Quality assessment of an interferon-gamma release assay for tuberculosis infection in a resource-limited setting

verfasst von: Nguyen TL Hang, Naoki Ishizuka, Naoto Keicho, Le T Hong, Do B Tam, Vu TX Thu, Ikumi Matsushita, Nobuyuki Harada, Kazue Higuchi, Shinsaku Sakurada, Luu T Lien

Erschienen in: BMC Infectious Diseases | Ausgabe 1/2009

Abstract

Background

When a test for diagnosis of infectious diseases is introduced in a resource-limited setting, monitoring quality is a major concern. An optimized design of experiment and statistical models are required for this assessment.

Methods

Interferon-gamma release assay to detect tuberculosis (TB) infection from whole blood was tested in Hanoi, Viet Nam. Balanced incomplete block design (BIBD) was planned and fixed-effect models with heterogeneous error variance were used for analysis. In the first trial, the whole blood from 12 donors was incubated with nil, TB-specific antigens or mitogen. In 72 measurements, two laboratory members exchanged their roles in harvesting plasma and testing for interferon-gamma release using enzyme linked immunosorbent assay (ELISA) technique. After intervention including checkup of all steps and standard operation procedures, the second trial was implemented in a similar manner.

Results

The lack of precision in the first trial was clearly demonstrated. Large within-individual error was significantly affected by both harvester and ELISA operator, indicating that both of the steps had problems. After the intervention, overall within-individual error was significantly reduced (P < 0.0001) and error variance was no longer affected by laboratory personnel in charge, indicating that a marked improvement could be objectively observed.

Conclusion

BIBD and analysis of fixed-effect models with heterogeneous variance are suitable and useful for objective and individualized assessment of proficiency in a multistep diagnostic test for infectious diseases in a resource-constrained laboratory. The action plan based on our findings would be worth considering when monitoring for internal quality control is difficult on site.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2334-9-66) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

NTLH participated in supervising the on-site implementation of the study, drafting the paper or substantially revising it. NI was responsible for making conception, design and overall supervision of the study, analysis and interpretation of data, drafting the paper or substantially revising it. NK participated in making conception and design of the study, analysis and interpretation of data, drafting the paper or substantially revising it. LTH carried out the immunoassays. DBT participated in on-site implementation. VTXT carried out the immunoassays. IM participated in technical transfer and supervision. NH was responsible for technical transfer and supervision. KH was responsible for technical transfer and supervision. SS participated in conception and design of the study. LTL participated in conception, design and supervision of the study. All authors read and approved the manuscript.

Background

Assuring quality is essential for clinical laboratories in the field of infectious diseases. Beneficiaries are not only patients obtaining a diagnosis on site but also future patients receiving benefits of clinical research supported by qualified laboratories. Quality assurance in modern laboratories is realized by total quality management including external quality assurance (EQA) and internal quality control (IQC) [13].
In most resource-constrained countries, however, regulations on quality assurance have not been laid down by the authorities and accuracy and precision of clinical measurements have not been monitored systematically [4]. Under such disadvantageous circumstances, when important but rather complicated testing for infectious diseases is undertaken, we cannot easily be confident that the skill has been transferred and maintained properly until the procedure becomes familiar and stably performed in accordance with a desirable quality control system [5]. During this vulnerable period, how to assess proficiency of the testing effectively and objectively, and how to assure and improve the quality are open issues to be addressed.
Currently, immunoassay is commonly used to make a serological diagnosis of infectious diseases involving human immunodeficiency virus, a variety of hepatitis virus and other sexually transmitted or blood-borne pathogens [6, 7], which are serious problems in the developing world. Enzyme linked immunosorbent assay (ELISA) is often used to make diagnosis of these diseases in the clinical laboratories. Because of the complexity of the method, however, quality control of these assay systems is challenging [8]. In this context, trend of point of care (POC) tests that facilitate immediate and on-site diagnosis as well as early treatment of infectious diseases has been emphasized [7]. However, their usage in resource-constrained countries is still hampered by high cost and difficulties in testing for high throughput screening and thus laboratory-based immunoassays would be irreplaceable in many fields.
Recently, a two-step immunoassay to detect tuberculosis (TB) infection has also been developed and used extensively [9]. It consists of whole blood stimulation with TB-specific antigens followed by measurement of interferon-gamma using ELISA. Our objective in the present study is to demonstrate that the quality of laboratory tests can be assessed objectively even in a resource-constrained laboratory if the optimum design of experiments and appropriate statistical models are chosen. As a result of this attempt, we experienced marked improvement of the quality of this multi-step immunoassay made by more than one laboratory staff member in a hospital of Viet Nam. We proposed a general plan to evaluate skills of laboratory staff members efficiently and quantitatively to perform qualified immuno-diagnostic testing especially for infectious diseases until such time as they establish a total quality management system by themselves.

