Quantification of water and lipid density with dual-energy mammography: validation in postmortem breasts

verfasst von: Sabee Molloi, Huanjun Ding, Hyo-Min Cho, Daniel Beidokhti

Erschienen in: European Radiology | Ausgabe 2/2021

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Breast cancer is the most common cancer in women and the second leading cause of cancer death. It is well known that breast density is an important risk factor for breast cancer and also can be used to personalize screening and for assessment of treatment response. Breast density has previously been correlated to volumetric water density. The purpose of this study is to validate the accuracy and precision of dual-energy mammography in measuring water density in postmortem breasts.

Methods

Twenty pairs of postmortem breasts were imaged using dual-energy mammography with energy-sensitive photon-counting detectors. Chemical analysis was used as the reference standard to assess the accuracy of dual-energy mammography in measuring volumetric water and lipid density. Images from different views and contralateral breasts were used to assess estimate of precision for water and lipid volumetric density measurements.

Results

The measured volumetric water and lipid density from dual-energy mammography and chemical analysis were in good agreement, where the standard errors of estimates (SEE) of both were calculated to be 2.1%. Volumetric water and lipid density measurements from different views were also in good agreement, with a SEE of 1.3% and 1.1%, respectively.

Conclusions

The results indicate that dual-energy mammography can be used to accurately measure volumetric water and lipid density in breast tissue. Accurate quantification of volumetric water density is expected to enhance its utility as a risk factor for breast cancer and for assessment of response to therapy.

