Reproducibility of length measurements of the lower limb by using EOS™

verfasst von: A. Clavé, D. G. Maurer, N. S. Nagra, F. Fazilleau, C. Lefèvre, E. Stindel

Erschienen in: MUSCULOSKELETAL SURGERY | Ausgabe 2/2018

Einloggen, um Zugang zu erhalten

Abstract

Background

Lower-limb discrepancy following total hip arthroplasty is the third-most common reason for patient dissatisfaction in orthopaedic surgery. Therefore, accurate planning and evaluation methods are mandatory. The main aim of this study was to evaluate the reliability of the EOS™ system by establishing and comparing the reproducibility of lower-limb automatic and manual 3D measurements. We hypothesized that the reproducibility of the lower-limb measurements is similar regardless of the method used and with an agreement higher than 0.95 for the length parameters.

Materials and methods

This study utilized an EOS radiological database of 112 patients. Two independent observers performed two rounds of lower-limb measurements twice, either in manual 3D or automatic 3D mode. The intra- and inter-observer reproducibility was evaluated by the calculation of the intra-class coefficient for each measurement method. The methods were then compared.

Results

The intra- and inter-observer reproducibility for length measurements found with the manual and automatic 3D methods was always > 0.98. There was no significant difference in the reproducibility between the two measurement modes, with the exception of the offset, hip–knee–shaft, and neck–shaft angles.