Methods

Interferon-gamma release assay (IGRA) for diagnosis of TB infection

IGRA is a general method to measure interferon-gamma induced by Mycobacterium tuberculosis-specific antigens (TB-Antigen) for detecting TB infection. In the ELISA-based IGRA (QuantiFERON-TB Gold In-Tube™, Cellestis, Victoria, Australia), one milliliter of the whole blood was collected into the Nil tube for negative control, Mitogen for positive control, and TB-Antigen separately. The blood in the tubes was mixed and placed in the incubator for 18 hours at 37°C (Cool incubator NC-25B, Funakoshi, Tokyo, Japan). Approximately 200 μl of plasma were harvested from each tube after centrifugation (Kubota 2010, Kubota, Tokyo, Japan).
Interferon-gamma concentrations in the plasma were measured by ELISA, using microtiter plate washer and reader (Wellwash Plus Microplate Washer and Multiscan JX Microplate Reader, Thermo Electron Corporation, Vantaa, Finland) with the analysis software provided by the manufacturer (QuantiFERON-TB Gold Analysis Software, ver. 2.50, Cellestis). In this study, interferon-gamma concentrations obtained from this calculation were directly used for further analysis.

Study setting

Two trials were carried out in Hanoi TB and Lung Disease Hospital, Viet Nam. Between the first and second trial, statistical analysis was made and an intervention was planned to ensure counterchecking and correct questionable manipulations. Each trial consisted of two runs. In each run, three milliliters of blood were collected from volunteers after informed consent had been obtained. Study protocols using IGRA were approved by ethical committees of the Ministry of Health, Viet Nam and International Medical Center of Japan respectively.
Two laboratory members, A and B, performed either plasma harvest or ELISA operation or both: Harvest included labeling and placing plasma storage tubes properly and transferring plasma from centrifuged blood collection tubes to these tubes by pipetting. ELISA was a process including preparing reagents and transferring plasma samples into the microtiter plate. ELISA ended with calculation of interferon-gamma concentration. Because their roles were changed occasionally due to the limited manpower of the laboratory, their performance in both Harvest and ELISA was the subject to be analyzed.

Balanced incomplete block design (BIBD)

A single specimen obtained from routine blood collection was not sufficient to assess two staff members' performance. Because additional blood sampling was not easily accepted in many countries including Viet Nam, BIBD was attempted to obtain analytical information from small volume of plasma samples in this study: Of four possible combinations of harvester and ELISA operator, two combinations were cyclically chosen, using the limited amount of specimen. Allocation of observed combinations by BIBD in this study was described in Table 1. In each trial, there were two levels of Harvest (two different Harvesters), two levels of ELISA (two different ELISA operators) and 12 levels of Specimen (12 different blood donors).
Table 1
Allocation of observed combinations of Harvester and ELISA operator.
Sample
Specimen*
Harvest
ELISA
Data
Sample
Specimen
Harvest
ELISA
Data
1
1
A
A
Observed
7
4
A
A
Observed
2
1
A
B
Observed
 
4
A
B
Not observed
 
1
B
A
Not observed
8
4
B
A
Observed
 
1
B
B
Not observed
 
4
B
B
Not observed
 
2
A
A
Not observed
 
5
A
A
Not observed
 
2
A
B
Not observed
9
5
A
B
Observed
3
2
B
A
Observed
 
5
B
A
Not observed
4
2
B
B
Observed
10
5
B
B
Observed
5
3
A
A
Observed
 
6
A
A
Not observed
 
3
A
B
Not observed
11
6
A
B
Observed
 
3
B
A
Not observed
12
6
B
A
Observed
6
3
B
B
Observed
 
6
B
B
Not observed
*To each specimen, two measurements were assigned. This layout was repeated twice by using different sets of specimens in each trial.