Key Points

• Dual-energy mammography can be used to accurately measure water and lipid volumetric density in breast tissue.
• Improved quantification of volumetric water density is expected to enhance its utility for assessment of response to therapy and as a risk factor for breast cancer.
Literatur
1.
Zurück zum Zitat (2006) World Heatlh Organization Fact Sheet N297: cancer (2006) World Heatlh Organization Fact Sheet N297: cancer
2.
Zurück zum Zitat McCormack VA, dos Santos SI (2006) Breast density and parenchymal patterns as markers of breast cancer risk: a meta-analysis. Cancer Epidemiol Biomarkers Prev 15:1159–1169CrossRef McCormack VA, dos Santos SI (2006) Breast density and parenchymal patterns as markers of breast cancer risk: a meta-analysis. Cancer Epidemiol Biomarkers Prev 15:1159–1169CrossRef
3.
Zurück zum Zitat Boyd NF, Martin LJ, Bronskill M, Yaffe MJ, Duric N, Minkin S (2010) Breast tissue composition and susceptibility to breast cancer. J Natl Cancer Inst 102:1224–1237CrossRef Boyd NF, Martin LJ, Bronskill M, Yaffe MJ, Duric N, Minkin S (2010) Breast tissue composition and susceptibility to breast cancer. J Natl Cancer Inst 102:1224–1237CrossRef
4.
Zurück zum Zitat Boyd NF, Guo H, Martin LJ et al (2007) Mammographic density and the risk and detection of breast cancer. N Engl J Med 356:227–236CrossRef Boyd NF, Guo H, Martin LJ et al (2007) Mammographic density and the risk and detection of breast cancer. N Engl J Med 356:227–236CrossRef
5.
Zurück zum Zitat Sprague BL, Conant EF, Onega T et al (2016) Variation in mammographic breast density assessments among radiologists in clinical practice: a multicenter observational study. Ann Intern Med 165:457–464CrossRef Sprague BL, Conant EF, Onega T et al (2016) Variation in mammographic breast density assessments among radiologists in clinical practice: a multicenter observational study. Ann Intern Med 165:457–464CrossRef
6.
Zurück zum Zitat Ekpo EU, Ujong UP, Mello-Thoms C, McEntee MF (2016) Assessment of interradiologist agreement regarding mammographic breast density classification using the fifth edition of the BI-RADS Atlas. AJR Am J Roentgenol 206:1119–1123CrossRef Ekpo EU, Ujong UP, Mello-Thoms C, McEntee MF (2016) Assessment of interradiologist agreement regarding mammographic breast density classification using the fifth edition of the BI-RADS Atlas. AJR Am J Roentgenol 206:1119–1123CrossRef
7.
Zurück zum Zitat Spayne MC, Gard CC, Skelly J, Miglioretti DL, Vacek PM, Geller BM (2012) Reproducibility of BI-RADS breast density measures among community radiologists: a prospective cohort study. Breast J 18:326–333CrossRef Spayne MC, Gard CC, Skelly J, Miglioretti DL, Vacek PM, Geller BM (2012) Reproducibility of BI-RADS breast density measures among community radiologists: a prospective cohort study. Breast J 18:326–333CrossRef
8.
Zurück zum Zitat Byng JW, Boyd NF, Fishell E, Jong RA, Yaffe MJ (1994) The quantitative-analysis of mammographic densities. Phys Med Biol 39:1629–1638CrossRef Byng JW, Boyd NF, Fishell E, Jong RA, Yaffe MJ (1994) The quantitative-analysis of mammographic densities. Phys Med Biol 39:1629–1638CrossRef
9.
Zurück zum Zitat Sivaramakrishna R, Obuchowski NA, Chilcote WA, Powell KA (2001) Automatic segmentation of mammographic density. Acad Radiol 8:250–256CrossRef Sivaramakrishna R, Obuchowski NA, Chilcote WA, Powell KA (2001) Automatic segmentation of mammographic density. Acad Radiol 8:250–256CrossRef
10.
Zurück zum Zitat Carton AK, Li JJ, Chen S, Conant E, Maidment ADA (2006) Optimization of contrast-enhanced digital breast tomosynthesis. In: Astley SM, Brady M, Rose C, Zwiggelaar R, (eds) Digital mammography, proceedings. (Lecture Notes in Computer Science), pp 183–189 Carton AK, Li JJ, Chen S, Conant E, Maidment ADA (2006) Optimization of contrast-enhanced digital breast tomosynthesis. In: Astley SM, Brady M, Rose C, Zwiggelaar R, (eds) Digital mammography, proceedings. (Lecture Notes in Computer Science), pp 183–189
11.
Zurück zum Zitat Kaufhold J, Thomas JA, Eberhard JW, Galbo CE, Trotter DE (2002) A calibration approach to glandular tissue composition estimation in digital mammography. Med Phys 29:1867–1880CrossRef Kaufhold J, Thomas JA, Eberhard JW, Galbo CE, Trotter DE (2002) A calibration approach to glandular tissue composition estimation in digital mammography. Med Phys 29:1867–1880CrossRef
12.
Zurück zum Zitat Pawluczyk O, Augustine BJ, Yaffe MJ et al (2003) A volumetric method for estimation of breast density on digitized screen-film mammograms. Med Phys 30:352–364CrossRef Pawluczyk O, Augustine BJ, Yaffe MJ et al (2003) A volumetric method for estimation of breast density on digitized screen-film mammograms. Med Phys 30:352–364CrossRef
13.
Zurück zum Zitat Highnam R, Jeffreys M, McCormack V, Warren R, Smith GD, Brady M (2007) Comparing measurements of breast density. Phys Med Biol 52:5881–5895CrossRef Highnam R, Jeffreys M, McCormack V, Warren R, Smith GD, Brady M (2007) Comparing measurements of breast density. Phys Med Biol 52:5881–5895CrossRef
14.
Zurück zum Zitat Cardinal HN, Fenster A (1990) An accurate method for direct dual-energy calibration and decomposition. Med Phys 17:327–341CrossRef Cardinal HN, Fenster A (1990) An accurate method for direct dual-energy calibration and decomposition. Med Phys 17:327–341CrossRef