Conclusion

Our results indicate a very good reproducibility of EOS™ length measurement, regardless of the method used. Automated 3D mode is preferred for the collection of angular and offset measurements. Furthermore, manual mode measurements are not affected by surgical history.
Level of evidence IV.
Literatur
1.
Zurück zum Zitat Marmor S (2011) Causes de procédures médicolégales après prothèse totale de hanche. RCOT 97:752–757 Marmor S (2011) Causes de procédures médicolégales après prothèse totale de hanche. RCOT 97:752–757
2.
Zurück zum Zitat Nogler M, Kessler O, Prassl A, Donnelly B, Streicher R, Sledge JB et al (2004) Reduced variability of acetabular cup positioning with use of an imageless navigation system. Clin Orthop Relat Res 426:159–163CrossRef Nogler M, Kessler O, Prassl A, Donnelly B, Streicher R, Sledge JB et al (2004) Reduced variability of acetabular cup positioning with use of an imageless navigation system. Clin Orthop Relat Res 426:159–163CrossRef
3.
Zurück zum Zitat Sugano N, Takao M, Sakai T, Nishii T, Miki H (2012) Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA? Clin Orthop Relat Res 470:3054–3059CrossRefPubMedPubMedCentral Sugano N, Takao M, Sakai T, Nishii T, Miki H (2012) Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA? Clin Orthop Relat Res 470:3054–3059CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Gandhi R, Marchie A, Farrokhyar F, Mahomed N (2009) Computer navigation in total hip replacement: a meta-analysis. Int Orthop 33:593–597CrossRefPubMed Gandhi R, Marchie A, Farrokhyar F, Mahomed N (2009) Computer navigation in total hip replacement: a meta-analysis. Int Orthop 33:593–597CrossRefPubMed
5.
Zurück zum Zitat Escott BG, Ravi B, Weathermon AC, Acharya J, Gordon CL, Babyn PS et al (2013) EOS low-dose radiography: a reliable and accurate upright assessment of lower-limb lengths. J Bone Joint Surg (Am) 95:e1831–e1837CrossRef Escott BG, Ravi B, Weathermon AC, Acharya J, Gordon CL, Babyn PS et al (2013) EOS low-dose radiography: a reliable and accurate upright assessment of lower-limb lengths. J Bone Joint Surg (Am) 95:e1831–e1837CrossRef
6.
Zurück zum Zitat Terry MA, Winell JJ, Green DW, Schneider R, Peterson M, Marx RG et al (2005) Measurement variance in limb length discrepancy: clinical and radiographic assessment of interobserver and intraobserver variability. J Pediatr Orthop 25:197–201CrossRefPubMed Terry MA, Winell JJ, Green DW, Schneider R, Peterson M, Marx RG et al (2005) Measurement variance in limb length discrepancy: clinical and radiographic assessment of interobserver and intraobserver variability. J Pediatr Orthop 25:197–201CrossRefPubMed
7.
Zurück zum Zitat Guenoun B, Zadegan F, Aim F, Hannouche D, Nizard R (2012) Reliability of a new method for lower-extremity measurements based on stereoradiographic three-dimensional reconstruction. Orthop Traumatol Surg Res 98(5):506–513CrossRefPubMed Guenoun B, Zadegan F, Aim F, Hannouche D, Nizard R (2012) Reliability of a new method for lower-extremity measurements based on stereoradiographic three-dimensional reconstruction. Orthop Traumatol Surg Res 98(5):506–513CrossRefPubMed
8.
Zurück zum Zitat Pasquier G, Ducharne G, Ali ES, Giraud F, Mouttet A, Durante E (2010) Total hip arthroplasty offset measurement: is C T scan the most accurate option? Orthop Traumatol Surg Res 96:367–375CrossRefPubMed Pasquier G, Ducharne G, Ali ES, Giraud F, Mouttet A, Durante E (2010) Total hip arthroplasty offset measurement: is C T scan the most accurate option? Orthop Traumatol Surg Res 96:367–375CrossRefPubMed
9.
Zurück zum Zitat O’Connor KJ, Grady JF, Hollander M (1988) CT scanography for limb length determination. Clin Podiatr Med Surg 5:267–274PubMed O’Connor KJ, Grady JF, Hollander M (1988) CT scanography for limb length determination. Clin Podiatr Med Surg 5:267–274PubMed
10.
Zurück zum Zitat Charpak G (1996) Prospects for the use in medicine of new detectors of ionizing radiation. Bull Acad Natl Med 180:161–168PubMed Charpak G (1996) Prospects for the use in medicine of new detectors of ionizing radiation. Bull Acad Natl Med 180:161–168PubMed
12.
Zurück zum Zitat Dubousset J, Charpak G, Skalli W, Kalifa G, Lazennec JY (2007) EOS stereo-radiography system: whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose. Rev Chir Orthop Repar Appar Mot 93:141–143CrossRef Dubousset J, Charpak G, Skalli W, Kalifa G, Lazennec JY (2007) EOS stereo-radiography system: whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose. Rev Chir Orthop Repar Appar Mot 93:141–143CrossRef
14.
Zurück zum Zitat Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRefPubMed Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRefPubMed
15.
Zurück zum Zitat Fisher RA (1925) Statistical methods for research workers. Statistical methods for research workers. Oliver & Boyd, Edinbourgh, PA Fisher RA (1925) Statistical methods for research workers. Statistical methods for research workers. Oliver & Boyd, Edinbourgh, PA
16.
Zurück zum Zitat Heaver C, St Mart J-P, Nightingale P, Sinha A, Davis ET (2013) Measuring limb length discrepancy using pelvic radiographs: the most reproducible method. Hip Int 23:391–394CrossRefPubMed Heaver C, St Mart J-P, Nightingale P, Sinha A, Davis ET (2013) Measuring limb length discrepancy using pelvic radiographs: the most reproducible method. Hip Int 23:391–394CrossRefPubMed
17.
Zurück zum Zitat Delin C, Silvera S, Bassinet C, Thelen P, Rehel J-L, Legmann P et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed Delin C, Silvera S, Bassinet C, Thelen P, Rehel J-L, Legmann P et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed
18.
19.
Zurück zum Zitat Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A et al (2011) Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproductibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res 97:373–380CrossRefPubMed Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A et al (2011) Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproductibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res 97:373–380CrossRefPubMed
20.
Zurück zum Zitat Sariali E, Lazennec JY, Khiami F, Gorin M, Catonne Y (2009) Modification of pelvic orientation after total hip replacement in primary osteoarthritis. Hip Int 19:257–263CrossRefPubMed Sariali E, Lazennec JY, Khiami F, Gorin M, Catonne Y (2009) Modification of pelvic orientation after total hip replacement in primary osteoarthritis. Hip Int 19:257–263CrossRefPubMed
21.
Zurück zum Zitat Blondel B, Parratte S, Tropiano P, Pauly V, Aubaniac J-M, Argenson J-N (2009) Pelvic tilt measurement before and after total hip arthroplasty. Orthop Traumatol Surg Res 95:568–572CrossRefPubMed Blondel B, Parratte S, Tropiano P, Pauly V, Aubaniac J-M, Argenson J-N (2009) Pelvic tilt measurement before and after total hip arthroplasty. Orthop Traumatol Surg Res 95:568–572CrossRefPubMed
22.
Zurück zum Zitat Barbier O, Ollat D, Versier G (2012) Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty. Orthop Traumatol Surg Res 98:398–404CrossRefPubMed Barbier O, Ollat D, Versier G (2012) Interest of an intraoperative limb-length and offset measurement device in total hip arthroplasty. Orthop Traumatol Surg Res 98:398–404CrossRefPubMed
23.
Zurück zum Zitat Barbier O, Skalli W, Mainard L, Mainard D, Computer Assisted Orthopedic Surgery–France (CAOS-France) (2014) The reliability of the anterior pelvic plane for computer navigated acetabular component placement during total hip arthroplasty: prospective study with the EOS imaging system. Orthop Traumatol Surg Res 00:S287–S291CrossRef Barbier O, Skalli W, Mainard L, Mainard D, Computer Assisted Orthopedic Surgery–France (CAOS-France) (2014) The reliability of the anterior pelvic plane for computer navigated acetabular component placement during total hip arthroplasty: prospective study with the EOS imaging system. Orthop Traumatol Surg Res 00:S287–S291CrossRef
Metadaten
Titel
Reproducibility of length measurements of the lower limb by using EOS™
verfasst von
A. Clavé
D. G. Maurer
N. S. Nagra
F. Fazilleau
C. Lefèvre
E. Stindel
Publikationsdatum
01.11.2017
Verlag
Springer Milan
Erschienen in
MUSCULOSKELETAL SURGERY / Ausgabe 2/2018
Print ISSN: 2035-5106
Elektronische ISSN: 2035-5114
DOI
https://doi.org/10.1007/s12306-017-0518-4

Weitere Artikel der Ausgabe 2/2018

MUSCULOSKELETAL SURGERY 2/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.