Erschienen in: Cardiovascular Ultrasound 1/2020

Open Access 01.12.2020 | Research

Fetal cardiac function at intrauterine transfusion assessed by automated analysis of color tissue Doppler recordings

verfasst von: Lotta Herling, Jonas Johnson, Kjerstin Ferm-Widlund, Fredrik Bergholm, Peter Lindgren, Sven-Erik Sonesson, Ganesh Acharya, Magnus Westgren

Erschienen in: Cardiovascular Ultrasound | Ausgabe 1/2020

Abstract

Background

Fetal anemia is associated with a hyperdynamic circulation and cardiac remodeling. Rapid intrauterine transfusion (IUT) of blood with high hematocrit and viscosity into the umbilical vein used to treat this condition can temporarily further affect fetal heart function. The aim of this study was to evaluate the short-term changes in fetal myocardial function caused by IUT using automated analysis of cine-loops of the fetal heart obtained by color tissue Doppler imaging (cTDI).

Methods

Fetal echocardiography was performed before and after IUT. cTDI recordings were obtained in a four-chamber view and regions of interest were placed at the atrioventricular plane in the left ventricular (LV), right ventricular (RV) and septal walls. Myocardial velocities were analyzed by an automated analysis software to obtain peak myocardial velocities during atrial contraction (Am), ventricular ejection (Sm), rapid ventricular filling (Em) and Em/Am ratio was calculated.
Myocardial velocities were converted to z-scores using published reference ranges. Delta z-scores (after minus before IUT) were calculated. Correlations were assessed between variables and hemoglobin before IUT.

Results

Thirty-two fetuses underwent 70 IUTs. Fourteen were first time transfusions. In the LV and septal walls, all myocardial velocities were significantly increased compared to normal values, whereas in the RV only Sm was increased before IUT (z-scores 0.26–0.52). In first time IUTs, there was a negative correlation between LV Em (rho = − 0.61, p = 0.036) and LV Em/Am (rho = − 0.82, p = 0.001) z-scores and hemoglobin before IUT. The peak myocardial velocities that were increased before IUT decreased, whereas LV Em/Am increased significantly after IUT.

Conclusions

This study showed that peak myocardial velocities assessed by cTDI are increased in fetuses before IUT reflecting the physiology of hyperdynamic circulation. In these fetuses, the fetal heart is able to adapt and efficiently handle the volume load caused by IUT by altering its myocardial function.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
IUT
Intrauterine transfusion
cTDI
Color tissue Doppler imaging
MCA PSV
Middle cerebral artery peak systolic velocity
MoM
Multiple of median
Hb
Hemoglobin
ROI
Region of interest
AV-plane
Atrioventricular plane
LV
Left ventricular/left ventricle
RV
Right ventricular/right ventricle
IVS
Interventricular septum
Am
Peak myocardial velocity during atrial contraction
Sm
Peak myocardial velocity during ventricular ejection
Em
Peak myocardial velocity during rapid ventricular filling
SD
Standard deviation
CI
Confidence interval
BMI
Body mass index
GA
Gestational age
Bpm
Beats per minute

Background

Chronic anemia increases the demands on the fetal heart to increase cardiac output and maintain adequate tissue oxygenation. This hyperdynamic cardiovascular adaptation is characterized by cardiomegaly, increased blood flow velocities, decreased vascular resistance and, if left untreated, cardiac dysfunction, hydrops and eventually death [14]. To prevent fetal deterioration, anemia is treated with intravascular intrauterine transfusions (IUT) [5, 6]. However, IUT temporarily also subjects the fetal heart to another type of strain when a large, high-viscosity volume is infused during a short period of time.
Current knowledge about cardiac function in anemic fetuses undergoing IUT is mainly derived from assessment of blood flow, whereas surprisingly little is known about the performance of the myocardium. Decreased cardiac output has been demonstrated following IUT when assessing blood flow, but data also indicate that the decrease might be transient [79]. Only a few studies have focused on how the myocardium and the pump function of the fetal heart is affected during anemia and IUT. Michel et al. showed an increase in peak systolic myocardial strain in anemic fetuses that partially normalized after IUT [10, 11].
Color tissue Doppler imaging (cTDI) is an echocardiographic technique evaluating myocardial motion that can be used to derive information on variables describing fetal cardiac function such as longitudinal myocardial velocities and cardiac cycle time intervals [1214]. Despite limitations, cTDI is a promising method for assessing fetal cardiac function. However, its routine clinical application is hampered by a cumbersome and time-consuming off-line analysis that has been partly solved through the use of an automated analysis of myocardial velocities [1517].
Thus, the aim of this study was to evaluate myocardial function before and after IUT using cTDI with an automated analysis software to identify physiological changes caused by short-term volume load in fetuses under-going transfusion.

