Prospective comparison of dual-energy CT aortography using 70% reduced iodine dose versus single-energy CT aortography using standard iodine dose in the same patient

verfasst von: William P. Shuman, Ryan B. O’Malley, Janet M. Busey, Mario M. Ramos, Kent M. Koprowicz

Erschienen in: Abdominal Radiology | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare dual-energy computed tomography (DECT) aortography using a 70% reduced iodine dose to single-energy CT (SECT) aortography using a standard iodine dose in the same patient.

Methods

Twenty-one patients with a prior SECT aortography using standard iodine dose had DECT aortography using 70% reduced iodine dose. Section 120 kVp images were compared to DECT images reconstructed at both 50 and 77 keV. Reviewers measured image noise and attenuation in the aorta at eight locations from proximal to distal and subjectively scored vascular enhancement on a four-point scale. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The volume CT dose index (CTDIvol) for each exam was recorded.

Results

Mean iodine dose was 50 g for SECT and 15 g for DECT (70% reduction). Mean aortic attenuation was similar for section 120 kVp (350 ± 67 HU) and DECT 50 keV (338 ± 57 HU, p = 0.547) but was lower at 77 keV (152 ± 23 HU). Measured image noise was greatest at 50 keV (12 ± 5 HU) and was lowest at 77 keV (7 ± 2 HU, p > 0.001). There was no difference in SNR or CNR between 120 kVp and 50 keV (p > 0.05). Mean subjective vascular enhancement scores for SECT were between good and excellent (3.33–3.69), and for DECT at 50 keV were between moderate and good (2.54–2.93, p < 0.0001). CTDIvol was 13.6 mGy for SECT and 13.1 mGy for DECT (p = 0.637).