15.
Zurück zum Zitat Molloi S, Ducote JL, Ding H, Feig SA (2014) Postmortem validation of breast density using dual-energy mammography. Med Phys 41:081917CrossRef Molloi S, Ducote JL, Ding H, Feig SA (2014) Postmortem validation of breast density using dual-energy mammography. Med Phys 41:081917CrossRef
16.
Zurück zum Zitat Ducote JL, Molloi S (2010) Quantification of breast density with dual energy mammography: an experimental feasibility study. Med Phys 37:793–801CrossRef Ducote JL, Molloi S (2010) Quantification of breast density with dual energy mammography: an experimental feasibility study. Med Phys 37:793–801CrossRef
17.
Zurück zum Zitat Ducote JL, Klopfer MJ, Molloi S (2011) Volumetric lean percentage measurement using dual energy mammography. Med Phys 38:4498–4504CrossRef Ducote JL, Klopfer MJ, Molloi S (2011) Volumetric lean percentage measurement using dual energy mammography. Med Phys 38:4498–4504CrossRef
18.
Zurück zum Zitat Ding HJ, Ducote JL, Molloi S (2012) Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system. Med Phys 39:1289–1297CrossRef Ding HJ, Ducote JL, Molloi S (2012) Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system. Med Phys 39:1289–1297CrossRef
19.
Zurück zum Zitat Johnson T, Ding H, Le HQ, Ducote JL, Molloi S (2013) Breast density quantification with cone-beam CT: a post-mortem study. Phys Med Biol 58:8573–8591CrossRef Johnson T, Ding H, Le HQ, Ducote JL, Molloi S (2013) Breast density quantification with cone-beam CT: a post-mortem study. Phys Med Biol 58:8573–8591CrossRef
20.
Zurück zum Zitat Ding H, Johnson T, Lin M et al (2013) Breast density quantification using magnetic resonance imaging (MRI) with bias field correction: a postmortem study. Med Phys 40:122305CrossRef Ding H, Johnson T, Lin M et al (2013) Breast density quantification using magnetic resonance imaging (MRI) with bias field correction: a postmortem study. Med Phys 40:122305CrossRef
21.
Zurück zum Zitat Cho HM, Ding H, Kumar N, Sennung D, Molloi S (2017) Calibration phantoms for accurate water and lipid density quantification using dual energy mammography. Phys Med Biol 62:4589–4603CrossRef Cho HM, Ding H, Kumar N, Sennung D, Molloi S (2017) Calibration phantoms for accurate water and lipid density quantification using dual energy mammography. Phys Med Biol 62:4589–4603CrossRef
22.
Zurück zum Zitat Aslund M, Cederstrom B, Lundqvist M, Danielsson M (2006) Scatter rejection in multislit digital mammography. Med Phys 33:933–940CrossRef Aslund M, Cederstrom B, Lundqvist M, Danielsson M (2006) Scatter rejection in multislit digital mammography. Med Phys 33:933–940CrossRef
23.
Zurück zum Zitat Fredenberg E, Hemmendorff M, Cederstrom B, Aslund M, Danielsson M (2010) Contrast-enhanced spectral mammography with a photon-counting detector. Med Phys 37:2017–2029CrossRef Fredenberg E, Hemmendorff M, Cederstrom B, Aslund M, Danielsson M (2010) Contrast-enhanced spectral mammography with a photon-counting detector. Med Phys 37:2017–2029CrossRef
24.
Zurück zum Zitat Fredenberg E, Lundqvist M, Cederstrom B, Aslund M, Danielsson M (2010) Energy resolution of a photon-counting silicon strip detector. Nucl Instrum Methods Phys Res Sect A 613:156–162CrossRef Fredenberg E, Lundqvist M, Cederstrom B, Aslund M, Danielsson M (2010) Energy resolution of a photon-counting silicon strip detector. Nucl Instrum Methods Phys Res Sect A 613:156–162CrossRef
25.
Zurück zum Zitat Johansson H, von Tiedemann M, Erhard K et al (2017) Breast-density measurement using photon-counting spectral mammography. Med Phys 44:3579–3593CrossRef Johansson H, von Tiedemann M, Erhard K et al (2017) Breast-density measurement using photon-counting spectral mammography. Med Phys 44:3579–3593CrossRef
26.
Zurück zum Zitat Molloi S, Ding H, Feig S (2015) Breast density evaluation using spectral mammography, radiologist reader assessment, and segmentation techniques: a retrospective study based on left and right breast comparison. Acad Radiol 22:1052–1059CrossRef Molloi S, Ding H, Feig S (2015) Breast density evaluation using spectral mammography, radiologist reader assessment, and segmentation techniques: a retrospective study based on left and right breast comparison. Acad Radiol 22:1052–1059CrossRef
27.
Zurück zum Zitat Chow CK, Venzon D, Jones EC, Premkumar A, O'Shaughnessy J, Zujewski J (2000) Effect of tamoxifen on mammographic density. Cancer Epidemiol Biomarkers Prev 9:917–921PubMed Chow CK, Venzon D, Jones EC, Premkumar A, O'Shaughnessy J, Zujewski J (2000) Effect of tamoxifen on mammographic density. Cancer Epidemiol Biomarkers Prev 9:917–921PubMed
Metadaten
Titel
Quantification of water and lipid density with dual-energy mammography: validation in postmortem breasts
verfasst von
Sabee Molloi
Huanjun Ding
Hyo-Min Cho
Daniel Beidokhti
Publikationsdatum
26.08.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2021
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-07179-9

Weitere Artikel der Ausgabe 2/2021

European Radiology 2/2021 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.