Methods

Pregnant women undergoing IUTs were recruited from the Center for Fetal Medicine at Karolinska University Hospital between 2009 and 2016. Exclusion criteria were multiple pregnancy, major fetal structural/chromosomal abnormality or intraperitoneal transfusion. Gestational age was based on measurement of crown rump length or biparietal diameter in the first trimester or biparietal diameter in the second trimester according to local guidelines. The study was approved by the Stockholm Regional Ethics Committee (DNr 2012/895–31/4 and 2009/1617–31/2) and all women gave their informed consent.
Maternal characteristics, medical history and perinatal outcome were recorded. A transabdominal ultrasound examination was performed for the estimation of fetal weight and measurement of peak systolic velocity in the middle cerebral artery (MCA PSV) using a Voluson E8 ultrasound machine with a 2–5 MHz curved array transducer (GE Healthcare, Zipf, Austria). MCA PSV was recorded and adjusted to multiple of median (MoM) for each gestational age to non-invasively assess the likelihood of fetal anemia as described by Mari et al. [18, 19]. Fetal hydrops was defined as the presence of fluid in at least two body compartments, i.e. ascites, subcutaneous edema, pleural or pericardial effusion.
IUT was performed according to standard techniques and the local protocol for intravascular transfusions has been described previously [20]. Briefly, a 20-gauge needle was inserted into the umbilical vein, either in the intrahepatic portion or at the cord insertion in women with an anterior placenta. Fetal analgesia with alfentanil (0.015 mg/kg) intramuscularly with a 22-gauge needle and fetal paralysis with succinylcholine (5 mg/kg) were applied when transfusion included penetration of the fetal skin with the needle. No antibiotics or tocolysis were given to the mother. Fetal blood sampling was done and packed red blood cells with a hematocrit of 75–85% were transfused intravenously. The volume of blood given was based on fetal weight, estimated fetoplacental blood volume, donor hematocrit and severity of anemia [21, 22]. Before transfusion a complete blood count was obtained by a hematology instrument (PocH-100i, Sysmex Corp.). The hemoglobin value was transformed to multiple of median (Hb MoM). Fetal anemia was defined as mild (0.84–0.65 MoM), moderate (0.65–0.55 MoM) or severe (< 0.55 MoM) [19].
Fetal echocardiography was performed before and within 2.5 h after IUT (median 69 min, interquartile range 54–79) by two experienced operators (KFW and LH) using a GE Vivid-i ultrasound imaging system with a 3S-RS (1.9–3.8 MHz) phased array transducer (GE Vingmed Ultrasound AS, Horten, Norway) or a Vivid S6 ultrasound imaging system with a M4S-RS (1.9–4.1 MHz) phased array transducer (GE CV Ultrasound, Haifa, Israel). All cTDI recordings were obtained in an apical or basal four-chamber view and cine-loops of consecutive cardiac cycles were stored for off-line analysis. Acquisition was optimized to obtain a high frame rate and the incident angle was kept as close to the long-axis of the heart as possible (deviation always <30o). Off-line analysis was performed in EchoPAC version 201 (GE Vingmed Ultrasound AS, Horten, Norway). Regions of interest (ROIs) were placed at the atrioventricular plane (AV-plane) in the left (LV) and right ventricular (RV) walls and the interventricular septum (IVS) (Fig. 1). ROI size was adjusted to gestational age [15]. With the aim to reduce variability all ROIs were placed by one operator (LH) and positioned with the AV-plane reaching approximately one third into the ROI at end-systole.
The myocardial velocity traces were subsequently exported from EchoPAC to MATLAB (R2015a, MathWorks, MA, USA), as a text file covering all available velocity data and further analyzed by an automated algorithm also developed into an automated analysis software. This is a fully automated procedure and no manual marking of the traces is needed once the velocity traces are transferred to MATLAB. Moreover, the algorithm compares all identified heart beats and uses that information in identification of cardiac time events. It can also compare different segments, i.e. the IVS, LV and RV walls, over time for more exact definitions [2325].
The peak myocardial velocity during atrial contraction (Am), ventricular ejection/systole (Sm) and rapid ventricular filling/early diastole (Em) were automatically obtained (Fig. 1). The Em/Am ratio was calculated. Fusion of Em and Am can take place at high heart rates, and such fusion events are detected by the algorithm. When fusion events are detected in a cardiac cycle the Em and Am cannot be separated and the value is omitted. All cardiac cycles available were evaluated by the algorithm and an average from all available cardiac cycles calculated. All measurements were performed for the LV, RV and IVS separately.

Statistical analysis

Data analysis was performed using IBM SPSS Statistics for Windows, version 23.0 (IBM Corp., Armonk, N.Y., USA) and MATLAB. Continuous variables were presented as mean ± standard deviation (SD) or median (interquartile range) as appropriate. Categorical variables were presented as absolute values and percentage [n (%)].
All cTDI variables were converted to z-scores [(measurement – predicted mean)/ predicted SD] by using published gestational age specific reference data obtained from normal pregnancies by the same technique [17]. Delta (Δ) z-scores were calculated as (z-score after IUT - z-score before IUT).
Mean z-scores and 95% confidence intervals (CI) were presented on measurements before IUT in all IUTs to assess differences between values compared to reference data. Spearman’s correlation (rho) was calculated between the Hb and MCA PSV MoMs before IUT, on one hand, and cTDI z-scores on the other hand. Changes in variables before and after IUT were assessed using Wilcoxon signed rank test for related samples for first and all IUTs. Statistical significance was set to p < 0.05.