Conclusion

70% Reduced iodine DECT aortography may result in similar aortic attenuation, CNR, SNR, and lower although acceptable subjective image scores when compared to standard iodine SECT aortography in the same patient.
Literatur
1.
Zurück zum Zitat Hagiwara S, Saima S, Negishi K, et al. (2007) High incidence of renal failure in patients with aortic aneurysms. Nephrol Dial Transplant 22:1361–1368CrossRefPubMed Hagiwara S, Saima S, Negishi K, et al. (2007) High incidence of renal failure in patients with aortic aneurysms. Nephrol Dial Transplant 22:1361–1368CrossRefPubMed
2.
Zurück zum Zitat Parmer SS, Fairman RM, Karmacharya J, et al. (2006) A comparison of renal function between open and endovascular aneurysm repair in patients with baseline chronic renal insufficiency. J Vasc Surg 44:706–711CrossRefPubMed Parmer SS, Fairman RM, Karmacharya J, et al. (2006) A comparison of renal function between open and endovascular aneurysm repair in patients with baseline chronic renal insufficiency. J Vasc Surg 44:706–711CrossRefPubMed
3.
Zurück zum Zitat Chang CF, Lin CC (2013) Current concepts of contrast-induced nephropathy: a brief review. J Chin Med Assoc 76:673–681CrossRefPubMed Chang CF, Lin CC (2013) Current concepts of contrast-induced nephropathy: a brief review. J Chin Med Assoc 76:673–681CrossRefPubMed
4.
Zurück zum Zitat Laville M, Juillard L (2010) Contrast-induced acute kidney injury: How should at-risk patients be identified and managed? J Nephrol 23:387–398PubMed Laville M, Juillard L (2010) Contrast-induced acute kidney injury: How should at-risk patients be identified and managed? J Nephrol 23:387–398PubMed
5.
Zurück zum Zitat Nyman U, Almen T, Aspelin P, et al. (2005) Contrast-medium-induced nephropathy correlated to the ratio between dose in gram iodine and estimated GFR in ml/min. Acta Radiol 46:830–842CrossRefPubMed Nyman U, Almen T, Aspelin P, et al. (2005) Contrast-medium-induced nephropathy correlated to the ratio between dose in gram iodine and estimated GFR in ml/min. Acta Radiol 46:830–842CrossRefPubMed
6.
Zurück zum Zitat Schoellnast H, Tillich M, Deutschmann MJ, et al. (2004) Aortoiliac enhancement during computed tomography angiography with reduced contrast material dose and saline solution flush: influence on magnitude and uniformity of the contrast column. Investig Radiol 39:20–26CrossRef Schoellnast H, Tillich M, Deutschmann MJ, et al. (2004) Aortoiliac enhancement during computed tomography angiography with reduced contrast material dose and saline solution flush: influence on magnitude and uniformity of the contrast column. Investig Radiol 39:20–26CrossRef
7.
Zurück zum Zitat Shuman WP, Chan KT, Busey JM, Mitsumori LM, Koprowicz KM (2016) Dual-energy CT aortography with 50% reduced iodine dose versus single-energy CT aortography with standard iodine dose. Acad Radiol 23:611–618CrossRefPubMed Shuman WP, Chan KT, Busey JM, Mitsumori LM, Koprowicz KM (2016) Dual-energy CT aortography with 50% reduced iodine dose versus single-energy CT aortography with standard iodine dose. Acad Radiol 23:611–618CrossRefPubMed
8.
Zurück zum Zitat Kulkarni NM, Sahani DV, Desai GS, Kalva SP (2012) Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: advantage of low-kiloelectron volt monochromatic images. J Vasc Interv Radiol 23:879–886CrossRefPubMed Kulkarni NM, Sahani DV, Desai GS, Kalva SP (2012) Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: advantage of low-kiloelectron volt monochromatic images. J Vasc Interv Radiol 23:879–886CrossRefPubMed
9.
Zurück zum Zitat Beeres M, Trommer J, Frellesen C, et al. (2016) Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging. Int J Cardiovasc Imaging 32:137–144CrossRefPubMed Beeres M, Trommer J, Frellesen C, et al. (2016) Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging. Int J Cardiovasc Imaging 32:137–144CrossRefPubMed
10.
Zurück zum Zitat Pinho DF, Kulkarni NM, Krishnaraj A, Kalva SP, Sahani DV (2013) Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: image quality, enhancement, diagnosis and radiation dose. Eur Radiol 23:351–359CrossRefPubMed Pinho DF, Kulkarni NM, Krishnaraj A, Kalva SP, Sahani DV (2013) Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: image quality, enhancement, diagnosis and radiation dose. Eur Radiol 23:351–359CrossRefPubMed
11.
Zurück zum Zitat Carrascosa P, Capunay C, Rodriguez-Granillo GA, et al. (2014) Substantial iodine volume load reduction in CT angiography with dual-energy imaging: insights from a pilot randomized study. Int J Cardiovasc Imaging 30:1613–1620CrossRefPubMed Carrascosa P, Capunay C, Rodriguez-Granillo GA, et al. (2014) Substantial iodine volume load reduction in CT angiography with dual-energy imaging: insights from a pilot randomized study. Int J Cardiovasc Imaging 30:1613–1620CrossRefPubMed
12.
Zurück zum Zitat He J, Wang Q, Ma X, Sun Z (2015) Dual-energy CT angiography of abdomen with routine concentration contrast agent in comparison with conventional single-energy CT with high concentration contrast agent. Eur J Radiol 84:221–227CrossRefPubMed He J, Wang Q, Ma X, Sun Z (2015) Dual-energy CT angiography of abdomen with routine concentration contrast agent in comparison with conventional single-energy CT with high concentration contrast agent. Eur J Radiol 84:221–227CrossRefPubMed
13.
Zurück zum Zitat Yuan R, Shuman WP, Earls JP, et al. (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography—a prospective randomized trial. Radiology 262:290–297CrossRefPubMed Yuan R, Shuman WP, Earls JP, et al. (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography—a prospective randomized trial. Radiology 262:290–297CrossRefPubMed
14.
Zurück zum Zitat Godoy MC, Heller SL, Naidich DP, et al. (2009) Dual-energy MDCT: comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies. Eur J Radiol 79:e11–e17CrossRef Godoy MC, Heller SL, Naidich DP, et al. (2009) Dual-energy MDCT: comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies. Eur J Radiol 79:e11–e17CrossRef
15.
Zurück zum Zitat Agrawal MD, Oliveira GR, Kalva SP, Pinho DF, Arellano RS, et al. (2016) Prospective comparison of reduced-iodine-dose virtual monochromatic imaging dataset from dual-energy CT angiography with standard-iodine-dose single-energy CT angiography for abdominal aortic aneurysm. Am J Roentgenol 207:W125–W132CrossRef Agrawal MD, Oliveira GR, Kalva SP, Pinho DF, Arellano RS, et al. (2016) Prospective comparison of reduced-iodine-dose virtual monochromatic imaging dataset from dual-energy CT angiography with standard-iodine-dose single-energy CT angiography for abdominal aortic aneurysm. Am J Roentgenol 207:W125–W132CrossRef
Metadaten
Titel
Prospective comparison of dual-energy CT aortography using 70% reduced iodine dose versus single-energy CT aortography using standard iodine dose in the same patient
verfasst von
William P. Shuman
Ryan B. O’Malley
Janet M. Busey
Mario M. Ramos
Kent M. Koprowicz
Publikationsdatum
13.01.2017
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 3/2017
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-016-1041-z

Weitere Artikel der Ausgabe 3/2017

Abdominal Radiology 3/2017 Zur Ausgabe

Classics in Abdominal Imaging

The “erased charcoal” sign

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.