Outliers

To identify outliers, Mahalanobis distance D was calculated, which took the distance from the mean and correlation into account [10]. When D > 2.0, the value of that observed pair was regarded as outlier.

A fixed-effect model and three-way analysis of variance (ANOVA)

To assess effects of factors of interest and error variance, we used a fixed-effect model:
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equa_HTML.gif
of which,
yijk: Interferon-gamma concentration in the plasma
μ: Grand mean of all measurements
αi: Harvest with i levels: i = 1, 2 (= A and B)
βj: ELISA with j levels: j = 1, 2 (= A and B)
γk: Blood specimen with k levels: k = 1, 2,..., 12
εijk: Within-individual error; following normal distribution with mean = 0 and variance = σ2: N (0, σ2)
In this clinical setting, effects of interaction terms were not considered in the above model, because harvesting plasma and performing ELISA are independent steps and it is unlikely that the exchanging of staff roles in itself could increase the chances of error.

Analysis of heterogeneous error variance affected by a given factor

To determine whether individuals of Harvest or ELISA affect within-individual error, we assessed a fixed-effect model with heterogeneous variance of error in the following way:
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equb_HTML.gif
where error follows the normal distribution N(0, σij 2).
Error variance affected by Harvesters was evaluated in the following formula:
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equc_HTML.gif
Similarly, the following formula was used for error variance affected by ELISA operators:
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equd_HTML.gif

Coefficient of variation (CV) before and after intervention

Error variance εijk that included sources of Harvest and ELISA was calculated in a simple one-way ANOVA model adjusted by specimen. Based on the following formula, CVs of the two trials were assessed:
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Eque_HTML.gif
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equf_HTML.gif
CV should not be larger than 20% in any types of immunoassay [8].

Assessment of heterogeneous variance between the two trials

To analyze overall within-individual error between the two trials statistically, we used a fixed-effect model with heterogeneous variance of error, under the assumption that α and β were fixed throughout the trials. The effect of each blood specimen γ was expected to be different between the two trials.
https://static-content.springer.com/image/art%3A10.1186%2F1471-2334-9-66/MediaObjects/12879_2009_Article_831_Equg_HTML.gif
of which, εijk1 and εijk2 were within-individual errors of the first and the second trial respectively. On the above assumption, εijk1 and εijk2 would be heterogeneous error between the trials.
Calculation of Mahalanobis distance, three-way ANOVA and estimation of heterogeneous variance were performed by SAS version 9.1 (SAS Institute Cary, NC, USA). Differences in error variance of two trials and error variance affected by a given factor were considered to be significant when P-value was less than 0.05.

Results

Evaluation of outliers and three-way ANOVA in the first trial

Out of 72 measurements obtained from the first trial, seven outliers were identified: One was in Nil condition, three in TB-Antigen and three in Mitogen (Mahalanobis D = 2.64 to 4.69).
To assess effects of individuals for Harvest and ELISA and character of errors involved in the first trial, we first performed three-way ANOVA using a fixed-effect model, in which three factors, Harvest, ELISA and individual blood specimens may have possible effects on the interferon-gamma concentration respectively. This model decomposes the total variance into between-individual error (or bias) and within-individual error (or imprecision). Herein, "between-individual error" indicates deviation in interferon-gamma values caused by the difference between Harvesters or ELISA operators, and "within-individual error" represents fluctuation of interferon-gamma values measured by a single Harvester or ELISA operator.
As shown in Table 2, mean square error indicating magnitude of within-individual error was large in all conditions of the first trial, which was indicated by remarkably large CV (> 20%) for Nil, TB-Antigen and Mitogen. Furthermore, in the condition of Mitogen, the mean-square value directing the effect of ELISA, or "between-individual error", was significantly large (P = 0.017). In the other two conditions, the effects of ELISA and Harvest were also considerably large but did not reach significant levels, as compared with the corresponding mean square errors. These findings indicate that their performance is unstable. Problems specific to ELISA and Harvest should be considered, although not statistically significant in all conditions.
Table 2
Three-way analysis of variance in the first trial.
 