Results

This study included 32 singleton fetuses that underwent a total of 70 IUTs with a mean of 2.2 ± 1.7 (range, 1–8) transfusions per fetus. cTDI recordings were obtained both before and after IUT in 66, and in four only before IUT. Fourteen of the 70 IUTs were first time transfusions, whereof one was not examined after IUT. In one of these first transfusions there was no hemoglobin value available before IUT. For details see Fig. 2.
Demographic details and clinical characteristics of the study population, transfusions and pregnancy outcome are shown in Table 1. There were 27 fetuses with anemia due to alloimmunization undergoing 63 IUTs. The most common maternal antibody was anti-D (n = 22) of which nine in combination with another antibody. Five fetuses underwent seven transfusions due to other causes. One fetus had a suspected Diamond-Blackfan anemia that was postnatally confirmed. Postnatal investigation showed suspicion of Chronic Infantile Neurological and Auricular syndrome in one baby but this was not genetically verified [26]. In three babies the etiology of the anemia remained unknown and persisted postnatally in one case.
Table 1
Clinical characteristics of the study population (n = 32), intrauterine transfusions (n = 70) and pregnancy outcomes
Variable
Value
Maternal data
 Age (years)
31.2 ± 5.3
 BMI (kg/m2)
24.1 ± 3.3
 Nulliparous
4 (12.5)
 Immunizationa
27 (84.4)
IUT data
 Number of IUTs per fetus
2.2 ± 1.7
Site of tranfusion:
 Intrahepatic
41 (58.6)
 Cord insertion in placenta
29 (41.4)
 Mean GA at IUT (weeks)
28.6 ± 4.7
 Hemoglobin before IUT (g/L)
88.6 ± 28.7
 Hemoglobin after IUT (g/L)
141.7 ± 23.9
Degree of anemia: (n = 69)
 None
20 (29.0)
 Mild
23 (33.3)
 Moderate
9 (13.0)
 Severe
17 (24.6)
 Infused volume (mL)
47.1 ± 23.9
Pregnancy outcome
 GA at delivery (weeks)
35.7 ± 1.4
 Pre-term delivery < 37 weeks
28 (87.5)
Mode of delivery:
 Normal vaginal delivery
11 (34.4)
 Elective cesarean section
8 (25.0)
 Emergency cesarean section
13 (40.6)
 Birth weight (g)
2793 ± 443
 Female babies
15 (46.9)
 Cord arterial pH (n = 13)
7.27 ± 0.10
 5-min Apgar score < 7
2 (6.3)
Data are given as mean ± SD or n (%). BMI, body mass index; IUT, intrauterine transfusion; GA, gestational age
aFive fetuses with other causes to anemia
There were 28 preterm deliveries but only two women delivered before week 34 + 0. Fetal hydrops was present at three (4.3%) IUTs in three fetuses. Isolated ascites was present at eight (11.4%) IUTs in six fetuses.
Of the 136 ultrasound examinations performed before or after IUT, the position of the heart was apical in 59 (43%) scans and basal in 77 (57%) scans. The frame rate was 207.7 ± 13.1 frames/sec with a minimum of 173 frames/sec. When extreme outliers in cTDI variables were identified the individual cTDI recordings and velocity traces were visually inspected. Values were omitted if the trace was of insufficient quality for the functioning of the algorithm. This together with missing values due to fusion of Em and Am waves resulted in a mean of 68 observations per variable before IUT (range, 64–70). An average of 9.1 ± 1.7 cardiac cycles were analyzed per myocardial velocity trace.

Myocardial velocities before IUT compared to reference ranges

Myocardial velocities before all 70 IUTs are presented in Table 2. Am, Sm and Em were all significantly increased in the LV and IVS, whereas in the RV only Sm reached statistical significance compared to our reference standard. Details for myocardial velocities in the LV wall are illustrated in Fig. 3.
Table 2
Myocardial velocities before intrauterine transfusion (n = 70). Absolute values and z-scores
Variable
LV
IVS
RV
Mean (±SD)
Mean z-score (95% CI)
Mean (±SD)
Mean z-score (95% CI)
Mean (±SD)
Mean z-score (95% CI)
(cm/sec)
 
(cm/sec)
 
(cm/sec)
 
Myocardial velocities
 Am
4.27 (±1.46)
0.31a (0.06–0.56)
4.03 (±1.01)
0.28a (0.08–0.48)
5.83 (±1.46)
0.12 (−0.16–0.40)
 Sm
2.77 (±1.11)
0.34a (0.08–0.61)
2.84 (±0.68)
0.52a (0.28–0.77)
3.66 (±1.14)
0.30a (0.03–0.57)
 Em
3.55 (±1.19)
0.48a (0.24–0.73)
3.38 (±0.96)
0.26a (0.05–0.47)
4.20 (±1.28)
0.12 (−0.12–0.35)
Em/Am ratio
0.90 (±0.36)
0.20 (−0.05–0.44)
0.87 (±0.28)
−0.01 (− 0.21–0.18)
0.76 (±0.26)
0.09 (− 0.15–0.34)
Values are presented as absolute values and z-scores created from published reference data
aindicates significance according to 95% confidence interval (CI)
LV, left ventricular wall; IVS, interventricular septum; RV, right ventricular wall; Am, peak myocardial velocity during atrial contraction; Sm, peak myocardial velocity during ventricular ejection; Em, peak myocardial velocity during rapid ventricular filling; SD, standard deviation

Correlation between myocardial velocities and Hb and MCA PSV before IUT

When analyzing observations made before first IUTs in the 12 cases with available Hb values and cTDI measurements, we observed a negative correlation between Hb and LV Em (rho = − 0.61, p = 0.036) as well as LV Em/Am (rho = − 0.82, p = 0.001, Fig. 4). When analyzing all IUTs, weaker negative correlations were found between Hb and Sm, Em, Em/Am in LV and Sm in IVS (rho = − 0.25 - -0.49, p < 0.05).
Correlation analysis between z-scores of cTDI variables and MCA PSV before first IUTs showed a strong positive correlation between MCA PSV and LV Em (rho = 0.66, p = 0.03) as well as LV Em/Am (rho = 0.84, p = 0.001, Fig. 5). When analyzing all IUTs, weaker positive correlations were found with LV Em and Em/Am (rho = 0.34, 0.39, p < 0.05).

