Erschienen in: International Ophthalmology 5/2017

21.10.2016 | Original Paper

Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens

verfasst von: Manuel Garza-Leon, Hugo A. Fuentes-de la Fuente, Ana V. García-Treviño

Erschienen in: International Ophthalmology | Ausgabe 5/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of the study was to assess the reproducibility of ocular biometry using the IOLMaster-700 in a healthy population.

Methods

This is a prospective, cross-sectional, observational reproducibility study. Ocular biometry was performed three times on each of 45 studied eyes. Flattest meridian (Kf) and the steepest meridian (Ks), central corneal thickness, axial length, anterior chamber depth, aqueous depth, lens thickness, and white-to-white distances were recorded. Reproducibility was evaluated using the coefficient of variation (CV), the within subject standard deviation, and the intraclass correlation coefficient (ICC).

Results

There was a high reproducibility in all parameters; CV was between 0.3 and 1 %, and the ICC was higher than 0.87 in all measurements.

Conclusions

IOLMaster-700 showed high reproducibility for ocular biometry.
Literatur
3.
Zurück zum Zitat Kaswin G, Rousseau A, Mgarrech M, Barreau E, Labetoulle M (2014) Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard. J Cataract Refract Surg 40(4):593–600. doi:10.1016/j.jcrs.2013.09.015 CrossRefPubMed Kaswin G, Rousseau A, Mgarrech M, Barreau E, Labetoulle M (2014) Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard. J Cataract Refract Surg 40(4):593–600. doi:10.​1016/​j.​jcrs.​2013.​09.​015 CrossRefPubMed
4.
Zurück zum Zitat Srivannaboon S, Chirapapaisan C, Chonpimai P, Koodkaew S (2014) Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer. J Cataract Refract Surg 40(5):709–715CrossRefPubMed Srivannaboon S, Chirapapaisan C, Chonpimai P, Koodkaew S (2014) Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer. J Cataract Refract Surg 40(5):709–715CrossRefPubMed
5.
Zurück zum Zitat Shen P, Zheng Y, Ding X, Liu B, Congdon N, Morgan I, He M (2013) Biometric measurements in highly myopic eyes. J Cataract Refract Surg 39(2):180–187CrossRefPubMed Shen P, Zheng Y, Ding X, Liu B, Congdon N, Morgan I, He M (2013) Biometric measurements in highly myopic eyes. J Cataract Refract Surg 39(2):180–187CrossRefPubMed
6.
Zurück zum Zitat Rončević MB, Bušić M, Čima I, Elabjer BK, Bosnar D, Miletić D (2011) Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation. Graefe’s Arch Clin Exp Ophthalmol 249(1):69–75CrossRef Rončević MB, Bušić M, Čima I, Elabjer BK, Bosnar D, Miletić D (2011) Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation. Graefe’s Arch Clin Exp Ophthalmol 249(1):69–75CrossRef
7.
Zurück zum Zitat Nakhli FR (2014) Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye. Saudi J Ophthalmol 28(4):287–291CrossRefPubMedPubMedCentral Nakhli FR (2014) Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye. Saudi J Ophthalmol 28(4):287–291CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Moyal L, Abrieu-Lacaille M, Bonnel S, Sendon D, de Rivoyre B, Berguiga M, Rambaud C, Froussart-Maille F, Rigal-Sastourne JC (2015) Comparison of two different surgical treatments of presbyopia for hyperopic patients over 55 years old: Presbylasik (Supracor) and Prelex (presbyopic lens exchange). J Fr Ophtalmol 38(4):306–315. doi:10.1016/j.jfo.2014.10.010 CrossRefPubMed Moyal L, Abrieu-Lacaille M, Bonnel S, Sendon D, de Rivoyre B, Berguiga M, Rambaud C, Froussart-Maille F, Rigal-Sastourne JC (2015) Comparison of two different surgical treatments of presbyopia for hyperopic patients over 55 years old: Presbylasik (Supracor) and Prelex (presbyopic lens exchange). J Fr Ophtalmol 38(4):306–315. doi:10.​1016/​j.​jfo.​2014.​10.​010 CrossRefPubMed
11.
Zurück zum Zitat Brown RH, Zhong L, Lynch MG (2014) Clear lens extraction as treatment for uncontrolled primary angle-closure glaucoma. J Cataract Refract Surg 5(40):840–841CrossRef Brown RH, Zhong L, Lynch MG (2014) Clear lens extraction as treatment for uncontrolled primary angle-closure glaucoma. J Cataract Refract Surg 5(40):840–841CrossRef
12.
Zurück zum Zitat Man X, Chan NC, Baig N, Kwong YY, Leung DY, Li FC, Tham CC (2015) Anatomical effects of clear lens extraction by phacoemulsification versus trabeculectomy on anterior chamber drainage angle in primary angle-closure glaucoma (PACG) patients. Graefe’s Arch Clin Exp Ophthalmol 253(5):773–778CrossRef Man X, Chan NC, Baig N, Kwong YY, Leung DY, Li FC, Tham CC (2015) Anatomical effects of clear lens extraction by phacoemulsification versus trabeculectomy on anterior chamber drainage angle in primary angle-closure glaucoma (PACG) patients. Graefe’s Arch Clin Exp Ophthalmol 253(5):773–778CrossRef
16.
Zurück zum Zitat Oshika T, Sugita G, Miyata K, Tokunaga T, Samejima T, Okamoto C, Ishii Y (2007) Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration. Br J Ophthalmol 91(2):185–188. doi:10.1136/bjo.2006.099945 CrossRefPubMed Oshika T, Sugita G, Miyata K, Tokunaga T, Samejima T, Okamoto C, Ishii Y (2007) Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration. Br J Ophthalmol 91(2):185–188. doi:10.​1136/​bjo.​2006.​099945 CrossRefPubMed