Nil
TB-Antigen
Mitogen
Mean (IU/ml)
0.7821
5.4013
15.3638
Harvest
   
   Mean Square
0.0000
1.5252
51.2656
   F value
0.0000
0.2200
2.5600
   P value
0.9984
0.6482
0.1404
ELISA
   
   Mean Square
1.8838
12.3026
161.3535
   F value
1.2600
1.7800
8.0700
   P value
0.2847
0.2112
0.0175
Error
   
   Mean Square
1.4741
6.8935
19.9916
   Root Mean Square
1.2142
2.6255
4.4712
Coefficient of Variation (%)
155.2476
48.6099
29.1022

Analysis of heterogeneous error variance in the first trial

We then analyzed which factor affected within-individual error. Because two laboratory members were involved in each step of this experiment, we assumed that within-individual error, i.e. error variance, could be different depending on the personnel in each step. Thus, we chose a fixed-effect model with heterogeneous variance of error affected by Harvest and ELISA (Table 3).
Table 3
Error variance affected by Harvester (left) and error variance affected by ELISA operator (right) in the first trial.
εi.
Harvester
P value
ε.j
ELISA operator
P value
Nil
A:1.9150
B:0.0160
0.0040
Nil
A:0.0036
B:3.2723
0.0244
TB-Antigen
A:2.9897
B:9.7114
0.2546
TB-Antigen
A: 0.1270
B:15.2216
0.0830
Mitogen
A:33.5782
B: 5.6792
0.2780
Mitogen
A:41.1154
B: 3.0221
0.3584
In Nil condition, difference in error variance was statistically significant between Harvesters A and B (P = 0.0040), when error variance caused by ELISA operator was not considered. Difference of error variance caused by ELISA operators A and B was also significant (P = 0.024), when error variance caused by Harvester was not taken into account. These findings imply that under the model, the error variance was affected significantly by different Harvesters or ELISA operators, respectively.

Intervention

By means of the above-mentioned statistical analysis of the first trial, we identified several points to be improved: a) there was a considerable number of outliers. Within-individual error was large and between-individual error was also comparably large, and b) within-individual error was affected by both Harvesters and ELISA operators at least when Nil was measured.
Based on these results, an intervention was introduced: 1) reviewing all procedures of Harvest and ELISA, 2) reconsidering and strengthening standard operation procedures, 3) checking working condition of machines, and 4) developing a checklist for countercheck. First, we attempted to find out which procedure of harvesting and ELISA operation would be unstable and all questionable manipulations were listed up. Essential laboratory skills, such as mixing the solution by pipetting, were reviewed. Secondly, standard operation procedures were rechecked and corrected seeing that the laboratory personnel were handling three blood collection tubes and three other plasma storage tubes from each blood donor at a time, they should take every care to identify the tubes during Harvest and ELISA and to confirm the right position of corresponding tubes. Thorough instruction for handling ELISA plates and tubes with manipulation of the pipette was given to avoid carry-over error or contamination. After intensive discussions, more attention was paid to basic laboratory practice and reduction of preventable mistakes. Thirdly, performance of the ELISA plate washer and reader and the quality of distilled water were also checked. Technical requirements from the manufacturer, such as temperature for reagent reservation, time of incubation, were strictly followed. Finally, a checklist for the countercheck of each step was developed for practical use.

General assessment by CV before and after intervention

To assess the overall improvement after intervention, CV was compared between the two trials. Because variation due to Harvest and ELISA was of interest, CV adjusted by the effect of specimens was calculated and used. The CV had decreased remarkably in each condition of the second trial, as compared with that of the first trial, indicating the overall improvement of test performance after intervention (Table 4).
Table 4
CV adjusted by specimen in the two trials.
 
CV (%)
Condition
1st trial
2nd trial
Nil
150.5036
2.1661
TB-Antigen
48.6219
2.3967
Mitogen
38.1630
9.8776

Evaluation of outliers and three-way ANOVA in the second trial

In the second trial, only one outlier was seen in Nil condition (Mahalanobis D = 2.59); the number of outliers was lower than that of the first trial.
We then proceeded to analyze the change of parameters that had possibly contributed to overall improvement of test performance. As shown in Table 5, both mean square error and mean-square values showing effects of Harvest and of ELISA were markedly lower in the second trial. The former implies the decrease in within-individual error and the latter shows the reduction of between-individual error. The latter change was also clearly shown when differences of least square means between Harvesters and between ELISA operators in each condition of the second trial were compared with those in the first trial (Figure 1).
Table 5
Three-way analysis of variance in the second trial.
 