Comparison of myocardial velocities before and after IUT

Fetuses receiving their first IUTs showed a significant decrease in Am and Sm in the LV as well as Sm and Em in IVS following IUT with median Δ z-scores − 0.69- -1.50; p < 0.05.
In the group of fetuses receiving their first time transfusion that had Hb values as well as pre- and post-transfusion cTDI measurements recorded (n = 12), the median Hb MoM was 0.60 (range, 0.14–1.08). Three fetuses had mild anemia, six fetuses had severe anemia and three fetuses had no anemia. In the sub-group of anemic fetuses (n = 9) the only variables that showed a significant change after transfusion was Sm which decreased from 1.84 cm/s to 1.53 cm/s in the LV (p = 0.05) and 2.50 cm/s to 1.74 cm/s in the IVS (p = 0.017).
When assessing all IUTs with available data before and after IUT, we found a significant decrease in heart rate from 140.4 ± 9.1 bpm before to 132.7 ± 14.4 bpm (p = 0.001) following transfusion. All peak myocardial velocities except Am and Em in the RV decreased significantly after IUT (median Δ z-scores − 0.34- -0.97; p < 0.01). The LV Em/Am ratio increased after IUT (median Δ z-score + 0.58; p < 0.001) whereas the other ratios did not change. Table 3. Details for the LV wall are also illustrated in Fig. 3.
Table 3
Difference in myocardial velocities before and after intrauterine transfusion (n = 66). Absolute values and z-scores
Variable
LV
IVS
RV
Median b
(IQ)
Median b z-score
(IQ)
Median b
(IQ)
Median b z-score
(IQ)
Median b
(IQ)
Median b z-score
(IQ)
(cm/sec)
 
(cm/sec)
 
(cm/sec)
 
Myocardial velocities
 Am
−1.03 (1.52)
−0.75a (1.54)
− 0.62 (1.41)
− 0.61a (1.29)
−0.47 (2.36)
− 0.40 (1.73)
 Sm
−0.52 (1.13)
− 0.62a (1.30)
−0.49 (0.64)
− 0.97a (1.18)
−0.45 (1.23)
− 0.49a (1.45)
 Em
−0.41 (1.08)
− 0.38a (1.12)
−0.34 (1.14)
− 0.34a (1.16)
−0.11 (2.00)
− 0.13 (1.61)
 Em/Am ratio
+ 0.23 (0.55)
+ 0.58a (1.43)
+ 0.05 (0.50)
+ 0.19 (1.55)
+ 0.002 (0.32)
+ 0.01 (1.35)
Values are presented as absolute values and z-scores created from published reference data
aindicates significance according to Wilcoxon signed rank test for related samples p < 0.01
LV, left ventricular wall; IVS, interventricular septum; RV, right ventricular wall
b, difference between measurement after minus before IUT; IQ; interquartile range
Am, peak myocardial velocity during atrial contraction; Sm, peak myocardial velocity during ventricular ejection; Em, peak myocardial velocity during rapid ventricular filling

Discussion

This study showed that all peak myocardial velocities in the LV and IVS and Sm in the RV wall are increased before IUT compared to normal values, reflecting the hyperdynamic circulation. The same myocardial velocities, as well as the heart rate, also decreased significantly after IUT. We also demonstrated a negative correlation between Hb concentration and LV Em and LV Em/Am before transfusion in all IUTs and more prominently when only first time IUTs were assessed. This study also showed that the effect of IUT on the fetal myocardium can be assessed using cTDI with automated analysis.
The increase in myocardial velocities before IUT, is in accordance with the hyperdynamic circulation associated with anemia that has been described by other investigators recording fetal blood flow velocities [2, 27, 28]. Interestingly, myocardial velocity during rapid ventricular filling (Em) seemed increasingly enhanced compared to the myocardial velocity during atrial contraction (Am) as the degree of anemia increased, suggesting an enhanced ability for relaxation. This resulted in a strong negative correlation (rho = − 0.82) between Hb and LV Em/Am ratio before first transfusions. Blood flow velocities in the MCA, the thoracic aorta and across the mitral and aortic valves have previously been demonstrated to correlate with hemoglobin levels in anemic fetuses [2, 19, 29]. In accordance with our present observations, Rizzo et al. have also shown a trend towards increased E/A ratios compared to normal controls using the measurement of blood flow velocities across the atrioventricular valves before IUT in a small group of previously transfused fetuses [8].
We observed a decrease of systolic and diastolic myocardial velocities and a significant increase in LV Em/Am ratio after IUT in accordance with the findings reported using fetal cardiac blood flow velocity variables [8]. A decrease in cardiac output immediately after IUT has previously been described by other investigators [7, 8, 30]. The most obvious explanation to the decrease in several of the peak myocardial velocities after IUT is the reversal of the hyperkinetic circulation as hemoglobin level and oxygen carrying capacity increase. However, the reduction of myocardial velocities was greater than expected suggesting other effects on cardiac function. This might be due to the fast volume expansion with packed red blood cells that might also have an immediate effect on afterload [31] known to be poorly tolerated by the fetal heart, and result in a decreased cardiac output [32, 33] and, thus, possibly further diminishing myocardial velocities.
The strength of this study is that it describes fetal cardiac physiology in fetuses before and after IUT with a focus on myocardial function that has not been well studied. It also demonstrates the potential of automation in facilitating the use of cTDI in the evaluation of fetal myocardial function that can be applied as a tool for the assessment of fetal cardiac physiology under different conditions.
Our study does have some limitations. Multiple measurements were performed in the same fetus when evaluating all IUTs, which may introduce some bias when assessing the correlation between myocardial velocities and degree of fetal anemia. Therefore, we evaluated first IUTs separately even though the numbers were few. The study group was also heterogeneous with varied etiologies and varying degrees of anemia that might influence the results. It is also important to acknowledge that evaluation of cardiac function has been performed close after, and with varying intervals from, the IUT and that several investigators did not demonstrate changes in cardiac variables when longer time had elapsed [1, 9, 34]. Moreover, cTDI has limitations, such as angle dependency influencing velocity measurements and subjectivity in ROI placements, although the latter might be solved by new automation techniques, such as artificial intelligence technology.
However, the current study demonstrated that the rapid physiological adaptation at fetal transfusions well can be monitored by the automated method we utilize in the present study. It remains to be established how this promising method will perform in more chronic disorders such as intrauterine growth restriction. Preliminary studies suggest that the method is sensitive enough to evaluate physiological adaptation at this condition.

