Skip to main content
Erschienen in: Abdominal Radiology 4/2020

12.02.2020 | Technical

Quantitative attenuation accuracy of virtual non-enhanced imaging compared to that of true non-enhanced imaging on dual-source dual-energy CT

verfasst von: Sanaz Javadi, Sherif Elsherif, Priya Bhosale, Corey T. Jensen, Rick R. Layman, Megan C. Jacobsen, Ott Le, Sun Jia, Rajan J. Parikh, Eric P. Tamm

Erschienen in: Abdominal Radiology | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the quantitative attenuation and reliability of virtual non-contrast (VNC) images of the abdomen acquired from multiphasic scans with a dual-energy computed tomography (DECT) system and compare it with that of true non-enhanced images (TNC) on second- (Flash) and third- (Force) generation DECT scanners.

Methods

This retrospective study was approved by the institutional review board and included 123 patients with pancreatic cancer who had undergone routine clinical multiphasic DECT examinations at our institution using Flash and Force scanners between March and August 2017. VNC images of the abdomen were reconstructed from late arterial phase images. For every patient, regions-of-interest were defined in the aorta, fluid-containing structures (gallbladder, pleural effusion, and renal cysts > 10 mm), paravertebral muscles, subcutaneous fat, spleen, pancreas, renal cortex, and liver (eight locations) on TNC and VNC images. The mean attenuation of VNC was compared with TNC by organ for each CT scanner using an equivalence test and the Bland–Altman plot. The mean attenuations for TNC or VNC were compared between the Force and Flash CT scanners using a two-sample t test.

Results

The VNC attenuation of organs on the Force scanner was lower than was that on the Flash, and the mean attenuation difference in different organs on the Force was closer to 0. The estimated means of TNC and VNC were equivalent for an equivalence margin of 10 on the Force scanner.

Conclusion

VNC images in DECT are a promising alternative to TNC images. In clinical scenarios in which non-enhanced CT images are required but are not available for accurate diagnosis, VNC images can potentially serve as an alternative to TNC images without the radiation exposure risks.
Literatur
2.
Zurück zum Zitat Alvarez RE, Macovski A (1976) Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21 (5):733–744CrossRef Alvarez RE, Macovski A (1976) Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21 (5):733–744CrossRef
3.
Zurück zum Zitat Mileto A, Marin D, Alfaro-Cordoba M, Ramirez-Giraldo JC, Eusemann CD, Scribano E, Blandino A, Mazziotti S, Ascenti G (2014) Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study. Radiology 273 (3):813–820. https://doi.org/10.1148/radiol.14140171 CrossRefPubMed Mileto A, Marin D, Alfaro-Cordoba M, Ramirez-Giraldo JC, Eusemann CD, Scribano E, Blandino A, Mazziotti S, Ascenti G (2014) Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study. Radiology 273 (3):813–820. https://​doi.​org/​10.​1148/​radiol.​14140171 CrossRefPubMed
8.
Zurück zum Zitat Frellesen C, Azadegan M, Martin SS, Otani K, D’Angelo T, Booz C, Eichler K, Panahi B, Kaup M, Bauer RW, Vogl TJ, Wichmann JL (2018) Dual-Energy Computed Tomography-Based Display of Bone Marrow Edema in Incidental Vertebral Compression Fractures: Diagnostic Accuracy and Characterization in Oncological Patients Undergoing Routine Staging Computed Tomography. Invest Radiol 53 (7):409–416. https://doi.org/10.1097/rli.0000000000000458 CrossRefPubMed Frellesen C, Azadegan M, Martin SS, Otani K, D’Angelo T, Booz C, Eichler K, Panahi B, Kaup M, Bauer RW, Vogl TJ, Wichmann JL (2018) Dual-Energy Computed Tomography-Based Display of Bone Marrow Edema in Incidental Vertebral Compression Fractures: Diagnostic Accuracy and Characterization in Oncological Patients Undergoing Routine Staging Computed Tomography. Invest Radiol 53 (7):409–416. https://​doi.​org/​10.​1097/​rli.​0000000000000458​ CrossRefPubMed
12.
13.
Zurück zum Zitat De Cecco CN, Darnell A, Macías N, Ayuso JR, Rodríguez S, Rimola J, Pagés M, García-Criado Á, Rengo M, Laghi A, Ayuso C (2013) Virtual Unenhanced Images of the Abdomen With Second-Generation Dual-Source Dual-Energy Computed Tomography: Image Quality and Liver Lesion Detection. Invest Radiol 48 (1):1–9. https://doi.org/10.1097/rli.0b013e31826e7902 CrossRefPubMed De Cecco CN, Darnell A, Macías N, Ayuso JR, Rodríguez S, Rimola J, Pagés M, García-Criado Á, Rengo M, Laghi A, Ayuso C (2013) Virtual Unenhanced Images of the Abdomen With Second-Generation Dual-Source Dual-Energy Computed Tomography: Image Quality and Liver Lesion Detection. Invest Radiol 48 (1):1–9. https://​doi.​org/​10.​1097/​rli.​0b013e31826e7902​ CrossRefPubMed
17.
Zurück zum Zitat Ananthakrishnan L, Rajiah P, Ahn R, Rassouli N, Xi Y, Soesbe TC, Lewis MA, Lenkinski RE, Leyendecker JR, Abbara S (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol (NY) 42 (3):702–709. https://doi.org/10.1007/s00261-016-1036-9 CrossRef Ananthakrishnan L, Rajiah P, Ahn R, Rassouli N, Xi Y, Soesbe TC, Lewis MA, Lenkinski RE, Leyendecker JR, Abbara S (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol (NY) 42 (3):702–709. https://​doi.​org/​10.​1007/​s00261-016-1036-9 CrossRef
Metadaten
Titel
Quantitative attenuation accuracy of virtual non-enhanced imaging compared to that of true non-enhanced imaging on dual-source dual-energy CT
verfasst von
Sanaz Javadi
Sherif Elsherif
Priya Bhosale
Corey T. Jensen
Rick R. Layman
Megan C. Jacobsen
Ott Le
Sun Jia
Rajan J. Parikh
Eric P. Tamm
Publikationsdatum
12.02.2020
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 4/2020
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-020-02415-8

Weitere Artikel der Ausgabe 4/2020

Abdominal Radiology 4/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.