Nil
TB-Antigen
Mitogen
Mean (IU/ml)
0.2308
0.3071
11.0017
Harvest
   
   Mean Square
0.0000
0.0000
3.4225
   F value
1.0000
0.1000
3.2100
   P value
0.3409
0.7572
0.1036
ELISA
   
   Mean Square
0.0000
0.0000
0.0770
   F value
1.0000
0.4000
0.0700
   P value
0.3409
0.5393
0.7937
Error
   
   Mean Square
0.0000
0.0001
1.1707
   Root Mean Square
0.0050
0.0079
1.0330
Coefficient of Variation (%)
2.1661
2.5615
9.3898

Analysis of heterogeneous error variance affected by harvester and ELISA operator in the second trial

In contrast to the first trial, there were no significant differences of error variance affected by Harvesters or ELISA operators (Table 6). This finding showed that the heterogeneous error variance indicating personnel-dependent unstableness was small enough in each step of the second trial.
Table 6
Error variance affected by Harvester (left) and error variance affected by ELISA operator (right) in the second trial.
εi.
Harvester
P value
ε.j
ELISA operator
P value
Nil
NE* (A > B)
 
Nil
A:0.000014
B:0.000035
0.4832
TB-Antigen
A:0.0001
B:0.00002
0.2391
TB-Antigen
NE (A > B)
 
Mitogen
A:2.1184
B: 0.1788
0.3291
Mitogen
A:1.9065
B: 0.2747
0.33347
*NE = not estimable by this calculation.

Assessment of heterogeneous variance between the two trials

We further evaluated the decrease in overall within-individual error statistically. For this purpose, we used a fixed-effect model with heterogeneous variance between the two trials. Under the assumption that influence of Harvest and ELISA was not changed between the two trials, estimated overall error variances of the two trials were compared. As shown in Table 7, values indicating the overall within-individual error in all conditions had significantly decreased in the second trial (P < 0.0001).
Table 7
Difference in estimated overall within-individual error between the two trials.
 
Estimated overall error variance (εijk)
Condition
1st trial
2nd trial*
Nil
1.3866
0.000025
TB-Antigen
6.9003
0.000062
Mitogen
35.3152
1.0814
*P < 0.0001