Conclusions

This study showed that peak myocardial velocities assessed by cTDI are increased in fetuses before IUT reflecting the physiology of hyperdynamic circulation. It also showed that the fetal heart is able to adapt and efficiently handle the volume load caused by IUT by altering its myocardial function.
With advances in automation, cTDI could be more readily applicable to study changes in fetal cardiac physiology.

Acknowledgements

To all women participating in this study and to all colleagues at Karolinska University Hospital involved in their care.
This research complies with the guidelines for human studies in accordance with the World Medical Association, Declaration of Helsinki. All participants have given their written informed consent and the study protocol was approved by the Stockholm Regional Ethics Committee (DNr 2012/895–31/4 and 2009/1617–31/2).
Not applicable.

Competing interests

LH, KFW, PL, SES, GA and MW have no conflict of interest to declare.
JJ has a few patents related to the subject in the article:
  • Lundback, S., Johnson, Jonas Heart cluster state machine simulating the heart. US Patent 20080154142 Washington, DC: U.S. 2006. 41.
  • Lundback, S., Johnson, Jonas, State space model of a heart (SSM). US Patent 8,244,510 Washington, DC: U.S. 2007. 42 [7]
  • Lundback, S., Johnson, Jonas, State machine user and validation interface system. US Patent 8,943,429 Washington, DC: U.S. 2008. 44 [17].
  • Lundback, S., Johnson, Jonas, Edfors Anders, Computer based system adapted to create a representation of the pumping action of a heart. US Patent 7,239,987. Washington, DC: U.S. 2001. 45.
  • Lundback, S., Johnson, Jonas, Bergholm Fredrik, A cardiac state system, US Patent Pending. Washington, DC: U.S. 2014.
FB also has one patent (mentioned in the list above):
  • Lundback, S., J. Johnson., Bergholm Fredrik, A cardiac state system, US Patent Pending. Washington, DC: U.S. 2014.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Copel JA, Grannum PA, Green JJ, Belanger K, Hanna N, Jaffe CC, et al. Fetal cardiac output in the isoimmunized pregnancy: a pulsed Doppler-echocardiographic study of patients undergoing intravascular intrauterine transfusion. Am J Obstet Gynecol. 1989;161(2):361–5.PubMed Copel JA, Grannum PA, Green JJ, Belanger K, Hanna N, Jaffe CC, et al. Fetal cardiac output in the isoimmunized pregnancy: a pulsed Doppler-echocardiographic study of patients undergoing intravascular intrauterine transfusion. Am J Obstet Gynecol. 1989;161(2):361–5.PubMed
2.
Zurück zum Zitat Hecher K, Snijders R, Campbell S, Nicolaides K. Fetal venous, arterial, and intracardiac blood flows in red blood cell isoimmunization. Obstet Gynecol. 1995;85(1):122–8.PubMed Hecher K, Snijders R, Campbell S, Nicolaides K. Fetal venous, arterial, and intracardiac blood flows in red blood cell isoimmunization. Obstet Gynecol. 1995;85(1):122–8.PubMed
3.
Zurück zum Zitat Moise KJ Jr. Management of rhesus alloimmunization in pregnancy. Obstet Gynecol. 2008;112(1):164–76.PubMed Moise KJ Jr. Management of rhesus alloimmunization in pregnancy. Obstet Gynecol. 2008;112(1):164–76.PubMed
4.
Zurück zum Zitat Davis LE, Hohimer AR. Hemodynamics and organ blood flow in fetal sheep subjected to chronic anemia. Am J Phys. 1991;261(6 Pt 2):R1542–8. Davis LE, Hohimer AR. Hemodynamics and organ blood flow in fetal sheep subjected to chronic anemia. Am J Phys. 1991;261(6 Pt 2):R1542–8.
5.
Zurück zum Zitat Rodeck CH, Kemp JR, Holman CA, Whitmore DN, Karnicki J, Austin MA. Direct intravascular fetal blood transfusion by fetoscopy in severe rhesus isoimmunisation. Lancet. 1981;1(8221):625–7.PubMed Rodeck CH, Kemp JR, Holman CA, Whitmore DN, Karnicki J, Austin MA. Direct intravascular fetal blood transfusion by fetoscopy in severe rhesus isoimmunisation. Lancet. 1981;1(8221):625–7.PubMed
6.
Zurück zum Zitat Zwiers C, van Kamp I, Oepkes D, Lopriore E. Intrauterine transfusion and non-invasive treatment options for hemolytic disease of the fetus and newborn - review on current management and outcome. Expert Rev Hematol. 2017;10(4):337–44.PubMed Zwiers C, van Kamp I, Oepkes D, Lopriore E. Intrauterine transfusion and non-invasive treatment options for hemolytic disease of the fetus and newborn - review on current management and outcome. Expert Rev Hematol. 2017;10(4):337–44.PubMed
