sie unseren Fach-Newsletterund bleiben Sie gut informiert.