Discussion

We have demonstrated that a study design BIBD and statistical analysis using fixed-effect models with heterogeneous variance of error are useful for objective and quantitative assessment of laboratory testing for the first time. A series of experiments in our study clearly showed that proficiency of the personnel was improved by an appropriate intervention between the first and second trials of a two-step ELISA-based immunoassay for tuberculosis newly introduced to a resource-constrained laboratory.
Design of clinical experiments including block designs can be used to estimate effect of factors and their possible interaction [10]. In block designs including BIBD, introduction of blocks usually provides extra precision for comparison of other factors, while difference between blocks is of no intrinsic interest [10]. In our proficiency testing, variation of individual specimens was not the point of interest, but analysis of the other two factors, Harvest and ELISA was of importance. Roles of laboratory members are occasionally changed because of limited manpower. In such a case, our analysis is indispensable for assessment of their individual skills in each step of the testing, since this kind of approach has not been evaluated by the conventional IQC methods [11].
Previous studies in clinical fields other than laboratory medicine showed the advantage of BIBD over the sample size [1214]. In the present study, this design enabled us to evaluate essential components of the blood testing procedure systematically without collecting an extra specimen from each donor. If all combinations of Harvesters and ELISA operators were to be tested at the same time, a twice-larger volume of blood should be collected from each volunteer, however, obtaining consent of this often causes difficulties in a country where blood sampling is not easily accepted. We have shown furthermore that this design is suitable for clinical settings in which many different specimens are to be handled at the same time.
In the first trial before intervention, we found that within-individual error was large and between-individual error tended to be so. However, a number of outliers also affected both within- and between-individual errors. The cause of outliers was probably due to mixing up of specimen tubes or contamination of samples resulting from unfamiliar handling of multiple samples, although this was not easily determined [15, 16]. Using a fixed-effect model with the heterogeneity of error variance, we further illustrated that within-individual error was affected by Harvesters and ELISA operators. The results indicated that there were problems with both steps of Harvest and ELISA, and with both laboratory members, and this represented a strong motivation to improve the skills of the laboratory personnel in both steps of Harvest and ELISA.
After timely intervention including checkup of all steps and standard operation procedures, marked improvement was observed in all parameters including CV, a general parameter for precision of measurements [8]. In case of IGRA in this study, CV should be kept less than 10% [17, 18] and in the second trial, this criterion was met satisfactorily.
We propose as a consequence the following action plan to improve diagnostic capacity in resource-constrained settings. This could be generalized not only for complicated immunoassay for infectious disease but also for other kinds of clinical tests:
  • Set the target CV derived from simple one-way ANOVA model of specimen (for example, 10%). This value should be defined before the commencement of study.
  • Design experiment to evaluate between- and within-individual error.
  • Conduct experiment.
  • Analyze data with ANOVA model with and without heterogeneous error variance.
  • If CV exceeds the target, review the operating procedures.
  • Conduct experiment a second time.
  • Consequently analyze data to ascertain any improvement.
  • Return to step 5 until CV becomes less than the target.
In-house quality control for effective transfer of skills is a topic of interest in our proposal and this should be carried out easily, at a low cost, whilst assuring objective and quantitative assessment in a clinical laboratory where resources such as reagents, manpower and feasibility of sample collection are limited. Our plan meets the above requirement. Measurements could be sent via the internet and analyzed in a statistical way by a joint-research facility inside or outside the country and an immediate feedback should be sent in an appropriate manner. Such continuous efforts to share information are important to maintain quality levels over a long distance [19].
In this age of evidence-based medicine and development of new diagnostic technologies, quality of laboratory tests is essential. There is an urgent need for validation and standardization of the new assays before they are adopted into clinical diagnostics [20]. Until such time as an effective quality control system is established, our approach is valuable to assure the quality of laboratory tests for timely diagnosis and treatment of infectious diseases. Another favorable design or analytical method might be suggested by others in the future studies, seeing that no standard way of quality monitoring has been proposed so far. We expect that the successful experience gathered in the present study will provide useful information for further comparison and discussion.
Our study has some limitations. It was obvious that outliers influenced statistical analysis in the first trial and exact causes of error in each condition were not clearly specified by the present analysis itself [15]. Through repeated experiments, the causes of error might be clearer, although all errors in our study decreased dramatically after a single intervention. We should also emphasize in conclusion, that a number of procedures should be combined to establish a total quality assurance system.

Conclusion

In a setting where a modern quality control system has not been entirely established, a laboratory test could be assessed quantitatively and such objective assessment is helpful for quality improvement of the test, if an appropriate design of experiment and statistical method are chosen. The design of experiment BIBD and analytical models for ANOVA were useful for objective assessment of individual skills in each stage of a multi-step immunoassay for tuberculosis in a laboratory with limited resources. A proposed plan to assess the level of proficiency might be useful for skill improvement of clinical testing especially for infectious diseases when monitoring is difficult to assure the sustainability of the technology transferred.

Acknowledgements

The authors thank Dr. Hiroshi Ohara (IMCJ-BMH Medical Collaboration Center) and Dr. Pham Huu Thuong (Hanoi Tuberculosis and Lung Disease Hospital) for supporting management. The authors also thank Kazuko Tanabe D.V.M. for her critical reading of this manuscript. This work was supported by grants from the Program of Founding Research Centers for Emerging and Reemerging Infectious Diseases, MEXT and International Cooperation Research Grant, the Ministry of Health, Labor and Welfare, Japan. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