7.
Zurück zum Zitat Moise KJ Jr, Mari G, Fisher DJ, Huhta JC, Cano LE, Carpenter RJ Jr. Acute fetal hemodynamic alterations after intrauterine transfusion for treatment of severe red blood cell alloimmunization. Am J Obstet Gynecol. 1990;163(3):776–84.PubMed Moise KJ Jr, Mari G, Fisher DJ, Huhta JC, Cano LE, Carpenter RJ Jr. Acute fetal hemodynamic alterations after intrauterine transfusion for treatment of severe red blood cell alloimmunization. Am J Obstet Gynecol. 1990;163(3):776–84.PubMed
8.
Zurück zum Zitat Rizzo G, Nicolaides KH, Arduini D, Campbell S. Effects of intravascular fetal blood transfusion on fetal intracardiac Doppler velocity waveforms. Am J Obstet Gynecol. 1990;163(4 Pt 1):1231–8.PubMed Rizzo G, Nicolaides KH, Arduini D, Campbell S. Effects of intravascular fetal blood transfusion on fetal intracardiac Doppler velocity waveforms. Am J Obstet Gynecol. 1990;163(4 Pt 1):1231–8.PubMed
9.
Zurück zum Zitat Copel JA, Grannum PA, Belanger K, Green J, Hobbins JC. Pulsed Doppler flow-velocity waveforms before and after intrauterine intravascular transfusion for severe erythroblastosis fetalis. Am J Obstet Gynecol. 1988;158(4):768–74.PubMed Copel JA, Grannum PA, Belanger K, Green J, Hobbins JC. Pulsed Doppler flow-velocity waveforms before and after intrauterine intravascular transfusion for severe erythroblastosis fetalis. Am J Obstet Gynecol. 1988;158(4):768–74.PubMed
10.
Zurück zum Zitat Michel M, Schmitz R, Kiesel L, Steinhard J. Fetal myocardial peak systolic strain before and after intrauterine red blood cell transfusion--a tissue Doppler imaging study. J Perinat Med. 2012;40(5):545–50.PubMed Michel M, Schmitz R, Kiesel L, Steinhard J. Fetal myocardial peak systolic strain before and after intrauterine red blood cell transfusion--a tissue Doppler imaging study. J Perinat Med. 2012;40(5):545–50.PubMed
11.
Zurück zum Zitat Michel M, Schmitz R, Entenmann A, Heinig J, Steinhard J. Correlation between cardiac tissue Doppler and peripheral vascular Doppler in the anemic fetus - an additional method to determine the transfusion threshold? Ultraschall Med. 2015;36(1):35–9.PubMed Michel M, Schmitz R, Entenmann A, Heinig J, Steinhard J. Correlation between cardiac tissue Doppler and peripheral vascular Doppler in the anemic fetus - an additional method to determine the transfusion threshold? Ultraschall Med. 2015;36(1):35–9.PubMed
13.
Zurück zum Zitat Elmstedt NN, Johnson JJ, Lind BB, Ferm-Widlund KK, Herling LL, Westgren MM, et al. Reference values for fetal tissue velocity imaging and a new approach to evaluate fetal myocardial function. Cardiovasc Ultrasound. 2013;11(1):29.PubMedPubMedCentral Elmstedt NN, Johnson JJ, Lind BB, Ferm-Widlund KK, Herling LL, Westgren MM, et al. Reference values for fetal tissue velocity imaging and a new approach to evaluate fetal myocardial function. Cardiovasc Ultrasound. 2013;11(1):29.PubMedPubMedCentral
14.
Zurück zum Zitat Nii M, Roman KS, Kingdom J, Redington AN, Jaeggi ET. Assessment of the evolution of normal fetal diastolic function during mid and late gestation by spectral Doppler tissue echocardiography. J Am Soc Echocardiogr. 2006;19(12):1431–7.PubMed Nii M, Roman KS, Kingdom J, Redington AN, Jaeggi ET. Assessment of the evolution of normal fetal diastolic function during mid and late gestation by spectral Doppler tissue echocardiography. J Am Soc Echocardiogr. 2006;19(12):1431–7.PubMed
15.
Zurück zum Zitat Herling L, Johnson J, Ferm-Widlund K, Lindgren P, Acharya G, Westgren M. Automated analysis of color tissue Doppler velocity recordings of the fetal myocardium using a new algorithm. Cardiovasc Ultrasound. 2015;13:39.PubMedPubMedCentral Herling L, Johnson J, Ferm-Widlund K, Lindgren P, Acharya G, Westgren M. Automated analysis of color tissue Doppler velocity recordings of the fetal myocardium using a new algorithm. Cardiovasc Ultrasound. 2015;13:39.PubMedPubMedCentral
16.