17.
Zurück zum Zitat Korynta J, Bok J, Cendelin J, Michalova K (1999) Computer modeling of visual impairment caused by intraocular lens misalignment. J Cataract Refract Surg 25(1):100–105CrossRefPubMed Korynta J, Bok J, Cendelin J, Michalova K (1999) Computer modeling of visual impairment caused by intraocular lens misalignment. J Cataract Refract Surg 25(1):100–105CrossRefPubMed
19.
Zurück zum Zitat Al-Refaie A, Bata N (2010) Evaluating measurement and process capabilities by GR&R with four quality measures. Measurement 43(6):842–851CrossRef Al-Refaie A, Bata N (2010) Evaluating measurement and process capabilities by GR&R with four quality measures. Measurement 43(6):842–851CrossRef
20.
Zurück zum Zitat Senol S (2004) Measurement system analysis using designed experiments with minimum α–β Risks and n. Measurement 36(2):131–141CrossRef Senol S (2004) Measurement system analysis using designed experiments with minimum α–β Risks and n. Measurement 36(2):131–141CrossRef
22.
Zurück zum Zitat Grulkowski I, Liu JJ, Zhang JY, Potsaid B, Jayaraman V, Cable AE, Duker JS, Fujimoto JG (2013) Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers. Ophthalmology 120(11):2184–2190. doi:10.1016/j.ophtha.2013.04.007 CrossRefPubMed Grulkowski I, Liu JJ, Zhang JY, Potsaid B, Jayaraman V, Cable AE, Duker JS, Fujimoto JG (2013) Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers. Ophthalmology 120(11):2184–2190. doi:10.​1016/​j.​ophtha.​2013.​04.​007 CrossRefPubMed
23.
Zurück zum Zitat Lopez de la Fuente C, Sanchez-Cano A, Segura F, Pinilla I (2014) Comparison of anterior segment measurements obtained by three different devices in healthy eyes. BioMed Res Int 2014 Lopez de la Fuente C, Sanchez-Cano A, Segura F, Pinilla I (2014) Comparison of anterior segment measurements obtained by three different devices in healthy eyes. BioMed Res Int 2014
24.
Zurück zum Zitat Gonvers M, Bornet C, Othenin-Girard P (2003) Implantable contact lens for moderate to high myopia: relationship of vaulting to cataract formation. J Cataract Refract Surg 29(5):918–924CrossRefPubMed Gonvers M, Bornet C, Othenin-Girard P (2003) Implantable contact lens for moderate to high myopia: relationship of vaulting to cataract formation. J Cataract Refract Surg 29(5):918–924CrossRefPubMed
25.
Zurück zum Zitat Zaldivar R, Davidorf JM, Oscherow S (1998) Posterior chamber phakic intraocular lens for myopia of -8 to -19 diopters. J Refract Surg 14(3):294–305PubMed Zaldivar R, Davidorf JM, Oscherow S (1998) Posterior chamber phakic intraocular lens for myopia of -8 to -19 diopters. J Refract Surg 14(3):294–305PubMed
26.
27.
Zurück zum Zitat Martin R, Ortiz S, Rio-Cristobal A (2013) White-to-white corneal diameter differences in moderately and highly myopic eyes: partial coherence interferometry versus scanning-slit topography. J Cataract Refract Surg 39(4):585–589. doi:10.1016/j.jcrs.2012.11.021 CrossRefPubMed Martin R, Ortiz S, Rio-Cristobal A (2013) White-to-white corneal diameter differences in moderately and highly myopic eyes: partial coherence interferometry versus scanning-slit topography. J Cataract Refract Surg 39(4):585–589. doi:10.​1016/​j.​jcrs.​2012.​11.​021 CrossRefPubMed
28.
Zurück zum Zitat De la Parra-Colín P, Garza-León M, Barrientos-Gutierrez T (2014) Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers. Int Ophthalmol 34(1):27–33CrossRefPubMed De la Parra-Colín P, Garza-León M, Barrientos-Gutierrez T (2014) Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers. Int Ophthalmol 34(1):27–33CrossRefPubMed
29.
Zurück zum Zitat Bao F, Wang Q, Cheng S, Savini G, Lu W, Feng Y, Yu Y, Huang J (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33(6):576–581CrossRefPubMed Bao F, Wang Q, Cheng S, Savini G, Lu W, Feng Y, Yu Y, Huang J (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33(6):576–581CrossRefPubMed
30.
Zurück zum Zitat Huang J, Liao N, Savini G, Li Y, Bao F, Yu Y, Yu A, Wang Q (2015) Measurement of central corneal thickness with optical low-coherence reflectometry and ultrasound pachymetry in normal and post-femtosecond laser in situ keratomileusis eyes. Cornea 34(2):204–208. doi:10.1097/ICO.0000000000000329 CrossRefPubMed Huang J, Liao N, Savini G, Li Y, Bao F, Yu Y, Yu A, Wang Q (2015) Measurement of central corneal thickness with optical low-coherence reflectometry and ultrasound pachymetry in normal and post-femtosecond laser in situ keratomileusis eyes. Cornea 34(2):204–208. doi:10.​1097/​ICO.​0000000000000329​ CrossRefPubMed
Metadaten
Titel
Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens
verfasst von
Manuel Garza-Leon
Hugo A. Fuentes-de la Fuente
Ana V. García-Treviño
Publikationsdatum
21.10.2016
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 5/2017
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-016-0380-7

Weitere Artikel der Ausgabe 5/2017

International Ophthalmology 5/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.