NTLH participated in supervising the on-site implementation of the study, drafting the paper or substantially revising it. NI was responsible for making conception, design and overall supervision of the study, analysis and interpretation of data, drafting the paper or substantially revising it. NK participated in making conception and design of the study, analysis and interpretation of data, drafting the paper or substantially revising it. LTH carried out the immunoassays. DBT participated in on-site implementation. VTXT carried out the immunoassays. IM participated in technical transfer and supervision. NH was responsible for technical transfer and supervision. KH was responsible for technical transfer and supervision. SS participated in conception and design of the study. LTL participated in conception, design and supervision of the study. All authors read and approved the manuscript.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat Westgard JO: Basic Planning for Quality. 2000, Madison: Westgard QC Inc Westgard JO: Basic Planning for Quality. 2000, Madison: Westgard QC Inc
2.
Zurück zum Zitat Kazmierczak SC: Laboratory quality control: using patient data to assess analytical performance. Clin Chem Lab Med. 2003, 41: 617-27. 10.1515/CCLM.2003.093.CrossRefPubMed Kazmierczak SC: Laboratory quality control: using patient data to assess analytical performance. Clin Chem Lab Med. 2003, 41: 617-27. 10.1515/CCLM.2003.093.CrossRefPubMed
3.
Zurück zum Zitat Westgard JO, Barry PL, Hunt MR, Groth T: A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem. 1981, 27: 493-501.PubMed Westgard JO, Barry PL, Hunt MR, Groth T: A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem. 1981, 27: 493-501.PubMed
4.
Zurück zum Zitat Heuck CC, Deom A: Healthcare in the developing world: Need for appropriate laboratory technology. Clin Chem. 1991, 37: 490-6.PubMed Heuck CC, Deom A: Healthcare in the developing world: Need for appropriate laboratory technology. Clin Chem. 1991, 37: 490-6.PubMed
5.
Zurück zum Zitat Smith PG, Morrow RH: Field trials of health interventions in developing countries: A toolbox. 1996, Oxford: Macmillan Press Limited, 2 Smith PG, Morrow RH: Field trials of health interventions in developing countries: A toolbox. 1996, Oxford: Macmillan Press Limited, 2
6.
Zurück zum Zitat Peruski AH, Peruski LF: Immunological methods for detection and identification of infectious disease and biological warfare agents. Clin Diagn Lab Immunol. 2003, 10: 506-13.PubMedPubMedCentral Peruski AH, Peruski LF: Immunological methods for detection and identification of infectious disease and biological warfare agents. Clin Diagn Lab Immunol. 2003, 10: 506-13.PubMedPubMedCentral
7.
Zurück zum Zitat Peeling RW, Mabey D, Herring A, Hook EW: Why do we need quality-assured diagnostic tests for sexually transmitted infections. Nat Rev Microbiol. 2006, 4: S7-19. 10.1038/nrmicro1569.CrossRefPubMed Peeling RW, Mabey D, Herring A, Hook EW: Why do we need quality-assured diagnostic tests for sexually transmitted infections. Nat Rev Microbiol. 2006, 4: S7-19. 10.1038/nrmicro1569.CrossRefPubMed
8.
9.
Zurück zum Zitat Pai M, Riley LW, Colford JM: Interferon-gamma assays in the immunodiagnosis of tuberculosis: a systematic review. Lancet Infect Dis. 2004, 4: 761-76. 10.1016/S1473-3099(04)01206-X.CrossRefPubMed Pai M, Riley LW, Colford JM: Interferon-gamma assays in the immunodiagnosis of tuberculosis: a systematic review. Lancet Infect Dis. 2004, 4: 761-76. 10.1016/S1473-3099(04)01206-X.CrossRefPubMed
10.
Zurück zum Zitat Armitage P, Berry G, Matthews JNS: Statistical method in medical research. 2002, Malden: Blackwell science Ltd, 4CrossRef Armitage P, Berry G, Matthews JNS: Statistical method in medical research. 2002, Malden: Blackwell science Ltd, 4CrossRef
11.
Zurück zum Zitat Williams GW, Schork MA: Basic statistics for quality control in the clinical laboratory. Crit Rev Clin Lab Sci. 1982, 17: 171-99. 10.3109/10408368209107035.CrossRefPubMed Williams GW, Schork MA: Basic statistics for quality control in the clinical laboratory. Crit Rev Clin Lab Sci. 1982, 17: 171-99. 10.3109/10408368209107035.CrossRefPubMed