Zurück zum Zitat Herling L, Johnson J, Ferm-Widlund K, Bergholm F, Lindgren P, Sonesson SE, et al. Automated analysis of fetal cardiac function using color tissue Doppler imaging. Ultrasound Obstet Gynecol. 2018;52(5):599–608.PubMed Herling L, Johnson J, Ferm-Widlund K, Bergholm F, Lindgren P, Sonesson SE, et al. Automated analysis of fetal cardiac function using color tissue Doppler imaging. Ultrasound Obstet Gynecol. 2018;52(5):599–608.PubMed
17.
Zurück zum Zitat Herling L, Johnson J, Ferm-Widlund K, Bergholm F, Elmstedt N, Lindgren P, et al. Automated analysis of fetal cardiac function using color tissue Doppler imaging in second half of normal pregnancy. Ultrasound Obstet Gynecol. 2019;53(3):348–57.PubMed Herling L, Johnson J, Ferm-Widlund K, Bergholm F, Elmstedt N, Lindgren P, et al. Automated analysis of fetal cardiac function using color tissue Doppler imaging in second half of normal pregnancy. Ultrasound Obstet Gynecol. 2019;53(3):348–57.PubMed
18.
Zurück zum Zitat Mari G, Adrignolo A, Abuhamad AZ, Pirhonen J, Jones DC, Ludomirsky A, et al. Diagnosis of fetal anemia with Doppler ultrasound in the pregnancy complicated by maternal blood group immunization. Ultrasound Obstet Gynecol. 1995;5(6):400–5.PubMed Mari G, Adrignolo A, Abuhamad AZ, Pirhonen J, Jones DC, Ludomirsky A, et al. Diagnosis of fetal anemia with Doppler ultrasound in the pregnancy complicated by maternal blood group immunization. Ultrasound Obstet Gynecol. 1995;5(6):400–5.PubMed
19.
Zurück zum Zitat Mari G, Deter RL, Carpenter RL, Rahman F, Zimmerman R, Moise KJ Jr, et al. Noninvasive diagnosis by Doppler ultrasonography of fetal anemia due to maternal red-cell alloimmunization. Collaborative Group for Doppler Assessment of the blood velocity in anemic fetuses. N Engl J Med. 2000;342(1):9–14.PubMed Mari G, Deter RL, Carpenter RL, Rahman F, Zimmerman R, Moise KJ Jr, et al. Noninvasive diagnosis by Doppler ultrasonography of fetal anemia due to maternal red-cell alloimmunization. Collaborative Group for Doppler Assessment of the blood velocity in anemic fetuses. N Engl J Med. 2000;342(1):9–14.PubMed
20.
Zurück zum Zitat Tiblad E, Kublickas M, Ajne G, Bui TH, Ek S, Karlsson A, et al. Procedure-related complications and perinatal outcome after intrauterine transfusions in red cell alloimmunization in Stockholm. Fetal Diagn Ther. 2011;30(4):266–73.PubMed Tiblad E, Kublickas M, Ajne G, Bui TH, Ek S, Karlsson A, et al. Procedure-related complications and perinatal outcome after intrauterine transfusions in red cell alloimmunization in Stockholm. Fetal Diagn Ther. 2011;30(4):266–73.PubMed
21.
Zurück zum Zitat Rodeck CH, Nicolaides KH, Warsof SL, Fysh WJ, Gamsu HR, Kemp JR. The management of severe rhesus isoimmunization by fetoscopic intravascular transfusions. Am J Obstet Gynecol. 1984;150(6):769–74.PubMed Rodeck CH, Nicolaides KH, Warsof SL, Fysh WJ, Gamsu HR, Kemp JR. The management of severe rhesus isoimmunization by fetoscopic intravascular transfusions. Am J Obstet Gynecol. 1984;150(6):769–74.PubMed
22.
Zurück zum Zitat Leduc L, Moise KJ Jr, Carpenter RJ Jr, Cano LE. Fetoplacental blood volume estimation in pregnancies with Rh alloimmunization. Fetal Diagn Ther. 1990;5(3–4):138–46.PubMed Leduc L, Moise KJ Jr, Carpenter RJ Jr, Cano LE. Fetoplacental blood volume estimation in pregnancies with Rh alloimmunization. Fetal Diagn Ther. 1990;5(3–4):138–46.PubMed
23.
Zurück zum Zitat Bergholm F. En algoritm för att finna maxima och minima i "periodisk" kolvliknande rörelse. Stockholm: KTH Technology and Health; 2016. Bergholm F. En algoritm för att finna maxima och minima i "periodisk" kolvliknande rörelse. Stockholm: KTH Technology and Health; 2016.
24.
Zurück zum Zitat Bergholm F. Kommentarer kring hastighetssignaler från AV-planets rörelse som är uppmätta i foster. Stockholm: KTH Technology and Health; 2016. Bergholm F. Kommentarer kring hastighetssignaler från AV-planets rörelse som är uppmätta i foster. Stockholm: KTH Technology and Health; 2016.
25.
Zurück zum Zitat Lundback S, Johnson J. inventorsState machine user and validation interface system. US patent. 2008;8(943):429. Lundback S, Johnson J. inventorsState machine user and validation interface system. US patent. 2008;8(943):429.
26.
Zurück zum Zitat Finetti M, Omenetti A, Federici S, Caorsi R, Gattorno M. Chronic infantile neurological cutaneous and articular (CINCA) syndrome: a review. Orphanet J Rare Dis. 2016;11(1):167.PubMedPubMedCentral Finetti M, Omenetti A, Federici S, Caorsi R, Gattorno M. Chronic infantile neurological cutaneous and articular (CINCA) syndrome: a review. Orphanet J Rare Dis. 2016;11(1):167.PubMedPubMedCentral
27.
Zurück zum Zitat Rightmire DA, Nicolaides KH, Rodeck CH, Campbell S. Fetal blood velocities in Rh isoimmunization: relationship to gestational age and to fetal hematocrit. Obstet Gynecol. 1986;68(2):233–6.PubMed Rightmire DA, Nicolaides KH, Rodeck CH, Campbell S. Fetal blood velocities in Rh isoimmunization: relationship to gestational age and to fetal hematocrit. Obstet Gynecol. 1986;68(2):233–6.PubMed
28.
Zurück zum Zitat Oepkes D, Vandenbussche FP, Van Bel F, Kanhai HH. Fetal ductus venosus blood flow velocities before and after transfusion in red-cell alloimmunized pregnancies. Obstet Gynecol. 1993;82(2):237–41.PubMed Oepkes D, Vandenbussche FP, Van Bel F, Kanhai HH. Fetal ductus venosus blood flow velocities before and after transfusion in red-cell alloimmunized pregnancies. Obstet Gynecol. 1993;82(2):237–41.PubMed
29.
Zurück zum Zitat Oberhoffer R, Grab D, Keckstein J, Hogel J, Terinde R, Lang D. Cardiac changes in fetuses secondary to immune hemolytic anemia and their relation to hemoglobin and catecholamine concentrations in fetal blood. Ultrasound Obstet Gynecol. 1999;13(6):396–400.PubMed Oberhoffer R, Grab D, Keckstein J, Hogel J, Terinde R, Lang D. Cardiac changes in fetuses secondary to immune hemolytic anemia and their relation to hemoglobin and catecholamine concentrations in fetal blood. Ultrasound Obstet Gynecol. 1999;13(6):396–400.PubMed
30.
Zurück zum Zitat Teyssier G, Fouron JC, Sonesson SE, Bonnin P, Skoll A. Circulatory changes induced by isovolumic increase in red cell mass in fetal lambs. Arch Dis Child Fetal Neonatal Ed. 1998;79(3):F180–4.PubMedPubMedCentral Teyssier G, Fouron JC, Sonesson SE, Bonnin P, Skoll A. Circulatory changes induced by isovolumic increase in red cell mass in fetal lambs. Arch Dis Child Fetal Neonatal Ed. 1998;79(3):F180–4.PubMedPubMedCentral
31.
Zurück zum Zitat Kilby MD, Szwarc RS, Benson LN, Morrow RJ. Left ventricular hemodynamic effects of rapid, in utero intravascular transfusion in anemic fetal lambs. J Matern Fetal Med. 1998;7(1):51–8.PubMed Kilby MD, Szwarc RS, Benson LN, Morrow RJ. Left ventricular hemodynamic effects of rapid, in utero intravascular transfusion in anemic fetal lambs. J Matern Fetal Med. 1998;7(1):51–8.PubMed
32.
Zurück zum Zitat Reller MD, Morton MJ, Reid DL, Thornburg KL. Fetal lamb ventricles respond differently to filling and arterial pressures and to in utero ventilation. Pediatr Res. 1987;22(6):621–6.PubMed Reller MD, Morton MJ, Reid DL, Thornburg KL. Fetal lamb ventricles respond differently to filling and arterial pressures and to in utero ventilation. Pediatr Res. 1987;22(6):621–6.PubMed
33.
Zurück zum Zitat Hawkins J, Van Hare GF, Schmidt KG, Rudolph AM. Effects of increasing afterload on left ventricular output in fetal lambs. Circ Res. 1989;65(1):127–34.PubMed Hawkins J, Van Hare GF, Schmidt KG, Rudolph AM. Effects of increasing afterload on left ventricular output in fetal lambs. Circ Res. 1989;65(1):127–34.PubMed
34.
Zurück zum Zitat Sikkel E, Klumper FJ, Oepkes D, Teunissen AK, Meerman RH, Le Cessie S, et al. Fetal cardiac contractility before and after intrauterine transfusion. Ultrasound Obstet Gynecol. 2005;26(6):611–7.PubMed Sikkel E, Klumper FJ, Oepkes D, Teunissen AK, Meerman RH, Le Cessie S, et al. Fetal cardiac contractility before and after intrauterine transfusion. Ultrasound Obstet Gynecol. 2005;26(6):611–7.PubMed
Metadaten
Titel
Fetal cardiac function at intrauterine transfusion assessed by automated analysis of color tissue Doppler recordings
verfasst von
Lotta Herling
Jonas Johnson
Kjerstin Ferm-Widlund
Fredrik Bergholm
Peter Lindgren
Sven-Erik Sonesson
Ganesh Acharya
Magnus Westgren
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
Cardiovascular Ultrasound / Ausgabe 1/2020
Elektronische ISSN: 1476-7120
DOI
https://doi.org/10.1186/s12947-020-00214-1

Weitere Artikel der Ausgabe 1/2020

Cardiovascular Ultrasound 1/2020 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.