12.
Zurück zum Zitat Fleiss JL: The design and analysis of clinical experiments. 1999, New York: John Wiley & Sons, IncCrossRef Fleiss JL: The design and analysis of clinical experiments. 1999, New York: John Wiley & Sons, IncCrossRef
13.
Zurück zum Zitat Endres HG, Hucke C, Holland-Letz T, Trampisch HJ: A new efficient trial design for assessing reliability of ankle-brachial index measures by three different observer groups. BMC Cardiovasc Disord. 2006, 6: 33-10.1186/1471-2261-6-33.CrossRefPubMedPubMedCentral Endres HG, Hucke C, Holland-Letz T, Trampisch HJ: A new efficient trial design for assessing reliability of ankle-brachial index measures by three different observer groups. BMC Cardiovasc Disord. 2006, 6: 33-10.1186/1471-2261-6-33.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Varma AAO, Chilton NW, Kutscher AH: Studies in the Design and Analysis of Dental Experiments: 13. Balanced Incomplete Block Design. J Dent Res. 1970, 49: 1166-1171.CrossRefPubMed Varma AAO, Chilton NW, Kutscher AH: Studies in the Design and Analysis of Dental Experiments: 13. Balanced Incomplete Block Design. J Dent Res. 1970, 49: 1166-1171.CrossRefPubMed
15.
Zurück zum Zitat Marks V: False-positive immunoassay results: a multicenter survey of erroneous immunoassay results from assays of 74 analytes in 10 donors from 66 laboratories in seven countries. Clin Chem. 2002, 48: 2008-16.PubMed Marks V: False-positive immunoassay results: a multicenter survey of erroneous immunoassay results from assays of 74 analytes in 10 donors from 66 laboratories in seven countries. Clin Chem. 2002, 48: 2008-16.PubMed
16.
Zurück zum Zitat Ricós C, García-Victoria M, de la Fuente B: Quality indicators and specifications for the extra-analytical phases in clinical laboratory management. Clin Chem Lab Med. 2004, 42: 578-82. 10.1515/CCLM.2004.100.PubMed Ricós C, García-Victoria M, de la Fuente B: Quality indicators and specifications for the extra-analytical phases in clinical laboratory management. Clin Chem Lab Med. 2004, 42: 578-82. 10.1515/CCLM.2004.100.PubMed
18.
Zurück zum Zitat Menzies D, Pai M, Comstock G: Meta-analysis: new tests for the diagnosis of latent tuberculosis infection: areas of uncertainty and recommendations for research. Ann Intern Med. 2007, 146: 340-54.CrossRefPubMed Menzies D, Pai M, Comstock G: Meta-analysis: new tests for the diagnosis of latent tuberculosis infection: areas of uncertainty and recommendations for research. Ann Intern Med. 2007, 146: 340-54.CrossRefPubMed
19.
Zurück zum Zitat Coloma J, Harris E: Sustainable transfer of biotechnology to developing countries: fighting poverty by bringing scientific tools to developing-country partners. Ann N Y Acad Sci. 2008, 1136: 358-68. 10.1196/annals.1425.014.CrossRefPubMed Coloma J, Harris E: Sustainable transfer of biotechnology to developing countries: fighting poverty by bringing scientific tools to developing-country partners. Ann N Y Acad Sci. 2008, 1136: 358-68. 10.1196/annals.1425.014.CrossRefPubMed
20.
Zurück zum Zitat Lalvani A, Meroni PL, Millington KA, Modolo ML, Plebani M, Tincani A, Villalta D, Doria A, Ghirardello A: Recent advances in diagnostic technology: applications in autoimmune and infectious diseases. Clin Exp Rheumatol. 2008, 26 (1 Suppl 48): S62-6.PubMed Lalvani A, Meroni PL, Millington KA, Modolo ML, Plebani M, Tincani A, Villalta D, Doria A, Ghirardello A: Recent advances in diagnostic technology: applications in autoimmune and infectious diseases. Clin Exp Rheumatol. 2008, 26 (1 Suppl 48): S62-6.PubMed
Metadaten
Titel
Quality assessment of an interferon-gamma release assay for tuberculosis infection in a resource-limited setting
verfasst von
Nguyen TL Hang
Naoki Ishizuka
Naoto Keicho
Le T Hong
Do B Tam
Vu TX Thu
Ikumi Matsushita
Nobuyuki Harada
Kazue Higuchi
Shinsaku Sakurada
Luu T Lien
Publikationsdatum
01.12.2009
Verlag
BioMed Central
Erschienen in
BMC Infectious Diseases / Ausgabe 1/2009
Elektronische ISSN: 1471-2334
DOI
https://doi.org/10.1186/1471-2334-9-66

Weitere Artikel der Ausgabe 1/2009

BMC Infectious Diseases 1/2009 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.